Application Note. Author. Abstract. Small Molecule Pharmaceuticals. Sonja Krieger Agilent Technologies, Inc. Waldbronn, Germany

Similar documents
Method Transfer from an Agilent 1200 Series LC to an Agilent 1260 Infinity II LC

Method Transfer from an Agilent 1260 Infinity LC to an Agilent 1260 Infinity II LC

Emulation of the Agilent 1100 Series LC Through Waters Empower Software Analysis of an Analgesic Mixture

Seamless Method Transfer from an Agilent 1260 Infinity Bio-inert LC to an Agilent 1260 Infinity II Bio-inert LC

Reducing Cycle Time for Charge Variant Analysis of Monoclonal Antibodies

Automated Scouting of Stationary and Mobile Phases Using the Agilent 1290 Infinity II Method Development Solution

Agilent 1290 Infinity Binary LC System with ISET Emulation of a Waters Alliance 2695 LC Applying Concave, Convex, and Linear Gradients

Agilent 1290 Infinity Binary LC System with ISET - Emulation of the Waters Alliance 2695 LC system analyzing b-blockers

Technical Overview. Author. Abstract. A.G.Huesgen Agilent Technologies, Inc. Waldbronn, Germany

Agilent 1290 Infinity Quaternary LC Support of Columns with 2.1 to 4.6 mm ID to 1200 bar

Agilent 1290 Infinity LC with ISET under Waters Empower control Emulation of Agilent 1100 Series LC

Protein Separation with ph Gradients Using Composite Buffer Systems Calculated by the Agilent Buffer Advisor Software

Fraction Analysis of Cysteine Linked Antibody-Drug Conjugates Using Hydrophobic Interaction. chromatography. Agilent 1260 Infinity II Bio-Inert System

Determination of Asarinin in Xixin (Asari Radix Et Rhizoma)

Application Note. Author. Abstract. Pharmaceuticals. A.G.Huesgen Agilent Technologies, Inc. Waldbronn, Germany

Technical Overview. Author. Abstract. Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany

High-Throughput LC/MS Purification of Pharmaceutical Impurities

Disulfide Linkage Analysis of IgG1 using an Agilent 1260 Infinity Bio inert LC System with an Agilent ZORBAX RRHD Diphenyl sub 2 µm Column

ISET. Agilent 1290 Infinity Binary LC with ISET, emulation of the Waters Alliance 2695 LC system analyzing aromatic acids. Application Note.

Comparing gradient transfer of isocratic hold and delay volume addition using the Agilent 1290 Infinity LC with ISET

Performance characteristics of the 1260 Infinity Quaternary LC system

Application Note. Author. Abstract. Pharmaceuticals. Detlef Wilhelm ANATOX GmbH & Co. KG. Fuerstenwalde, Germany mau

Application Note. Environmental. Agilent 1290 Infinity Binary LC. with ISET and fine-tuning. Agilent 1290 Infinity Binary LC.

Performance Characteristics of the Agilent 1220 Infinity Gradient LC system

Charge Heterogeneity Analysis of Rituximab Innovator and Biosimilar mabs

Benefits of 2D-LC/MS/MS in Pharmaceutical Bioanalytics

Fast Method Development Using the Agilent 1290 Infinity Quaternary LC System with Column Selection Valve

Application Note. Author. Abstract. Biopharmaceuticals. Verified for Agilent 1260 Infinity II LC Bio-inert System. Sonja Schneider

High-resolution Analysis of Charge Heterogeneity in Monoclonal Antibodies Using ph-gradient Cation Exchange Chromatography

Molecular Characterization of Biotherapeutics The Agilent 1260 Infi nity Multi-Detector Bio-SEC Solution with Advanced Light Scattering Detection

Quality-by-Design-Based Method Development Using an Agilent 1290 Infinity II LC

Developing Purification Strategies for the Agilent 1260 Infinity II Preparative LC/MSD System

Improved Sensitivity with the Agilent 1290 Infinity Evaporative Light Scattering Detector (ELSD)

Clone Selection Using the Agilent 1290 Infinity Online 2D-LC/MS Solution

Minimizing Sample Carryover Using the Multiwash Function of the Agilent 1290 Infinity II Multisampler

Application Note. Author. Abstract. Biotherapeutics and Biologics. Sonja Schneider Agilent Technologies, Inc. Waldbronn, Germany

Peptide Mapping: A Quality by Design (QbD) Approach

Application Note. Authors. Abstract. Biopharmaceuticals

Application Note. Author. Abstract. Small Molecule Pharmaceuticals & Generics. A.G.Huesgen Agilent Technologies, Inc. Waldbronn, Germany

Application Note. Authors. Abstract. Introduction. Pharmaceutical

Technical Overview. Authors. Abstract. Verified for Agilent 1260 Infinity II LC Bio-inert System

Robustness of the Agilent Ultivo Triple Quadrupole LC/MS for Routine Analysis in Food Safety

Technical Overview. Author. Abstract. A.G.Huesgen Agilent Technologies, Inc. Waldbronn, Germany

Analysis of Nucleosides Using an Agilent Infinity II High Speed UHPLC with the 6130 Single Quadrupole Mass Selective Detector

Characterize mab Charge Variants by Cation-Exchange Chromatography

Size Exclusion Chromatography of Biosimilar and Innovator Insulin Using the Agilent AdvanceBio SEC column

Analytical Instrument Qualification According to USP <1058>: Requirements and Examples for the Agilent 1290 Infinity LC System

A Comprehensive Workflow to Optimize and Execute Protein Aggregate Studies

Fast, Low Pressure Analysis of Food and Beverage Additives Using a Superficially Porous Agilent Poroshell 120 EC-C18 Column

Fast Analysis of Environmental Phenols with Agilent Poroshell 120 EC-C18 Columns

Fast and High Resolution Analysis of Intact and Reduced Therapeutic Monoclonal Antibodies (mabs)

Separate and Quantify Rituximab Aggregates and Fragments with High-Resolution SEC

Agilent 1260 Infinity II Prime LC ACHIEVE CONVENIENCE EVERY DAY IN EVERY WAY

Analysis of amoxicillin and five impurities on the Agilent 1220 Infinity LC System

Characterization of mab aggregation using a Cary 60 UV-Vis Spectrophotometer and the Agilent 1260 Infinity LC system

Scale-up with the Agilent SD-1 Purification System analytical and preparative runs on a single system

High-throughput and Sensitive Size Exclusion Chromatography (SEC) of Biologics Using Agilent AdvanceBio SEC Columns

Method Translation in Liquid Chromatography

Reversed-phase Separation of Intact Monoclonal Antibodies Using Agilent ZORBAX Rapid Resolution High Definition 300SB-C8 1.

Analysis of TCM injections using the Agilent 1290 Infinity LC system

Automated QbD-Based Method Development and Validation of Oxidative Degraded Atorvastatin

EPA 8330A Analysis of Explosives Using Agilent Poroshell 120 EC-CN and EC-C18 Columns

New Advances in UHPLC -Resolution, Speed & Sensitivity

Automated alternating column regeneration on the Agilent 1290 Infinity LC

AdvanceBio HIC: a Hydrophobic HPLC Column for Monoclonal Antibody (mab) Variant Analysis

AGILENT INFINITYLAB. Maximum efficiency for your liquid chromatography workflows

Techniques for Reducing the Effects of Sample Solvent on UHPLC Analyses

Peptide Mapping. Hardware and Column Optimization

Application of Agilent AdvanceBio Desalting-RP Cartridges for LC/MS Analysis of mabs A One- and Two-dimensional LC/MS Study

Application Note. Author. Abstract. Environmental. Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany

Agilent 1260 Infinity Bio-inert Quaternary LC System. Infinitely better for bio-molecule analysis

Quantification of genotoxic "Impurity D" in Atenolol by LC/ESI/MS/MS with Agilent 1200 Series RRLC and 6410B Triple Quadrupole LC/MS

Rapid UHPLC Analysis of Reduced Monoclonal Antibodies using an Agilent ZORBAX Rapid Resolution High Definition (RRHD) 300SB-C8 Column

Characterize Fab and Fc Fragments by Cation-Exchange Chromatography

Simplifying Methods Transfer: Novel Tools for Replicating Your Established Methods on an ACQUITY Arc System

Application Note. Biopharma. Authors. Abstract. James Martosella, Phu Duong Agilent Technologies, Inc Centreville Rd Wilmington, DE 19808

Agilent 1290 Infinity II 2D-LC Solution Biopharmaceutical Polymer Analysis. WCBP Jan 2017 Washington, DC

EPA Method 543: Selected Organic Contaminants by Online SPE LC/MS/MS Using the Agilent Flexible Cube

Controlling Agilent 1200 Series Rapid Resolution LC systems through Waters Empower chromatography data software. Technical Overview.

Agilent 1290 Infinity II LC SET NEW BENCHMARKS IN EFFICIENCY WITH THE NEXT GENERATION OF UHPLC

The Agilent 1260 Infinity BioInert Quaternary Pump. Scope of a low-pressure mixing UHPLC pump with Bio-Inert Capabilities

Improving Retention Time Precision and Chromatography of Early Eluting Peptides with Acetonitrile/Water Blends as Solvent B

Agilent 1220 Infinity II LC GET ON THE FAST TRACK TO HIGHEST EFFICIENCY AT A PRICE YOU CAN AFFORD

Analytical HPLC solutions

Agilent 1220 Infinity LC. Infinitely more affordable

Intact & Subunit Purity

Fast and High-Resolution Reversed-Phase Separation of Synthetic Oligonucleotides

Agilent AdvanceBio SEC Columns for Aggregate Analysis: Instrument Compatibility

Sean M. McCarthy and Martin Gilar Waters Corporation, Milford, MA, U.S. INTRODUCTION EXPERIMENTAL RESULTS AND DISCUSSION

Monitoring of Mammalian Cell Culture Media with HILIC LC/MS

Agilent 1260 Infinity II LC ASPIRE TO NEW LEVELS OF EFFICIENCY EVERY DAY IN EVERY WAY

Agilent 1290 Infinity II LC MORE EFFICIENCY MORE FREE TIME. #EfficientUHPLC

NISTmAb characterization with a high-performance RP chromatography column

Cell Clone Selection Using the Agilent Bio-Monolith Protein A Column and LC/MS

Agilent 218 Purification System. Purify your way

Agilent 1260 Infinity LC & LC/MS Purification Systems. Purify your way

Heart-cut 2D-LC/MS approach for pharmaceutical impurity identification using an Agilent 6540 Q-TOF LC/MS System

Ultrapure water: LC-MS suitability tests

Qualification of High-Performance Liquid Chromatography Systems

Transferring a Method from a Conventional 3.5 µm or 5 µm to an Agilent Poroshell 120 Column to Improve Analytical Speed and Resolution

Transcription:

Method Transfer from an Agilent 1100 Series Quaternary LC to an Agilent 1260 Infinity II LC Proof of Equivalency for the Analysis of Antihistaminic Drugs Application Note Small Molecule Pharmaceuticals Author Sonja Krieger Agilent Technologies, Inc. Waldbronn, Germany Abstract The transfer of conventional LC methods from legacy equipment to new instrumentation is an important topic in all laboratories. This Application Note shows the transfer of a conventional LC method for the analysis of antihistaminic drugs from an Agilent 1100 Series Quaternary LC to an Agilent 1260 Infi nity II LC, and proves that equivalent results in terms of retention times and resolution are obtained. Also, the conventional LC method is transferred to UHPLC conditions optimized for resolution as well as for speed using the 1260 Infi nity II LC. This offers the possibility of increasing peak resolution, while enabling time and solvent savings. Agilent 1260 Infinity II LC Agilent 1100 Series LC

Introduction Instrument-to-instrument method transfer is an important topic for all laboratories throughout different industries 1. Especially for validated methods in the pharmaceutical industry, instrument-to-instrument method transfer is compulsory, but it is also important in QA/QC in other industries. One example of instrument-to-instrument method transfer is the transfer of conventional LC methods from older equipment such as the Agilent 1100 Series Quaternary LC, to new instruments such as the Agilent 1260 Infinity II LC. This Application Note shows the analysis of antihistaminic drugs using a 1100 Series Quaternary LC. The method is transferred to the 1260 Infinity II LC, and equivalent results in terms of retention time and resolution are obtained. With a pressure range of up to 600 bar, the 1260 Infinity II LC enables performing UHPLC analyses using Agilent InfinityLab Poroshell columns. The transfer of the conventional LC analysis of antihistaminic drugs to UHPLC conditions, optimized for resolution as well as for speed, will be shown. Histamine is an endogenous biogenic amine that acts as a neurotransmitter in the nervous system and as a local mediator in the gut, skin, and immune system 2. It evokes complex physiological changes that occur through four receptors; the first histamine receptor (H1 receptor) is widely expressed throughout the body, and is responsible for most of the symptoms caused by histamine in allergies 2. Antihistaminic drugs (H1 receptor antagonists) have long been used in the treatment of nausea, vomiting, and allergies 2. Experimental Instrumentation The Agilent 1260 Infinity II LC comprised the following modules: Agilent 1260 Infinity II Quaternary Pump (G7111B) Agilent 1260 Infinity II Vialsampler (G7129A) with integrated column compartment, 3.0 µl heater (Option #063) and sample cooler (Option #100) Agilent 1260 Infinity II Diode Array Detector WR (G7115A) with standard 10-mm flow cell, (G1315-60022) The Agilent 1100 Series Quaternary LC comprised the following modules: Agilent 1100 Quaternary Pump (G1311A) Agilent 1100 Degasser (G1379A) Agilent 1100 Autosampler (G1313A) Agilent 1100 Thermostatted Column Compartment (G1316A) Agilent 1100 Diode Array Detector (G1315B) with standard flow cell, 10 mm (G1315-60022) Software Agilent OpenLAB CDS Version 2.1 (availability planned for September 2016) Columns Agilent ZORBAX SB-C18, 4.6 150 mm, 5 µm (p/n 883975-902) Agilent InfinityLab Poroshell 120 EC-C18, 3.0 150 mm, 2.7 µm (p/n 693975-302T) Agilent InfinityLab Poroshell 120 EC-C18, 3.0 50 mm, 2.7 µm (p/n 699975-302T) Chemicals All solvents were LC grade. Acetonitrile was purchased from Merck (Darmstadt, Germany). Fresh ultrapure water was obtained from a Milli-Q Integral system equipped with a 0.22-μm membrane point-of-use cartridge (Millipak, EMD Millipore, Billerica, MA, USA). Potassium dihydrogen phosphate, tripelenamine hydrochloride, chlorpheniramine maleate, tetracaine, and promethazine hydrochloride were purchased from Sigma-Aldrich (Steinheim, Germany). Sample A mixture of the antihistaminic drugs tripelenamine, chlorpheniramine, tetracaine, and promethazine was prepared in 25 mm potassium dihydrogen phosphate in water, ph 3/acetonitrile (90/10; v/v) at a concentration of 50 µg/ml. 2

Methods Table 1. Chromatographic conditions for conventional LC analysis. Parameter Description Column Agilent ZORBAX SB-C18, 4.6 150 mm, 5 µm Solvent A) 25 mm Potassium dihydrogen phosphate in water, ph 3 B) Acetonitrile Gradient 10 %B at 0 minutes, 55 %B at 25 minutes Stop time 25 minutes Post time 10 minutes Flow rate 1.5 ml/min Temperature 40 C Injection volume 10.0 µl Detection 204/4 nm, reference 360/100 nm, data rate 10 Hz Table 2. Chromatographic conditions for UHPLC analysis optimized for resolution. Parameter Description Column Agilent InfinityLab Poroshell 120 EC-C18, 3.0 150 mm, 2.7 µm Solvent A) 25 mm Potassium dihydrogen phosphate in water, ph 3 B) Acetonitrile Gradient 10 %B at 0 minutes, 55 %B at 13.3 minutes Stop time 13.3 minutes Post time 6 minutes Flow rate 1.2 ml/min Temperature 40 C Injection volume 4.25 µl Detection 204/4 nm, reference 360/100 nm, data rate 40 Hz Table 3. Chromatographic conditions for UHPLC analysis optimized for speed. Parameter Description Column Agilent InfinityLab Poroshell 120 EC-C18, 3.0 50 mm, 2.7 µm Solvent A) 25 mm Potassium dihydrogen phosphate in water, ph 3 B) Acetonitrile Gradient 10 %B at 0 minutes, 55 %B at 2.13 minutes Stop time 2.13 minutes Post time 1 minute Flow rate 2.5 ml/min Temperature 40 C Injection volume 4.25 µl Detection 204/4 nm, reference 360/100 nm, data rate 80 Hz 3

Results and Discussion This Application Note shows the analysis of antihistaminic drugs using a conventional LC method on a 1100 Series Quaternary LC. This method is transferred to a 1260 Infinity II LC for proof of equivalency. Additionally, the conventional LC method for the analysis of antihistaminic drugs is transferred to UHPLC conditions optimized for resolution as well as for speed using the 1260 Infinity II LC. Figure 1 shows the conventional LC analysis of the four antihistaminic drugs tripelenamine, chlorpheniramine, tetracaine, and promethazine on the 1100 Series Quaternary LC. Excellent retention time and area precision was obtained (Table 4). Figure 2 shows the conventional LC analysis of the antihistaminic drugs run on the 1260 Infinity II LC, and Table 5 presents the corresponding retention time and area precision. In terms of retention time and area precision, the 1260 Infinity II LC outperforms the excellent 1100 Series Quaternary LC. In addition, a slight increase in resolution was observed. 320 300 280 260 240 220 200 180 160 140 120 100 80 60 40 20 0 0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 Figure 1. Conventional LC analysis of antihistaminic drugs on an Agilent 1100 Series Quaternary LC; overlay of 10 consecutive runs. Table 4. Conventional LC analysis of antihistaminic drugs on an Agilent 1100 Series Quaternary LC; retention time and area precision were determined from 10 consecutive runs. Compound RT (min) RT RSD (%) Area Area RSD (%) Resolution Tripelenamine 9.26 0.05 1,033.5 0.26 64.6 Chlorpheniramine 9.60 0.05 1,048.3 0.28 2.0 Tetracaine 11.86 0.05 577.2 0.26 13.8 Promethazine 13.49 0.03 2,014.5 0.25 9.1 320 300 280 260 240 220 200 180 160 140 120 100 80 60 40 20 0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 Figure 2. Conventional LC analysis of antihistaminic drugs on an Agilent 1260 Infinity II LC; overlay of 10 consecutive runs. 4

Figure 3 and Table 6 compare the retention times of the antihistaminic drugs analyzed on the 1100 Series Quaternary LC and the 1260 Infinity II LC. With a maximum deviation of 1.7 %, excellent agreement of retention times was observed. This proves the equivalency of the 1260 Infinity II LC compared to the 1100 Series Quaternary LC for the analysis of the antihistaminic drugs. Table 5. Conventional LC analysis of antihistaminic drugs on an Agilent 1260 Infinity II LC; retention time and area precision were determined from 10 consecutive runs. Compound RT (min) RT RSD (%) Area Area RSD (%) Resolution Tripelenamine 9.10 0.01 1,018.9 0.21 63.3 Chlorpheniramine 9.48 0.02 1,007.0 0.21 2.3 Tetracaine 11.84 0.02 508.9 0.17 14.5 Promethazine 13.48 0.01 2,017.1 0.14 9.2 Agilent 1260 Infinity II LC Agilent 1100 Series LC Figure 3. Conventional LC analysis of antihistaminic drugs on an Agilent 1100 Series Quaternary LC and an Agilent 1260 Infinity II LC. Table 6. Conventional LC analysis of antihistaminic drugs on an Agilent 1100 Series Quaternary LC and an Agilent 1260 Infinity II LC; comparison of retention times. Compound RT Deviation (min) RT Deviation (%) Tripelenamine 0.16 1.7 Chlorpheniramine 0.12 1.3 Tetracaine 0.02 0.2 Promethazine 0.02 0.1 5

Instrument control and data analysis was performed using Agilent OpenLAB CDS Version 2.1. Figure 4 shows an impression of the data analysis in OpenLAB CDS Version 2.1. OpenLAB CDS Version 2.1 offers a single software system for liquid chromatography, gas chromatography, and mass spectrometry. It provides a new, flat user interface, and enables advanced technical support for compliance reporting. It also provides customized and interactive reporting with drag and drop template creation, and enables review of relevant data only. In combination with the 1260 Infinity II LC, Agilent InfinityLab supplies were used. The Agilent InfinityLab Quick Connect fittings (p/n 5067-6166, with a 0.17 105 mm capillary) and Agilent Quick Turn fittings (p/n 5067-5966) enable tool-free, fast, and easy column installation, ensuring a perfect column connection independent of the user. The setup of the 1260 Infinity II LC on the Agilent InfinityLab Flex Bench rack (p/n 5043-1252) enables efficient use of lab space and an ergonomic approach with easy access to the instrument. The pressure range of up to 600 bar of the 1260 Infinity II LC allows UHPLC analyses using InfinityLab Poroshell columns. When ordering a 1260 Infinity II LC, the customer has the choice between different InfinityLab Poroshell columns that can be delivered with the system, for example, the Agilent InfinityLab Poroshell 120 EC-C18, 3.0 150 mm, 2.7 µm column (p/n 693975-302T) or the Agilent InfinityLab Poroshell 120 EC-C18, 3.0 50 mm, 2.7 µm column (p/n 699975-302T). For the analysis of antihistaminic drugs, those columns allow the method transfer to UHPLC conditions optimized for resolution and speed, respectively. Figure 4. Impression of the data analysis in Agilent OpenLAB CDS Version 2.1. 6

Figure 5 and Table 7 show the analysis of the antihistaminic drugs under UHPLC conditions optimized for resolution. Making full use of the pressure range of the 1260 Infinity II LC, the InfinityLab Poroshell 120 EC-C18, 3.0 150 mm, 2.7 µm column can be operated at a relatively high flow rate of 1.2 ml/min. Under these conditions, the resolution of tripelenamine and chlorpheniramine was increased from 2.3 with the conventional LC method to 5.4 with the UHPLC method optimized for resolution. Also, the analysis time could be decreased by 47 %, and the solvent use reduced by 57 %. Deploying the InfinityLab Poroshell 120 EC-C18, 3.0 50 mm, 2.7 µm column, the analysis of antihistaminic drugs can be transferred to UHPLC conditions optimized for speed, as shown in Figure 6 and Table 8. Compared to the conventional LC method, the resolution of tripelenamine and chlorpheniramine was still increased from 2.3 to 2.9. At the same time, the analysis time and solvent use could be greatly reduced by 91 and 86 %, respectively. 440 400 360 320 280 240 200 160 120 80 40 0 0.5 1.5 2.5 3.5 4.5 5.5 6.5 7.5 8.5 9.5 10.5 11.5 12.5 Figure 5. UHPLC analysis optimized for resolution of antihistaminic drugs on an Agilent 1260 Infinity II LC; overlay of 10 consecutive runs. Table 7. UHPLC analysis optimized for resolution of antihistaminic drugs on an Agilent 1260 Infinity II LC; retention time and area precision were determined from 10 consecutive runs. Compound RT (min) RT RSD (%) Area Area RSD (%) Resolution Tripelenamine 5.53 0.05 523.7 0.05 97.2 Chlorpheniramine 5.93 0.03 515.5 0.10 5.4 Tetracaine 7.43 0.04 260.6 0.09 21.0 Promethazine 8.46 0.03 1,028.2 0.04 13.6 560 520 480 440 400 360 320 280 240 200 160 120 80 40 0 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8 0.9 1.0 1.1 1.2 1.3 1.4 1.5 1.6 1.7 1.8 1.9 2.0 2.1 Figure 6. UHPLC analysis optimized for speed of antihistaminic drugs on an Agilent 1260 Infinity II LC; overlay of 10 consecutive runs. 7

Conclusion The transfer of a conventional LC method for the analysis of antihistaminic drugs from an Agilent 1100 Series Quaternary LC to an Agilent 1260 Infinity II LC showed a maximum retention time deviation of 1.7 % as well as a slight increase in resolution, and thereby proves the equivalency of the 1260 Infinity II LC compared to the 1100 Series Quaternary LC for the analysis of the antihistaminic drugs. With a pressure range of up to 600 bar, the 1260 Infinity II LC enables UHPLC analysis using Agilent InfinityLab Poroshell columns. By transfer of the conventional LC method to UHPLC conditions optimized for resolution as well as for speed, peak resolution was increased, and at the same time analysis time and solvent use were reduced. Table 8. UHPLC analysis optimized for speed of antihistaminic drugs on an Agilent 1260 Infinity II LC; retention time and area precision were determined from 10 consecutive runs. Compound RT (min) RT RSD (%) Area Area RSD (%) Resolution Tripelenamine 1.16 0.00 254.4 0.19 53.0 Chlorpheniramine 1.23 0.00 252.1 0.23 2.9 Tetracaine 1.46 0.05 126.7 0.25 9.8 Promethazine 1.63 0.04 500.0 0.19 6.4 References 1. Agilent 1290 Infinity with ISET, Agilent Technologies User Manual, part number G4220-90314, 2015. 2. Panula, et al., International Union of Basic and Clinical Pharmacology. XCVIII. Histamine Receptors, Pharmacological Reviews 2015, 67, 601 655. www.agilent.com For Research Use Only. Not for use in diagnostic procedures. This information is subject to change without notice. Agilent Technologies, Inc., 2016 Published in the USA, June 1, 2016 5991-6914EN