Low-level Detection of Ethanol, 1,4-Dioxane, and Other Oxygenates Using the Eclipse Purge-and-Trap Sample Concentrator

Similar documents
A Presentation by OI Analytical. Oxygenate Analysis on the Eclipse Purge-and-Trap Sample Concentrator

Optimization of Volatile Petroleum Hydrocarbons Analysis by Automated Headspace Using Method Development Tools

Improved Performance and Productivity of Purge-and-Trap Analyses

Amy Nutter, Applications Chemist; Teledyne Tekmar P a g e 1

US EPA Method 8260 with the Atomx XYZ and the Thermo Scientific TRACE 1310 GC and ISQ MS

US EPA Method 8260 With the Tekmar Lumin and the Agilent 7890B GC/5977A MS

Using the Pulsed Flame Photometric Detector (PFPD) for Low-level Analysis of Organophosphorus Pesticides

USEPA Methods and 8021B using the Eclipse Purge-and-Trap (P&T) Sample Concentrator and PID/ELCD Tandem Detectors

USEPA Methods and 8021B using the Eclipse Purge-and-Trap (P&T) Sample Concentrator and PID/ELCD Tandem Detectors

US EPA Method Atomx XYZ and Agilent 7890B GC/5977A MS

Roger Bardsley, Applications Chemist; Teledyne Tekmar P a g e 1

Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1

Using Alternative Carrier Gas for US EPA Application Note. Abstract. Introduction. Experimental-Instrument Conditions

The Analysis of Oxygenates by EPA Method 8260

Trap Selection, Fast GC, and Troubleshooting Strategy for Purge & Trap GC Volatiles

Determination of Volatile Organic Compounds (VOCs) in Various Drinking Water Sources by GC/MS

Thermal Desorption Technical Support Note 87: A cryogen-free method for monitoring trace greenhouse gases in air. Key Words:

Improving GC-MS Method Robustness and Cycle Times Using Capillary Flow Technology and Backflushing

Olefin Analyzer Solutions for Polymer Grade Ethylene and Propylene

Capillary GC Troubleshooting: a Practical Approach

C146-E096 TD-20. Shimadzu Thermal Desorption System

A Rapid Method for Trace Analysis of Organophosphorus Pesticides in Drinking Water

EPA Method 524 for Determination of VOCs in Drinking Water Using Agilent 5975T LTM GC/MSD with Static Headspace

Analysis of Semivolatiles by GC/FID using the Ultra Inert Inlet Liners with Wool

A GC MS Purge-and-Trap Method Comparison Study for MTBE Analysis in Groundwater

New Developments in Thermal Desorption (TD) Tube & Canister Technologies for Collection & Analysis of Soil Gas

CryoTrap. For Agilent Technologies 7890/6890/6850 GC Systems. CryoTrap

Analysis of Suspected Flavor and Fragrance Allergens in Cosmetics Using the 7890A GC and Capillary Column Backflush Application

CHROMATOGRAPHY/MASS SPECTROMETRY

Determination of 1,4-Dioxane in Drinking Water by Gas Chromatography/Mass Spectrometry (GC/MS) with Selected Ion Monitoring (SIM)

Velocity XPT Sample Concentrator. Accelerated Purge and Trap Technology for Maximum Productivity In Your Laboratory

PAL SPME Fibers Optimized for Automation

PAL SPME Fibers Optimized for Automation

Chromatography Applications and Technology

Solutions for monitoring VOCs in air

Gas Chromatography SSI-GC-003

Rapid Microchamber Tests for Screening Chemical Emissions from Car Trim in Accordance with ISO

Determination of Volatile Organic Compounds (VOCs) present in the interiors of car by using GCMS/MS with static and dynamic headspace

Introducing the Eclipse: A New Generation of Purge-and-Trap Sample Concentrators

The Use Of Automated Backflush on the 7890A/5975A GC-MS System. Application. Setup Example Using Biomarkers. Authors. Abstract.

Drinking Water Semivolatiles Analysis using the 6890N/5975B inert GC/MSD. Application. Author. Abstract. Instrument Operating Parameters.

FOOD PESTICIDE RESIDUES LABORATORY REFERENCE NO INSPECTOR DATE

Confirmation of THC in Oral Fluids Using High-Resolution 2-D GC/MS Application Note

Fast GC Analyses of Volatiles

A New Horizon of Performance Agilent G1888 Network Headspace Sampler

The Importance of High Speed Data Acquisition in the Design and Implementation of GC/MS Applications

CTC automated sample injectors for gas chromatography, distributed by Agilent Technologies

Determination of volatile compounds in automotive interior materials by thermal desorption GC-MS

Analysis of 16 Pesticide Residues in Broccoli Using CarbonX dspe QuEChERS AOAC Kits Using GC/MS

Reduced carryover and increased reproducibility in the analysis of cold pressed pink grapefruit essential oil

Gas Chromatography Assignment Chem 543/443

Direct Injection of Fish Oil for the GC-ECD Analysis of PCBs: Results Using a Deans Switch With Backflushing. Application. Author.

GC Troubleshooting. Frank Michel, Katherine K. Stenerson T HCE

Ghost Peaks in Gas Chromatography Part 2: The Injection Port

Analysis of an Acrylic Copolymer using Pyrolysis-GC/MS

Agilent Consumables Improve Performance with Shimadzu GC/MS

Neue Möglichkeiten zur Überwachung der Umgebungsluft mit Thermodesorption und GC/MS

Markes International Thermal Desorption G8125A, G8126A, G8127A, G8128A, G8131A, G8132A, G8136A, G8133A, MKI-MicroC Site Preparation Checklist

TurboMatrix. Thermal Desorption Solutions

Volatile Organic Compounds in Solids PBM

Passive Monitoring: A Guide to Sorbent Tube Sampling for EPA Method 325. Nicola Watson July 2015

Application Note: Overview. Introduction

Pesticide Analysis in a Complex Matrix using a 2D GC/MS System

NWTPH-Gx. Volatile Petroleum Products Method for Soil and Water

Analysis of styrene leached from polystyrene cups using GCMS coupled with Headspace (HS) sampler

An Examination of the Presence, Formation, and Transformation of Volatile Halogenated Organic Species in Wastewater Extracts Using GC-ICP-MS

TDLAS. Application Note. Tunable Diode Laser Absorption Spectroscopy (TDLAS) for Carbon Dioxide Measurements In Natural Gas Processing

Historical Review: Fast & micro Gas Chromatography

Improving Productivity in Your Environmental and Food Testing Labs with Agilent 7890A/5975C GC/MS System

MS Certified Vials. Pre-cleaned and certified vials for mass spectrometry

ENVIRONMENTAL ANALYSIS. Solutions for Your Analytical Business Markets and Applications Programs. Solution Note

THERMAL SAMPLING TECHNIQUES FOR THE ANALYSIS OF POLYMERS BY GC-MS

Volatile Organic Compounds in Solids PBM

Fast On-Site Mine Safety Analysis by the Agilent 490 Micro GC

Environmental Air Monitoring

A Complete Solution for Aroma Characterization

Improved Gas Chromatograph Method for the Analysis of Trace Hydrocarbon Impurities in 1, 3-Butadiene

Volatile Organic Compounds in Soil PBM

Calibration Standards for Measurement of Toxic Organics in Biogas and Biomethane Using Permeation Tubes

Protocol for Quantitative Determination of Residual Solvents in Cannabis Concentrates Prepared by: Amanda Rigdon, May 23 rd, 2016

Methods for Increased Sample Throughput with Dirty Samples

The Agilent Metabolomics Dynamic MRM Database and Method

Agilent Ultimate Plus Fused Silica Tubing

MEASUREMENT OF VOC AND SVOC EMITTED FROM AUTOMOTIVE INTERIOR MATERIALS BY THERMAL DESORPTION TEST CHAMBER METHOD

Prefractionator for Reliable Analysis of the Light Ends of Crude Oil and other Petroleum Fractions

Agilent 7667A Mini Thermal Desorber MEASURE VOCS IN GAS MATRIX WHEN MOBILITY MATTERS

Available Laboratory Methods for Soil Gas Analysis

SEMI-VOLATILE ORGANICS IN DRINKING WATER using EPA Method optimized for greater sensitivity

The Analysis of 4-Vinyl-Cyclo-Hexene (VCH) in Butadiene with the DVLS LGI Injector

MARKET FOCUSED - environmental ORGANOCHLORINE PESTICIDES

Multiresidue Pesticide Analysis with the Agilent Intuvo 9000 GC and Agilent 7000 Series Mass Spectrometer

ANALYTICAL METHODOLOGY FOR GROUNDWATER AND SOIL ASSESSMENT GUIDELINES

Forensic Toxicology. Analysis of Ethanol in Blood using Master SHS Static Headspace Sampler and Master GC Gas Chromatograph APPLICATION NOTE.

Interlaboratory Validation Study for the Analysis of Nitrosamines in Drinking Water using GC-MS/MS

What Does the Future Hold for Ambient Air Monitoring Regulations?

Liners and consumables for OPTIC Inlet. Liner Selection Guide

Sample Matrix Influence on GC/MS/MS Multiresidue Pesticide Analysis

Quantitative determination of residual 2-(2-chloroethoxy) ethanol (CEE) in quetiapine fumarate by gas chromatogaraphy

HYDROGEN AS A CARRIER GAS. Gas Chromatography new frontiers: can Hydrogen substitute Helium?

NON-TARGETED ANALYSIS OF EMISSIONS FROM TOBACCO HEATING PRODUCTS

Transcription:

Application Note 25370206 Low-level Detection of,, and Other Oxygenates Using the Eclipse Purge-and-Trap Sample Concentrator Keywords Eclipse Fuel Oxygenates Methanol Model 4552 Purge and Trap Sample Concentrator Water Management Water/Soil Autosampler Presented at the 2006 Pittsburgh Conference on Analytical Chemistry and Applied Spectroscopy, Orlando, FL March 12 17, 2006 Introduction Fuel oxygenates such as methyl tert-butyl ether (MTBE) and tert-butyl alcohol (TBA) have been used as additives in reformulated gasoline to boost octane ratings and make the fuel burn more cleanly. Approximately 30% of the nation's gasoline contain some type of oxygenate additive. Concerns have been raised about the toxicity of MTBE and other additives as they enter the water table due to a spill or leak from underground storage tanks. As a result, several states have been requiring the use of ethanol as an alternative fuel additive. Replacing MTBE with ethanol has led to the need for a reliable analytical method that can be fully automated and provide low detection limits for ethanol in water. Using ethanol as an alternative fuel in the future will also require new test methods. Standard purge-and-trap (P&T) methods can be used to concentrate and analyze for ethanol and other fuel oxygenates. However, low ethanol concentrations are difficult to detect using conventional P&T settings due to ethanol's high solubility in water and low purge efficiency. Before now, the practical limit of detection for ethanol has been approximately 100 ppb (µg/l). This application note describes how the detection limits for ethanol and 1,4-dioxane can be lowered to 10 ppb by modifying standard P&T operating conditions. Variables such as trap adsorbent material, sample temperature, analytical trap temperature during purge, and water management fitting temperature during desorb were investigated for their effects on detection limits and peak shapes. The selective ion-monitoring (SIM) mode on a gas chromatograph/mass spectrometer (GC/MS) was also examined as a potential technique to further enhance sensitivity. Figure 1. OI Analytical Eclipse 4660 Purge-and-Trap Sample Concentrator Chromatograms of ethanol and 1,4-dioxane at 10 ppb are shown and all instrument operating parameters are provided. Experimental These experiments were conducted using the OI Analytical Model 4660 Eclipse Purge-and-Trap Sample Concentrator (Figure 1) and Model 4552 Water/Soil Autosampler. Analyses were performed using an Agilent 6890N GC and 5975 Inert MS with ChemStation to acquire and process

the data. Table 1 lists all instrument operating conditions before they were optimized for low-level detection of ethanol and 1,4-dioxane. Table 1. Typical operating conditions for analysis of volatile organic compounds (VOC) before conditions were optimized to enhance responses of ethanol and 1,4-dioxane. Asterisks mark the variables chosen for adjustment. Eclipse Purge-and-Trap Sample Concentrator Trap* #10 Tenax /silica gel/carbon molecular sieve Purge time and trap temperature* 11 minutes at ambient laboratory temperature Dry purge Not necessary with the patented Cyclone Water Management fitting Desorb preheat temperature On at 180 C Desorb time and trap temperature 0.5 minutes at 190 C Bake time and temperature 10 minutes at 210 C Cyclone Water Management* 110 C at purge, 0 C (ambient) at desorb, 240 C at bake Sparge mount temperature 40 C Sample temperature* 40 C Valve oven temperature 110 C Transfer line temperature 110 C Agilent 6890N GC and 5973 Inert MS Column Restek Rtx -624, 30 m x 0.25 mm x 1.4-µm film Inlet Split/splitless, 35:1 split ratio, 28 ml/minute split flow, 1-mm liner Inlet temperature 240 C Carrier gas Helium at 0.8 ml/minute Oven program 45 C (hold for 4.5 minutes), 12 C/minute to 100 C (hold for 0 minute), 25 C/minute to 240 C (hold for 1.3 minutes) Solvent delay 1.5 minutes MS acquisition mode* Scan 35 260 amu MS quad temperature 150 C MS source temperature 230 C A standard mixture was prepared for the P&T optimization tests which contained ethanol, 1,4-dioxane, and six additional oxygenate compounds. The initial variable testing was performed using this standard mixture at a concentration of 100 ppb. Two internal standards, fluorobenzene and 1-bromo-3-chloropropane, were added at 20 ppb for the calibration and statistical MDL studies. Figure 2 shows a typical total ion chromatogram (TIC) of ethanol, 1,4-dioxane, and the oxygenate compounds listed in Table 2. The inserts illustrate the chromatography of ethanol and 1,4-dioxane prior to optimization of the selected P&T variables. 2

7 4 5 6 1 2 3 8 100 ppb m/z = 45 100 ppb m/z = 88 Figure 2. Total ion chromatogram (TIC) of ethanol, 1,4-dioxane, and six additional oxygenate compounds at 100 ppb using the typical operating conditions listed in Table 2. The inserts show the extracted ion current profiles (EICP) of ethanol and 1,4-dioxane prior to optimization of P&T variables. 3

Results and Discussion Table 2. Compound names, quantitation ions, and retention times used in the analysis No. Compound Name Quantitation Ion Retention Time 1 45 2.88 2 Acetone 43 3.33 3 tert-butanol (TBA) 59 3.95 4 Methyl tert-butyl ether (MTBE) 73 4.16 5 Diisopropyl ether (DIPE) 45 4.89 6 Ether tert-butyl ether (ETBE) 59 5.42 7 tert-amyl methyl ether (TAME) 73 7.02 9 88 8.25 IS Fluorobenzene 96 7.30 IS 1-Bromo-3-chloropropane 77 9.81 Optimization of Purge-and-Trap Variables Several variables of the Eclipse Sample Concentrator were selected to determine their effects on the detection limits and peak shapes of ethanol and 1,4-dioxane. The other six oxygenate compounds can usually be detected at 1 ppb or lower with excellent chromatography using normal P&T conditions. 1 Previous studies revealed that retention of water by the adsorbent material of the analytical trap can be detrimental to the chromatography of polar compounds. 2 Water can interfere with the fast desorption of polar compounds from the analytical trap, creating a wide thermal desorption band and producing irregular peak shapes. P&T variables, such as trap type, sample temperature, trap temperature during purge, and water management fitting temperature during desorb, were selected for optimization due to their impact on water retention by the trap and subsequent chromatography. Trap Type Three traps, #7, #10, and #11, were chosen to determine their effects on peak shapes and recoveries of ethanol and 1,4-dioxane. The #7 trap consists of Tenax as its only component. Tenax traps compounds with boiling points above 35 C, and is highly nonpolar and hydrophobic compared to other trap adsorbent materials. The #10 trap contains three layers of adsorbent materials, Tenax, silica gel, and carbon molecular sieve. Carbon molecular sieve and silica gel are excellent trapping materials for highly volatile compounds such as the six light gases, but silica gel actively retains water due to its hydrophilic nature. The #11 trap, also called the VOCARB 3000, contains Carbopack B, Carboxen 1000, and Carboxen 1001. It is known for its ability to bind a wide range of compounds from the six light gases to the higher boiling compounds such as naphthalene. It also resists absorption of water and methanol, but in some instruments may require a dry purge of one to six minutes to completely remove the retained water. Figure 3 illustrates the relative recoveries of ethanol and 1,4-dioxane using all three trap types, and the amount of water thermally desorbed to the GC from each trap (m/z 18). The relative recoveries for ethanol and 1,4-dioxane were highest using the #7 trap. Compared to the #10 or #11 traps, the #7 trap retained the least amount of water, produced the best peak shapes, and provided the highest response. This result can be attributed to the chemical properties of Tenax, the only adsorbent material in the #7 trap, which has a lower affinity for water than silica gel, carbon molecular sieve, or the VOCARB trap. 2 4

m/z = 45 #7 Trap #7 Trap m/z = 88 EICP Color Trap Black #7 Green #11 Red #11 + dry purge Blue #10 Water m/z = 18 #7 Trap Figure 3. Overlaid EICP for ethanol and 1,4-dioxane using the #7, #10, and #11 traps. The bottom chromatogram shows the overlaid EICP for water (m/z = 18) using all three traps. The #7 trap (Tenax) produced the best peak shapes and highest responses for ethanol and 1,4-dioxane, and transferred the least amount of water to the GC. The #11 trap was run with and without a dry purge step, and transferred the same amount of water using both sets of conditions illustrating the effectiveness of the patented Cyclone Water Management device. 5

Sample Temperature and 1,4-dioxane are extremely polar compounds with low purge efficiency (recovery) due to their high solubility in water. Their purge efficiencies can be significantly improved by heating the sample which forces the volatile compounds out of the liquid and into the gas phase for collection on the trap. The test mixture was analyzed using three sample temperatures: ambient laboratory temperature, 40 C, and 60 C. Figure 4 shows that using a 60 C sample temperature during purge improved recoveries noticeably and produced the highest responses for both ethanol and 1,4-dioxane. Further heating of the sample can produce even higher responses, but the additional water transferred to the trap causes peak shape to deteriorate, offsetting the advantage of increased response. m/z = 45 60 C 40 C Ambient m/z = 88 60 C 40 C Ambient Figure 4. Overlaid EICP of ethanol and 1,4-dioxane using three different sample temperatures (ambient laboratory temperature, 40 C, and 60 C) and a #7 trap. The Eclipse s patented Infra-Sparge Sample Heater provided accurate and reliable sample heating, and direct temperature feedback through a thermocouple located directly in the sample. 6

Trap Temperature During Purge When the analytical trap retains a significant amount of water during purge, the water can create interference by widening the thermal desorption band for the polar compounds, and become a major cause of peak tailing. Previous investigations have found that increasing the trap temperature during purge minimized the amount of water retained on the trap and improved peak shape of some polar compounds3 2 To determine the optimum trap temperature for ethanol and 1,4-dioxane during purge, three trap temperatures were tested: ambient laboratory temperature, 30 C, and 40 C. A 10-ppb standard mix was analyzed to determine the effects of trap temperature on peak shapes and responses of the two test compounds. Figure 5 illustrates how increasing trap temperature produces improved peak shapes for these two polar compounds. Increasing the trap temperature during purge from the ambient laboratory temperature (standard condition) to 40 C resulted in the most improved peak shapes, and peak area counts remained approximately constant. It is important to note that although this technique may improve performance for selected polar compounds, when the trap is held at even slightly elevated temperatures during the purge step, some of the most volatile compounds, such as the six gases, can be lost due to trap breakthrough. 10 ppb m/z = 45 Ambient 10 ppb m/z = 88 Ambient 30 C 30 C 40 C 40 C Figure 5. EICP of ethanol and 1,4-dioxane at 10 ppb using three different trap temperatures during purge (ambient laboratory temperature, 30 C, and 40 C). The slight increase in trap temperature during purge minimized water retention on the trap and narrowed the thermal desorption band for each analyte, but could also cause loss of the gases due to trap breakthrough. 7

Water Management Temperature The Eclipse Sample Concentrator is equipped with the patented Cyclone Water Management fitting (WMF), which efficiently removes the water transferred to and retained on the analytical trap during the purge step. The WMF employs an inert, nickel-plated, near-ambient cyclone chamber positioned at the exit of the adsorbent trap during the desorb step to maximize water removal. During the purge step the WMF is held at 110 C, and all water purged from the sample is transferred directly onto or through the trap, depending on the adsorbent chosen. The WMF is then cooled to the ambient laboratory temperature prior to the desorb step, and the water present in the desorption gas is condensed as it passes through the WMF. In addition to condensation, the cyclone chamber has a unique geometry that forces the desorbing gas into a turbulent vortex that aids in the physical separation of the gas and liquid (water) phases. Finally, during bake, the WMF is heated to 240 C, removing the condensed water and reconditioning the device for the next sample. Since polar compounds such as ethanol are hydrophilic, some portion of these analytes may also be cold-trapped along with the water. This is one of the causes of the high detection limits for ethanol when using standard conditions. The normal setting for the WMF during desorb is 0 C (ambient laboratory temperature). Raising this temperature above ambient transfers more water to the GC during desorb, and it can also noticeably increase the mass of some polar compounds transferred as well. Three WMF temperatures (ambient laboratory temperature, 40 C, and 70 C) during desorb were tested to determine the effect on recoveries of ethanol and 1,4-dioxane. Increasing the WMF temperature to 40 C during desorb increased the ethanol response by approximately 32%, indicating that as much as one-third of the ethanol desorbed from the trap had been condensed in the WMF at ambient temperature. An additional slight increase of about 3% was seen when increasing the temperature to 70 ºC, but this perceived advantage was offset by the additional water transferred to the GC. The peak area counts for 1,4-dioxane remained approximately the same at all three temperatures, indicating that there was no loss of this compound to the WMF when the temperature was set to ambient during desorb. Table 3 shows the peak area counts for ethanol and 1,4-dioxane at the three test temperatures. Table 3. Responses of ethanol and 1,4-dioxane at 100 ppb using three different water management fitting temperatures during desorb. response was improved by approximately 32% when the WMF was heated to 40 C. Response of 1,4-dioxane remained constant at all temperatures. WMF Temperature at Desorb Compound 0 C (Ambient) 40 C 70 C 26,954 35,781 36,988 40,904 39,033 40,861 8

Figure 6 illustrates that the WMF at ambient temperature during desorb transferred the least amount of water to the GC, and that higher temperatures transferred greater amounts of water. The relative difference in water transferred illustrates the effectiveness of the WMF under standard conditions. Figure 7 illustrates the changes in response produced by each test variable. Using these results, the standard P&T operating conditions can be optimized to lower the detection limits for ethanol and 1,4-dioxane to 10 ppb by making the adjustments shown in Table 4. Figure 8 illustrates the response and peak shape of ethanol (m/z = 45) and 1,4-dioxane (m/z = 88) at 10 ppb using the fully optimized conditions. Figure 6. Overlaid EICP of water (m/z = 18) transferred to the GC using three different WMF temperatures during desorb: ambient laboratory temperature, 40 C, and 70 C. Significantly less water was transferred to the GC when the WMF temperature is set at 0 C (ambient) during desorb, illustrating the effectiveness of the WMF under standard conditions. Water m/z = 18 Ambient 70 C 40 C Trap Type Water Management Fitting Temperature during Desorb 600000 45000 500000 40000 400000 35000 Response 300000 Response 30000 200000 100000 25000 0 #10 #11 Using Dry Purge #11NoDryPurge #7 20000 0 C (Ambient) 40 C 70 C Sample Temperature Trap Temperature during Purge Increasing trap temperature during purge improved peak shape significantly, but had little effect on overall response. 500000 180000 450000 160000 400000 350000 140000 120000 300000 Response 250000 200000 Response 100000 80000 150000 60000 100000 40000 50000 20000 0 0 Ambient 40 C 60 C Ambient 30 C 40 C Figure 7. Relative responses of ethanol and 1,4-dioxane using individually optimized P&T conditions. Best overall performance was achieved using a #7 (Tenax) trap, heating the sample to 60 C during purge, and using a WMF setting of 40 C during desorb. Raising the trap temperature to 40 C during purge improved peak shape significantly, but had little effect on overall response. 9

Table 4. Optimized P&T conditions for ethanol and 1,4-dioxane Parameter Standard Optimized Trap type #10 #7 Sample temperature 40 C 60 C Trap temperature during purge Ambient 40 C Water management fitting temperature during desorb Ambient 40 C 10 ppb m/z = 45 10 ppb m/z = 88 Figure 8. EICP of ethanol (m/z = 45) and 1,4-dioxane (m/z = 88) at 10 ppb using the fully optimized P&T conditions: #7 (Tenax) trap, sample temperature of 60 ºC, trap temperature set to 40 ºC during purge, and WMF temperature of 40 ºC during desorb. Calibration and MDL Studies A standard mixture of ethanol and 1,4-dioxane was used to run a calibration curve from 5 ppb to 150 ppb. Fluorobenzene and 1-bromo-3-chloropropane were used as internal standards at a concentration of 20 ppb. The percent relative standard deviations (% RSD) of the response factors were 9.4% for ethanol and 12.4% for 1,4-dioxane, and met the requirement of USEPA Method 8260 of less than 15%. The R 2 values determined using linear regression were also found to be within USEPA requirements at 0.9981 for ethanol and 0.9985 for 1,4-dioxane. The response factors from the calibration curve were used to calculate the statistical method detection limit (MDL) for both compounds using the procedure outlined in the Appendix to Part 136 Definition and Procedure for the Determination of the Method Detection Limit Revision 1.11. To determine the MDL, seven replicates of the 10-ppb ethanol and 1,4-dioxane standard were analyzed. A concentration of 10 ppb was chosen due to the analytes signal-to-noise ratios of 9.5 and 10.9, respectively. The standard deviations of the calculated concentrations of ethanol and 1,4-dioxane were used with the Student s t-test to calculate each compound s MDL. A statistical MDL of 1.8 ppb for ethanol and 2.1 ppb for 1,4-dioxane were calculated using the optimized operating conditions of the Eclipse Sample Concentrator. Table 5 lists the statistical results from the calibration curve and the MDL study. 10

Table 5. Calibration results from 5 to 150 ppb and MDL study of seven replicates of 10-ppb ethanol and 1,4-dioxane using the fully optimized P&T conditions Compound Calibration MDL Study % RSD R 2 Conc. Average (ppb) Standard Deviation (ppb) Statistical MDL (ppb) 9.4 0.9981 9.6 0.58 1.8 12.4 0.9985 10.3 0.68 2.1 Estimated Limit of Quantitation in the SIM Mode The Agilent 5975 Inert MS was switched to the Selected Ion Monitoring (SIM) mode to determine whether the limit of quantitation for ethanol could be lowered further. Monitoring ions 45, 31, and 32 during the elution of a 0.5-ppb ethanol standard resulted in a peak with signal-to-noise ratio of approximately 35, indicating that when using the SIM mode, ethanol can be detected and quantified at single-digit ppb concentrations. Figure 9 illustrates the repeatability of ethanol at 0.5 ppb using the SIM mode. Percent RSDs of the area counts over the seven replicates analyses were 3.6% for ethanol and 5.1% for 1,4-dioxane. 10 ppb SIM mode m/z = 45 % RSD = 3.6 Figure 9. Overlaid EICP (m/z = 45) from seven replicate analyses of ethanol at 0.5 ppb analyzed using the optimized P&T conditions and Selected Ion Monitoring (SIM) 11

Analysis of Methanol Using Optimized Purge-and-Trap Operating Conditions The optimized conditions were also used to evaluate the performance of methanol when analyzed by P&T. The scan range on the MS was changed to 29 to 260 amu to capture methanol s quantitation ions of 31, 32, and 29. Figure 10 illustrates the EICP of methanol at 250 ppb using the fully optimized conditions shown in Table 4. Using the operating conditions before optimization (Table 1), methanol could not be detected at 250 ppb. Figure 10. EICPs for methanol at 250 ppb analyzed on the Eclipse Sample Concentrator using the fully optimized conditions described in Tables 1 and 4. The ions shown are m/z 29, 31, and 32. Methanol 250 ppb m/z = 31, 32, and 29 Conclusions Increasing demands for cleaner burning fuels and use of ethanol as both a fuel additive and a potential alternative to gasoline have led to the need for an analytical method that is capable of detecting and reliably quantifying ethanol at extremely low concentrations. By optimizing the operating conditions of the Eclipse P&T Sample Concentrator both ethanol and 1,4-dioxane can be detected at 10 ppb with no degradation of peak shape. A calibration curve from 5 to 150 ppb passed all acceptance criteria for USEPA Method 8260, and statistical MDLs were 1.8 ppb and 2.1 ppb, respectively. Using the SIM mode allowed detection of ethanol in the single-digit ppb range. References 1. OI Analytical Application Note 1996, Analysis of Fuel Oxygenates Using the Eclipse Purge-and-Trap Sample Concentrator. Available from OI Analytical, P.O. Box 9010, College Station, TX, 77842-9010. 2. McCaffrey, C. A.; MacLachlan, J.; Brookes, B. I. Analyst 1994, 119, 897 902. 3. Helmig, D.; Greenberg, J. P. Journal of Chromatography 1994, 677, 123 132. 4. Volatile Organic Compounds by Gas Chromatography/Mass Spectrometry (GC/MS), Revision 2. Test Methods for Evaluating Solid Waste, Physical/Chemical Methods; SW-846; U.S. Environmental Protection Agency, Office of Solid Waste (OSW) Methods Team: Washington, DC; December 1996; Method 8260B. Agilent and ChemStation are registered trademarks of Agilent Technologies, Inc. Carboxen is a trademark of Supelco, Inc. Rtx is a registered trademark of Restek Corporation. Tenax is a registered trademark of Enka Research Institute Arnhem. VOCARB is a registered trademark of U.S. Filter Corporation. P.O. Box 9010 College Station, TX 77842-9010 Tel: (979) 690-1711 FAX: (979) 690-0440 www.oico.com