Determination of Total Petroleum Hydrocarbons in Rubble and Soils by Accelerated Solvent Extraction and GC-FID

Similar documents
SpeedExtractor E-916 Determination of Total Petroleum Hydrocarbons in Soil, Sediment, and Waste Samples using the SpeedExtractor E /2009

Automated Solid-Phase Extraction of Organochlorine Pesticides from Drinking Water. Pranathi R Perati Thermo Fisher Scientific, Sunnyvale, CA, USA

Automated Solid-Phase Extraction of Organochlorine Pesticides from Drinking Water

Concentration of Human Hormones in Drinking Water Using Solid Phase Extraction and Analysis by High Performance Liquid Chromatography

Improving Retention Time Precision and Chromatography of Early Eluting Peptides with Acetonitrile/Water Blends as Solvent B

Extraction of POPs from Environmental Samples using Accelerated Solvent Extraction (ASE )

In-Cell and External Hydrolysis of Biomass for Carbohydrate Determinations Using Automated Solvent Extraction

Analysis of Illegal Dyes in Food Matrices Using Automated Online Sample Preparation with Liquid Chromatography-Mass Spectrometry

Thermo Scientific Dionex ASE 150 and 350 Systems. Unsurpassed Extraction Technology Accelerated Solvent Extraction

Extraction of Phthalates from Solid and Liquid Matrices

Determination of Diesel Range Organics (DROs) Using SPE and GC/FID by Method 8015D*

EXTRACTION OF BIOLOGICAL TISSUES FOR AROMATIC AND CHLORINATED HYDROCARBONS

Jonathan R. Beck and Charles T. Yang; Thermo Fisher Scientific, San Jose, CA

The Benefits of SOLAμ Technology in Sample Preparation. Jon Bardsley and Ken Meadows Thermo Fisher Scientific, Runcorn, UK

A Fast and Robust Linear ph Gradient Separation Platform for Monoclonal Antibody (MAb) Charge Variant Analysis

Analysis of Fracking Flowback Water From the Marcellus Shale Using In-line Conductivity, Automated Dilution, and Ion Chromatography

Simultaneous Extraction of PAHs and PCBs from Tissue and Soil Samples Using a New Accelerated Solvent Extraction (ASE) Procedure

Direct Injection of Fish Oil for the GC-ECD Analysis of PCBs: Results Using a Deans Switch With Backflushing. Application. Author.

Nitrogen Determination in Soils and Plants by Flash Combustion using Argon as Carrier Gas

Large Volume Injection of Organochlorine Pesticide Extracts Using a Programmable Temperature Vaporizer in the Solvent Split Mode

Introduction. Yan Liu, Victor Barreto, Rosanne Slingsby, Chris Pohl Thermo Fisher Scientific, Sunnyvale, CA, USA

A Rapid Method for Trace Analysis of Organophosphorus Pesticides in Drinking Water

Determination of Persistent Organic Pollutants (POPs) in Fish

Alex J. Zhang, Pranathi Perati, and Monika Verma; Thermo Fisher Scientific, Sunnyvale, CA, USA

ANALYSIS OF PESTICIDE RESIDUES IN DRINKING WATER AS PER BUREAU OF INDIAN STANDARDS USING THE AGILENT 7000 GC/MS/MS WITH PESTICIDES ANALYZER

Determination of 1,4-Dioxane in Drinking Water by Gas Chromatography/Mass Spectrometry (GC/MS) with Selected Ion Monitoring (SIM)

TargetQuan 3 Software. Leading the way in regulatory. POPs quantification. Bullet Bullet Bullet

Thermo Scientific EASY-nLC 1200 System. Leading in simplicity. and performance

Development of Ultra-fast ph-gradient Ion Exchange Chromatography for the Separation of Monoclonal Antibody Charge Variants

Quantitative LC-MS Analysis of 14 Benzodiazepines in Urine Using TraceFinder 1.1 Software and High Resolution Accurate Mass

Determination of Manganese in Brine ( 30% NaCl)

SpeedExtractor E-916 Extraction of Spices using the SpeedExtractor E-916 for the determination of Pesticides

Analytical Method for aldrin, endrin and dieldrin (Targeted to Agricultural, Animal and Fishery Products)

Analysis of Glyphosate and AMPA in Environmental Water Samples by Ion Chromatography and On-Line Mass Spectrometry with Minimal Sample Preparation

for water and beverage analysis

Application Note No. 067/2011

Determination of organochlorine pesticide residues in eggs by gel-permeation chromatography (GPC) cleanup and GC-ECD

Method Optimisation in Bottom-Up Analysis of Proteins

Determination of Oxalate and Other Anions in Bayer Liquor Using Ion Chromatography

Highlighted Capillary IC Applications

HIC as a Complementary, Confirmatory Tool to SEC for the Analysis of mab Aggregates

Nano LC at 20 nl/min Made Easy: A Splitless Pump Combined with Fingertight UHPLC Nano Column to Boost LC-MS Sensitivity in Proteomics

Analysis of 16 Pesticide Residues in Broccoli Using CarbonX dspe QuEChERS AOAC Kits Using GC/MS

Optimization of Volatile Petroleum Hydrocarbons Analysis by Automated Headspace Using Method Development Tools

A Novel Screening Method for Anthropogenic Sewage Pollutants in Waste Water, Ground Water and Drinking Water Samples by LC HRAM Analysis

Sprint analysis of lubricating oils using the Thermo Scientific icap 7600 ICP-OES

Claudio De Nardi, 1 Marta Kozak, 2 David Kasper 3 1. Thermo Fisher Scientific, Dreieich, Germany; 2 Thermo Fisher Scientific, San Jose, CA (USA); 3

Direct Carbohydrate Analysis in Beverages and Foods Using Pulsed Amperometric Detection or Charged Aerosol Detection

MARKET FOCUSED - environmental ORGANOCHLORINE PESTICIDES

Fast mass transfer Fast separations High throughput and improved productivity Long column lifetime Outstanding reproducibility Low carryover

High-speed, High-resolution Oligonucleotide Separations Using Small Particle Anion-Exchangers

Direct Carbohydrate Analysis in Beverages and Foods Using Pulsed Amperometric Detection or Charged Aerosol Detection

Analysis of Monoclonal Antibody and Related Substances using a New Hydrophobic Interaction Chromatography (HIC) Column

Analysis of PCBs in Sewage sludge for the characterization study of ERM CC392

Determination of chemical contaminants in marine fish by GCMS/MS using QuEChERS as an extraction method

CORESTA Recommended Method No. 57

Research Chemicals. HONEYWELL RETURNABLE CONTAINER PROGRAM EUROPE Reduce Costs and Accelerate Your Research

NWTPH-Gx. Volatile Petroleum Products Method for Soil and Water

TECHNICAL INFORMATION

A Novel ph Gradient Separation Platform for Monoclonal Antibody (MAb) Charge-Variant Analysis

CHROMATOGRAPHY/MASS SPECTROMETRY

Thermo Scientific Dionex ICS-900 Ion Chromatography System

Gas Chromatography Assignment Chem 543/443

The Importance of High Speed Data Acquisition in the Design and Implementation of GC/MS Applications

MICROWAVE SOLVENT EXTRACTION TECHNIQUES FOR VARIOUS REGULATORY COMPOUNDS

FOOD PESTICIDE RESIDUES LABORATORY REFERENCE NO INSPECTOR DATE

Analysis of Trace Elements in Seawater Using the Thermo Scientific icap 7000 Series ICP-OES Duo

PESTICIDE SCREENING APPLICATION SOLUTION

Total Petroleum Hydrocarbons By Gas Chromatography in Colombian Waters and Soils

Determination of Trace Concentrations of Oxyhalides and Bromide in Municipal and Bottled Waters Using a Compact Ion Chromatography System

Organohalide Pesticides in Water by GC/µECD with Agilent J&W DB-CLP1 and DB-CLP2

Determination of Sulfite in Food and Beverages by Ion Exclusion Chromatography with Pulsed Amperometric Detection

Determination of Monocrotophos, Diazinon, Malathion, EPN, and Methamidaphos from Aqueous Samples Using Atlantic HLB SPE Disks

ANALYTICAL STANDARDS AND REAGENTS

Extraction Products Separator Paper 188 Solid Phase Extraction (SPE) 189 Extraction Thimbles 192. Extraction Products

David Steiniger, 1 Dwain Cardona, 1 Cristian Cojocariu, 2 Paul Silcock, 2 Alexander Semyonov, 1 Jason Cole 1 1

MS Certified Vials. Pre-cleaned and certified vials for mass spectrometry

Passion. Power. Productivity. Corporate Headquarters

TechnicalNOTE METABOLYNX EXACT MASS DEFECT FILTER: RAPIDLY DISCRIMINATE BETWEEN TRUE METABOLITES AND FALSE POSITIVES INTRODUCTION

Minimizing Carbon Dioxide Contamination in Anion Reagent-Free Ion Chromatography

Agilent Ultimate Plus Fused Silica Tubing

The Simultaneous Analysis of Benzene, Toluene, Ethylbenzene, o,m,p-xylenes and Total Petroleum Hydrocarbons in Soil by GC-FID after Ultra-Sonication

Optimizing Productivity and Reliability for Monocyclic Aromatic Hydrocarbon Purity Analysis According to ASTM D7504 on the Agilent 8890 GC System

Thermo Scientific EASY-Spray Technology. Plug-and-spray with. State-of-the-art performance

High-Resolution Charge Variant Analysis for Top-Selling Monoclonal Antibody Therapeutics Using a Linear ph Gradient Separation Platform

Chromatography Column Performance and Data Analysis Success Guide. Hints and Tips for Better Purifications

Method of determination of phthalates by gas chromatography / mass spectrometry in wines

Automation for Improving the Workflows for LC-MS/MS. Francois Espourteille, Ph.D. Manager, Applications

Validation Data for the Analysis of Chlorotalonil in Leek Using Mini-Luke Method Followed by GC-QqQ-MS/MS

High-Resolution Separation of Intact Monoclonal Antibody Isoforms

Corn Syrup Analysis E-1a- 1. ACETALDEHYDE and ISOVALERALDEHYDE (Gas Chromatography)

Analytically Monitoring the Effect of Fracturing Activity

Parallel LC with Capillary PS-DVB Monolithic Columns for High-Throughput Proteomics

Implementation of a Walk-Up High- Pressure Capillary Ion Chromatograph for the Fast Separation of Pharmaceutical Relevant Inorganic Anions and Cations

Thermo Scientific icap 7000 Plus Series ICP-OES. Gain more power. experience more performance

Development of an industrial standard method for Total Petroleum Hydrocarbon (TPH) analysis

Gas Chromatography SSI-GC-003

Simplifying Complex Multi-Residue Pesticide Methodology in GC-MS/MS

Liner Style* Function Advantages Disadvantages Recommended For Straight Low surface area for less activity Least expensive Low activity

Automated sample preparation followed by sensitive analysis by GC-MS/MS for environmental contaminants in surface waters

Transcription:

CONSULENZE AMBIENTALI SPA Determination of Total Petroleum Hydrocarbons in Rubble and Soils by Accelerated Solvent Extraction and GC-FID Fabrizio Galbiati, 1 Luca Teli 2 1 Thermo Fisher Scientific (Schweiz) AG, Reinach, Switzerland 2 Consulenze Ambientali SpA, Scanzorosciate, Italy Customer Application Note 120 Key Words Pressurized Fluid Extraction, UNI EN 14039:2005, Hydrocarbons >C 12, Hydrocarbons C 10, In-Line Clean-Up, Rocket Evaporator, Gas Chromatography, Flame Ionization Detection Goal To demonstrate an accelerated solvent extraction and GC-FID procedure for total petroleum hydrocarbons (TPH) in contaminated soils and rubble. Introduction The hydrocarbon contamination of soil that occurs through the spillage or leakage of petrol, diesel, or lubricants is one of the most common forms of environmental pollution. Hydrocarbons have been routinely determined by infrared spectroscopy after extraction with halogenated solvents such as 1,1,2-trichloro-1,2,2-trifluoroethane (Freon 113) or tetrachloromethane. Because of ozone depletion concerns, the use of chlorofluorocarbons has been banned or severely limited in many countries. Alternative methods like Soxhlet extraction (United States Environmental Protection Agency (U.S. EPA) Method 3540) and ultrasonic extraction (U.S. EPA Method 3550) are presently used for the extraction of hydrocarbons from soils prior to their analytical determination. Those techniques are very labor intensive and suffer from high solvent consumption, often requiring 18 hours and 500 ml of solvent per sample. Accelerated Solvent Extraction was developed to reduce the amount of solvent and time required for extraction of organic contaminants from solid samples. With Accelerated Solvent Extraction, extractions can be completed in as little as 20 minutes and use less than 50 ml of solvent per sample. Accelerated Solvent Extraction uses elevated temperature to improve the extraction efficiency of organic contaminants from solid samples. This technique also uses elevated pressure to keep solvents in a liquid state above their boiling point in order to achieve rapid and reproducible extractions. Due to the use of elevated temperature, interferences may be extracted along with desired analytes during those conventional extraction processes. These unwanted co-extractables may interfere with analyte detection or decrease instrument performance. Traditionally, chromatographic techniques such as gel-permeation chromatography (GPC) or a glass column packed with specific adsorbents are used to clean sample extracts prior to separation and analysis. For example, the determination of the hydrocarbon content in solid waste according to the UNI EN 14039:2005 requires a postextraction clean-up in a glass column with Florisil. Recent advances using Accelerated Solvent Extraction systems, as described in several publications, 1-18 include procedures for selective removal of interferences during sample extraction, thus combining extraction and purification into a single step. This advance eliminates the need to perform offline sample clean up procedures and reduces the amount of time spent on sample preparation prior to analysis with GC. The method reported here is applicable to soils with a hydrocarbon content (>C 12 ) between 25 and 4000 mg/kg and waste with a hydrocarbon content (C 10 ) between 50 and 8000 mg/kg expressed on a dry matter basis. All hydrocarbons with a boiling range of approximately 175 C to 525 C, e.g. n-alkanes from C 10 H 22 to C 40 H 82, isoalkanes, cycloalkanes, alkyl benzenes, alkyl naphthalenes, and polycyclic aromatic compounds are determined as total petroleum hydrocarbons (TPH).

2 Experimental Equipment An ED53 oven from Binder was used for drying the samples. A Sartorius analytical balance was used for weighing the dry soil samples and the standards. A Fritsch Pulverisette planetary ball mill was used to grind the samples. The extractions were carried out with the Thermo Scientific Dionex ASE 350 Accelerated Solvent Extractor (P/N 083114, 120 V, or P/N 083146, 240 V) (Figure 1) equipped with 34 ml stainless steel extraction cells (P/N 068089). The extracts were collected in 60 ml double-ended Thermo Scientific Rocket Flip-Flop tubes (P/N 076360) (Figure 2) and directly concentrated in a 2 ml autosampler glass vial (Thermo Scientific Chromacol VAGK ISP: GC 2-SVW + 9-SCK(B)-ST1) with the Thermo Scientific Rocket Evaporator (P/N 075904, 120 V or 082766, 240 V) (Figure 3). Figure 1. Dionex ASE 350 Accelerated Solvent Extractor. Figure 2. Rocket Flip-Flop Vial. Figure 3. Rocket Evaporator. The samples were analyzed with a Thermo Scientific FOCUS GC Gas Chromatograph equipped with a split/ splitless injector, a Thermo Scientific TRACE TR-5 GC column (30 m 0.25 mm 0.25 µm, P/N 260E142P), and a flame ionization detector (FID). Solvents and Standards Pesticide-grade hexane (Fluka, P/N 34484) was the extraction solvent for ASE. Thermo Scientific Dionex ASE Prep Diatomaceous Earth (DE) (P/N 062819) was used. Silica gel (70-230 mesh for column chromatography, P/N 453337) was purchased from Carlo Erba Reagents. n-decane (P/N D901), n-dodecane (P/N D221104) and n-tetracontane (P/N 87087) were purchased from Sigma-Aldrich. n-octadecanoic acid octadecyl ester (P/N 46408) was purchased from Fluka. Diesel standard (without additives, DIN H53, P/N CA03009000) and mineral oil standard (without additives, DIN H53, P/N CA03009010) were purchased from Dr. Ehrenstorfer GmbH. Mineral Oil Standard Mixture Type A and B at a concentration of ~8000 mg/ml in heptane (P/N 69246, exact content on the label) was purchased from Fluka. Sodium sulfate (P/N 31481) was purchased from Sigma-Aldrich. Standard solutions with concentrations of 4000.0, 2000.0, 1000.0, 500.0, 200.0, and 100.0 mg/l were prepared from an 8000 mg/l stock solution of both diesel and mineral oil standards. The silica gel was activated by heating at 550 C for at least 4 h. The test solution of n-octadecanoic acid octadecyl ester (C 36 H 72 O 2 ) was prepared by dissolving 100 mg of n-octadecanoic acid octadecyl ester in 100 ml n-hexane. The clean-up efficiency of each batch of silica gel was checked by adding 10 ml of the test solution to a clean-up column filled with 2.0 g of silica gel and 2 g of sodium sulfate. After chromatographic analysis of the eluate the percent recovery of the n octadecanoic acid octadecyl ester should not exceed 5%. The retention time window (RTW) standard solution for C 10 was prepared by weighing 30 mg of n tetracontane into a 1 L volumetric flask, followed by the addition of n-hexane and 30 μl (about 21 mg) of n-decane. The retention time window (RTW1) standard solution for >C 12 was prepared by weighing 30 mg of n-tetracontane into a 1 L volumetric flask, followed by the addition of n-hexane and 35 μl (about 26 mg) of n-dodecane.

Extraction, Concentration, and Measurement A cellulose filter (Thermo Scientific, P/N 056780) was placed in the bottom of a 34 ml extraction cell, followed by 5 g of activated silica gel and another cellulose filter. Four gram samples of soil from residential areas, 2 g of soil from industrial areas, or 2 g of rubble were mixed in a glass beaker with a sufficient amount of Dionex ASE Prep DE and the resulting mixture carefully poured into the extraction cell. Any empty volume was filled with Dionex ASE Prep DE. The Accelerated Solvent Extractor system was programmed according to the conditions reported in Table 1. Table 1. Conditions for Accelerated Solvent Extraction. Solvent n-hexane Temperature 100 C Static Cycles 1 Extraction Time 5 min Rinse Volume 60% Purge Time Total Extraction Time per Sample Total Solvent Volume per Sample Table 2. Gas chromatograph conditions. Carrier Gas Flow Rate Oven Temperature Helium 90 s 20 min 40 ml The extracts were directly collected into Flip-Flop vials and evaporated on the Rocket Evaporator using the preprogrammed Low Boiling Point Method until nearly dry. The leftover was reconstituted with 1 ml of the RTW or RTW1 solution according to the type of analysis requested and injected in the GC. The GC conditions are summarized in Table 2. Injector Temperature (SSL) 300 C Injection Mode Splitless Time Splitflow 2.0 ml/min, constant flow 45 C (hold 1 min), 25 C/min to 340 C (hold 2 min) Splitless 0.80 min 50 ml/min Detector Temperature (FID) 340 C Injected Volume 2 µl Results and Discussion Soils are an extremely complex matrix to analyze, particularly when they originate from contaminated sites. The nature and composition of the soil has a significant impact on the extraction efficiency and the sensitivity of the method. The typical soil consists of approximately 45% mineral, 5% organic matter, 20 30% water, and 20 30% air. In particular, the organic matter contains high molecular weight, non-volatile material that can remain in the GC system thus leading to poor analytical performance. Many analysts use extensive sample preparation techniques to extract and concentrate the compounds of interest from this unwanted nonvolatile material. These extraction and concentration techniques can however become time consuming and costly. Construction and demolition rubble samples are also of concern because of their possible contamination with hydrocarbons and other compounds like polycyclic aromatic hydrocarbons (PAHs) and polychlorinated biphenyls (PCBs). According to the UNI EN 14039:2005, the soil sample is extracted with an acetone - n-heptane mixture. Acetone is then removed from the extract with water. The resulting organic phase is subjected to solid phase extraction with Florisil to remove the organic matter such as humus and humic compounds, and the eluate is analyzed by GC-FID. The same norm also mentions Accelerated Solvent Extraction as an alternative extraction procedure; however, it is still followed by a manual clean-up with Florisil. Several recent publications describe procedures for selective removal of interferences during Accelerated Solvent Extraction, thus combining extraction and purification into a single step. 3

4 Due to the compositional complexity of petroleum products, it is impossible to assess the extent of petroleum hydrocarbon contamination by separately measuring the concentration of each hydrocarbon contaminant. One parameter that is currently widely used for expressing the total concentration of nonpolar petroleum hydrocarbons in soil is termed TPH. It is a non-specific, method-defined parameter that is determined by GC-FID. The total peak area of resolved and unresolved components in the chromatogram range delimited by the retention times of n-decane and n-tetracontane is integrated (Figure 4). Only semi- and non-volatile hydrocarbons are therefore included in the TPH parameter. The first extractions were done using spiked soils that were previously heated for 4 h at 800 C in order to completely remove the organic matter. The TPH spike levels were 25, 50, and 750 mg/kg. Twenty samples were extracted for the fraction >C 12, and twenty samples for the fraction C 10. The test results showed that spike recovery rates were between 84.0% and 101% for >C 12 and 79.0% and 96.6% for C 10. The UNI EN 14039:2005 additionally prescribes a maximum RSD of 9.14% for a low contaminated soil and 8.13% for a highly contaminated soil (corresponding to 697 mg/kg and 1818 mg/kg, respectively). For rubble, the maximum RSD is 6.05%, corresponding to 7841 mg/kg. With a maximum RSD of 5.7% for >C 12 and 5.5% for C 10, both parameters were in full compliance with the requirements of the European norm. A typical chromatogram for a 750 mg/kg spiked sample is showed in Figure 4. All the results are summarized in Tables 3 and 4. The concentration of the contaminated soil was around 4000 mg/kg as shown in Figure 5. The sample was processed according to the procedure described in the Extraction, Concentration, and Measurement section. The diesel oil fraction, corresponding approximately to C 11 -C 18, can be clearly recognized as the main pollutant in this sample. Table 3. Average recovery and RSD for >C 12. Concentration mg/kg 25 50 750 88.9 105.0 99.3 81.8 106.8 101.3 84.4 105.0 100.5 87.6 99.1 101.2 90.8 99.8 99.4 85.2 94.7 102.2 81.9 101.6 97.8 79.9 107.2 99.4 87.8 95.5 103.7 78.1 106.2 102.2 76.1 101.2 100.0 90.1 98.4 100.0 89.8 89.7 101.6 81.5 103.3 100.0 90.8 102.1 104.7 75.4 91.3 103.6 84.2 92.7 99.8 85.9 95.2 95.8 79.0 109.3 100.4 80.9 95.9 98.2 Average Recovery % 84.0 100.0 100.5 RSD 5.8 5.7 2.1 Table 4. Average recovery and RSD for C 10. Concentration mg/kg 25 50 750 77.9 101.6 102.0 80.3 101.3 100.7 80.9 90.2 95.9 82.7 98.6 98.0 88.8 101.5 93.5 81.1 90.6 93.7 78.6 96.6 86.8 77.6 104.5 94.9 76.5 95.3 96.3 73.5 98.9 93.4 72.7 91.8 97.9 84.2 94.9 86.7 85.6 87.1 93.9 73.1 99.0 100.3 78.8 103.5 98.6 75.5 89.0 104.3 81.4 94.1 102.0 83.4 89.1 94.1 70.2 103.3 89.3 77.3 100.8 105.0 Average Recovery % 79.0 96.6 96.4 RSD 6.0 5.7 5.4 Response [mvolts] C40 C10 Figure 4. Chromatogram for 750 mg/kg spiked soil sample after ASE-extraction and concentration. Minutes

5 Response [mvolts] C40 C10 Figure 5. Chromatogram for a heavily contaminated soil sample after ASE and concentration. Minutes Conclusion When using Accelerated Solvent Extraction with in-cell cleanup to extract hydrocarbons from contaminated soils, the spike recovery is between 84.0% and 100.5%. The selective removal of interferences with the in-cell cleanup avoids time-consuming and costly post-extraction manual purification procedures. Processing a sample using Accelerated Solvent Extraction requires only 20 min and only 40 ml of solvent. The Rocket Evaporator eliminates the need for cumbersome nitrogen stream evaporation. By using the Flip-Flop system the samples can be concentrated directly into the GC vial. Acknowledgments Special thanks to Luca Teli of Consulenze Ambientali SpA (a UNI CEI EN ISO/IEC 17025:2005 accredited company), Scanzorosciate, Italy, who performed the work on the soil samples and his supervisor Raffaella Gibellini for the permission to use the data for this Customer Application Note. References 1. Ezzell, J.; Richter, B.; Francis, E. Selective Extraction of Polychlorinated Biphenyls from Fish Tissue Using Accelerated Solvent Extraction, American Environmental Laboratory, 1996, 8(12), 12-13. 2. Thermo Fisher Scientific Inc., Selective Extraction of PCBs from Fish Tissue Using Accelerated Solvent Extraction (ASE), Application Note 322, LPN 0764, Sunnyvale, CA, 1996. 3. Thermo Fisher Scientific Inc., Determination of PCBs in Large-Volume Fish Tissue Samples Using Accelerated Solvent Extraction (ASE), Application Note 342, LPN 1204, Sunnyvale, CA, 2000. 4. Björklund, E.; Muller, A.; von Holst, C. Comparison of Fat Retainers in Accelerated Solvent Extraction for the Selective Extraction of PCBs from Fat-Containing Samples, Anal. Chem., 2001, 73, 4050-4053. 5. Sporring, S.; Björklund, E. Selective Pressurized Liquid Extraction of Polychlorinated Biphenyls from Fat- Containing Food and Feed Samples: Influence of Cell Dimensions, Solvent Type, Temperature and Flush Volume, J. Chromatogr., A., 2004, 1040, 155-161. 6. Haglund, P.; Sporring, S.; Wiberg, K.; Björklund, E. Shape-Selective Extraction of PCBs and Dioxins from Fish and Fish Oil Using In-Cell Carbon Fractionation Pressurized Liquid Extraction, Anal. Chem., 2007, 79, 2945-2951. 7. Thermo Fisher Scientific Inc., Determination of Perchlorate in Vegetation Samples Using Accelerated Solvent Extraction (ASE) and Ion Chromatography, Application Note 356, LPN 1830, Sunnyvale, CA, 2006. 8. Gentili, A.; Perret, D.; Marchese, S.; Sergi, M.; Olmi, C.; Curini, R. Accelerated Solvent Extraction and Confirmatory Analysis of Sulfonamide Residues in Raw Meat and Infant Foods by Liquid Chromatography Electrospray Tandem Mass Spectrometry, J. Agric. Food Chem., 2004, 52, 4614-4624. 9. Björklund, E.; Sporring, S.; Wiberg, K.; Haglund, P.; von Holst, C. New Strategies for Extraction and Clean-up of Persistent Organic Pollutants from Food and Feed Samples Using Selective Pressurized Liquid Extraction, Trends in Anal. Chem., 2006, 25(4), 318-325.

10. Hussen, A.; Westbom, R.; Megersa, N.; Retta, N.; Mathiasson, L.; Björklund, E. Optimisation of Pressurized Liquid Extraction for the Determination of p,p -DDT and p,p -DDE in Aged Contaminated Ethiopian Soils, Anal. Bioanal. Chem., 2006, 386(5), 1525-1533. 11. Hussen, A.; Westbom, R.; Megersa, N.; Mathiasson, L.; Björklund, E. Development of a Pressurized Liquid Extraction and Clean-up Procedure for the Determination of Alpha-Endosulfan, Beta-Endosulfan and Endosulfan Sulfate in Aged Contaminated Ethiopian Soils, J. Chromatogr. A, 2006, 1103, 202-210. 12. Sporring, S.; von Holst, C.; Björklund, E. Selective Pressurized Liquid Extraction of PCB s from Food and Feed Samples: Effects of High Lipid Amounts and Lipid Type on Fat Retention, Chromatograpia, 2006, 64(9-10), 553-557. 13. Wiberg, K.; Sporring, S.; Haglund, P.; Björklund, E. Selective Pressurized Liquid Extraction of Polychlorinated Dibenzo-p-dioxins, dibenzofurans and dioxin-like polychlorinated biphenyls from Food and Feed Samples, J. Chromatogr. A, 2007, 1138, 55-64. 14. Hussen, A.; Westbom, R.; Megersa, N.; Mathiasson, L.; Björklund, E. Selective Pressurized Liquid Extraction for Multi-residue Analysis of Organochlorine Pesticides in Soil, J. Chromatogr. A, 2007, 1152, 247-253. 15. Thermo Fisher Scientific Inc., Fast Determination of Acrylamide in Food Samples Using Accelerated Solvent Extraction (ASE) Followed by Ion Chromatography with UV or MS Detection, Application Note 409, LPN 1497, Sunnyvale, CA, 2003. 16. Thermo Fisher Scientific Inc., Extraction and Cleanup of Acrylamide in Complex Matrices Using Accelerated Solvent Extraction (ASE) Followed by Liquid Chromatography, Tandem Mass Spectrometry (LC-MS/MS), Application Note 358, LPN 1922, Sunnyvale, CA, 2007. 17. Poerschmann, J.; Carlson, R. New Fractionation Scheme for Lipid Classes Based on In-Cell Fractionation Using Sequential Pressurized Liquid Extraction, J. Chromatogr. A, 2006, 1127, 18-25. 18. Poerschmann, J.; Trommler, U.; Biedermann, W.; Truyen, U.; Lucker, E. Sequential Pressurized Liquid Extraction to Determine Brain-Originating Fatty Acids in Meat Products as Markers in Bovine Spongiform Encephalopathy Risk Assessment Studies, J. Chromatogr. A, 2006, 1127, 26-33. Customer Application Note 120 www.thermofisher.com/chromatography 2016 Thermo Fisher Scientific Inc. All rights reserved. Freon is a registered trademark of DuPont Corp. Binder is a trademark of BINDER GmbH. Sartorius is a trademark of Sartorius AG. Fritsch and Pulverisette are trademarks of Fritsch GmbH. Fluka is a trademark of Fluka AG. Carlo Erba Reagents is a trademark of Carlo Erba Reagents S.A.S. Sigma-Aldrich is a registered trademark of Sigma-Aldrich Co. is a trademark of LGC Limited. Florisil is a registered trademark of U.S. Silica Company Corp. All other trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is presented as an example of the capabilities of Thermo Fisher Scientific products. It is not intended to encourage use of these products in any manners that might infringe the intellectual property rights of others. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details. Africa +43 1 333 50 34 0 Australia +61 3 9757 4300 Austria +43 810 282 206 Belgium +32 53 73 42 41 Brazil +55 11 3731 5140 Canada +1 800 530 8447 China 800 810 5118 (free call domestic) 400 650 5118 CAN71843-EN 08/16S Denmark +45 70 23 62 60 Europe-Other +43 1 333 50 34 0 Finland +358 10 3292 200 France +33 1 60 92 48 00 Germany +49 6103 408 1014 India +91 22 6742 9494 Italy +39 02 950 591 Japan +81 6 6885 1213 Korea +82 2 3420 8600 Latin America +1 561 688 8700 Middle East +43 1 333 50 34 0 Netherlands +31 76 579 55 55 New Zealand +64 9 980 6700 Norway +46 8 556 468 00 Russia/CIS +43 1 333 50 34 0 Singapore +65 6289 1190 Sweden +46 8 556 468 00 Switzerland +41 61 716 77 00 Taiwan +886 2 8751 6655 UK/Ireland +44 1442 233555 USA +1 800 532 4752