Oxygen Demand, Chemical

Similar documents
Oxygen Demand, Chemical

By Authority Of THE UNITED STATES OF AMERICA Legally Binding Document

Oxygen Demand, Chemical

Oxygen Demand, Chemical

Dimethylphenol Method Method LR (0.23 to mg/l NO 3 N or 1.00 to mg/l NO 3 ) TNTplus 835

DR/4000 PROCEDURE. OXYGEN DEMAND, Chemical. Colorimetric Measurement

By Authority Of THE UNITED STATES OF AMERICA Legally Binding Document

Dimethylglyoxime Method Method to 6.0 mg/l Ni TNTplus 856

DR/4000 PROCEDURE. OXYGEN DEMAND, Chemical. Colorimetric Measurement, 0 to 40 mg/l COD

USEPA 1 Periodate Oxidation Method 2 Method to 20.0 mg/l Mn (HR) Powder Pillows

Method (0.1 to 8.0 mg/l Cu) TNTplus 860

By Authority Of THE UNITED STATES OF AMERICA Legally Binding Document

Nitrogen, Simplified TKN (s-tkn ) DOC

Phosphorus, Reactive (Orthophosphate)

Scope and application: For water, wastewater, seawater, brine solutions, produced waters and hydraulic fracturing waters.

Nitrogen, Simplified TKN (s-tkn )

Phosphorus, Reactive (Orthophosphate)

Method 8017 (0.7 to 80.0 µg/l) Powder Pillows. Scope and Application: For water and wastewater; digestion is required to determine total cadmium.

USEPA 1,2 Bicinchoninate Method 3 Method 8506 (CuVer 1) and Method 8026 (CuVer 2) 0.04 to 5.00 mg/l Cu Powder Pillows or AccuVac Ampuls

3. Add 0.4 ml of. 7. Use a TenSette

Method to 500 µg/l CH 2 O Powder Pillows

Instrument Adapters Light shield DR 6000, DR 5000 DR 3900 LZV849 DR 3800, DR 2800 LZV646 DR or (A)

1-(2-Pyridylazo)-2-Naphthol PAN Method 1 Method to mg/l Mn (LR) Powder Pillows

3. Add 0.4 ml of. 7. Use a TenSette

DR/4000 PROCEDURE. IRON, Ferrous. Using Powder Pillows

Sulfate. USEPA 1 SulfaVer 4 Method 2 Method to 70 mg/l SO 4. Powder Pillows. Test preparation. Before starting.

Hardness, Calcium/Magnesium

Table 163 Instrument-specific information. Instrument Pour-thru Kit Cell orientation Adapter. DR 6000 LQV Arrow faces right

Phosphorus, Reactive (Orthophosphate)

Color, True and Apparent, LR

Sulfate DOC SulfaVer 4 Method 1. Method (0 70, and mg/l) Powder Pillows. Test procedure

Calcium and Magnesium; Chlorophosphonazo Rapid Liquid Method Method to 1000 µg/l Ca and Mg as CaCO 3 (ULR) Pour-Thru Cell

PHOSPHORUS, REACTIVE (0.00 to 5.00 mg/l PO 4

Scope and application: For water and wastewater. Turbid samples are filtered during the analysis.

1,5-Diphenylcarbohydrazide Method* (Powder Pillows or AccuVac Ampuls) USEPA accepted for wastewater analyses** Using Powder Pillows

Amperometric Titrator assembly. Digital Titrator. Beaker, low-form, 250-mL. Stir bar, octagonal, Teflon-coated, 50.8 x 7.9 mm

SOP-C-102 Determination of Chemical Oxygen Demand

FerroVer Method (Powder Pillows or AccuVac Ampuls) USEPA approved for reporting wastewater analysis (digestion is required; see Section 2 * )

3. Fill a 50-mL. 2. Press: 1 ENTER The display will show mg/l, Al and the ZERO icon.

Direct ISE Method Method to 4.00 mg/l NO 3 N Nitrate ISE

TPTZ Method * (Powder Pillows or AccuVac Ampuls) Using Powder Pillows

1,10 Phenanthroline Method * (Powder Pillows or AccuVac Ampuls) Using Powder Pillows

Oxygen Demand, Biochemical

Application Notes for COD Analysis DETERMINATION OF CHEMICAL OXYGEN DEMAND (COD) IN WATER AND WASTE WATER.

PaSEC Colorimeter Instructions DR850

Approved for NPDES (Editorial Revision 1978) Silica, Dissolved (Colorimetric)

CEL212 Environmental Engineering

METHOD #: Approved for NPDES (Issued 1971) Oxygen, Dissolved (Modified Winkler, Full-Bottle Technique) ANALYTE: CAS # O Oxygen

SOP_Chem_001 Chemical Analysis_ Chemical Oxygen Demand. Standard Operation Procedure Chemical Oxygen Demand Closed Reflux, Titrimetric Method

NACA Laboratory Equipment and Supplies

Direct ISE Method Method to 10.0 mg/l F Fluoride ISE

Why Hach Reagents? Alphabetical Listing of Reagents

USEPA Membrane Filtration Method Method m-tec. Scope and Application: For potable water, nonpotable water, recreation water and wastewater.

Ground Water & UST Samples: Containers, Preservation and Hold Times Table

DR/4000 PROCEDURE OXYGEN SCAVENGERS. 4. Fill a second sample cell (the blank) with 25 ml of deionized water.

METHOD DETERMINATION OF TURBIDITY BY NEPHELOMETRY. Edited by James W. O'Dell Inorganic Chemistry Branch Chemistry Research Division

Technical Process Bulletin

BOD/CBOD, TSS ph OPERATOR BASICS. Presented by: Marcy Bolek - Alloway

SUMMARY OF STANDARDS FOR WATER ANALYSIS ON COURSE WAT-E2120 SPRING 2017

Total Dissolved Solids Dried at 180ºC. Total Volatile Dissolved Solids at 550 C

Rev Experiment 10

3.0 DETECTION LIMIT 3.1. See Procedure ALQAP Procedure for the Determination of Uncertainty

Auburn University Department of Civil Engineering Introduction to Environmental Engineering CIVL Laboratory #6 Chemical Oxygen Demand (COD)

Heterotrophic Bacteria

Surface Water Samples: Containers, Preservation and Hold Times Table

METHOD 7550 OSMIUM (ATOMIC ABSORPTION, DIRECT ASPIRATION)

SOP-C-130 Determination of Total Solids and Percent Solids

CEE 311 Environmental Engineering I Fall 2015 Dr. Jagadish Torlapati BIOCHEMICAL OXYGEN DEMAND (BOD)

Standard Operating Procedure for GLNPO Turbidity: Nephelometric Method

Coliforms, Total and E. coli

IRON (Colorimetric) 2. Muffle Furnace: Equipped with pyrometer and capable of operating at controlled temperatures up to 600 C

Corn Starch Analysis B-58-1 SULFUR DIOXIDE

STP (off-line) Water Technologies & Solutions analytical procedure. colorimetric method DCA reagent 1 (5/5/1) 2 to 15 mg/l as STP

3. Fill a 50-mL. 2. Press: 1 ENTER The display will show mg/l, Al and the ZERO icon.

OBSERVATIONS OF AN OPERATOR IN THE LAB IMPROVING LABORATORY RESULTS. Shane Jordan-Hill. Merck Millipore

AUTOMATED WATER ANALYSIS

DR 1900 PORTABLE SPECTROPHOTOMETER

ENGI Environmental Laboratory. Lab #2. Solids Determination. Faculty of Engineering & Applied Science

STEPS TO DMR-QA SUCCESS. BY: Marcy Bolek

Cuvette Test LCK 555 BOD 5 /BOD [n]

Cuvette Test LCK 555 BOD 5 /BOD [n]

INSTRUCTION MANUAL. ba76126e01 03/2013. photoflex Series TEST INSTRUCTIONS AND SPECIFICATIONS, PROG. V 2.02 Y

Analysis of dissolved nitrite (NO 2- )

Cadmium Testing and On Site Calibration for Water Testing Detection Range: ppm

RELIABILITY OF TOC RESULTS IN ULTRAPURE WATER ANALYSIS

NPOC, VOC, TC-IC and TDN using Shimadzu TOC-V Protocols. Water Quality Analysis Laboratory at the University of New Hampshire

ImmuLux Human IL-6 Fluorescent ELISA Kit

Application Note AN1501

Suspended Metals Digestion Procedure

Standard Methods for the Examination of Water and Wastewater

Procine CRP ELISA Kit

Benefits of On-line Monitoring of Carbon, Nitrogen and Phosphorus

Cuvette Test LCK 555 BOD 5 /BOD [n]

Mouse RF IgM ELISA Kit

Module 12: Laboratory Overview Answer Key

SOP-C-109 Determination of Filterable Residue

:00 am 6:00 pm Mon-Thu (MT) 6:00 am 5:00 pm Fri (MT)

MAHATMA GANDHI MISSION S JAWAHARLAL NEHRU ENGINEERING COLLEGE, AURANGABAD. (M.S.)

Uranium Testing and Site Calibration for Water Testing Detection Range: 2 60ppb

Transcription:

without chloride removal, 10067 Manganese III Reactor Digestion Method (without chloride removal) 1 30 to 1000 mg/l COD Mn Scope and Application: For water and wastewater 1 U.S. Patent 5,556,787 DOC316.53.01102 Method 10067 Test preparation How to use instrument-specific information The Instrument-specific information table displays requirements that may vary between instruments. To use this table, select an instrument then read across to find the corresponding information required to perform this test. Table 1 Instrument-specific information Instrument Light shield Adapter DR 5000 DR 2800 LZV646 DR 2700 LZV646 DR/2500 DR/2400 5945700 Before starting the test: DR 2800 and DR 2700 only: Install the light shield in Cell Compartment #2 before performing this test. If the sample contains chloride, use the chloride removal method. To determine if the sample contains chloride, use Quantab Titrator Strips for low range chloride. If the sample COD value is not between 30 and 1000 mg/l, dilute the sample with deionized water to obtain this range. Multiply the final result by the dilution factor. Homogenize the sample for even distribution of solids and better accuracy and reliability. Stability of the reagent blank allows for reuse. Verify the reagent blank quality by measuring the absorbance of the blank vs. a clean COD vial filled with deionized water. The absorbance range should be about 1.4 1.5. If the sample boils during the digestion, the vial is not properly sealed. Test results will be invalid. Use finger cots to handle hot sample cells.spilled reagent will affect test accuracy and is hazardous. Do not run tests with spilled vials. See the DRB200 User Manual for selecting pre-programmed temperature applications. Page 1 of 6

Collect the following items: Description Quantity Blender 1 DRB200 Reactor 1 Light Shield 1 Manganese III COD Reagent Vials, 20 1000 mg/l COD 1 Pipet, TenSette (0.1 to 1.0 ml) 1 Pipet tips for TenSette Pipet 2 Test Tube Rack 1 Water, deionized varies See Consumables and replacement items for reorder information. Sample preparation and measurement Stored Programs 432 COD Mn III Start 1. Select the test. Insert an adapter or a light shield if required (see Instrument-specific information). 2. Turn on the DRB200 Reactor and heat to 150 C or use the COD program. 3. Homogenize 100 ml of sample for 30 seconds in a blender. If suspended solids are present, continue to mix the sample while pipetting. Pipet 0.5 ml of homogenized sample into one Mn III COD vial (the prepared sample) and 0.5 ml of deionized water into another Mn III COD vial (the blank). Page 2 of 6

Sample preparation and measurement (continued) 4. Cap and invert several times to mix. 5. Insert the vials in the DRB200 Reactor at 150 C. Close the protective cover. Digest for one hour. Digest more resistant organics and the blank for up to four hours. 6. Remove the vials and place them in a cooling rack for two minutes. If a vial develops a colorless upper layer and a purple lower layer, invert the vial several times to mix and proceed. 7. Cool the vials to room temperature in a cool water bath or with running tap water. This takes several minutes. Zero 8. Invert the vials several times to mix. 9. Wipe the blank and insert it into the 16-mm round cell holder. Wipe the blank and insert it into the 16-mm round cell holder. 10. ZERO the instrument. The display will show: 0 mg/l COD Mn 11. Wipe the sample and insert it into the 16-mm round cell holder. READ the results in mg/l COD Mn. Interferences Inorganic materials may also be oxidized by trivalent manganese and constitute a positive interference when present in significant amounts. Chloride is the most common interference. If chloride is known to be present in significant levels, the chloride needs to be removed with the vacuum pretreatment device. A simple way to determine if chloride will affect test results is to run routine samples with and without the chloride removal, then compare results. Other inorganic interferences (i.e., nitrite, ferrous iron, sulfide) are not usually present in significant amounts. If necessary, these interferences can be corrected after determining their concentrations with separate methods and adjusting the final COD test results accordingly. Ammonia nitrogen is known to interfere in the presence of chloride; it does not interfere if chloride is absent. Page 3 of 6

Sample collection, preservation and storage Accuracy check Collect samples in clean glass bottles. Use plastic bottles only if they are known to be free of organic contamination. Test biologically active samples as soon as possible. Homogenize samples containing solids to assure representative samples. Samples treated with concentrated sulfuric acid to a ph of less than 2 (about 2 ml per liter) and refrigerated at 4 C may be stored up to 28 days. Correct test results for volume additions. Standard solution method Note: Refer to the instrument user manual for specific software navigation instructions. Required for accuracy check: Potassium acid phthalate (KHP) Deionized water 1. Prepare an 800-mg/L COD standard solution as follows: a. Dry 0.6808 g of KHP overnight at 120 C. b. Dissolve the KHP in 1 liter of deionized water. 2. Use 0.50 ml of the 800-mg/L solution in place of the sample. 3. To adjust the calibration curve using the reading obtained with the standard solution, navigate to Standard Adjust in the software. Instrument DR 5000 DR 2800 DR 2700 DR/2500 DR/2400 Navigate to: OPTIONS>MORE>STANDARD ADJUST OPTIONS>MORE>STANDARD ADJUST OPTIONS>MORE>STANDARD ADJUST OPTIONS>STANDARD ADJUST OPTIONS>STANDARD ADJUST Method performance 4. Turn on the Standard Adjust feature and accept the displayed concentration. If an alternate concentration is used, enter the concentration and adjust the curve to that value. Program Instrument Standard Precision 95% Confidence Limits of Distribution Sensitivity ΔConcentration per 0.010 ΔAbs 432 DR 5000 600 mg/l COD 576 624 mg/l COD 8 mg/l COD DR 2800 600 mg/l COD 576 624 mg/l COD 8 mg/l COD DR 2700 600 mg/l COD 576 624 mg/l COD 8 mg/l COD DR/2500 500 mg/l COD 479 521 mg/l COD 8 mg/l COD DR/2400 500 mg/l COD 479 521 mg/l COD 8 mg/l COD Page 4 of 6

Estimated detection limit (EDL) The EDL for program 432 is 4mg/L COD. The EDL is the calculated lowest average concentration in a deionized water matrix that is different from zero with a 99% level of confidence. Summary of method Chemical Oxygen Demand (COD) is defined as... a measure of the oxygen equivalent of the organic matter content of a sample that is susceptible to oxidation by a strong chemical oxidant (APHA Standard Methods, 19th ed., 1995). Trivalent manganese is a strong, non-carcinogenic chemical oxidant that changes quantitatively from purple to colorless when it reacts with organic matter. It typically oxidizes about 80% of the organic compounds. Studies have shown that the reactions are highly reproducible and test results correlate closely to Biochemical Oxygen Demand (BOD) values and hexavalent chromium COD tests. None of the oxygen demand tests provide 100% oxidation of all organic compounds. A calibration is provided which is based on the oxidation of Potassium Acid Phthalate (KHP). A different response may be seen in analyzing various wastewaters. The KHP calibration is adequate for most applications. The highest degree of accuracy is obtained when test results are correlated to a standard reference method such as BOD or one of the chromium COD methods. Special waste streams or classes will require a separate calibration to obtain a direct mg/l COD reading or to generate a correction factor for the precalibrated KHP response. The sample digestion time can be extended up to four hours for samples that are difficult to oxidize. Test results are measured at 510 nm. Consumables and replacement items Required reagents Description Quantity/Test Unit Catalog number Manganese III COD Reagent Vials, 20 1000 mg/l COD 1 25/pkg 2623425 Water, deionized varies 4 L 27256 Required apparatus Description Quantity/Test Unit Catalog number Blender, 2-speed, 120 VAC 1 each 2616100 Blender, 2-speed, 240 VAC 1 each 2616102 Blender Container, 50 250 ml 1 2/pkg 2674800 DRB200 Reactor, 110 V, 15 x 16 mm 1 each LTV082.53.40001 DRB200 Reactor, 220 V, 15 x 16 mm 1 each LTV082.52.40001 Pipet, TenSette, 0.1 to 1.0 ml 1 each 1970001 Pipet Tips, for TenSette Pipet 19700-01 2 50/pkg 2185696 Test Tube Rack 1 each 1864100 Pipet Tips, for TenSette Pipet 19700-01 varies 1000/pkg 2185628 Page 5 of 6

Recommended standards and apparatus Description Unit Catalog number COD Standard Solution, 800-mg/L COD 200 ml 2672629 Oxygen Demand Standard for BOD, COD, TOC, 10-mL ampules 16/pkg 2833510 Potassium Acid Phthalate, ACS 500 g 31534 Wastewater Standard, Influent Inorganics, for NH 3 N, NO 3 N, PO 4, COD, SO 4, TOC 500 ml 2833149 Goggles, Safety vented each 2550700 Gloves, Chemical Resistant, 9-9½ 1 1 pair 2410104 1 other sizes are available Optional reagents and apparatus Description Unit Catalog number Titrator Strips, Quantab, for low range chloride 40 tests 2744940 Finger cots 2/pkg 1464702 Flask, volumetric class A 1 L 1457453 ph Paper, 0 14 ph range 100/pkg 2601300 Balance, analytical, 80 g capacity, 115 VAC each 2936701 Weighing Paper, 76x76 mm 500/pkg 1473800 Oven, Laboratory 120 VAC/60Hz 1 1428900 Oven, Laboratory 240 VAC/50Hz 1 1428902 COD Standard Solution, 300-mg/L 200 ml 1218629 COD Standard Solution, 300-mg/L 500 ml 1218649 COD Standard Solution, 1000-mg/L 200 ml 2253929 Standard Methods, most recent edition 2270800 FOR TECHNICAL ASSISTANCE, PRICE INFORMATION AND ORDERING: In the U.S.A. Call toll-free 800-227-4224 Outside the U.S.A. Contact the HACH office or distributor serving you. On the Worldwide Web www.hach.com; E-mail techhelp@hach.com HACH COMPANY WORLD HEADQUARTERS Telephone: (970) 669-3050 FAX: (970) 669-2932 Hach Company, 2007. All rights reserved. Printed in the U.S.A. Updated February 2008, Edition 5