Purity Determination and DSC Tzero Technology

Similar documents
High Heating Rate DSC

Physical Aging and Fragility of Amorphous Sucrose by DSC

Impact of Tzero Technology on DSC Resolution

Physical Aging and Fragility of Amorphous Polyethylene Terephthalate

Impact of Tzero Technology on the Measurement of Weak Transitions

Characterization of Pharmaceutical Materials by Thermal Analysis

TA INSTRUMENTS DIFFERENTIAL SCANNING CALORIMETERS

Thermal Analysis. Dr. Lidia Tajber School of Pharmacy and Pharmaceutical Sciences, Trinity College Dublin

i Series Thermal Analysis Systems

G16 - THERMAL ANALYSIS

HYDROGEN CONCENTRATION DETERMINATION IN PRESSURE TUBE SAMPLES USING DIFFERENTIAL SCANNING CALORIMETRY (DSC)

SEMATECH Provisional Test Method for Evaluating Bulk Polymer Samples of UPW Distribution System Components (DSC and TGA Methods)

Interpretation of the Crystallization Peak of Supercooled Liquids Using Tzero DSC

Differential Scanning Calorimetry DSC 200 F3

Amorphous Content of Common Pharmaceutical Materials

THERMAL ENERGY STORAGE SYSTEMS STEARIC / LAURIC ACID MIXTURES AS PHASE CHANGE MATERIALS

MICRO THERMAL ANALYSIS CALIBRATION, REPEATABILITY AND REPRODUCIBILITY

A comparison of a novel single-pan calorimeter with a conventional heat-flux differential scanning calorimeter

New Castle, DE USA. Lindon, UT USA. Hüllhorst, Germany. Shanghai, China. Beijing, China. Tokyo, Japan. Seoul, South Korea.

Standard Test Method for Transition Temperatures of Polymers By Differential Scanning Calorimetry 1

A Beginner s Guide to Differential Scanning Calorimetry (DSC)

The Study of Solid-Liquid Phase Equilibria With Phase Technology Analyzers

Nova Geração de Instrumentos de Análise Térmica

TA INSTRUMENTS. Thermal Analysis

APPLICATIONS OF HIGH SENSITIVITY DSC

DSC Step Annealing for Fingerprinting Molecular Structure in Poly(vinylidene fluoride)

Thermal Analysis Investigation of a Poly(Lactic Acid) Biodegradable Plastic

DSC limit of detection: < 0.1 µw (depending upon the sensor) System time constant: 0.6 s to 3 s (depending upon the sensor)

REALIZATION OF THE FREEZING POINTS OF INDIUM, TIN AND ZINC USING STAINLESS STEEL CASED CELLS

Introduction to TG/DTA/DSC. Thermal Processing Technology Center Illinois Institute of Technology

The Q5000 Automatic Sample Processor

TA INSTRUMENTS THERMOGRAVIMETRIC ANALYZERS

A vapor (or gas) needs a completely enclosed container to have a definite volume; it will readily take on any shape imposed.

Differential Scanning Calorimeter. DSC-60 Plus Series C160-E013A

New Castle, DE USA Lindon, UT USA Pittsburgh, PA USA Elstree, United Kingdom Shanghai, China Beijing, China Taipei, Taiwan Tokyo, Japan Seoul, Korea

A chemical change indicates the onset of oxidation or degradation of the material. 2,3 There are

EXPERIMENT 15 FREEZING POINT: A COLLIGATIVE PROPERTY OF SOLUTIONS

Differential Scanning Calorimetry DSC 404 F3

Standard Test Method for Oxidative-Induction Time of Polyolefins by Differential Scanning Calorimetry 1

DIFFERENTIAL SCANNING CALORIMETER. Chip-DSC 100 Chip-DSC 10

DIFFERENTIAL SCANNING CALORIMETER. Chip-DSC 100 Chip-DSC 10

Characterization of Polymorphic Transitions in a Pharmaceutical by DSC and MDCS

Solubility of Small-molecule Drugs into Polymer Excipients in Hot Melt Extruded Dosage Forms

DIFFERENTIAL SCANNING CALORIMETER. Chip-DSC 100 Chip-DSC 10

Methodology for DSC calibration in high heating rates

Crystallization kinetics of PHB and its blends 3.1 Introduction

Simultaneous Thermal Analyzer STA 449 F5 Jupiter

Collected Applications Thermal Analysis PHARMACEUTICALS C

Collected Applications Thermal Analysis ELASTOMERS. Volume C

Quantitation of amorphous content

Analyzing & Testing. Guarded Hot Plate - GHP. Method, Instrumentation, Applications. GHP 456 Titan. Leading Thermal Analysis

Differential Scanning Calorimetry DSC 404 F1

The Use of Thermal Analysis in the Development and Characterisation of Materials for Energy Storage -Thermochemical Energy storage

A Short Reviews of Thermal Techniques

OPEN YOUR EYES TO A WORLD OF OPPORTUNITY IN THERMAL ANALYSIS. DSC 4000/6000/8000/8500 Comprehensive Solutions for Differential Scanning Calorimetry

Analyzing & Testing. Thermogravimetric Analysis TGA. Method, Technique, Applications

INFLUENCE OF EXPERIMENTAL CONDITIONS ON DATA OBTAINED BY THERMAL ANALYSIS METHODS

Method Development and Validation for Online UV-Dissolution Methods Using Fiber-Optic Technology

Plastics Differential scanning calorimetry (DSC) Part 2: Determination of glass transition temperature and glass transition step height

Realization of ITS-90 from K through K: One Company s Approach

Instructions for using the PerkinElmer DSC-8500

HAZARDOUS EVALUATION OF ORGANIC BLOWING AGENTS FOR TIRES

Analyzing & Testing. Differential Scanning Calorimetry. Method, Technique, Applications DSC 204 F1. Leading Thermal Analysis

DIFFERENTIAL SCANNING CALORIMETER DSC PT 1000

T H E R M A L A N A L Y S I S

ISO INTERNATIONAL STANDARD

ISO INTERNATIONAL STANDARD. Plastics Differential scanning calorimetry (DSC) Part 2: Determination of glass transition temperature

Physical pharmacy. dr basam al zayady

TGA Calibrations, Adjustments & Maintenance. Queens University Belfast 16/02/12

Part 1: General principles

Thermal properties of components of lightweight wood-framed assemblies at elevated temperatures. Bénichou, N. ; Sultan, M. A.

Carbon Doping of MgB 2 by Toluene and Malic-Acid-in-Toluene

DIL 402 CD Double/Differential Dilatometer

STA 2500 Regulus Simultaneous Thermal Analysis

Thermal cover:layout 1 1/18/11 3:56 PM Page 2 TA Instruments

Simultaneous Thermal Analysis STA 2500 Regulus

Thermoset Curing Schedule and its Affect on the Final Properties

STUDY OF POLYMORPHIC TRANSFORMATION OF ORNIDAZOLE DRUG BY DIFFERENTIAL SCANNING CALORIMETRY AND OTHER COMPLEMENTARY TECHNIQUES

Thermal Analysis Premium DSC 3+ STAR e System Innovative Technology Versatile Modularity Swiss Quality

EXPERIMENTAL APPLICATIONS OF DIFFERENTIAL SCANNING CALORIMETRY TO TEST AND CHARACTERIZE POLYMERS AT THE UNIVERSITY OF WASHINGTON BOTHELL

Differential Scanning Calorimetry DSC 404 F1/F3 Pegasus

CCT/10-19 A procedure for estimating ideal freezing temperatures using the gradient of the freezing plateau

Atmosphere Effect on Soldering of Flip Chip Assemblies. C. C. Dong Air Products and Chemicals, Inc. U.S.A.

SOLVING ADHESION AND DIE BUILDUP PROBLEMS IN EXTRUSION COATING

S. J. Nolan and R. Savin

Introduction. From the combination of TG and DTA or HDSC one receives a broad range of information, such as:

TA INSTRUMENTS. Thermal Analysis

How Does Temperature Non-uniformity of an Annealing Furnace Affect SPRT Stability?

FURTHER DETAIL FROM DSC THERMOGRAMS

Thermal Value Excellence

CHM-202 General Chemistry and Laboratory II Laboratory 2 Molar Mass by Freezing Point Depression

Simultaneous Oxygen, Nitrogen and Hydrogen Determination of Metals. Dennis Lawrenz, John Hawkins

No change similar. 5. Furnace temp resolution should be Specification deleted. No change above. 8. DSC measuring range should be preferably

Comprehensive Analysis of Oxidized Waxes by Solvent and Thermal Gradient Interaction Chromatography and Two-Dimensional Liquid Chromatography

VALIDATION OF ANALYTICAL PROCEDURES: METHODOLOGY *)

Available online at Scholars Research Library

Study on crystallization behaviour of co-polyamide 66 containing triaryl phosphine oxide

Heat Transfer in Polymers

Decomposition of Li 2 CO 3 in existence of SiO 2 in mould flux of steel casting

Differential Scanning Calorimetry DSC 3500 Sirius

Transcription:

Purity Determination and DSC Tzero Technology R. Bruce Cassel, Ph.D. TA Instruments, 19 Lukens Drive, New Castle DE 1972, USA ABSTRACT The determination of absolute purity by DSC has been an accepted technique since the mid 196s. One source of error in the purity method, that has limited its applicability and accuracy, is uncorrected thermal lag. New DSC instrumentation incorporates these corrections directly into the heat flow signal. This superior approach leads to improved peak resolution and to more accurate purity analysis. It is likely that this advance will allow purity analysis to be performed at faster heating rates and the purity method will be extended to higher levels of purity. INTRODUCTION The melting of a pure material takes place over an infinitely narrow temperature range. The result is a sharp melting spike on a DSC trace at a temperature characteristic of that material. For a material that only has a small amount of impurity (that dissolves in the melt, but not in the solid) the melting point is depressed and the melting range is broadened. This is the basis of the calorimetric purity determination by observed melting point depression. The equations that predict the rate of heat flow versus temperature for melting a material (of given molecular weight, melting point, heat of fusion and purity) are derived from thermodynamic relationships, and predate the existence of commercial DSC s (1). A. P. Gray adapted these relationships to the non-ideal measuring systems of DSC s, in the 196 s (2). This purity method is thoroughly discussed and evaluated in ASTM Special Technical Publication (STP) Number 838 (3), and there is an ASTM standard method E 928 that details procedures and precautions with its use (4). Although there are numerous theoretical and practical limitations of the method, it has proven a useful tool for quickly determining the absolute purity of purified organic materials, where the only impurities present are ones that are difficult to separate through ordinary separation techniques. The beauty of the method is that it determines from a single DSC run the extent of impurity to within 5 or 1 percent independently of which impurities are present. Figure 1 shows the melting endotherms of phenacetin (pacetophenetidide) at three levels of impurity obtained from NIST for purposes of interlaboratory study (5). The purities are 99.97, 99.5 and 99. %. It was clear from the earliest efforts to utilize the theoretical equations, that the thermal resistances and capacitances that are inherent to the DSC and sample encapsulation, affect the shape of a sharp melting peak. This is because the heat flow into the melting sample must flow across the thermal resistances of the DSC cell and across the small air gap that separates the DSC sensor-surface from the sample itself. This heat flow produces a temperature drop, known as thermal lag. Thus, the recorded 1 TA295

temperature (of the sensor) is not the same as the actual sample temperature. Moreover, during melting of a pure material the actual sample temperature stops increasing during the melt and stays constant until the last bit of sample is melted. Meanwhile, the DSC set point temperature continues to increase linearly at the designated heating rate. The sample pan, which is in good contact with the sample, is thus being programmed at a nonlinear rate, producing a heat flow effect that distorts the peak shape. One way to correct this thermal lag is to correct the purity calculation for the heat capacity of the sample pan and the thermal resistances in the system. To obtain these parameters with sufficient accuracy requires that the DSC user make additional DSC measurements of the contact resistance (R ) and the heat capacity of the pan (Cp(pan)) at the temperature of the purity analysis. R is obtained from the leading edge slope of the melting peak of a pure material melting in the same temperature range and using the same pan type. Cp(pan) can be obtained from the pan weight and the specific heat capacity of the pan material. Another approach is to estimate these corrections and use conditions (small sample size, very slow heating rate) to minimize the effect. Besides taking more time, these conditions result in a much lower heat flow signal, resulting in noise and baseline curvature errors. These limitations have reduced the practicality of the technique and -1 Heat Flow T4P (W/g) -3 99.% 99.5% 99.97% -5 126 127 128 129 13 131 132 133 134 135 136 Figure 1 - DSC Melting Curves of Phenacetin at Three Purity Levels tempted users to take short cuts, resulting in compromised accuracy. INSTRUMENTATION Tzero TM Technology makes use of an alternative DSC cell design enabling the thermal resistances and capacitances to be accurately measured as part of a standard calibration process. This capability is accomplished by using a DSC sensor with an 2 TA295

additional Tzero thermocouple positioned between the sample and reference sensor as the set point controller for the DSC, as shown in Figure 2. This physical arrangement, known as Tzero Technology, compensates for the thermal lags in the DSC. The heat flow is described by a four-term heat flow equation, rather than by the single-term equation conventionally assumed (6). Advanced Tzero Technology employs a model of the entire DSC-Sample system and utilizes a more complete, four-term heat flow equation that effectively compensates for all thermal lag effects (7). As a result, all DSC melting peaks are automatically corrected for the thermal lag related errors. This is seen by comparing samples of high purity as shown in Figure 3. Comparing the onset and peak temperatures on these two plots (taken from the same DSC run, but using different sensor signals) shows how the peak shape, especially the leading edge, that should be vertical in this case, is more accurately represented using Advanced Tzero Technology. It is this leading edge of the purity peak that is analyzed in the purity software. While the improvements in the DSC signals to remove peak distortion has the most dramatic improvement effect on the purity determination, there are some other aspects of Tzero Technology that warrant mention. First, the use of the four-term heat flow equation also results in a flatter DSC baseline, since two of the four terms remove effects of asymmetry in the DSC. This asymmetry is the cause of baseline curvature, a dramatic limitation of some DSC instrumentation. Second, improvements in the design of the furnace and cooling system of the Tzero DSC result in a much more reproducible baseline. This is because the cooling accessory is coupled to the furnace by nickel rods that offer a constant heat flow path despite the dimensional changes inherent in such extremes of temperature. The differential expansion is responsible in part for the shifting baselines of traditional DSC designs. Since baseline effects limit the sensitivity of a DSC, the result of this improvement in baseline reproducibly is improved sensitivity for measuring small samples. Also, by reducing the negative aspects of extreme temperature gradients, the Silver Base for Cell Lid #1 Measuring Chamber Silver Base for Cell Lid #2 Chambers for Temperature Conditioning of Purge Gas Tzero Sensor Furnace 54 Nickel Cooling Rods Cooling Flange Figure 2 - Tzero TM Sensor (Left) Q Series TM DSC Cell (Right) nickel rod coupling enables boosting the heating and cooling power, so that the system is capable of a greater range of heating and cooling rates, and of temperatures. By 3 TA295

improving the furnace control feedback loop, using the Tzero sensor and heater (PID) control, the equilibration time has also been dramatically improved. PURITY DETERMINATION A traditional DSC curve and purity calculation are seen in Figure 4. The sample is a 1 mg sample of over-the-counter ibuprofen pain medication. The sample weight has been adjusted to take into account the inert ingredients in the formulation. These ingredients are sufficiently dissimilar to the active ingredient that there is no mutual dissolution either in the solid or liquid phase. Hence, the weight fraction of the excipient does not affect the purity analysis. This lack of melting depression by the inert ingredients is normally the case for drug formulations, since without mutual dissolution chemical interaction is unlikely. Were it otherwise, drug-excipient interaction would reduce product shelf life. The heating rate used in Figure 4 is the common 1 C/min (3). The sample was hermetically sealed to ensure there is no loss of volatiles, the heat of which might effect the purity determination. Unfortunately, as with most hermetic pans, the mass of the pan is greater than that of a standard pan, and the coupling to the DSC is poorer. Both of these would ordinarily be negatives for a purity analysis, since the thermal lag corrections are proportional to these factors. However, because of Advanced Tzero TM Technology both of these effects are addressed in the DSC signals. The purity calculation data in Figure 4 is the van t Hoff plot, the fitting of the partial area data versus the temperature corresponding to those partial areas. That is, each of the data points reflects a determination of the fraction of the total area, the fraction of the material that has melted up to that temperature. For high purity samples the partial area measurements are very small. (The corresponding points on the DSC curve are closely spaced in the region of the pre-melting tail.) Any errors in the DSC signal may cause scatter or bias in the partial area data. Hence, to obtain reliable purity determination data requires much of a DSC. 2 2 156.6 C 28.98J/g 156.63 C 28.84J/g Heat Flow T1 (W/g) Heat Flow T4P (W/g) 158. C 156.83 C -6 152 154 156 158 16 162 Tzero -6 152 154 156 158 16 162 Figure 3-3.5 mg Indium by Traditional DSC (left) and Using Advanced Tzero TM 4 TA295

5 TA295

High resolution. The area of each incremental partial area slice of the peak should represent only the data in that narrow temperature range. If a DSC lacks resolution it means that each data point has been smeared (averaged) to some extent with its neighbors. The Q1 DSC with Advanced Tzero Technology is shown to give the highest resolution of any DSC when compared using the TAWN resolution index (6). This is a critical need for purity analysis. Low Noise. Noise in the DSC signal will cause scatter in the partial area data and reduce the accuracy of the determination. Noise will also prevent the analysis of small samples, that are preferred for this analysis, and often a small sample is all that is available. (If the DSC data is smoothed, it will change the results of the purity calculation.) Straight Baselines. As can be seen in Figure 4, the melting starts very gradually. From theoretical considerations, a 99 % pure material is 2 % melted by 3 ºC below the melting point (7). So in fact, every DSC melting peak with an impurity has a long, gradual, low-temperature tail that is part of the overall melting. It is important to start the peak baseline well below the extrapolated onset to the peak. Baseline curvature or drift will not only adversely affect the total area of the peak, but more importantly, it will bias the partial area data, that when fitted to the van t Hoff equation will cause error in the determination of purity. One of the biggest limitations to accurate purity analysis is the distortion of melting peak shape by the temperature gradient between sample and sensor. Even correcting this in the purity software doesn t completely correct for this problem because of uncertainties in leading edge slope values used. Advanced Tzero automatically addresses these effects by using the full four-term heat flow equation, evaluated through calibration. Moreover, removing the distortion provides additional benefits for purity determination especially for high purity materials. That is, for these analyses the information of the impurity is mostly held in the small portion of the leading edge of the melting peak. So, to obtain the best analysis, the user selects limits for the partial area calculation that cover this portion of the melting peak (see Figure 5, right hand side). From the Advanced Tzero data from a Q1, this can be seen as the data up to the.2 72 -. -.2 7 Heat Flow (W/g) -.4 -.6 -.8 Purity: 98.6 mol % Melting Point: 71.83 C Depression:.53 C 66-1. Delta H: 36.39 kj/mol Correction: 5.64 % -1.2 Mol. Weight: 26.28 g/mol Cell Const: 1.28 Onset Slope: -12.89 mw/ C 64-1.4 Total Area / Partial Area 2 4-1.6 62 45 5 55 6 65 7 75 8 Exo up TA Instruments Figure 4 - Purity Analysis of Ibuprofen 6 TA295 68

inflection point of the leading edge. But for a conventional DSC it is difficult to assess this point because of the heat flow distortion. The left side of Figure 6 shows the leading edge of the melting peak for the same DSC run but analyzing the traditional one term heat flow relationship. This is the signal analyzed by all conventional DSC s including power compensation and heat flux designs. Detecting reaction products caused by interactions between the active drug and another potential formulation component is one of the uses of the DSC purity determination (8). The procedure for this calls for determining the purity of the active component, then grinding it with all the components, aging the mixture at an elevated temperature, and then rerunning the mix to assess additional impurities. If there is no loss of purity it is an indication that the components in the formulation are stable and will 1 1 134.3 C 134.33 C Heat Flow T1 (W/g) -1-3 Heat Flow T4P (W/g) -1-3 135.1 C -5 133. 133.5 134. 134.5 135. 135.5 136. 134.52 C -5 133. 133.5 134. 134.5 135. 135.5 136. Figure 5 - Melting of 99.97% Phenacetin Using Traditional DSC (Left) And Using Advanced Tzero TM DSC (Right) resist shelf-life degradation of the active component. Figure 6 shows two samples of an over-the-counter ibuprofen formulation, one from a container well past the expiration date. The lower temperature curve is from the aged sample and indicates a change in its purity. (This figure is meant to demonstrate the technique, and not meant to infer that the lower purity product is in any way hazardous or less effective.) CONCLUSIONS DSC has long been used to determine purity in pharmaceutical components and formulations. To obtain accurate data requires careful technique. Q Series DSCs, known as Advanced Tzero Technology, make the gathering of accurate purity data easier. By compensating for thermal lag errors, the Q1 DSC, with Advanced Tzero TM Technology reduces many of the prior limitations of the purity technique. REFERENCES 1. F. D. Rossini, Chemical Thermodynamics, John Wiley & Sons, New York, 195, p. 294 and 454. 2. W. P. Brennan, M. D. DiVito, R. L. Fyans, A. P. Gray, An Overview of the Calorimetric Purity Method in Purity Determinations by Thermal Analysis, R. L. Blaine and C. K. Schoff (Eds.), American Society for Testing and Materials Special Technical Publication 838, West Conshohochen, PA, 1984, pp. 5-15. 7 TA295

3. Purity Determinations by Thermal Analysis, R. L. Blaine and C. K Schoff (Eds.), Special Technical Publication 838, West Conshohocken, PA., 1984. 4. E 928, Mole Percent Impurity by Differential Scanning Calorimetry, American Society for Testing and Materials, West Conshohocken, PA 5. SRM 1514 Thermal Analysis Purity Sample Set, NIST, Gaithersburg, MD. 6. R. L. Danley, P. A. Caulfield, DSC Baseline Improvements by a Heat Flow Measurement Technique, Proceedings of the 29 th Conference of the North American Thermal Analysis Society, 21, pp. 667-672. 7. United States Pharmacopeia, XXII edition, U.S. Pharmacopeial Convention, Rockville, MD, 199, pp. 1615-1617 8. D. Giron, C. Goldbronn, Journal of Thermal Analysis, 1995, 44, p. 217..5 Heat Flow T4P (W/g). -.5-1. -1.5. 56 58 6 62 64 66 68 7 72 74 76 Figure 6 -Ibuprofen Formulation. Two samples: One Past the Manufacturer s Expiration Date KEYWORDS differential scanning calorimetry, melting, purity, pharmaceuticals 8 TA295