Comparing the Agilent 2100 Bioanalyzer Performance to Traditional DNA Analysis Techniques

Similar documents
Quantitative analysis of PCR fragments with the Agilent 2100 Bioanalyzer. Application Note. Odilo Mueller. Abstract

EZ Vision DNA Dye as Loading Buffer

Electrophoresis and the Agilent Bioanalyzer. Advanced Biotechnology Lab I Florida Atlantic University January 23, 2008

Optimizing crna fragmentation for microarray experiments using the Agilent 2100 bioanalyzer. Application Deborah Vitale.

EZ-Vision DNA Dye as Loading Buffer

LabChip 90 System with DataViewer Software

Efficient Method for Isolation of High Quality Concentrated Cellular RNA with Extremely Low Levels of Genomic DNA Contamination Application

G E N A X X O N. b i o s c i e n c e

Performance characteristics of the High Sensitivity DNA kit for the Agilent 2100 Bioanalyzer

E-Gel precast agarose gels

Table of contents. I. Description...2. Kit Components...2. Storage...2. Required reagents and equipment...2. V. Protocol...3. Example Experiment...

Quantitation of ssdna using OliGreen Fluorescent Stain

The Agilent Total RNA Isolation Kit

NGS Sample QC with the Agilent 2200 TapeStation. Rainer Nitsche Application Engineer Agilent Technologies, Inc.

High sensitivity quality control of RNA samples using the RNA 6000 Pico LabChip kit. Application. Rüdiger Salowsky Anna Henger.

High sensitivity quality control of RNA samples using the RNA 6000 Pico LabChip kit. Application. Rüdiger Salowsky Anna Henger.

Best Quantification Practices with the Agilent 5200 Fragment Analyzer System

How does electrophoresis work? The gel is made from agarose, DNA is a negative molecules, Molecules sort based on: Charge, Size, shape.

Product Information FluoroStain DNA Fluorescent Staining Dye (Green, 10,000X)

Focal Points. Application Note FP-153. Next Generation Gel Imaging with GelRed and GelGreen Dyes and GelDoc-It Imaging System

Imaging Nucleic Acid Gels on the Odyssey Fc Imager

LabChip GXII: Antibody Analysis

10X ligation buffer ligase 1 vector DNA insert DNA H 2 O. 10 µl Total Volume. 10X ligation buffer ligase 1 vector DNA insert DNA

NUCLEIC ACID PURIFICATION KITS FAST SIMPLE QUALITY CONVENIENT REPRODUCIBLE

Agarose gel electrophoresis of DNA fragments

DNA Visualizer Extraction Kit

This article reprinted from: Dooley, M. M Restriction endonuclease digestion of a plasmid.

Getting the Most from Your DNA Analysis from Purification to Downstream Analysis. Eric B. Vincent, Ph.D. February 2013

Identification of Unknown Plasmid Code Named 681A18

Application Note. Genomics. Abstract. Authors. Kirill Gromadski Ruediger Salowsky Susanne Glueck Agilent Technologies Waldbronn, Germany

DNA Labeling Kits Fluorescein-dCTP and -dutp Instruction Manual

Alternative to 2D gel electrophoresis OFFGEL electrophoresis combined with high-sensitivity on-chip protein detection.

DNA Labeling Kits Texas Red -dctp and -dutp Instruction Manual

Report on the Verification of Performance of a DNA Extraction Method for Maize Grains

Quality and Quantity Assessment of Nucleic Acids and Proteins. Ebru Serinsöz Linke, MD/PhD

Author. Abstract. Introduction. Agriculture

Application Note 18 RNA/DNA/Protein Sample Preparation METHODS AND MATERIALS INTRODUCTION

Embrace the Future of Electrophoresis

Imaging Nucleic Acid Gels on the Odyssey Fc Imager

Ligation Independent Cloning (LIC) Procedure

Highly Resolved Separation of DNA Fragments on the Fragment Analyzer

Short Tandam Repeat (D1S58) Detection Kit (for Academic Instructions) Product # 54600

About us. Cyanagen s.r.l. has a certified Quality System ISO QUALITY CERTIFIED. Green Stain DNA Loading Dye Rev00

Application of Molecular Biology tools for cloning of a foreign gene

Lecture Four. Molecular Approaches I: Nucleic Acids

Fully Automated Library Quantification for Illumina Sequencing on the NGS STAR

Restriction Analysis of Purified para-r

Evaluating the Agilent 4200 TapeStation System for High Throughput Sequencing Quality Control

All-In-One Precast Agarose Gel Electrophoresis Kit (2x9-Well)

Nucleic acid-free silica-matrix: Regeneration of DNA binding columns

E-Gels Fast, agarose gel electrophoresis for every throughput need

CloneDirect Rapid Ligation Kit

GREEN STAIN DNA LOADING DYE DNA LOADING DYE CONTAINING FLUORESCENT STAIN FOR NUCLEIC ACID DETECTION IN GELS

Lab 7: Running an Agarose Gel for the Restriction Digests

LAB 6: Agarose Gel Electrophoresis of Restriction Digested Plasmid DNA

Comparison of different methods for purification analysis of a green fluorescent Strep-tag fusion protein. Application

Lab 5: Shark Attacks, Again! DNA Fingerprinting to the Rescue

AGAROSE GEL ELECTROPHORESIS. Assiut University

Rapid, Accurate and Flexible DNA Quantitation Using the QuantiFluor dsdna System on the MANTIS Liquid Handler

Nucleic Acid Electrophoresis APPLICATION GUIDE

Fisher (Fairlawn, NJ) and Sigma-Aldrich (St. Louis, MO) and were used without further. (Promega) and DpnI (New England Biolabs, Beverly, MA).

GeNei TM Gel Extraction Teaching Kit Manual

ReliaPrep DNA Clean-Up and Concentration System

Accurate Nucleic Acid Analysis is Critical to Successful Sequencing. Steve Siembieda VP Commercialization November 10, 2016

Highly Resolved Separation of DNA Fragments on the Fragment Analyzer

Methods in virus diagnosis PCR techniques

Sapphire. Biomolecular Imager THE NEXT GENERATION OF LASER-BASED IMAGING

Agilent 2200 TapeStation System ACCELERATE YOUR SAMPLE QC WITH CONFIDENCE

E-Gel 96 Automated agarose gel electrophoresis as easy as Plug & Play

Zymo-Spin Technology. The microcentrifuge column that revolutionized nucleic acid clean-up

TruSeq ChIP Sample Preparation

Implementation of Automated Sample Quality Control in Whole Exome Sequencing

High purity plasmids. Introduction

DIRECTION FOR USE. Art. No / /25 reactions in vitro Diagnosticum for birds

Appendix IV Version

Dig System for Starters

Introduction. Kit components. 50 Preps GF-PC-050. Product Catalog No. 200 Preps GF-PC Preps GF-PC Preps SAMPLE

Review on cloning technique. Lecture 9

XactEdit Cas9 Nuclease with NLS User Manual

Genomic Sequencing. Genomic Sequencing. Maj Gen (R) Suhaib Ahmed, HI (M)

Microfluidic Assays for High Throughput Screening of Protein Quality in Bioprocess Development

BIOL/CHEM 475 Spring 2007 RESTRICTION ENDONUCLEASES AND BACTERIOPHAGE λ

GenBuilder TM Cloning Kit User Manual

Gel Electrophoresis and Analysis

GENOMICS WORKFLOW SOLUTIONS THAT GO WHERE THE SCIENCE LEADS. Genomics Solutions Portfolio

maxxbond kit are bio-degradable and non-toxic for Product information

For in vitro Veterinary Diagnostics only. PCR Detection Kit for Fowl Adenovirus.

qpcr Kit, DNA-free Product components 100 rxn 250 rxn Product description

Cat. # MK600. For Research Use. ApopLadder Ex. Product Manual. v201608da

ITS Sequencing in Millepora. 10/09 Subcloning DNA Fragments into pbluescript Preparation of pbluescript Vector

Hy-Fy High Fidelity Mix (x2)

Polymerase Chain Reaction

Importance of Molecular Genetics

HiPer Random Amplification of Polymorphic DNA (RAPD) Teaching Kit

electrophoresis tech Performance Comparison of the Experion Automated Electrophoresis System and SDS-PAGE for Protein Analysis

Innovations To Meet Your Needs

7.02 Recombinant DNA Methods Spring 2005 Exam Study Questions Answer Key

Validation Study of FUJIFILM QuickGene System for Affymetrix GeneChip

Texas A&M University-Corpus Christi CHEM4402 Biochemistry II Laboratory Laboratory 8: DNA Restriction Digest (II) and DNA Sequencing (I)

Chapter 10 (Part II) Gene Isolation and Manipulation

Transcription:

Comparing the Agilent 2100 Bioanalyzer Performance to Traditional DNA Analysis Techniques Application Note Author Deborah Vitale Agilent Technologies, Inc. Palo Alto, CA, USA Abstract This Application Note compares the performance of the Agilent 2100 bioanalyzer to the performance of the most commonly used techniques for nucleic acid separation, detection and quantification. A standardized mass ladder and two restriction digests were used in this study for comparisons based on sensitivity, accuracy and reproducibility. Separations were performed on the Agilent 2100 bioanalyzer in conjunction with the DNA 7500 and 12000 LabChip kits and by electrophoresis on precast 1 agarose gels stained with ethidium bromide or SYBR Gold nucleic acid stain. Gels were imaged and analyzed with a fluorescence gel imaging system. In addition, samples were quantitated through fluorescence measurements made with the PicoGreen dsdna quantitation kit. The advantages of better detection sensitivity, sizing accuracy and reproducibility, coupled with a rapid and automated analysis indicate that analyses performed with the Agilent 2100 bioanalyzer are superior to those using the leading alternatives.

Introduction As the frontiers of genetic, biological and biochemical research continue to evolve, the modern laboratory must be conscious of the time and cost of analyses, as well as the quality and reproducibility of data. One of the most common yet tedious techniques performed in biological labs is the characterization of DNA via slab gel electrophoresis. This technique is typically qualitative. For a simple check of the samples a polaroid photograph taken over a UV light source is sufficient to identify general sample quality and to estimate fragment size and quantity against molecular size and mass markers. However, more quantitative measurements, which are frequently necessary for samples used in cloning, sequencing or gene expression experiments, require imaging apparatus. The most popular options are fluorescence scanners or digital cameras, which interface with gel analysis software designed to accurately determine the size and amount of the DNA in specific bands. Typically, this imaging software requires a significant amount of manual intervention to generate results. Although quantitative information on individual sample fragments can be extracted from gels, it is not very reliable and further measurements must be made on a fluorimeter or UV spectrophotometer to ensure accurate concentrations. These instruments, however, only provide information on the concentration of the total sample. They do not provide any information regarding sample constituents or potential contaminants, and use of these instruments can be very time intensive while consuming large quantities of precious sample. Until now, there have not been any alternative technologies available to reduce the time, effort and amount of sample needed to size and quantitate DNA fragments. The Agilent 2100 bioanalyzer is the first commercially available chip based nucleic acid analysis system. Through the use of microfluidic technology, only 1 μl of nucleic acid sample is required for separation in micro-channels that are filled with a sieving polymer and a fluorescent dye. When an electrical voltage is applied to the microchip, the sample migrates through these microchannels etched in the chip surface. As sample moves, DNA fragments of different sizes separate according to their mass. Intercalating dye within the sieving matrix allows the migrating DNA fragments to be detected. A PC connected to the instrument controls the separations, records the fragment sizes and concentrations, and allows real-time data analysis and manipulation. This integration streamlines the entire workflow process (Table 1). Analysis with the Agilent 2100 bioanalyzer yields several important advantages compared to traditional separation, imaging and analysis techniques. Analysis of DNA fragments, including chip preparation, separation, detection and data analysis, is performed in a more timely manner. Prepackaged kits, standardized sample preparation and automated analysis yield more accurate and reproducible data due to decreased manual intervention. Three different DNA kits are available to size a broad range of DNA fragments and cover a range of fragment sizes from 25 to 12,000 bp. Most importantly, the bioanalyzer has better sensitivity and performance than the common competitive techniques currently available, using minimal amounts of sample. Materials and Methods Chemicals and reagents pbr322 digested with Bst NI, and Dra I restriction enzyme were purchased from New England Biolabs (Beverly, MA). Low DNA Mass and 1 kb Plus DNA were obtained from Life Technologies (Rockville, MD). Adenovirus 2 DNA and 6X Blue Ficol loading dye were acquired from Sigma (St. Louis, MO). PicoGreen dsdna quantitation kit and 10,000X SYBR Gold nucleic acid stain were purchased from Molecular Probes Inc. (Eugene, OR). Ethidium bromide at the concentration of 10 mg/ml, 0.5M EDTA and 10X TBE were ordered from Amresco (Solon, OH). Reliant 1 agarose precast gels were obtained from Promega (Madison, WI). DNA 7500 LabChip kits and DNA 12000 LabChip kits were supplied by Agilent Technologies. Table 1. Comparison of the time requirements to run chips on the Agilent 2100 bioanalyzer versus comparable alternative techniques. For precise quantitation of samples qualitatively checked on agarose gels, solution-based fluorescence measurements are often used. These methods require an additional 45 minutes to prepare the samples, perform the measurements and analyze the data. Chip Preparation (incl. loading of samples) 5 minutes 10 minutes Separation 30 minutes 30 180 minutes Staining/destaining 15 minutes Scanning/analysis 20 minutes Total 35 minutes 75 225 minutes Gel 2

Sample preparation and quantitation The Dra I digestion of Adenovirus 2 DNA was prepared by combining 40 μg of Ad2 with 246 units of Dra I for 3 hours at 37 C. The reaction was stopped by adding EDTA to a final concentration of 10 mm. The Ad2/DraI sample was then aliquoted and diluted with deionized water to the following concentrations: 250 ng/μl, 100 ng/μl, 50 ng/μl, 10 ng/μl, 2.5 ng/μl, and 0.5 ng/μl. The 1 μg/μl stock of pbr322/bst NI digest was diluted to the following concentrations: 500 ng/μl, 250 ng/μl, 50 ng/μl, 5 ng/μl, 0.5 ng/μl, and 0.05 ng/μl. Low DNA mass ladder was alliquoted at the following concentrations: 117.5 ng/μl, 58.75 ng/μl, 14.69 ng/μl, 3.67 ng/μl, 0.92 ng/μl, and 0.23 ng/μl. For comparison, all samples and dilutions were quantitated using the PicoGreen dsdna quantitation kit from Molecular Probes, in 200 μl volumes on a 96-well micro titer plate. Kit reagents were prepared according to the recommended PicoGreen protocols. Fluorescence measurements were made on the CytoFluor II Microplate Fluorescence Reader (PerSeptive Biosystems). The concentrations determined by the fluorescence measurements were used as the reference sample concentrations. Quantitation measurements from the other comparative techniques were compared against these values. Chip preparation The Low DNA Mass and pbr322/bst NI samples were analyzed with the DNA 7500 LabChip kit for sizing and quantitation. The Ad2/Dra I digest sample was analyzed with the DNA 12000 LabChip kit. All chips were prepared according to the instructions provided with the DNA 7500 and DNA 12000 LabChip kits. Gel preparation, staining and imaging Loading dye was diluted 1:5 with 1X TBE buffer. One microliter of ladder or sample was mixed with five microliters of diluted dye, and then loaded onto a 1 agarose precast gel according to the Reliant gel instructions. The gels were run in 1X TBE at 100 V for approximately 1.5 hours. Following separation, the gels were stained with either ethidium bromide or SYBR Gold nucleic acid stain for 15 minutes. The staining buffer was made by adding 5 μl of 10 mg/ml ethidium bromide or 10,000X SYBR Gold to 50 ml of 1X TBE. After staining, the gels were destained to remove background fluorescence by rinsing with 50 ml of 1X TBE, twice. Immediately following destaining, the gels were imaged on the FluorImager 595 (Molecular Dynamics) 610 RG filter. The PMT voltage was set to 900V for a two minute scan at normal detection sensitivity. Gel images were generated with a digital resolution of 16 bits per pixel, with the pixel size set to 100 microns. The gel image was then analyzed with the FragmeNT software. Results and Discussion A B C 500 ng/µl 5 ng/µl 0.05 ng/µl 500 ng/µl In many biological experiments where nucleic acids are manipulated in multistep procedures, the introduction of impurities, nonspecific generation of PCR products, and incomplete digests, can have drastic effects on the outcome of an experiment. The ability to detect and accurately characterize small amounts of fragments enables the scientist to optimize reactions and actively guide the course of experimentation. The DNA 7500 LabChip kit is designed to size and quantitate doublestranded DNA fragments from 100 7,500 bp while the DNA 12000 LabChip kit sizes fragments as large as 12 kbp. Sensitivity and linearity System performance was evaluated by comparing the results obtained with the Agilent 2100 bioanalyzer to gel electrophoresis on 1 agarose gels stained with either SYBR Gold or ethidium bromide. Comparison between the chip and the gel images of pbr322/bst NI digest dilutions run at concentrations ranging from 500 ng/μl to 0.05 ng/μl (Figure 1) clearly shows that the bioanalyzer is able to detect fragments that cannot be positively identified on the agarose 5 ng/µl 0.05 ng/µl 500 ng/µl 5 ng/µl 0.05 ng/µl Figure 1. Gel image comparison of pbr332/bst NI digest dilutions run on: A) the Agilent 2100 bioanalyzer, B) a 1 agarose gel stained with SybrGold nucleic acid stain, and C) a 1 agarose gel stained with ethidium bromide. For sizing comparison on the gels, the six pbr322/bst NI samples were run with the DNA 7500 LabChip kit ladder in lane number 1. Both sets of agarose gels were scanned with the FluorImager 595. 3

gels. Dilutions of Low DNA Mass, which contain standardized fragment concentrations, were used to determine the limits of detection for each system. The bioanalyzer was able to consistently detect 0.02 ng of fragment. This is five times more sensitive than gels stained with SYBR Gold and twenty five times more sensitive than gels stained with ethidium bromide, which have detection limits of 0.1 ng and 0.5 ng respectively. Although high sensitivity is desirable, accurate quantitation can only be obtained within the linear range of the assay for the Agilent 2100 Bioanalyzer. Low DNA Mass fragments ranging from 0.02 ng to 48 ng were used to determine the linearity of the bioanalyzer (Figure 2). The DNA 7500 assay was found to be linear from 0.5 50 ng/μl (R 2 = 0.9988). Such high accuracy can be achieved with the Agilent 2100 bioanalyzer because both external standards, in the form of a DNA ladder, and internal standards, in the form of mass and size markers which are run in every lane, are used to calibrate the sizing and quantitation of each sample. This eliminates slight differences as each sample is sequentially run through the separation channel. Typically, internal standards are not run with samples on a gel. Additional variability can also occur when loading samples into submarine gels and in the staining following the separation. Subsequently, it is commonly accepted that quantitation off of gels is not a reliable method of accurately measuring sample concentration. Quantitative analysis Accurate and reproducible quantitation is important for many procedures including depositing precise amounts of PCR products onto microarrays, determining the correct amount of cloning insert needed relative to vector for a ligation reaction or calculating the correct amount of DNA template needed for a sequencing reaction. The ability of the bioanalyzer to accurately determine fragment concentration, while simultaneously sizing and checking sample integrity, is a valuable advantage over current competitive technologies. Quantitation accuracy and reproducibility were Fluorescence area 900 800 700 600 500 400 300 200 100 Linearity: 0.02 48 ng DNA 0 20 DNA (ng) R 2 = 0.9988 determined by analysis of low DNA mass ladder fragments at six concentrations ranging from 1.2 ng/μl to 24.2 ng/μl Table 2). For these measurements the precision was between 85 and 90, whereas the percent standard deviation was between 4 and 11. The quantitation derived from the agarose gels was considerably less accurate and reproducible (data not shown). Only at concentrations above 50 ng/μl did the accuracy approach 50, while reproducibility even at this concentration was very poor due to the high variability in staining. Figure 2. Linear relationship between the expected DNA fragment concentration and the area under the fluorescence electropherogram. Although the assay is linear over three orders of magnitude, the recommended DNA assay specifications for concentration lay between 0.5 ng 50 ng/μl. Table 2. Six low mass ladder fragments were analyzed with the Agilent 2100 bioanalyzer to determine the quantitation accuracy and reproducibility of the DNA 7500 assay. 40 60 Conc. (ng/μl) Chip conc. (n = 60) dev. 1.2 1.1 9.4 0.13 10.6 2.4 2.1 11.8 0.14 5.9 4.9 4.2 13.7 0.23 4.8 9.3 8.5 13.3 0.43 4.5 14.8 12.8 15.5 0.65 4.4 24.2 21.2 14.4 1.36 5.6 4

Sizing analysis Dilutions of pbr322/bst NI digest samples, which contain five fragments ranging from 121 1,857 bp, were used to determine the sizing accuracy and reproducibility of the DNA 7500 assay compared to the analysis of agarose gels (Table 3). On average, the data generated with the bioanalyzer showed lower percentage errors in comparison to agarose gel analysis. The reproducibility of the bioanalyzer was also slightly better than that of the ethidium bromide and the SYBR Gold stained gels. The performance of the DNA 12000 LabChip kit for sizing analysis was compared to agarose gel analysis using dilutions of an Ad2/DraI digest that is comprised of twelve fragments ranging from 119 bps to 9228 bps (Figure 3). Not only are the bands in the bioanalyzer gel-like image more defined and more uniform, all of the bands are also consistently detected (Figure 4). The 119 bp fragment does not resolve from the 149 bp fragment on the agarose gels, even when 250 ng of sample are loaded onto the gel. The ability to detect and accurately size both very small and very large digest fragments with the same assay is a unique advantage of the bioanalyzer. Table 4 contains the data comparing the DNA 12000 assay performance to the gel analysis with FragmeNT. The sizing accuracy of all three techniques was very good. The DNA 12000 assay showed similar reproducibility compared to agarose gels stained with SYBR Gold and ethidium bromide. Table 3. Five fragments from the pbr322/ Bst N I digest were used to compare the sizing accuracy and reproducibility of the Agilent 2100 bioanalyzer to that achieved by gel analysis. (bp) Fluorescence Chips DNA 7500 assay Chip Dev. 1 Agarose gels SYBR Gold stained Gel Dev. 1 Agarose gels EtBr stained Gel Dev. 121 112 7.2 5.9 5.2 133-9.7 12.9 9.7 115 5.2 4.6 4.0 383 380 0.9 1.3 0.3 459-19.8 71.2 15.5 399-4.3 17.4 4.4 929 932-0.4 3.0 0.3 996-7.2 19.9 2.0 958-3.1 29.0 3.0 1,058 1073-1.4 53.0 4.9 1,128-6.6 28.1 2.4 1,097-3.7 35.6 3.2 1,857 1650 11.2 45.3 2.8 1,985-6.9 34.3 1.7 1,921-3.4 46.1 2.4 500 400 300 200 100 0 30 35 40 45 50 55 60 65 70 75 80 85 90 95 Time (s) Figure 3. Electropherogram of Ad2/DraI at 50 ng/μl analyzed on the Agilent 2100 bioanalyzer. A B C 100 ng/µl 10 ng/µl 2.5 ng/µl 100 ng/µl 10 ng/µl 2.5 ng/µl 100 ng/µl 10 ng/µl 2.5 ng/µl Figure 4. Gel image comparison of adenovirus 2 DraI digest dilutions run on A) the Agilent 2100 bioanalyzer, B) a 1 agarose gel stained with SYBR Gold nucleic acid stain, and C) a 1 agarose gel stained with ethidium bromide. For sizing comparison on the gels, the six Ad2/Dra I samples were run with the DNA 12000 LabChip kit ladder in lane number 1. Both sets of agarose gels were scanned with the FluorImager 595. 5

Conclusion The Agilent 2100 bioanalyzer shows excellent performance for quantitative analysis and sizing of DNA fragments. The advantages of the bioanalyzer boldly stand out with respect to sensitivity, linear dynamic range and sizing capability over a broad range of fragment sizes. The bioanalyzer performed fragment quantitation at 85 or better, with a percent standard deviation less than or equal to 10. The same precision could not be attained with the fluorescence scanner. Over a size range of 100 to 12000 bps, the percent standard deviation of both the DNA 7500 and DNA 12000 assays was found to be 5 or lower, while the accuracy was 90 or better. Scanned gels stained with ethidium bromide or SYBR Gold had typically higher percentage sizing errors. The performance of the bioanalyzer was found to be superior to agarose gel analysis. Additionally, the bioanalyzer consumes considerably less sample and is less labor intensive. As the separations are performed, the DNA fragments are detected and analyzed in real-time, and the digital data storage facilitates easy data exchange. There are no additional staining, de-staining or imaging steps before data extraction. Automation of both the separation and data analysis makes the Agilent 2100 bioanalyzer easy to use, while the superior performance makes it an ideal tool for the analysis of nucleic acids. Table 4. System performance of the Agilent 2100 bioanalyzer versus that of scanned agarose gels with respect to accuracy and reproducibility, over a broad range of Ad2/Dra I digest fragment sizes. Note that the 119 bp fragment was only detected by the Agilent 2100 bioanalyzer. (bp) Chips DNA 12000 assay Chip (n = 216) dev. 119 113 5.0 0.5 0.4 1 agarose gels SYBR Gold stained Gels (n = 54) dev. 1 agarose gels EtBr stained Gels (n = 54) dev. 149 153-2.5 0.6 0.4 137 8.1 2.3 1.7 1.3 12.3 2.9 2.2 641 656-2.2 12.9 2.0 672-4.9 6.9 1.0 671-4.7 7.6 1.1 815 831-1.9 15.3 1.8 829-1.7 6.1 0.7 819-0.5 21.9 2.7 1,195 1,205-0.1 37.4 3.1 1,261-5.5 32.5 2.6 1232-3.1 28.4 2.3 2,058 1,937 5.9 99.9 5.2 2,120-3.0 23.4 1.1 2,099-2.0 15.2 0.7 2,800 2,626 6.2 58.1 2.2 2,905-3.8 118.7 4.1 2,848-1.7 27.4 1.0 3,588 3,343 6.8 36.4 1.1 3,683-2.7 55.7 1.5 3,634-1.3 59.8 1.7 4,182 4,145 0.9 32.8 0.8 4,316-3.2 49.6 1.2 4,247-1.6 56.3 1.3 4,845 4,719 2.6 31.3 0.7 4,805 0.8 178.1 3.7 4,732 2.3 319.5 6.8 6,297 5,793 8.0 78.9 1.4 6,642-5.8 172.2 2.6 6,332-0.6 366.6 5.8 9,228 8,508 7.8 338.7 4.0 10,367-12.4 206.2 2.0 10,349-12.1 262.0 2.5 www.agilent.com/chem For Research Use Only. Not for use in diagnostic procedures. This information is subject to change without notice. Agilent Technologies, Inc., 2000, 2016 Published in the USA, April 1, 2016 5980-0549E