IgG Purity/Heterogeneity and SDS-MW Assays with High- Speed Separation Method and High Throughput Tray Setup

Similar documents
Simplifying CE-SDS Data Processing

IgG Purity and Heterogeneity Assay Kit For the PA 800 Plus Pharmaceutical Analysis System. Application Guide

Capillary Electrophoresis of Proteins

Application Note. Biopharm. Author. Abstract. Christian Wenz Agilent Technologies, Inc. Waldbronn, Germany

Unified Workflow for Monoclonal Antibody Charge Heterogeneity, Purity, and Molecular Weight Analysis

Analysis of Monoclonal Antibody Charge Variants by Capillary Zone Electrophoresis

Impurity Detection with a New Light Emitting Diode Induced Fluorescence Detector Coupled to the Agilent 7100 Capillary Electrophoresis System

Assay of IgG Purity and Heterogeneity using High-Resolution Sodium Dodecyl Sulfate Capillary Gel Electrophoresis

Optimization of cief to Resolve Challenging MAb Samples of Clinical Immunodiagnostic Relevance

Repeatability of C100ht Biologics Analyzer, for High Throughput Glycan Screening

Separation of Recombinant Human Erythropoietin (rhepo) using the European Pharmacopoeia Method on the PA 800 plus Pharmaceutical Analysis System

deltadot ANTIBODY ANALYSIS Application Note: Antibodies using LABEL FREE INTRINSIC IMAGING LONDON United Kingdom

High-Throughput Glycan Screening to Remove Biopharmaceutical Development Bottlenecks

LabChip GXII: Antibody Analysis

SCIEX Anion Analysis Kit For P/ACE MDQ and P/ACE MDQ plus Capillary Electrophoresis Systems. Instruction Guide

Rediscover the Strengths of Your PA 800 Plus - A Multi-Attribute Protein Analyzer

The Importance of Temperature Control in Capillary Electrophoresis of Carbohydrates

Automated and quantitative analysis of biologics

Analysis of Intact Monoclonal Antibodies Using an M3 MicroLC with the TripleTOF 6600

More Samples. Better Decisions Faster.

Automated and Quantitative Analysis of Biologics

HT Pico Protein Express LabChip Kit, Frequently Asked Questions

Application Note. Author. Abstract. Biopharmaceuticals. Verified for Agilent 1260 Infinity II LC Bio-inert System. Sonja Schneider

Comparison of Different Brands of Carrier Ampholytes for Monoclonal Antibody Charge Heterogeneity Analysis by Capillary Isoelectric Focusing

Knock! Knock! Who s there? Maurice Who? SungAe Suhr Park, Mee Ko, Eleanor Le, Janice Chen Amgen Inc.

Answers, Simple and Streamlined. Solutions for Biotherapeutic Intact Mass Analysis

CE Oligonucleotide Analysis Kit Instruction Manual

Streamlined and Complete LC-MS Workflow for MAM

E-Coli Cell Lysate By High Performance Capillary Electrophoresis

LIFluor EnhanCE. for use with dsdna 1000 and dsdna 20,000 Kits

Capillary Electrophoresis in Quality Control PART II: CE-SDS: Method Development and Robustness

Separate and Quantify Rituximab Aggregates and Fragments with High-Resolution SEC

Application Note # ET-20 BioPharma Compass: A fully Automated Solution for Characterization and QC of Intact and Digested Proteins

PA 800 plus Catalog. Systems/Software

TECHNICAL INFORMATION

Universal Solution for Monoclonal Antibody Quantification in Biological Fluids Using Trap-Elute MicroLC-MS Method

Fully Automated C100HT Biologics Analyzer Sample Preparation on a Biomek i5 Multichannel Workstation

Improving the Detection Limits for Highly Basic Neuropeptides Using CESI-MS

Improving Sensitivity for an Immunocapture LC-MS Assay of Infliximab in Rat Plasma Using Trap-and-Elute MicroLC-MS

BioBA High Capacity Enrichment Sample Preparation Kit:

Improving Sensitivity in Bioanalysis using Trap-and-Elute MicroLC-MS

Automating the C100ht Biologics Analyzer Sample Preparation on a Biomek i5 Liquid Handler Platform

Developing Quantitative UPLC Assays with UV

Ultra-Sensitive Quantification of Trastuzumab Emtansine in Mouse Plasma using Trap-Elute MicroLC MS Method

High-resolution Analysis of Charge Heterogeneity in Monoclonal Antibodies Using ph-gradient Cation Exchange Chromatography

Isotopic Resolution of Chromatographically Separated IdeS Subunits Using the X500B QTOF System

ACCURACY THE FIRST TIME. Bring Your Esoteric Assays In-House Routinely SCIEX DIAGNOSTICS MASS SPECTROMETERS

Clone Selection Using the Agilent 1290 Infinity Online 2D-LC/MS Solution

Biotherapeutic Non-Reduced Peptide Mapping

"VUPNBUFE BOE RVBOUJUBUJWF

quantitate host cell DNA with high sensitivity and throughput resdnaseq Host Cell DNA Quantitation Systems: CHO, E.

Journal of Chemical and Pharmaceutical Research, 2017, 9(3): Research Article

One-stop cief and CE-SDS. One-day method dev. Results in a snap. FDA, PDQ.

Contaminant bovine IgG assay

Microfluidic Assays for High Throughput Screening of Protein Quality in Bioprocess Development

OptiFlow Quant Solution for High-Sensitivity Quantitation

Application Guide. PA 800 plus Pharmaceutical Analysis System. SDS-MW Analysis. A51970AA April 2009

Biotherapeutic Peptide Mapping Information Dependent Acquisition (IDA) Method

High-throughput and Sensitive Size Exclusion Chromatography (SEC) of Biologics Using Agilent AdvanceBio SEC Columns

Applications. Capillary Electrophoresis. Cell Lysate Analysis. Analysis of E. Coli Lysate. Technology

Mass Spectrometry Analysis of Liquid Chromatography Fractions using Ettan LC MS System

Separation of Native Monoclonal Antibodies and Identification of Charge Variants:

Highly Resolved Separation of DNA Fragments on the Fragment Analyzer

electrophoresis tech Performance Comparison of the Experion Automated Electrophoresis System and SDS-PAGE for Protein Analysis

TOYOPEARL GigaCap Series

Highly Resolved Separation of DNA Fragments on the Fragment Analyzer

A Robust cief Method: Intermediate Precision for the ph 5-7 Range

Development and Validation of the 3500 Series Genetic Analyzer for Human Identification

Separation of SDS-Protein Complexes Using the CE-SDS Protein Kit ezid

Column for High Performance,High-Binding Capacity Ion Exchange Chromatography:TSKgel SuperQ-5PW and Its Applications

Automated and quantitative

Reducing Cycle Time for Charge Variant Analysis of Monoclonal Antibodies

Sean M. McCarthy and Martin Gilar Waters Corporation, Milford, MA, U.S. INTRODUCTION EXPERIMENTAL RESULTS AND DISCUSSION

Encompassing Method Development, Small Scale Purification, and Post-Purification Analysis with the Investigator SFC System

Technical tips for ELISA and multiplex

ssdna 100-R Kit Care and Use Instructions AG June 2015

PRODUCT DATA SHEET. Carboxylated Fluorescent Gold Nanoparticles. Description. Characteristics

Size Exclusion BioHPLC Columns

Disulfide Linkage Analysis of IgG1 using an Agilent 1260 Infinity Bio inert LC System with an Agilent ZORBAX RRHD Diphenyl sub 2 µm Column

Contaminant bovine transferrin assay

Application Note. Abstract. Author. Biopharmaceutical

Automated Scouting of Stationary and Mobile Phases Using the Agilent 1290 Infinity II Method Development Solution

Contaminant human transferrin assay

Mobile Phase Optimization in SEC Method Development

MultiQuant Software Version 3.0


10 Reasons to Switch from SDS-PAGE to the Agilent 2100 Bioanalyzer

This is not unique to CE. It took for instance about 25 years for HPLC to establish itself properly in the pharmaceutical industry.

A Comprehensive Workflow to Optimize and Execute Protein Aggregate Studies

Application Note. Biopharma. Authors. Abstract. James Martosella, Phu Duong Agilent Technologies, Inc Centreville Rd Wilmington, DE 19808

Proteomics PA 800 (Beckman) ProteomeLab System (Detect Instrument Picture to Jump to that section of SOP)

Human IgE ELISA Antibody Pair Kit

LabChip 90 System with DataViewer Software

CSE Separation of Cancer Cell Lysate

Pharmacokinetic Analysis of the mab Adalimumab by ELISA and Hybrid LBA/LC/MS: A Comparison Study

The Simple Western Approach to Vaccine and Clinical Protein Research

Standard Sensitivity NGS Fragment Analysis Kit (1 bp 6000 bp) User Guide (DNF ) (DNF )

High Sensitivity NGS Fragment Analysis Kit (1 bp 6000 bp) User Guide (DNF ) (DNF )

Rhinophase -AB, Zirconia-based Monoclonal Antibody Purification System

SDS-CGE Method Development for Purified Protein Vaccine Antigens

Transcription:

IgG Purity/Heterogeneity and SDS-MW Assays with High- Speed Separation Method and High Throughput Tray Setup High Throughput Methods to Maximize the Use of the PA 800 Plus system Jose-Luis Gallegos-Perez SCIEX Separations, Framingham, MA Introduction The PA 800 Plus Pharmaceutical System provides a comprehensive, automated, and quantitative solution for the characterization and analysis of therapeutic proteins. The application menu for the PA 800 Plus includes, among others, the IgG Purity/Heterogeneity assay and the SDS-gel molecular weight (SDS-MW) analysis.,2 The first method allows for the resolution of reduced and non-reduced immunoglobulins by size and to subsequently quantify the heterogeneity and impurities that may be present in IgG preparations. The SDS-MW analysis is designed to provide an estimation of the molecular weight of a protein in a sample. There are two types of analysis methods that can be used with these assays: ) The high resolution (HR) method with an inlet to detection window capillary length of 20.0 cm, using the capillary cartridge in a left-to-right configuration. 2) The high-speed (HS) method that uses the capillary cartridge in the right-to-left configuration with a sample introduction inlet to detection window capillary length of 0.0 cm. Both methods are used with a tray configuration that allows running up to 2 samples before having to change reagents. However, other tray setups are possible to optimize and maximize the number of sample runs. This report illustrates the use of the HS method and proposes a High Throughput Tray (HTT) setup to maximize the number of samples per run. We demonstrate that HS and HR analysis methods can be used with this optimized tray array. Advantages of the HS Analysis Method with the High Throughput Tray Setup The HS method provides faster separation time (-20 min vs. 0 min of the HR method); it represents 72 samples/day instead of 8/day. PA 800 Plus system The use of an HTT setup allows for analysis of up to 8 samples (2 using the conventional tray array) before changing chemical reagents. It represents savings in consumables and time. Using the HS method and high throughput tray setup saves time and reagents and helps maximize the use of the PA 800 Plus. Experimental Procedure Reagents: Separations using the IgG Purity Heterogeneity Assay Kit and SDS-MW Kit were performed as described in their respective manuals.,2 Iodoacetamide (IAA) and 2- mercaptoethanol were purchased from Sigma-Aldrich and used without any further purification. Ultrapure water (type ) was obtained from a Milli-Q Direct Ultrapure Water System. Bovine Serum Albumin (BSA) and Myoglobin (Myo) were purchased from Sigma-Aldrich. Capillary: A bare fused-silica capillary was used for separation of monoclonal antibody samples; 0 m i.d. x 0 m o.d. x 0.2 cm total length. Inlet-to-detection window equal to 20 cm for the HR method and 0 cm for the HS method. p

Sample Preparation: IgG control standard: A vial containing 9 L of the IgG control standard was combined with 2 L of the 0 kda internal standard and L of 2-mercaptoethanol in a fume hood. The vial was mixed thoroughly and centrifuged at 00g for min. The vial was sealed with Parafilm and heated to 70 C for 0 minutes, after which the vial was cooled for minutes in a room-temperature water bath. Non-reduced IgG sample preparation: A vial containing 9 L of the IgG control was combined with 2 L of the 0 kda internal standard and L of a 20 mm IAA solution. The vial was mixed thoroughly and centrifuged at 00g for minute. The vial was sealed with Parafilm and heated to 70 C for 0 minutes, after which the vial was cooled for minutes in a room-temperature water bath. HTT configuration: HR method: Figure 2 shows the position of the buffer vials in each inlet and outlet buffer tray using the HTT setup. In this configuration the buffer vials in each row are enough for 8 consecutive experiments, therefore it is possible to run up to 8 individual samples. (8 runs per row x rows). Method parameters are shown in Figure. SDS-MW Size Standard: 0 L of molecular weight size standard, 8 L of sample buffer, 2 L of internal standard, and L of 2-mercaptoethanol were combined in a vial. The mixture was heated to 00 C for minutes and placed into a room-temperature water bath to cool down for five minutes before injection. BSA and Myo were diluted with the SDS-MW sample buffer to a final concentration of 0. mg/ml each. For non-reduced protein preparation, a vial containing 9 L of sample solution was combined with 2 L of 0 kda internal standard and L of 20 mm IAA solution. The vial was mixed thoroughly and centrifuged at 00g for minute. The vial was sealed with Parafilm and heated to 70 C for min. For the reduced protein sample, 9 L of the sample was combined with 2 L of the 0 kda internal standard, and L of 2- mercaptoethanol. The mixture was heated to 00 C for min. HS method with conventional tray setups Because the HS method uses the capillary cartridge in the rightto-left configuration, the buffer tray on the right must be used as the inlet tray while the buffer tray on the left must be used as the outlet tray (see Figure ). The polarity for injection and separation should be normal. Using the conventional tray configuration, a maximum of 2 individual samples can be analyzed without reloading the trays with fresh solutions. Outlet Inlet Figure : Image of the inlet and outlet buffer vial tray set-up for the HS method in the PA800 Plus. Gel-R Gel-S NaOH HCl Gel-R Gel-S NaOH HCl Gel-R Gel-S NaOH HCl Gel-R Gel-S NaOH HCl 2 Gel-R Gel-S NaOH HCl Gel-R Gel-S NaOH HCl INLET BUFFER TRAY Method HR HS* Tray on the left Tray on the right 2 OUTLET BUFFER TRAY * Use only rows - (see text) Figure 2: Scheme of the inlet and outlet buffer vial tray setup for the HTT configuration using the HR or the HS method. Ensure the position of the buffer trays in the instrument according to the method being used. (HS method for no more than 0 individual samples, see text). Gel-S = Gel used for separation, Gel-R = Gel used for rinsing, both are of the same gel type. p 2

HS method: If less than 0 individual samples are being analyzed, buffer trays must be set up as shown in Figure 2 using rows from to only. IMPORTANT: In this case, the F-column in the SAMPLE TRAY must be kept empty, with no samples to avoid a tray collision during analysis. Follow the method parameters shown in Figure. HS method and HTT setup to maximize the number of sample analyses: To use the HTT configuration with the HS method for analysis of up to 8 samples requires switching reagents in positions C and D in the outlet and inlet buffer trays to avoid tray collision when samples are injected (see Figure ). All columns in the sample tray (from A to F) can be loaded with samples. For this specific case, use the method parameters show in Figure. 8 7 2 8 7 Sample Tray Sample Gel-R NaOH 2Gel-S HCl Sample Table indicates the volume of reagents used in each buffer vial and their number for all 8 runs using the HTT setup. Table. Volume in Buffer Vials. Reagents Vol (ml) Number of vials (8 runs) Water. 2 Gel. Water. 8 Waste (water) 0.8 2 NaOH. HCl. Instrument: Separations were performed using a PA 800 Plus Pharmaceutical Analysis System from SCIEX. PDA detector settings were configured according to the application guides. The instrument was controlled using 2 Karat software v0. The sample and capillary storage temperatures were set at ºC and 2ºC respectively. CE Instrument setups for the HR and HS methods with the HTT configuration are shown in Figures, and. Waste Waste HGel-S 2 O Waste Waste Waste Waste Gel-S Waste Waste Waste Waste Gel-S Waste Waste Waste Waste Gel-S Waste Waste 2 Waste Waste Gel-S Waste Waste Waste Waste Gel-S Waste Waste 2 Waste Gel-R NaOH Gel-S HCl Waste Waste Gel-S Waste Outlet Buffer Tray Inlet Buffer Tray Figure : Scheme of the inlet and outlet buffer vial tray setup for the HTT configuration using the HS method to inject up to 8 individual samples. p

Figure : High Resolution (HR) Method Parameters with HTT configuration. p

Figure : High-Speed (HS) Method with HTT configuration for up to 0 individual samples p

Figure : High-Speed Method with HTT configuration for up to 8 individual samples p

Results and Discussion HR and HS methods using the HTT setup To demonstrate assay repeatability using the HTT setup, the SDS-MW Size Standard was used. For the HR and HS methods, 8 sample replicates were injected using the conventional (C)-tray array and 8 more samples for the HTT configuration. Figure 7 shows profiles obtained in the experiments with a) HR method and b) HS method. It is evident that both tray setups produce the same analyte profile and that no ghost peaks, artifacts, or any other spurious signal is present in the HTT configuration. This result demonstrates that the HTT setup can be implemented with high-sensitivity or high-resolution methods, therefore the IgGpurity/heterogeneity or the SDS-MW applications could be run under these experimental conditions. Application of the HR and HS Methods with the IgG/Purity Heterogeneity Assay Using the HTT configuration, separation reproducibility of the HR and HS methods was evaluated. Eight replicates of the IgG control standard were injected under non-reducing and reducing conditions. Under non-reducing conditions (Figure 8) the peaks observed correspond to the 0 kda internal standard, the intact antibody (IgG) and the non-glycosylated heavy chain (NG). In addition, impurities that are known to be present in the IgG control standard, such as heavy-heavy (HH) and 2 heavy light chain (2HL) are resolved from the whole antibody using either method. Under reducing conditions (Figure 9), the peaks observed correspond to the 0 kda internal standard (IS), the IgG light chain (LC), IgG non-glycosylated heavy chain (NG) and the IgG heavy chain (HC). As in the previous case, all peaks can be easily identified using either method. Figure 7: Repeatability of HR and HS Methods using Conventional-tray and HTT setups. Stacked data illustrating 8 separations per tray setup using the a) HR method or b) HS method. To facilitate easier analysis, electropherograms were offset by 0.02 AU. Samples were alternately injected using the conventional (C) or the HTT setup. The order in which the samples were injected is indicated by the number. Figure 8: Comparison of HR and HS methods with the HTT setup for IgG Control Standard under Non-reducing Conditions. Peak identification: 0 kda is the Internal Standard, IgG is the intact antibody, NG the non-glycosylated heavy chain, HH and 2HL are the heavy-heavy (HH) and 2 heavy light chain impurities. p 7

Table 2. Accuracy and Repeatability of the HS and HR Methods under Non-Reducing Conditions Parameter Migration Time Area Percent Method HR HS HR HS Injection NG IgG NG IgG NG IgG NG IgG 2.2 27.0.2. 7.9 88. 7.7 88.27 2 2. 27...8 7.88 88.2 7.7 88.28 2.70 27.9.8.2 7.87 88. 7.87 88. 2.77 27.2.2. 7.89 88. 7.7 88.2 2.8 27.2..8 7.88 88.8 7.7 88.2 2.89 27.8.8.7 7.89 88. 7.7 88.7 7 2.9 27.0.0.7 7.79 88. 7. 88. 8 2.0 27.07.29. 7.88 88.9 7.77 88.2 Mean 2.7 27.2.0.29 7.87 88. 7.7.29 SD 0.2 0. 0.08 0.08 0.0 0.0 0.0 0.08 %CV 0. 0. 0.2 0. 0. 0.0 0.80 0. Table. Accuracy and Repeatability of the HS and HR Methods under Reducing Conditions Figure 9: Comparison of HR and HS methods with the HTT setup for IgG Control Standard under Reducing Conditions. Peak identification: 0 kda is the Internal Standard, LC is the Light Chain, NG the Non-glycosylated heavy chain and HC the Heavy Chain. Migration times and area percent were determined for the main identified peaks (NG, intact IgG, LC, and HC) (Tables 2 and ). These results indicate that the evaluated parameters show high reproducibility for 8 separations using either the HR or the HS method. Reproducibility for migration time was always less 0. %CV and lower than.7 %CV for the peak area percent. As expected, resolution was slightly lower in the HS mode due to the somewhat shorter capillary length; however adjacent peak pairs with close migration times, such as NG/IgG or NG/HC, can be resolved and unequivocally identified and quantified. This demonstrates that the HS method can be used with complete confidence under standardized and validated conditions. Molecular Weight Estimation of Myoglobin (Myo) and Bovine Serum Albumin (BSA) The SDS-MW analysis kit is designed for the separation and estimation of molecular weight (MW) of a protein in a range between 0 and 22 kda. The use of the HS method, for this experiment, was compared to the conventional HR method. The HTT setup was used for all experiments. The molecular weight of an unknown protein can be estimated by making a comparison against a standard curve created using proteins of known sizes. For this reason, a protein sizing standard mixture was injected in parallel to the Myo and BSA samples. The sizing standard reagent contained six proteins: 0, 20,, 0, 00 and 22 kda. Sample mixtures containing the internal standard (IS), Myo and BSA were treated under reducing or non-reducing conditions and three replicates of each condition were analyzed using the PA800 Plus instrument (Figure 0). In all cases, analyte peaks were well resolved using either the high speed or the high resolution method. Molecular weights for Myo and BSA were estimated using the qualitative analysis feature in the 2 Karat p 8

Control and Analysis Software. Results listed in Table indicate high reproducibility (%CV < 0.2, n = ) for the MW estimation of Myo or BSA, under non-reducing and reducing conditions. Similarly, the difference of the MW values determined by either method (HS or HR) was minimal with an absolute value difference lower than 0.2 kda. Therefore, both methods can be used under optimized experimental conditions producing comparable results. Table. Determination of Molecular Weight for Myoglobin and Bovine Albumin Serum (in kda) Non-reduced Reduced HR method HS method HR method HS-Method Injection Myo BSA Myo BSA Myo BSA Myo BSA.9.9.0.720.989 8.7.9 8. 2.99.9.0.20.989 8.7.87 8.829.99.9.9.299.989 8.8.808 9.02 Mean.9.9.99..989 8.7.87 8.72 SD 0.020 0.000 0.00 0.2 0.000 0.09 0.09 0. %CV 0. 0.000 0.2 0.2 0.000 0.00 0.0 0. Conclusions The IgG Purity/Heterogeneity and SDS-MW assays can be performed using a high speed (HS) separation method with an inlet to detection window of 0.0 cm, using the capillary cartridge in the right to left configuration. In comparison to the more highly resolving conventional High Resolution (HR) method, results are achieved faster saving assay time. The use of a high throughput tray (HTT) setup in order to maximize the number of samples per run sequence does not cause deterioration in the separation. The combination of the HS method with the HTT setup is an alternative when in need of higher throughput sample analysis. References Figure 0: Determination of Protein Molecular Weight using HR and HS methods. Peak identification: 0 kda is the Internal Standard, Myo is myoglobin and BSA is Bovine Serum Albumin. Three injections under reduced and non-reduced conditions were analyzed. A size standard mixture was run before to inject the samples with the HS or HR-methods.. PA 800 Plus Pharmaceutical Analysis System-IgG Purity/Heterogeneity Assay Application Guide. SCIEX part number A97AD, January 20. 2. PA 800 Plus Pharmaceutical Analysis System-SDS-MW- Analysis Assay Application Guide. SCIEX part number A970AD, January 20 20 AB SCIEX. For Research Use Only. Not for use in diagnostic procedures. AB SCIEX is doing business as SCIEX. The trademarks mentioned herein are the property of AB SCIEX Pte. Ltd. or their respective owners. AB SCIEX is being used under license. Document number: RUO-MKT-02-2-A p 9