The Analysis of 4-Vinyl-Cyclo-Hexene (VCH) in Butadiene with the DVLS LGI Injector

Similar documents
Improved Gas Chromatograph Method for the Analysis of Trace Hydrocarbon Impurities in 1, 3-Butadiene

APPLICATION NOTE. UOP : Analysis of Trace CO and CO 2 in bulk H 2 and Light Gaseous Hydrocarbons by GC

Gas Chromatography SSI-GC-003

SCION GC Series. The Gas Chromatographers Choice for Separations

SCION GC Series. The Gas Chromatographers Choice for Separations

Hydrocarbon Processing Industry Gas Chromatographs. Standard GC Analyzer Solutions

Prefractionator for Reliable Analysis of the Light Ends of Crude Oil and other Petroleum Fractions

Gas Chromatography Assignment Chem 543/443

The Importance of High Speed Data Acquisition in the Design and Implementation of GC/MS Applications

Performance characteristics of the 1260 Infinity Quaternary LC system

A Rapid Method for Trace Analysis of Organophosphorus Pesticides in Drinking Water

Fast On-Site Mine Safety Analysis by the Agilent 490 Micro GC

METHOD 310A - DETERMINATION OF RESIDUAL HEXANE THROUGH GAS CHROMATOGRAPHY

Analysis of Permanent Gases and Extended Hydrocarbons in Natural Gas and Natural Gas Liquids by Capillary Gas Chromatography

Determination of Volatile Organic Compounds (VOCs) in Various Drinking Water Sources by GC/MS

Ghost Peaks in Gas Chromatography Part 2: The Injection Port

Analysis of Suspected Flavor and Fragrance Allergens in Cosmetics Using the 7890A GC and Capillary Column Backflush Application

Improving GC-MS Method Robustness and Cycle Times Using Capillary Flow Technology and Backflushing

Fast Refinery Gas Analysis Using the Agilent 490 Micro GC QUAD

Optimization of Volatile Petroleum Hydrocarbons Analysis by Automated Headspace Using Method Development Tools

7890A An Essential Partner for the Trace

The Use Of Automated Backflush on the 7890A/5975A GC-MS System. Application. Setup Example Using Biomarkers. Authors. Abstract.

Using a Micro GC to measure, monitor, and optimize your process. ASTS, Houston June 2015 Shannon Coleman Agilent Technologies

Technical Overview. Author. Abstract. A.G.Huesgen Agilent Technologies, Inc. Waldbronn, Germany

Agilent Gas Phase Solutions

METHOD TRANSLATION OF HJ FOR INTUVO

Olefin Analyzer Solutions for Polymer Grade Ethylene and Propylene

Residual Solvents in medical cannabis analysis using the SRI 8610C FID GC June 2014

2.1 This procedure provides steps and conditions for the determination of Vapor Space Organic Compounds in ground waters, and industrial waste waters.

Determination of Diesel Range Organics (DROs) Using SPE and GC/FID by Method 8015D*

Improved Performance and Productivity of Purge-and-Trap Analyses

High temperature Simdist analysis using the new MXT-1 HT Simdist phase. Jaap de Zeeuw and Gary Stidsen Restek corporation, Bellefonte, PA, USA

Mach-RGA. The Fast Refinery Gas Analyzer. Analyzer

Roger Bardsley, Applications Chemist; Teledyne Tekmar P a g e 1

Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1

Turnkey configurations for gas chromatographic analysis, including instrumentation, methods, supplies and service. Analysis of Natural Gas

Pesticide Analysis in a Complex Matrix using a 2D GC/MS System

Chem 321 Lecture 23 - Liquid Chromatography 11/19/13

SCION GC Series. The Gas Chromatographers Choice for Separations

ADVANCES IN MICRO GAS CHROMATOGRAPHY

Standard Test Methods for Aromatics in Mineral Spirits by Gas Chromatography 1

Fast Detailed Hydrocarbon Analysis by Modified ASTM Method D6730

Protein-Pak Hi Res HIC Column and HIC Protein Standard

Impact of GC Parameters on The Separation Part 4: Choice of Film Thickness

Optimizing Productivity and Reliability for Monocyclic Aromatic Hydrocarbon Purity Analysis According to ASTM D7504 on the Agilent 8890 GC System

Protocol for Quantitative Determination of Residual Solvents in Cannabis Concentrates Prepared by: Amanda Rigdon, May 23 rd, 2016

Low-level Detection of Ethanol, 1,4-Dioxane, and Other Oxygenates Using the Eclipse Purge-and-Trap Sample Concentrator

Nitrosamines Analysis in Drinking Water Using GC/MS/MS Meeting Equivalence to EPA Method 521

Fast Method Development Using the Agilent 1290 Infinity Quaternary LC System with Column Selection Valve

Direct Injection of Fish Oil for the GC-ECD Analysis of PCBs: Results Using a Deans Switch With Backflushing. Application. Author.

Large Volume Injection of Organochlorine Pesticide Extracts Using a Programmable Temperature Vaporizer in the Solvent Split Mode

Performance Characteristics of the Agilent 1220 Infinity Gradient LC system

Fast Analysis of Extended Natural Gas using the New Transportable INFICON Micro GC Fusion

Agilent Consumables Improve Performance with Shimadzu GC/MS

Methane baseline concentrations and sources in shallow aquifers from the shale. gas-prone region of the St. Lawrence Lowlands (Quebec, Canada)

ADVANCES IN MICRO GAS CHROMATOGRAPHY

ABB MEASUREMENT & ANALYTICS DATA SHEET. PGC5007 Total sulfur analyzer

Analysis of an Acrylic Copolymer using Pyrolysis-GC/MS

Historical Review: Fast & micro Gas Chromatography

Forensic Toxicology. Analysis of Ethanol in Blood using Master SHS Static Headspace Sampler and Master GC Gas Chromatograph APPLICATION NOTE.

Application Note. Author. Abstract. Pharmaceuticals. Detlef Wilhelm ANATOX GmbH & Co. KG. Fuerstenwalde, Germany mau

CryoTrap. For Agilent Technologies 7890/6890/6850 GC Systems. CryoTrap

ADVANCES IN MICRO GAS CHROMATOGRAPHY

Agilent 490 Micro GC. The Measure Anywhere Micro GC February 2011

Simplifying Methods Transfer: Novel Tools for Replicating Your Established Methods on an ACQUITY Arc System

Reducing Cycle Time for Charge Variant Analysis of Monoclonal Antibodies

Fast Gas Chromatography using Heated Headspace Gas Auto Sampling Techniques: Polyethylene Pellets, Product

Analysis of Semivolatiles by GC/FID using the Ultra Inert Inlet Liners with Wool

EPA Method 543: Selected Organic Contaminants by Online SPE LC/MS/MS Using the Agilent Flexible Cube

TurboMatrix Headspace and Headspace Trap Samplers

Technical Article. Innovative Chromatography Solutions. The Rtx -CLPesticides Column Set Lets You Make the Choice

FASTER SMALLER SMARTER EASIER GREENER

High-resolution Analysis of Charge Heterogeneity in Monoclonal Antibodies Using ph-gradient Cation Exchange Chromatography

Analysis of USP <467> Residual Solvents using the Agilent 7697A Headspace Sampler with the Agilent 7890B Gas Chromatograph

New Developments in Thermal Desorption (TD) Tube & Canister Technologies for Collection & Analysis of Soil Gas

Rapid Microchamber Tests for Screening Chemical Emissions from Car Trim in Accordance with ISO

GC Troubleshooting. Frank Michel, Katherine K. Stenerson T HCE

Field Tests of Chloroform Collection/Analysis Methods

Alumina BOND/MAPD PLOT Columns!

MARKET FOCUSED - environmental ORGANOCHLORINE PESTICIDES

MicroSAM / Sitrans CV

IC-D3160 Ion Chromatograph

Scale-up with the Agilent SD-1 Purification System analytical and preparative runs on a single system

Separation and Quantitation of a Mixture of Alcohols by Gas Chromatography

Application Note. Abstract. Authors. Environmental

PROCESS MOISTURE ANALYZERS Measuring moisture in gas or HC liquids in hazardous areas

Improving Productivity in Your Environmental and Food Testing Labs with Agilent 7890A/5975C GC/MS System

The Agilent StreamSelect LC/MS Solution: Increasing the Throughput of a Triple Quadrupole Mass Spectrometer

GC Method Development

Improving Retention Time Precision and Chromatography of Early Eluting Peptides with Acetonitrile/Water Blends as Solvent B

Agilent 1290 Infinity II 2D-LC Solution Biopharmaceutical Polymer Analysis. WCBP Jan 2017 Washington, DC

Field Analytical Method Development for C 6 -C 9 Range Hydrocarbon Measurement

US EPA Method Atomx XYZ and Agilent 7890B GC/5977A MS

High Throughput Sub-4 Minute Separation of Antibodies using Size Exclusion Chromatography

GC Inlets An Introduction

Volatile Organic Carbon Contaminated Site Assessment

Physical Stability of a Silica- Based Size Exclusion Column for Antibody Analysis

Two Headspace Solutions for the Analysis of Dissolved Gases in Water by Method RSKSOP-175

SIMPLIFY AND ACCELERATE GAS ANALYSIS. Micro GC Fusion GAS ANALYZER

Chromatography Applications and Technology

Transcription:

The Analysis of 4-Vinyl-Cyclo-Hexene (VCH) in Butadiene with the DVLS LGI Injector Introduction In synthetic rubber manufacturing, 4-Vinylcyclohexene (VCH) can be found as a byproduct from the chemical processing of 1,3-Butadiene. VCH is a dimer of 1,3-Butadiene and is normally present in Butadiene. It is generally controlled below the 500-1000 ppmw level per product specifications. VCH is a very toxic and colorless liquid having a strong odor in normal conditions. Application Note Authors: Lenny Kouwenhoven, Director R&D of Da Vinci Laboratory Solutions Anita Ruissen, Application Specialist of Da Vinci Laboratory Solutions VCH specification and test method Control over VCH is important for Butadiene consumers. As indicated the product specification for VCH is between 500 and 1000 ppm. At present there is no known inhibitor for the Butadiene dimer reaction. It is strictly a time/ temperature phenomenon. The lower the temperature, the slower the rate of dimer formation. Conversely, the higher the temperature, the faster the rate of dimer formation. To minimize dimer formation the transit and residence/storage times should be strongly shortened. Also storing Butadiene at low temperatures will minimize the amount of dimer formed. Dimer removal is currently done via distillation. To accurately determine the VCH content in Butadiene Da Vinci Laboratory Solutions developed the Liquefied Gas Injector (LGI). Application Description The method uses the Liquefied Gas Injector. The sample is injected under pressure directly onto the column. The Butadiene sample remains in liquid phase, at room temperature and without contact with transfer lines, vaporizers or valves. As a result all limitations of the conventional sample introduction techniques are resolved. Figure One: the DVLS Liquefied Gas Injector The chromatography is based on boiling point separation. The total amount is reported in parts per million. The system setup is based on the ASTM D7756-11: Standard Test Method for Residues in Liquefied Petroleum (LP) Gases by Gas Chromatography with Liquid, On-Column Injection. The GC is equipped with the Liquefied Gas Injector as displayed in Figure One, an on-column injector and a solvent vapor exit. Boosting Laboratory Efficiency

Figure Two shows the configuration of retention gap and columns. Sample is injected into a 5 meter Sulfinert coated stainless steel capillary. The retention gap is connected to a 3 meter non polar retaining column, with an exit for flushing the Butadiene light ends. Subsequently, the exit is closed and the flow is switched to the non-polar analytical column for the elution of VCH. In the same analysis another important parameter can be determined: Tertiary Butyl Catechol. Table One shows the LGI settings, Table Two shows typical settings of the gas chromatograph and column details. Figure Two: Column Configuration Analytical Results The VCH was dissolved in Toluene and four concentrations of VCH in Pentane were prepared; 24.7 ppm, 247 ppm, 494 ppm and 988 ppm. Figures Three, Four, Five show the chromatogram of 125 ppm VCH. Table Three lists the precision found. Equilibration Time Oven Program Run Time Back COC Inlet He 1 min 45 C (2.0 min), 25 C/min 250 C (0 min) 12.2 min 55 C (2.0 min), 25 C/min 250 C (0 min) Injection Time 50 ms Flow 4mL/min Pre Injection Delay 1 sec Septum Purge Flow 12 ml/min Post Injection Delay Solvent Vent Stop Flow 1 sec 10 sec 0 sec Table Two: GC Parameters Table One: LGI Parameters Concentration VCH 24.7 ppm 247 ppm 494 ppm 988 ppm Average Counts 3543 34305 69821 143264 Standard Deviation 31 359 1039 1378 % RSD 0.9 1.0 1.5 1.0 Table Three: Precision of four concentrations VCH in Pentane

Figure Three: Chromatogram of 24.7 ppm VCH in Pentane Figure Four: Chromatogram of 247 ppm VCH in Pentane

Figure Five: Chromatogram of 494 ppm VCH in Pentane Figure Six: Chromatogram of 988 ppm VCH in Pentane

Figure Seven: Overlay of four concentrations of VCH in Pentane Conclusion A method developed by DVLS uses the new Liquefied Gas Injector (LGI) to inject Butadiene under pressure, in liquid phase directly on the analytical GC column. Analytical results have demonstrated that the LGI technique is a safe, fast and accurate method for the determination of VCH in Butadiene. In the same analysis Tertiary Butyl Catechol can be determined. The repeatability is better than 2% relative and the lower detection limit is far below 1 ppm. References: 1. ASTM D7756-11 :Standard Test Method for Residues in Liquefied Petroleum (LP) Gases by Gas Chromatography with Liquid, On- Column Injection 2. The analysis of Contaminants in Liquefied Gases by Gas Chromatography by Lenny Kouwenhoven and Anita Ruissen, Petro Industry News, October/November 2011 3. A Safe and Fast Solution for Accurate Quantification of Heavy Residues in LPG by Gas Chromatography, Representative Liquefied Gas Sample Introduction via High Pressure On-Column Injection into a Gas Chromatographic System, by Lenny Kouwenhoven and Anita Ruissen, Petro Industry News, August/September 2012 DA VINCI LABORATORY SOLUTIONS B.V. P.O. Box 12103, 3004 GC Rotterdam - The Netherlands T: +31 (0)10 258 1870 - F: +31 (0)10 258 1879 - E-mail: solutions@davinci-ls.com www.davinci-ls.com