After reviewing this tutorial, the user should be able to Topics to be covered

Similar documents
PRINCIPLES OF OPERATION

REFERENCES FOR CURRENT NRC REGULATIONS: PARTS 20 AND 35.

QUALITY CONTROL AND ASSURANCE OF Tc-99m GENERATORS AND KITS APPLIED IN SAEC LABORATORIES

STUDY OF THE ELUTION OF 99 Mo- 99m Tc GENERATORS PRODUCED AT IPEN-CNEN/SP

Simplified Method for Determining the Radiochemical Purity of 99m Tc-MAG3

Determination of 99 Mo contamination in 99m Tc elute obtained from 99 Mo / 99m Tc- generator

USP Chapter 823 USP 32 (old) vs. USP 35 (new)

Complies with the monograph for Parenteral Preparations and with that for Radiopharmaceuticals.

Nuclear Pharmacy. Jim Velez Director, Nuclear Pharmacy (206) (206) Pager

THALLOUS ( 201 Tl) CHLORIDE INJECTION: Revised Final text for addition to The International Pharmacopoeia (January September 2009)

THE INTERNATIONAL PHARMACOPOEIA

2014 Meeting of Mid-Atlantic States Radiation Control Programs

Regulatory aspects: Radiopharmacy (GMP)

Q & A on USP General Chapter <823> Revision Proposal Webinar held on 21 February 2011

PHYSICO-CHEMICAL, BIOLOGICAL OR MICROBIOLOGICAL TESTS OF MEDICINAL PRODUCTS

Validation of Analytical Methods used for the Characterization, Physicochemical and Functional Analysis and of Biopharmaceuticals.

Characterization and Quality Control Analysis of 99m Tc-Bicisate

The instant thin-layer chromatography (ITLC) method of

OFFICE OF ENVIRONMENTAL HEALTH AND RADIATION SAFETY (EHRS)

Cold kit with higher 99m Tc activities - A Chromatographic Study. Introduction. Correspondence author: RADIOPHAMATHTICAL, ORIGINAL ARTICLE

Different Models for Cost-Effective Fractionation of Bone Radiopharmaceuticals; Methylene Diphosphonate (MDP) & Hydroxymethylene Diphosphonate (HDP)

Technescan MAG3. NAME OF THE MEDICINE Technescan MAG3 Kit for the Preparation of Technetium ( 99m Tc) Mertiatide.

Radioisotopes production technology -EGYPT experience

Lead cation exchange and alumina columns for concentration of 99m Tc-pertechnetate

Ge / 68 Ga GENERATOR* HOW TO DO PET AS SIMPLY AS SPECT? EXCELLENCE DEDICATED TO NUCLEAR MEDICINE, HEALTHCARE & ENVIRONMENT

A new two-strip TLC method for the quality control of technetium-99m mercaptoacetyltriglycine

Recovery Profiles of Molybdate Solutions Through Varying Dowex TM -1 Resin Amounts, and Translation to Mo99/Tc99m Separation

CQIE PET PROCEDURES. American College of Radiology Clinical Research Center. Centers for Quantitative Imaging Excellence LEARNING MODULE

RADIATION SAFETY AND PROTECTION

Proposed Revision of USP General Chapter Radiopharmaceuticals for Positron Emission Tomography Compounding <823>

VALIDATION OF ANALYTICAL PROCEDURES: METHODOLOGY *)

REVISED DRAFT FOR COMMENT

SUMMARY OF PRODUCT CHARACTERISTICS (PHYTATE) mg. Quantity per vial Function. Tc-Technephyte GBq information

Estimation of the Radiation Effective Dose for Workers in 99m Tc Generator Production

US Regulatory Requirements: Clinical & Clinical Research Production, SPECT & PET Radiopharmaceuticals (RPs)

Radiopharmaceuticals overview of requirements

Sensitivity Motivation. Flexible Geometry High Sensitivity SPECT System for Small Animals and Plants

Ultra-Technekow FM Mo-99/Tc-99m generator. Product Information

in94070.txt at Page 1 of 5 2/24/00

OFFICE OF ENVIRONMENTAL HEALTH AND RADIATION SAFETY (EHRS)

Draft Guidelines for Radiopharmacy

Functionality. Normal Operation

S U M M AR Y O F P R O D U C T C H AR AC T E R I S T I C S

University of Maryland Baltimore. Radiation Safety Procedure

GOOD LABORATORY PRACTICE Agilent Cary 8454 UV-Visible Spectroscopy System

PosiRx. automated radiopharmaceutical system

TBIA101 LABELED WITH 68

Development of 99m Tc production from (n,γ) 99 Mo

Radiation Safety Manual

SUPPLEMENTAL ALPHABETICAL LISTING OF ISOTOPES ATOMLAB DOSE CALIBRATORS. FOR CALIBRATORS WITH 6.0 V and HIGHER SOFTWARE updated 3/8/06

METHOD DETERMINATION OF TURBIDITY BY NEPHELOMETRY. Edited by James W. O'Dell Inorganic Chemistry Branch Chemistry Research Division

Latest USP Initiatives: Monographs, General Chapters, and Compounding

IRON (Colorimetric) 2. Muffle Furnace: Equipped with pyrometer and capable of operating at controlled temperatures up to 600 C

MATERIAL SAFETY DATA SHEET DRAXIMAGE DTPA MATERIAL SAFETY DATA SHEET. Kit for the Preparation of Technetium Tc99m Pentetate Injection

Chapter 3 Obtaining an Authorization to Possess and Use Radioactive Material

procedu Education And Continuing - Page 1 of 27 -

Radiopharmacies, Radioisotopes and Accelerators Team BEST developments

Automating EPA 6020 Compliant Analysis with the Agilent 7900 ICP-MS and ESI prepfast Autodilution System

Quantification of genotoxic "Impurity D" in Atenolol by LC/ESI/MS/MS with Agilent 1200 Series RRLC and 6410B Triple Quadrupole LC/MS

Regional or Global Molybdenum-99 Supply Indian Perspective

Instructions and Checklist for the Preparation of Data for the Pediatric Administered Activity Survey

Impurity Control in the European Pharmacopoeia

METHOD 7550 OSMIUM (ATOMIC ABSORPTION, DIRECT ASPIRATION)

RADIOACTIVE MATERIALS SAFETY REFRESHER EXAM

DNA Damage Quantification Kit

Analytical Method of Limit Test for Hexachlorobenzene. in Picloram TC and Method Validation Data*

for IND and RDRC Regulated PET Compounding

DNA Damage Quantification Kit

Chapter 7 Transportation of Radioactive Material

á1225ñ VALIDATION OF COMPENDIAL PROCEDURES

RADIOPHARMACY GMP QC HPLC VALIDATION

Appendix-2 Country Report on Isotope Production and Its Applications

Standard Operation Procedure (SOP) for the Extraction and Instrument Analysis of Siloxanes from Cartridges

Dose-on-Demand Biomarker Generator System for Molecular Imaging

Plant DNA/RNA Extraction Kit. User Manual

A Simple Rapid and Sensitive Method Development for Quantification of Quetiapine Fumarate in Bulk and Dosage Forms Using RP-HPLC

HEALTH PHYSICS Emergency Procedures and Levels of Action

MOLECULAR TESTING: VERIFYING/VALIDATING INSTRUMENTS, REAGENTS AND ASSAYS. Richard L. Hodinka, Ph.D.

THE UNIVERSITY OF CHICAGO APPLICATION FOR NON-HUMAN USE OF RADIOACTIVE MATERIAL

Cadmium Testing and On Site Calibration for Water Testing Detection Range: ppm

EPA APPROVED METHOD FOR TURBIDITY ANALYSIS 360 x 90, Turbidity Method 10258

CH241 Experiment #1 (Weeks of September 11, 18, and 25, 2017)

VICH Topic GL2 (Validation: Methodology) GUIDELINE ON VALIDATION OF ANALYTICAL PROCEDURES: METHODOLOGY

APPENDIX III SAMPLE LAB REPORT. Experiment 1. High Performance Liquid Chromatography. Alfred E. Neuman

Product Code: PA-E01 Directions For Use Introduction Intended Use

GAIA PEPTIDE LABELLING FLUIDIC PROCESSOR

Radiolabeling and Quality Control of 99mTc-rituximab Standard SLN Biopsy Criteria

prepfast Autodilution Autocalibration Auto QC Dilution Elemental Scientific

Radionuclide Release at Fukushima

Revised Regulations for Medical Use of Byproduct Material Review of U.S. NRC s revised 10 CFR Part 35

Imaging Requirements (RTOG 1106/ACRIN 6697) Adam Opanowski, CNMT, PET, NCT, R.T.(N)(ARRT) Imaging Analyst ACR Core Laboratory

Latest Developments in USP Monographs and the Compounding of Sterile Radiopharmaceuticals. Jim Ponto, MS, RPh, BCNP

Atomlab. Dose Calibrators Wipe Test Counter IODEX. The Clinical Advantage. fast, accurate measurements. Bring science,

Overview of Generic PET Applications. Sema Basaran, Ph.D. The 34 th Annual High Country Nuclear Medicine Conference.

Technology Note. Monitoring TOC in ultra-pure laboratory water. Control of impurity levels. Figure 1 Control of impurities

The use of personal protective equipment in nuclear medicine

InterRad - Interdisciplinary Radiation Research

ACCURACY THE FIRST TIME. Reliable testing with performance, productivity and value combined AB SCIEX 3200MD SERIES MASS SPECTROMETERS

Size Exclusion Chromatography of Biosimilar and Innovator Insulin Using the Agilent AdvanceBio SEC column

GMP and Quality Control: An Industry Perspective. Steven S. Zigler, Ph.D. PETNET Pharmaceuticals, Inc. Knoxville, Tennessee

Transcription:

After reviewing this tutorial, the user should be able to list the various types of impurities found in the eluate collected from the Mo- 99/Tc-99m generator and give an example of each identify the various classes of impurities found in radiopharmaceuticals and give an example of each describe in detail how QC testing is performed on the eluate and on Tc-99m radiopharmaceuticals state the specifications for each of these tests and what action is required if there is a test failure list the 4 QC tests performed on a dose calibrator and required frequency of performance describe how each test is performed, state the legal specification, and describe the action required if there is a test failure state the difference between agreement states and non-agreement states know the basics of licensure by the NRC and State regulatory agencies list the various requirements for maintaining log books in the lab and the required frequency of entry be familiar with basic rules of radiation safety, e.g., to describe what precautions and action must be taken when there is a spill or contamination is found on a technologist's hands define "reportable event" according to the new rules, and to state what one must do to document and report the occurrence. Topics to be covered Rationale for Performing Quality Control: Criterion Description of a Non-Diagnostic Study and how to avoid it Impurities in Generators and Radiopharmaceuticals: Classification and Detection QC Testing of a Mo/Tc Generator QC Testing in Radiopharmaceuticals Quality Control of Dose Calibrators

Classification of Impurities in the Hot Lab TYPE EXAMPLE EFFECT Radionuclidic Mo-99 High Radiation dose; Poor Image Quality Radiochemical HR Tc Poor Image Quality, Altered Radiation Dose Chemical Al 3+ Altered Radiation Dose Image Quality Method of Quantification of Impurities in the Hot Lab TYPE EXAMPLE METHOD Radionuclidic Mo-99 Dose Calibrator or Multi-Channel Analyzer Radiochemical HR Tc Thin Layer Chromatography Chemical Al 3+ Colorimetric or other methods Required Quality Control Testing of a Mo/Tc Generator TEST FREQUENCY SPECIFICATIONS Mo Breakthrough First Elution <0.15 µci Mo/mCi Tc at t administration Al 3+ Breakthrough Not Required <10 ppm of Al 3+ ; may be expressed as µg/ml Hydrolyzed Reduced Tc Optional, but recommended < 5% recommended

Performance of the Mo-99 Breakthrough Test Mo-99 is assayed directly in the special lead pig supplied by manufacturer of your dose calibrator. Tc-99m is then assayed directly in the plastic sleeve. Activity (uci) of Mo-99 is divided by activity (mci) of Tc-99m to obtain a ratio. If this ratio is <0.15 uci Mo-99 per mci of Tc-99m at time of administration, the generator eluate has passed the Mo-99 Breakthrough Test. As a rule of thumb, if the ratio is <0.038 at time of elution, the material will be suitable for injection for at least 12 hours. Aluminum Ion Breakthrough Al 3+ ion is measured colorimetrically. A drop of the eluate is placed on one end of a special test paper; a drop of a standard solution of Al 3+, concentration 10 ppm, is placed on the other end of the test strip. If the color at the center of the drop of eluate is less red than that of the standard solution, the eluate has passed the Al 3+ Ion Breakthrough Test. Units may be also be expressed as µg/ml.

Radiopharmaceuticals: Quality Control Radiochemical Impurity Chemical Form Free Tc -1 Pertechnetate, TcO 4 Hydrolyzed Reduced Tc 99mTc MAG3 99mTc HMPAO probably TcO(OH) 2.H 2 O, a hydrated Tc-oxide Possible impurity is 99mTc tartrate Stereochemical impurities Thin Layer Chromatography Systems of Tc-99m Compounds For Tc-99m DTPA, MDP, GH, PYP, HDP, MAA, MIAA, SC: CHROMATOGRAPHY SYSTEM IMPURITY MEASURED Silica Gel/0.9% saline Hydrolyzed Reduced Tc Paper/acetone or paper/mek Free Tc (Pertechnetate) Tc-99m Thin Layer Chromatography: Separation on Strip CHROMATOGRAPHY SYSTEM Silica Gel/ Saline Paper/acetone or paper /MEK ITLC-SA/20% saline ITLC-SG/water SEPARATION OF RADIOCHEMICAL SPECIES HR Tc in bottom half; all other species in top half Free Tc in top half; all other species in bottom half. HR Tc in bottom half; all other species in top half Free Tc in top half; all other species in bottom half Thin Layer Chromatography Systems Tc-99m Hepatobiliary Agents CHROMATOGRAPHY SYSTEM IMPURITY MEASURED ITLC-SA/20% saline Free Tc (Pertechnetate) ITLC-SG/water Hydrolyzed Reduced Tc Exception to the Rule For insoluble Tc-99m Radiopharmaceuticals (e.g., Tc-MAA, Tc-SC), it is only necessary to test for the presence of free Tc. No simple system can effectively separate colloidal HR Tc from an insoluble product, so HR Tc cannot be measured in these products. We therefore ignore its presence.

Strip Design Calculations % Impurity = CPM in Half of Strip X 100 CPM in Whole Strip Labeling Efficiency = 100% - % HR Tc - % Free Tc Migration Patterns

What is the radiochemical purity of Tc-MDP for strips below? Numbers represent "Counts per Minute" for each half of strip. Answer: Migration Patterns of Tc-MDP Results: % free Tc = 1,000/(1,000 + 99,000) = 1 % % HR Tc = 1,000/(1,000 + 99,000) = 1 % Therefore, radiochemical purity of Tc-99m MDP is 98% since % Tc-MDP = 100 % - 2 % = 98% DOSE CALIBRATORS: MANDATORY QUALITY CONTROL TESTS TEST FREQUENCY REGULATORY AGENCY Accuracy Annually NRC/State Dept Of Nuclear Safety Constancy Daily NRC/State Dept Of Nuclear Safety Linearity Quarterly NRC/State Dept Of Nuclear Safety Geometry At installation NRC/State Dept Of Nuclear Safety

Accuracy Test This test is designed to show that the calibrator is giving correct readings throughout the entire energy scale that we are likely to encounter. Low, medium, and high energy standards (usually Co-57, Ba-133 or Cs-137, and Co- 60, respectively), are measured in the dose calibrator using appropriate settings. Standard and measured values are compared. Standard Energy (kev) expected value measured value (mci) (mci) Co-57 122 2.48 2.48 Cs-137 662 3.38 3.29 Co-60 1,332 1.55 1.52 Constancy Test This test measures precision and is designed to show that a long-lived source, usually 30 y Cs-137, yields reproducible readings on a daily basis on all isotope settings we are likely to use. The Cs-137 source is placed in the dose calibrator. Activity is then measured on the Cs-137 setting and all other settings used on a daily basis. Values are recorded in the dose calibrator logbook and are compared with recent values to determine if instrument is maintaining constancy on a daily basis Reading (µci) Isotope Monday Tuesday Wednesday Thursday Friday Cs-137 123 124 122 126 124 Ga67 223 224 222 224 226 Tl201 163 164 162 166 164 Tc99m 243 242 244 246 244 I131 313 314 312 316 314 I123 193 192 194 196 194 In111 283 284 282 286 284 Xe133 433 434 432 436 434

Linearity Test This test is designed to prove that the dose calibrator readout is linear for sources varying from the mci range through the mci range. A high activity Tc- 99m source (50-300 mci) is measured at T 0 and at predetermined time intervals up to 72 hours. Expected and actual measurements are compared (and may be analyzed graphically) to determine if the instrument is linear throughout the activity range we are likely to encounter. Elapsed time Expected reading Measured reading (hr) (mci) (mci) 0 300 300 1 267 272 2 238 241 3 212 209 6 150 148 12 75 72.4 24 18.75 19.1 48 1.17 1.19 72 0.073 0.074 78 0.036 0.037 Geometry Test This test is designed to show that correct readings can be obtained regardless of the sample size or geometry. One ml of Tc-99m in a 10 ml syringe (activity 25 mci) is measured in the dose calibrator and the value obtained is recorded. The activity is then diluted with water to 2 ml, 3 ml, 5 ml, and 10 ml. At each of these points a reading is taken and the value recorded. Data are then evaluated to determine the effect of sample geometry on the dose calibrator reading. If instrument is geometry-dependent, it may be necessary to routinely correct readings obtained when using calibrator. This test must also be performed after repair, recalibration or after moving instrument.

The following table displays the data collected during a Geometry Test. The dose calibrator has passed the test. Sample Volume Activity (mci) 0.5 25.5 1.0 25.3 2.0 25 3.0 24.8 4.0 24.7 5.0 24.5 6.0 24.4 Specifications for All Dose Calibrator QC Tests Deviation from standard or expected values must be within + 10%. If Deviation > 10%, then obligation is to record value, note repair or recalibration of instrument, retest, and record new values. In addition to the above steps, every dose must be corrected mathematically until the instrument is repaired. There is NO LONGER a reporting requirement.