Size-Exclusion Chromatography (SEC) Optimization Guide

Similar documents
ACQUITY UPLC Protein BEH SEC Columns and Standards

Analysis of biomolecules by SEC and Ion-Exchange UPLC

Agilent AdvanceBio SEC Columns for Aggregate Analysis: Instrument Compatibility

2012 Waters Corporation 1

R R Innovation Way P/N SECKIT-7830 Newark, DE 19711, USA Tel: Fax: Website: Published in November 2013

Developing Quantitative UPLC Assays with UV

Application Note. Authors. Abstract. Biopharmaceuticals

MassPREP On-Line Desalting Cartridge

Protein-Pak Hi Res HIC Column and HIC Protein Standard

Lifetime stability of size exclusion chromatography columns for protein aggregate analysis

H-Class & H-Class Bio AutoBlend and AutoBlend plus. Jean-Michel Plankeele May Waters Corporation 1

Developing Robust and Efficient IEX Methods for Charge Variant Analysis of Biotherapeutics Using ACQUITY UPLC H-Class System and Auto Blend Plus

A Comprehensive Workflow to Optimize and Execute Protein Aggregate Studies

Separate and Quantify Rituximab Aggregates and Fragments with High-Resolution SEC

Mobile Phase Optimization in SEC Method Development

TSK-GEL BioAssist Series Ion Exchange Columns

ACQUITY Arc TM System

Performance characteristics of the 1260 Infinity Quaternary LC system

EASY-Spray Columns. Guidance for column set up and installation Tips to maximize column lifetime

Understanding the Effects of Common Mobile Phase Additives on the Performance of Size Exclusion Chromatography

'Tips and Tricks' for Biopharmaceutical Characterization using SEC

Simplifying Methods Transfer: Novel Tools for Replicating Your Established Methods on an ACQUITY Arc System

A universal chromatography method for aggregate analysis of monoclonal antibodies

Size Exclusion BioHPLC columns Ion Exchange BioHPLC columns

NISTmAb characterization with a high-performance RP chromatography column

ACQUITY UPLC. I-Class PLUS WHAT SEPARATES YOU FROM EVERYONE ELSE

ION EXCHANGE KIT FOR MAB SEPARATIONS

Agilent MetaCarb 87H Organic Acids Column H + Form User Manual Part No. A5214

Multiple Detector Approaches to Protein Aggregation by SEC

Application Note. Author. Abstract. Biopharmaceuticals. Verified for Agilent 1260 Infinity II LC Bio-inert System. Sonja Schneider

CHARACTERIZATION & PURIFICATION

High-resolution Analysis of Charge Heterogeneity in Monoclonal Antibodies Using ph-gradient Cation Exchange Chromatography

Size Exclusion BioHPLC Columns

Method Development Considerations for Reversed-Phase Protein Separations

High-throughput and Sensitive Size Exclusion Chromatography (SEC) of Biologics Using Agilent AdvanceBio SEC Columns

Performance Characteristics of the Agilent 1220 Infinity Gradient LC system

High Throughput Sub-4 Minute Separation of Antibodies using Size Exclusion Chromatography

ACQUITY Arc VERSATILITY WITHOUT COMPROMISE

BioHPLC columns. Tim Rice Biocolumn Technical Specialist

Solutions Guide. MX Series II Modular Automation for Nano and Analytical Scale HPLC And Low Pressure Fluid Switching Applications

The Agilent 1260 Infinity BioInert Quaternary Pump. Scope of a low-pressure mixing UHPLC pump with Bio-Inert Capabilities

Reversed-phase Separation of Intact Monoclonal Antibodies Using Agilent ZORBAX Rapid Resolution High Definition 300SB-C8 1.

Columns for Biomolecules BioLC Column Lines

CX-1 ph Gradient Buffer

AdvanceBio HIC: a Hydrophobic HPLC Column for Monoclonal Antibody (mab) Variant Analysis

Biotherapeutic Method Development Guide

Automatic Combination of Orthogonal Separation Modes for Multiple Components in Mixtures

Hichrom Limited 1 The Markham Centre, Station Road, Theale, Reading, Berkshire, RG7 4PE, UK Tel: +44 (0) Fax: +44 (0)

Qualification of UPLC system IRequest UPLC Community Tips & tricks Waters Corporation 1

ProSEC 300S. Protein Characterization columns

Column for High Performance,High-Binding Capacity Ion Exchange Chromatography:TSKgel SuperQ-5PW and Its Applications

Molecular Characterization of Biotherapeutics The Agilent 1260 Infi nity Multi-Detector Bio-SEC Solution with Advanced Light Scattering Detection

4/4/2013. BioHPLC columns. Paul Dinsmoor Biocolumn Technical Specialist. April 23-25, Size Exclusion BioHPLC Columns

Maximizing Chromatographic Resolution of Peptide Maps using UPLC with Tandem Columns

Biotherapeutic Method Development Guide

Introduction Ron Majors is a Senior Scientist at Agilent. Bill Champion is a chemist in Agilent s HPLC Columns tech support group.

Separation of Recombinant Human Erythropoietin (repo) Using Agilent Bio SEC-3

Higher Order mab Aggregate Analysis using New Innovative SEC Technology

Protein Separation with ph Gradients Using Composite Buffer Systems Calculated by the Agilent Buffer Advisor Software

Fast mass transfer Fast separations High throughput and improved productivity Long column lifetime Outstanding reproducibility Low carryover

for Acclaim Carbamate Column

Sean M. McCarthy and Martin Gilar Waters Corporation, Milford, MA, U.S. INTRODUCTION EXPERIMENTAL RESULTS AND DISCUSSION

Physical Stability of a Silica- Based Size Exclusion Column for Antibody Analysis

Ultrahydrogel Columns

Agilent MetaCarb Ca PLUS Carbohydrate Column Ca ++ Form User Manual Part No. A5205

ADVANCE ACCURACY AND PRODUCTIVITY FOR FASTER ANALYSIS

Rapid UHPLC Analysis of Reduced Monoclonal Antibodies using an Agilent ZORBAX Rapid Resolution High Definition (RRHD) 300SB-C8 Column

Application of Agilent AdvanceBio Desalting-RP Cartridges for LC/MS Analysis of mabs A One- and Two-dimensional LC/MS Study

Validation of a Dual Wavelength Size Exclusion HPLC Method with Improved Sensitivity to Detect Aggregates of a Monoclonal Antibody Biotherapeutic

Disulfide Linkage Analysis of IgG1 using an Agilent 1260 Infinity Bio inert LC System with an Agilent ZORBAX RRHD Diphenyl sub 2 µm Column

Choosing the Right Calibration for the Agilent Bio SEC-3 Column

Chromatography column for therapeutic protein analysis

ADVANCE ACCURACY AND PRODUCTIVITY FOR FASTER ANALYSIS

HIC as a Complementary, Confirmatory Tool to SEC for the Analysis of mab Aggregates

Characterization of mab aggregation using a Cary 60 UV-Vis Spectrophotometer and the Agilent 1260 Infinity LC system

HiTrap Desalting, 5 ml

Superdex 200 HR 10/30 INSTRUCTIONS

The New Standard in High Resolution Size Exclusion

Size exclusion chromatography for Biomolecule analysis: A "How to" Guide

Thermo Scientific. MAbPac HIC-10. Product Manual. P/N: June Part of Thermo Fisher Scientific

Reducing Cycle Time for Charge Variant Analysis of Monoclonal Antibodies

Econo-Pac High Capacity Ion Exchange Cartridges, 5 ml Instruction Manual. Catalog Numbers , ,

Simple Techniques for Improving the Isolation of Synthetic Peptides Jo-Ann Jablonski Principal Scientist Waters Corporation

Charge Heterogeneity Analysis of Rituximab Innovator and Biosimilar mabs

Size Exclusion Chromatography of Biosimilar and Innovator Insulin Using the Agilent AdvanceBio SEC column

Data File. HiLoad Superdex 30 prep grade HiLoad Superdex 75 prep grade HiLoad Superdex 200 prep grade. Pre-packed gel filtration columns

mab Titer Analysis with the Agilent Bio-Monolith Protein A Column

Chem 321 Lecture 23 - Liquid Chromatography 11/19/13

( ) IONPAC ATC-HC ANION TRAP COLUMN

PRODUCT DATA SHEET. Carboxylated Gold Nanoparticles. Description. Features. Storage. Applications. Handling. Characteristics

Application Note. Biopharma. Authors. Abstract. James Martosella, Phu Duong Agilent Technologies, Inc Centreville Rd Wilmington, DE 19808

Userguide: SepFast Gel Filtration columns

Clone Selection Using the Agilent 1290 Infinity Online 2D-LC/MS Solution

Thermo Scientific EASY-Spray LC Columns Optimization - Red flags to watch for and how to troubleshoot your workflow Estee Toole

ph gradient analysis of IgG1 therapeutic monoclonal antibodies using a 5 µm WCX column

EAG.COM MATERIALS SCIENCES APPLICATION NOTE. By J. Tompkins 1, T. Spurgeon 1, R. Tobias 1, J. Anders 1, E. Butler-Roberts 2, and M.

Characterization of IgG monomers & their aggregates

Agilent 1260 Infinity Bio-inert Quaternary LC System. Infinitely better for bio-molecule analysis

Techniques for Improving the Efficiency of Large Volume Loading in Preparative Liquid Chromatography

Transcription:

Size-Exclusion Chromatography (SEC) Optimization Guide 1

Table of Contents I. Introduction II. ACQUITY UPLC Instrumentation Considerations for Successful SEC Analysis III. Importance of Developing a Robust UPLC-based SEC Method a. Choosing the Appropriate ACQUITY UPLC Protein BEH SEC Column b. Method Development IV. Considerations When SEC Separated Components Tail and/or Resolution Decreases a. Connecting an ACQUITY UPLC Protein BEH SEC Column to an ACQUITY UPLC System b. LC System Dispersion Effects on a UPLC SEC Separation c. Determining LC System Dispersion Volume V. Maximizing ACQUITY UPLC Protein BEH SEC Column Life a. Mobile Phase Preparation and Use b. Avoiding Microbial Contamination of the Solvent Delivery System c. Cleaning ACQUITY UPLC Systems d. Minimizing Sample Particulate Contamination of Protein BEH SEC Columns e. Preventing Sample Formulation Constituent Contamination of Protein BEH SEC Columns f. Column Storage VI. Summary and References VII. Ordering Information 2

I. Introduction Introduction

I. Introduction Size-exclusion chromatography (SEC) is a technique commonly used throughout the development and commercialization of biotherapeutic proteins. Traditional HPLC-based SEC methods involve the use of columns packed with >3 µm particles containing an appropriate pore size distribution to allow for the separation of proteins of different hydrodynamic volumes (i.e., Stokes Radii). Ideally, the proteins are separated based solely on their relative size in solution. To achieve this, there should be no secondary interactions (e.g., hydrophobic or ionic) between the proteins and the SEC particles. In 2010, Waters was the first company to introduce the use of SEC columns containing <3 µm particles (patent pending) as an effective way to increase the resolution of separated components and to decrease analysis times. This technology, as implemented on an appropriately designed UPLC System, enabled the separation and quantitation of monoclonal antibody (mab) high molecular weight species (>300,000 Da), mab monomer (150,000 Da), and lower molecular weight species (e.g., <100,000 Da) in <8 minutes (Figure 1). It is the objective of this document to detail the instrument, method, and column-use considerations that are important to obtaining robust UPLC-based SEC separations of proteins with maximum column lifetimes.

Figure 1 mab mononer: @ 150k Daltons Conditions System: ACQUITY UPLC H-Class Bio 0.008 0.007 0.006 0.005 HMW mab dimer total area = 0.46% USP Rs between dimer and monomer = 2.35 LMW mab clip 1: @ 100k Daltons Column: Injection: 5 µl SEC eluent: ACQUITY UPLC Protein BEH SEC, 200Å, 1.7 µm, 4.6 x 300 mm (p/n: 186005226) 25 mm sodium phosphate with 0.15 M sodium chloride, ph 6.8 0.004 0.003 0.002 HMW mab dimer: @ 300k Daltons LMW mab clip 2: @ 50k Daltons Flow rate: Temp.: 30 C 0.4 ml/min UV detection: 280 nm 0.001 0.000 4.0 4.5 5.0 5.5 6.0 6.5 7.0 7.5 8.0 min ACQUITY UPLC BEH SEC Separation of a biotherapeutic monoclonal antibody (mab). 5

II. ACQUITY UPLC Instrumentation Considerations for Successful SEC Analysis ACQUITY UPLC Instrumentation

II. ACQUITY UPLC Instrumentation Considerations for Successful SEC Analysis Several factors contribute to the generation of high quality and reproducible UPLC-based, protein SEC separations. The selection of an appropriately configured Waters UPLC Instrument and the proper attachment of the SEC column to the LC system are major factors contributing to the generation of successful chromatography. Table 1 includes the recommended Waters UPLC System components for use in SEC separations that are frequently performed using salt-containing mobile phases that lack organic solvent. As noted in Figure 2, the use of non-recommended ACQUITY UPLC System components can have a significantly negative effect on the quality of the SEC separation. (For more information, reference: Guidelines for Routine Use and Maintenance of Ultra-Performance Size-Exclusion and Ion-Exchange Chromatography Systems, p/n: 720004182en on www.waters.com).

Figure 2 Lift here for Table 1 AU AU 0.015 0.010 0.005 0.000 0.030 0.020 0.010 0.000 Titanium 5 mm flow cell = 1 hour Analytical Teflon AF 10 mm flow cell 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 min ACQUITY UPLC Protein BEH SEC 200Å Chromatogram of bovine serum albumin (BSA) using an ACQUITY UPLC Photodiode Array Detector that shows the adverse effect of the flow-cell material of construction on peak shape. The BSA monomer exhibits extensive peak tailing when the flow cell consists of Teflon AF material. Conditions System: Column: Sample: Injection: 4 µl SEC eluent: Flow rate: Temp.: 30 C ACQUITY UPLC H-Class Bio ACQUITY UPLC Protein BEH SEC, 200Å, 1.7 µm, 4.6 x 300 mm (p/n: 186005226) Bovine serum album at 5 mg/ml in water 25 mm sodium phosphate with 0.15 M sodium chloride, ph 6.8 0.4 ml/min UV detection: 280 nm 9

Table 1. ACQUITY UPLC System Considerations for Size-Exclusion Chromatography (SEC). SYSTEM TYPE SYSTEM OVERVIEW Solvent Delivery Considerations for SEC Sample Management Considerations for SEC ACQUITY UPLC System Uses a Binary Solvent Manager (BSM) coupled to an autosampling sample manager. The BSM includes solvent degassing, switching, blending, and delivery elements. For SEC applications, Waters recommends a 10x increase in seal wash. The Sample Manager includes sample transport mechanics, sample temperature control, and sample injection fluidics. Sample injection volumes of 0.1 250 µl, in either a fixed-loop or partial-loop mode. For more viscous samples, reduce syringe draw rates. Waters does not recommend use of non-volatile buffers over 500 mm in needle wash solvents. For SEC applications, Waters recommends 90% aqueous/10% organic mixtures for seal washes. ACQUITY UPLC H-Class System Flexibility of Quaternary Solvent Blending and Auto Blend Plus Technology with direct-inject autosampling designed for routine sample analysis using non-salt containing eluents. The Quaternary Solvent Manager (QSM) is a high pressure pump that can simultaneously pump four degassed solvents (A, B, C, and D). For SEC applications, Waters recommends a 10x increase in seal wash. The Sample Manager-FTN uses a flow-through-needle design to aspirate samples. Follow Waters service guidelines for removing salt deposits and for routine maintenance of the sample syringe for SEC applications. For more viscous samples, reduce syringe draw rates. Waters does not recommend use of non-volatile buffers over 500 mm in needle wash solvents. For SEC applications, Waters recommends 90% aqueous/10% organic mixtures for seal washes. ACQUITY UPLC H-Class Bio System Flexibility of Quaternary Solvent Blending and Auto Blend Plus Technology with direct-inject autosampling designed for the separation of biomolecules. The system's flow path is engineered with non-stainless steel, iron-free materials to maximize system robustness. Constructed of bio-inert materials, the bio-quaternary solvent manager is a high pressure pump designed to accommodate the aqueous, high-ionic-strength solvents used in biopharmaceutical applications. No changes are needed for SEC with NaCl. (Reference technical brief: Size Exclusion Using ACQUITY UPLC H-Class Bio System, p/n: 720003602en on www.waters.com) The biosm-ftn uses a flow-through needle design to aspirate samples. The biocompatible needle, made of MP35N material, serves as part of the injection flow path when the sample is pushed onto the column. Waters does not recommend use of non-volatile buffers over 500 mm in needle wash solvents. For SEC applications, Waters recommends 90% aqueous/10% organic mixtures for seal washes. ACQUITY UPLC I-Class System Comprised of a Binary Solvent Manager (BSM) and a sample manager with fixed-loop or flowthrough-needle options. This system has the lowest system dispersion specifications and the lowest sample carryover for demanding LC-MS applications. The BSM is a high pressure pump that can simultaneously pump and proportion two degassed solvents (A1 or A2 with B1 or B2). The BSM utilizes two serial flow solvent delivery mechanisms operating in parallel to enable automatic linear and non-linear gradient profile implementation. For SEC applications, Waters recommends a 10x increase in seal wash. The Sample Manager-FTN uses a flow-through needle design to aspirate samples and hold it in the sample needle in preparation for injecting the sample onto the column. Sample injection volumes of 0.1 10 µl are supported by the standard configuration. Sample temperature is regulated over a range of 4 to 40 C using peltier technology. For more viscous samples, reduce syringe draw rates. Waters does not recommend use of non-volatile buffers over 500 mm in needle wash solvents. For SEC applications, Waters recommends 90% aqueous/10% organic mixtures for seal washes.

Column Heaters for SEC Accommodates one column, 20 150 mm length x 1.0 4.6 mm internal diameter (I.D.). 30-CHC: Accommodates multiple columns, serially connected columns supported (up to six); 20 300 mm length x 2.1 7.8 mm I.D., plus either a guard column or an in-line filter (20 mm maximum). UV/Vis Detector and Flow Cell Considerations for SEC ACQUITY UPLC TUV Detector: 500 nl (analytical flow cell volume); Waters recommends use of titanium (1.5 µl, 5 mm path length) or stainless steel optical flow cells (1.5 µl, 5 mm path length) when performing SEC applications under aqueous conditions. Other flow cell material (i.e., Teflon ) may cause peak tailing. ACQUITY Refractive Index Detector: 1.3 µl (flow cell volume). ACQUITY PDA Detector: Waters recommends TaperSlit for SEC applications. Typical UPLC System Dispersion Volume: No Column Heater 10 15 µl CH-A: Single column, up to 4.6 mm internal diameter (I.D.), up to 150 mm in length with filter or guard column. CH-30A: Single column, up to 4.6 mm I.D., up to 300 mm length with filter or guard column. 30-CHC: Accommodates multiple columns, serially connected columns supported (up to six); 20-300 mm length x 2.1 7.8 mm I.D., plus either a guard column or an in-line filter (20 mm maximum). ACQUITY UPLC TUV Detector: 500 nl (analytical flow cell volume); Waters recommends use of titanium (1.5 µl, 5 mm path length) or stainless steel optical flow cells (1.5 µl, 5 mm path length) when performing SEC applications under aqueous conditions. Other flow cell material (i.e., Teflon) may cause peak tailing. ACQUITY Refractive Index Detector: 1.3 µl (flow cell volume). ACQUITY PDA Detector: Waters recommends TaperSlit for SEC applications. 10 15 µl CH-A: Single column, up to 4.6 mm internal diameter (I.D.), up to 150 mm in length with filter or guard column. CH-30A: Single column, up to 4.6 mm I.D., up to 300 mm length with filter or guard column. 30-CHC: Accommodates multiple columns, serially connected columns supported (up to six); 20 300 mm length x 2.1 7.8 mm I.D., plus either a guard column or an in-line filter (20 mm maximum). ACQUITY UPLC TUV Detector: 500 nl (analytical flow cell volume); Waters recommends use of titanium (1.5 µl, 5 mm path length) or stainless steel optical flow cells (1.5 µl, 5 mm path length) when performing SEC applications under aqueous conditions. Other flow cell material (i.e., Teflon) may cause peak tailing. ACQUITY Refractive Index Detector: 1.3 µl (flow cell volume). ACQUITY PDA Detector: Waters recommends TaperSlit for SEC applications. 10 15 µl CH-A: Single column, up to 4.6 mm internal diameter (I.D.), up to 150 mm length with filter or guard column. ACQUITY UPLC TUV Detector: 500 nl (analytical flow cell volume); Waters recommends use of titanium (1.5 µl, 5 mm path length) or stainless steel optical flow cells (1.5 µl, 5 mm path length) when performing SEC applications under aqueous conditions. Other flow cell material (i.e., Teflon) may cause peak tailing. ACQUITY Refractive Index Detector: 1.3 µl (flow cell volume). ACQUITY PDA Detector: Waters recommends TaperSlit for SEC applications. 10 15 µl

III. Importance of Developing a Robust UPLC-Based BEH SEC Method UPLC-Based BEH SEC Method

III. Importance of Developing a Robust UPLC-Based BEH SEC Method a. Choosing the Appropriate ACQUITY UPLC Protein BEH SEC Column A number of factors need to be evaluated in the development of a UPLC Protein BEH SEC Separation Method that can reliably separate and accurately quantitate the various constituents contained in a protein sample. SEC separates compounds primarily based on their relative size in solution. Figure 3 shows calibration curves generated on ACQUITY UPLC Protein BEH SEC Columns of different pore sizes. This data is based on the use of proteins and peptides of known molecular weights. Chromatographers can use this data to select the most appropriate Protein BEH SEC column for a specific application performed on an appropriately configured Waters ACQUITY UPLC System.

Figure 3 MW 1,000,000 100,000 10,000 IgM (900 kda) Ovalbumin (44 kda) Blue Dextran Myoglobin (17 kda) RNase A (13.7 kda) Aprotinin (6.5 kda) Thyroglobulin (669 kda) Conalbumin (75 kda) IgG (150 kda) Amylglucosidase (97 kda) ACQUITY UPLC Protein BEH SEC, 125Å, 1.7 µm: 1K 80K Daltons ACQUITY UPLC Protein BEH SEC, 200Å, 1.7 µm: 10K 450K Daltons ACQUITY UPLC Protein BEH SEC, 450Å, 2.5 µm: 100K 1500K Daltons 1000 Angiotensin frag 1-7 (899 Da) 100 Allantoin (158 Da) Uracil (112 Da) 0.4 0.6 0.8 1.0 Normalized Retention Volume (Vr/VC) Calibration curves for ACQUITY UPLC Protein BEH SEC, 125Å, 200Å, and 450Å Columns. 15

b. Method Development A number of factors need to be evaluated in SEC method development. Ideally, SEC separations are based on differences in the size of native proteins in solution. For this reason, size-exclusion chromatography of biomolecules is usually performed under aqueous, physiological conditions. However, the presence of secondary interactions can obscure the desired sizebased separation and accurate quantitation of the resolved species. To minimize secondary interactions, the mobile phase and separation conditions need to be evaluated. It should be noted that the conditions used for the SEC separation can alter the protein structure and state. The concentration and type of salt used, as well as, the mobile-phase ph, can affect the three-dimensional structure and the protein-protein interactions. For these reasons, evaluation of a SEC method should be performed with the actual sample to be analyzed. (For more information, reference Method Development for Size-Exclusion Chromatography of Monoclonal Antibodies and Higher Order Aggregates, p/n: 720004076en on www.waters.com). 16

Figure 4 ph 6.0 [NaCl] 0.015 0.010 0.005 0.000 0.015 150 mm 250 mm 350 mm Monomer HMW LMW1 LMW2 6.5 7.0 7.5 UV absorbance @ 280 nm 0.010 0.005 0.000 0.015 0.010 0.005 0.000 0.015 0.010 0.005 0.000 4 5 6 7 8 4 5 6 7 8 4 5 6 7 8 min min min The effect of ph and ionic strength on the separation of a mab using an ACQUITY UPLC Protein BEH SEC, 200Å, 1.7 µm, 4.6 x 300 mm Column. 17

IV. Considerations When SEC Separated Components Tail and/or Resolution Decreases Tailing and/or Resolution Decreasing

IV. Considerations When SEC Separated Components Tail and/or Resolution Decreases a. Connecting an ACQUITY UPLC Protein BEH SEC Column to an ACQUITY UPLC System It is critical that the ACQUITY UPLC Protein BEH SEC Column be correctly installed onto the UPLC System. Figure 5 details the effects of proper vs. improper column connections on chromatographic performance. It is strongly recommended that new column inlet and outlet connections be made whenever a new ACQUITY UPLC Protein BEH SEC Column is installed onto the UPLC System. For more information on proper column connections, reference these videos on www.waters.com: How to Install a UPLC Column onto an ACQUITY UPLC System with an Active Column Preheater How to Connect an ACQUITY UPLC Column to an ACQUITY UPLC System

Figure 5 Proper Fit Improper Fit - Leak Improper Fit - Void Expected peak shape Expected retention Expected signal intensity Expected pressure Expected peak shape Higher retention Varied signal intensity Lower pressure Poor peak shape Expected retention Low signal intensity Expected pressure Effects of column-to-lc-system connections on chromatographic performance. 21

The effect of a purposely-created, 600 µl gap between the ACQUITY UPLC System solvent line and the BEH SEC column inlet is shown in Figure 6. This gap leads to a subtle increase in peak tailing, which can result in variability in the integrated area of the low-level LMW1 peak eluting as a trailing peak after the large mab monomer peak. The HMW and LMW2 peaks are unaffected as they are baseline resolved from the monomer peak. 22

Figure 6 0.008 0.007 0.006 600 µm Void Gap No Gap mab mononer: @ 150k Daltons LMW mab clip 1: @ 100k Daltons 1.22% LMW1 1.13% LMW1 Conditions System: Column: ACQUITY UPLC H-Class Bio with Tunable UV (TUV) detector ACQUITY UPLC Protein BEH SEC, 200Å, 1.7 µm, 4.6 x 300 mm (p/n: 186005226) UV (280 nm) 0.005 0.004 0.003 0.002 HMW mab dimer: @ 300k Daltons 0.51% HMW 0.50% HMW LMW mab clip 2: @ 50k Daltons 0.33% LMW2 0.33% LMW2 Sample: Injection: 5 µl SEC eluent: mab at 3 mg/ml in SEC mobile phase 25 mm sodium phosphate with 0.15 M sodium chloride, ph 6.8 0.001 Flow rate: 0.4 ml/min 0.000 Temp.: 30 C 4.0 4.5 5.0 5.5 6.0 6.5 7.0 7.5 8.0 min UV detection: 280 nm Effect of improper column-to-system connection on a mab separation. 23

b. LC System Dispersion Effects on a Waters ACQUITY UPLC Protein BEH SEC Separation In SEC, analytes elute within a single column volume during the isocratic separation. It is therefore critical to appreciate that the total LC system volume, including the injector, tubing, and detector volumes, affect the obtained separation. In general, and as shown in Figures 7 and 8, the lower the total LC system dispersion volume relative to the column volume, the narrower the peaks. The more dramatic effect of LC system dispersion volume on the ability to resolve challenging and closely eluting peaks (e.g., mab monomer from LMW mab clip 1) is shown in Figure 8. Most often, and for ACQUITY UPLC Protein BEH SEC separations, it is recommended that the total UPLC system dispersion volume be less than 15 µl. 24

Figure 7 4.6 x 150 mm, 1.7 µm, extra-column dispersion (4σ) = 7.4 µl LMW1> Rs = 1.13, p/v = 1.26, %area = 7.44 0.006 0.005 0.004 0.003 0.002 0.001 0.000 2.6 HMW Monomer LMW1 LMW2 4.6 x 150 mm, 1.7 µm, extra-column dispersion (4σ) = 23.9 µl LMW1> Rs = ND, p/v = 1.00, %area = 8.74 0.006 0.005 0.004 0.003 0.002 0.001 0.000 2.6 2.8 3.0 3.2 3.4 3.6 3.8 4.0 4.2 min 2.8 3.0 3.2 3.4 3.6 3.8 4.0 4.2 min 4.6 x 300 mm, 1.7 µm, extra-column dispersion (4σ) = 7.4 µl LMW1> Rs = 2.10, p/v = 3.66, %area = 4.60 0.008 0.006 0.004 0.002 0.000 5 4.6 x 300 mm, 1.7 µm, extra-column dispersion (4σ) = 23.9 µl LMW1> Rs = 1.69, p/v = 1.98, %area = 4.94 0.008 0.0045 0.0030 0.0015 0.000 5 6 7 8 min 6 7 8 min Effect UPLC system dispersion and SEC column length have on mab component resolution. Conditions System: Column: Sample: Injection: SEC eluent: Flow rate: Temp.: 30 C ACQUITY UPLC H-Class Bio with Tunable UV (TUV) detector ACQUITY UPLC Protein BEH SEC, 200Å, 1.7 µm, 4.6 x 150 mm (p/n: 186005225) 2 mg/ml mab IgG1 (Infliximab) 1 µl for single, 4.6 x 150 mm column; 2 µl for 4.6 x 300 mm column (i.e., two 4.6 x 150 mm columns in series) 20 mm sodium phosphate with 350 mm sodium chloride, ph 6.8 0.35 ml/min UV detection: 280 nm 25

c. Determining the LC System Dispersion Volume 1. Replace the column with a Zero-Volume Union (p/n: 700002636). The LC tubing should be 0.005" I.D. 2. Purge all LC solvent, wash, and purge lines with water, then 50/50 water/acetonitrile. 3. Set the detector to 273 nm and collect data at >40 points per second. 4. Flow rate: 0.5 ml/min. 5. Run time: 1 min. 6. Sample: 0.16 mg/ml caffeine in 50/50 water/acetonitrile. 7. Injection volume: 0.5 µl. 8. Inject three mobile-phase blanks followed by five caffeine sample injections. 9. To calculate the LC system volume: Measure the caffeine peak width (in minutes) at 4.4% peak height. Multiply the peak width by the flow rate to determine the peak volume width in ml. Multiply the peak volume width in ml by 1000 to determine the peak volume width in µl. Note: The average ACQUITY UPLC H-Class System volume dispersion, when measured using 4-Sigma Method, should be = 10.0 µl when used with CH-A, = 12 µl for CM-A, and <22.0 µl for 30CH-A column heater. If your value is greater, determine the source(s) of the deleterious extra peak dispersion volume and correct. 26

Figure 8 AU AU 0.30 0.20 0.10 0.00 0.25 0.20 0.15 0.10 0.05 0.00 USP Res= 2.37 USP Res= 1.37 Aggregates Aggregates Monomers Monomers H-Class BIO UPLC System Dispersion: <12 µl Alliance HPLC System Dispersion: @30 µl 2 4 6 8 10 12 14 16 18 20 22 24 26 28 min Effect of LC system dispersion on an ACQUITY UPLC Protein BEH SEC, 200Å, 1.7 µm Column separation of a human polyclonal IgG antibody. Conditions Column: Sample: Injection: 5 µl ACQUITY UPLC Protein BEH SEC, 200Å, 1.7 µm, 4.6 x 300 mm (p/n: 186005226) Humanized polyclonal lgg, 3 mg/ml in mobile phase Mobile phase: 25 mm sodium phosphate with 0.15 M sodium chloride, ph 6.8 Flow rate: Temp.: 30 C 0.2 ml/min UV detection: 280 nm Note: Unlike the mab sample used in Figures 1, 4, 6, and 7, the sample shown here did not contain either the LMW clip 1 (i.e., 100,000 Da) or LMW clip 2 (i.e., 50,000 Da). 27

V. Maximizing ACQUITY UPLC Protein BEH SEC Column Life Maximizing ACQUITY UPLC Protein BEH SEC Column Life

V. Maximizing ACQUITY UPLC Protein BEH SEC Column Life With proper care in preparing and handling the SEC mobile phase and samples, ACQUITY UPLC Protein BEH SEC Columns can deliver stable performance for >1000 sample injections (Figure 9). However, the performance and lifetime of an ACQUITY UPLC BEH SEC Column will deteriorate if the samples and/or mobile phase introduced into the column contain particulates. To maximize column lifetime, high quality filtered water (e.g., Milli-Q Millipak 0.22 µm filtered water) should be used, and mobile-phase buffers should be passed through a <0.2 µm filter. It is recommended that mobile-phase buffers be filtered using a sterile disposable filter. It is further recommended that sintered glass filter supports be avoided when filtering mobile phases due to the potential of introducing silicates which could alter column performance. Use of sterile 0.2-µm-filter units is recommended.

Figure 9 AU 0.016 0.008 0.000 ACQUITY UPLC BEH SEC, 200Å, 1.7 µm, 4.6 x 150 mm, 2493627917137 Injection # 1 2 3 4 5 min One hundred overlays from 1000 repetitive injections of commercial Vectibix mab formulation on a Waters ACQUITY UPLC Protein BEH SEC, 200Å, 1.7 µm, 4.6 x 150 mm Column. Conditions System: Column: Sample: Injection: 3.5 µl Mobile phase: Flow rate: Temp.: 30 C UV detection: ACQUITY UPLC H-Class Bio with Tunable UV (TUV) detector ACQUITY UPLC Protein BEH SEC, 200Å, 1.7 µm, 4.6 x 150 mm (p/n: 186005225) Vectibix, 10 mg/ml diluted 1:1 with water 0.1 M sodium phosphate with 0.1 M sodium chloride, ph 7.0 0.5 ml/min 280 nm 31

a. Mobile Phase Preparation and Use Mobile phases with ph values in the physiological range that do not contain organic solvent can support microbial growth. Implementation of the following recommendations will help minimize microbial contamination, thus avoiding premature column failure. Clean laboratory glassware properly and dry at >170 C for at least 1 hour to avoid potential microbial growth in remaining water droplets. The use of sterile containers is advised. Use only high purity water (18.2 MΩ cm) when preparing SEC mobile phases. If bottled water is used, it should be opened the day of use. Always filter prepared mobile phase through compatible 0.22 µm or smaller membrane filters. The use of sterile filters and containers is also recommended. Never top-off mobile-phase bottles. Always change bottles when replacing the mobile phase. Replace low-ionic-strength mobile phases (<150 mm) every 2-3 days. High-ionic-strength mobile phases (>150 mm) should be replaced every two weeks. This longer interval is acceptable because high salt concentrations will inhibit microbial growth. All mobile-phase bottles should be visually inspected daily for microbial growth and/or particulates. Microbial growth may form a film on the bottle surface and may be observed by swirling the bottle. It is highly recommended to remove and not use the Waters ACQUITY UPLC Mobile-Phase Line Sinkers when performing SEC with 100% aqueous mobile phases since they may become contaminated with microbes, thus contaminating freshly prepared solutions of filtered mobile phase. 32

Note: Waters Corporation will honor an existing service contract at sites that have removed the UPLC Mobile-Phase Line Sinkers provided that the account s SOP states that all SEC mobile phases must be filtered and contained in sterile glassware. If microbial growth is observed, immediately discontinue SEC analysis and perform the LC cleaning procedure detailed later in this document. Additional Considerations: To prevent potential precipitation of the buffer from the mobile phase in the LC system, good laboratory practice suggests maintaining a low flow rate (e.g., 0.1 ml/min) through the system after the SEC column has been removed and stored as recommended. If the UPLC System will be idle for more than 2 days, and after the SEC column has been removed and stored as recommended, prime each mobile-phase line for 15 minutes at 1 ml/min with high purity water followed by flushing and storage in 70/30 isopropanol (IPA)/water to prevent microbial growth. 33

b. Avoiding Microbial Contamination of the Solvent Delivery System Microbial contamination of the UPLC System may cause BEH SEC column fouling, resulting in an unacceptable separation (Figure 10). The column pressure may also increase significantly. Note that injection of particulates (e.g., microbes) will result in premature failure of any HPLC or UPLC Column. However, failure may occur sooner on columns containing <3 µm particles. If the mobile-phase reservoir becomes contaminated with microbes, the microbes will enter the solvent delivery flow path, and may contaminate fresh mobile phase. The mobile-phase line sinkers are one location that commonly harbors microbes. 34

Figure 10 For sample and separation conditions, reference Waters ACQUITY UPLC Protein BEH SEC Columns and Standards Care and Use Manual, p/n: 720003385en on www.waters.com. Effect of microbial growth on an ACQUITY UPLC Protein SEC separation of the BEH200 SEC Protein Standard Mix (p/n: 186006519). Contamination was confirmed by analysis of the column inlet frit (Figures 11 and 12). 35

Figures 11 and 12 Scanning electron micrograph of the inlet frit removed from an ACQUITY UPLC Protein BEH SEC, 200Å, 1.7 µm Column contaminated with bacteria. a) b) Mobile-phase line sinker (figure a) from a microbe-contaminated SEC mobile-phase bottle. The sinker was aseptically removed, suspended in sterile PBS, then 500 µl was applied to a microbe supporting agar plate. Figure b confirms the presence of microbes contained in the removed SEC mobile-phase line sinker. 36

c. Cleaning ACQUITY UPLC Systems To clean an ACQUITY UPLC System used in 100% aqueous SEC methods, the use of 100% IPA, as used in the standard UPLC cleaning protocols, should be replaced by the use of 70% isopropanol (IPA) or 70% ethyl alcohol (EtOH) due to their ability to kill bacteria. One hundred percent concentration of either IPA or EtOH does not penetrate the cell wall of bacteria and is therefore not an effective bactericide nor are methanol-containing solutions (Reference 3). Waters has experienced situations where "microbe-contaminated-mobile-phase filters" (a.k.a. sinkers) have resulted in rapid particulate fouling of a new ACQUITY UPLC Protein BEH SEC Column even when freshly prepared and filtered mobile phase is used. This is why Waters recommends that mobile phase sinkers NOT be used for this SEC application as previously detailed. 37

There are three different protocols for cleaning/purging ACQUITY UPLC Systems that can be used under different circumstances. 1. For a system that is in continuous use with sterile filtered mobile phase: The best practice for the daily use of buffers starts with fresh mobile phase in a new bottle (no topping off). After installing fresh mobile phases, wet prime (a function of Empower Software) the buffer lines for 4 minutes if the system was used the day before. If the system was not used the day before, wet prime all lines for 6-10 minutes. This may be accomplished by using the system startup function and changing the prime time as needed. This function also primes the seal wash, purge, and needle wash solvent lines. Keep in mind that when the ACQUITY UPLC System is wet primed, under the control of the ACQUITY UPLC console, the column can remain on the system connected to the pump and detectors. The Empower solvent priming function diverts the priming solvents directly to waste. This makes it very easy to purge your lines with mobile phase or the other solvent sequences described below. It does require the movement of your solvent lines, but if you catch a bubble during the process, the wet priming command will eliminate it. 2. For a system that has been left unused for more than 2-3 days: If your system is left in buffer with or without flow for more than 2-5 days, the lifetime of your column may benefit by wet priming all buffer lines with water for 4 minutes, then with 70% IPA for 6-10 minutes before going back to water for 4 minutes and then your buffer for 4 minutes before starting another analysis. It is further recommended to leave your system in 70% IPA (after priming the buffer lines with water) when not in use. When restarting, wet prime your system with 70% IPA for 3-4 minutes then water for 4 minutes before wet priming your system with your desired buffer. 38

3. For routine periodic cleaning for preventive maintenance or when the system is contaminated: Table 2 is a summary of the cleaning protocol and reagents required to execute this procedure, which can be used at regular intervals (every 3-6 months) for preventive maintenance. a. Disconnect the column and attach a V-Detail Zero Volume Union (p/n: 700002636) to the column inlet and outlet lines. b. For each step of the cleaning protocol, monitor the system pressure to keep it <1000 psi to prevent damage to the detector flow cell. If necessary, replace the backpressure device with large bore waste tubing. c. If the system has recently been used with mobile-phase buffers or salts, it should be flushed thoroughly with 100% high-purity water prior to the introduction of organic solvents. d. Place A, B, C, D, seal wash, sample wash, and purge lines into a clean bottle containing 70% isopropanol as the cleaning solvent. Note: If there is an incompatibility between the mobile phase or wash solvents with 70% isopropanol, flush first with the appropriate intermediate solutions to ensure compatibility. 39

e. Prime solvent lines A, B, C, and D for 5 minutes each. The Empower Prime command vigorously forces solvents through the lines and system at 4 ml/min to assist in the cleaning process. f. Prime the seal wash. g. Prime the wash solvent for 200 seconds and the purge solvent for 40 cycles. h. Purge the system at 0.2 ml/min for 10 minutes using 25% A, 25% B, 25% C, and 25% D. i. Repeat steps d through h using 100% methanol as the cleaning solvent, using a flow rate of 1.0 ml/min for step h. j. Repeat steps d through h using 100% high-purity water as the cleaning solvent, using a flow rate of 1.0 ml/min for step h. Caution: If a nebulization-based detector is connected (MS or ELSD), take it off line before performing step k. Direct the flow from the union outlet to waste. k. Repeat steps d through h using 10% aqueous phosphoric acid as the cleaning solvent, using a flow rate of 0.2 ml/min for step h. Caution: To avoid damage, do not place the seal wash line in this solution. Place the seal wash line in high purity water. l. Repeat steps d through h using 100% high-purity water as the cleaning solvent, using a flow rate of 1.0 ml/min for step h. m. If applicable, reconnect the MS or ELS detector. n. Repeat steps d through h using 100% methanol with a flow rate of 1.0 ml/min for step h. 40

Table 2. ACQUITY UPLC System Cleaning Protocol Summary. Check ( ) when step is completed. Eluent Flow Rate Prime Lines A, B, C, D (5 min each) Prime Wash Solvent (200 sec) Prime Purge Solvent (40 cycles) Prime Seal Wash Purge A/B/C/D 25/25/25/25 (10 min) IPA 0.2 ml/min Methanol 1 ml/min High Purity Water 1 ml/min Leave seal wash line in high purity water. If used, remove nebulization-based detectors (MS or ELSD) from flow path. 10% Aqueous 0.2 ml/min Phosphoric Acid High Purity Water 1 ml/min If used, return nebulization-based detectors to flow path. Methanol 1 ml/min Required Cleaning Solvents 100% high purity water Milli-Q water or equivalent 100% methanol Fisher Methanol, Optima #A454-4 or equivalent 70% isopropanol (IPA) in high purity water (v,v) Fisher 2-Propanol, Optima #A464-4 or equivalent 10% phosphoric acid (aqueous) Prepare 10% (v/v) using HPLC-grade 85% phosphoric acid and high-purity water 41

d. Minimizing Sample Particulate Contamination of Protein BEH SEC Columns The performance and lifetime of an ACQUITY UPLC Protein BEH SEC Column can deteriorate due to the accumulation of particulates. These include insoluble protein aggregates contained in the injected sample. As a result, it is recommended to filter all protein-containing samples using sterile, low protein binding, 0.2 µm syringe filters. Alternatively, ACQUITY UPLC Sample Vials containing sample can be centrifuged so that any insoluble particulates form a solid pellet at the bottom of the vial. Note: Be sure to adjust the ACQUITY UPLC needle depth to prevent sampling any particulates located at the bottom of the centrifuged vial. For more information on setting the needle depth and minimum volume in vials, reference "Sample Vials and Accessories", p/n: 720001818en on www.waters.com. e. Preventing Sample Formulation Constituent Contamination of Protein BEH SEC Columns Studies have found a variety of chemical and physical conditions that can cause the creation of biotherapeutic protein aggregates. To minimize this occurrence, excipients are typically added to therapeutic protein formulations. However, these excipients can cause inaccurate SEC quantitation due to changes in retention, peak shape, or resolution. While a variety of SEC column cleaning protocols may be used to restore a column's performance, this can be time consuming and ultimately ineffective. As a last resort, column replacement is often required, resulting in higher costs and delays. As indicated in Figure 13, the use of guard columns may protect Protein BEH SEC columns from deterioration due to chemical contamination. For more information, reference Improving the Lifetime of UPLC Size-Exclusion Chromatography Columns Using Short Guard Columns, p/n: 720004034en on www.waters.com. 42

Figure 13 9000 8000 7000 6000 USP plate count 5000 4000 3000 2000 1000 Column and guard Column without guard 0 0 100 200 300 400 500 600 700 800 900 1000 Injection number Effect of using an ACQUITY UPLC Protein BEH SEC Guard Column (4.6 x 30 mm) on column efficiency. The arrows indicate where the guard column was changed, noting how the column life was extended (top/red) vs. a control that did not use a guard column (bottom/blue). 43

f. Column Storage To prevent the potential of microbial growth in SEC columns that will be removed from your system and stored, we recommend the following: 1. Start by using a solvent line that is already in 100% water. Then using this line, flush the existing mobile phase from the column with approximately 20-column volumes of HPLC grade and filtered water (e.g., 50 ml for 4.6 x 150 mm columns) to remove any buffer and salts from the SEC column. 2. After preparing the UPLC System as needed, flush the SEC column with approximately 20-column volumes of filtered 20% methanol/80% water to prevent bacterial growth. The column can then be safely stored at room temperature. Note: It is important that all salt and/or protein be flushed from the BEH SEC Column BEFORE storage in 20% MeOH. Failure to remove salt and/or protein may cause precipitation in the column when it is exposed to the 20% organic storage solvent. 44

VI. Summary and References Summary and References

VI. Summary and References Summary With the practices described, robust ACQUITY UPLC Protein BEH SEC analyses can be achieved, offering faster and higher resolution separations than those obtained using traditional HPLC-based SEC methods. To maximize the lifetime of an ACQUITY UPLC Protein BEH SEC Column, it is important to avoid microbial growth in the mobile phase as well as to ensure that injected samples are free of particulates. The use of BEH SEC guard columns can be an effective means to protecting your analytical SEC column from deterioration due to these various types of failure mechanisms. References 1. Hong, P.; Koza, S.M.; Bouvier, E.S.P. A Review: Size-exclusion chromatography for the analysis of protein biotherapeutics and their aggregates. J. Liq. Chrom. Rel. Tech. 2012, 35(20), 2923-2950. 2. Bouvier, E.S.P.; Koza, S.M. Advances in size-exclusion separations of proteins and polymers by UHPLC. Trends Analyt. Chem. 2014, 63, 85-94. 3. Guideline for Disinfection and Sterilization in Healthcare Facilities, 2008, William A. Rutala, David J. Weber, M.D., and the Healthcare Infection Control Practices Advisory Committee. US CDC, Atlanta, Georgia. (https://www.cdc.gov/hicpac/pdf/ guidelines/disinfection_nov_2008.pdf)

VII. Ordering Information Ordering Information

VII. Ordering Information ACQUITY UPLC Protein BEH SEC Columns and Guard Kits Description Configuration Particle Size Dimension P/N ACQUITY UPLC Protein BEH SEC, 450Å Column and Standard 2.5 µm 4.6 x 150 mm 176002996 ACQUITY UPLC Protein BEH SEC, 450Å Column 2.5 µm 4.6 x 150 mm 186006851 ACQUITY UPLC Protein BEH SEC, 450Å Column and Standard 2.5 µm 4.6 x 300 mm 176002997 ACQUITY UPLC Protein BEH SEC, 450Å Column 2.5 µm 4.6 x 300 mm 186006852 ACQUITY UPLC Protein BEH SEC, 450Å Guard Kit 2.5 µm 4.6 x 30 mm 186006850 ACQUITY UPLC Protein BEH SEC, 200Å Column 1.7 µm 2.1 x 150 mm 186008471 ACQUITY UPLC Protein BEH SEC, 200Å Column and Standard 1.7 µm 4.6 x 150 mm 176003904 ACQUITY UPLC Protein BEH SEC, 200Å Column 1.7 µm 4.6 x 150 mm 186005225 ACQUITY UPLC Protein BEH SEC, 200Å Column and Standard 1.7 µm 4.6 x 300 mm 176003905 ACQUITY UPLC Protein BEH SEC, 200Å Column 1.7 µm 4.6 x 300 mm 186005226 ACQUITY UPLC Protein BEH SEC, 200Å Guard Kit 1.7 µm 4.6 x 30 mm 186005793 ACQUITY UPLC Protein BEH SEC, 125Å Column and Standard 1.7 µm 4.6 x 150 mm 176003906 ACQUITY UPLC Protein BEH SEC, 125Å Column 1.7 µm 4.6 x 150 mm 186006505 ACQUITY UPLC Protein BEH SEC, 125Å Column and Standard 1.7 µm 4.6 x 300 mm 176003907 ACQUITY UPLC Protein BEH SEC, 125Å Column 1.7 µm 4.6 x 300 mm 186006506 ACQUITY UPLC Protein BEH SEC, 125Å Guard Kit 1.7 µm 4.6 x 30 mm 186006504 ELSD Outlet Tubing (0.004" I.D. x 6" length) Tubing 430001562 0.005 x 1.75" SEC UPLC Connection Tubing, 2/pk Tubing 186006613 NOTE: Size-exclusion chromatography may require modifications to an existing ACQUITY UPLC System. Please reference Size-Exclusion and Ion-Exchange Chromatography of Proteins using the ACQUITY UPLC System (p/n: 715002147) or Size Exclusion and Ion-Exchange Chromatography of Proteins using the ACQUITY UPLC H-Class System (p/n: 715002909) on www.waters.com for specific recommendations.

XBridge Protein BEH SEC HPLC and UHPLC Columns Description Configuration Particle Size Dimension P/N XBridge Protein BEH SEC, 125Å Column with BEH125 SEC Protein Standard Mix Guard Column 3.5 µm 7.8 x 30 mm 176003591 XBridge Protein BEH SEC, 125Å Column with BEH125 SEC Protein Standard Mix Column 3.5 µm 7.8 x 150 mm 176003592 XBridge Protein BEH SEC, 125Å Column with BEH125 SEC Protein Standard Mix Column 3.5 µm 7.8 x 300 mm 176003593 XBridge Protein BEH SEC, 200Å Column with BEH200 SEC Protein Standard Mix Guard Column 3.5 µm 7.8 x 30 mm 176003594 XBridge Protein BEH SEC, 200Å Column with BEH200 SEC Protein Standard Mix Column 3.5 µm 7.8 x 150 mm 176003595 XBridge Protein BEH SEC, 200Å Column with BEH200 SEC Protein Standard Mix Column 3.5 µm 7.8 x 300 mm 176003596 XBridge Protein BEH SEC, 450Å Column with BEH450 SEC Protein Standard Mix Guard Column 3.5 µm 7.8 x 30 mm 176003597 XBridge Protein BEH SEC, 450Å Column with BEH450 SEC Protein Standard Mix Column 3.5 µm 7.8 x 150 mm 176003598 XBridge Protein BEH SEC, 450Å Column with BEH450 SEC Protein Standard Mix Column 3.5 µm 7.8 x 300 mm 176003599 Straight Connection Tubing and End-fittings for XBridge Protein BEH SEC Column U-Bend Connection Tubing and End-fittings for XBridge Protein BEH SEC Column Tubing and End-fittings Tubing and End-fittings WAT022681 WAT084080 ACQUITY UPLC BEH SEC Protein Standards Description P/N BEH125 SEC Protein Standard Mix 186006519 BEH200 SEC Protein Standard Mix 186006518 BEH450 SEC Protein Standard Mix 186006842 49

2017 Waters Corporation. Waters, The Science of What s Possible, ACQUITY, UPLC, Alliance, and Empower are registered trademarks of Waters Corporation. Auto Blend Plus is a trademark of Waters Corporation. Teflon is a registered trademark of The Chemours Company. Vectibix is a registered trademark of Amgen. Milli-Q is a registered trademark of EMD Millipore Corporation. Optima is a trademark of Fisher Chemical. All other trademarks are the property of their respective owners. November 2017 720006067EN KP.2Cool 50