Bio-Monolith Protein G Column - More Options for mab Titer Determination

Similar documents
mab Titer Analysis with the Agilent Bio-Monolith Protein A Column

Analysis of Intact and C-terminal Digested IgG1 on an Agilent Bio MAb 5 µm Column

Separate and Quantify Rituximab Aggregates and Fragments with High-Resolution SEC

Faster Separations Using Agilent Weak Cation Exchange Columns

Application Note. Biopharma. Authors. Abstract. James Martosella, Phu Duong Agilent Technologies, Inc Centreville Rd Wilmington, DE 19808

Separation of Recombinant Human Erythropoietin (repo) Using Agilent Bio SEC-3

High-resolution Analysis of Charge Heterogeneity in Monoclonal Antibodies Using ph-gradient Cation Exchange Chromatography

High-throughput and Sensitive Size Exclusion Chromatography (SEC) of Biologics Using Agilent AdvanceBio SEC Columns

Agilent AdvanceBio SEC Columns for Aggregate Analysis: Instrument Compatibility

Application Note. Author. Abstract. Biopharmaceuticals. Verified for Agilent 1260 Infinity II LC Bio-inert System. Sonja Schneider

Application Note. Authors. Abstract. Biopharmaceuticals

Charge Heterogeneity Analysis of Rituximab Innovator and Biosimilar mabs

Cell Clone Selection Using the Agilent Bio-Monolith Protein A Column and LC/MS

Rapid UHPLC Analysis of Reduced Monoclonal Antibodies using an Agilent ZORBAX Rapid Resolution High Definition (RRHD) 300SB-C8 Column

LOABeads PrtA Magnetic bead purification of antibodies Capacity 45 mg IgG/ml

Reducing Cycle Time for Charge Variant Analysis of Monoclonal Antibodies

TECHNICAL BULLETIN. HIS-Select HF Nickel Affinity Gel. Catalog Number H0537 Storage Temperature 2 8 C

A Comprehensive Workflow to Optimize and Execute Protein Aggregate Studies

Fast and High-Resolution Reversed-Phase Separation of Synthetic Oligonucleotides

Reversed-phase Separation of Intact Monoclonal Antibodies Using Agilent ZORBAX Rapid Resolution High Definition 300SB-C8 1.

PROCEDURE FOR USE NICKEL NTA Magnetic Agarose Beads (5%)

Multiple Detector Approaches to Protein Aggregation by SEC

Optimizing Protein Separations with Agilent Weak Cation-Exchange Columns

Choosing the Right Calibration for the Agilent Bio SEC-3 Column

Cell Culture Optimization Using an Agilent Bio-Monolith Protein A Column and LC/MS

AdvanceBio HIC: a Hydrophobic HPLC Column for Monoclonal Antibody (mab) Variant Analysis

Seamless Method Transfer from an Agilent 1260 Infinity Bio-inert LC to an Agilent 1260 Infinity II Bio-inert LC

PROTEIN L AGAROSE BEADS

Clone Selection Using the Agilent 1290 Infinity Online 2D-LC/MS Solution

Agilent Biocolumns. Titer Determination. Application Compendium

Molecular Characterization of Biotherapeutics The Agilent 1260 Infi nity Multi-Detector Bio-SEC Solution with Advanced Light Scattering Detection

Column for High Performance,High-Binding Capacity Ion Exchange Chromatography:TSKgel SuperQ-5PW and Its Applications

ProPac Elite WCX, 5 μm Particle, for Fast, High Resolution Protein and mab Analysis. July 2018

Ni-NTA Agarose. User Manual. 320 Harbor Way South San Francisco, CA Phone: 1 (888) MCLAB-88 Fax: 1 (650)

Chromatography column for therapeutic protein analysis

Size Exclusion BioHPLC Columns

Refold SK Protein Refolding Kit

Purification of GST-tagged proteins using PureCube Glutathione Agarose

TECHNICAL BULLETIN. Ni-CAM HC Resin High Capacity Nickel Chelate Affinity Matrix. Product No. N 3158 Storage Temperature 2 8 C

Disulfide Linkage Analysis of IgG1 using an Agilent 1260 Infinity Bio inert LC System with an Agilent ZORBAX RRHD Diphenyl sub 2 µm Column

rprotein A GraviTrap Protein G GraviTrap rprotein A/Protein G GraviTrap

Strep-Spin Protein Miniprep Kit Catalog No. P2004, P2005

Strep-Spin Protein Miniprep Kit Catalog No. P2004 & P2005

Characterization of mab aggregation using a Cary 60 UV-Vis Spectrophotometer and the Agilent 1260 Infinity LC system

SERVA Ni-NTA Magnetic Beads

1. Cross-linking and cell harvesting

Size Exclusion BioHPLC columns Ion Exchange BioHPLC columns

ADVANCE ACCURACY AND PRODUCTIVITY FOR FASTER ANALYSIS

CX-1 ph Gradient Buffer

Protein Separation with ph Gradients Using Composite Buffer Systems Calculated by the Agilent Buffer Advisor Software

Thermo Scientific. MAbPac HIC-10. Product Manual. P/N: June Part of Thermo Fisher Scientific

Affi-Gel Protein A MAPS II Kit Instruction Manual

ADVANCE ACCURACY AND PRODUCTIVITY FOR FASTER ANALYSIS

Separation of Monoclonal Antibodies Using TSKgel HPLC Columns

Jan 25, 05 His Bind Kit (Novagen)

Ab SpinTrap Ab Buffer Kit

MagExtactor -His-tag-

Lifetime stability of size exclusion chromatography columns for protein aggregate analysis

Technical Note Detection of post-immunoprecipitation proteins by Western blot using the Quick Western Kit IRDye 680RD

GST Fusion Protein Purification Kit

TSK-GEL BioAssist Series Ion Exchange Columns

GORE. Protein Capture Devices. Operating Instructions Contents. Product Description Device Characteristics... 2 General Handling...

His-Spin Protein Miniprep

Mobile Phase Optimization in SEC Method Development

phab Amine and Thiol Reactive Dyes

AminTRAP HIS Prepacked Column

FOR RESEARCH USE ONLY. NOT FOR HUMAN OR DIAGNOSTIC USE.

ION EXCHANGE KIT FOR MAB SEPARATIONS

Cdc42 Activation Assay Kit

Size Exclusion Chromatography of Biosimilar and Innovator Insulin Using the Agilent AdvanceBio SEC column

Arf6 Activation Assay Kit

GST Elution Buffer. (Cat. # ) think proteins! think G-Biosciences

Developing Quantitative UPLC Assays with UV

HOOK 6X His Protein Purification (Bacteria)

Glutathione Agarose Resin User s Guide

Protein A Agarose Immunoprecipitation Kit

NISTmAb characterization with a high-performance RP chromatography column

Analysis of Atrazine in Drinking Water at the ppb Level Using New Agilent Reversed Phase LC Columns. Application. Author. Abstract.

Columns for Biomolecules BioLC Column Lines

Rab5 Activation Assay Kit

Human IgG ELISA Kit. Strip well format. Reagents for up to 96 tests

High-Resolution Analysis and Purification of Oligonucleotides with the DNAPac PA-100 Column

The preparation of native chromatin from cultured human cells.

HOOK 6X His Protein Purification (Yeast)

Comparison of different methods for purification analysis of a green fluorescent Strep-tag fusion protein. Application

Characterize mab Charge Variants by Cation-Exchange Chromatography

phab Amine and Thiol Reactive Dyes

RISE Program Workshop in Protein Purification

Impurity Detection with a New Light Emitting Diode Induced Fluorescence Detector Coupled to the Agilent 7100 Capillary Electrophoresis System

Fast and High Resolution Analysis of Intact and Reduced Therapeutic Monoclonal Antibodies (mabs)

RheB Activation Assay Kit

Bacteria Genomic DNA Purification Kit

Glutathione Resin. (Cat. # , , , ) think proteins! think G-Biosciences

RayBio Genomic DNA Magnetic Beads Kit

Validation of a Dual Wavelength Size Exclusion HPLC Method with Improved Sensitivity to Detect Aggregates of a Monoclonal Antibody Biotherapeutic

Rapid GST Inclusion Body Solubilization and Renaturation Kit

Preparative HPLC is still the

3. Close the bottom end of the column and apply the packing device on top. Pump water through the upper adaptor to remove air.

Human IgG Antigen ELISA Kit

AFFINITY GST PURIFICATION

Transcription:

Bio-Monolith Protein G Column - More Options for mab Titer Determination Application Note Biologics and Biosimilars Author Phu T. Duong Agilent Technologies, Inc. Introduction In recent years, monoclonal antibodies (mabs) have become one of the major biopharma products in response to the need to treat various diseases. These antibodies have been engineered with a specific genetic makeup for better targeting of disease agents. During the development of these antibodies, Protein A and G analytical affinity columns are used for determining their titer or concentration from various cell culture supernatants, to select the high-yield clone. An inert polymeric monolith is used as the support for both the Protein A and Protein G columns. Both columns have high affinity for antibodies, and so, they bind only to antibodies in cell-culture supernatants. However, they have different selectivity, as indicated in Table 1. This application note introduces the Agilent Bio-Monolith Protein G column. The column is designed for high speed and high loading capacity. Data are presented to show the linearity with high specificity. Linearity analysis reveals the column s capability for accurate quantitation analysis of mabs in cell-culture supernatant. In addition, lifetime analysis data indicate that the column can be highly reproducible, and has a long lifetime with stable and low backpressure. The Bio-Monolith Protein G column is complementary to the Bio-Monolith Protein A column, to provide more options for titer determination of monoclonal antibodies.

Table 1. Binding affinity of Agilent Bio-Monolith Protein A and G for different IgG subclasses [1,2]. Antibody Protein A Protein G Human Human IgG1 ++++ ++++ Human IgG2 ++++ ++++ Human ++++ Human IgG4 ++++ ++++ Human IgA ++ Human IgD ++ Human IgE ++ Human IgM ++ Mouse Mouse IgG1 + ++ Mouse IgG2a ++++ ++++ Mouse IgG2b ++++ +++ Mouse + +++ Mouse IgM +/ Antibody fragments Human Fab + + Human F(ab )2 + + Human scfv + + Human Fc + + Human K + + Human l + + ++++ = Strong affinity +++ = Moderate affinity ++ = Weak affinity + = Slight affinity = No affinity Materials and Methods Sodium phosphate monobasic monohydrate (Sigma-Aldrich, Corp., p/n S3522) (MW 137.99) Sodium phosphate dibasic dihydrate (Sigma, p/n 71643) (MW 177.99) Citric acid monohydrate (Sigma, p/n C7129) (MW 21.14) Sodium chloride (Sigma, p/n S5886) (MW 58.44) Phosphoric acid, 85 wt % in H 2 O, 99.99% trace metal basis (Aldrich, p/n 345245) Sodium hydroxide solution, 5 to 52% in water, eluent for IC (FLUKA 7264) (MW 4) Glycine, for electrophoresis, 99% (Sigma, p/n G8898) (MW 75.7) Glacial acetic acid (Sigma, p/n A9967) (MW 6.5) Hydrochloric acid, 36.5 to 38.%, BioReagent, for molecular biology (Sigma, p/n H1758) Escherichia coli cell-lysis kit (Sigma p/n CB5). Chinese hamster ovary (CHO)-cell supernatant and lysate, insect-cell supernatant, humanized CHO-cell derived monoclonal antibody (IgG2 and ) from CreativeBio Labs. Method Mobile phase A is the binding and washing buffer, containing 5 mm sodium phosphate buffer, ph 7.4. Two stock solutions (.2 M) are made from 27.6 g sodium phosphate monobasic monohydrate in 1 L deionized water, and 35.6 g sodium phosphate dibasic dihydrate, also in 1 L of deionized water. To make 2 L sodium phosphate containing 5 mm sodium phosphate buffer and 5 mm sodium chloride, ph 7.: 1. Mix 195 ml of sodium phosphate monobasic monohydrate stock solution and 35 ml sodium phosphate dibasic dihydrate stock solution. Agitate using a magnetic stirrer. 2. Add 5.8 g sodium chloride, and continue stirring. 3. Add 1 L deionized water. 4. Measure the ph of the solution, and adjust to ph 7. with NaOH or 3 M phosphoric acid. 5. Pour the solution into a 2 L volumetric flask, and add water to the mark. Mobile phase B is an eluting buffer containing.1 M citric acid, ph 2., made by dissolving 21 g citric acid monohydrate in about 6 ml of water with gentle stirring. Adjust to ph 2. with 1 M HCl, and then dilute solution with water to 1 L in a volumetric flask. CHO-cell supernatant, lysate, insect-cell lysate (all samples were spun), and humanized monoclonal IgG1, IgG2, and were purchased from Creative BioLabs, New York. Escherichia coli lysate was prepared according to the recommended protocol of Sigma-Aldrich, Corp [3]. The kit contains CelLytic B, a bacterial lysis reagent (5 ml), lysozyme solution (1 1 ml), benzonase (25, units), and protease inhibitor cocktail (5 ml). The lysate is prepared by adding 1. g cell paste to 1 ml CelLytic B reagent,.2 ml lysozyme,.1 ml protease inhibitors, and 5 units of benzonase. The mixture is vortexed briefly, and mixed for 1 minutes (by hand or on the shaker) to ensure further extraction of the soluble proteins. The mixture is then centrifuged at 5, g for 1 minutes to pellet any insoluble materials. The soluble protein fraction (supernatant) is carefully removed from the cell debris (pellet at the bottom of the tube). The protein concentration in the supernatant can be estimated using the Bradford protein assay, in this case 4 mg/ml. 2

Some supernatant was then spiked with IgG1, IgG2, or as described. Supernatants were diluted with binding buffer (or buffer A) to a final concentration of 1 mg/ml, and spiked with 2 mg/ml purified humanized single monoclonal antibody such as IgG1, IgG2, or. Conditions Column: Agilent Bio-Monolith Protein G, 5.2 mm diameter, 4.95 mm long (p/n 569-3639) Binding buffer: A, 5 mm sodium phosphate, ph 7.4 Eluting buffer: B,.1 M citric acid, ph 2. Sample: see chromatograms Inj vol: see chromatograms Flow rate: 1. ml/min (or see chromatograms) Gradient: Time (min) % A 1.4 1.5 2. 2.1 1 4.2 1 Temp: 25 C Detector: UV, 28 nm System: Agilent 126 Infinity Bio-inert Quaternary LC Results and Discussion Specificity and selectivity As indicated in Table 1, the Protein G column has higher binding affinity for different humanized monoclonal antibody subclasses than the Protein A column, and only Protein G has affinity for the subclass. Data in Figure 1A demonstrate the specificity of the Bio-Monolith Protein G column and its ability to monitor titer, that is, the presence and concentration of antibodies in the supernatant. The column was injected with the sample containing purified recombinant humanized mab,, which was spiked in the supernatant of CHO cells. This mab was expressed in the CHO cell line. The data show that was the only protein captured and eluted from the column, at about 1.6 minutes at 1. ml/min, whereas all host-cell proteins were not captured by the column, and eluted at the flow-through peak. To show the different selectivity between Protein A and Protein G columns, both were separately injected with monoclonal antibodies IgG1, IgG2, and. Both IgG1 and IgG2 were captured by the two columns (data not shown) but was not captured by Bio-Monolith Protein A. was eluted as a flow-through peak, and was only captured and eluted by the Bio-Monolith Protein G column (compare Figures 1A and 1B). Absorbance (UV 28 nm) Absorbance (UV 28 nm) 4 35 3 25 2 15 1 5 1, 8 6 4 2 A Bio-Monolith Protein A No selectivity for flow-through B.5 1. 1.5 2. 2.5 3. 3.5 4. Bio-Monolith Protein G Selectivity includes Unbound (CHO host-cell proteins) no retained.5 1. 1.5 2. 2.5 3. 3.5 4. Figure 1. A) No binding of on Bio-Monolith Protein A column. was eluted as the flow-through peak. 2 mg/ml humanized mixed with binding buffer (3 µl was injected onto the column). B) The Agilent Bio-Monolith Protein G column quickly captured only from a harvested cell culture spiked with (5 µl of 2. mg/ml mixed with 1 mg/ml CHO-cell supernatant was injected onto the column). A more rugged test was performed to confirm that the specificity of the Bio-Monolith Protein G column was such that it had no binding affinity for host-cell proteins. Host-cell protein samples from E. coli cell lysate, CHO-cell lysate, supernatant, and insect-cell lysate were used. These lysates contained host-cell proteins that were extracted by lysing buffers that contained sodium dodecyl sulfate, a chemical that greatly influences nonspecific binding to the column. These samples did not contain antibody, only host-cell proteins. The test was run because rugged samples give nonspecific results if the column is not well designed. Figure 2 shows no evidence of any protein from any host-cell supernatant being absorbed by the column. All host-cell proteins were eluted as flow-through peaks. The data suggest that the Bio-Monolith Protein G column did not have any affinity for host-cell proteins. 3

Absorbance (UV 28 nm) Absorbance (UV 28 nm) Absorbance (UV 28 nm) Absorbance (UV 28 nm) 1, 8 6 4 2 8 6 4 2 1,4 1,2 1, 8 6 4 2 1,4 1,2 1, 8 6 4 2 A Insect-cell lysate B C D.5 1. 1.5 2. 2.5 3. 3.5 CHO-cell supernatant.5 1. 1.5 2. 2.5 3. 3.5 CHO-cell lysate E. coli lysate.5 1. 1.5 2. 2.5 3. 3.5.5 1. 1.5 2. 2.5 3. 3.5 Figure 2. Specificity analysis using the Agilent Bio-Monolith Protein G column with 1 mg/ml host-cell proteins diluted with binding buffer. The injection volume was 5 µl for all samples. A) insect cell lysate, B) CHO-cell supernatant, C) CHO-cell lysate, D) E. coli lysate. Accurate quantitation Accurate quantitation of mab titer is essential during the early stages of development when the cell line is selected, and also during manufacture when the amount of mab in the cell-culture supernatant determines the optimum harvest time. To demonstrate the ability of the Bio-Monolith Protein G column in the accurate quantitation of mabs, different amounts (µg) of purified IgGs were injected onto the column. Data of peak areas versus amounts of IgGs generated were used to construct the linearity lines, to determine the accuracy of the analysis. Figure 3 shows the linearity of peak areas from the Protein G column, showing that the column can be used for quantitation of mab in harvest cell-culture media with different concentration ranges. The column was injected as low as 2 µg IgG. The signal-to-noise ratio was no higher than 1:1 for 2 µg (data not shown). The maximum loading capacity for this column is approximately 4 to 5 µg IgG (data not shown), which covers the range of concentrations achieved during cell-line selection and production. Peak area (*S) 2, 18, 16, 14, 12, 1, 8, 6, 4, 2, 5 1 15 2 IgG Amount (µg) IgG1 IgG2 R 2 =.9959 R 2 =.996 R 2 =.9989 Figure 3. Agilent Bio-Monolith Protein G column quantitates monoclonal antibodies. The linearity lines include peak area data from 25 to 2 µg IgG. 4

Wider loading range than another vendor s protein G column Figure 4A shows the linearity comparison for a wide loading range of between the Bio-Monolith Protein G column and another vendor s 2.1 3 mm, 4,Å, protein G column. The Agilent column generated a linear loading range from 25 to 2 µg, whereas the other column could only generate linear data between 25 and 1 µg, as recommended by the other vendor s literature. The reason for not having a linear line at a higher loading range was because the column was not able to retain all material at the higher loading range. In fact, at a 2 µg loading range, the other vendor s protein G column had a break-through peak for the mab (some was not retained on the column and eluted as a flow-through peak), as shown in Figure 4B. 2, 18, 16, A R 2 =.9987 Agilent Bio-Monolith Protein G Other vendor protein G Peak area (*S) 14, 12, 1, 8, 6, 4, 2, Maximum loading amount recommended by other vendor Agilent Bio-Monolith Protein G column with linear calibration over wider sample load Absorbance (UV 28 nm) 12 1 8 6 4 5 1 15 2 25 Amount (µg) B Agilent Bio-Monolith Protein G column 2 µg/injection Absorbance (UV 28 nm) 12 1 8 6 4 Other vendor protein G column cannot be absorbed (flow-though) 2 µg/injection 2 2.5 1. 1.5 2. 2.5 3. 3.5 4..5 1. 1.5 2. 2.5 3. 3.5 4. Figure 4. A) Linear loading range comparison between Agilent Bio-Monolith Protein G column and another vendor s protein G column. B) Under a 2 µl load, the other vendor s column shows flow-through that is not observed with the Bio-Monolith Protein G. Thus, the Bio-Monolith Protein G has a higher loading capacity. 5

Rapid separations The Bio-Monolith Protein G is designed for rapid separations. Its capability for fast mab capture and elution at various flow rates is demonstrated with in Figure 5, with 1., 1.5, 2., and 2.5 ml/min flow rates (The column can be operated up to 3. ml/min, data not shown.). Table 2 shows the flow rates and operating gradients. The retention of the peak was shortened when the flow rate increased. However, the relative peak areas between all flow rates were similar. This indicates that the column generated similar recovery at all flow rates (Table 3). Table 2. Flow rates and operating gradients used in an assessment of retention capability of the Agilent Bio-Monolith Protein G column. Time (min) %A %B 1. ml/min 1.4 1.5 1 1.7 1 1.8 1 4.2 1 1.5 ml/min 1.3 1.4 1 1.3 1 1.4 1 3. 1 2. ml/min 1.2 1.3 11.9 1 1. 1 2.1 1 2.5 ml/min 1.1 1.3 1.7 1.8 1 1.7 1 Absorbance (UV 28 nm) Absorbance (UV 28 nm) Absorbance (UV 28 nm) Absorbance (UV 28 nm) 5 4 3 2 1 5 4 3 2 1 5 4 3 2 1 5 4 3 2 1.44.53.69.185.12 1 %B.612.5 1. 1.5 1 %B 2. ml/min.815 %B.5 1. 1.5 1 %B 1.76 2.5 ml/min %B %B.5 1. 1.5 2. 1 %B 1.518 %B 1.5 1. 1.5 2. 2.5 3. 8 6 4 2 Buffer B% 2. %B %B 1 8 6 4 2 Buffer B% 1.5 ml/min 2.5 3. %B 1 8 6 4 2 Buffer B% 1. ml/min 3.5 4. Figure 5. Binding of on the Agilent Bio-Monolith Protein G column evaluated at several flow rates. Injection 5 µl (2. mg/ml mixed with 5 mg/ml CHO-cells host proteins). %B 1 8 6 4 2 Buffer B% Table 3. Agilent Bio-Monolith Protein G column delivers similar relative percentages of flow-through and peaks for all flow rates. Flow rate (ml/min) Total area (mauu*s) Flow-through peak area (*S) Flow-through peak relative area (%) peak area (*S) 2.5 798 521 65.3 277 34.7 2. 1,56 79 67.1 347 32.9 1.5 1,39 932 67.1 458 32.9 1. 2,69 1392 67.3 677 32.7 peak relative area (%) 6

Figure 6 shows the linearity line of the column s backpressure. When the column was tested at.5 ml/min intervals, increasing from 1 to 2.5 ml/min, the backpressure increased linearly. The column has a maximum backpressure of 15 bar. A typical separation flow rate of the Bio-Monolith Protein G column is 1. ml/min. With the instrument flow rate set at 1. ml/min, the column backpressure was ~24 bar. When the flow rate was increased to 2.5 ml/min, the column backpressure increased to ~6 bar. There was minimal effect on IgG binding on the column at maximum backpressure, as indicated earlier. The compatibility of different elution buffers Figure 7 shows compatibility with different eluting buffers for the Bio-Monolith Protein G column. The peak can be eluted by many different acidic eluents. Table 4 shows the strengths and ph of the acidic eluents. These eluents were capable of eluting from the column with similar retention times (similar data were also observed for other IgGs), except with 12 mm HCl. The retention of the peak was longer, compared to the retention time of the peak eluted by other eluting buffers. This eluting buffer, when its concentration was increased to.1 M, gained enough strength to elute with retention similar to other eluting buffers. Pressure (bar) 7 6 5 4 3 2 R 2 =.999 It is also noteworthy that each acidic eluent generated an peak with slightly different peak width and tailing factor. It is suggested that, depending on the IgG, acidic eluents and their strengths will have different influences on retention time, as well as peak width and tailing factor. Therefore, depending on the IgG and expectations of the data, acidic eluents and their concentrations must be determined experimentally. 1.5 1. 1.5 2. 2.5 3. Flow rate (ml/min) Figure 6. Flow rate versus back pressure. The backpressure of the Agilent Bio-Monolith Protein G column increased linearly when the flow rate also increased linearly. Absorbance (UV 28 nm) 175 15 125 1 75 5 25 1,2,3,4, and 5 are peaks 1 2 4 3 5 Table 4. The compatibility and effect of different acidic eluents on IgG. Peak no. Acid PW 5% TF 1.1 M citric acid ph 2..58 1.68 24 2.1 M HCl.53 1.58 24 3 5% acetic acid.71 1.85 24 4.1 M glycine.75 1.82 24 5 12 mm HCl.68 1.69 24 Pressure (bar) at 1 ml/min.5 1. 1.5 2. 2.5 3. 3.5 4. Figure 7. Humanized peak eluted by different acidic eluents from the Agilent Bio-Monolith Protein G column. 7

Recovery of column performance after clean-in-place Figure 8 shows the Bio-Monolith Protein G column can recover its full performance after clean-in-place (CIP). The column was injected with after it was injected for more than 1, injections with spiked in CHO-cell supernatant and lysate. Data show that when the column was dirty, peak width of was broaden and peak height was reduced (compare Figure 8A and 8B, before and after cleanin-place). The data also show that some was not captured by the column and was eluted as a flow-through peak (Figure 8, Panel A). After the column was cleaned, it regained its full performance and was fully captured (see Figure 8B). Figure 9 shows the linearity comparison data of a wide loading dynamic range before and after the column was properly cleaned. peak areas of before and after CIP were very comparable and highly linear. The data indicated that the column was effectively cleaned and compatible with the cleaning protocol described in Table 5. Absorbance (UV at 28 nm) Absorbance (UV at 28 nm) 1 8 6 4 2 1 8 6 4 2 A Before clean B After clean 1.563.5 1. 1.5 2. 2.5 3. 3.5 4. 1.562.5 1. 1.5 2. 2.5 3. 3.5 4. Figure 8. A) Agilent Bio-Monolith Protein G column injected with after more than 1 injections. Column was dirty. B) Column was cleaned and regained its full performance. Table 5. Protocol for clean-in-place for the Agilent Bio-Monolith Protein G column. To prevent contaminants from entering the rest of the column, it should be in a reversed-flow direction for the first step of the protocol at.2 to.5 ml/min. Step Solution Washing column volume (CV) 1.1 M NaOH 1 to 2 2 Deionized water 1 to 2 3.5 M Sodium phosphate buffer, ph 7.4 1 to 2 4 Re-equilibrate column with binding buffer 5 8

In some cases, simple regeneration of the monolithic column is not sufficient. Sample molecules might not fully elute from the column or might precipitate within it. This build-up of contaminants in the column can cause loss of resolution and binding capacity, increased backpressure, or complete blockage. A specific CIP protocol should be designed according to the type of contaminants that are present in the sample. More regeneration suggestions are in the user guide. 2 1.9 1.8 1.7 1.6 1.5 A Peak Area (*S) 2, 18, 16, 14, 12, 1, 8, 6, 4, 2, 5 1 15 2 Amount (µg) Before CIP After CIP R² =.9997 R² =.9959 Figure 9. Comparison of linearity provided by the Agilent Bio-Monolith Protein G column across a wide dynamic loading range before and after clean-in-place. Lifetime and reproducibility Figures 1 and 11 show the results of 1, consecutive injections of and CHO-cell supernatant plus lysate onto the Bio-Monolith Protein G column. The column received 4 injections with CHO-cell supernatant plus lysate, followed by 1 injections of. This sequence was set for 1, injections continuously without stopping for cleaning. The integrity of peak retention time and peak area (Figure 9), as well as peak tailing factor and peak width of (Figure 1), remained nearly unchanged without compromising the column s performance in terms of binding, separation, and eluting capacity. Figure 11 shows that the peak width and tailing factor had minimal impact during the course of the 1,-injection study. Peak area of 1.4 2 4 6 8 1, No. of injections 1,9 1,7 1,5 1,3 1,1 9 7 5 3 B 1 2 4 6 8 1, No. of injections Figure 1. Reproducibility of the Agilent Bio-Monolith Protein G column over 1, injections without clean-in-place. A) Retention time. B) Peak area of. Ten injections were recorded after 4 injections, and over 1, injections were made. Retention time and peak width of remained unchanged (standard deviation = 2.5, n = 1). Tailing factor (5%) 5. 4.5 4. 3.5 3. 2.5 2. 1.5 1..5 Tailing of Peak width 2 4 6 8 1, 1,2 No. of injections Figure 11. The consistency of peak width and tailing factor of the peak over 1, injections on the Agilent Bio-Monolith Protein G column..5.45.4.35.3.25.2.15.1.5 Peak width 9

Conclusions The Agilent Bio-Monolith Protein G column has a high affinity for monoclonal antibody subclasses. It is evident that the column can capture and accurately quantitate across a wide loading linearity range for mab from supernatants. The columns can be used effectively to quantitate the amount of monoclonal antibody with various flow rates, without sacrificing data. The operating backpressure was significantly low, illustrating that the Bio-Monolith Protein G column can be operated with <6 bar HPLC instruments. The flexibility of the column with different acidic eluents enables easy and straightforward experimental design. Agilent Bio-Monolith Protein A and G columns are thus complementary, thus Protein G has affinity for mabs that do not bind to Protein A, and vice versa. These columns provide more options for rapid titer determination for a wider range of mab variants. References 1. Richman, D. D.; Cleveland, P. H.; Oxman, M. N.; Johnson, K. M. The binding of 1. Staphylococci protein A by the sera of different animal species. J. Immunol. 1982, 128, 23-235. 2. Frank, M. B. Antibody Binding to Protein A and Protein G beads; 5. In Molecular Biology Protocols; Frank, M. B., Ed.; Oklahoma Medical Research Foundation, Oklahoma City, USA, 1997. 3. Anon. CelLytic B Plus Kit. Catalog numbers CB5 and CB5. Technical Bulletin. Sigma-Aldrich, Corp. St. Louis, MO, USA. For More Information These data represent typical results. For more information on our products and services, visit our Web site at www.agilent.com/chem. www.agilent.com/chem For Research Use Only. Not for use in diagnostic procedures. Agilent shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material. Information, descriptions, and specifications in this publication are subject to change without notice. Agilent Technologies, Inc., 215, 217 Printed in the USA November 2, 217 5991-694EN