AAMI Quality Systems White Paper: Comparison of 21 CFR Part 820 to ISO 13485:2016 1

Similar documents
AAMI Quality Systems White Paper

Correspondence Between ISO 13485:2016 and 21 CFR Part 820 QMS Requirements

Quality Management System MANUAL. SDIX, LLC Headquarters: 111 Pencader Drive Newark, Delaware 19702

ISO 13485:2016 Mandatory documentation requirements & MyEasyISO

Medical devices Quality management systems Requirements for regulatory purposes

21 CFR Ch. I ( Edition) Personnel Design controls Identification Traceability.

ISO 13485:2016 The Final Countdown

YOU RE CLOSER THAN YOU THINK

FDA 21 CFR Part 820 vs. ISO 13485:2016 Comparison Table created by greenlight.guru

ISO 13485:2016 Medical Devices-Quality Management Systems- Requirements for Regulatory Purposes. Theresa McCarthy

Checklist for the assessment based on the standards

QUALITY MANUAL ECO# REVISION DATE MGR QA A 2/25/2008 R.Clement J.Haislip B 6/17/2008 T.Finneran J.Haislip

SCHEDULE M-III [See Rule 76]

Medical QMS Quality Manual

Food and Drug Administration, HHS Pt Sec Scope Definitions Quality system.

PART 820 QUALITY SYSTEM REGULATION. 21 CFR Ch. I ( Edition)

INTERNATIONAL STANDARD

ISO 9001:2015. Quality Management System. Manual

CO-ORDINATION OF NOTIFIED BODIES PPE Regulation 2016/425 RECOMMENDATION FOR USE

AS 9100 Rev C Quality Systems Manual AS-050C-QM

Design Quality. Indu Lakshman

QUALITY SYSTEM MANUAL

Wilson-Hurd ATF Quality Manual

25 D.L. Martin Drive Mercersburg, PA (717)

ISO9001:2008 SYSTEM KARAN ADVISER & INFORMATION CENTER QUALITY MANAGEMENT SYSTEM SYSTEM KARAN ADVISER & INFORMATION CENTER

AEROSPACE STANDARD. Quality Systems - Aerospace - Model for Quality Assurance in Design, Development, Production, Installation and Servicing

EPICOR, INCORPORATED QUALITY ASSURANCE MANUAL

Clause Map IATF 16949:2016 to ISO/TS 16949:2009

ASQ Tappan Zee Section Medical Devices, New Regulations and Standards. 26 th September / V. Fischer / Rev. 01

ISO /TS 29001:2010 SYSTEMKARAN ADVISER & INFORMATION CENTER SYSTEM KARAN ADVISER & INFORMATION CENTER

TURBO MACH A DIVISION OF VT SAA

QUALITY MANUAL QM0492 ISO SUPPLEMENT III

ISO 9001:2000 Revision G

HACCP audit checklist

ISO INTERNATIONAL STANDARD. Medical devices Quality management systems Requirements for regulatory purposes

ISO Introduction and Background

QM-1 QUALITY MANAGEMENT SYSTEMS MANUAL. Revision 10

Machined Integrations, LLC

MDSAP Purchasing Process

Supplier Quality Survey. 1. Type of Business: g) Commodities supplied? Supplier Changes/comments: 2. Headcount breakdown by group: Purchasing

Expert Commentary on BS EN ISO 13485:2016, Medical devices Quality management systems Requirements for regulatory purposes

RULES FOR A QUALITY STANDARDS SYSTEM (QSS) IN MARITIME ACADEMIES/INSTITUTIONS

Vendor Qualification Survey

Document Number: QM001 Page 1 of 19. Rev Date: 10/16/2009 Rev Num: 1. Quality Manual. Quality Manual. Controlled Copy

Auditing Within the Culture of the Medical Device Industry & ISO 13485:2016 Overview. ASQ Quality Conference 2017 for Region 5 October 25, 2017

Quality Manual Revision: C Effective: 03/01/10

Revision. Quality Manual. Multilayer Prototypes. Compliant to ISO / AS9100 Rev C

14620 Henry Road Houston, Texas PH: FX: WEB: QUALITY MANUAL

OPERATIONS MANUAL ISO 9001 Quality Management System

SQF 2000 Code. 6th Edition AUGUST A HACCP-Based Supplier Assurance Code for the Food Manufacturing and Distributing Industries

AEROSPACE STANDARD. (R) Inspection and Test Quality Systems Requirements for Aviation, Space, and Defense Organizations RATIONALE

SECTION L QUALITY ASSURANCE. 1) designating and managing quality control functions, including:

NMT Specialized Machining Inc & NMT General Machining Inc AS 9100 Rev C Quality Systems Manual

Odyssey Electronics, Inc Fairlane Livonia, MI 48150

c) Have personnel been appointed to supervise the production operations across all shifts in order to ensure the product quality?

INTERNATIONAL STANDARD

INTERNATIONAL STANDARD

Osprey Technologies, LLC. Quality Manual ISO9001:2008 Rev -

QUALITY MANUAL. This document defines the requirements, processes, structure and documentation for the Teledyne DGO Quality Management System.

Quality system implementation from a manufacturers viewpoint

QUALITY MANUAL. Number: M-001 Revision: C Page 1 of 18 THIS DOCUMENT IS CONSIDERED UNCONTROLLED UNLESS ISSUED IDENTIFIED AS CONTROLLED

ISO 9001: 2000 (December 13, 2000) QUALITY MANAGEMENT SYSTEM DOCUMENTATION OVERVIEW MATRIX

Quality Manual TABLE OF CONTENTS APPROVAL SIGNATURE PAGE AMENDMENT RECORD SCOPE EXCLUSIONS 2.1 3

Quality Manual ISO 9001:2000

TUV SUD BABT PRODUCTION QUALITY CERTIFICATION SCHEME

Specification for Quality Programs for the Petroleum, Petrochemical and Natural Gas Industry

DECISION TO ACCREDIT

Table of Contents Table of Contents... 2 Introduction to Manual... 3 Scope... 3

Supplier Quality Requirements SQR-1.0 Rev. 3

WHO Prequalification of In Vitro Diagnostics Programme

QMS Team: MR and all HODs (Internal Auditors) MR March 10. Quality policy Define quality policy The Steering committee Objectives and targets

UR Startup Inc. ISO Audit Checklist. conducted for. Conducted on (Date and Time) 02 Aug :06 PM. Inspected by Andy Dion

UNIT 10 CLAUSE-WISE INTERPRETATION OF ISO 22000: 2005

GENERIC QUALITY ASSURANCE REQUIREMENTS FOR: BUILT TO PRINT ITEMS, ITEMS TO STANDARD AND OFF THE SHELF ITEMS

Guide for Distributors of Medical Devices

Quality Assurance Policy and Procedures

ISO 22000:2005 Standard INTERNATIONAL STANDARDS REGISTRATIONS

Quality Manual QM -07 Approval: D. Wheeler. AARD Spring & Stamping Quality Manual. Quality Manual. Page 1 of 24

SUPPLIER SURVEY FORM Instructions

Quality Management System Manual

EUROPEAN COMMISSION HEALTH AND CONSUMERS DIRECTORATE-GENERAL. EudraLex. The Rules Governing Medicinal Products in the European Union

DEC STD ISO Quality Systems - Model for Quality Assurance in Design /Development, Production, Installation and Servicing

INSTRUCTIONS FOR CERTIFICATION OF FACTORY PRODUCTION CONTROL

ISO 22000:2005 SYSTEMKARAN ADVISER & INFORMATION CENTER SYSTEM KARAN ADVISER & INFORMATION CENTER FOOD SAFETY MANAGEMENT SYSTEM ISO 22000:2005

Comparison Matrix ISO 9001:2015 vs ISO 9001:2008

Luminus Devices, Inc Quality Management Systems Manual ISO 9001:2008

The following is an example systems manual from a low volume (TE, but not an automotive supplier) company.

POLYCRAFT INCORPORATED QUALITY MANUAL Quality Manual QM -10 Approval: D. Wheeler.

Review of HPRA Draft Guide to Distribution of Medical Devices, including in vitro diagnostic Medical Devices

Comparison of Requirements ISO/TS vs. AS9100 and ISO 13485

UNANNOUNCED EU MEDICAL DEVICE AUDITS BY NOTIFIED BODIES: IMPACT ON SUPPLIERS

MANUAL QUALITY CONTROL & QUALITY ASSURANCE

ASTM D10-F02 Workshop FDA Regulatory Perspective. Patrick Weixel FDA Center for Devices and Radiological Health

Texas International Oilfield Tools, Ltd.

Perry Johnson Registrars, Inc. Licensed Copy #2 RECYCLING INDUSTRY OPERATING STANDARD. Prepared for ISRI Services Corporation

Supplier Quality Agreements

QUALITY MANUAL. Revision K. JADE PRECISION MEDICAL COMPONENTS, LLC 3063 B Philmont Avenue Huntingdon Valley, PA 19006

Changes in the EXCiPACT 2017 Edition Requirements from the 2012 Edition

SUPPLIER QUALITY MANUAL

Quality System Regulation (QSR) Training

Transcription:

AAMI s White Paper Comparison of 21 CFR Part 820 to ISO 13485:2016 February 2017 AUTHORS Seb Clerkin, GMP Advisory Services Nicola Martin, Owner, Nicola Martin Consulting Jack Ward, Owner, Ward Sciences and Consulting LLC Tonya Wilbon, s Specialist, FDA, Center for Devices and Radiological Health, Office of In Vitro Diagnostics and Radiological Health A. ISO 13485:2016 What is it? ANSI/AAMI/ISO 13485:2016, Medical devices Quality management systems Requirements for regulatory purposes, specifies requirements for a Quality Management System when an organization needs to demonstrate its ability to provide medical devices and related services that consistently meet customer requirements and regulatory requirements applicable to medical devices and related services. This International Standard specifies requirements for a Quality Management System that can be used by an organization involved in one or more stages of the life cycle of a medical device, including design and development, production, storage and distribution, installation, servicing and final decommissioning and disposal of medical devices, and design and development, or provision of associated activities (e.g., technical support). The requirements in this International Standard may also be used by suppliers or other external parties providing product (e.g., raw materials, components, subassemblies, medical devices, sterilization services, calibration services, distribution services, maintenance services) to such organizations. The supplier or external party can voluntarily choose to conform to the requirements of this International Standard or can be required by contract to conform. AAMI AAMI s White Paper: Comparison of 21 CFR Part 820 to ISO 13485:2016 1

Note that this 2016 revision to ISO 13485 supersedes and replaces the 2003 revision of ISO 13485. Medical device manufacturers that were conforming to the 2003 revision of this standard must achieve conformance to the 2016 revision no later than 3 years after the March 2016 publication of ISO 13485:2016. B. Importance of ISO 13485 to medical device manufacturers Compliance with ISO 13485 is often seen as the first step in achieving compliance with European regulatory requirements. The conformity of Medical Devices and In-vitro Diagnostic Medical Device according to European Union (EU) Directives 93/42/ EEC, 90/385/EEC, and 98/79/EEC must be assessed before sale is permitted. The preferred method to prove conformity is the certification of the Quality Management System according ISO 13485 and ISO 14971 by a Notified Body. The result of a positive assessment is the certificate of conformity allowing the Conformité Européenne (CE) mark and the permission to sell the medical device in the European Economic Area (EEA). C. Comparing different Quality Management Systems standards and regulations Comparing Which Standards/Regulations? Use These References 1. ISO 13485:2003 to ISO 13485:2016 1. Annex A of ISO 13485:2016 (page 27) 2. ISO 13485:2016 to ISO 9001:2015 2. Annex B of ISO 13485:2016 (page 31) 3. ISO 13485:2003 to 21 CFR Part 820 3. AAMI Compendium, 3rd edition (Table 3.1) 4. ISO 13485:2016 to 21 CFR Part 820 4. This AAMI white paper (Table 1) D. Table 1 Additional information Note that in several key ISO 13485:2016 clauses in Table 1 (Design Controls and CAPA, for example), the authors have included an estimate of resources that would be required to update resultant processes to meet the ISO 13485:2016 standard. These estimates are provided as initial assumptions only and it is recommended that each organization should conduct a directed analysis to determine actual resources that would be required for a specific Quality Management System. Estimate of Effort to Update Existing 21 CFR Part 820 QS Management System Level Examples updates required to existing documents Major updates to existing documents or new documents required Moderate Clause 8.2.6: New requirement to document test equipment used Clauses 7.3.7: Design Validation plans required 2 AAMI s White Paper: Comparison of 21 CFR Part 820 to ISO 13485:2016 AAMI

Table 1 Comparison of 21 CFR Part 820 to ISO 13485:2016 21 CFR Part 820 820.1 Scope 0.1 Introduction, General 0.2 Clarification of Concepts 0.3 Process Approach 0.4 Relationship with ISO 9001 0.5 Compatibility with other management systems 1 Scope 2 Normative references 820.5 Quality system 0.1 Introduction, General 1 Scope 4.1 Quality management system, General requirements 820.20 Management responsibility 820.20(a) Quality policy 820.20(b) Organization 5.1 Management commitment 5.2 Customer focus 5.3 Quality policy 5.4.1 Quality objectives 4.2.1 Documentation requirements, General 5.1 Management commitment 5.3 Quality policy The Regulation addresses exemptions, applicability to foreign manufacturers, and the authority under which the Regulation was promulgated. The Regulation also addresses possible conflicts between compliance with the Regulation and with other FDA regulations. Explanations of the meaning of as appropriate in the Standard are similar to where appropriate in the Regulation with the Standard including the necessity to manage risk and comply with applicable regulatory requirements if a requirement is deemed appropriate. The Standard also includes explanations of the terms: risk, documented, product, regulatory requirement, shall, should, may, and can. The Standard addresses regulatory exclusions from design controls (based on scope of operations). The Standard explains its relationship with ISO 9001 and references ISO 9000:2015 with regard to the fundamentals of s and vocabulary. The Regulation does not reference these standards, although it does use some terms and concepts that appear in these standards. The Standard explains the process approach. The Regulation has no similar explanation. The Standard addresses compatibility with management systems for other types of activities. The Regulation does not. ISO 13485:2016 includes its potential application by suppliers to medical device manufacturers and adds a greater emphasis on the life cycle of a medical device. The Regulation and the Standard specify that manufacturers or organizations have flexibility in structuring their according to their device type, organization size and structure, needs, and situation. The Regulation s focus is the provision of safe and effective medical devices. The Standard has an explicit focus on meeting customer requirements, in addition to meeting regulatory requirements. The Standard has more detailed requirements regarding measurable quality objectives. The Regulation and the Standard require a quality policy. The Standard specifies criteria of a quality policy. 6.1 Provision of resources The Regulation specifies the establishment of an organizational structure. The Standard does not have a comparable requirement. AAMI AAMI s White Paper: Comparison of 21 CFR Part 820 to ISO 13485:2016 3

820.20(b)(1) Responsibility and authority 820.20(b)(2) Resources 820.20(b)(3) Management representative 820.20(c) Management review 820.20(d) Quality planning 820.20(e) Quality system procedures 5.5.1 Responsibility and authority 5.1 Management commitment 6.1 Provision of resources 5.5.2 Management representative 5.1 Management commitment 5.6 Management review 5.4.2 Quality management system planning 4.2.1 Documentation requirements, General 4.2.2 Quality manual The requirements of the Regulation and the Standard are similar. No significant differences. The requirements of the Regulation and the Standard are similar. The requirements for appointment of and responsibilities of a management representative are similar. The Regulation requires the management representative to report on the performance of the. The Standard requires the management representative to report on the effectiveness of the Quality Management System and any need for improvement. The Standard also has an additional requirement for promotion of awareness of regulatory and Quality Management System requirements throughout the organization. Requirements to conduct periodic management reviews are similar. The Regulation specifies conducting management reviews with sufficient frequency and specifies the documentation of the date of the review and the results. The Standard addresses specific topics that should be covered during management reviews including assessing opportunities for improvement and the need for changes to the Quality Management System, including the quality policy and quality objectives. Requirements of the Regulation and the Standard for quality planning are similar. The Regulation has specific requirements on how to define the quality plan and requirements to establish how the requirement for quality will be met. The Regulation requires establishment of procedures and instructions and an outline of the structure of the documentation used in the, but does not require a quality manual. The Standard requires establishment of a quality manual, which can include or reference procedures and documents necessary to ensure the effective planning, operation, and control of the organization s processes. 820.22 Quality audit 8.2.4 Internal audit Requirements of the Regulation and the Standard for quality audits are similar. The Regulation includes specific requirements for management having responsibility for the area audited to review the audit reports and specific requirements to document the dates of the quality audit. 820.25 Personnel, (a) General 6.2 Human resources, General Requirements of the Regulation and the Standard for personnel are similar. The Standard references establishing competence, providing needed training, and ensuring awareness of personnel. 4 AAMI s White Paper: Comparison of 21 CFR Part 820 to ISO 13485:2016 AAMI

820.25(b) Training 820.30 Design Controls, (a) General 820.30(b) Design and development planning 6.2 Human resources, General 6.4.1(b) Work environment 7.3.1 Design and development, General Comments: Although the Standard and the Regulation are mostly consistent regarding the requirements of Design Control, it is recommended that all 21 CFR Part 820.30 Design Control procedures be reviewed to determine the extent of any required updates. 7.3.2 Design and development planning Comments: Design and development planning procedures should be reviewed to ensure adequate compliance with 7.3.2(c), 7.3.2(e), and 7.3.2(f), which are not explicitly required in the Regulation. Requirements of the Regulation and the Standard are similar. The Regulation has additional requirements to make personnel aware of the device defects that could result from improper job performance, and to make personnel who perform verification and validation activities aware of defects and errors they may encounter in performing their jobs. The Standard specifically addresses competence for temporary personnel working in controlled environmental conditions. The Regulation specifies requirement of procedures for identifying training needs. The Standard specifies requirements for documenting the process to establish competency. The Regulation requires design and development procedures for design of Class III, Class II, and certain Class I devices in order to ensure that specified design requirements are met. The Standard requires that the organization shall document procedures for design and development of all devices regardless of device class or risk. The requirements for design and development planning are similar, but the Standard now requires explicit documented plans for: 7.3.2 a) the design and development stages; 7.3.2 b) the review(s) needed at each design and development stage; 7.3.2 c) the verification, validation, and design transfer activities that are appropriate at each design and development stage; 7.3.2 d) the responsibilities and authorities for design and development; 7.3.2 e) the methods to ensure traceability of design and development outputs to design and development inputs; 7.3.2 f) the resources needed, including necessary competence of personnel. AAMI AAMI s White Paper: Comparison of 21 CFR Part 820 to ISO 13485:2016 5

820.30(c) Design input 820.30(d) Design output 820.30(e) Design review 7.3.3 Design and development inputs Comments: Design and development inputs procedures should be reviewed to ensure adequate compliance with 7.3.3(a), 7.3.3(b), 7.3.3(c), 7.3.3(d), and 7.3.3(e), which are not explicitly required in the Regulation. 7.3.4 Design and development outputs Comments: Design and development outputs procedures should be reviewed to ensure adequate compliance with 7.3.4(b), which is not explicitly required in the Regulation. 7.3.5 Design and development review Comments: Design and development review procedures should be reviewed to ensure adequate compliance with 7.3.5(b), which is not explicitly required in the Regulation. The requirements for design inputs are similar. The Regulation includes explicit requirement that design input requirements are approved with a dated approval. The Standard now requires that design and development inputs include: 7.3.3 a) functional, performance, usability, and safety requirements, according to intended use; 7.3.3 b) applicable regulatory requirements and standards; 7.3.3 c) applicable output(s) of risk management; 7.3.3 d) as appropriate, information derived from previous similar designs; 7.3.3 e) other requirements essential for design and development of the product and processes. ISO 13485:2016 has added that product requirements are able to be verified or validated. Requirements for design output are similar. The Regulation requires that design outputs are documented, reviewed, and approved prior to release with a dated signature. The Standard now requires that design outputs shall: 7.3.4 a) meet the input requirements for design and development; 7.3.4 b) provide appropriate information for purchasing, production, and service provision; 7.3.4 c) contain or reference product acceptance criteria; 7.3.4 d) specify the characteristics of the product that are essential for its safe and proper use. Requirements for design review are similar. The Regulation requires the use of an independent reviewer, and also that the results of a design review, including identification of the design, the date, and the individual(s) performing the review shall be documented in the design history file. The Standard now requires that design and development reviews: 7.3.5 a) evaluate the ability of the results of design and development to meet requirements; 7.3.5 b) identify and propose necessary actions. 6 AAMI s White Paper: Comparison of 21 CFR Part 820 to ISO 13485:2016 AAMI

820.30(f) Design verification 820.30(g) Design validation 7.3.6 Design and development verification Comments: Design and development verification procedures should be reviewed to ensure adequate compliance with 7.3.6, and specifically with acceptance criteria, statistical techniques, with rationale for sample size and... connected to, or have an interface with, other medical devices... which are not explicitly required in the Regulation. The Standard explicitly requires documented verification plans and includes details of the content of these plans. 7.3.7 Design and development validation Quality System procedures and resultant processes to meet the ISO 13485:2016 Standard: Moderate Comments: Design and development validation procedures should be reviewed to ensure adequate compliance with 7.3.7, and specifically with acceptance criteria, statistical techniques, with rationale for sample size and The rationale for the choice of product used in validation shall be recorded. and... connected to, or have an interface with, other medical devices.... The Standard explicitly requires documented validation plans and includes details of the content of these plans. Requirements for design verification are similar. The Regulation includes specific documentation requirements, specifically that the results of the design verification, including identification of the design, method(s), the date, and the individual(s) performing the verification shall be documented in the design history file. The Standard now requires the following for design and development verification: Design and development verification shall be performed in accordance with planned and documented arrangements to ensure that the design and development outputs have met the design and development input requirements. The organization shall document verification plans that include methods, acceptance criteria, and, as appropriate, statistical techniques with rationale for sample size. If the intended use requires that the medical device be connected to, or have an interface with, other medical device(s), verification shall include confirmation that the design outputs meet design inputs when so connected or interfaced. Records of the results and conclusions of the verification and necessary actions shall be maintained (see 4.2.4 and 4.2.5). Requirements for design validation are similar. The Regulation has additional requirements for risk analysis and software validation and includes specific documentation requirements, specifically that the results of the design validation, including identification of the design, method(s), the date, and the individual(s) performing the validation shall be documented in the design history file. The Standard now requires the following for design and development validation: Design and development validation shall be performed in accordance with planned and documented arrangements to ensure that the resulting product is capable of meeting the requirements for the specified application or intended use. The organization shall document validation plans that include methods, acceptance criteria, and, as appropriate, statistical techniques with rationale for sample size. Design validation shall be conducted on representative product. Representative product includes initial production units, batches, or their equivalent. The rationale for the choice of product used for validation shall be recorded (see 4.2.5). (continued on next page) AAMI AAMI s White Paper: Comparison of 21 CFR Part 820 to ISO 13485:2016 7

820.30(g) Design validation (continued from previous page) 820.30(h) Design transfer 7.3.8 Design and development transfer Comments: Design and development transfer procedures should be reviewed to ensure adequate compliance with 7.3.8, and specifically with These procedures shall ensure that design and development outputs are verified as suitable for manufacturing before becoming final production specifications and that production capability can meet product requirements., which are not explicitly required in the Regulation. As part of design verification and validation, the organization shall perform clinical evaluations or performance evaluations of the medical device in accordance with applicable regulatory requirements. A medical device used for clinical evaluation or performance evaluation is not considered to be released for use to the customer. If the intended use requires that the medical device be connected to, or have an interface with, other medical device(s), validation shall include confirmation that the requirements for the specified application or intended use have been met when so connected or interfaced. Validation shall be completed prior to release for use of the product to the customer. Records of the results and conclusions of the validation and necessary actions shall be maintained (see 4.2.4 and 4.2.5). Requirements for design transfer are similar. The Standard now requires the following for design and development transfer: The organization shall document procedures for transfer of design and development outputs to manufacturing. These procedures shall ensure that design and development outputs are verified as suitable for manufacturing before becoming final production specifications and that production capability can meet product requirements. Results and conclusions of the transfer shall be recorded (see 4.2.5). 8 AAMI s White Paper: Comparison of 21 CFR Part 820 to ISO 13485:2016 AAMI

820.30(i) Design changes 820.30(j) Design history file 7.3.9 Control of design and development changes Comments: The Standard now explicitly requires that the review of design and development changes includes an evaluation of the effect of the changes on constituent parts and product in process or already delivered, inputs or outputs of risk management, and product realization processes. 7.3.10 Design and development files Comments: The Regulation explicitly identifies the design history file as the compilation of records which describes the design history of a finished device. The Standard has a general requirement for design and development files; however, the EU Medical Device Directive requires a technical file or design dossier. Requirements for design changes are similar. The Standard now requires the following for design changes: The organization shall document procedures to control design and development changes. The organization shall determine the significance of the change to function, performance, usability, safety, and applicable regulatory requirements for the medical device and its intended use. Design and development changes shall be identified. Before implementation, the changes shall be: a) reviewed; b) verified; c) validated, as appropriate; d) approved. The review of design and development changes shall include evaluation of the effect of the changes on constituent parts and product in process or already delivered, inputs or outputs of risk management, and product realization processes. Records of changes, their review, and any necessary actions shall be maintained (see 4.2.5). The Standard has a general requirement for maintaining design and development files as follows: The organization shall maintain a design and development file for each medical device type or medical device family. This file shall include or reference records generated to demonstrate conformity to the requirements for design and development and records for design and development changes. AAMI AAMI s White Paper: Comparison of 21 CFR Part 820 to ISO 13485:2016 9

820.40 Document controls 820.50 Purchasing controls, (a) Evaluation of suppliers, contractors and consultants 820.50(b) Purchasing data 4.2.1 Documentation requirements, General 4.2.3 Medical device file 4.2.4 Control of Documents 4.2.5 Control of Records Comments: The Standard lists the documents that would be included in the medical device file. The content of the medical device file would include the records contained within the device medical record, and additional requirements. 7.4.1 Purchasing process Comments: The Standard has greater detail on the supplier selection criteria. The Standard has specific requirements on supplier reevaluation and ongoing monitoring. 7.4.2 Purchasing information Requirements are similar. The Standard has additional requirements for control of documents of external origin and for the protection of confidential health information. The Standard has separate clauses for the control of documents and records. The Regulation does not differentiate. Requirements for evaluation of suppliers are similar. The Regulation has an explicit requirement to evaluate, select, and control contractors and consultants. Requirements for purchasing data (the Regulation) and purchasing information (the Standard) are similar. The Regulation and the Standard require that manufacturers pursue an agreement with suppliers to provide notification of any changes to the product or service. 820.60 Identification 7.5.8 Identification Requirements for identification are similar. The Standard explicitly addresses identification of returned devices. 10 AAMI s White Paper: Comparison of 21 CFR Part 820 to ISO 13485:2016 AAMI

820.65 Traceability 7.4.2 Purchasing information 7.5.9.1 Traceability, General 820.70 Production and Process Controls, (a) General, (b) Production and process changes 820.70(c) Environmental control, (d) Personnel, (e) Contamination control 820.70(f) Buildings 7.1 Planning of product realization 7.5.1 Control of production and service provision Comments: The Standard requires qualification of infrastructure. 6.3 Infrastructure 6.4.1 Work environment 6.4.2 Contamination Control 6.3 Infrastructure 7.5.1 Control of production and service provision Moderate Comments: The Standard requires qualification of infrastructure. The Standard specifies that infrastructure requirements shall also include requirements to prevent product mix-up and ensure orderly product handling. The Regulation has traceability requirements for life supporting, life sustaining, and implantable devices through distribution to the initial consignee. It achieves general traceability of devices through device history records. The Standard explicitly requires some degree of traceability for all devices and greater traceability for active implantable and implantable medical devices, including traceability beyond distributors. Requirements for developing, controlling, and monitoring production processes are similar. The Regulation explicitly requires verification or validation of changes to specifications, methods, processes, and procedures. The organization shall establish documented requirements for risk management throughout product realization. Records arising from risk management shall be maintained (see 4.2.4). NOTE 1 A document specifying the processes of the Quality Management System (including the product realization processes) and the resources to be applied to a specific product, project, or contract can be referred to as a quality plan. NOTE 2 The organization may also apply the requirements given in 7.3 to the development of product realization processes. NOTE 3 See ISO 14971 for guidance related to risk management. No significant differences. Requirements for buildings (the Regulation) and infrastructure (the Standard) are similar. AAMI AAMI s White Paper: Comparison of 21 CFR Part 820 to ISO 13485:2016 11

820.70(g) Equipment 820.70(h) Manufacturing material 820.70(i) Automated processes 820.72 Inspection, measuring and test equipment 6.3 Infrastructure 7.5.1 Control of production and service provision Comments: The Standard includes equipment under infrastructure. 7.5.2 Cleanliness of product 7.5.6 Validation of processes for production and service provision Comments: The Standard states that the level of software validation is proportionate to the risk associated with the software. 7.6 Control of monitoring and measuring devices Requirements for equipment (the Regulation) and infrastructure (the Standard) are similar. The Regulation has additional requirements for periodic inspections to ensure adherence to maintenance schedules, and making equipment adjustment limitations readily available. The Standard has an additional requirement pertaining to supporting services such as communication systems. Requirements for manufacturing materials (the Regulation) and cleanliness of product (the Standard) are similar. Requirements are similar. The Regulation includes specific documentation requirements. Requirements for inspection, measuring, and test equipment (the Regulation) and control of monitoring and measuring devices (the Standard) are similar. The Standard has more detailed requirements for assessing the impact of out-ofcalibration equipment on product in distribution, and addresses software capability requirements when used in monitoring and measurement applications. 12 AAMI s White Paper: Comparison of 21 CFR Part 820 to ISO 13485:2016 AAMI

820.75 Process validation 820.80 Receiving, in-process, and finished device acceptance, (a) General, (b) Receiving acceptance activities, (c) In-process acceptance activities, (d) Final acceptance activities 820.80(e) Acceptance records 820.86 Acceptance status 7.5.6 Validation of processes for production and service provision Moderate Comments: The Standard specifies the requirement to have documented procedures for validation of processes. 7.1 Planning of product realization 7.4.3 Verification of purchased product 7.5.10 Customer property 8.2.6 Monitoring and measurement of product 7.1 Planning of product realization 8.2.6 Monitoring and measurement of product 7.5.8 Identification Comments: The Standard has reference to UDI requirements. Requirements for process validation are similar. The Regulation has additional explicit requirements for monitoring, controlling, and documenting validated processes during routine production and having qualified operators perform validated processes. The Standard lists specific items that shall be contained within validation procedures. Requirements are similar. The Regulation requires devices to be controlled or quarantined until release. The Standard explicitly requires satisfactory completion of all planned arrangements before product release and service delivery. The Standard also specifically addresses the care and control of customer property; the Regulation has no corresponding requirement. The Standard directly links verification of purchased product to risk and levels of supplier evaluations. Requirements are similar. Requirements for acceptance status (the Regulation) and status identification (the Standard) are similar. AAMI AAMI s White Paper: Comparison of 21 CFR Part 820 to ISO 13485:2016 13

820.90 Nonconforming product 820.100 Corrective and preventive action 820.120 Device labeling 8.3.1 Control of nonconforming product, General 8.3.2 Actions in response to nonconforming product detected before delivery 8.3.3 Actions in response to nonconforming product detected after delivery 8.3.4 Rework 8.4 Analysis of data 8.5.2 Corrective action 8.5.3 Preventive action Moderate Comment: Risk-based approach will need to be clearly demonstrated when determining new CAPAs, timelines, and priorities. 7.5.1 Control of production and service provision Requirements for nonconforming product are similar. The Regulation requires determination of the need for an investigation, and a signature whenever concessions are approved for use. The Standard specifically addresses nonconformity in delivered devices. Requirements for corrective and preventive action are similar. The Regulation includes specific procedural requirements. The Standard treats corrective action and preventive action separately. The new Standard requires a risk-based approach to timelines and action taken. The Regulation has requirements for label integrity, inspection, storage, operations, and control numbers, which are more explicit and detailed than the Standard. The new Standard requires defined operations for labeling. The Regulation requires a specific, documented label release prior to use of the label. 14 AAMI s White Paper: Comparison of 21 CFR Part 820 to ISO 13485:2016 AAMI

820.130 Device packaging 820.140 Handling, 820.150 Storage 7.5.1 Control of product and service provision 7.5.11 Preservation of product Moderate Comment: If packaging alone cannot provide preservation, the methods for preserving the product will need to be demonstrated. The Regulation requires that packaging and shipping containers be designed to protect devices. The new Standard requires that suitable packaging and shipping containers are designed and constructed, and conditions needed, if packaging alone cannot provide preservation. 7.5.11 Preservation of product Requirements for handling and storage (the Regulation) and preservation of product (the Standard) are similar. The Regulation specifically addresses stock rotation for product that may deteriorate over time. 20.160 Distribution 7.2.1 Determination of requirements related to the product 7.2.2 Review of requirements related to the product 7.5.9.2 Particular requirements for implantable medical devices 820.170 Installation 7.5.3 Installation activities No significant differences. 820.180 Records, General requirements 820.180(a) Confidentiality 820.180(b) Record retention period The Regulation requires distribution procedures and records of distribution. The Standard only requires distribution records for implantable and active implantable devices. Under the Regulation, contract review is required before release of devices for distribution. Under the Standard, contract review is required when orders are received. 4.2.5 Control of records The Regulation and the Standard have similar requirements for record legibility and storage. The Regulation specifically requires backup of records stored in data processing systems. The Regulation also requires manufacturers to make records available to company officials and FDA employees. The Standard has no corresponding requirement. None 4.2.4 Control of documents 4.2.5 Control of records The Regulation has a provision for manufacturers to mark records during an inspection to identify those containing confidential information. The Standard has no corresponding provision. Requirements for retention of documents and records are similar. 820.180(c) Exceptions None The Regulation exempts reports of internal audits, supplier audits, and management reviews from review during FDA inspections. The Standard has no comparable exemptions. AAMI AAMI s White Paper: Comparison of 21 CFR Part 820 to ISO 13485:2016 15

820.181 Device master record 820.184 Device history record 820.186 Quality System record 820.198 Complaint files 4.2.1 Documentation requirements, General 4.2.3 Medical device file 7.1 Planning of product realization Comment: The Device Master Record and Device Master File can be the same document once the requirements from both the Regulation and the new Standard are incorporated. 7.5.1 Control of production and service provision None 8.2.2 Complaint handling Moderate 820.200 Servicing 7.5.4 Servicing activities 8.4 Analysis of data 820.250 Statistical techniques 8.1 Measurement, analysis, and improvement, General Requirements are similar for maintaining a record or file for each type or model of device, which includes procedures and specifications for production processes, packaging, labeling, and quality assurance. The Regulation assigns the name Device Master Record to these records/files. The Standard assigns the name Medical Device File. The Regulation and the Standard have similar requirements for maintaining records of production history. The Regulation assigns the name Device History Record to these records. In addition, the Standard includes explicit requirements for documenting sterilization processes. The Regulation has no comparable explicit requirement; however, the expectation is that sterilization processes will be documented. The Standard does not require a Record (QSR) but does require the types of records and documents that could be kept in the QSR. The requirements for complaint handling activities are similar. In addition, the Regulation requires that certain complaints be investigated and that complaints reportable under 21 CFR Part 803 be identified. The Regulation s recordkeeping requirements are more stringent. The new Standard requires a justification for no investigation, where the old version required justification where no corrective action was taken. Requirements are similar. Both require service records to be analyzed to determine if they meet the definition of a compliant. In addition, the Regulation has more detailed requirements for information to be included in service records. Requirements are similar. In addition, the Regulation requires establishing procedures for using statistical techniques and sampling methods, and for basing sampling plans on valid statistical rationale. 16 AAMI s White Paper: Comparison of 21 CFR Part 820 to ISO 13485:2016 AAMI