Trace Element Analysis of Industrial Wastewater and Sewage with TXRF. Bruker Nano GmbH Berlin, Germany, February 23 rd, 2011

Similar documents
TXRF for Trace Elemental Analysis of Air, Land, and Water. Bruker AXS Inc., Madison, WI, USA November 2012

S4 T STAR Biopharmaceutical production: Rapid nutrient and contaminant analysis of cell culture media by TXRF spectroscopy

Advanced Materials Analysis with Micro-XRF for SEM

STUDY OF ELEMENT FINGERPRINTING IN GOLD DORÉ BY GLOW DISCHARGE MASS SPECTROMETRY

S2 RANGER LE: Analysis of Light Elements in Cement, Slags and Feldspar

Radial, Axial or Dual View ICP: Which Do You Choose? Manny Almeida Teledyne Leeman Labs, Inc. Hudson, NH

Direct Analysis of Photoresist by ICP-MS. Featuring the Agilent Technologies 7500s ICP-MS

Automating EPA 6020 Compliant Analysis with the Agilent 7900 ICP-MS and ESI prepfast Autodilution System

The image part with relationship ID rid4 was not found in the file. Welcome

ESI SC-FAST Sample Introduction Solutions for PerkinElmer ICP-OES and ICP-MS Systems

The Agilent Atomic Spectroscopy portfolio for Environmental applications AA, MP-AES, ICP-OES, ICP-MS & ICP-QQQ

Table I: MCL and MRL Concentrations for Contaminants Monitored Under the Safe Drinking Water Act National Primary Drinking Water Regulations

On site heavy metal testing and monitoring using monochromatic XRF methods

Expanded Coating Analysis Performance for electroless Ni Plating and Bath Analysis with Micro-XRF

Functionality. Normal Operation

Sprint analysis of lubricating oils using the Thermo Scientific icap 7600 ICP-OES

ElvaX Light for Petrochemical analysis

Analysis of inhomogeneous samples and trace element detection in alloys using QUANTAX Micro-XRF on SEM

Atomic Spectroscopy Solutions for Environmental Applications

Increased Sample Throughput for ICP-OES Applied to U.S. EPA Method 200.7

Amplifying ICPMS Productivity Using Discrete Sampling Technology

Analysis of Mineral and Heavy Metal Content in Beverages Using the Teledyne Leeman Labs Prodigy Plus ICP-OES

- Particulate matter (PM10, 2.5, 1) -Coarse region (2.5µm < dp <10µm)

New Advances using Handheld XRF Technology for the Prevention of Flow-Accelerated Corrosion (FAC)

The Determination of Trace Elements in Stainless Steel by Forked Platform GFAAS

Multi-Element Analysis of Trace Metals in Animal Feed using ICP-OES

Collaborative Efforts for Creating a Robust Trace Metals Analytical Procedure for Flue Gas Desulfurization Wastewaters by ICP-MS

SIMPLIFIED MATERIALS ANALYSIS VIA XRF. Ravi Yellepeddi and Didier Bonvin Thermo Fisher Scientific (Ecublens) SARL, Ecublens/Switzerland

Analysis of steel and its alloys using the GB/T standard and an Agilent 5100 ICP-OES in dual view mode

DETERMINATION OF ELEMENTS IN DRINKING WATER AS PER BUREAU OF INDIAN STANDARDS 10500, & USING THE AGILENT 5100 ICP-OES

Characterization of Surface Metals on Silicon Wafers by SME-ICP-MS. Featuring the Agilent Technologies 7500s ICP-MS

Layer Thickness Analysis of Thin Metal Coatings with. Bruker Nano Analytics, Berlin, Germany Webinar, June 8 th 2017

ElvaX ProSpector in Exploration & Mining

XRF DRIFT MONITORS DATA CALIBRATION MATERIAL

Analytical Techniques for Grade and Quality Control in Coal Mining. Dr Paul O Meara XRD Application Specialist

EU water analysis using the Thermo Scientific icap 7400 ICP-OES Duo

SECTIOn 4. Claritas PPT Multi-Element Standards for ICP-MS

SPECTRAL INTERFERENCE IN X-RAY FLUORESCENCE ANALYSIS OF COMMON MATERIALS

Analysis of Toxic Trace Elements in Coffee Bean Products by HR ICP-OES

Lecture Document for Metal Analysis (by the JICA Expert Team) Theoretical Training for AAS

EU Water Analysis Using the Thermo Scientific icap 7400 ICP-OES Duo

ANALYSIS OF TCLP EXTRACTS BY X-RAY FLUORESCENCE

Assaying and Analysis Choices in Practice

Composition Analysis of Animal Feed by HR ICP-OES

Multi-element determination in pharmaceutical preparations using the Thermo Scientific icap Q ICP-MS

Analysis of Environmental Samples with the Agilent 730-ES Following US EPA Guidelines

Determination of Elemental Bad Actors (As, Hg and Si) in Petrochemical Samples Using ICP-AES and ICP-MS

INA-X System for SNMS and SIMS

Guideline regarding analysis on Fujitsu Group specified chemical substances

Gallium Arsenide monocrystalline

Application Note. Abstract. Authors. Environmental

Merits of EPA Method 1640:

Quality Monitoring of Natural Water from Southwest District of Moscow Using Total Reflection X-ray Fluorescence Analysis

Core Analysis with the Tracer

Analysis of inorganic impurities in lubricating oils by ICP-MS

Susan Miller Brick Industry Association. John Sanders National Brick Research Center

Improved Antimony Recoveries in Soil Matrices by 3050B/6010B. DataChem Laboratories, Inc. Robert P. Di Rienzo Jeffery S. Ward John T.

Trace Elements in Petrochemical Samples: Achieving Excellent Analytical Results

ENVIRONMENTAL CHEMICAL ANALYSIS III

Ultra-Trace Analysis of Beryllium in Water and Industrial Hygiene Samples by ICP-MS Application

Determination of trace and ultratrace elements by ICP-SFMS

Analysis of 14 Trace Elements in Stainless Steel

ICP-OES. Varian Vista-MPX CCD Simultaneous ICP-OES

Analytically Monitoring the Effect of Fracturing Activity

Portable Elemental Analysis Food safety and agriculture solutions. Innovation with Integrity. Handheld XRF

Applied Comparison in Coating Analysis Micro-XRF and Handheld XRF

Inductively coupled plasma (ICP) spectroscopy for online measurements of trace metal emissions at the 250 kw PACT furnace

The Determination of Toxic Metals in Waters and Wastes by Furnace Atomic Absorption

Environmental Impacts of Human Activities Monitored by PTEs Contamination of Stream Sediments along the Wad Malyan River

Analysis of wear metals and contaminants in engine oils using the 4100 MP-AES

Microanalysis with high spectral resolution: the power of QUANTAX WDS for SEM

Improving Sampling Techniques for the Analysis of Lead in Groundwater: Determining Optimal Filtration Conditions. Contract Number: SR00-055

Bottom Ash Data Week 45

Determination of Elemental Impurities in Graphite-based Anodes using the Agilent 5110 ICP-OES

Blank Optimization Using Ultrapure Water Suitable for Trace Ion Analysis.

Thermo Scientific X-ray product range. For chemical and phase analysis of solids, liquids and powders

Thermo Scientific X-ray product range. For chemical and phase analysis of solids, liquids and powders

Bottom Ash Data Week 52

X-ray Fluorescence Spectrometry X-ray Fluorescence Spectrometry (XRF) is a non-destructive, quantitative technique for determining chemical

X-Ray Fluorescence Lab Report

ANALYTICAL PERFORMANCE OF A HANDHELD EDXRF SPECTROMETER WITH MINIATURE X-RAY TUBE EXCITATION

Reduction of the Sample Size in the Analysis of Rock by EDXRF

Bottom Ash Data Week 47

Accurate. Robust. Reliable. AGILENT 710 SERIES ICP-OES

Accurate. Robust. Reliable. AGILENT 710 SERIES ICP-OES

Certificate of Analysis

The Determination of Alloying Elements in High-Carbon Steel using Radially Viewed Inductively Coupled Plasma Optical Emission Spectrometry (ICP-OES)

Certificate of Analysis

S8 TIGER Series 2. Lab Report XRF 144. Innovation with Integrity. Accurate Quality Control of Tool Steels XRF

icap 7000 Plus Series ICP-OES Streamlined performance and ultra-low ICP-OES detection limits for routine laboratories

Agilent Atomic Spectroscopy Solutions for Energy & Chemicals AA, MP-AES, ICP-OES, ICP-MS.

Analysis of Metals in Water, Stream Sediments and Floodplain Soils Collected March 21-23, 2005 from the Bayou Creek System. David J.

Trace Metals in Waters by GFAAS, in Accordance with U.S. EPA and Health Canada Requirements

Determination of 22 Elements Following US EPA Guidelines with a New Megapixel CCD ICP-OES

Certificate of Analysis First issued: May 2008 Version: May 2008

ABSTRACT: Clean sampling and analysis procedures were used to quantify more than 70 inorganic chemical constituents (including 36 priority

Agilent 5100 ICP-OES DUAL VIEW ICP-OES MINUS THE WAIT

Analysis of Trace Metals in Tobacco and Tobacco Ash using Atomic Emission Spectroscopy

VHG Labs. Standards and Supplies for Spectrochemical Analysis QUALITY. TRACEABILITY. DEPENDABILITY. WORLDWIDE

US EPA Method using the Thermo Scientific icap 7600 ICP-OES Duo

Transcription:

Trace Element Analysis of Industrial Wastewater and Sewage with TXRF Bruker Nano GmbH Berlin, Germany, February 23 rd, 2011

Welcome Today s Topics TXRF how does it work? Wastewater analysis Water monitoring with bioindicators Comparison with Atomic Spectroscopy methods Interactive Q & A Speakers Dr. Hagen Stosnach Applications Scientist TXRF Berlin, Germany Dr. Armin Gross Global Product Manager TXRF Berlin, Germany

TXRF How does it work?

Principles of total reflection X-ray fluorescence (TXRF) spectroscopy Total reflection X-ray fluorescence spectroscopy monochromator detector Beam angle: 0 o / 90 o X-ray tube sample disc Samples must be prepared on a reflective media Polished quartz glass or polyacrylic glass disc Dried to a thin layer, or as a thin film or microparticle

Principles of total reflection X-ray fluorescence spectroscopy Samples for TXRF Powders: Direct preparation or as suspension Liquids: Direct preparation Always as a thin film, micro fragment or suspension of a powder Necessary sample amount: Low µg respectively µl range Simple quantification Matrix effects are negligible due to thin layer Quantification is possible by internal standardization

Principles of total reflection X-ray fluorescence spectroscopy In TXRF the samples are prepared as thin films or layers Matrix effects are negligible Quantification is possible

Principles of total reflection X-ray fluorescence spectroscopy In TXRF the samples are prepared as thin films or layers Matrix effects are negligible Quantification is possible TXRF detects elements from Na(11) to U(92) The element sensitivities depend on the atomic number K-lines Element sensitivity L-lines The sensitivity factors are calibrated ex works Quantification requires the addition of one standard element Atomic number

Principles of total reflection X-ray fluorescence spectroscopy Mo tube offers lowest LLD Element sensitivity W tube suitable for elements with Z = 41 to 53 (Nb to I) 2 1,5 Si 1 0,5 0 10 20 30 40 50 60 70 80 90 Atomic number K (Mo) L (W) K (Mo) L (Mo)

The instrument S2 PICOFOX Benchtop TXRF spectrometer S2 PICOFOX Metal-ceramic X-ray tube Mo anode Air-cooled Optionally other tubes available Multilayer monochromator XFlash silicon drift detector Electro-thermally cooled 149 ev @ MnKα 100 kcps Automatic version 25 sample cassette

Wastewater analysis by TXRF spectrometry

Sample preparation Suspensions and particulate matter Suspensions can be analyzed right after dilution Dilute sample with distilled water Add internal standard Homogenize Pipette on carrier

Sample preparation Suspensions and particulate matter Dry through heat/vacuum Load the instrument Start data aquisition

Industrial wastewater Cement Situation Discharge of wastewater from a cement plant Threshold value of 500 µg/l Pb to be fulfilled by law Analytical task Filter water with fine particles Thickened sludge as a suspension (cc) pd

Industrial wastewater Cement Measurements and sample preparation Filter water Sludge Filtration Dilution 1:1 Int. Std., 10 ppm Co TXRF

Industrial wastewater Cement Results Filtered samples (with particles or clear) fulfill threshold values Raw sludge is close or above the limits filtration removes lead-containing solids The analysis of further elements is possible (not shown) Pb (µg/l) Filtered sewage Filtered with particles Sludge filtered Sludge suspension Conc. (sample 1) n.d. 4.0 9.4 411 LLD (sample 1) n.d. 2.5 2.0 9.5 Conc. (sample 2) n.d. n.d. 13 938 LLD (sample 2) n.d. n.d. 2.2 16

Industrial wastewater Petro Situation Control of wastewater in the petrochemical industry Analytical task Analysis of wastewater special interest in V, Mn, Ni, Hg (cc) Leonard

Industrial wastewater Petro Measurements and sample preparation Wastewater Int. Std., 10 ppm Ga TXRF

Industrial wastewater Petro Results The elements V, Mn, Ni and Hg could be detected down low ppb The analysis of further elements is possible (not shown) Conc. (µg/l) Sample 1 Sample 2 Sample 3 Sample 4 V < NWG 1.9 79 15 Mn 104 1.3 904 9.9 Ni 1.3 0.68 < NWG 5.4 Hg < NWG 0.46 < NWG 4.0

Industrial wastewater Oil/solvent mixtures Situation Before discharging liquid mixtures of oils and solvents the content of heavy metals has to be verified Analytical task All sewages have to be tested for the entire element range Therefore, all samples were analyzed by two instruments subsequently: W unit for elements from Mo to Sb, Mo unit for all other elements (cc) Michael Meding

Industrial wastewater Oil/solvent mixtures Measurements and sample preparation Sewage water Oil/solvent mix. Oil samples Weighing (100-500 mg) Dilution, 500 µl CHCl 3 Int. std., 50 ppm Ga Int. standard Co, oil-based, 50 ng TXRF

Industrial wastewater Oil/solvent mixtures Results sewage All elements from P to Hg were quantitatively analyzed Detection limits were < 10 ppb for most elements (Z > 19) Detection limits measured by the W unit were higher but fulfilled the required limits (e.g. Cd < 80 ppb, Sb < 160 ppb)

Industrial wastewater Oil/solvent mixtures Results oil samples and mixtures Best detection limits were in the range from 15 to 60 ppb (e.g. Br, Se, As, Zn, Ni, Cu) LLD measured by the W unit depend on sample type: Cd: 0.3 ppm in solvent, 1 2 ppm in oils and mixtures Sb: 0.6 ppm in solvent, 4 ppm in oils Mo: 0.2 ppm in solvent, 1 ppm in oils and mixtures Improved sample preparation for achieving lower LLD is possible

Industrial wastewater TiO 2 production Situation Before discharging the concentrations of several heavy metals they must be analyzed in order to fulfill legislative norms Analytical task Wastewater has to be analyzed with regard to limit values for Cr < 0,5 mg/l Cd < 5,0 µg/l Hg < 1,0 µg/l Concentrations in the mg/l range of additional elements (Fe, Ti and V) have to be supervised

Industrial wastewater TiO 2 production Measurements and sample preparation Wastewater Wastewater Int. Std., 10 ppm Ga Dilution 1:10 or more (filtration) TXRF ICP-MS

Industrial wastewater TiO 2 production Fast screening of wastewater delivers comparable results of TXRF and ICP-MS TXRF: Cd and Hg (conc. range 30 to 70 ppb) were not detectable due to line overlaps ICP-MS: S, Cl, Br, Rb, Sr were not detectable Recovery (%) 200% 150% 100% 50% 0% K Ca Ti V Cr Mn Fe Ni Cu Zn Pb

Industrial wastewater TiO 2 production Detection limits 1.000 100 LLD (µg/l) 10 1 S Cl K Ca Ti V Cr Mn Fe Ni Cu Zn As Se Br Rb Sr Pb

Water monitoring with bioindicators

Water monitoring with bioindicators Introduction Bioindicators are biological species used to monitor the health and integrity of an environment or ecosystem monitor for changes of metal pollutions in surface waters and sediments General questions Accumulation through the nutrition chain? Impact of sublethal metal concentrations? Methods for fast screening?

Bioaccumulation of metals in the trophic chain Element mass fraction increases at log scale Secondary consumer Tertiary consumer Primary consumer Primary producer

Metal screening of single Daphnia Zooplankton Daphnia as bioindicator Describes biological availibility of heavy metals, indicator for water quality Measurements of single organisms (< 100 µg) after selection by type, sex, growth status Trace element analysis of single organisms by ICP-OES and AAS impossible

Metal screening of single Daphnia Sample preparation Rinsing of Daphnia Freeze drying Selection of individual organisms, weighing and size measurement Fix Daphnia on quartz carrier, add HNO 3 and internal standard Cold Plasma Ashing (2 h, 180ºC) no loss of volatile trace element compounds CPA reduces organic matrices and lowers background counts TXRF measurement Typically 1000 s

Metal screening of single Daphnia Direct correlation of metal concentration and dry weight of bioindicator organism mass fraction (ug/g DW) 1000 100 10 Ni mass fraction (ug/g DW) 100 10 Pb R 2 = 0,9601 1 10 100 1000 10000 1000 DW (mg) Fe R 2 = 0,928 1 10 100 1000 10000 100 DW (mg) Cu mass fraction (ug/g DW) 100 10 mass fraction (ug/g DW) 10 R 2 = 0,8101 1 10 100 1000 10000 DW (mg) R 2 = 0,7918 1 10 100 1000 10000 DW (mg)

Water monitoring with bioindicators What are biofilms? Dry sample microscopy photo Wet sample underwater photo 1 mm 10 mm Microorganisms Water surface area EPS, extracellular polymeric substances Water flow Growth carrier Courtesy of UFZ Magdeburg

Significance of biofilms Negative impact of biofilms Bioaccumulation of harmful chemicals in aquatic systems Biogenic corrosion of metals, concrete etc. Biofouling in pipeline systems Germination of water pipelines Contamination with pathogenes in hospitals Application of biofilms Self-purification of waters Binding and removal of toxic materials Biotechnological use Removal of organics and N-compounds in wastewater treatment plants

Biofilms Sampling location Elbe watershed Harz mountain range mining area since bronce age Schlüsselstollen Schlenze Saale Flow direction Courtesy of UFZ Magdeburg

Biofilms Sampling location Sampling location Schlüsselstollen Gallery was built in 1879 Drains the mining district at a length of 31 km Courtesy of UFZ Magdeburg

Biofilms Sampling Grown on round polycarbonate slide Courtesy of UFZ Magdeburg

Biofilms Sample preparation for TXRF Sampling and transport (4 º C) Rinse with clean water Freeze drying Homogenization Weighing ca. 500 µg at Ultrabalance Digestion with HNO 3 /H 2 O 2 Internal standard Sc 10 µl sample aliquot Drying on a hot plate 80 º C Analysis with TXRF

Accumulation factors in water and biofilms Element Accumulation factor water* Accumulation factor biofilm* Pb 1400 2200 Zn 630 18 Cu 11 140 Ni 8 7 *) relation of gallery to creek water Biofilms: Pb > Cu > Zn > Ni > Fe > Ca

Sorption effects in biofilms Immediate accumulation of Pb and Ni during growth Release during longer exposition Slow accumulation of Cu and Zn

Outlook Biofilms Further investigations are planned on a river platform (meso cosmos) with flow lines simulating natural flow behaviours Acknowledgement Margarete Mages, Helmholtz Centre for Environmental Research (UFZ) Magdeburg, Germany

Comparison with Atomic Spectroscopy

TXRF vs. AAS/ICP-OES TXRF Small footprint/portable Microgram sample size %-ppb levels Single std. calibration Low maintenance Non-destructive Fast learning curve Fast sample prep AAS/ICP-OES Large, fixed installation Milligram sample size ppm sub ppb levels Multi-standard calibration High maintenance Destructive Long learning curve Laborious sample prep

Want ICP-MS? Bruker 820-MS Low DLs <1ppt for some elements in solution Wide range of elements (>75) Isotopic analysis capability

Innovative Ion optics 90 º design Ionen mirror Parabolic electrostatic field Reflects and focusses ions (patented) Hollow structure of the ion mirror Photons and uncharged particles pass without contact Arrangement of turbo pump optimal conditions for vacuum removes unwanted particles

Convenient technology Collision Reaction Interface CRI Bruker s answer to common interferences? The Collision/Reaction Interface (CRI) H 2 gas injected to react with Ar + and ArX + molecular ions He gas used to separate other interferences Improved detection limits for several key elements ( 40 Ca, 56 Fe, 75 As, 80 Se) Increased productivity, most suitable for routine applications Fast and reproducible switching of CRI modes

Application labs in your neighborhood Application labs with TXRF and ICP-MS at work or coming soon Berlin, Germany Coventry, UK Paris, France Milton, Canada Houston, Texas, USA Melbourne, Australia Seoul, Korea

Q & A Any Questions? Please type in the questions you may have for our speakers in the Questions form and click Submit

Thank you for your attention! Upcoming: 14 th International Conference on Total Reflection X-ray Fluorescence and Related Methods 06 10 June 2011, Dortmund, Germany www.bruker.com www.s2picofox.com

Copyright Bruker Corporation. All rights reserved. www.bruker.com