Edward F. Askew PhD Askew Scientific Consulting

Similar documents
Overview of the EPA Office of Water s Alternate Test Procedure Program

Hach Method Luminescence Measurement of Dissolved Oxygen (LDO ) in Water and Wastewater

STEPS TO DMR-QA SUCCESS. BY: Marcy Bolek

TNI s NATIONAL ENVIRONMENTAL LABORATORY ACCREDITATION PROGRAM (NELAP) Changing to the 2009 Standard

A Summary of the Changes in the Newly Promulgated GC/MS Method for Volatile Organics in Wastewater

Florida Department of Environmental Protection. Field and Lab Audits

How to use the TNI standards to meet essential QC elements in the MUR

Abraxis Method 546 Guide

Phosphorus, Reactive (Orthophosphate)

Monitoring Quality of Wastewater and Storm Water Data Dealing with Multiple Permits, States, and Analytical Laboratories

Will Accrediting Authorities allow PT Providers to reissue a result report if the lab identifies a reporting problem? Sponsor: Bill Hall Resolution:

Verifying the Reliability of EPA Method 314 to Measure Perchlorate at Sub ppb Levels vs New EPA Method Options

ENZYMATIC NITRATE ANALYSIS FOR EPA REGULATED LABORATORIES USING NITRATE REDUCTASE A GUIDE TO UPDATING REGULATORY METHODS FOR CWA NITRATE REPORTING

Microbac Laboratories, Inc.

Phosphorus, Reactive (Orthophosphate)

Hach Method Revision 1.0 January, Hach Company 5600 Lindbergh Drive Loveland, Colorado 80538

Why Hach Reagents? Alphabetical Listing of Reagents

A2LA. R231 Specific Requirements: Threat Agent Testing Laboratory Accreditation Program. December 6, 2017

recovery and between 0% and 10% relative standard deviation.

Table I: MCL and MRL Concentrations for Contaminants Monitored Under the Safe Drinking Water Act National Primary Drinking Water Regulations

Trace Metals in Waters by GFAAS, in Accordance with U.S. EPA and Health Canada Requirements

METHOD DETERMINATION OF TURBIDITY BY NEPHELOMETRY. Edited by James W. O'Dell Inorganic Chemistry Branch Chemistry Research Division

Regulation 85. Navigating Regulation 85 Part II Lab Methods and Data Submission

Application Note AN1501

Collaborative Efforts to Implement On Line Analyzer Technology for Regulatory Total Residual Chlorine Monitoring

Overview CWA Methods Update Rule Method August 2017 Adrian Hanley, U.S. EPA

Sulfate. USEPA 1 SulfaVer 4 Method 2 Method to 70 mg/l SO 4. Powder Pillows. Test preparation. Before starting.

Method 0040 for Volatile Total Organics as Unspeciated Mass in Stack Gas Emissions

DR/4000 PROCEDURE. IRON, Ferrous. Using Powder Pillows

CT DEP Reasonable Confidence Protocols. Glen Breland, VP Member CT DEP QA/QC Workgroup

Validation of Analytical Methods used for the Characterization, Physicochemical and Functional Analysis and of Biopharmaceuticals.

A24 CALA Checklist for Microbiology Revision 4.1 February 16, 2018

PHOSPHORUS, REACTIVE (0.00 to 5.00 mg/l PO 4

Laboratory Developed Tests. William Castellani, MD Inter-regional Commissioner Clinical Pathologist, Penn State Hershey Medical Center

Method Update Rule and Sufficiently Sensitive Methods Rule: What You Need to Know

Total Dissolved Solids Dried at 180ºC. Total Volatile Dissolved Solids at 550 C

DR/4000 PROCEDURE. OXYGEN DEMAND, Chemical. Colorimetric Measurement, 0 to 40 mg/l COD

Page 1 of 26 Revision October 2001

METHOD VALIDATION: AN ESSENTIAL COMPONENT OF THE MEASUREMENT PROCESS. Presented by : Jerry Parr, The NELAC Institute

How to Rock your Lab Audit!

2010 PNCWA Conference Bend, OR

DR/4000 PROCEDURE. OXYGEN DEMAND, Chemical. Colorimetric Measurement

ASTM D19 Method Validation Procedures William Lipps Analytical & Measuring Instrument Division July, 2015

TECHNICAL GUIDANCE FOR THE VALIDATION OF METHODS USED BY CHEMICAL LABORATORIES IN THE FOOD, WATER AND RELATED INDUSTRIES

Medical Cannabis Laboratory Approval Program

QAM-A-103 Data Reporting by the Laboratory Manager

Nonylphenols and Ethoxylates in Water by LC/MS/MS

By Authority Of THE UNITED STATES OF AMERICA Legally Binding Document

Analytical Method of Limit Test for Hexachlorobenzene. in Picloram TC and Method Validation Data*

10. Validated Normal Phase HPLC Method for the Determination. Fulvestrant is primarily used in the treatment of hormone receptor

CRITERIA FOR VALIDATION OF METHODS USED BY CHEMICAL LABORATORIES AND RELATED INDUSTRIES

By Authority Of THE UNITED STATES OF AMERICA Legally Binding Document

3. Fill a 50-mL. 2. Press: 1 ENTER The display will show mg/l, Al and the ZERO icon.

Investigation Requirements for Ethanol-Blended Fuel Releases Petroleum Remediation Program

Environmental Water Testing: Surface Water, Groundwater, Hard Water, Wastewater, & Seawater

Verification of Method XXXXXX

Simple Mathematics and How It Impacts You

A Rapid Method for Trace Analysis of Organophosphorus Pesticides in Drinking Water

Oxygen Demand, Chemical

EPA Methods Update. What s new and what s coming in EPA ICP-MS methods?

The 2016 TNI Standard. Module 2: Quality Systems Module 4: Method Validation Module 5: Microbiology

3. Add 0.4 ml of. 7. Use a TenSette

SOP-C-102 Determination of Chemical Oxygen Demand

Document Review 1 2 3

Turbidity by Nephelometric Determination

DRPK-005

Challenges and roadblocks adopting new laboratory Accreditation Standards. Ron Coss Orange County Sanitation District

GROUNDWATER MONITORING QUALITY ASSURANCE/QUALITY CONTROL PLAN LOWER YAKIMA VALLEY GWMA INITIAL CHARACTERIZATION

Dimethylphenol Method Method LR (0.23 to mg/l NO 3 N or 1.00 to mg/l NO 3 ) TNTplus 835

QAM-I-113 Operation and Calibration of the Dissolved Oxygen Meter

MODULE 2A GENERAL MANAGEMENT SYSTEM REQUIREMENTS

BSK Labs Frequently Asked Questions

By Authority Of THE UNITED STATES OF AMERICA Legally Binding Document

EPA APPROVES IN-SITU OPTICAL RDO METHODS

MOLECULAR TESTING: VERIFYING/VALIDATING INSTRUMENTS, REAGENTS AND ASSAYS. Richard L. Hodinka, Ph.D.

DETERMINATION OF METHOD UNCERTAINTY A PRACTICAL EXAMPLE FOR THE ANALYSIS OF TRIHALOMETHANES IN DRINKING WATER USING PURGE&TRAP GC-MS

ANALYTICAL REPORT. Attn: Richard Boelter

TOC600 ANALYZER S TANDARD OPERATING P ROCEDURES

Standard Operating Procedure (SOP)

Method 8017 (0.7 to 80.0 µg/l) Powder Pillows. Scope and Application: For water and wastewater; digestion is required to determine total cadmium.

Quality Assurance / Quality Control (QA / QC)

Analysis of Heavy Metals in Iron-based Fertilizers by HR ICP-OES

DoD Training Environmental Laboratory Accreditation Program Requirements for 8330B

Neilson Research Corporation received 42 sample(s) on 6/18/2018 for the analyses presented in the following report.

The Fundamentals of Understanding and Evaluating Laboratory Environmental Data

Molecular Diagnosis Challenges & Solutions. Using Molecular Kits or Laboratory Developed Tests (Home Brew), Emphasis on Validation

Module 12: Laboratory Overview Answer Key

ASTM D19.06 SUBCOMMITTEE

Analysis of Environmental Samples with the Agilent 730-ES Following US EPA Guidelines

3. Add 0.4 ml of. 7. Use a TenSette

Laboratory Accreditation Test Validation: A Brave New World for Anatomic Pathology

Guide to Fulfillment of Validation and Verification of Examination Requirements

1,5-Diphenylcarbohydrazide Method* (Powder Pillows or AccuVac Ampuls) USEPA accepted for wastewater analyses** Using Powder Pillows

US EPA Method using the Thermo Scientific icap 7600 ICP-OES Duo

WARNING LETTER AUG 3, 2016

New Standard Methods and ASTM Methods for the Determination of Total Nitrogen in Aqueous Samples

METHOD 7550 OSMIUM (ATOMIC ABSORPTION, DIRECT ASPIRATION)

SOP-C-130 Determination of Total Solids and Percent Solids

Committee D19 0n Water Subcommittee D19.05 on Methods for Analysis of Inorganic Substances in Water

Preparing for an On-Site Audit

Transcription:

Edward F. Askew PhD Askew Scientific Consulting

DO YOU KNOW WHAT IT IS?? (Are the Auditors Listening??) Askew Scientific Consulting LLC 2

Requires specific test procedure: Protocol for Review and Validation of Alternate Test Procedures for Regulated Organic and Inorganic Analytes in Wastewater Under EPA s Alternate Test Procedure Program Once recommended, must be promulgated to be a rule!!! ATP Follow 40 CFR 136.6 Refer to Redding Memo Complete Equivalency Askew Scientific Consulting LLC 3

Question 1: What type of DO probe do you use for BODs? What method allows you in writing to use an optical DO probe?? SM 5210 B Footnotes (Hach or InSitu) Question 2: What method do you use for CODs? What method allows the use of Silver Sulfate?? SM 5220 D EPA 410.4 Revision 2 Askew Scientific Consulting LLC 4

Auditor-Laboratory What Should I Look For In My Test Kit???? Askew Scientific Consulting LLC 5

Fill a sample cell with 10 ml of sample. Add the contents of one Reagent Powder Pillow to the cell. A blue color develops if phosphorus is in the sample. Immediately close the sample cell. Shake vigorously for 20 30 seconds. Old Test Kit EPA Letter 1996 & 1999 EPA Approval Letter only for EPA Methods 365.1& 365.2 Add 5.0 ml of sample to a vial. (Vial contains 5 ml liquid reagent) Cap and invert to mix. Using a funnel, add the contents of one dry packet to the vial. Cap the vial tightly and shake for 10-15 seconds. Note: The powder will not completely dissolve. New Test Kit 2017 Askew Scientific Consulting LLC 6

40 CFR part 136.6 Askew Scientific Consulting LLC 7

Flexibility to Modify Methods - 40 CFR 136.6 https://www.epa.gov/cwa-methods/alternate-testprocedures If you use a modification to an approved 40 CFR Part 136 method and document the modification as described at 40 CFR 136.6, you will no longer receive or require a letter from EPA. The promulgated 136.6, as modified by the May 18, 2012 Methods Update Rule allows the regulated community more flexibility to modify approved methods without EPA review, provided certain requirements are met. This regulation allows the analytical community greater flexibility to modify approved methods to lower the costs of measurements, overcome matrix interferences, or otherwise improve the analysis without EPA review. Laboratories that modify Part 136 methods may be private, public or commercial and may conduct analyses for one or more clients or facilities. Askew Scientific Consulting LLC 8

My Auditor Is ONSITE!!!! Askew Scientific Consulting LLC 9

Method modifications: If the underlying chemistry and determinative technique in a modified method are essentially the same as an approved part 136 method, then the modified method is an equivalent and acceptable alternative to the approved method provided the requirements of this section are met. However, those who develop or use a modification to an approved (part 136) method must document that the performance of the modified method, in the matrix to which the modified method will be applied, is equivalent to the performance of the approved method. Supporting documentation must, if applicable, include the Routine initial demonstration of capability and Ongoing QC including Determination of precision and accuracy, Detection limits, Matrix spike recoveries. Askew Scientific Consulting LLC 10

My Auditor Wants More Information About The Modified Method s Use For Regulatory Reporting Under the CWA Askew Scientific Consulting LLC 11

Modified Method Client or Auditor Developer Laboratory Use Purchasing Laboratory Askew Scientific Consulting LLC 12

The Supplier The Laboratory Askew Scientific Consulting LLC 13

Richard Redding Memo: Flexibility to Modify CWA Methods, November 20, 2007 https://www.epa.gov/sites/production/files/2015-08/documents/cwa-methodflexibility_memo_11-20-2007.pdf Developer Responsibilities Provide the laboratory with a side-by side method comparison table The developer should provide to its customers an in-depth comparison of the modified method with the EPA approved method, and document the comparison in a two-column method comparison table. The two-column method comparison table shall include the number and title of each method, the latest revision date of the modified method and a detailed discussion of each of the 17 topics required by the standard EPA method format. Askew Scientific Consulting LLC 14

The developer should provide to their clients the modified method written in the standard EPA format: http://www.epa.gov/waterscience/methods Provide a copy of the data comparing the modified method performance to the approved method to demonstrate that the method is capable of yielding reliable data for compliance monitoring purposes. Test results from validation of a modified method are used to demonstrate that the modified method produces results are equivalent to results produced by the EPA-designated approved method. Equivalency is established by demonstrating that the modified method produces results that meet or exceed the QC acceptance criteria of the EPA-designated approved method. Verify that all items of the "Equivalency Checklist" are met: Askew Scientific Consulting LLC 15

1. Concentrations of calibration standards. Document the range of the concentrations of material used to establish the relationship between response of the measurement system and analyte concentration. 2. %RSD or correlation coefficient of calibration regression. 3. Performance range tested with units. 4. Sample(s) used in initial demonstration have the recommended preservative, where applicable. 5. Sample(s) used in initial demonstration met recommended holding times, where applicable. 6. Interferences. 7. Document the qualitative identification criteria used. 8. Performance evaluation studies performed for analytes of interest, where available. Askew Scientific Consulting LLC 16

9. Latest study sponsor or title 10.Latest study number. 11.Analysis of external reference material 12.Results of analyses on reference material from a source different from that used to prepare the calibration standards, if applicable. 13.Sources of external reference material, if applicable. 14.Surrogates used, if applicable. 15.Concentrations of surrogates, if applicable. 16.Recoveries of surrogates appropriate to the proposed use, if applicable. 17.Sample preparation. 18.Clean-up procedures. 19.Method blank result. 20.Matrix (reagent water, drinking water, effluent) 21.Matrix spikes. 22.Spiking system, appropriate to the method and application. Askew Scientific Consulting LLC 17

23.Spike concentrations (with units corresponding to the final sample concentration) and recoveries. 24.Source of spiking material. 25.Number of replicate spikes 26.Initial demonstration of capability. 27.Precision (analyte by analyte) 28.Duplicates. 28.Bias (analyte by analyte). 29.Detection limit (with units; analyte by analyte). 30.Confirmation of detection limit, if applicable. 31.Quantitation limit (with units; analyte by analyte) Minimum level (ML), practical quantitation level (PQL) or limit of quantitation (LOQ). 32.Qualitative confirmation. Askew Scientific Consulting LLC 18

My Auditor is Breathing Down My Neck!! Askew Scientific Consulting LLC 19

Askew Scientific Consulting LLC 20

%RSD or correlation coefficient of calibration regression. Performance evaluation studies performed for analytes of interest, where available. Analysis of external reference material Results of analyses on reference material from a source different from that used to prepare the calibration standards, if applicable. Sources of external reference material, if applicable. Askew Scientific Consulting LLC 21

Method blank result. Matrix (reagent water, drinking water, effluent) Matrix spikes. Spiking system, appropriate to the method and application. Spike concentrations (with units corresponding to the final sample concentration) and recoveries. Source of spiking material. Askew Scientific Consulting LLC 22

Number of replicate spikes Initial demonstration of capability. Duplicates. Bias (analyte by analyte). Detection limit (with units; analyte by analyte). Estimated Confirmation of detection limit, if applicable. Askew Scientific Consulting LLC 23

Auditor Has The Checklist Out!!! Askew Scientific Consulting LLC 24

Require Supplier to Provide: Equivalency Report that Answers ALL requirements in 40 CFR 136.6 and Richard Redding s Memo!! https://www.epa.gov/sites/production/files/2015-08/documents/cwa-method-flexibility_memo_11-20-2007.pdf Method in EPA Format Product Insert Does Not Do This!!!! Additional Nice Items to Have: Check List for Laboratory Management/QC Staff Check List for Laboratory Staff Askew Scientific Consulting LLC 25

Table 1: Laboratory Method Equivalency Checklist Item from Reding Memo Have a detailed Standard Operating Procedure (SOP) available. (Guidance provided in Spectroquant method) Performing and document an initial demonstration of capability. (Required in Spectroquant method) Verify the modified method by analyzing and documenting 3-7 representative effluents (QC in the Spectroquant method meets this requirement as per 40 CFR part 136.6(b)(2)(i)(A)). Check Off (Yes/No) The facility/lab is to show they can get the modified method to work and that it gets comparable results for their effluent. (QC in the Spectroquant method meets this requirement as per 40 CFR part 136.6(b)(2)(i)(A)) A demonstration of calibration linearity or use of a calibration curve. (required in Spectroquant method) Periodic calibration verification (Required in Spectroquant method) An ongoing demonstration of performance (ongoing precision and recovery (OPR) and a blank with each sample batch (Required in Spectroquant method) A demonstration of the method detection limit (MDL) (Required in Spectroquant method) Matrix spike and matrix spike duplicate for each discharge the first time that the sample of the discharge is analyzed and at a frequency of 5% thereafter (Required in Spectroquant method) Meeting the quality control (QC) specifications of the method. (See Spectroquant method) Keep on hand the modified method manufacturer's supporting data available for review when the manufacturer has developed the method modification. Askew Scientific Consulting LLC 26

MilliporeSigma http://www.emdmillipore.com/usepa Environmental Express http://www.envexp.com/products/1- Wet_Chemistry/PW-Solids_Testing/TDSSW- StableWeigh_for_TDS Askew Scientific Consulting LLC 27

Askew Scientific Consulting LLC 28