Characterize Fab and Fc Fragments by Cation-Exchange Chromatography

Similar documents
Characterize mab Charge Variants by Cation-Exchange Chromatography

Cell Culture Optimization Using an Agilent Bio-Monolith Protein A Column and LC/MS

Cell Clone Selection Using the Agilent Bio-Monolith Protein A Column and LC/MS

Clone Selection Using the Agilent 1290 Infinity Online 2D-LC/MS Solution

mab Titer Analysis with the Agilent Bio-Monolith Protein A Column

Application of Agilent AdvanceBio Desalting-RP Cartridges for LC/MS Analysis of mabs A One- and Two-dimensional LC/MS Study

Application Note. Author. Abstract. Biotherapeutics and Biologics. Sonja Schneider Agilent Technologies, Inc. Waldbronn, Germany

Precise Characterization of Intact Monoclonal Antibodies by the Agilent 6545XT AdvanceBio LC/Q-TOF

PLRP-S Polymeric Reversed-Phase Column for LC/MS Separation of mabs and ADC

Analysis of Monoclonal Antibody (mab) Using Agilent 1290 Infinity LC System Coupled to Agilent 6530 Accurate-Mass Quadrupole Time-of-Flight (Q-TOF)

Profiling Glycosylation of Monoclonal Antibodies at Three Levels Using the Agilent 6545XT AdvanceBio LC/Q TOF

Separation of Native Monoclonal Antibodies and Identification of Charge Variants:

Characterization of Glycosylation in the Fc Region of Therapeutic Recombinant Monoclonal Antibody

Analysis of Intact and C-terminal Digested IgG1 on an Agilent Bio MAb 5 µm Column

Charge Heterogeneity Analysis of Rituximab Innovator and Biosimilar mabs

High-resolution Analysis of Charge Heterogeneity in Monoclonal Antibodies Using ph-gradient Cation Exchange Chromatography

Agilent Biocolumns. Titer Determination. Application Compendium

Identifying Monoclonal Antibody Mutation Sites Using the Agilent 1290 Infinity II 2D-LC Solution with Q-TOF LC/MS

Reducing Cycle Time for Charge Variant Analysis of Monoclonal Antibodies

Analysis of Monoclonal Antibody Digests with the Agilent 1290 Infinity 2D-LC Solution Part 2: HILIC RPLC-MS

Protein Separation with ph Gradients Using Composite Buffer Systems Calculated by the Agilent Buffer Advisor Software

Size Exclusion Chromatography/ Mass Spectrometry Analysis of Antibody Drug Conjugates Using the Agilent 1260 Infinity II Bio-Inert LC

Benefits of 2D-LC/MS/MS in Pharmaceutical Bioanalytics

Fast and Efficient Peptide Mapping of a Monoclonal Antibody (mab): UHPLC Performance with Superficially Porous Particles

Disulfide Linkage Analysis of IgG1 using an Agilent 1260 Infinity Bio inert LC System with an Agilent ZORBAX RRHD Diphenyl sub 2 µm Column

Analysis of Monoclonal Antibody Digests with the Agilent 1290 Infinity 2D-LC Solution

High-Throughput LC/MS Purification of Pharmaceutical Impurities

Analysis of Monoclonal Antibody N-glycans by Fluorescence Detection and Robust Mass Selective Detection Using the Agilent LC/MSD XT

Seamless Method Transfer from an Agilent 1260 Infinity Bio-inert LC to an Agilent 1260 Infinity II Bio-inert LC

A Comprehensive Approach for Monoclonal Antibody N-linked Glycan Analysis from Sample Preparation to Data Analysis

An Integrated Workflow for Intact and Subunits of Monoclonal Antibody Accurate Mass Measurements

Fraction Analysis of Cysteine Linked Antibody-Drug Conjugates Using Hydrophobic Interaction. chromatography. Agilent 1260 Infinity II Bio-Inert System

Developing Purification Strategies for the Agilent 1260 Infinity II Preparative LC/MSD System

Application Note. Author. Abstract. Biopharmaceuticals. Verified for Agilent 1260 Infinity II LC Bio-inert System. Sonja Schneider

Emulation of the Agilent 1100 Series LC Through Waters Empower Software Analysis of an Analgesic Mixture

Separate and Quantify Rituximab Aggregates and Fragments with High-Resolution SEC

N-Glycan Profiling Analysis of a Monoclonal Antibody Using UHPLC/FLD/Q-TOF

Strategies for the Separation and Characterization of Protein Biopharmaceuticals

Characterizing Antibody-Drug Conjugates Using Micro Pillar Array Columns Combined with Mass Spectrometry (µpac -MS)

Analysis of Monoclonal Antibodies and Their Fragments by Size Exclusion Chromatography Coupled with an Orbitrap Mass Spectrometer

Put the PRO in Protein Characterization

Application Note. Authors. Abstract. Biopharmaceuticals

Application Note. Authors. Abstract. Ravindra Gudihal Agilent Technologies India Pvt. Ltd. Bangalore, India

Analysis of Nucleosides Using an Agilent Infinity II High Speed UHPLC with the 6130 Single Quadrupole Mass Selective Detector

Reproducible LC/MS Peptide Separation Using Agilent AdvanceBio Peptide Plus Columns

Application Note. Biopharma. Authors. Abstract. James Martosella, Phu Duong Agilent Technologies, Inc Centreville Rd Wilmington, DE 19808

Method Transfer from an Agilent 1200 Series LC to an Agilent 1260 Infinity II LC

Heart-cut 2D-LC/MS approach for pharmaceutical impurity identification using an Agilent 6540 Q-TOF LC/MS System

Rapid UHPLC Analysis of Reduced Monoclonal Antibodies using an Agilent ZORBAX Rapid Resolution High Definition (RRHD) 300SB-C8 Column

Agilent 1290 Infinity Binary LC System with ISET Emulation of a Waters Alliance 2695 LC Applying Concave, Convex, and Linear Gradients

Monoclonal Antibody Analysis on a Reversed-Phase C4 Polymer Monolith Column

Antibody Analysis by ESI-TOF LC/MS

Robustness of the Agilent Ultivo Triple Quadrupole LC/MS for Routine Analysis in Food Safety

Peptide Mapping of Glycoprotein Erythropoietin by HILIC LC/MS and RP-LC/MS

Determination of Asarinin in Xixin (Asari Radix Et Rhizoma)

Application Note. Abstract. Author. Biopharmaceutical

Confident Pesticides Analysis with the Agilent LC/Triple Quadrupole and TOF/QTOF Solutions

Monoclonal Antibody Charge Variants Identification by Fully Automated Capillary Isoelectric Focusing Mass Spectrometry

Fast and High-Resolution Reversed-Phase Separation of Synthetic Oligonucleotides

EPA Method 543: Selected Organic Contaminants by Online SPE LC/MS/MS Using the Agilent Flexible Cube

Critical Quality Attribute Assessment by Peptide Mapping Using LC/MS with Superficially Porous Columns

Reversed-phase Separation of Intact Monoclonal Antibodies Using Agilent ZORBAX Rapid Resolution High Definition 300SB-C8 1.

mab and ADC Analysis Shanhua Lin, Ph.D. The world leader in serving science

Mass Spectrometry Analysis of Liquid Chromatography Fractions using Ettan LC MS System

Faster Separations Using Agilent Weak Cation Exchange Columns

Peptide Mapping: A Quality by Design (QbD) Approach

Ensure your Success with Agilent s Biopharma Workflows

LC-MS Analysis of Digested mabs Using a Reversed-Phase Desalting Cartridge and High-Resolution, Accurate Mass Spectrometry (HRAMS)

Thermo Scientific. MAbPac HIC-10. Product Manual. P/N: June Part of Thermo Fisher Scientific

mabs and ADCs analysis by RP

Agilent 1290 Infinity II 2D-LC Solution Biopharmaceutical Polymer Analysis. WCBP Jan 2017 Washington, DC

LC/MS Based Quantitation of Intact Proteins for Bioanalytical Applications

CQA Assessment by Peptide Mapping Using LC/MS with an AdvanceBio Peptide Plus Column

Application Note. Authors. Abstract. Introduction. Pharmaceutical

Columns for Biomolecules BioLC Column Lines

Monitoring of Mammalian Cell Culture Media with HILIC LC/MS

Scale-up with the Agilent SD-1 Purification System analytical and preparative runs on a single system

AdvanceBio HIC: a Hydrophobic HPLC Column for Monoclonal Antibody (mab) Variant Analysis

Authors. Introduction. Experimental. Environmental Analysis

Minimizing Sample Carryover Using the Multiwash Function of the Agilent 1290 Infinity II Multisampler

ENVIRONMENTAL ANALYSIS. Solutions for Your Analytical Business Markets and Applications Programs. Solution Note

Fast and High Resolution Analysis of Intact and Reduced Therapeutic Monoclonal Antibodies (mabs)

Application Note. Abstract. Authors. Introduction. Environmental

Molecular Characterization of Biotherapeutics The Agilent 1260 Infi nity Multi-Detector Bio-SEC Solution with Advanced Light Scattering Detection

High-throughput and Sensitive Size Exclusion Chromatography (SEC) of Biologics Using Agilent AdvanceBio SEC Columns

High-Resolution Charge Variant Analysis for Top-Selling Monoclonal Antibody Therapeutics Using a Linear ph Gradient Separation Platform

Quantification of Host Cell Protein Impurities Using the Agilent 1290 Infinity II LC Coupled with the 6495B Triple Quadrupole LC/MS System

Chromatographic Workflows for Biopharmaceutical Characterization

Isotopic Resolution of Chromatographically Separated IdeS Subunits Using the X500B QTOF System

Application Note. Author. Introduction. Ravindra Gudihal Agilent Technologies India Pvt Ltd

MassPREP On-Line Desalting Cartridge

Performance characteristics of the 1260 Infinity Quaternary LC system

Quantification of genotoxic "Impurity D" in Atenolol by LC/ESI/MS/MS with Agilent 1200 Series RRLC and 6410B Triple Quadrupole LC/MS

Characterization of mab aggregation using a Cary 60 UV-Vis Spectrophotometer and the Agilent 1260 Infinity LC system

Technical Overview. Author. Abstract. Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany

Optimizing Protein Separations with Agilent Weak Cation-Exchange Columns

Agilent 1290 Infinity Quaternary LC Support of Columns with 2.1 to 4.6 mm ID to 1200 bar

Development of Analysis Methods for Therapeutic Monoclonal Antibodies Using Innovative Superficially Porous Particle Biocolumns

Fast Online Desalting of mabs Using a Reversed-Phase Desalting Cartridge for LC-MS Analysis

UHPLC for Large Bio-Therapeutics

Transcription:

Characterize Fab and Fc Fragments by Cation-Exchange Chromatography Application Note Biologics and Biosimilars Authors Isabel Vandenheede, Emmie Dumont, Pat Sandra, and Koen Sandra Research Institute for Chromatography (RIC) President Kennedypark 6 B-85 Kortrijk Belgium Maureen Joseph Agilent Technologies, Inc. Abstract This application note describes the use of the Agilent Bio MAb weak cation-exchange column in measuring the mab fragments obtained by papain digestion and the subsequent mass spectrometric characterization of the generated Fab and Fc variants following peak collection. Introduction Monoclonal antibodies (mabs) are a rapidly growing class of therapeutics used in the treatment of cancer and autoimmune diseases. Cation-exchange chromatography (CEX) is a powerful technology for the characterization of these molecules. Another application note described the CEX separation of the charge variants of the intact monoclonal antibody trastuzumab []. Trastuzumab, marketed as Herceptin since 998, is in widespread use in the treatment of HER positive breast cancer. The work presented here describes the use of the Agilent Bio MAb weak cationexchange (WCX) column to separate charge variants of papain digested trastuzumab and the subsequent characterization of the Fab and Fc variants by mass spectrometry (MS).

Experimental Materials Acetonitrile and water were obtained from Biosolve B. V. (Valkenswaard, The Netherlands). NaCl, -(N-morpholino)ethanesulfonic acid (MES), papain, and tris(-carboxyethyl)phosphine (TCEP) were purchased from Sigma-Aldrich, Corp. (St. Louis, MO, USA). Humanized monoclonal antibody trastuzumab, marketed as Herceptin, was obtained from Roche (Basel, Switzerland). Sample preparation Herceptin was incubated at 7 C for hours after adding papain. Following incubation, the papain digest buffer was exchanged to CEX mobile phase A and subsequently diluted to mg/ml. The ph-stressed mab was obtained by incubating Herceptin at 7 C for days in M tris ph 9.. Collected CEX fractions were optionally reduced at room temperature using TCEP to a final concentration of 5 mm. Instrumentation Cation-exchange measurements were performed on: Agilent Series Degasser (GA) Agilent Series Binary Pump (GA) Agilent Series Autosampler (G67A) Agilent Series Thermostat for autosampler (GB) Agilent Series Thermostatted Column Compartment (G6A) Agilent Series Diode Array Detector (G5A) Agilent Series Analytical Fraction Collector (G6C) LC/MS measurements were performed on an Agilent 65 Ultra High Definition (UHD) Accurate-Mass Q-TOF with Agilent Jet Stream LC/MS (G65A) and an Agilent 9 Infinity LC System equipped with: Agilent 9 Infinity Binary pump (GB) Agilent 9 Infinity Autosampler (G6A) Agilent 9 Infinity Thermostat for autosampler (GB) Agilent 9 Infinity Thermostatted Column Compartment (G6C) Agilent 9 Infinity Diode Array Detector (GA) Software Agilent OpenLAB CDS ChemStation revision C..5 (5) Agilent MassHunter for instrument control (B5.) Agilent MassHunter for data analysis (B6.) Agilent BioConfirm software for MassHunter (B6.) Instrumental conditions Cation-exchange conditions Column: Agilent Bio MAb NP5,. 5 mm, 5 µm, PEEK (p/n 59-) Mobile phase: A) mm MES ph 5.6 B) mm MES ph 5.6 + mm NaCl Flow rate: 7 µl/min Injection volume: 6 µl ( µg on-column) Column temperature: C Gradient: Time (min) % B 9.5 8 5 5.5 8 Detection: UV at nm/8 nm Fraction collection: Time-based Online desalting LC/MS conditions Cartridge: On-line desalting cartridge,. mm Mobile phase: A).% formic acid in water (v/v) B).% formic acid in acetonitrile (v/v) Flow rate: µl/min Injection volume: µl Needle wash solvent: 6% Acetonitrile, 5% water, 5% isopropanol Autosampler temperature: 7 C Gradient: Time (min) % B 5.5 5 8. 8.. 5 5 5 Q-TOF source: Agilent Jet Stream positive ionization mode Drying gas temperature: C Drying gas flow rate: 8 L/min Nebulizer pressure: 5 psig Sheath gas temperature: 5 C Sheath gas flow rate: L/min Nozzle voltage:, V Capillary voltage:,5 V Fragmentor voltage: V Q-TOF detection: Mass range, amu Data acquisition range: 5 to, m/z High resolution mode ( GHz) Data acquisition rate: spectrum per s Profile acquisition Diverter valve: Time (min) Flow to waste Flow to MS.5 Flow to waste

Results and Discussion The structures of an antibody and its Fab and Fc fragments generated following papain digestion are displayed in Figure. Figure shows the UV nm CEX profiles of the papain digest of nonstressed and ph-stressed Herceptin obtained on the Agilent Bio MAb weak cation-exchange column. Due to the lower isoelectric point of the Fab and Fc fragments compared to the intact antibody, the ph of the mobile phase needs to be lowered compared to the mobile phase buffer at ph 7.65 used for intact mab CEX analysis. An MES buffer at ph 5.6 was used to obtain optimal separation of the Fab and Fc. The CEX peaks indicated in Figure were collected and further characterized by MS following online desalting. Lc Hc Fab Papain Fc Figure. Structure of a monoclonal antibody and its Fab and Fc fragments. A mab is composed of two light (Lc) and two heavy chains (Hc) connected through disulfide bonds. Papain cleaves the mab at the hinge region with the generation of the Fab and Fc fragment. The Fc region is N-glycosylated and contains complex bi-antennary glycans. mau,, 8 6 mau,, 8 6 Papain digest of nonstressed mab 8 6 8 Papain digest of ph-stressed mab 5 8 6 8 Figure. Cation-exchange separation of charge variants of papain-digested ph-stressed and nonstressed Herceptin. The annotated peaks were collected and further characterized using mass spectrometry following online desalting. 6 7 8 min min

The deconvoluted spectra of the four collected CEX peaks of the nonstressed sample are shown in Figure. The spectra indicate that the Fab and Fc fragments correspond to, respectively, CEX peaks and. The typical Fc glycosylation pattern is apparent in CEX peak. The Dalton mass difference between CEX peaks and is indicative of a deamidation of the. Indeed, a deamidation involving the conversion of asparagine to aspartic acid renders the more acidic, explaining its earlier elution. CEX peak is identified as the with loss of the basic tripeptide KTH at the papain cleavage site. This represents an artifact of the sample preparation. Counts Counts Counts Counts 6 8 CEX peak 7,7.5 7 6 5 6 CEX peak - KTH + deamidation of LC 7 CEX peak 7,67.7.8.6.. 6, 6,6 6,8 7, 7, 7, 7,6 7,8 8, 8, 8, 8,6 6 GF/GF.6 CEX peak 5,98. GF/GF GF/GF. 5,89.8 GF/GF 5,..8 G/GF GF/GF. 5,67.7 5,6. Fc fragment 5,8 5, 5, 5, 5,6 5,8 5, 5, 5, 5,6 5,8 5, Figure. Deconvoluted spectra of the four collected cation-exchange peaks of nonstressed papain digested Herceptin. The abbreviations G, GF, GF, and GF refer to the N-glycans attached to the Fc fragment.

The deconvoluted spectra of the four collected CEX peaks of the ph-stressed sample are shown in Figure. The most abundant CEX peak 6 is identified as the deamidated Fab fragment. The spectra reveal an extra mass difference of Dalton between CEX peaks 5 and 6. Additional MS analysis following TCEP reduction of peaks 5 and 6 (Figure 5) demonstrated a deamidation of the heavy chain portion of the on top of the deamidation of the light chain for peak 5. Conclusions The Agilent Bio MAb weak cation-exchange column successfully separated charge variants of a papain-digested monoclonal antibody. Following peak collection, charge variants could be further characterized by mass spectrometry in combination with online desalting. 6 CEX peak 5 7.8 CEX peak 6.6.. 6 CEX peak 7 6 Counts Counts Counts Counts 769.8 768.9 767.7 + deamidation of LC and HC + deamidation of LC 6, 6,6 6,8 7, 7, 7, 7,6 7,8 8, 8, 8, 8,6 8,8 9, 9, 6 GF/GF.5 GF/GF 5,98. GF/GF CEX peak 8 GF/GF 5,8. 5,.5.5 G/GF GF/GF 5,67. 5,6.5 Fc fragment 5,8 5, 5,6 5, 5, 5,8 5, 5,6 Figure. Deconvoluted spectra of the four collected cation-exchange peaks of ph-stressed papain digested Herceptin. Counts 5 LC of,. Deamidated HC of,.7 TCEP reduced CEX peak 5 6 Deamidated LC of HC of.6,.5,.8....8.6.. TCEP reduced CEX peak 6,,,,6,8,,,,6,8 Figure 5. Deconvoluted spectra of TCEP reduced cation-exchange peaks 5 and 6. 5

Reference. I. Vandenheede, E. Dumont, P. Sandra, K. Sandra, M. Joseph. Characterize mab Charge Variants by Cation-Exchange Chromatography. Agilent Technologies, Inc. Application Note, publication number 599-57EN,. For More Information These data represent typical results. For more information on our products and services, visit our Web site at www.agilent.com/chem. www.agilent.com/chem Agilent shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material. Information, descriptions, and specifications in this publication are subject to change without notice. Agilent Technologies, Inc., Printed in the USA November, 599-57EN