SEC-MS with Volatile Buffers for Characterization of Biopharmaceuticals

Similar documents
Development of Ultra-fast ph-gradient Ion Exchange Chromatography for the Separation of Monoclonal Antibody Charge Variants

Fast and High Resolution Analysis of Intact and Reduced Therapeutic Monoclonal Antibodies (mabs)

Evaluation of Acclaim HPLC Columns using the National Institute of Standards and Technology Standard Reference Material 870

Fast Online Desalting of mabs Using a Reversed-Phase Desalting Cartridge for LC-MS Analysis

Analysis of Monoclonal Antibodies and Their Fragments by Size Exclusion Chromatography Coupled with an Orbitrap Mass Spectrometer

Improving Retention Time Precision and Chromatography of Early Eluting Peptides with Acetonitrile/Water Blends as Solvent B

LC-MS Analysis of Digested mabs Using a Reversed-Phase Desalting Cartridge and High-Resolution, Accurate Mass Spectrometry (HRAMS)

ph gradient analysis of IgG1 therapeutic monoclonal antibodies using a 5 µm WCX column

Lifetime stability of size exclusion chromatography columns for protein aggregate analysis

High-Throughput Peptide Mapping with the Vanquish UHPLC System and the Q Exactive HF Mass Spectrometer

A Fast and Robust Linear ph Gradient Separation Platform for Monoclonal Antibody (MAb) Charge Variant Analysis

A universal chromatography method for aggregate analysis of monoclonal antibodies

A Complete Workflow Solution for Intact Monoclonal Antibody Characterization Using a New High-Performance Benchtop Quadrupole- Orbitrap LC-MS/MS

MAbPac RP Column. High-performance reverse phase chromatography column for monoclonal antibody analysis

Application of Agilent AdvanceBio Desalting-RP Cartridges for LC/MS Analysis of mabs A One- and Two-dimensional LC/MS Study

Analysis of Monoclonal Antibody and Related Substances using a New Hydrophobic Interaction Chromatography (HIC) Column

Thermo Scientific EASY-nLC 1200 System. Leading in simplicity. and performance

PLRP-S Polymeric Reversed-Phase Column for LC/MS Separation of mabs and ADC

Chromatography column for therapeutic protein analysis

Developing Quantitative UPLC Assays with UV

HIC as a Complementary, Confirmatory Tool to SEC for the Analysis of mab Aggregates

Thermo Scientific Peptide Mapping Workflows. Upgrade Your Maps. Fast, confident and more reliable peptide mapping.

Thermo Scientific MAbPac HIC Columns. Novel Hydrophobic Interaction HPLC Columns. Designed for Monoclonal Antibody Analysis

Analysis of Monoclonal Antibodies and Their Fragments by Size-Exclusion Chromatography Coupled with an Orbitrap Mass Spectrometer

Rapid UHPLC Analysis of Reduced Monoclonal Antibodies using an Agilent ZORBAX Rapid Resolution High Definition (RRHD) 300SB-C8 Column

High-Resolution Charge Variant Analysis for Top-Selling Monoclonal Antibody Therapeutics Using a Linear ph Gradient Separation Platform

Thermo Scientific Solutions for Intact-Protein Analysis. Better, Faster Decisions for. Biotherapeutic Development

Rapid Peptide Mapping via Automated Integration of On-line Digestion, Separation and Mass Spectrometry for the Analysis of Therapeutic Proteins

Nano LC at 20 nl/min Made Easy: A Splitless Pump Combined with Fingertight UHPLC Nano Column to Boost LC-MS Sensitivity in Proteomics

Jonathan R. Beck and Charles T. Yang; Thermo Fisher Scientific, San Jose, CA

Application Note. Author. Abstract. Biotherapeutics and Biologics. Sonja Schneider Agilent Technologies, Inc. Waldbronn, Germany

Method Optimisation in Bottom-Up Analysis of Proteins

Disulfide Linkage Analysis of IgG1 using an Agilent 1260 Infinity Bio inert LC System with an Agilent ZORBAX RRHD Diphenyl sub 2 µm Column

High-speed, High-resolution Oligonucleotide Separations Using Small Particle Anion-Exchangers

NISTmAb characterization with a high-performance RP chromatography column

A Novel ph Gradient Separation Platform for Monoclonal Antibody (MAb) Charge-Variant Analysis

Clone Selection Using the Agilent 1290 Infinity Online 2D-LC/MS Solution

Concentration of Human Hormones in Drinking Water Using Solid Phase Extraction and Analysis by High Performance Liquid Chromatography

Mass Spectrometry Analysis of Liquid Chromatography Fractions using Ettan LC MS System

Simultaneous Quantitation of a Monoclonal Antibody and Two Proteins in Human Plasma by High Resolution and Accurate Mass Measurements

Application Note. Authors. Abstract. Biopharmaceuticals

Analysis of Monoclonal Antibody (mab) Using Agilent 1290 Infinity LC System Coupled to Agilent 6530 Accurate-Mass Quadrupole Time-of-Flight (Q-TOF)

A highly sensitive and robust 150 µm column to enable high-throughput proteomics

Characterization of intact monoclonal antibody with microfluidic chip electrophoresis mass spectrometry

Analysis of Glyphosate and AMPA in Environmental Water Samples by Ion Chromatography and On-Line Mass Spectrometry with Minimal Sample Preparation

Precise Characterization of Intact Monoclonal Antibodies by the Agilent 6545XT AdvanceBio LC/Q-TOF

columns PepSwift and ProSwift Capillary Monolithic Reversed-Phase Columns

Technical Overview. Authors. Abstract. Verified for Agilent 1260 Infinity II LC Bio-inert System

Thermo Scientific EASY-nLC 1000 System. Effortless, split-free nanoflow UHPLC Top performance in LC/MS

High-resolution Analysis of Charge Heterogeneity in Monoclonal Antibodies Using ph-gradient Cation Exchange Chromatography

for water and beverage analysis

Charge Heterogeneity Analysis of Rituximab Innovator and Biosimilar mabs

Fast and High-Resolution Reversed-Phase Separation of Synthetic Oligonucleotides

Seamless Method Transfer from an Agilent 1260 Infinity Bio-inert LC to an Agilent 1260 Infinity II Bio-inert LC

Analysis of Illegal Dyes in Food Matrices Using Automated Online Sample Preparation with Liquid Chromatography-Mass Spectrometry

Chromatographic Workflows for Biopharmaceutical Characterization

Characterize Fab and Fc Fragments by Cation-Exchange Chromatography

Fast mass transfer Fast separations High throughput and improved productivity Long column lifetime Outstanding reproducibility Low carryover

User Guide. SMART Digest Kit

Thermo Scientific Dionex ICS-900 Ion Chromatography System

Characterize mab Charge Variants by Cation-Exchange Chromatography

Antibody subunit analysis workflow on a quadrupole-orbitrap mass spectrometer: from optimized sample preparation to data analysis

Separate and Quantify Rituximab Aggregates and Fragments with High-Resolution SEC

Monoclonal Antibody Analysis on a Reversed-Phase C4 Polymer Monolith Column

Simple charge variant profile comparison of an innovator monoclonal antibody and a biosimilar candidate

Quantitative LC-MS Analysis of 14 Benzodiazepines in Urine Using TraceFinder 1.1 Software and High Resolution Accurate Mass

Key Words Pesticides analysis, food safety, TSQ Endura, TraceFinder, MRL, tsrm, residue analysis

Implementation of a Walk-Up High- Pressure Capillary Ion Chromatograph for the Fast Separation of Pharmaceutical Relevant Inorganic Anions and Cations

Simple, Robust, High Quality Intact Mass Analysis A Biosimilars Case Study

BIOANALYTICAL STRATEGY FOR IN VITRO METABOLITE SCREENING WITH EXACT MASS USING THE Q-Tof micro. Jose M. Castro-Perez 1, Carina Leandersson 2

A Comprehensive Workflow to Optimize and Execute Protein Aggregate Studies

MassPREP On-Line Desalting Cartridge

Separation of Native Monoclonal Antibodies and Identification of Charge Variants:

Method Development Considerations for Reversed-Phase Protein Separations

ACQUITY UPLC H-Class Bio System

Thermo Scientific TSQ Endura Triple-Stage Quadrupole Mass Spectrometer. Extreme quantitative value. with unprecedented ease

Characterization of mab aggregation using a Cary 60 UV-Vis Spectrophotometer and the Agilent 1260 Infinity LC system

Using HRAM Survey Analysis Combined with Rapid MS2 Data to Develop a Fragmentation Based Detection Workflow for Structure ID Acquisition

MAbPac SCX-10 Column for Monoclonal Antibody Variant Analysis and Characterization

Thermo Scientific Q Exactive HF Orbitrap LC-MS/MS System. Higher-Quality Data, Faster Than Ever. Speed Productivity Confidence

Thermo Scientific TSQ Endura Triple-Stage Quadrupole Mass Spectrometer. Extreme quantitative value. with unprecedented ease

Isotopic Resolution of Chromatographically Separated IdeS Subunits Using the X500B QTOF System

Profiling Glycosylation of Monoclonal Antibodies at Three Levels Using the Agilent 6545XT AdvanceBio LC/Q TOF

The Benefits of SOLAμ Technology in Sample Preparation. Jon Bardsley and Ken Meadows Thermo Fisher Scientific, Runcorn, UK

Fraction Analysis of Cysteine Linked Antibody-Drug Conjugates Using Hydrophobic Interaction. chromatography. Agilent 1260 Infinity II Bio-Inert System

Application Note # ET-20 BioPharma Compass: A fully Automated Solution for Characterization and QC of Intact and Digested Proteins

Thermo Scientific EASY-Spray Technology. Plug-and-spray with. State-of-the-art performance

High-Resolution Separation of Intact Monoclonal Antibody Isoforms

Use of alternative chromatographic phases and LC-MS for characterization of N-glycans from NISTmAb RM 8671

SMART Digest compared to classic in-solution digestion of rituximab for in-depth peptide mapping characterization

Fast and Efficient Peptide Mapping of a Monoclonal Antibody (mab): UHPLC Performance with Superficially Porous Particles

Size Exclusion Chromatography of Biosimilar and Innovator Insulin Using the Agilent AdvanceBio SEC column

A comprehensive CE/ESI-MS solution for BioPharma applications

Introduction. Yan Liu, Victor Barreto, Rosanne Slingsby, Chris Pohl Thermo Fisher Scientific, Sunnyvale, CA, USA

ADVANCE ACCURACY AND PRODUCTIVITY FOR FASTER ANALYSIS

IdeS-cleaved mab subunit analysis with LC-HRAM-MS: a quick and accurate comparison of biosimilar and originator biotherapeutics

A Novel Screening Method for Anthropogenic Sewage Pollutants in Waste Water, Ground Water and Drinking Water Samples by LC HRAM Analysis

Reducing Cycle Time for Charge Variant Analysis of Monoclonal Antibodies

Transcription:

SE-MS with Volatile uffers for haracterization of iopharmaceuticals Martin Samonig, Remco Swart Thermo Fisher Scientific, Germering, Germany pplication Note 1133 Key Words Monoclonal ntibodies, MbPac SE-1, iocompatible UHPL, iotherapeutics haracterization, iopharma Goal Prove the suitability of the biocompatible Thermo Scientific Vanquish Flex system for size exclusion chromatography separations with mass spectrometry compatible volatile buffers. Introduction Size exclusion chromatography (SE) is a routine method for the characterization of biopharmaceuticals. The conventional SE setup uses UV detection, due to high salt concentrations in commonly used buffers. In this study, the separation of proteins by denaturing SE with direct coupling to a mass spectrometer (MS) was investigated. mobile phase containing acetonitrile, trifluoroacetic acid, and formic acid was used, which also allowed a direct molecular weight measurement by mass spectrometry. The denaturing SE method with a separation based on size differences provides an alternative to the commonly used (ion pair) reversed-phase chromatography (IP-RP). Especially for large proteins, IP-RP peaks often have significant tailing, due to secondary interactions with the column. This effect can be reduced with elevated column temperatures (>70 ), but can be problematic for temperature-sensitive proteins. SE can be used at room temperature with satisfying peak shape also for large proteins. Trifluoroacetic acid (TF) is a common ion-pairing reagent that is often used to increase the separation efficiency for proteins in reversed-phase separations. 1 In SE buffers, the addition of TF reduces the protein-protein and protein-column interactions and is used instead of salts to avoid these negative effects. 2 However, it has been known that TF suppresses the mass spectrometry signal; therefore, a very low TF concentration of 0.05% was used. Experimental System evaluation was performed using the HPL protein standard mixture (Sigma-ldrich hemie GmbH). The commercially available monoclonal antibodies rituximab (F. Hoffmann-La Roche, Ltd) and denosumab (mgen) were used for this study, and disulfide bonds were reduced by incubation for 30 minutes at 60 with 5 mm Tris (2-carboxyethyl) phosphine hydrochloride (TEP) to separate light and heavy chain.

2 Equipment Vanquish Flex UHPL system consisting of: System ase (P/N VF-S01-) Quaternary Pump F (P/N VF-P20-) Split Sampler FT (P/N VF-10-) olumn ompartment H (P/N VH-10-) iode rray etector HL (P/N VH-10-) LightPipe Flow ell, Standard, 10 mm (P/N 6083.0100) Vanquish MS onnection Kit (P/N 6720.0405) Thermo Scientific Q Exactive HF Hybrid Quadrupole-Orbitrap Mass Spectrometer Experimental onditions - HPL olumn Thermo Scientific MbPac SE-1, 5 µm, 300 Å (4 x 300 mm (P/N 074696) Mobile Phase : 0.1% F and 0.05% TF in 3/7 acetonitrile/water (v/v), (P/N acetonitrile TS-51101) Gradient Isocratic Flow Rate 0.2 ml/min Temperature 25 Injection Volume 1 µl etection 214 nm ata ollection Rate: 10 Hz Response Time 0.4 s Flow ell 10 mm LightPipe Experimental onditions - MS Source HESI-II Sheath Gas Pressure 35 psi uxiliary Gas Flow 10 arbitrary units apillary Temperature 250 S-lens RF Voltage 60 Source Voltage 3.5 kv Full MS Parameters Full MS Mass Range 1500 5000 / 400 3000 Resolution Settings 15.000 / 240.000 Target Value 3e6 Max Injection Time 200 ms Microscans 10 / 1 SI 10 100 ev ata Processing Thermo Scientific ionex hromeleon hromatography ata System (S) software version 7.2 SR3 was used for data acquisition, and the data analysis was performed using Thermo Scientific Proteineconvolution software version 3.0. Results and iscussion The mixture of four standard proteins (cytochrome c, myoglobin, ribonuclease, and transferrin) was used to evaluate the setup. The separation and corresponding mass spectrum for each protein is shown in Figure 1. The measured mass after deconvolution has a mass deviation to the calculated theoretical mass of 0.75 1.23 ppm and shows the outstanding mass accuracy of the Q Exactive mass spectrometer. The concentration of the additive TF is a critical parameter in this method and has to be optimized for protein separations; it varies between 0.01 and 0.2%. The quaternary pump of the Vanquish Flex system enables a straightforward mobile phase optimization using one channel containing an adequate concentration of TF. ll common therapeutic mb products are dissolved in formulation buffers with MS-incompatible components like polysorbate 80 and salts to stabilize the protein. The established SE-MS setup is able to separate the small buffer components from the protein, can be used as a desalting step prior to mass detection and provides an alternative to IP-RP separations. (Figure 2 and 2). The isocratic separation allows fast separations, due to the absence of an equilibration step and the low carryover of SE separations. This method can also be used to separate the intact mb from the detached chains for product related impurity characterizations, due to a retention time difference of 0.7 minutes from the intact mb to the next eluting peak of the heavy chain. The disulfide bridges of the mb were reduced with TEP to separate light (23 ka) and heavy chain (50 ka). The two chains were baseline separated and the measured protein spectra are shown in Figure 2 and 2.

12.22 3 Protein omplexes >100 ka 9.10 11.32 12.90 2 4 6 8 10 12 14 16 18 20 22 24 Transferrin 78021.648 a 998.1834 z=17 1304.9305 z=13 1884.4521 z=9 Myoglobin 16940.9851 a 1.23 ppm 2422.5803 2826.1767 z=6 Ribonuclease 13673.2702 a 0.75 ppm 1244.8513 z=11 1521.2604 z=9 1955.6193 2281.2214 z=6 2737.4656 z=5 941.7172 z=13 1224.0306 z=10 1748.0402 ytochrome 12222.2113 a 0.92 ppm 2446.8534 z=5 3000 Figure 1. Separation of a mixture of the four standard proteins and the measured mass spectra for cytochrome c (1), myoglobin (1), ribonuclease, (1) and transferrin (1). The protein complexes at RT = 8 min can be minimized with increased TF concentrations (~ 0.1%), with the drawback of decreased signal intensities of the protein spectra.

4 8.58 15.71 Unreduced Sample 9.27 11.10 Reduced Sample 5 10 15 20 25 5 10 15 20 25 Intact Monoclonal ntibody (mb), 8.58 min Heavy hain (Hc), 9.27 min 2000 3000 4000 5000 1000 1500 2000 Formulation uffer Light hain omponents, 15.71 min (Hc), 11.10 min 2000 2500 3000 3500 1000 1500 2000 Figure 2. Separation of the unreduced and reduced monoclonal antibody rituximab.

onclusion The use of TF and F as additives for volatile SE buffers enables the possibility for MS detection and delivers sufficient separation efficacy to characterize biopharmaceuticals. The biocompatible flow path of the Vanquish Flex system and the solvent flexibility of the quaternary pump delivers the ideal system for SE-MS analysis and even establishes the possibility of method scouting. References 1. Thermo Scientific pplication Note 591: L/MS nalysis of the Monoclonal ntibody Rituximab Using the Q Exactive enchtop Orbitrap Mass Spectrometer. 2013. 2. rakawa, T., et al., The critical role of mobile phase composition in size exclusion chromatography of protein pharmaceuticals. J Pharm Sci, 2010, 99(4), 1674-92. pplication Note 1133 www.thermofisher.com/chromatography 2016 Thermo Fisher Scientific Inc. ll rights reserved. ll trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is presented as an example of the capabilities of Thermo Fisher Scientific products. It is not intended to encourage use of these products in any manners that might infringe the intellectual property rights of others. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details. frica +43 1 333 50 34 0 ustralia +61 3 9757 4300 ustria +43 810 282 206 elgium +32 53 73 42 41 razil +55 11 3731 5140 anada +1 800 530 8447 hina 800 810 5118 (free call domestic) 400 650 5118 N71707-EN 0616S enmark +45 70 23 62 60 Europe-Other +43 1 333 50 34 0 Finland +358 10 3292 200 France +33 1 60 92 48 00 Germany +49 6103 408 1014 India +91 22 6742 9494 Italy +39 02 950 591 Japan +81 6 6885 1213 Korea +82 2 3420 8600 Latin merica +1 561 688 8700 Middle East +43 1 333 50 34 0 Netherlands +31 76 579 55 55 New Zealand +64 9 980 6700 Norway +46 8 556 468 00 Russia/IS +43 1 333 50 34 0 Singapore +65 6289 1190 Sweden +46 8 556 468 00 Switzerland +41 61 716 77 00 Taiwan +886 2 8751 6655 UK/Ireland +44 1442 233555 US +1 800 532 4752