If All Else Fails Read the Directions. Disclosure. Manufacturer s Instructions. Presenter: Sue Lafferty CHICA Canada Conference, May 2010

Similar documents
Medical Device Reprocessing Assessment - General (Cleaning, Disinfection, and/or Sterilization)

6 Handling, collection, and transport of contaminated items General rationale Separation of waste and reusable items at point of

6 Handling, collection, and transport of contaminated items General rationale Separation of waste and reusable items at point of

New National Standard of Canada CAN/CSA-Z Canadian medical device reprocessing

Objectives. The Perioperative Nurse s Role 6/24/2016. Instrument Processing & Sterilization Chesapeake Perioperative Consortium 2016

Ideal Sterilization Method. Sterilization. Sterilization Practices in Healthcare Facilities

Manner of Assessment Code (check all that apply) & Surveyor Notes. Yes N/A. Yes. Yes N/A N/A. Yes N/A. N/A Yes N/A

3/26/2014 SPAULDING CLASSIFICATION OBJECTIVES PRINCIPLES OF DISINFECTION AND STERILIZATION IN THE DENTAL SETTING PROCESSING SEMI CRITICAL OBJECTS

PRINCIPLES OF DISINFECTION AND STERILIZATION IN THE DENTAL SETTING OBJECTIVES

applying a unique combination of disciplines consulting, preclinical, toxicology,

QUALITY ASSURANCE IN AN MDRD

Flexible Endoscopes: The A,B,C s of Monitoring Manual Cleaning Efficacy!

Clostridium difficile Update Dr. Michelle Alfa, Winnipeg Sponsored by ARJO

MDRAO ONLINE COURSE FOURTH EDITION 2018

Understanding ANSI/AAMI ST 91. Flexible and Semi-Rigid Endoscope Reprocessing in Healthcare Facilities

CSA Z314 Series Standards Understanding the Standards

Monitoring Flexible Endoscope Channels to Assure Cleaning

ANSI/AAMI ST79 A2:2009 Key Changes in the 2009 Amendment

How is Sterilization Validation Performed by the Medical Device Manufacturer?

Is Your Sterile Processing Department Ready for Survey?

THE TERMS VALIDATION AND VERIFICATION ARE FREQUENTLY

THE BASICS OF STERILIZATION

Reprocessing of Single Use Devices (SUDs) MEDEC Regulatory Conference May 9, 2016

Christina Bradley Endoscope Decontamination A Webber Training Teleclass. Christina Bradley

July 17, 2013 SOURCE: IPC Standards and Projects

disinfectionandsterilization.org

3034 Owen Drive Antioch, TN USA Fax: Maybachstraße 10. Revised 01/17

Lotus. Reprocessing instructions for reusable instrumentation

Overview: High Level Disinfection

Lotus. Reprocessing instructions for reusable instrumentation. French/Swiss protocol

Low Temperature Processing. An Update on Guidelines, Standards and Requirements

8/23/2012. Continuing Education Contact Hours. How much of your formal education focused on sterilization or sterile processing?

Principles of Disinfection and Sterilization in the outpatient setting

Principles of Disinfection and Sterilization in the outpatient setting

Competency Guide: Care and Handling of Rigid Endoscopes

Validation of cleaning and sterilization of reusable medical devices

ARABIC CANADIAN MEDICAL CENTER CSSD POLICIES AND PROCEDURES

Administrative Policies and Procedures. Cross Reference: Date Issued: 2/14 Date Reviewed: 9/14 Date: Revised: 12/14 Attachment: None Page of 1 of 6

6/12/2016. Objectives. Manufacturers Instructions for Use: Impact on Patient Safety. Background

3.03 Understand support services

St Boniface Research Centre Winnipeg, Manitoba Canada

MDRAO Importance of Bedside & Manual Precleaning CONMED.CA

Disinfection & Sterilization in the Ambulatory Surgery Center

CLEANING AND HANDLING OF MICROPORT INSTRUMENTS The following languages are included in this packet:

3.03 Understand support services

*Slides updated to include newer guidelines in February 2016 by Shelby Lassiter, BSN, RN, CPHQ, CIC

Loaner Programme Guidance

Loaner Programme Guidance

Facility: Sterile Processing Assessment Date: CRITERIA ANSI/AAMI COMPLIANT Number Y N N/A Point of Use Gross contaminant is removed with

Sterile Services and Theatres Working in Harmony

Guide for Cleaning, Sterilization and Storage of Reusable Introducers

CDWA3 / CDWA4 CDWAH. Page 4. BASKET ZONE: water pressure, spray system performance, detergent chemistry and concentration, temperature.

FREQUENTLY ASKED QUESTIONS IPAC Lapses

Not to be reproduced - AGPAL & QIP

Microfibre cleaning: is it all its been cracked up to be?

Best Practices for Reprocessing Endoscopes

DUAL SPECIES BIOLOGICAL INDICATOR SPORE STRIPS For Monitoring Ethylene Oxide (EO), Steam and Chemiclave Processes

Surgical Instruments - Instructions for use

CARE SERVICE USER GUIDE. Sonicision System

STERILIZATION VALIDATION IN THIS SECTION REPROCESSING VALIDATIONS FOR REUSABLE MEDICAL DEVICES. Radiation Sterilization Validation

Monitoring your decontamination and reprocessing cycle

Guide for Cleaning, Sterilization and Storage of Introducers for Caldera Medical Implants

Amalgatome SD Sterilization Tray User Guide

House Keeping. Introduction to ANSI/AAMI ST91:2015: Flexible and Semi-Rigid Endoscope Processing in Health Care Facilities. Welcome!

Water for Instrument Processing

Products made in Switzerland

B. The Facility Manager is responsible for ensuring all staff are adequately trained and experienced in autoclave sterilization procedures.

Disinfection & Sterilization

Putting Confidence in Your Hands.

Loaner Program Guidance

The Reusable Surgical Instrument Cycle

ISO Part 1 and Part 2 Compliance Requirements. Cathriona O Neill

Medical Device Reprocessing Quality Assurance. 3M All Rights Reserved.

Steam Sterilization Monitoring in the New CSA Standard (Z )

HANDLING OF WRIGHT DISPOSABLE PROPHECY ANKLE INSTRUMENTS

Best Practices for High Level Disinfection Part I

Best Practice Monitoring for Sterile Processing Professionals

Manufacturer s IFUs. What Are They and Why Are They So Important?

DEFENDO. Sterile Single-use Valves. Procedure

Quality Control EQUIPMENT CONTROL. Monitor critical performance characteristics of sterilization, decontamination or cleaning equipment

Choose a category. You will be given the answer. You must give the correct question. Click to begin. Instructions for using this template.

Sterilization Service Revolution Service Enhancement in HA Hospitals

8/27/2015. Learning Objectives. Purpose of Sterilization. To destroy all living pathogenic microorganisms

3/11/ M Sterile U Network 3M Sterile U Web Meeting March 11, Welcome! Topic: An Overview of ANSI/AAMI ST58: 2013

Entellus Medical Sinuscope INSTRUCTIONS FOR USE

Sterile Packaging & The End User

RIGID ENDOSCOPE OPERATING INSTRUCTIONS

Welcome! How do I get a CE Certificate?

Standards of Infection Prevention in Reprocessing Flexible Gastrointestinal Endoscopes Webinar FAQ

Endoscopy and OR Solutions SYSTEM 1 TM. Express STERILE PROCESSING SYSTEM SAFETY, ASSURANCE AND EFFICIENCY. One Integrated Approach to Healthcare

SUBJECT: Detection of Blood (hemoglobin) residue inside the biopsy channel of an endoendoscope

Overview and Introduction Annex 1 Revision

Recommendations for Physical Space for Decontamination Areas

What You Need CDHO's website, RCDSO's website, advisors, PHO's website (Grand Rounds, Training) Infection Prevention and Control February 20, 2018

Obtaining Manufacturer s IFU: Challenges and Barriers

BIOTECH-GERMANDE ENDOSCOPE REPROCESSING WITH BRUSHLESS CLEANING PROCESS CONDITIONS. Lionel PINEAU PhD

Loaner Instrumentation

What Dentistry Can Learn from AAMI ST79. Best Practices for device and instrument reprocessing

PRODUCT CATALOGUE. for gke Steri-Record and gke Clean-Record Chemical indicators, PCDs and accessories for cleaning and sterilization

PART 5. Sterilisation Consumables RFTHPV

Transcription:

If All Else Fails Read the Directions The Importance of Medical Device Manufacturer s Instructions Presenter: Sue Lafferty CHICA Canada Conference, May 2010 Disclosure Member of 3M Canada Speakers Bureau: Received honorarium for speaking at Medical Device Reprocessing Seminars 2 3 1

Manufacturer has received more than 47 reports of cribs being mis-assembled with the mattress platform being used as a side rail, including 27 reports of babies becoming entrapped, resulting in one death. 4 An eyelid retractor made from bending a 3.2-cm paper-clip Retraction of post-traumatic swollen eyelids with a pair of paperclips. Paper-clip eyelid retractors are simple, cheap, readily available selfmade eyelid retractors, and are helpful in opening closed oedematous eyelids in both trauma and postoperative situations. No patients suffered from any undue pressure or complications from the paper-clip eyelid retractors. Cheng LHH, Kumar P.Ann R Coll Surg Engl. 2008 April; 90(3): 253. 5 Reamer and guide wire following reprocessing 6 2

Objectives: Following this presentation participants will: State the importance of obtaining and following manufacturer s instructions Have an awareness of requirements for manufacturer s instructions List 3 tools that can be used when requesting and reviewing instructions 7 Standards, Guidelines and Regulations Requiring Manufacturer s Instructions Canadian Standards Association sterilization standards Accreditation Canada PIDAC guidelines Health Canada Provincial Departments of Health 8 Definition: Manufacturer s Instructions The written directions provided by the manufacturer or distributor of a product that contain the necessary information for the safe and effective use of the product. CSA Z314.8-08 Decontamination of Reusable Medical Devices 9 3

Definition: Validation Manufacturer s instructions must be validated Validation: a documented procedure for obtaining, recording and interpreting i the results required to establish that a process will consistently yield product complying with predetermined specifications. CSA Z314.3-09 Effective sterilization in health care facilities by the steam process. 10 Also Known As: IFUs-Instructions for use MIFUs-Manufacturer s instructions for use DFUs-Directions for use MDFUs-Manufacturer s directions for use 11 Why MIFUs are required? 12 4

Medical devices have become increasingly complex 13 14 Disinfectants, sterilants and cleaning solutions, if used incorrectly, can damage devices or result in a contaminated device 15 5

Cleaning, disinfection and sterilization equipment can malfunction if improperly installed, operated or monitored *Bed pan washer-disinfector-hot and cold water feed lines were reversed and sprayer arms were improperly installed which lead to inefficient cleaning by the equipment. Ultra-sonic cleaning equipment was installed and assumed to be working correctly. Testing, when done, revealed that 2 of the 4 transducers were not functioning. **Outbreak of surgical complications (TASS) caused by a steam sterilizer that was improperly maintained according to MIFU. *Alfa MJ, Olson N, Buelow-Smith L. Simulated-use testing of bedpan and urinal washer disinfectors: Evaluation of Clostridium difficile spore survival and cleaning efficacy. AJIC, Feb. 2008, Vol. 36, No.1: 5-11. **Hellinger WC, Hasan SA, Bacalis LP et al. Outbreak of toxic anterior segment syndrome following cataract surgery associated with impurities with autoclave steam moisture. ICHE. March 2006, Vol. 27, No. 3: 294-298. 16 17 Health Canada Notice Hospitals shall have in place: Procedures to ensure that reusable devices are cleaned, disinfected and sterilized according to manufacturer s instructions. A mechanism to regularly review these procedures and ensure that they are being followed. A requirement, at the time of purchase, that manufacturers include complete instructions and, where necessary, adequate training for the cleaning disinfecting and sterilizing of reusable devices. A procedure to report to Health Canada any cases in which the manufacturer does not provide adequate instructions. Health Canada Health Products and Food Branch. Notice to Hospitals: Inadequate cleaning/sterilization of EZ clean monobloc acetabular reamers a reuseable medical device, April 21, 2004 18 6

Important for: Reusable medical devices Products used to reprocess medical devices (e.g. cleaning and disinfectant products) Equipment used to reprocess medical devices Products used for disinfectant or sterilization monitoring Packaging for medical devices (e.g. wrappers, pouches, containers) 19 Format for MIFUs Product Labels Product Inserts User manuals Electronic or printed material Official updates to original instructions (e.g. responses to product alerts, ongoing validation studies) 20 MIFUs for Reusable Medical Devices: Device-specific, legible and understandable Clearly indicate which parts need to be disassembled Can be achieved within the facility s resources In accordance with intended use of the device State if device is immersible Specify sterilization cycles (method, time,temperature) Specify if there is a limit to # of times a device can be reprocessed or if reprocessing will degrade the device. CSA Z414.3 Effective Sterilization in Health Care Facilities by the Steam Process, 2009 21 7

Reusable Medical Device Instructions Validated cleaning methods and reprocessing instructions Soaking Manual or Automated Cleaning (rinsing, drying) Lubrication Assembly/wrapping Schematic drawings Provision of in-service to staff reprocessing the device Routine maintenance schedules 22 Disinfectant Products Require a Drug Identification Number (DIN) or Natural Product Number (NPN) from Health Canada 23 Precaution for use of product Storage Vapour control PPE Disposal Accidental exposure spill F/U List of active ingredients Disinfectant Label Information Microbicidal claims: list of organisms killed Manufacturer Product X Manufacturer contact information Expiration Date Mixing or Opening Date DIN/NPN Directions for use: Preparation: dilution or activation or RTU Limitations of use (e.g. only in specific AER) Compatible materials Level of disinfection provided Concentration, temperature, contact time, activation, efficacy testing Rinsing: volume & type of water 24 8

Cleaning Products Instrument detergents Enzymatic cleaning agents Cleaning accessories e.g. brushes, sponges, etc. Note: Cleaning products do not require a Drug Identification Number (DIN) 25 26 Equipment Used in Reprocessing Ultrasonic cleaners Automated lumen cleaners Packaging heat sealers Washer-disinfectors Automated endoscope reprocessors (AERs) Automated drying equipment/cabinets Sterilizers 27 9

Equipment requirements/instructions Medical Device License from Health Canada for Class II and higher Establishment License for Class 1 Devices If the equipment (e.g. sterilizer, AER) requires a chemical product, the chemical must have a DIN Installation instructions What types of devices the equipment has been validated to process (e.g. lumened, porous devices) Methods to test that equipment is operating according to its specifications Schedule and details of preventive maintenance program 28 Sterilization Packaging: Single use or reusable textile wraps Sterilization pouches Rigid instrument container or container system 29 Packaging Requirements and MIFUs Evidence of: Medical Device License Compatibility with sterilization method and devices to be sterilized Heat/chemical resistant Allows for sterilant penetration Instructions for closure or sealing Ability to aseptically dispense package contents Sealing, flexibility Prevents contamination during storage/handling Intact seal, strength, shelf life data Lint-free and particulate-free Non toxic/non leaching Dyes and labels Reprocessing & Maintenance for reusable containers or wraps 30 10

Disinfectant/Sterilization Monitoring Test Strips to measure efficacy of disinfectants Biological Indicators Chemical Indicators 31 Disinfectant Test Strips Specific to the disinfectant and the minimum effective concentration of the disinfectant Quality assurance procedure to ensure each batch of strips will perform effectively Clear criteria for determining pass/fail of disinfectant concentration 32 Biological Indicators: Appropriate for the method of sterilization e.g. steam: Geobacillus stearothermophilus, ETO: Bacillus atrophaeus Clear directions for placement in sterilizer Detailed instructions for handling Method of incubation Clear criteria for indication of positive or negative result 33 11

Chemical Indicators Appropriate for the method of sterilization Class of indicator (1-6) What variables does the indicator measure e.g. steam: time, temperature, pressure? What are the stated values or endpoints measured? Clear criteria for pass/fail result (ease of reading) 34 How to Obtain MIFUs Ensure it is a requirement prior to purchase of any instrument, equipment, reprocessing product or device Request from distributor, vendor or manufacturer Subscribe to a commercially available medical device manufacturer s instruction data base Ensure your organization has a process to review or receive manufacturer s alerts, cautions, revisions or recalls of product and that you are on the distribution list 35 Problems with Manufacturer s Instructions 36 12

Instructions may be unclear, nonspecific and vague 37 Instructions may be unclear due to poor or inaccurate translation from another language OR applicable to standards in another country. 38 Instructions may be generic and not specific to the device, equipment or product. One size does not fit all 39 13

Instructions may be written in technical jargon or terms not well understood by users 40 Tools to Assist You CSA/ISO 17664 Sterilization of Medical Devices - Information to be Provided by the Manufacturer for the Processing of Resterilizable Medical Devices Health Canada DRAFT GUIDANCE DOCUMENT: Information to Be Provided by Manufacturers for the Reprocessing and Sterilization of Reusable Medical Devices, December 2006 http://www.hc-sc.gc.ca/dhp-mps/md-im/applicdemande/guide-ld/md_dgd_manrep_im_eld_fabreteng.php 41 Tools: MEDEC Checklist for Medical Device Manufacturer s Validated Instructions for Reprocessing of Reusable Medical Devices, January 2010 42 14

43 44 45 15

If Problems Encountered Notify responsible/affected departments in your organization Notify the manufacturer Notify Health Canada Medical Devices http://www.hc-sc.gc.ca/dhp-mps/compli-conform/probreport-rapport/rep_md_prob-rap_inc_im_tc-tm-eng.php 46 Take Home Message MIFUs are essential to effective and safe reprocessing of reusable medical devices MIFUs are not always readily available but can be obtained Make no assumptions: Know and communicate specifically to the manufacturer what is required 47 Thank You 48 16