Water-Water Titration for the MicroCal ITC200

Similar documents
Calcium-EDTA Tutorial for the MicroCal ITC200

Standard Operating Procedure

DNA RESTRICTION ANALYSIS

Family Secrets. Part 4 Testing for the HD Gene

Polymer Block Cleaning Kit

PrepFiler Express and PrepFiler Express BTA Forensic DNA Extraction Kits

Procedure for PCR Amplification for Casework

Human DNA Alu Amplification by Polymerase Chain Reaction (PCR)* Laboratory Procedure

Purpose: Drive library fragments to ISPs (beads) and populate the beads with empcr. Quantify the yield using the Qubit..

168 West Main Street, #922, New Market, Md Revision 5.2 6/6/2017

Revision 5.1 4/17/17

Purpose: Enrich ISPs, perform second Qubit assessment, Load Ion Torrent chip.

CHEMICAL ENGINEERING LABORATORY CHEG 4137W. Bioreactor

PAT940 Smart Titrator

User Guide. Part No.: BPLUS10. High Pass Plus TM Gel Cassettes. High Pass DNA Size Selection. Doc. #460045

An Oxidation-Reduction Titration: The Reaction of Fe 2+ and Ce 4+

CHEMICAL ENGINEERING SENIOR LABORATORY CHEG Bioreactor

User Guide BLF-7510 BUF High-Pass TM DNA Size Selection. Sage Science Part Nos:

Contamination Prevention and Decontamination

Aurora Reusable Cartridge Handling Manual

Affymetrix Processing of ChIP or DamID DNA to GeneChip Drosophila Tiling 2.0R Array

TruSeq Custom Amplicon Low Input Workflow on the MiniSeq System

minipcr TM Genes in Space Food Safety Lab: Mars Colony at Risk!

Techniques for Biotechnology!

Department of Laboratories St. Louis, MO ISSUE DATE: June 2007 REVISION DATE: May 2012 REVIEWED DATE: May 2018 PRINCIPLE:

User Guide PAC-20KB PAC-30KB. High-Pass TM DNA Size Selection. Sage Science Part Nos:

Pro live Owner s Manual. For Owner s Manual updates, warranty information, and support, please visit:

Aims of this case study

User Guide PAC-20KB PAC-30KB. High-Pass TM DNA Size Selection. of SMRTbell TM templates for PacBio RS Sequencing. Sage Science Part Nos:

Procedure for Use of the 3130XL Genetic Analyzer

TRACE 1300 and TRACE 1310 Gas Chromatographs. NoVent Microfluidics. Installation Guide Revision A March 2013

Vinmetrica SC-100A User Manual

TruSeq Custom Amplicon Low Input Dual Strand Workflow on the MiniSeq System

Please read and follow the instructions below carefully to ensure proper setup and avoid problems.

InDrop Single Cell Reverse Transcription

Investigating Photosynthesis Using Algal Balls

edna PROTOCOL SAMPLE COLLECTION Caren Goldberg and Katherine Strickler, Washington State University Revised November 2015

BIOSCAN* ATP method for biocide (toxicant) evaluations

ESSENTIAL FOR SURVIVAL

SUPER FRESHNESS METER KV-202. Instruction Manual

Hydro-wind Education Kit ASSEMBLY GUIDE

Ocean Water. Evaluation copy

isolight System Guide For Research Use Only. Not for use in diagnostic procedures.

PORTABLE ph METER PRICE CODE NUMBER MODEL 100

How to Set Up and Run Gel Electrophoresis

*This instrument is intended for laboratory use only

For detailed instructions about the TruSeq Custom Amplicon library preparation methods, refer to your reference guide.

WESTPAC Training Workshop Research and Monitoring of the Ecological Impacts of Ocean Acidification on Coral Reef Ecosystems

Reagent Kit Selection Guide

BioSpecimen Exchange for Neurological Disorders (BioSEND) Parkinsonism Biomarker Subtypes Training Webinar

Let s Move It! Gel Electrophoresis Using Food Dye Student Guide

Technical Bulletin No. MK3176. PREPARED: July, 2010

IDENTIFYING UNKNOWN SUBSTANCES

Agilent High Sensitivity DNA Kit Guide

Reagent Kit Guide RNA 6000 Nano Assay. Edition April 2003

CHEMICAL/BACTERIA KIT - FLUIDS

TaqMan Advanced mirna Assays

BioSpecimen Exchange for Neurological Disorders, BioSEND

L-JACK Termination Procedure

Agilent 3100 OFFGEL Fractionator Kit

Methods and Materials Manual Of Ion Chromatography Using DX-100 IC. Collecting Samples. LDPE bottle with cap (pre-rinsed with DI water)

To whom it may concern, February 21, 2011

ExcelGel 2-D Homogeneous Instructions

BIOSCAN* ATP method for monitoring of biofilm/sessile microorganisms in cooling systems

Learning Basic Laboratory Skills

ABI Prism 310 Cycle Sequencing Protocol Economy Version

Fast, easy, reliable, ultra-pure transfection grade plasmid DNA Maxiprep using a microcentrifuge spin-column.

Fast, easy, reliable, ultra-pure transfection grade plasmid DNA Maxiprep using a microcentrifuge spin-column.

Let s Move It! Gel Electrophoresis Using Food Dye Student Guide

edna PROTOCOL SAMPLE COLLECTION Caren Goldberg and Katherine Strickler, Washington State University Revised January 2017

QUICK REFERENCE GUIDE

Quick Guide. Lesson 1 Cheek Cell DNA Template Preparation

Calcium and Magnesium; Chlorophosphonazo Rapid Liquid Method Method to 1000 µg/l Ca and Mg as CaCO 3 (ULR) Pour-Thru Cell

DNA FINGERPRINTING 3/2003 TEACHER PREPARATION AND INSTRUCTION GUIDE FOR USE WITH CAROLINA BLU STAIN

MSES consultants, inc.

Electrophoresis 101 MiniLab

HT RNA Pico Sensitivity LabChip Kit LabChip GX/GXII User Guide

Electrophoresis 101 Student Worksheet

Table 163 Instrument-specific information. Instrument Pour-thru Kit Cell orientation Adapter. DR 6000 LQV Arrow faces right

Antibiotic-Resistant Bacteria

INSTRUCTION & REFERENCE MANUAL Model no. ODY-1720NX

EPK- EasyTest Patch Test Kit Manual

Protocol for BelloCell-500AP Operation ver.1.0

Recommended Procedure for Sampling Gastrointestinal Flexible Endoscopes

Maxwell RSC Cell DNA Purification Kit

Basic Biotechnology Kit

Western Blot Protocol

The fastest, easiest, most reliable method for purification of up to 1.2 mg of ultra-pure endotoxin-free plasmid DNA.

Maxwell RSC Whole Blood DNA Kit

Applied Biosystems QuantStudio 12K Flex Real- Time PCR System: OpenArray Plate

Dissolved Oxygen Probe (Order Code: DO-BTA)

PCR Detection of Genetically Modified (GM) Foods Protocol

Note: The syringe technique is ideal for water samples collected in sterilized media bottles and transported back to the lab for filtration.

Agilent 2100 Bioanalyzer (PSC 533)

ON-SITE ANALYSIS PIPELINE FLUIDS BACTERIA KIT

USER MANUAL. ExStik CL200A. Waterproof Total Residual Chlorine Tester. Patent Pending. Additional User Manual Translations available at

GroundStation Module User Manual V1.0

QUICK REFERENCE GUIDE. Easy operation fast results. Closer to Care.

Monoclonal antibodies

Set-up, programming and analysis MagNA Pure 96 (Roche)

Transcription:

Water-Water Titration for the MicroCal ITC200 Goal: Water-Water titrations are a good way to check for possible problems, especially the cleanliness of the ITC and your skills in cell filling. If a water-water titration works well the instrument is in good working order and your cell filling procedure is correct. Getting Started: 1. The computer and the ITC200 are normally left on. If they have been switched off for any reason makes sure that you boot them in the order shown below. If you don't, you may encounter communication problems between computer and ITC. Turn on the ITC 200 computer. Turn on the Micro-Cal ITC200. Launch the ITC 200 with Origin software. 2. The software will establish a connection with the instrument, and the instrument will then heat the cell to 30 C. You can change the temperature setting right away to 25 C. This will save you time later. 3. Under the Instrument Controls tab (see figure below) change the Set Point temperature to 25 C and click "Set Jacket Temp". 4. Fill 5 to 10 ml fresh MiliQ water into a clean 15 ml conical tube. 1

5. If you are using the water-water titration for trouble shooting you should replace the water in the water wash bottle and the contense of the buffer wash bottle with fresh MiliQ water and refreshen the Methanol in the Methanol wash bottle. Loading Sample Cell and Syringe: Preparing the ITC 200 for sample loading 1. Remove the white cap from the cell and insert the cleaning apparatus into the sample cell. Make certain that the cleaning apparatus is pushed firmly into the sample cell. 2. Attach the fill adapter to the injection syringe. To do this the syringe should be firmly seated in the wash position. Align the inner and outer ring so that you can see the hole the connector is going to be screwed into. Stabilize the aligned assemblies with your fingers and screw the connector in. It should fit snugly. If it is too loose the drying of the syringe might not work correctly leaving Methanol in the syringe and the sample will not be drawn into the syringe right. If you overtighten it you will break the syringe! (Note: This is how the instrument should have been left by the previous user) 2

3. Click on the Instrument Controls tab. (See Figure above) 4. Under the Instrument Controls Washing Module, click on the "Cell and Syringe Wash" button. The Wash Module will first flush the cell with water. Once it is done with the cell it will sound a small beep. After that it will flush the syringe with water and methanol and dry it with air. Loading the Sample Cell: 5. After the beep, while the syringe wash is going on you can start filling the cell. 6. Use the dedicated loading syringe to fill/empty the cell. It has a protective sleeve over the needle to avoid scratching the sample cell. Do not use any other syringe!!!! 7. Insert the empty loading syringe into to the cell until it gently touches the bottom. Pull the syringe up a millimeter so that it is no longer pressing against the bottom. Remove any remaining liquid from the cell. 8. Wash the loading syringe 3 times with fresh, clean MiliQ water. 9. Fill the sample cell with fresh MiliQ water from the 15 ml conical. To fill the cell, slowly draw up ~300μl water into the loading syringe. Avoid air bubbles. If you inject air bubbles into the sample cell your experiment will not work! Slowly insert the syringe into the cell until the tip gently touches the bottom of the cell. Pull the syringe up slightly so that it is no longer pressing against the bottom. Begin to inject the water into the cell. Once you have injected ~150 μl into the cell, inject the rest of solution in short pulses to dislodge any bubbles that may be clinging to the walls of the cell. 3

Continue in this manner until you see the water solution well up around the syringe. Draw the syringe out of the cell and gently place it on the lip between the plastic collar and metal of the cell. Draw up any excess liquid. Soak the cell in the water for 2 minutes. 10. Remove the water from the sample cell using the loading syringe and discard the solution. 11. Refill the sample cell with fresh water from the 15 ml conical in the same way as described above. Note: If you do the water-water titration for trouble shooting you should refill the reference cell with fresh MiliQ water as well. Loading the titration syringe: 12. Fill a 200 μl PCR tube with 80 μl water out of the same 15 ml conical. Do not close the lid. Insert the filled, open tube into the tube holder. Make sure it sits all the way down with the lid fitting into the slot provided (see photo below). 13. To fill the syringe, go to the Instrument Controls tab and click Syringe Fill. 14. Follow the prompts. You will be asked to place the syringe pipette into the rest position. After that the plunger of the syringe will be moved down. 4

Next you will be asked to place the syringe into the loading position. Be careful when doing this and make sure that the needle does not get bent when inserted into the PCR tube with the titrant. Next, the washing module will fill the syringe. Observe the process carefully. If a large air bubble remains on the top of the solution you can restart the loading process but leave the syringe in the loading position for all steps. This will most likely eliminate the bubble. 15. Disconnect the fill adapter from the syringe. 16. Insert the syringe pipette into the cell. Rest position PCR tube with titrant Load position 5

Setting Run Parameters 17. Click on the Advanced Experimental Design tab and enter the parameters shown below in the indicated fields. Don't forget to change the filename. 18. Click the Setup tab and change the path for data storage to your folder. 6

Starting the Run: 19. Switch off the overhead light if you used it. 20. Double check that the fill adapter is disconnected and press the start button. 21. The system will adjust to the chosen temperature. 22. Once the temperature is reached, the inner assembly of the titration syringe will begin to spin. 23. The system will adjust the reference power. 24. Once the reference power is stabilized the run will begin. Note: If it takes much longer than 10 min to adjust the reference power, something might not be quite right. 25. Check that you switched off the overhead light (it will heat the cell from above, making it harder to stabilize the reference power). 26. You can consider reloading the cell, since air bubbles might affect the stabilization of the reference power. To do that, stop the run, return the syringe pipette into load position, and remove the sample from the cell using the sample loading syringe. Reload the cell, following the procedure above. 7

Result: The plot below shows a typical result for a normal water-water titration. Clean up after the run: 27. Remove the syringe pipette from the cell and insert firmly into the wash position. 28. Attach the fill adapter to the syringe pipette as described above. 29. Remove the solution from the cell using the loading syringe. You can keep the solution to double check the concentration. 30. Insert the long needle of the cell cleaning apparatus into the sample cell and push down carefully until the O-ring has sealed. 31. Go to the instrument control tab and preform a "cell and syringe wash". 32. Leave the instrument in this configuration. If you run into any problems or have questions about ITC let Dr.Erbse know. Annette Erbse, Office C316, phone 2-0528, erbse@colorado.edu 8