Determination of Oxalate and Other Anions in Bayer Liquor Using Ion Chromatography

Similar documents
Determination of Manganese in Brine ( 30% NaCl)

Minimizing Carbon Dioxide Contamination in Anion Reagent-Free Ion Chromatography

Introduction. Yan Liu, Victor Barreto, Rosanne Slingsby, Chris Pohl Thermo Fisher Scientific, Sunnyvale, CA, USA

Practical guidance for switching between anion and cation analysis systems

Effect of Temperature on IC Signal Noise

Determination of Inorganic Anions in Acid Rain Using a Dedicated High-Pressure Capillary Ion Chromatography System

Analysis of Fracking Flowback Water From the Marcellus Shale Using In-line Conductivity, Automated Dilution, and Ion Chromatography

Complete Customer Support

Implementation of a Walk-Up High- Pressure Capillary Ion Chromatograph for the Fast Separation of Pharmaceutical Relevant Inorganic Anions and Cations

Options Eluent Regeneration (RFIC-ER ) Column Heater In-Line Vacuum Degas Full Chromeleon Chromatography Data System

Configuring the Dionex Modular HPIC system with an inline water purifier and inline calibration for trace anion determinations in ultrapure water

Thermo Scientific Dionex ICS-900 Ion Chromatography System

Configuring the Dionex Integrion HPIC system for trace anion determinations in ultrapure water

columnsi o n P a c C S 1 6 C a t i o n - E x c h a n g e C o l u m n

( ) IONPAC ATC-HC ANION TRAP COLUMN

An Automated System for At-Line Process Analysis of Biopharmaceutical Fermentation Reactions

Reducing Carbonate Interference in Anion Determinations with the Carbonate Removal Device (CRD)

Determination of Trace Concentrations of Oxyhalides and Bromide in Municipal and Bottled Waters Using a Compact Ion Chromatography System

columns IonPac ICE-AS1 Ion-Exclusion Column Superior Chromatographic Performance Separates weakly ionized acids by differences in pk a

PRODUCT MANUAL FOR THE. ELECTROLYTIC ph MODIFIER (EPM) EPM, P/N

Passion. Power. Productivity. Corporate Headquarters

Dionex IonPac AS16 Column

Advanced Features New Advanced Automation Features Precise Performance Temperature Control Speed Application Flexibility

Dionex IonPac AS19 Columns

Fast Ion Determinations and Efficient Separations of Complex Samples Using 4 µm Particle-Size Ion Exchange Columns

Analysis of Glyphosate and AMPA in Environmental Water Samples by Ion Chromatography and On-Line Mass Spectrometry with Minimal Sample Preparation

In-Cell and External Hydrolysis of Biomass for Carbohydrate Determinations Using Automated Solvent Extraction

Fast mass transfer Fast separations High throughput and improved productivity Long column lifetime Outstanding reproducibility Low carryover

Determination of Perchlorate in Drinking Water Using a Reagent-Free Ion Chromatography System

Dionex Eluent Generator Cartridges

New Developments in Capillary Ion Chromatography Systems with Electrochemical Detection and Their Applications

Dionex IonPac AS11-HC-4µm

Continuously Regenerated Trap (CR-TC 500, Capillary)

Capillary Ion Chromatography and High-Pressure Ion Chromatography: Applications and Solutions

Optimize Your Process Operations by Improving Process Monitoring

Dionex IonPac AS14 Columns

Dionex IonPac AS11-HC-4µm

Next generation suppressor for ion chromatography. Thermo Scientific Dionex ERS 500 Electrolytically Regenerated Suppressor

Improved Determination of Trace Perchlorate in Drinking Water Using 2D-IC

Thermo Scientific. Dionex IonPac AS9-HC. Column Product Manual. P/N: May Part of Thermo Fisher Scientific

Eluent Generator Cartridges

Determination of Total Nitrogen and Phosphorus in Environmental Waters

Determination of Perchlorate in Drinking Water by Ion Chromatography

Enjoy Industry-Leading Support

Product Manual for the Dionex IonPac Ultra Trace Anion Concentrator (UTAC)

Determination of Trace Copper, Nickel, and Zinc in Boiling Water Reactors Using Nonsuppressed Conductivity Detection

Application Note 168

Product Manual for Dionex IonPac AS24A and AG24A Columns

Assay for Citrate and Phosphate in Pharmaceutical Formulations Using a High-Pressure Compact Ion Chromatography System

Thermo Scientific Dionex ICS-4000 Integrated Capillary HPIC System. Always ready, better results. capillary high-pressure ion chromatography

Integrated IC Systems Installation Qualification Dionex Corporation

Determination of Cations in Hydraulic Fracturing Flowback Water from the Marcellus Shale

Fluorine and Chlorine in Cu concentrates and ores by Metrohm Combustion IC

Product Manual for Dionex IonPac AS4A-SC and AG4A-SC Columns

For new orders of the following parts discussed in this manual, please use the updated part numbers listed below.

Concentration of Human Hormones in Drinking Water Using Solid Phase Extraction and Analysis by High Performance Liquid Chromatography

for IonPac AG9-HC IonPac AS9-HC

Thermo Scientific Dionex DRS 600 Dynamically Regenerated Suppressor

Application Note 171 INTRODUCTION

Determination of halogens in coal using combustion ion chromatography

Determination of Sulfite in Food and Beverages by Ion Exclusion Chromatography with Pulsed Amperometric Detection

Determination of Total and Potential Sulfate and Total Chloride in Fuel-Grade Butanol Per ASTM D

columns PepSwift and ProSwift Capillary Monolithic Reversed-Phase Columns

User Manual. Dionex Guardcap H Revision 02 March For Research Use Only. Not for use in diagnostic procedures.

Sprint analysis of lubricating oils using the Thermo Scientific icap 7600 ICP-OES

Dionex DRS 600 Suppressor Dionex ERS 500e Suppressor Dionex ERS 500 Carbonate Suppressor

Configuring a High-Pressure Dedicated Capillary IC System for Electrochemical Detection

Determination of Morpholine, Ethanolamine, and Hydrazine in Simulated Nuclear Power Plant Wastewater

Chem 321 Lecture 23 - Liquid Chromatography 11/19/13

Dual EGC A simple solution for complex carbohydrate analysis

A Novel ph Gradient Separation Platform for Monoclonal Antibody (MAb) Charge-Variant Analysis

Thermo Scientific Dionex ICS HPIC Capillary and Analytical System. High resolution, fast analysis high-pressure ion chromatography

AS Autosampler. of front panel control.

HPAE-PAD analysis of galactosyl-oligosaccharidecontaining samples using dual eluent generation cartridges

Dionex IonPac CS16-4µm and Dionex IonPac CS16-Fast-4µm

Sensitive Determination of Hexavalent. Brian De Borba, Lipika Basumallick, Jeffrey Rohrer

PRODUCT MANUAL. IONPAC AG14A GUARD COLUMN (4 x 50 mm, P/N ) (3 x 30 mm, P/N )

MMS TM 300. (MicroMembrane TM Suppressor 300)

A Fast and Robust Linear ph Gradient Separation Platform for Monoclonal Antibody (MAb) Charge Variant Analysis

Analysis of Disinfection Byproducts by Ion Chromatography

Two-Dimensional LC Protein Separation on Monolithic Columns in a Fully Automated Workflow

Nano LC at 20 nl/min Made Easy: A Splitless Pump Combined with Fingertight UHPLC Nano Column to Boost LC-MS Sensitivity in Proteomics

Identification of RNA Linkage Isomers by Anion-Exchange Purification with ESI-MS of Automatically-Desalted Phosphodiesterase-II Digests

Determination of Trace Anions in Ultrapure Water Using Capillary Ion Chromatography

electrochemical detection

Dionex Guardcap Products

Investigations into Method Improvements for Hexavalent Chromium. Presented by Aaron Rose Thermo Fisher/Dionex Products For 2012 Watercon

Introduction. Three different cases for different dosage forms and pharmacopeias, respectively, must be distinguished (Table 1).

Automated Solid-Phase Extraction of Organochlorine Pesticides from Drinking Water. Pranathi R Perati Thermo Fisher Scientific, Sunnyvale, CA, USA

Thermo Scientific. Dionex CRS 500. Product Manual. P/N: March Part of Thermo Fisher Scientific

A HIGHLY VERSATILE IC SYSTEM CAPABLE OF EVERY CHALLENGE THAT COMES YOUR WAY THERMO SCIENTIFIC DIONEX ICS HPIC SYSTEM

Improving Retention Time Precision and Chromatography of Early Eluting Peptides with Acetonitrile/Water Blends as Solvent B

Errata. Product Manual for Dionex IonPac Cryptand Columns

Thermo Scientific. Dionex IonPac CS17. Product Manual. P/N: November Part of Thermo Fisher Scientific

Automated Solid-Phase Extraction of Organochlorine Pesticides from Drinking Water

Dionex IonPac CS19 Columns

Determination of trace anions in ultrapure water using capillary ion chromatography

Parallel LC with Capillary PS-DVB Monolithic Columns for High-Throughput Proteomics

HIC as a Complementary, Confirmatory Tool to SEC for the Analysis of mab Aggregates

CX-1 ph Gradient Buffer

Transcription:

Application Note 206 Determination of Oxalate and Other Anions in Bayer Liquor Using Ion Chromatography Introduction The Bayer Process is used to isolate alumina (Al 2 O ) from bauxite. In this process, the manufacturer dissolves bauxite in sodium hydroxide to extract the aluminum from other components which are not soluble in sodium hydroxide, mainly oxides. 1 The solution is referred to as Bayer liquor, and the temperature of this process depends on the mineral composition of the bauxite. The next steps precipitate aluminum hydroxide [Al(OH) ] that is used to form alumina, which is used in the aluminum smelting process. Alumina is formed by placing Al(OH) crystals in a kiln and driving off the water, a process referred to as calcination. Many variables determine the success of calcination to form alumina, and one is the size of the Al(OH) crystals. Large crystals are preferred, and after precipitation only the large crystals are put into the kiln for calcination. Small crystals are returned to the precipitator. Oxalate can interfere with crystal formation, and therefore a measurement of the oxalate concentration helps predict the success of forming large crystals from a given Bayer liquor. Ion chromatography (IC) is the method of choice for determining oxalate as well as other organic acids and inorganic anions in Bayer liquor, though the sample s high ionic strength and metal content make it a challenging sample. To address these challenges, a direct injection method was designed using a Reagent Free Ion Chromatography (RFIC ) system with in-line sample cleanup to accurately and reproducibly measure oxalate in Bayer liquor. The online sample preparation removes most of the aluminum so that it does not contaminate either the column or suppressor. The remaining aluminum oxalate retained by the preparation devices is recovered by rinsing with a small volume of sodium fluoride. Fluoride is used to promote ligand exchange with oxalate for aluminum. This online sample preparation method allows an accurate, reproducible, and rugged determination of oxalate in Bayer liquor.

Equipment Dionex ICS-2000 or ICS-000 RFIC system AS40 or AS autosampler CIRA 11AXCS Electrolytic Purifier (Trovion, P/N 91218) CIRA installation kit (Trovion, P/N 9120) PC-100 Pump Controller (Trovion, P/N 90100) consisting of a 2-channel peristaltic pump and -way selection valve 1 L plastic container (P/N 06291) 4 L plastic container (P/N 06292) Chromeleon 6.8 Chromatography Data System For ICS-2000: IonPrep Controller (Trovion, P/N 9120) consisting of a 10-port valve and constant current power supply For ICS-000: AM Automation Manager (P/N 06176) consisting of a 10-port valve and a -way valve PS-0 Power Controller (Trovion, P/N 9000) Reagents and Standards Deionized water (DI), Type I reagent grade, 18 MΩ-cm resistivity. Sodium chloride (NaCl) Malonic acid [CH 2 (COOH) 2 ] Sodium sulfate (Na 2 SO 4 ) Sodium oxalate (Na 2 C 2 O 4 ) Sodium fluoride (NaF) All compounds should be ACS reagent grade or better from reliable sources. Conditions Guard Column: IonPac AG17, 4 0 mm (P/N 0684) Analytical Column: IonPac AS17, 4 20 mm (P/N 0682) Trap Column: IonPac UTAC LP1 (P/N 06079) Eluent Source: EGC II KOH cartridge (P/N 08900) with CR-ATC (P/N 060477) Eluent Concentration: Time (min) KOH conc (mm) 7.0 6 0.0 6.0 6.0 14 8.0 14 12.0 0 12.1 0 17.9 0 18.0 6 Flow Rate: 1.0 ml/min Inj. Volume: 10 µl Temperature: 0 ºC Pressure: 2100 psi Detection: Suppressed conductivity, ASRS 00 suppressor, external water mode, 124 ma 2 Determination of Oxalate and other Anions in Bayer Liquor using Ion Chromatography

Preparation of Solutions and standards Eluent Solution The eluent generator (EG) produces the eluent using the EluGen EGC II KOH cartridge and deionized water supplied by the pump. The eluent concentration is controlled by Chromeleon software. Because the IonPac AS17 column set does not produce 2000 psi of backpressure, backpressure tubing must be added to achieve the 2100 2200 psi backpressure that will allow the EG degasser to function properly. 0.1% Sodium Fluoride Dissolve 1 g of NaF in 1000 ml of deionized water. Place the solution in a 1 L plastic container. Standard Solutions Stock Standards Prepare 1000 mg/l standards for each of the anions and organic acids in DI water. Standards should be prepared from the highest purity compounds available. Table 1 provides the amounts needed to prepare 100 ml of each standard. Concentrated standards are stable for at least one month when stored at 4 C. Table 1. Masses of Compounds Used to Prepare 100 ml of 1000 mg/l Standard Standard Compound Amount (g) Chloride Sodium chloride 0.16 Malonate Malonic acid 0.102 Sulfate Sodium sulfate 0.148 Oxalate Sodium oxalate 0.12 Working Standards Prepare working standards on the day of use from the stock standards. For this experiment, four mixed standards were made for system calibration. The concentrations of those standards are shown in Table 2. Sample Preparation Prediluted (200-fold with deionized water) Bayer liquor samples were kindly provided by a customer. To remove the high concentration of aluminum from the sample in an automated manner prior to injection, use a CIRA 11AXCS Electrolytic Anion Purifier. The CIRA 11AXCS consists of two depletion chambers where one chamber removes all ions to produce UPW necessary for flushing sample from the loop. The other chamber is used to remove metals, including aluminum, from the sample. To ensure that there is no secondary retention of oxalate in the matrix removing chamber, flush the chamber with 0.1% sodium fluoride to elute any remaining aluminum oxalate. System Preparation and Setup The following steps describe fluidic connections of the 10-port valve on both the IonPrep Controller (ICS 2000 installation) and the Automation Manager (ICS 000 installation). 1. After installing the columns and suppressor with the suppressor effluent entering the conductivity cell, locate the in-line filter/pressure relief assembly in the CIRA installation kit (P/N 9120). Connect the filter assembly to the CD outlet port and to CIRA 11 AXCS (P/N 91218) port #1. 2. Connect CIRA port #4 to 10-port valve port #1 with an appropriate length of 0.020 i.d. (orange) tubing.. Connect CIRA port #2 to 10-port valve port #6 with an appropriate length of 0.020 i.d. (orange) tubing. 4. Connect CIRA port # to injection valve port S with an appropriate length of 0.020 i.d. (orange) tubing.. Connect 10-port valve port #8 to the -way valve outlet port labeled C, located in the rear panel of the PC 100 Pump Controller with an appropriate length of 0.020 i.d. (orange) tubing. Connect the -way valve port labeled B to the 1 L plastic container (0.1% NaF solution). Plug port labeled A with a ¼-28 end fitting. Table 2. Standard Concentrations for Method Calibration and Calibration Results Analyte Concentration (mg/l) Results Level 1 Level 2 Level Level 4 % R 2 Offset Slope Chloride 1 10 0 100.000-0.0074 0.270 Malonate 1 10 0 99.97 0.0467 0.0891 Sulfate 1 10 0 99.999 0.068 0.192 Oxalate 1 10 0 99.90 0.0762 0.114 Application Note 206

If using an Automation Manager (AM), the NaF line is connected to the port labeled open on the AM -way valve. Plug the port labeled closed with a ¼-28 end fitting. 6. Connect 10-port valve port #9 to the peristaltic pump inlet with an appropriate length of 0.020" i.d. (orange) tubing. 7. Install the remaining channel of the peristaltic pump inlet to the 4 L external water reservoir (for ASRS external water mode). Connect the pump outlet line to the rest of the regenerant ports, preferably CR-ATC regen in and out to ASRS regen in. 8. Install a sample loop (10 µl) between ports #2 and # of the 10-port valve. 9. Install a 2 µl loop between ports #10 and #7 of the 10-port valve. 10. Install the autosampler lines (in and out) to ports #4 and #, respectively, of the 10 port valve. 11. Install the concentrator column on the injection valve in place of an injection loop. 12. Now looking at the diagram below, connect the ASRS Regen waste line to CIRA port #. 1. Install CIRA waste line at port #6. ICS-2000 IonPrep Controller and PC-100 Pump Controller Set power selector located at the rear of the controller to an appropriate voltage (120 V or 240 V). In addition to main power input, connect the 24 V DC power adapter to the IonPrep Controller rear panel. Use appropriate power cord to power up the controller. The IonPrep Controller supplies a constant current to the CIRA. The IonPrep 10-port valve is used for sample injection and NaF addition. The PC-100 Pump Controller delivers DI water to regeneration chambers of the CR-ATC, ASRS, and CIRA. The second channel delivers the NaF solution to the IonPrep Controller. The -way valve is used for NaF line selection. TTL Connections Connect TTL1 and TTL2 cables of the IC to IonPrep Valve TTL and Pump Controller Valve TTL located on the rear panels of the two components. If required, use Relay1 and Relay2 of the IC to control CIRA and pump ON/OFF by connecting Relay cables to appropriate TTL/Relay connections at the rear panel of the two components. In normal operation, the CIRA and pump are switched ON and OFF manually. P C Guard Analytical Water (PC-100) CR-ATC L CD ASRS W S Eluent Generator Waste Concentrator Filter CIRA 1 4 2 DI Water 6 Waste Sample In Waste 1 7 PC-100 9 NaF B C A 261 CIRA Power Connection Locate the CIRA cable in the CIRA Installation Kit. Connect this cable to CIRA and to CIRA Power Cable port at the rear of the IonPrep. Use default Medium setting for routine applications. Turn CIRA power ON only when PC-100 Pump flow is on. ICS-000 Automation Manager, PS-0 Power Controller, and PC-100 Pump Controller Set power selector located at the rear of the PC-100 and PS-0 to an appropriate voltage (120 V or 240 V). The AM 10-port valve is used for sample injection and NaF addition. The PS-0 supplies a constant current to the CIRA. The PC-100 Pump Controller delivers DI water to regeneration chambers of the CR-ATC, ASRS, and CIRA. The second channel delivers NaF solution to the AM. The -way valve is used for NaF line selection. Figure 1. Diagram of the matrix elimination setup. TTL Connections Connect TTL2 cables of the IC to the Pump Controller Valve TTL located on the rear panel of the controller. If required, use Relay1 and Relay2 of the IC to control CIRA and pump ON/OFF by connecting Relay 4 Determination of Oxalate and other Anions in Bayer Liquor using Ion Chromatography

cables to appropriate TTL/Relay connections on the rear panel of the PS-0 and PC-100. In normal operation, the CIRA and pump are switched ON and OFF manually. CIRA Power Connection Locate the CIRA cable in the CIRA Installation Kit. Connect this cable to CIRA and to CIRA Power Cable port at the rear of the PS-0. Use the default Medium setting for routine applications. Turn CIRA power ON only when PC-100 Pump flow is on. Fluidic Flow The ASRS effluent from the CD outlet flows into CIRA port #1 and through port #4 where all ions are removed [ultrapure water (UPW) is produced]. In A position (10 port), the purified effluent flows through the 10-port valve ports #1, #10, #7, and #6, respectively. Sample or standard solution is loaded at this time. The purified effluent continues to flow into the CIRA and the injection valve (Inject position) and out to waste. The matrix elimination begins when the 10-port and injection valves switch to B and load positions, respectively. The purified effluent from port #1 is diverted to ports #2, #, and #6 where the sample is flushed. The sample band flows into CIRA port #2 and through # where all metals are removed. The anions continue with the flow from the CIRA to the concentrator column. During this time, the -way valve switches to the B position, which allows the NaF solution to flow into the 2 µl loop. After 1.0 min of sample flush, the 10-port valve returns to the A position where NaF starts to flow into CIRA ports #2 and #. Typically, ml is required to completely flush the sample to the concentrator column. The analysis starts immediately after the injection valve switches to the inject position. Valve Control Program Time TTL1 TTL2 Injection (min) (10-port) (-way) (6-port) -7.0* V (A) V (A) Inject -6.0 Load -.0 0 V (B) 0 V (B) -4.0 V (A) V (A) 0.0 Inject *sample loading must be completed Results and Discussion Separation To keep analysis times short, the IonPac AS17, a lower capacity anion-exchange column, was chosen to separate oxalate from sulfate, malonate, chloride, and other anions. Figure 2 shows a separation of four concentration levels of a mixed standard of chloride, malonate, sulfate, and oxalate. The large early eluting peak is fluoride, from the subsequent injection of 0.1% sodium fluoride during the matrix elimination. If the separation of less retained organic acids such as acetate and formate is required, a longer separation can be performed using an IonPac AS19 while omitting the addition of the sodium fluoride. The AS19 separation was developed by a customer who also uses a carbonate removal device (CRD 200) in their method to simplify the quantification of some of the anions. There is also a published report of using an IonPac AS11-HC column to separate formate, acetate, propionate, oxalate, succinate, glutarate, fluoride, chloride, and sulfate in Bayer liquor. 1 Sample Analysis Prior to sample analysis the system was calibrated with the standards in Table 2. Calibration results are shown in the same table. The results showed good linearity for each ion in the chosen concentration range. 100 µs 1 2 Columns: IonPac AS17, IonPac AG17 Concentrator: UTAC-LP1 Eluent Source: EGC II KOH with CR-ATC Gradient: See Conditions Flow Rate: 1.0 ml/min Inj. Volume: 10 µl Temperature: 0 C Detection: Suppressed conductivity, ASRS 00, external water mode, 124 ma Peaks: 1. Fluoride 0.1% NaF 2. Chloride 1,, 10 and 0 mg/l. Malonate 1,, 10 and 0 4. Sulfate 1,, 10 and 0. Oxalate 1,, 10 and 0 20 0 6 9 12 1 18 Minutes 262 Figure 2. Overlay of chromatograms of the blank and four standards at concentrations of 1,, 10, and 0 mg/l. 4 Application Note 206

To test the method we analyzed 12 Bayer liquor samples provided by a customer. Table shows the sample analysis results. The customer confirmed that the results in Table were accurate. The chromatogram in Figure, an overlay of three injections of sample #1, and the results of those injections, presented in Table 4, show the method s precision. The customer that uses the IonPac AS19 separation with the sample preparation method in this application note, but without the co-injection of 0.1% sodium fluoride, has found the method to be accurate and rugged, with replacement of neither column nor suppressor in over 6 months of sample analysis. Conclusion This application note describes an accurate and rugged method using an RFIC system with online sample preparation to determine oxalate and other ions in Bayer liquor. This method can be very useful for aluminum producers in determining the potential success of a given Bayer liquor for forming large Al(OH) crystals prior to calcination. References 1) Xiao, J. B. J. of the Chilean Chem. Soc. 2006, 1 (), 964-967. Table. Bayer Liquor Sample Analysis Results (mg/l) Sample Name Chloride Malonate Sulfate Oxalate Sample #1 4.47 11.9 44.84 16.77 Sample #2 29.87 9. 7.48 11.88 Sample # 4.80 12.14 4.7 16.77 Sample #4 47.10 10.70 6.0 12.26 Sample # 29.64 16.19 108.72 21.1 Sample #6 2.8 10.09 80.28 19.1 Sample #7 2.47 16.47 114.21 27.40 Sample #8 4.68 12.08 4. 16.7 Sample #9.6 n.a. 8..0 Sample #10 6.1 1.88 1.48 2.10 Sample #11 n.a. 9.80 86.4 2.0 Sample #12.20 11.92 47.2 17.09 Analyte Table 4. Reproducibility for Three Injections of Sample #1 (RSD) Retention Time Amount Height Plate (EP) Chloride 0.074 0.680 0.74 0.102 Malonate 0.161 0.667 0.068 1.727 Sulfate 0.116 0.782 0.01 1.469 Oxalate 0.09 0.611 0.27 0.99 100 µs 1 2 Columns: IonPac AS17, IonPac AG17 Concentrator: UTAC-LP1 Eluent Source: EGC II KOH with CR-ATC Gradient: See Conditions Flow Rate: 1.0 ml/min Inj. Volume: 10 µl Temperature: 0 C Detection: Suppressed conductivity, ASRS 00, external water mode, 124 ma Sample: Bayer Liquor Peaks: 1. Fluoride 0.1% NaF 2. Chloride 4.47 mg/l. Malonate 11.9 4. Sulfate 44.84. Oxalate 16.77 20 0 6 9 12 1 18 Minutes 26 Figure. Overlay of three consecutive runs of sample #1. Reagent-Free and RFIC are trademarks and ASRS, Chromeleon, EluGen, and IonPac are registered trademarks of Dionex Corporation. 4 Passion. Power. Productivity. Dionex Corporation 1228 Titan Way P.O. Box 60 North America U.S./Canada (847) 29-700 South America Europe Austria (4) 1 616 1 2 Benelux (1) 20 68 9768 (2) 4294 Denmark (4) 6 6 90 90 France () 1 9 0 01 10 Germany (49) 6126 991 0 Asia Pacific Australia (61) 2 9420 2 China (82) 2428 282 India (91) 22 2764 27 Japan (81) 6 688 121 Korea (82) 2 26 280 Singapore (6) 6289 1190 Sunnyvale, CA Brazil () 11 71 140 Ireland () 1 644 0064 Italy (9) 02 1 62 1267 Sweden (46) 8 47 80 Taiwan (886) 2 871 66 LPN 2091 PDF 06/09 94088-60 Switzerland (41) 62 20 9966 United Kingdom (44) 1276 691722 (408) 77-0700 www.dionex.com 2009 Dionex Corporation