Trace Moisture Contamination in Ultra-High Purity Phosphine: Techniques for Measurement and Control

Similar documents
The MIDAC Low Moisture in Corrosive Gases Process Analyzer for Continuous On-Line Applications

Trace Water Vapor Analysis Using Cavity Ring-down Spectroscopy, Oscillator Quartz Crystal and Impedance Sensor Technologies

SPARK TM MOISTURE ANALYZER VALIDATION Performed in Conjunction with Air Products and Chemicals, Inc.

Background Statement for SEMI Draft Document #5765 Revision to SEMI C TEST METHOD FOR ppm AND ppb HUMIDITY STANDARDS

Background Statement for Ratification Ballot of SEMI Draft Document #5765 Revision to SEMI C TEST METHOD FOR ppm AND ppb HUMIDITY STANDARDS

Fpollutant-free zero air for analytical

LD8000. DESIGN REPORT v2. Trace Nitrogen in Argon or Helium analyzer

Aurora TDLAS analyzers for Hydrogen Recycle Applications

TDLAS. Application Note. Tunable Diode Laser Absorption Spectroscopy (TDLAS) for Carbon Dioxide Measurements In Natural Gas Processing

High Flow Delivery Systems for Bulk Specialty Gases

Fluorescence & UV- Visible Workshop

Avoiding oxygen and moisture contamination in ultra-high purity gases for semi-conductor manufacture

Onsite quantitative FTIR analysis of water in turbine oil

Process Analytics Using Quantum Cascade Laser

Aurora Trace HDLAS Moisture Analyzer

REMOVAL OF VOLATILE METALS FROM CORROSIVE GASES BY A NEW PURIFER

Measuring trace impurities in HyCO, regeneration hydrogen and inert gases Los Gatos Research (LGR)

PROCESS ANALYZERS FOR PURIFIED BULK GAS QC/QA: ENTERING THE PPT RANGE

Measuring trace impurities in HyCO, regeneration hydrogen and inert gases Los Gatos Research (LGR)

Field-tests at Kiviõli Keemiatööstus OÜ oil-shale processing plant in Kiviõli, Estonia

Proceedings of the 43 rd Annual ISA Analysis Division Symposium

Application Note. Hydrogen Purity Analysis by FTIR PROBLEM BACKGROUND

Greenhouse Gas Measurement Technology

Simplicity. Accuracy. Durability. Lotix. Automated High Temperature Combustion TOC Analyzer

Developing a Fluorinated Greenhouse Gas Stack Testing Method using FTIR

APPLICATION NOTE. UOP : Analysis of Trace CO and CO 2 in bulk H 2 and Light Gaseous Hydrocarbons by GC

Electron Capture Detector for GC-2010 Plus

GOOD LABORATORY PRACTICE Agilent Cary 8454 UV-Visible Spectroscopy System

Chem 344 slides for Kinetics Methods extras!

ETHYLENE PLANT DRY CRACKED GAS Product Code 54801

Lab 6 Measurement of Ozone

Aurora Moisture Analyzer New Light for Moisture Measurement in Carbon Dioxide

Olefin Analyzer Solutions for Polymer Grade Ethylene and Propylene

TDLAS. Application Note. Use of TDLAS Based Analyzers for Real-Time Process Control in Ethylene Oxide Production PROCESS INSTRUMENTS

APPLICATIONS BENEFITS INTELLIGENT

Lab 7 Measurement of Ozone

EDXRF APPLICATION NOTE IRON IN ALUMINUM # 1237

icap 7000 Plus Series ICP-OES Streamlined performance and ultra-low ICP-OES detection limits for routine laboratories

Co-Combustion of Wood and Straw in a Circulating Fluidized Bed Boiler reduction of NO x emission

Measuring moisture in ultra-high-purity gases

Fundamental Kinetics Database Utilizing Shock Tube Measurements

open path monitoring 법응용 대기가스측정및분석

Monitoring and control of binary gas mixtures from solid phase MOCVD sources using an acoustic sensor

ABB Measurement & Analytics Trace HCl measurements in coal-fired power plant emissions using Cavity Enhanced Laser Absorption Spectroscopy

D DAVID PUBLISHING. Hydrogen Gas Purifiers for Fuel Cells. 1. Introduction. 2. Scientific Approach

Trace HCl measurements in coal-fired power plant emissions using Cavity Enhanced Laser Absorption Spectroscopy

Moisture removal from hydrogen chloride (HCl) gas

On-Line Gas Analysis In Air Separation Plants

Water Vapor and Carbon Nanotubes

MOISTURE MEASUREMENT IN BIOMETHANE. Biomethane for Injection to Natural Gas Network

Simplicity. Accuracy. Durability. Lotix. Automated High Temperature Combustion TOC Analyzer

Maximizing Sample Recovery on the ACCQPrep HP125

Sapphire SF: Low-Noise Ultra-Narrow Bandwidth Performance Delivers Superior Raman Spectroscopy Results

CO Sensor Response Study NO & NOx Analyzer Validation. June 6, 2018

Detection of Plutonium with the Microwave Plasma Continuous Emissions Monitor*

Fusion UV/Persulfate TOC Analyzer

CEMS for Measurement of Ammonia, SO 3, and Low Level NOx

Advanced Spectroscopic Solutions for Fuel Cell Hydrogen Analysis

March 2016 ARM, Inc. white paper: Safe Approach to Gas Purification. Abstract

Characterization of Emissions and Occupational Exposure Associated with Five Cover Gas Technologies for Magnesium Die Casting

HydrocarbonEngineering. May2009. Reprinted from.

A Rapid Method for Trace Analysis of Organophosphorus Pesticides in Drinking Water

Determination of Hydrocarbons in Environmental Samples with Spectrum Two

INDUSTRIAL MEASUREMENTS USING FTIR GAS ANALYZER

Impact of electron irradiation on N- and O-enriched FZ silicon p-in-n pad detectors

The development of instrumentation for application to carbon capture and storage measurements

Utility of Variable Bandwidth Monochromators for Quantification of Fluorescent Probes in Produced Effluent Water

NANOCHEM Purifiers. Complete Purification Solutions with

MB3600. Versatile FT-NIR analyzer designed for your industry. The most reliable FT-NIR specifically designed for QA / QC.

The Proper Mechanism for Field Validating Moisture Analyzers. GE Measurement & Sensing GE Oil & Gas

WHITEPAPER: AMMONIA MEASUREMENT FOR COMBUSTION SOURCES

Proudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment Parker N2-04

Fiber and Electro-Optics Research Center Virginia Polytechnic Institute and State University Blacksburg, Virginia 24061

Modified Draft Indian Standard METHODS FOR CHEMICAL ANALYSIS OF STEELS

Title: Standard Operating Procedure for Measurement of CO in Ambient Air by Gas Filter Correlation (GFC)

Final Report. Characterisation of Sample Report. Job No 2016/11/12-34 AS No. 1234A. Client Example Contact Sample. Signed. Example report.

ABB MEASUREMENT & ANALYTICS DATA SHEET. PGC5007 Total sulfur analyzer

EPA APPROVED METHOD FOR TURBIDITY ANALYSIS 360 x 90, Turbidity Method 10258

GDMS. High Resolution GDMS.

INTERPLANT STANDARD - STEEL INDUSTRY SPECIFICATION FOR INFRA-RED PYROMETERS IPSS:

Properties of Inclined Silicon Carbide Thin Films Deposited by Vacuum Thermal Evaporation

Other GHGs. IPCC Climate Change 2007: The Physical Science Basis

FOUNDRY-MASTER Pro. First class laboratory optical emission spectrometer for complete professional metal analysisoes

Analysis of Mineral and Heavy Metal Content in Beverages Using the Teledyne Leeman Labs Prodigy Plus ICP-OES

Application Note: Product Pipeline Interface Detection by NIR Spectroscopy

Laser measurement of KCl vapor in 4 MW CFB boiler during straw combustion and ferric sulfate injection

Precision Humidity Solutions

Application Note 6: On-line Reaction Monitoring of Alcoholic Fermentation

Field-test at Betonstal Sp. z.o.o. Szczecin/ Poland

Analysis of strongly absorbing chromophores by UV-visible ATR spectroscopy

SYNTHESIS, CHARACTERIZATION AND PHOTOCATALYTIC ACTIVITY OF MgO NANOPARTICLES

Agilent 8453 Spectrophotometer and UV-visible ChemStation for Dissolution Testing. Solution Bundles for automated Dissolution Testing

Two Great Technologies Together At Last. The 700 Series NanoTrace Moisture And Oxygen Analyzers

Analysis campaign for measurement of oxy-combustion flue gas from Coal Fired Plants

Process Total Organic Carbon Analyzer

Ratio and Spectral Scanning

EMISSION MEASUREMENT TECHNICAL INFORMATION CENTER NSPS TEST METHOD. Method 10 - Determination of Carbon Monoxide Emissions from Stationary Sources

Revision of 30 April 2013 draft, 4 November 2013

Chapter-IV OPTICAL PROPERTIES

Development of clean coal technology by using the feature of oxy-fuel combustion

Transcription:

Trace Moisture Contamination in Ultra-High Purity Phosphine: Techniques for Measurement and Control M. Raynor, H. Funke, J. Yao, T. Watanabe and R. Torres, Matheson Tri-Gas, Advanced Technology Center, 1861 Lefthand Circle, Longmont, Colorado K. Bertness, Optoelectronics Division, National Institute of Standards and Technology, Broadway, Boulder, Colorado Pittsburgh Conference on Analytical Chemistry and Applied Spectroscopy, March 7-12, 2004, Chicago, Illinois

Introduction Gas purity is critical to the production of III-V compound semiconductor devices Trace O from impurities such as H 2 O, are incorporated into epitaxial layers during MOCVD processes cause lattice defects and undesirable oxide formation decrease luminescence of optoelectronic components Objectives: Investigate methods for measurement of trace H 2 O impurity in PH 3 Demonstrate moisture control through point-of-use purification

Measurement Techniques Requirements of Trace Analysis Instrumentation Sensitivity and selectivity for H 2 O in PH 3 at ppb levels Rapid and accurate response Rugged and capable of withstanding reactive gas Long term stability Ease of use and applicability to on-line monitoring Techniques Investigated Chilled mirror hygrometry Cavity ring-down spectroscopy FTIR spectroscopy

Chilled Mirror Hygrometry Frost point is defined as the temperature to which a volume of gas must be cooled such that it becomes saturated with respect to ice. Exponential relationship between frost point and H 2 O concentration is well documented in N 2 Fundamental measure of H 2 O concentration Capable of long-term accuracy and stability Calibration required in PH 3 Condensation temperature of the PH 3 matrix gas limits the ultimate sensitivity of the technique

Chilled Mirror Hygrometer Cryogenic Sensing Chamber and Mirror Assembly Cooling Unit PH 3 in The mirror temperature is cooled until a frost layer is established. The mirror temperature is adjusted to maintain a constant condensation layer on the mirror. The layer is monitored optically and displayed as a balance voltage PH 3 out Buck Research CR-3 Chilled Mirror Hygrometer

Setup for Evaluation of Chilled Mirror Hygrometer Purified N 2 in N 2 in dry N Moisture Generator To hood Moisture Permeation Tube use to establish frost layer MFC Pressure transducer Mass Flow Meter To Scrubber PH 3 sample cylinder Purifier Chilled Mirror Hygrometer Gas Cabinet FTIR Purge N 2

Calibration Curve for Moisture in Phosphine using the Chilled Mirror Hygrometer -84-82 -80-78 -76-74 -72-70 -68 Frost Point Centigrade 2000 1800 1600 1400 1200 1000 800 600 400 200 0 ppbv Water Calibration points from 77-1730 ppbv moisture in PH 3 (1 slpm, 14.7 psia) Percentage of N 2 in PH 3 kept below 1% for all calibration points System equilibrated for ~4 hours at each moisture concentration

Moisture Detection in Phosphine Cylinder Sources Using Chilled Mirror Hygrometer 278 ppb H 2 O in PH 3 59 ppb H 2 O in PH 3 [H 2 O] (ppbv) 300 250 200 150 100 50 0-78 -79-80 -81-82 -83-84 -85-86 -87 0 60 120 180 240 300 Time (min) Frost Point ( C) [H 2 O] (ppbv) 120 100 80 60 40 20 0 Frost layer established 0 50 100 150 200 Time (min) -82-83 -84-85 -86-87 -88 Frost Point ( C) PH 3 from cylinder source, (N5.7 purity) at 1 slpm Long equilibration time PH 3 from cylinder source, (Ultima 6N purity) at 1 slpm Once established, frost layer maintained during analysis

Detection Limit [H 2 O] (ppbv) 70 65 60 55 50 45 Develop frost layer Lost layer Develop frost layer Lost layer 0 50 100 150 200 Time (min) Develop frost layer Lost layer -84.4-84.6-84.8-85 -85.2-85.4-85.6-85.8-86 -86.2 Temperature ( C) Mirror could not be cooled below minimum mirror set-point of -86 C due to potential condensation of liquid PH 3 H 2 O concentration in PH 3 less than detection limit

Observations with Chilled Mirror Hygrometer Detection of H 2 O in PH 3 possible down to ~55 ppb Reestablishing frost layer is time consuming and requires user intervention Mirror contamination occurs over time and results in noisy readings Regular cleaning Long equilibration time

Principle of Cavity Ring-Down Spectroscopy Light Light Time a Time Light is injected at a frequency, matched to an absorption frequency of the impurity being measured When the cavity and the input frequency are in resonance, the cavity fills with light The output light increases, triggering the shutter on the input light The light in the cavity recirculates and decays exponentially

Schematic of CRDS System at NIST Trigger Electronics AOM 936 nm Laser Scope/Digitizer Mode Matching Optics Cavity PH 3 (toxic) HeNe Cavity Length Servo Loop PZT (shutter) (freq. shifting) AOM

Ring-down Signals Normalized Light Power (arb. units) 1 8 6 4 2 0.1 8 6 4 2 0.01 8 6 4 0 100 200 A B 1.0 0.8 0.6 0.4 0.2 0.0 0 300 400 Time (µs) A B 200 500 400 600µs 600 700µs τ ( ν ) a = c L tot t = ring-down time a = cavity length c = speed of light (in cavity medium) 1 ( ν L tot = total loss per pass ) Ring-down time depends on base losses (frequency independent at this scale) absorption losses (frequency dependent)

CRDS Results: Dry cavity nitrogen Loss per pass x10 6 12.45 12.40 12.35 Water in Nitrogen at ν ~ =10697.42 cm -1 27 ± 2 nmol/mol -4000-2000 0 2000 4000 Frequency Shift (MHz) τ ( ν ) a = c L tot Water concentration is directly proportional to the peak area. 27 ± 2 nmol/mol water in nitrogen (27 ppb) n = 1 ( ν ) Model absorption lines as Voigt profiles on a linear baseline. Number density is then: Area a S a = cavity length S = absorption line strength

PH 3 Interference at 10687.36 cm -1 Loss per pass x10 6 22 20 18 16 Water in Phosphine at ν ~ =10687.36 cm -1 600 ± 350 nmol/mol Measured spectrum Fit to Voigt profiles H 2 O absorption PH 3 absorption -4000-2000 0 2000 4000 Frequency Shift (MHz) For low concentrations of H 2 O, there is too much interference from unmapped PH 3 absorption to accurately deconvolve the absorption lines Detection limit in N 2 at this absorption line < 15 nmol/mol (15 ppb)

Survey of Strong H 2 O lines near 935 nm 30 25 20 PH 3 at 6.65 kpa (50 Torr) 10667.76 cm -1 10687.36 cm -1 10697.42 cm -1 10700.67 cm -1 Loss per pass x10 6 15 10667.6 10667.8 10687.2 10687.4 10687.6 10697.2 10697.4 10697.6 10700.6 10700.8 10701.0 Wavenumber (cm -1 )

CRDS Spectrum: 6.65 kpa PH 3 at 10667.76 cm -1 15.5 14.0 H 2 O mole fraction = 590 ± 160 nmol/mol Pressure-broadening coefficient = 28 ± 7 MHz/kPa Background also changes with pressure Loss per pass x10 6 15.0 14.5 14.0 13.5 13.6 13.2 12.8-8000 -6000-4000 -2000 0 2000 4000 Measured spectrum Fit to Voigt profiles H 2 O absorption PH 3 absorption 13.0-2000 0 2000 4000 Frequency Shift (MHz) Estimated detection limit = 50 nmol/mol (50 ppb)

CRDS Summary Potential of CRDS for trace H 2 O detection in PH 3 demonstrated by NIST researchers Challenge to find a moisture line in 935 nm region where PH 3 absorptions are minimal Estimate 50 ppb detection limit using line at 937.40350 nm (10667.76 cm -1 ) Future Investigate other spectral ranges to lower detection limits

FTIR Spectroscopy Based on absorption of IR radiation by moisture impurity in a flowing gas stream within a gas cell Non-destructive technique Gas wetted parts can be selected for compatibility with phosphine Demonstrated sensitivity with long path-length cell Requires optimization of instrument design and measurement conditions

Infrared Spectrum of Dry PH 3 and Moisture in N 2 H 2 O detected in 1520-1560 cm -1 region with minimal interference from PH 3 absorptions Absorbance -0.118-0.119-0.120-0.121-0.122-0.123-0.124 990 ppb moisture in N 2 3.0 2.5 2.0 1.5 Dry Phosphine -0.125-0.126-0.127-0.128 1580 1570 1560 1550 1540 1530 1520 Wavenumber (cm-1) Dry phosphine 1510 1500 1490 Absorbance 1.0 0.5 0.0-0.5-1.0-1.5-2.0-2.5 4000 3500 990 ppbv Moisture in N 2 3000 2500 2000 Wavenumber (cm -1 ) 1500 1000 500 Calibration spectra input into classical least squares software for quantitation 512 spectra, 4 cm -1 resolution to maximize signal to noise Quantpad analysis program

FTIR Response IR absorption of moisture follows Beer s law (Log I/I o = -abc) Linear response for H 2 O standards diluted in dry PH 3 matrix LOD REG : 33 ppbv Dry N 2 bench purge required for long term stability Air must be excluded from sampling system to prevent PH 3 decomposition products affecting results CLS Response [ppmv] 1.2 1 0.8 0.6 0.4 0.2 0 y = 1.0783x + 0.0063 0 0.2 0.4 0.6 0.8 1 1.2 Moisture Concentration [ppmv] LOD REG based on ordinary least squares method (SEMI C10-0698)

FTIR Detection of Moisture in PH 3 0.1 Dry PH 3 0.1 PH 3 Ultima 6N Grade from Cylinder Source Moisture [ppmv] 0.08 0.06 0.04 0.02 0 Standard Deviation of Moisture data in Dry PH 3 is ± 10 ppbv Moisture [ppmv] 0.08 0.06 0.04 0.02 Moisture Equilibrates at 50 ppbv -0.02 0 60 120 0 0 50 100 150 200 Time (min) Time (min)

FTIR Observations FTIR currently method of choice for quantification of trace H 2 O in PH 3 Careful selection of H 2 O bands required to optimize sensitivity Detection to ~30 ppb possible with low resolution instrument, an MCT detector and a 10 m pathlength cell Improved sensitivity (10-20 ppb) likely with dedicated FTIR analyzers optimized for on-line gas measurements

Moisture Control Dry down of delivery systems exposed to atmosphere difficult due to adsorptive properties of moisture Gas phase moisture concentrations may vary over cylinder life, due to enrichment of water in the liquid phase Point-of-use purification provides a solution for consistent gas purity Efficiency of purifier materials evaluated using FTIR

Removal of H 2 O in PH 3 by Point of Use Purifier Removal of 5 ppm H 2 O in PH 3 by Nanochem PHX Purifier at 0.4 slpm -0.115-0.116-0.117 ~5 ppm H 2 O in PH 3 (Purifier Inlet Challenge) <100 ppb H 2 O reached in 10 min H 2 O level less than FTIR detection limit after 80 min Absorbance -0.118-0.119-0.120-0.121-0.122 0.25-0.123-0.124 0.2-0.125-0.126 Purified PH 3-0.127 H 2 O (ppm ) 0.15 0.1-0.128 1560 1 ppm H 2 O in N 2 1550 1540 1530 1520 Wavenumber (cm -1 ) 1510 1500 0.05 0 0 20 40 60 80 100 Time (min) IR spectra showing features of 5 ppm inlet H 2 O challenge and absence of features in purified PH 3

Summary and Outlook Current techniques allow H 2 O detection at tens of ppbs Performance of some III-V devices may be effected by H 2 O at much lower levels than can be measured analytically Point-of-use purification important for consistent gas purity Next generation analyzers should provide single digit and sub-ppb level sensitivity Laser based systems have greatest potential of filling sensitivity requirement