Realization of ITS-90 from K through K: One Company s Approach

Similar documents
REALIZATION OF THE FREEZING POINTS OF INDIUM, TIN AND ZINC USING STAINLESS STEEL CASED CELLS

FIXED POINTS FOR SECONDARY LEVEL CALIBRATIONS

5914A/5915A/5916A 5917A/5918A/5919A Metal Freeze Point Cell. Users Guide

A High-Quality Platinum Resistance Thermometer to 661 C

A Study on the Stability of Standard Platinum Resistance Thermometer in the Temperature Range from 0 C through 720 C

Experimental Study and Computer Modeling of the Triple Point of Argon System

Experimental Study and Computer Modeling of the Triple Point of Argon System

5944/5945/5946/5947 Mini Metal Fixed Point Cells User s Guide

Triple Point of Water Cells

Temperature calibration

USING A MINI TRIPLE POINT OF WATER SYSTEM TO IMPROVE RELIABILITY IN A TEMPERATURE CALIBRATION LABORATORY

Primary Standards for Temperature Calibration

Available online Journal of Scientific and Engineering Research, 2018, 5(1): Research Article

IMPROVEMENTS RELATING TO THE CALIBRATION OF THERMOMETERS

How Does Temperature Non-uniformity of an Annealing Furnace Affect SPRT Stability?

Temperature Calibration Equipment and Services

2002 An Optimal Gallium Cell Interoperability Between A Cell Made In England And Intercompared At NIST, To It s Reference Cell

Tin-filled multi-entrance fixed point

SLIM QUARTZ & METAL CLAD CELLS User Maintenance Manual/Handbook

Reference thermometer Model CTP5000

5900E. Users Guide. Mercury Triple Point Cell

AM1760/1762 Secondary Standard Platinum Resistance Thermometer User s Guide

AM1760/1762 Secondary Standard Platinum Resistance Thermometer User s Guide

Report on NIM-NMC Bilateral Comparison: SPRT Calibration. Comparison from -190 o C to 420 o C

Thermometry laboratory

Thermometry, Chinese TAIPEI, CMS (Center for Measurement Standards)

AM1643 Full Immersion Platinum Resistance Thermometer User s Guide

The results agree with the manufacturer s specifications for the fixed point cells.

Phase Change Cells. T. Shane Topham, Gail Bingham, Harry Latvakoski, Mike Watson

A quality response to your thermal process requirements

HIGH TEMPERATURE STANDARD PLATINUM RESISTANCE THERMOMETER MODEL 96178

Datasheet Platinum Resistance Pt100 Dual Element Industrial Sensor Probe 6mm diameter x 150mm or 250mm probe lengths

AM1660 Precision Industrial Platinum Resistance Thermometer User s Guide

AM1730 Secondary Reference Platinum Resistance Thermometer User s Guide

PRIMARY LABORATORY COMPARISONS

Users Guide. Gallium Melting Point Cell

A TECHNICAL GUIDE AND STANDARD FOR THE USE OF GALLIUM MELT-POINT CELLS

Platinum Resistance Thermometer Pt100 Precision Probe - High accuracy platinum resistance thermometer probe, 6mm diameter x 250mm

4 RESISTANCE THERMOMETRY

5698 Working Standard Platinum Resistance Thermometer User s Guide

AM1880 Standard Platinum Resistance Thermometer User s Guide

WORKING STANDARD PLATINIUM RESISTANCE THERMOMETER MODEL 909 User Maintenance Manual/Handbook

AM1620 Precision Industrial Platinum Resistance Thermometer User s Guide

Specifications: Ice point resistance: Fundamental interval (0 C to 100 C): 38.5Ω (nominal)

Ceramic Ceramic No. of Lead/Wire Lead Allied RS order Dia Ø Length Wires Type Length Code code

ENGLISH Datasheet Platinum Resistance Pt100 & Pt1000 Thin Film Detectors Platinum Sensing Resistors Thin Film (Pt100 & Pt1000 Ohm)

DEVELOPMENT OF DURABLE GOLD VERSUS PLATINUM THERMOCOUPLES

Improving the reliability of temperature measurements up to 1550 C

ENGLISH Datasheet Platinum Resistance Pt100 4 wire class B Resistance Thermometer with DIN B Head

Mineral Insulated Platinum Resistance Thermometer PT100 Sensor - Available in 3mm or 6mm diameter x 150mm or 250mm probe lengths

Platinum Resistance Pt100 & Pt1000 Silicone Patch Sensors Element Encapsulated in A Silicone Rubber Patch with Self-Adhesive Aluminium Foil Backing

üthe Source for Calibration Professionals Solutions for Primary & Secondary Laboratories Temperature Calibration Equipment & Services

Pt100 (100 0 C) to IEC 751, Class B, 4 wire 3.0, 4.0, 4.5 or 6.0mm diameter 316 stainless steel sheath.

ENGLISH. 1metre 7/.02mm Teflon insulated twin twisted lead

Temperature Measuring Instrumentation. M H Joshipura Chemical Engineering Department IT, NU

ENGLISH Datasheet Platinum Resistance Pt1000 Class B Sensor with Teflon insulated lead

Platinum Resistance Pt100 Autoclave Drain Probe, 4 Wire Compatible with most common makes of Autoclaves

150 mm. 1/8 BSPP brass male compression fitting with locknut included

PROGRESS IN HIGH-TEPMERATURE FIXED POINTS USING METAL-CARBON EUTECTICS

Users Guide. Working Standard Platinum Resistance Thermometer

Platinum Ceramic Resistance Elements

THE GALLIUM POINT, AN ALTERNATIVE REFERENCE TEMPERATURE TO THE WATER TRIPLE POINT Tempmeko 2004

THERMOPHYSICAL MEASUREMENTS

Lab #1a Temperature Measurement

realizing ITS-90 temperature calibration equipment & services

ENGLISH Datasheet Platinum Resistance Pt100 Hygienic Thermometer, 1.5'' RJT style fitting

realizing ITS-90 temperature calibration equipment & services

Realization of the Temperature Scale in the Range from K (Hg Triple Point) to C(Cu Freezing Point) in Croatia

CHARACTERISTICS OF SELF-POWERED NEUTRON DETECTORS USED IN POWER REACTORS

Guide to the Realization of the ITS-90

ENGLISH. Datasheet. RS single element: Single 4 wire element, allowing connection to any Pt100 2, 3 or 4 wire instrument

Proven Capabilities. Features and Benefits. Watlow provides:

Thermometry, Spain, CEM (Centro Español de Metrologia), INTA (Instituto Nacional de Técnica Aeroespacial)

SnAg3.5 (CLF5013) Halide Free No Clean Core Wire

SURFACE TEMPERATURE SENSORS

Guide to the Realization of the ITS-90

Apparatus for Vacuum Heating of TEM Samples

Proven Capabilities. Features and Benefits. Watlow provides:

5. How to Select and Use the Right Temperature Sensor

Temperature Calibration Technology for Laboratory and Industry

SnAg3.0 (CLF5013) Halide Free No Clean Core Wire

TEMPSENS I N S T R U M E N T S. w w w. t e m p s e n s. c o m. -ve Temperature Bath. Liquid Bath. Dry Block Furnaces. Black Body Sources

üthe Source for Calibration Professionals Solutions for Primary & Secondary Laboratories Temperature Calibration Equipment & Services

In most industrial applications, temperatures are typically

MEDUSA R MODEL 999 User Maintenance Manual/Handbook

NANMAC. Quality Performance Solutions

New Type of Precision Thermometer Offers Unrivalled Accuracy and Stability

122 PROCEEDINGS OF THE OKLAHOMA

ELECTRICAL RESISTIVITY AS A FUNCTION OF TEMPERATURE

Construction of a Low Temperature Heat Exchanger Cell

3.3 Air temperature. Types of instruments

Automotive air or oil temperature Appliances cooking temperature. MOUNTING DIMENSIONS (for reference only)

MEHRAN UNIVERSITY OF ENGINEERING AND TECHNOLOGY, JAMSHORO Sensors & Actuator for Automatic Systems LAB # 01 RESISTIVE TEMPERATURE TRANSDUCERS

User Manual. Version 1.0 THERMISTOR CHARACTERISTICS TRAINER. Model No : (ITB - 06ACE)

Test of Abnormal Heat in Hydrogen Loaded Metal

ISA Seminars on the Web Live Experts on Hot Topics

Innovative Temperature Measurement in Heat Treatment Processes

Troubleshooting RTDs and Thermocouples

Meteorology 432. Thermometry Spring 2013

BUREAU OF INDIAN STANDARDS Draft Indian Standard

Transcription:

Realization of ITS-90 from 273.15 K through 1234.93 K: One Company s Approach Xumo Li, Steve Iman, Mike Hirst, Mingjian Zhao Hart Scientific, Inc. Abstract All of the fixed points of the International Temperature Scale of 1990 (ITS-90) in the range from 273.15 K to 1234.93 K have been realized in specially-designed, permanently-sealed cells. The purity of all of the metals used was 99.9999% or greater. Many improvements in design and techniques used were made in order to get the highest possible accuracy. Since it is important that all of the fixed points be traceable to NIST (National Institute of Standards and Technology), these fixed points were compared with NIST data by using two NIST-calibrated standard platinum resistance thermometers (SPRTs). The differences between the fixed point values and those measured at NIST were within 0.5 mk for the freezing points (FP) of tin and zinc, and within 1 mk for the FP of aluminum. The expanded uncertainties (coverage factor K=2) at the fixed points were 0.4 mk for the melting point (MP) of gallium, 1.0 mk for the FP of indium, tin, and of zinc, 4.0 mk for the FP of aluminum and 10 mk for the FP of silver. Two SPRTs were used to interpolate between the fixed points. A 25-ohm SPRT was used up to 933.473 K and a 0.25-ohm SPRT up to 1234.93 K. Introduction The International Temperature Scale of 1990 (ITS-90) became the official international temperature scale on January 1, 1990, superseding the International Practical Temperature Scale of 1968 (IPTS-68). During the last couple of years the efforts of our laboratory have been concentrated on establishing the new international temperature scale between 273.15 K and 1234.93 K. In order to realize all of the defining fixed points through this temperature range it was necessary to develop and manufacture high-quality furnaces and fixed-point cells including the triple point of water, the melting point of gallium, and the freezing points of indium, tin, zinc, aluminum, and silver. These fixed-point cells are of the sealed cell design to aid in the ease of use and convenience of intercomparison. In addition, two highly-stable SPRTs of the authors design were calibrated by NIST and were used as the standard interpolation instruments in realizing the ITS-90 temperature scale.

Standard Platinum Resistance Thermometer According to the ITS-90, the temperature between 13.8033 K and 961.78 C is defined by means of a platinum resistance thermometer calibrated at specified sets of defining fixed points. The uncertainties of the realization and dissemination of the ITS-90, therefore, depend closely on the stability of an SPRT. Special attention was paid in developing, testing and improving SPRTs. Three types of SPRTs have been developed and tested. The main specifications are shown in Table 1. The detailed test data are listed in Table 2 and Table 3. A 25-ohm SPRT (S/N 92017) and a 0.25-ohm SPRT (S/N 92256) were calibrated by NIST and used throughout these tests to compare our fixed-point data with NIST. The data given in the NIST Reports of Calibration for these SPRTs are listed in Table 4. The resistance ratios W(t) at the exact Table 1: Main specifications of three types of SPRTs. Model 5681 5684 5685 Temperature Range 189 C to 661 C 0 C to 1070 C 0 C to 1070 C Nominal Resistance at T. P. Water 25Ω 0.25Ω 2.5Ω Sensor Support Quartz cross Quartz strip with notches Diameter of Sensor Pt wire Protecting Sheath Quartz cross 0.07 mm 0.4 mm 0.2 mm Quartz Diameter: 7 mm Length: 520 mm Quartz Diameter: 7 mm Length: 680 mm Quartz Diameter: 7 mm Length: 680 mm Typical stability of R tp 1 mk/100h at 661 C 1 mk/100h at 1085 C 1 mk/100h at 1085 C The official maximum temperature of an SPRT as a defining interpolation instrument of the ITS-90 is 961.78 C, but these types of SPRTs were found to be stable up to at least 1070 C. The annealing temperature during the stability test was 1085 C. The lower temperature limit of these types of SPRTs can be as low as 189 C. In general, it is suggested that a 25-ohm SPRT be used in the range below 0 C. Table 2: Results of stability tests with 2.5-ohm SPRTs. R tp (ohms) Total Annealing time S/N 2001 S/N 2002 S/N 2004 S/N 2006 at 1085 C (hours) 2.475910 2.463342 2.469168 2.458491 25 2.475914 2.463354 2.469171 2.458499 50 2.475905 2.463357 2.469168 2.458504 75 Notes 2.475910 2.463355 2.469160 2.458514 After 1 thermal cycle 2.475917-2.469162 2.458523 After 1 thermal cycle Thermal cycle procedure: Put tested SPRT into a furnace at 962 C. Take the SPRT out of the furnace after twenty minutes and cool it in the air. Insert the SPRT into the furnace again, and after thirty minutes decrease the temperature of the furnace from 962 C to 480 C at a rate of 100 C per hour. Finally take the SPRT out of the furnace, cool it in the air and measure its R tp. temperatures of the freezing points of indium, tin, zinc, aluminum and silver were calculated by us

Table 3: Results of stability tests with 25-ohm SPRTs. R tp (ohms) Total Annealing time S/N 1008 S/N 1009 S/N 1010 S/N 1013 at 660 C (hours) 25.57653 25.70259 25.57474 25.53706 25 25.57641 25.70255 25.57452 25.53702 50 25.57639 25.70248 25.57450 25.53696 75 25.57644 25.70247 25.57449-100 Notes 25.57641 25.70256 25.57443 25.53962 After one thermal cycle 25.57641 25.70258 25.57443 25.53942 After one thermal cycle Thermal cycle procedure: Put a tested SPRT into a furnace at 660 C. Take the SPRT out of the furnace after twenty minutes and cool it in the air. Insert the SPRT into the furnace again, and after thirty minutes decrease the temperature of the furnace from 660 C to 480 C at a rate of 100 C per hour. Finally take the SPRT out of the furnace, cool it in the air and measure its R tp. according to the coefficients given in the NIST Reports. These calculated values are listed in Table 4 as well. Defining Fixed Points According to the ITS-90, the calibration of an SPRT from 273.16 K to 1234.93 K requires the use of defined fixed points. Actually, the fixed points are the most important part of the ITS-90. All of the Table 4: The NIST calibration data of two SPRTs used for comparison. SPRT S/N: 92017 92256 NIST Test No: 253699 255156 Date of Report 3 May 1994 13 April 1995 For zero-power For 1.0 ma For zero-power For 14.14 ma a 7 or a 6 1.5526301E-04 1.5988934E-04 3.6806657E-05 3.2513557E-05 b 7 or b 6 1.5007187E-05 1.7190604E-05 7.1081548E-05 6.1878236E-06 c 7 or c 6 2.2808655E-06 2.8945480E-06 2.1652828E-06 1.8774523E-06 d 5.1265611E-06 2.6454136E-06 W(29.7646 C) 1.11812035 1.11811977 W(156.5985 C) 1.60970212 1.60969863 W(231.928 C) 1.89264875 1.89264332 1.89283467 1.89283031 W(419.527 C) 2.56864562 2.56863636 2.56898418 2.56897629 W(660.323 C) 3.37558565 3.37557056 3.37610714 3.37609560 W(961.78 C) 4.28654566 4.28652538 defining fixed points in this range were realized in specially-designed, permanently-sealed cells manufactured at Hart Scientific.

Fixed-Point Cell The construction of all metal fixed-point cells is basically the same. Some of the materials used in construction are different, however. The particular metals used for each of the different type of cells are all 99.9999% pure or better. The gallium cell differs slightly in that a PTFE (Teflon) crucible is used instead of graphite. All cells are constructed using a quartz shell and reentrant well. Each cell is connected to a vacuum system after assembling. The cell is then evacuated at a selected temperature for more than 100 hours and during that period the cell is repeatedly purged with 99.999% pure argon. Once it has been determined that the cell is free from contaminants, it is filled with pure argon and sealed at a known atmospheric pressure exactly at the temperature of the fixed point. The sealed cell solves most problems associated with open cell designs. Information on each cell is outlined in Table 5. The actual temperature of the sensor of an SPRT calibrated at a fixed point might be different from the value assigned to the fixed point by the ITS-90 because of small pressure differences. Realization of the Fixed Points Realization of a fixed point to a high level of accuracy requires a strict envelopment, the temperatures of which must be very uniform, stable and controlled. The equipment used to realize these fixed points is listed in Table 6. The three-zone furnace used for the freezing points of indium, tin, zinc and aluminum is shown in Figure 1 and the heat-pipe furnace for the freezing point of silver and aluminum is shown in Figure 2. A typical axial temperature distribution in the furnace at the freezing point of zinc is shown in Figure 3. Table 5: Fixed point cell Thermometric Fixed Point Purity of Material, % Quantity, kg H 2 O TP Immersion Depth, cm 0.76 23 Ga MP 99.99999 0.45 15 In FP 99.9999+ 0.97 18 Sn FP 99.9999+ 0.96 18 Zn FP 99.9999+ 0.95 18 Al FP 99.9999+ 0.35 18 Ag FP 99.9999+ 1.35 18 Water from a purification system, with a resistivity of about 18 megohm-cm, was distilled into the cell. Immersion depth is calculated with the SPRT in its fully-immersed position in the cell. The distance from the midpoint of the sensor of the SPRT to the tip of the SPRT quartz sheath is about 1.5 cm. The actual distance from the upper surface of the metal to the bottom of the central well is 19.5 cm for the metal freezing points, and 16.5 cm for the melting point of gallium. All resistance measurements were made using a Guildline DC automatic bridge, model 6675, S/N 60956. The resistance of the 25-ohm SPRT was measured using DC currents of 1 ma and 1.41 ma. The resistance of the 0.25-ohm SPRT was measured using DC currents of 14.14 ma and 20.0 ma. These currents allowed the measurement results to be extrapolated to values at zero power. The re-

Table 6: Equipment used for fixed points. Fixed point uniformity Equipment used Temperature uniformity Triple point of water Model 7012 bath ±0.001 C Melting point of gallium Model 7011 bath ±0.001 C Freezing point of indium Model 9114 furnace, three zones ±0.02 C Freezing point of tin Model 9114 furnace, three zones ±0.02 C Freezing point of zinc Model 9114 furnace, three zones ±0.02 C Freezing point of aluminum Model 9114 furnace, three zones ±0.03 C Freezing point of silver Model 9115 furnace, heat pipe ±0.05 C sistance at the triple point of water was measured immediately after each measurement at the melting or freezing points described in this paper. The triple point of water is the most important fixed point of the ITS-90. The stability of all SPRTs should be checked at the triple point of water periodically. According to the ITS-90, the temperatures are interpolated in terms of the ratio of the resistance at a temperature t (R t ) and the resistance at the triple point of water (R tp ) in the range from 13.8033 K to 961.78 C [1]. The measurement at the triple point of water, therefore, always follows a calibration at any other fixed point. In practical work, using resistance ratios instead of absolute resistance insures the greatest possible accuracy[2]. The triple point of water cell is comprised of a borosilicate glass shell, high purity water, ice and water vapor. Once frozen, the cell is maintained for approximately 3 months with regular use in a model 7012 constant-temperature bath. The melting point of gallium (302.9146 K) is an important calibration point for biological, environmental, oceanographic, geological and energy research. The gallium cell as mentioned earlier has a Teflon crucible and an outer glass shell. The Teflon crucible allows for the 3.1% expansion of the gallium upon solidification. Two cells were used in the realization of the gallium point serial numbers 01 and 02. The cells were first placed in a cooler below the melting point to insure that the gallium was completely solidified. They were then placed in a specially-designed fixture suited for a model 7011 constant-temperature bath. The temperature was raised to a temperature 1 C higher than the melting point for about 20 minutes to partially melt the gallium to form a liquid-solid interface. The cells were then maintained in the bath at a temperature reduced to 0.2 C higher than the melting point. This procedure usually resulted in the melting curve lasting for seven days or longer. A typical melting curve is shown in Figure 4. The melting point of gallium realized in this way was compared with NIST data through the NIST-calibrated SPRT S/N 92017. The comparison results are listed in Table 7. The temperature differences from NIST s were within 0.3 mk for both cells, and the difference between the two cells was within 0.1 mk. All of the freezing points except tin were realized in a similar fashion. First, the temperature of the furnace was raised about 10 C higher than the freezing point. After the metal was completely melted, the furnace was set at a stable temperature 1 to 1.5 C higher than the freezing point and left overnight. The next morning, the furnace temperature was decreased slowly (0.1 to 0.15 C per minute). To monitor the metal sample temperature, a SPRT was inserted into the cell. The temperature of the metal sample decreased to a point lower than the freezing point before recalescence. The amount of this supercooling is different from metal to metal. Table 8 summarizes the amounts of

A/M 9114 Furnace Construction SPRT SPRT Equilibration Block Thermal Guard Assembly PRT Sensor Main Controller SET DOWN UP EXIT Top Zone Heater HART SCIENTIFIC Top Zone Controller R M OP1 OP2 PAR TOP ZONE Main Heater Differential TCs Freeze Point Cell PRIMARY ZONE Cell Support Container Differential TCs Thermal Block (3 zone subdivision) Bottom Zone Controller R M OP1 A/M PAR OP2 BOTTOM ZONE Bottom Zone Heater Water Cooling Coils Super Cooling Gas Supply (Argon) Figure 1 The three-zone fixed point furnace. HART SCIENTIFIC

9115 Furnace Construction Top Support Block Retaining Plate SPRT Thermal Guard Assembly Top Cover Cutout Thermocouple Cooling Coils Thermal Shunt Disk Heating Element Control and Cutout Thermocouple Metal Freeze Point Cell Basket and Cover Metal Freeze Point Cell Heat Pipe Ceramic Fiber Insulation Air Gap (Circulating Air) Bottom Support Block HART SCIENTIFIC Figure 2 The heat-pipe fixed point furnace.

427.5 Temperature ( C) 427.4 427.3 427.2 427.1 427 426.9 426.8 supercool observed in the laboratory. Immediately after recalescence the thermometer was removed from the furnace and a cold quartz tube was inserted into the fixed-point cell for one minute to initiate the freeze. At this point, the SPRT to be calibrated was introduced into the cell. Meanwhile, the furnace was kept at a stable temperature of 1 C below the freezing point. This procedure provides a very stable, long freezing plateau that typically lasts for more than ten hours. The changes in tem- Ohms 28.488 28.486 28.484 28.482 28.48 28.478 28.476 28.474 9:00 24-Jun 16:00 24-Jun 23:00 24-Jun 6:00 25-Jun 13:00 25-Jun 20:00 25-Jun 3:00 26-Jun 10:00 26-Jun 17:00 26-Jun 0:00:00 26-Jun 7:00 27-Jun 14:00 27-Jun 21:00 27-Jun 4:00 28-Jun 11:00 28-Jun 18:00 28-Jun 1:00 29-Jun 8:00 29-Jun 15:00 29-Jun 22:00 29-Jun 5:00 30-Jun 12:00 30-Jun 19:00 30-Jun 2:00 1-Jul 9:00 1-Jul 16:00 1-Jul 0 25 50 75 100 125 150 175 200 Height (mm) Figure 4 Axial temperature distribution in the furnace near the freezing point of zinc. Figure 3 A typical melting curve of gallium.

Table 7: The comparison of the melting point of gallium (SPRT S/N: 92017) Calibration Difference Cell S/N Results, W Ga NIST Data W (10 6 ) t (mk) Ga01 1.11812140 1.11812035 1.05 0.26 Ga02 1.11812132 0.97 0.25 Table 8: Supercool ranges for different freezing points observed in the laboratory. Fixed point Range of supercool ( C) FP of indium 0.05-0.15 FP of tin 1.5-5.0 FP of zinc 0.01-0.03 FP of aluminum 0.5-1.0 FP of silver 1.0-1.8 perature during the first half of the plateau were usually within ±0.2 to 0.3 mk. A typical freezing curve is shown in Figure 5. The procedure for the freezing point of tin is similar, with the following difference: when the temperature indicated by a thermometer immersed in the tin sample reached the freezing point, a cold gas current was forced upward around the outer surface of the cell until recalescence. Then the cold gas current was shut off. After recalescence the furnace was kept at a stable temperature of 1 C below the freezing point as with the other metals. The freezing points realized in the laboratory were compared with NIST data through two NIST calibrated SPRTs S/N 92017 and S/N 92256. The comparison results are summarized in Table 9. Uncertainty The uncertainty of these measurements is expressed here in accordance with the new NIST policy[3]. The components of standard uncertainty are listed in Table 10. The reproducibility, an uncertainty of category A, is calculated from repeated calibration 419.54 419.535 Temperature 419.53 419.525 419.52 419.515 419.51 5:35 6:00 6:25 6:50 7:15 7:40 8:05 8:30 8:55 9:20 9:45 10:10 10:35 11:00 11:25 11:50 12:15 12:40 1:05 1:30 1:55 2:20 2:45 3:10 3:35 4:00 4:25 4:50 5:15 5:40 6:10 6:35 7:00 7:25 7:50 8:15 8:40 Time Figure 5 A typical freezing curve of zinc.

data of a SPRT at the fixed point. Another important source of uncertainty stemmed from impurities in the sample. The nominal purity of all pure metal samples used in the work was 99.9999+%. The expanded uncertainties (U) are obtained by multiplying the combined standard uncertainties (u c ) by a coverage factor (k). We used the same value k=2asnist s new policy suggests. Conclusions Table 9: Comparison of freezing points realized in the laboratory with NIST data. Cell S/N W(t) calibrated at NIST data Difference the laboratory W (10 6 ) t (mk) In01 1.609697 01 1.60970212 5.11 1.34 In02 1.6096983 0 3.82 1.00 Sn01 1.89264802 1.89264875 0.73 0.20 Sn02 1.89264705 1.70 0.46 Zn01 2.56864470 2.56864562 0.92 0.26 Zn02 2.56864617 0.55 0.16 Al01 3.37558788 3.37558565 2.23 0.70 Al02 3.37558877 3.12 0.97 Al03 3.37558708 1.43 0.45 Ag01 4.28655863 4.28654566 12.97 4.57 Ag02 4.28655061 4.95 1.74 NIST data was taken from Table 4. The NIST data for the freezing point of silver was that of SPRT S/N 92256 and others were of S/N 92017. The uncertainty for the freezing point of silver is better than 10 mk, for the freezing point of aluminum better than 4 mk, and for all of the other fixed points in the range, better than 1 mk. The comparison of the data obtained was well within the estimated expanded uncertainties. Therefore, this laboratory has established the ITS-90 in the range from 273.15 K to 1234.93 K, with traceability to NIST. References 1. B.W. Mangum and G.T. Furukawa, Guidelines for realizing the International Temperature Scale of 1990", NIST Technical Note 1265, (1990). 2. Xumo Li, Producing the Highest Level of Accuracy from Standard Platinum Resistance Thermometers, Cal Lab, (July-August, 1995).

Table 10: Estimated uncertainties in the fixed points. Source of uncertainty Category A or B Uncertainty (mk) Ga In Sn Zn Al Ag Reproducibility A 0.1 0.2 0.2 0.24 1.0 2.0 Impurity of sample B 0.15 0.4 0.4 0.4 1.5 4.0 Non-linearity of the bridge B 0.04 0.04 0.04 0.06 0.08 0.12 Stemmed from the t. p. water B 0.04 0.07 0.08 0.11 0.16 0.23 Stemmed from the hydrostatic correction Stemmed from the pressure correction B 0.01 0.03 0.02 0.03 0.02 0.05 B 0.02 0.05 0.03 0.04 0.07 0.06 Combined uncertainty 0.19 0.46 0.46 0.49 1.81 4.48 Expanded uncertainty 0.38 0.92 0.92 0.98 3.62 8.96 3. Taylor, Barry N. and Kuyatt, Chris E., Guidelines for Evaluating and Expressing the Uncertainty of NIST Measurement Results, NIST Technical Note 1297, 1994 Edition.