Regulations governing the application of medical accelerators

Similar documents
ISO 13485:2016 Medical Devices-Quality Management Systems- Requirements for Regulatory Purposes. Theresa McCarthy

Medical Device Directive

International Standards and EU regulation of medical device software an update

SOFTWARE APPS AS MEDICAL DEVICES. The Regulatory Landscape 2015 WHITE PAPER WHITE PAPER PRODUCED BY MAETRICS

Medical Device Software: A new challenge

Short intro to the MDR & IVDR (C)2018 QAdvis AB

Design Quality. Indu Lakshman

FREQUENTLY ASKED QUESTIONS DURING ELECTRO-MEDICAL DEVICE MARKET ACCESS

(C) QAdvis. IEC and IEC how to make them work (and why so much attention on SW) QAdvis RMD, Prague, November 8th 2016

1 Preface Introduction... 13

Regulation of software a medical device

The Complete Guide to FDA Design Controls

Implications of the new MDR from a Product Testing and Certification Perspective

Automated System Validation By: Daniel P. Olivier & Curtis M. Egan

The challenges of software medical device regulation.

MEDICAL DEVICE. Technical file.

Medical Device Software Standards

Manufacturers. Factsheet for. of Medical Devices. Medical Devices Regulation (MDR) background. Act now to be ready on time!

Left to Our Own Devices Design Control & Risk Management Strategies for Combination Products

ASQ Tappan Zee Section Medical Devices, New Regulations and Standards. 26 th September / V. Fischer / Rev. 01

Manufacturers. Factsheet for. of Medical Devices. Medical Devices Regulation (MDR) background. Act now to be ready on time!

Software in Medical Devices

V&V Best Practices. CASSS, CMC Strategy Forum Steven W. Badelt, PhD Managing Partner Suttons Creek, Inc. SUTTONSCREEK.COM

Rocky Mountain Regulatory Affairs Society. Elements of and Maintenance/Remediation of the Design History File (DHF) March 15, 2017

The Risk Management + Design Controls Connection: What Device Makers Need to Know

BSI Audits for MDR Certification

Medical Device Regulation

Ready or Not: The New Medical Device Regulations Are Here!

Auditdata Your Partner in Audiology Solutions

Self-Care Medical Devices Framework. Dr. Simone Breitkopf Head HEOR, Governmental and Public Affairs

Experience with Directive 93/42/EEC (MDD)

AAMI Quality Systems White Paper

Post market Surveillance ISO EU Medical Device Regulation

Standard and Directives on Medical Devices

Regulatory Framework for Medical Device

Contents. List of Acronyms Preface

Manufacturers will not need EMC or conformity assessments but will need to illustrate the

Bridging gaps: medical device directive vs regulation. Geert Corstens 1 November 2018

QMS Aspects of the MDR (& IVDR)

EU MDR 10 Things Packaging Engineers Should Know

Development of Safety Related Systems

Regulatory Affairs in Medical Technology

Choices IEC rd Edition and Component Selection

Copyright GCI, LLC Remediation of Legacy Medical Devices

GS1 Ireland Healthcare User Group (HUG) Information Day

EU MDR Timeline. Dr. Christian B. Fulda Jones Day FDLI Annual Conference Access materials at fdli.org/annual2018

Technical documentation in the new MDR. Gabriele Catignoli, Andreas Purde

EURATOM Booklet. Ready for February 6th 2018?

A Continuous Improvement Approach for Medical Device Software Development Companies. Dr. Özden Özcan-Top - Dr. Fergal Mc Caffery

ISO 13485:2016. What s in it - and what does it mean?

Applying ISO14971 / IEC62304 / IEC A Practical Guide On How To Implement Risk Management

Safety cannot rely on testing

Medical Device Software Development:

SPECIFICATION STATEMENT OF WORK. Performance Specification for a radiotherapy linear accelerator

ehealth Suisse Checklists Addendum to the guideline for app developers, manufacturers and distributors

Accelerating Software Development

Technical Documentation

Medidée Services SA. Nano-Tera.ch. 05 February 2015 part 3. Regulatory fundamentals. Michael Maier

Results of the IEC Functional Safety Assessment

IVDR Breakout. Copyright 2017 BSI. All rights reserved.

Mapping Your Success Staying Current on Standards Under the EU Approach

Post Market Surveillance

Overview of the 2nd Edition of ISO 26262: Functional Safety Road Vehicles

Understanding IEC 62304

Food and Drug Administration Modernization Act of 1997: Modifications to the List of

Certificating a safety related part of a control system

Laboratorio di Tecnologie Biomediche

Model-Based Design for ISO Applications. April 2010

Contents. Regulatory Bodies... 13

Pre-Clinical Testing for Devices and Diagnostics

Panel Discussion: European Medical Device Regulations Preparing for the Storm Moderator: Lenita Y. Sims Spears, Senior Quality Consultant/Senior

Course Title ID Duration Basic Premium. Biological Evaluation of Medical Devices: A Risk-Based Approach N mins

Changes to the International Regulatory Environment

Concept of Risk Management in Medical Equipment Application of ISO in IEC rd Edition

Technical documentation. Capita Selecta proposed MDR

A REVIEW OF MARKET ENTRY REQUIREMENTS FOR RISK MANAGEMENT, WITH SPECIAL EMPHASIS ON FDA AND ISO COMPLIANCE

Essential Requirements Straumann Custom-made Dental Restorative Devices According to 93/42/EEC annex I as amended by 2007/47/EC

MDSAP AUDIT PROCESS. A Manufacturer s Perspective. Connie Hoy EVP Regulatory Affairs Cynosure, Inc.

RE: Docket No. FDA 2017 N 0041: Agency Information Collection Activities; Proposed Collection; Comment Request; Safety Assurance Case

IEC Functional Safety Assessment

FDA Medical Device HFE Guidance

11th International Workshop on the Application of FPGAs in Nuclear Power Plants

Medical Device Regulatory Framework 9 SEPTEMBER 2015 FUNDISA CONFERENCE JANE ROGERS

Checklist for the assessment based on the standards

The New EU Medical Device Regulation (MDR) An Active Device Lifecycle Approach Implementation and Remediation Activities

The Theranos 483s. What can we learn? David Amor, MS, CQA Principal, Medgineering

Conformity Assessment of Medical Devices Under The New MDR

The Quality System Regulation

CE marking is regarded as a passport for products to enter and circulate freely within the 30 counties forming part of the EEA.

Fill the gap! New product standard: IEC (C) 2016 QAdvis AB

CONSTRUCTION SECTOR STANDARDIZATION GUIDANCE DOCUMENT

FDA 21 CFR Part 820 vs. ISO 13485:2016 Comparison Table created by greenlight.guru

GENEESS Guideline. Best Practices for Electronics in Medical Devices. V1.3 October 2016

Co-ordination of Notified Bodies Medical Devices (NB-MED) on Council Directives 90/385/EEC, 93/42/EEC and 98/79/EC

REQUIREMENTS FOR ACCREDITATION OF INSPECTION BODIES

A roadmap to ISO implementation

U.S. Food and Drug Administration (FDA) Regulatory Pathways for Medical Devices

Commissioning of the Heidelberg Ion Beam Therapy Centre

Medical Devices for Clinical Use

ELECtro-Medical devices. in trusted hands. sgs is the world s leading

Transcription:

Regulations governing the application of medical accelerators in 50 minutes. marko.mehle@cosylab.com

2 1.The wonderland of STANDARDS AND REGULATIONS

3 Laws and standards Medical devices (and systems) are subject to regulations (laws!) it all starts here. to ensure safety and effectiveness of devices on the market. Every country/region has its own set of laws/regulations. Standards are written by (international, private) organizations. In theory, standards are NOT MANDATORY, but In practice, officially recognized standards need to be followed.

4 Medical Devices Directive 93/42/EEC* (replaced by new MDR) Code of Fed. Regulations Title 21 Dev. class I, IIa, IIb and III Quality Syst. EN13485 CE Marking Notified Body (private) Technical File Harmonized Standards EN Dev. class I, II and III. QMS Title 21 Part 820 510(k) or PMA FDA clearance DHF, DMR and DHR Recognized consensus std. AAMI/ANSI

5 93/42/EEC Contents Some important contents: Definitions. Classification into I, IIa, IIb or III. List of essential requirements for safety and effectiveness for all devices. Processes to be followed by manufacturers, such as quality, risk management, post market, etc. Instructs manufacturers to create and maintain a Technical File for every device Foresees the adoption of harmonized standards. Legal mechanism for market approval and entitles notified bodies.

6 Marking 1. Intended purpose and classification 2. Identify the essential requirements 3. Establish necessary processes in the organization 4. Develop the device accordingly 5. Verification and Validation Test against requirements (we ll talk about it latter) 6. Clinical evaluation: a) Find equivalent device and use available literature route and/or b) Clinical investigation route. 7. Finalize technical documentation 8. Contact notified body and perform certification 9. Monitor trough lifecycle

7 Harmonized Standards 1. Quality management system EN 13485:2016 Defines organization processes ++ Requires auditing the organization (not only the product!) 2. Risk management EN 14971:2012 3. Electrical safety EN 60601 series (next slide) 4. Software lifecycle EN 62306 5. Usability EN 62366 6. Other standards for specific applications (labeling, packaging, transport, environmental, etc.)

8 Simplified Standards Ecosystem FDA CFR Part 21 EC MDD 93/42/EEC ANSI/AAMI CEN/CENELEC QMS (820) (EN/AAMI) IEC 60601-1 Section 4: General Section 14: PEMS Section 16: ME Systems (EN) ISO 13485 (EN/AAMI) IEC 62304 Med. SW lifecycle 60601-2-1 Electron Linac 60601-1-2 EMC 14971 Risk Management 60601-2-64 Light Ion Beam 60601-2-68 Image guided 60601-1-3 X-ray 60601-1-6 Usability (EN/AAMI) IEC 62366 Application of Usability

9 Meet the 60601 family 60601-1:2007 General Requirements Sections 1 to 17 Collateral Standards 60601-1-2 EMC 60601-1-3 X-Ray 60601-1-6 Usability 60601-1-8 Alarms 60601-1-9 Envir. 60601-1-10 60601-1-11 Home 60601-1-12 Emergency Particular Standards 60601-2-1 Electron ACC 60601-2-3 Surgical.. 60601-2-8 therapeutic X- ray.. 60601-2-64 Light Ion Beam.. 60601-2-68 Image guided radiotherapy

10 Risk management: ISO 14971 Most widely recognized RM standard for medical devices (and required by law). Not only design phase => whole device lifecycle. Based on the following concepts: HAZARD - potential source of harm HARM - physical injury or damage to the health of people, or damage to property or the environment RISK - combination of the probability of occurrence of harm and the severity of that harm. A damage-centric system => it aims to reduce the risk of harm (contrast with FMEA!) Not only failures to devices: also misuse is covered.

11 Risk Management Process

12 Some important aspects Risk = (Probability, Severity). Traceability is Key: Every risk must be identified Every hazardous situation must be evaluated Every risk must be linked to mitigation Every mitigation must be verified (e.g. tested ) Risk Management activities must continue after the product goes to the market (!!) RM team members must be defined in advance, with proper qualifications/training. After each RM cycle, a RM report must be released

13 How does it look like?

14 Summary (I) Laws regulate the market of medical devices. Safety and effectiveness. Standards are used to demonstrate compliance with the laws. In EU, the process is called CE marking. It is about demonstrating fulfillment of Essential Requirements. Notified bodies. There are many standards; the most important ones are: Quality system 13485 Risk management 14971 The 60601 family Software lifecycle 62304 Design process must be documented.

15 2. Some practical aspects of a PT MACHINE AS A MEDICAL DEVICE

16 What device??? It is quite simple to think of an ECG machine, a thermometer or even an X-ray diagnostic. According to MDD, a PT system is a class IIb. medical device. But What is the medical device? Are the sub-components medical devices on their own? Also class IIb.? Do all the same standards apply for all subsystems and components? How to aggregate different components into a device with a single purpose?

A complex one Accelerator Beamline Machine Control System Machine Interlock Syst. Auxiliary systems Beam Diagnostics Patient Couch Gantry Positioning Lasers Nozzle Extender Collision Avoidance Control Electronics Imaging 2D X-Ray / CBCT Controls for Imaging Registration Software Patient protection system Fast interlocks Room locking, operator protection Scanning system Dose delivery Treatment Control System Medical Software, Interfaces to OIS

18 ME SYSTEMS The standard 60601-1, section 16 offers an interesting view. ME system composed of medical and non medical equipment. 3 Conditions: The ME SYSTEM as a whole shall not have unacceptable RISK. Within the PATIENT ENVIRONMENT, the level of safety for 60601-1 Outside PATIENT ENVIRONMENT, the level of safety for applicable standards. 60601-1 allows to exclude some parts of the system from the need to be developed according to medical standards.

19 System Architecture 60601 compliance is required PT System Might be excluded Beam Production and Transport Nozzle System Treatment and Positioning Control System Position Verification Accelerator Scanning Treatment Control X-Ray Beamline Dose Delivery Positioning Control X Ray tube Magnets SW Module 1 X Ray panel Beam Diag. SW Module 2 Registration SW Control System *but still needs to comply with all other applicable regulations for a device/system of its type Lasers

20 Use of Software Software plays a key role: system complexity is in fact driven by automation -> PEMS (60601-1 Section 14.) 1. EN 60601-1 PEMS clause applies, excepted: No BASIC SAFETY or ESSENTIAL PERFORMANCE* ISO14971 demonstrates no unacceptable risk Compliance requires that a development life-cycle be specified and followed 2. EN 62304 -> Software life cycle processes. Verification and Validation activities need to be conducted.

21 Goals of System Architecture To define a wise system architecture, to group components with similar characteristics and functions in subsystems. Risk management is used to determine overall risk and to find what components are high risk and which are low risk. Also, risk aggregation. Some system components may not need to comply with 60601-1 (and related standards). Classify software according to 62304 into classes A, B and C. Translate User Requirements into Engineering Requirements/Specifications.

22 Two types of requirements Product Requirements System/ Subsystem Reqs Functional and performance requirements corresponding to the intended use of the product. Applicable regulatory requirements The sources for product requirements are: User requirements formulated by clients respect their medical and business goals. Requirements known from the market. Regulatory requirements imposed or recommended by authority. derived from the product definition requirements (functional, performance and regulatory), translating them onto technical specification. Technical requirements for the entire system, potentially with notes for the user manual and service manual. Safety relevant requirements The sources for translating regulatory requirements into technical requirements are the corresponding standards. Safety relevant requirements are derived from risk control measures defined in the risk management process, and applicable safety and essential performance standards for medical devices.

23 V&V at first sight The infamous V-model : A concept from the Software world Building the thing right vs. building the right thing

24 Test case traceability Every requirement should be covered by single or multiple test cases Requirement Unique ID single or multiple Arch. and design Unique ID single or multiple single or multiple RM Mitigation Unique ID Risk Identification Risk Evaluation Risk Control RM Hazard Unique ID Test case Unique ID Test report Unique ID Risk Control Verification

25 Summary (II) A PT System is one of the most complex medical devices. The regulations and standards are easier to apply to simpler devices; for complex systems sometimes it is necessary to do some magic. 3 useful standards: 60601-1, sections 14 and 16 Risk Management EN 14971 IEC 60601-2-64 System level architecture plays a major role! Write good requirements; start EARLY in the process Traceability!

26 It s all about processeses 1. The laws and regulations request it 2. Technical standards require it 3. Good engineering practices recommend it 4. It gives you good tools and maneuvering space when things does not go as planned. 5. It s about finding the sweet spot between too much and too little :

THANK YOU! Marko Mehle COSYLAB

28 THOSE SLIDES AT THE END

29 Basic safety and essential performance Originally only safety, Essential Performance is new to the 3rd. Edition ESSENTIAL PERFORMANCE: performance necessary to achieve freedom from unacceptable RISK. [whether its absence or degradation causes unacceptable RISK.] BASIC SAFETY: freedom from unacceptable RISK directly caused by physical HAZARDS when ME EQUIPMENT is used under NORMAL CONDITION and SINGLE FAULT CONDITION. Only makes sense in the context of Risk Management! Effectiveness Essential performance Safety

30 V&V for FDA From 21. CFR 820.30 - Design controls Design verification: [ ] Design verification shall confirm that the design output meets the design input requirements. The results of the design verification, including identification of the design, method(s), the date, and the individual(s) performing the verification, shall be documented in the DHF. Design validation: [ ] Design validation shall ensure that devices conform to defined user needs and intended uses and shall include testing of production units under actual or simulated use conditions. Design validation shall include software validation and risk analysis, where appropriate. The results of the design validation,

31 V&V for MDD MDD 93/42/EEC, 12.1a: For devices which incorporate software or which are medical software in themselves, the software must be validated according to the state of the art taking into account the principles of development lifecycle, risk management, validation and verification.

32 Measurement of beam parameters The beam size shall be 3mm or less (1sigma) at the isocentre at the highest treatment energy at all gantry angles (Verification)

33 Measurement of beam parameters The maximum range of the beam in water shall be 37 g/cm^2 or greater for treatment, which is defined as 250MeV or greater in a water target. (Verification)

34 Mechanical tests and measurements The PPS shall be aligned with mechanical isocenter ±1mm. (Verification)