Experimental Protocol for Multiplex Fluorescent Blotting Using the ChemiDoc MP Imaging System

Similar documents
Clarity and Clarity Max Western ECL Substrates

Immun-Star WesternC Chemiluminescent Kit

western blotting tech

Multiplex Fluorescent Western Blot Starter Kit for the Bio- Rad ChemiDoc MP

PrimePCR Pricing and Bulk Discounts

Blotting and Immunodetection. Multiplex Fluorescent Blot Detection: A Troubleshooting Guide

PrimePCR Pricing and Bulk Discounts

and Criterion TGX Stain-Free Precast Gels

Imaging and Electrophoresis. Gel Doc EZ Imager. Accelerate your research. Push-button imaging in just 10 seconds.

PrimePCR Pricing and Bulk Discounts

ProteOn XPR36 Quantikinetics: Antibody Concentration and Detailed Kinetic Analysis in a Single Experimental Cycle

Superior Performance of Laemmli PAGE with Extended Shelf Life

Strep-tag Technology for Molecular Weight (MW) Determinations on Blots Using Precision Plus Protein Standards

StarBright Blue Fluorescent Secondary Antibodies

IgG TrueBlot Protocol for Mouse, Rabbit or Goatderived Antibodies - For Research Use Only

electrophoresis tech Mini-PROTEAN TGX Precast Gel: A Gel for SDS-PAGE with Improved Stability Comparison with Standard Laemmli Gels

ADDENDUM. Ready Gel System RESOURCE GUIDE. Applications:

Overview. Tools for Protein Sample Preparation, 2-D Electrophoresis, and Imaging and Analysis

Western Blotting Detection Reagents

ProteinChip IMAC30 Array (Immobilized Metal Affinity Capture) Instruction Manual. Catalog #C

Electrophoresis. Ready-to-Run Buffers and Solutions

electrophoresis tech Versatile Separation Capabilities of the PROTEAN i12 IEF System

Product Catalog # Description List Price (JPY) Primer Assays (desalted)

Pulsed Field Gel Electrophoresis

Chromatography Column Performance and Data Analysis Success Guide. Hints and Tips for Better Purifications

ChemiDoc MP Imaging System

Western Blot Tool Box

Ordering Information. ProteOn XPR36 System. Protein Interaction Analysis

Orexin A (HUMAN, MOUSE, RAT, PORCINE, OVINE,

This Document Contains:

ProteoMiner Protein Enrichment Technology

Western Blot Analysis

hfab Rhodamine Housekeeping Antibodies

Technical Note Detection of post-immunoprecipitation proteins by Western blot using the Quick Western Kit IRDye 680RD

Western Blot Protocol

Bio-Rad Laboratories BIOPLEX 2200 SYSTEM. BioPlex 2200 ANA Screen with MDSS. The first and only fully-automated, multiplexed ANA Screen

Bio-Plex Pro Mouse Cytokine, Chemokine, and Growth Factor Assays

Western Blotting Products. A Complete Source for All Your Blotting Needs

Bio-Plex Pro Mouse Cytokine, Chemokine, and Growth Factor Assays

Sensido Advance ECL Substrate (HRP) Solution A (Substrate Solution) Solution B (Peroxide Solution)

Technical Data Sheet. Sensido Plus ECL Substrate (HRP) Description. Advantages

ab Ran Activation Assay Kit

protein interaction analysis

electrophoresis tech Performance Comparison of the Experion Automated Electrophoresis System and SDS-PAGE for Protein Analysis

protein interaction analysis tech note 5367

KPL SignaLOCK ChemiWestern Kits (Film and Imager Analysis)

Get Your Fill. A full line of empty chromatography columns and accessories. Protein Purification

In-Cell Western Kits I and II

IR-Blot Secondary antibodies Rev01

IR-Blot Secondary antibodies Rev00

Innovations To Meet Your Needs

SYNGENE APPLICATION TRAINING GUIDE

Strep-tag detection in Western blots

pimago -biotin Phosphoprotein Detection

In-Gel Western Detection Using Near-Infrared Fluorescence

Standards for Electrophoresis and Blotting

Nuvia HR-S Strong Cation Exchange Media Instruction Manual

Discover the Importance of Third Party Quality Control

Kinase Reaction and Alkylation Protocol

ab Optiblot Fluorescent Western Blot Kit

Western BLoT Hyper HRP Substrate

Western BLoT Quant HRP Substrate

Western BLoT Chemiluminescence HRP Substrate

Western BLoT Ultra Sensitive HRP Substrate

Western blot: proteins separating technique, protocol, theory and trouble shooting

Western BLoT Quant HRP Substrate

Chromatography Success Guide. Tips and Tricks for Successful Preparative Chromatography and Protein Purification

Detection of Hog1 Phosphorylation in Candida albicans in Response to an Antifungal Protein Brigitte ME Hayes and Nicole L van der Weerden *

Segments of the obstructed intestinal loops were fixed in 4% paraformaldehyde

BioPlex 2200 Syphilis Total & RPR Assay

Reach greater highs. (and lows ) with new protein ladder choices. Fermentas now sold as. Thermo Scientific

Using Sapphire700 Stain and DRAQ5 Stain for Cell Number Normalization

Immunohematology. IH-1000 Fully Automated System for ID-Cards

Dolphin-Chemi Plus. Aim: To visualise and evaluate the performance of chemiluminescent immunoblots using Wealtec s Dolphin-Chemi plus image system

protein interaction analysis bulletin 6299

ONE-HOUR Western TM Multiplex Kit II

KPL LumiGLO Ultra Western Blotting Substrate

MATERIAL DATA SHEET. NOTE: Kit contains reagents sufficient for 10 x 30 μl reactions and 5 Western Blots (mini-gel format). Reagents Provided in Kit

Solutions for Western Blotting

Azure Biosystems Western Blotting Workflow

Optiblot ECL Ultra Detect Kit (1.2pg 2ng)

MATERIAL DATA SHEET. Reagents Provided in Kit

Bio-Plex suspension array system

Western BLoT Blocking Buffer (Fish Gelatin)

*with VIVID Technology*

Supplementary Material - Methods

Optiblot ECL Detect Kit (23 pg 187 ng)

1. Goat Anti-Caspase-3 (CPP32) Antibody, R&D systems (cat #AF-605-NA), 0.5ug/ml

Odyssey Western Blotting Kits LT

ReliaBLOT TM. IP/Western Blot Reagents Cat. No. WB120, rabbit

Introduction. Important Product Information. Number Description. Table of Contents

Cdc42 Activation Assay Kit

ab64254 Liquid Fast-Red Substrate Kit (75X)

RheB Activation Assay Kit

Bio-Plex Pro Human Inflammation Assays

Western BLoT Immuno Booster

QIAGEN Supplementary Protocol

Western blot troubleshooting guide

ab ERK1/2 (Thr202/Tyr204) In-Cell ELISA Kit

Protein molecular weight marker and anti-marker antibody for chemiluminescent detection on Western blots.

Transcription:

Experimental Protocol for Multiplex Fluorescent Blotting Using the ChemiDoc MP Imaging System Protocol Bulletin 6570 Stefanie L. Ritter and Donald G. Rainie, Deparment of Behavioral Neuroscience and Psychiatric Disorders, Yerkes National Primate Research Center, Emory University, Atlanta, GA 30322 Bio-Rad provides instruments and reagents for multiplex fluorescent western blotting (MFWB), a powerful technique that is gaining increased attention in the life science community. Like any other assay, MFWB requires careful assay development to guarantee high-quality data. Here, we introduce a protocol independently generated by Dr. Stefanie Ritter, a postdoctoral fellow in Dr. Donald Rainie s lab at Emory University. We hope this protocol and the included tips will help other scientists have a successful experience with MFWB. Background Fluorescence-based multiplex western blot detection is particularly suited for simultaneous detection of phosphorylated and total protein populations, as it allows for accurate and easy signal normalization within the same sample on the same blot. Moreover, the ability to simultaneously detect two proteins of interest increases the information that can be collected from your sample. This is especially important if you have only enough sample for one or two western blots. Fluorescence-based western blot detection is very sensitive but lacks the additional amplification achieved when using horseradish peroxidase (HRP) secondary antibodies and chemiluminescence-based detection. Importantly, the factors affecting signal-to-noise ratio after the detection of the protein(s) of interest must be optimized, including blocking conditions, antibody concentration, and the amount of experimental sample to be probed. Likewise, the fluorescence-based detection requires that care be taken to avoid exposing the membrane to dust and/or other environmental contaminants, which could introduce unwanted background. The ChemiDoc MP Imaging System is a high-resolution gel documentation system that allows fast, easy quantification of gels and blots. It uses a new-generation lighttight enclosure (the universal hood III) that minimizes contaminants. It contains LED-based illumination to detect fluorescent secondary antibodies in the visible range. Here, we describe a protocol using the ChemiDoc MP Imaging System to detect levels of phosphorylated and total populations of the extracellular signal-regulated kinase (ERK) protein in brain samples. The detection of phospho-erk (perk) using either traditional chemiluminescence-based detection or fluorescence-based detection are compared as well. Materials and Methods Harvest buffer 150 mm NaCl 20 mm Hepes 1x Halt Protease and Phosphatase Inhibitor Cocktail and 1x EDTA (Thermo Fisher Scientific) 6x Laemmli buffer Rat brain tissue Tissue homogenizer 95 C heat block 4 20% Mini-PROTEAN TGX Gels (Bio-Rad Laboratories) Tris-glycine SDS-PAGE running buffer Tris-glycine transfer buffer supplemented with 20% methanol PowerPac Power Supply and Mini Trans-Blot Cell (Bio-Rad Laboratories) Precision Plus Protein Standards (Bio-Rad Laboratories) Nitrocellulose membranes (Bio-Rad Laboratories) LF-PVDF membranes (Bio-Rad Laboratories) Methanol 10x Tris-buffered saline (TBS) 2 M Tris (25.4 g Trizma Cl + 4.72 g Trizma base) 1.5 M NaCl (87.66 g) Diluted in 1 L of dh 2 O 1x TBS (10x diluted in dh2 O) 1x TBS/T (10x diluted in dh 2 O w/ 0.1% Tween 20 added prior to use) Ponceau S 0.1 N NaOH Black or opaque blotting boxes (Rockland Immunochemicals/ LI-COR Biosciences) Lid to prevent dust contamination and/or evaporation Bovine serum albumin (BSA) Primary antibodies (see protocol) Secondary antibodies (see protocol) Forceps SuperSignal West Pico Chemiluminescent Substrate (Thermo Fisher Scientific) ChemiDoc MP Imaging System (Bio-Rad Laboratories)

Procedure Homogenize and process brain tissue in harvest 1 10 buffer. Rinse membranes 3 times for 5 min in 1x TBS/T. 2 3 4 5 6 7 8 9 Dilute samples in 6x Laemmli buffer, boil at 95 C for 5 min and load onto a 4 20% Mini-PROTEAN TGX Gel in the order listed in Figure 1. Run protein samples using SDS-PAGE and transfer to LF-PVDF using standard transfer conditions. Cut LF-PVDF membrane into two sections and reversibly stain with Ponceau S to examine transfer efficiency. Remove stain by washing the membrane with 0.1 N NaOH and rinse twice in 1x TBS. Block membranes for 30 min in freshly prepared 3% BSA in TBS with agitation at room temperature (do not include Tween in this step). Probe membranes with either mouse anti-perk (1:500, Santa Cruz Biotechnology, Inc.) (Figures 1A and 1B) or rabbit anti-total ERK (terk) (1:1,000, Cell Signaling Technology) (Figure 1C) diluted in 3% BSA in TBS/T. Process overnight with agitation at 4 C. Return membranes to an orbital shaker for 10 min at room temperature. Discard primary antibody solutions and wash membranes 3 times for 5 min in 1x TBS/T. Incubate membranes with their respective secondary antibodies. The membranes shown in Figure 1 were probed with the following secondary antibodies: A, anti-mouse HRP (1:8,000, Vector Laboratories); B, anti-mouse 680 (Alexa Fluor 1:4,000); C, anti-rabbit 549 (DyLight 1:4,000). The antibodies were diluted in 3% BSA in TBS/T and the membranes were incubated at room temperature for one hour with gentle agitation. 11 12 Incubate the membrane probed with HRP secondary antibody in SuperSignal West Pico Chemiluminescent Substrate for 5 min at room temperature with gentle agitation. Discard the substrate and image the blot on the ChemiDoc MP Imaging System using the Chemi Hi Sensitivity feature at a manual exposure of 1 60 sec, acquiring ten images throughout the duration of the imaging. Note: In Figure 1, the image captured at 6.8 sec was selected and saved, as it did not have any signal saturation. Rinse the fluorescent membranes (Figures 1B and 1C) with 1x TBS and keep in TBS until ready to image on the ChemiDoc MP Imaging System. Image the fluorescent membrane using a protocol to detect the IR 680 (Autoexpose, Faint Bands, 30 sec) and DyLight 549 (manual exposure, 360 sec) secondary antibodies. Note: The manual exposure time was chosen to avoid signal saturation and to increase the information gathered in one image. The gain of the images was adjusted to minimize background and a gain was selected in which perk bands at both 44 and 42 kd could be distinguished. A comparison of detection with fluorescence and chemiluminescence using the above protocol is shown in Figure 1.

A. Chemiluminescence of perk B. Fluorescence of perk C. Fluorescence of terk D. Fluorescence Merged Image 1 2 3 4 5 6 1 anti-perk 2 anti-mouse IR 680 1 anti-terk 2 anti-rabbit 549 Merged Fig. 1. Comparison of perk/terk detection methods. Lane 1, 15 µl of Precision Plus Protein All Blue Standards (1:10 dilution); lanes 2 5 and lane 14, 20 µl of experimental brain sample 2; lanes 6, 8, 13, and 15, 10 µl of 1x Laemmli buffer; lanes 9 and 10, 20 µl of control brain sample; lanes 11 and 12, 20 µl of experimental brain sample 1; lane 7, 5 µl Precision Plus Protein All Blue Standards (1:10 dilution). Probing of brain samples with either chemiluminescence-based detection of perk (A) or a multiplex fluorescence-based detection of perk and terk (B, C, and D) resulted in comparable patterns of detection. 1 anti-perk 2 anti-mouse HRP Important Information Choice of Membrane It is imperative to use a membrane with low autofluorescence for fluorescence-based antibody detection. Although nitrocellulose membranes are cheap and widely used, a number of companies have formulated low-fluorescence PVDF (LF-PVDF) membranes for use with fluorescence-based imaging. Nitrocellulose membranes autofluoresce in the visible and near infrared (near IR) spectrum, making background a more substantial issue and minimizing the detectable range useful for imaging (Figure 2A). In fact, nitrocellulose membranes may work only with the near IR or IR spectrum. In contrast, LF-PVDF membranes activated with methanol prior to imaging on the ChemiDoc MP Imaging System have little to no detectable background in the visible or near IR range (Figure 2B). Importantly, if a PVDF membrane is used, it will need to be activated with methanol. When using methanol, make sure the membrane does not dry out during the blotting and imaging process. Water may be sprinkled onto the membrane for the final imaging steps with the ChemiDoc MP Imaging System. A. Nitrocellulose Membranes B. Activated LF-PVDF Membranes Alexa Fluor 488 Alexa Fluor 488 Alexa Fluor 680 Alexa Fluor 680 Fig. 2. Comparison of the autofluorescence of nitrocellulose and LF-PVDF membranes for fluorescence-based imaging. Images show autofluorescence of nitrocellulose membranes (A) and LF-PVDF membranes (B) in the visible (top) and infrared (bottom) spectra. Note: Low and high gain refers to the gain settings for the image as displayed in the Image Transform options of the ChemiDoc MP Imaging System. Images were captured at 30 sec, in accordance with timing required for routine imaging detection.

Importance of Blocking Buffers Optimization of multiplex detection requires titrating the perk and terk antibody conditions in a variety of blocking buffers, including 5% milk diluted in a HEPES-buffered salt solution in H 2 O, 5% casein blocker in PBS or TBS, LI-COR Odyssey buffer, Rockland fluorescence buffer, or 3 5% BSA diluted in PBS or TBS. Generally, TBS is useful when working with phosphoproteins and is recommended by Cell Signaling Technologies (purveyor of most phosphoprotein antibodies used in this study). The goal is to increase the sensitivity of the antibody for the protein of interest while achieving specificity of the signal. As shown in Figure 3, the same sample probed with the same amount of primary antibody can vary widely in its pattern of detection depending on blocking and incubation conditions. This highlights that an antibody must be optimized on a case-by-case basis to achieve optimal results for imaging. A pstat3 Stat3 B pstat3 Stat3 Alexa Fluor 680, 22 sec Rockland block, Rockland+PBS/T 1 and 2 Rockland block, 5% milk 1 and 2 C pstat3 Stat3 D pstat3 Stat3 Rockland block, 5% BSA in PBS/T 1 and 2 Odyssey block, 5% BSA in PBS/T 1 and 2 Fig. 3. Comparison of the effect of blocking conditions on multiplex fluorescent detection of pstat3 and Stat3. Alternating samples of the Precision Plus Protein Unstained Standards and 20 µg of brain lysate were subjected to SDS-PAGE, transferred to LF-PVDF, and cut into four separate membranes for blocking buffer optimization. The membranes were blocked in various buffers such as Rockland blocking buffer (A C) or LI-COR Odyssey blocking buffer (D). After a 30 min incubation in blocking buffer, the membranes were incubated with rabbit anti-pstat3 and mouse anti-stat3 antibodies (Cell Signaling Technology) diluted in various buffers, 1:1 dilution of Rockland and PBST (A), 5% nonfat milk in 20 mm HEPES and 150 mm NaCl w/0.1% Tween (B), or 5% BSA in PBS/T (C and D), and incubated overnight at 4 C with agitation. Approximately 18 hr later, the primary antibody solutions were discarded, the membranes were washed 3x for 5 min in either PBS/T (A, C, and D) or milk (B), and the primary antibodies were detected with Alexa Fluor 680 (pstat3, 1:4,000 dilution) or Alexa Fluor 488 (Stat3, 1:4,000 dilution) in the same diluent as the primary antibody. Note: pstat3 images were transformed to emphasize the faint banding pattern across conditions, while Stat3 images were not transformed and thus represent the entire dynamic range for that image capture. Additionally, Stat3 may be directly compared across the four conditions, as the exposure time for the Stat3 channel (Alexa Fluor 488) is identical for all four. Overall, transitioning from traditional film and chemiluminescencebased western blot detection to fluorescence-based detection will most likely require adjusting the protocol and titrating antibodies for the new method of detection. It is imperative to optimize the signal-to-noise ratio of the antibody, especially for fluorescence-based detection, in order to measure a clear signal for the protein of interest.

General Tips 1 2 3 4 Wear gloves at all times while handling membranes. Use freshly cleaned boxes for incubations (clean with ethanol and dh 2 O prior to use). Keep blot boxes closed while incubating to avoid dust exposure. Cap all liquid reagents when not in use and filter solutions, if necessary, to remove particulates (might be helpful for milk-based blocking solutions). Optimize blocking and primary antibody conditions. See Figure 3 for specifics. Membranes may be blocked for 30 60 minutes prior to adding the primary antibody. Generally, incubating primary antibodies overnight at 4 C will facilitate detection of the protein of interest. If necessary, store the antibody at 4 C and use within a few days. However, bacteria/mold might begin to grow in the antibody container, which could lead to an artifact signal on the blots. Ideally, use fresh antibodies for imaging. Secondary antibody concentrations may also be optimized. For example, Stat3 can be detected when using Alexa Fluor 488 at 1:4,000 and 1:10,000 dilutions. The final concentration, again, will depend on the signal-to-noise ratio. While exposing the blots, choose exposure times that are just prior to signal saturation. For example, although the ERK bands shown in Figure 1 were exposed for 30 seconds using the Autoexpose feature for Faint Bands, a longer exposure time of 360 seconds could be used to capture more information in the image. However, there may be changes to the densitometry measurements between the two exposure times or changes to the Image Transformation features. This could be empirically determined. Precision Plus Protein standards are sold under license from Life Technologies Corporation, Carlsbad, CA for use only by the buyer of the product. The buyer is not authorized to sell or resell this product or its components. Alexa Fluor is a trademark of Life Technologies Corporation. LI-COR and Odyssey are trademarks of LI-COR, Inc. SuperSignal, Halt, and DyLight are trademarks of Thermo Fisher Scientific Inc. Trizma is a trademark of Sigma-Aldrich Corporation. Tween is a trademark of ICI Americas, Inc. Bio-Rad Laboratories, Inc. Life Science Group Web site www.bio-rad.com USA 800 424 6723 Australia 61 2 9914 2800 Austria 01 877 89 01 Belgium 09 385 55 11 Brazil 55 11 3065 7550 Canada 905 364 3435 China 86 21 6169 8500 Czech Republic 420 241 430 532 Denmark 44 52 10 00 Finland 09 804 22 00 France 01 47 95 69 65 Germany 089 31 884 0 Greece 30 210 9532 220 Hong Kong 852 2789 3300 Hungary 36 1 459 6100 India 91 124 4029300 Israel 03 963 6050 Italy 39 02 216091 Japan 81 3 6361 7000 Korea 82 2 3473 4460 Mexico 52 555 488 7670 The Netherlands 0318 540666 New Zealand 64 9 415 2280 Norway 23 38 41 30 Poland 48 22 331 99 99 Portugal 351 21 472 7700 Russia 7 495 721 14 04 Singapore 65 6415 3188 South Africa 27 861 246 723 Spain 34 91 590 5200 Sweden 08 555 12700 Switzerland 026 674 55 05 Taiwan 886 2 2578 7189 Thailand 1800 88 22 88 United Kingdom 020 8328 2000 Bulletin 6570 Rev A US/EG 14-0881 0614 Sig 1213