MassPREP On-Line Desalting Cartridge

Similar documents
Method Development Considerations for Reversed-Phase Protein Separations

Protein-Pak Hi Res HIC Column and HIC Protein Standard

Application of Agilent AdvanceBio Desalting-RP Cartridges for LC/MS Analysis of mabs A One- and Two-dimensional LC/MS Study

Mass Spectrometry Analysis of Liquid Chromatography Fractions using Ettan LC MS System

Improved SPE for UPLC/MS Determination of Diquat and Paraquat in Environmental

Illicit Drug Analysis in Urine Using 2D LC-MS/MS for Forensic Toxicology

EASY-Spray Columns. Guidance for column set up and installation Tips to maximize column lifetime

Guide to Trap Cartridge Care and Use

Maximizing Chromatographic Resolution of Peptide Maps using UPLC with Tandem Columns

Developing Quantitative UPLC Assays with UV

Rapid UHPLC Analysis of Reduced Monoclonal Antibodies using an Agilent ZORBAX Rapid Resolution High Definition (RRHD) 300SB-C8 Column

Biotherapeutic Method Development Guide

2012 Waters Corporation 1

Fast Online Desalting of mabs Using a Reversed-Phase Desalting Cartridge for LC-MS Analysis

Biotherapeutic Method Development Guide

Solutions Guide. MX Series II Modular Automation for Nano and Analytical Scale HPLC And Low Pressure Fluid Switching Applications

Method Development of a 2D LC-HRMS Extraction and Detection Method for Organophosphorus Flame Retardants in Environmental Water Samples

Clone Selection Using the Agilent 1290 Infinity Online 2D-LC/MS Solution

LC TRAPPING GUIDE & PRODUCTS

Hichrom Limited 1 The Markham Centre, Station Road, Theale, Reading, Berkshire, RG7 4PE, UK Tel: +44 (0) Fax: +44 (0)

Eoin F.J. Cosgrave, Matthew A. Lauber, Robert Birdsall, and Sean M. McCarthy Waters Corporation, Milford, MA, USA APPLICATION BENEFITS INTRODUCTION

Analysis of Monoclonal Antibody (mab) Using Agilent 1290 Infinity LC System Coupled to Agilent 6530 Accurate-Mass Quadrupole Time-of-Flight (Q-TOF)

( ) IONPAC ATC-HC ANION TRAP COLUMN

R R Innovation Way P/N SECKIT-7830 Newark, DE 19711, USA Tel: Fax: Website: Published in November 2013

Exploring Extra Sensitivity Using ionkey/ms with the Xevo G2-XS Q-Tof HRMS for Small Molecule Pharmaceutical Analysis in Human Plasma

Agilent Multiple Affinity Removal Spin Cartridges for the Depletion of High-Abundant Proteins from Human Proteomic Samples.

Sean M. McCarthy and Martin Gilar Waters Corporation, Milford, MA, U.S. INTRODUCTION EXPERIMENTAL RESULTS AND DISCUSSION

Ultrahydrogel Columns

for Acclaim Carbamate Column

Technical Overview. Author. Abstract. Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany

UPLC/MS/MS Method for the Quantification of Trastuzumab in Human Serum at the 5-nM Level Using Xevo TQD MS and ACQUITY UPLC H-Class System

ION EXCHANGE KIT FOR MAB SEPARATIONS

Improving Retention Time Precision and Chromatography of Early Eluting Peptides with Acetonitrile/Water Blends as Solvent B

Characterize Fab and Fc Fragments by Cation-Exchange Chromatography

Generating Automated and Efficient LC/MS Peptide Mapping Results with the Biopharmaceutical Platform Solution with UNIFI

Column for High Performance,High-Binding Capacity Ion Exchange Chromatography:TSKgel SuperQ-5PW and Its Applications

Techniques for Improving the Efficiency of Large Volume Loading in Preparative Liquid Chromatography

Application Note. Author. Abstract. Biotherapeutics and Biologics. Sonja Schneider Agilent Technologies, Inc. Waldbronn, Germany

Analysis of biomolecules by SEC and Ion-Exchange UPLC

ITSP Extraction of Cortisol from Plasma

Improving Sensitivity for an Immunocapture LC-MS Assay of Infliximab in Rat Plasma Using Trap-and-Elute MicroLC-MS

Automated On-Line SDS Removal from Diminutive Protein Samples Prior to Capillary HPLC Application Note 221

Thermo Scientific EASY-Spray LC Columns Optimization - Red flags to watch for and how to troubleshoot your workflow Estee Toole

ACQUITY UPLC. I-Class PLUS WHAT SEPARATES YOU FROM EVERYONE ELSE

Simplifying Methods Transfer: Novel Tools for Replicating Your Established Methods on an ACQUITY Arc System

BioTrap 500 C18/C8 INSTRUCTION MANUAL. for direct injection of up to 500 µl plasma in HPLC. Illustration of a BioTrap 500 particle

Fast mass transfer Fast separations High throughput and improved productivity Long column lifetime Outstanding reproducibility Low carryover

Development of a Clinical Research Method for the Measurement of Testosterone. Dihydrotestosterone

Proteomics Training Kit

MAbPac RP Column. High-performance reverse phase chromatography column for monoclonal antibody analysis

ProSwift and PepSwift Monolith Columns for Biomolecule Analysis

columns PepSwift and ProSwift Capillary Monolithic Reversed-Phase Columns

Development and evaluation of Nano-ESI coupled to a triple quadrupole mass spectrometer for quantitative proteomics research

Tabisam Khan 1, David R. Little 1, Liyu Yang 2, Jiwen Chen 2, Parya Nouri 2, Samy Tadros 2, Patrick J. Rudewicz 2

Introduction Ron Majors is a Senior Scientist at Agilent. Bill Champion is a chemist in Agilent s HPLC Columns tech support group.

Chromatography column for therapeutic protein analysis

A Modular Preparative HPLC System for the Isolation of Puerarin from Kudzu Root Extracts

Developing Purification Strategies for the Agilent 1260 Infinity II Preparative LC/MSD System

Gel filtration using ÄKTA start

DETERMINATION OF PROTEIN BINDING BY UPLC/MS/MS

The Impact of Spare Parts. Performance

Multi-Residue Pesticide Analysis in Ginseng Powder: Optimized Cleanup After QuEChERS Extraction for UPLC-MS/MS And GC-MS/MS Analysis

Martin Gilar Waters Corporation, Milford, MA, U.S. Origins of synthetic oligonucleotides impurities. Lab-scale isolation options

Ultra-Sensitive Quantification of Trastuzumab Emtansine in Mouse Plasma using Trap-Elute MicroLC MS Method

Automatic Combination of Orthogonal Separation Modes for Multiple Components in Mixtures

E f f ic i ent ly Com pa ring Bat c h e s o f a n Intac t Mono c lo na l A n t ibo dy using t h e

Analysis of Intact Monoclonal Antibodies Using an M3 MicroLC with the TripleTOF 6600

Minimizing Sample Carryover Using the Multiwash Function of the Agilent 1290 Infinity II Multisampler

Trastuzumab Glycan Batch-to-Batch Profiling using a UPLC/FLR/Mass Spectrometry Platform

[ Care and Use Manual ] Waters AccellPlus QMA and CM Bulk Media. I. calculating bulk media needs. II. packed column use. IIi.

Developing Robust and Efficient IEX Methods for Charge Variant Analysis of Biotherapeutics Using ACQUITY UPLC H-Class System and Auto Blend Plus

Strategies for Phospholipid Removal using Polymer-based SPE

EFFECTIVE SPE STRATEGIES FOR LC/MS DETERMINATION OF SUDAN DYES IN CHILI PRODUCTS

Chem 321 Lecture 23 - Liquid Chromatography 11/19/13

Agilent 1290 Infinity II 2D-LC Solution Biopharmaceutical Polymer Analysis. WCBP Jan 2017 Washington, DC

Agilent MetaCarb Ca PLUS Carbohydrate Column Ca ++ Form User Manual Part No. A5205

H-Class & H-Class Bio AutoBlend and AutoBlend plus. Jean-Michel Plankeele May Waters Corporation 1

Characterize mab Charge Variants by Cation-Exchange Chromatography

On-Line. User s Guide SPE CARTRIDGES. for Rapid Cleanup and Extraction of Analytes

Benefits of 2D-LC/MS/MS in Pharmaceutical Bioanalytics

Extraction of Vitamin B7 (Biotin) from Serum Using EVOLUTE

PLRP-S Polymeric Reversed-Phase Column for LC/MS Separation of mabs and ADC

Superose 6 HR 10/30 Superose 12 HR 10/30

Universal Solution for Monoclonal Antibody Quantification in Biological Fluids Using Trap-Elute MicroLC-MS Method

Bio-Scale Mini Affi-Prep Protein A Cartridges, 1 and 5 ml. Instruction Manual. Catalog # ,

The Role of Chemistry Consumables in Biopharmaceutical Product Development

BioHPLC columns. Tim Rice Biocolumn Technical Specialist

This User Bulletin does not apply to PROCISE clc instruments.

Cell Clone Selection Using the Agilent Bio-Monolith Protein A Column and LC/MS

Reducing Cycle Time for Charge Variant Analysis of Monoclonal Antibodies

Converting a Liquid-Liquid Extraction Method for Vitamin D to a 96-Well Plate Supported Liquid Extraction Format

Automated Scouting of Stationary and Mobile Phases Using the Agilent 1290 Infinity II Method Development Solution

ENVIRONMENTAL ANALYSIS. Solutions for Your Analytical Business Markets and Applications Programs. Solution Note

Parallel LC with Capillary PS-DVB Monolithic Columns for High-Throughput Proteomics

Improving Sensitivity in Bioanalysis using Trap-and-Elute MicroLC-MS

BIOANALYTICAL STRATEGY FOR IN VITRO METABOLITE SCREENING WITH EXACT MASS USING THE Q-Tof micro. Jose M. Castro-Perez 1, Carina Leandersson 2

Xevo G2-S QTof and TransOmics: A Multi-Omics System for the Differential LC/MS Analysis of Proteins, Metabolites, and Lipids

ACQUITY UPLC Protein BEH SEC Columns and Standards

Robustness of the Agilent Ultivo Triple Quadrupole LC/MS for Routine Analysis in Food Safety

Application Note. Author. Abstract. Environmental. Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany

Transcription:

CONTENTS I. INTRODUCTION II. INSTALLING THE MASSPREP ON-LINE DESALTING CARTRIDGE INTO THE SENTRY.1 X 1 MM GUARD COLUMN HOLDER III. RECOMMENDED LC/MS SYSTEM CONFIGURATION TO MINIMIZE MS SOURCE CONTAMINATION IV. SUGGESTED METHOD FOR MASS ANALYSIS USING MASSPREP ON-LINE DESALTING CARTRIDGE a. Instrumentation b. Separation Method V. GUIDELINES FOR USE AND TROUBLESHOOTING VI. MISCELLANEOUS I. INTRODUCTION Mass spectrometry is a powerful tool for intact protein analysis. However, the quality of analyses can be compromised by salts in the sample that interfere with ionization and/or form adducts with the investigated proteins. It is therefore important to remove or significantly minimize the introduction of extraneous salt ions into the mass analyzer. The use of on-line, desalting devices can efficiently perform this function with minimal sample handling. Waters MassPREP On-line Desalting Cartridges (.1 x 1 mm, P/N 186785) contain a polymer-based, reversed-phase packing material that can be used to effectively desalt protein samples prior to mass spectrometry. (Note: The purchase and use of Waters Sentry.1 x 1 mm Guard Column Holder Kit (P/N WAT97958) is required for this application as shown in Figure 1.) The phenyl phase material contained in these on-line devices successfully traps proteins allowing the salts to be washed to waste prior to elution of the desalted protein into the mass spectrometer. With an optimized LC/MS method, cycle times as low as 3 minutes are achievable. These on-line devices, when used with an appropriate System injector and injector wash solution, can be used for the sequential analyses of different samples (See Figures and 3). This Care and Use manual provides suggested methods for the successfully use of this product. For the latest list of MassPREP, BioSuite or NanoEase column offerings, go to www.waters.com/ lifesciences. II. INSTALLING THE MASSPREP ON-LINE DESALTING CARTRIDGE INTO THE SENTRY.1 X 1 MM GUARD COLUMN HOLDER 1. Unscrew the two halves of the Sentry.1 x 1 mm Guard Column Holder.. Remove the old MassPREP On-line Desalting Cartridge, if present.

3. Install the new MassPREP On-line Desalting Cartridge, noting the flow arrow direction on the device. This should match the flow arrow direction on the analytical column. 4. Reassemble the Sentry.1 x 1 mm Guard Column Holder. Hand-tighten only. 5. Connect the.1 x 1 mm Guard Column Holder containing the MassPREP On-line Desalting Cartridge onto or LC/MS System using appropriate connectors (e.g., PEEK tubing and connectors). Prior to use, the MassPREP On-line Desalting Cartridge reversed-phase packing material must be wet with organic solvent prior to equilibration with 95 Eluent A/5 Eluent B. This can be accomplished by running a blank injection (i.e., buffer injection followed by running the separations gradient) or by washing the column with a minimum of five column volumes of 5 Eluent B followed by a ten column volume equilibration at 95 Eluent A/5 Eluent B prior to sample injection. While equilibrating, check the holder and all connections for leaks. III. RECOMMENDED LC/MS SYSTEM CONFIGURATION TO MINIMIZE MS SOURCE CONTAMINATION The presence of non-volatile salts (e.g., NaCl) can suppress protein ionization and foul the ionization source of the mass spectometer. It is therefore important to remove or significantly minimize the introduction of these compounds into the mass analyzer. Figure details an LC/MS configuration that can effectively be used to load sample onto the column and elute salt to waste prior to valve switching and protein elution into the mass analyzer. Position 1: Sample Loading and Salt Elution to Waste Gradient MassPREP On-Line Desalting Cartridge 1 6 Waste Mass Spectrometer 5 3 4 Pump Recirculation Figure 1: Sentry Guard Holder (P/N WAT97958) and MassPREP On-line Desalting Cartridge (P/N 186785). Position : Sample Elution to Mass Spec Gradient MassPREP On-Line Desalting Cartridge Waste 1 6 Mass Spectrometer 5 3 4 Pump Recirculation Figure : LC/MS Valve Switching Configuration for use with MassPREP On-line Desalting Cartridge.

IV. SUGGESTED METHOD FOR MASS ANALYSIS USING MASSPREP ON-LINE DESALTING CARTRIDGE b. Separation method: Eluent A: H with.1 Formic Acid (Considerations to Minimize Protein Carryover Between Injections): injector design, composition of injector wash solution, and number of injector wash cycles between injections are some of the factors that can effectively minimize protein carryover between sample analyses. Waters MassPREP On-line Desalting Cartridges, when used with an appropriately performing System and injector wash protocol, can be used for the sequential analyses of different samples due to the lack of protein carryover from prior injections. Figure 3 shows the mass analyses results obtained from buffer blank and intact protein injections [bovine serum albumin (BSA) and IgG monoclonal antibody (mab)] processed through the MassPREP On-line Desalting Cartridge using the Waters instrumentation and separation method detailed below: a. Instrumentation System: Alliance 796 Separations Module Needle Wash Solution Composition: H (35) / Isopropanol (5) Acetonitrile (6) Number of Needle Washes: Single Needle Wash Time: Mass Spectrometer: 15 sec needle exterior/3 sec inject port Q-tof Ionization mode: ES + Capillary voltage: 33 Eluent B: Acetonitrile with.1 Formic Acid Flow:.4 ml/minute Initial Column Conditioning: Prior to use, the MassPREP On-line Desalting Cartridge reversed phase packing material must be wet with organic solvent prior to equilibration with 95 Eluent A/5 Eluent B. This can be accomplished by running a blank injection (i.e., buffer injection followed by running the separations gradient) or by washing the cartridge with a minimum of five column volumes of 5 Eluent B followed by a ten cartridge volume equilibration at 95 Eluent A/5 Eluent B prior to sample injection. Sample Loading/Salt Elution: Load sample onto cartridge while delivering 95 Eluent A/5 Eluent B. The protein sample will bind to the cartridge while the contaminating salt passes through unretained to waste. Sample Elution to Mass Spec: At approximatly..5 min, begin increase Eluent B concentration from 5 to 8 in 1.5 min. Hold at 8 B for an additional 1. min to allow protein to completely elute from cartridge into MS system. Column re-equilibration: At approximately 3. min, re-equilibration cartridge with ten cartridge volumes of 95 Eluent A/5 Eluent B prior to next sample injection. Cone voltages: 3 for BSA mass analysis 35 for mab mass analysis Source temp: 15 C Desolvation temp: 35 C Gas flow: Cone gas: 5 L/Hr 5 L/Hr 3

1 1 Injection 1_blank Injection 99_MAb 1 Injection 1 Injection.1..3.4.5.6.7.8.9 1. 1.1 1. 1.3 1.4 1.5 1.6 1.7 1.8 1.9. Time 1 Injection 1 blank 1 Injection 99 MAb 1 Injection 98 BSA 1 Injection 97 blank 1 1 14 16 18 4 6 8 3 3 34 36 38 m/z Figure 3: Results from final 4 injections on MassPREP On-line Desalting Cartridge from a 1 injection overnight sequence noting absence of protein carryover between injections. Test series consisted of repetitive injections of PBS, IgG monoclonal antibody (mab) and BSA. 4

V. GUIDELINES FOR USE AND TROUBLESHOOTING 1. The amount of intact protein to load onto a MassPREP On-line Desalting Cartridge to obtain acceptable mass analysis results depends upon a number of variables. These include the type of mass analyzer used, the ionization efficiency of analyzed protein and MS source, the potential ionization suppression effect of ion-pairing reagents used in the separation eluent (e.g., TFA), etc. In general, between 1 and 5 µg of injected protein such as BSA or an IgG monoclonal antibody should yield good mass detector signal and excellent protein recovery from the MassPREP On-line Desalting Cartridge. Proportionately less material should be injected when analyzing smaller molecular weight proteins to prevent overloading the On-line Desalting Cartridge and or mass detector.. The maximum recommended operating pressure for, when configured WITHOUT pre-column tubing, is approximately 15 psi. Use of pressure generating tubing immediately before the is required to generate sufficient UPLC System back pressure for proper functioning. Thus, the total reported UPLC System backpressure generated when BOTH the pre-column tubing and are configured for an application is approximately 57 psi at a flow of.5 ml/min with a 5 ACN/.1 FA eluent. 3. Results obtained with Waters MassPREP On-line Desalting Cartridges can vary depending upon the physical characteristics of the analyzed proteins as well as on the performance characteristics of the (or LC/MS System), injector, and injector wash protocol used. VI. MISCELLANEOUS Recommended Storage: For overnight storage, continuously flush the cartridge with the mobile phase at slow flow (e.g.,.1. ml/min). If the cartridge is not to be used for several days, store the cartridge in 5 acetonitrile (CH 3 CN) without any eluent additives such as formic acid or TFA. Do not store the MassPREP On-line Desalting Cartridges in buffered eluents. Completely seal the stored column to avoid evaporation and drying of the bed. Waters Corporation, Waters. The Science of What s Possible, and Alliance are registered trademarks of Waters Corporation. MassPREP, Sentry, BioSuite, NanoEase, and Q-Tof micro are trademarks of Waters Corporation. All other trademarks are property of their respective owners. 14 Waters Corporation. Produced in the U.S.A. June 14 71516EN Rev C. KP-PDF Waters Corporation 34 Maple Street Milford, MA 1757 U.S.A. T: 1 58 478 F: 1 58 87 199 www.waters.com