Size Exclusion Chromatography/ Mass Spectrometry Analysis of Antibody Drug Conjugates Using the Agilent 1260 Infinity II Bio-Inert LC

Similar documents
Precise Characterization of Intact Monoclonal Antibodies by the Agilent 6545XT AdvanceBio LC/Q-TOF

Fraction Analysis of Cysteine Linked Antibody-Drug Conjugates Using Hydrophobic Interaction. chromatography. Agilent 1260 Infinity II Bio-Inert System

Application Note. Author. Abstract. Biotherapeutics and Biologics. Sonja Schneider Agilent Technologies, Inc. Waldbronn, Germany

Clone Selection Using the Agilent 1290 Infinity Online 2D-LC/MS Solution

Application of Agilent AdvanceBio Desalting-RP Cartridges for LC/MS Analysis of mabs A One- and Two-dimensional LC/MS Study

Analysis of Monoclonal Antibody (mab) Using Agilent 1290 Infinity LC System Coupled to Agilent 6530 Accurate-Mass Quadrupole Time-of-Flight (Q-TOF)

Characterize Fab and Fc Fragments by Cation-Exchange Chromatography

PLRP-S Polymeric Reversed-Phase Column for LC/MS Separation of mabs and ADC

Seamless Method Transfer from an Agilent 1260 Infinity Bio-inert LC to an Agilent 1260 Infinity II Bio-inert LC

Characterize mab Charge Variants by Cation-Exchange Chromatography

Cell Clone Selection Using the Agilent Bio-Monolith Protein A Column and LC/MS

Separation of Native Monoclonal Antibodies and Identification of Charge Variants:

Reducing Cycle Time for Charge Variant Analysis of Monoclonal Antibodies

Cell Culture Optimization Using an Agilent Bio-Monolith Protein A Column and LC/MS

Profiling Glycosylation of Monoclonal Antibodies at Three Levels Using the Agilent 6545XT AdvanceBio LC/Q TOF

Disulfide Linkage Analysis of IgG1 using an Agilent 1260 Infinity Bio inert LC System with an Agilent ZORBAX RRHD Diphenyl sub 2 µm Column

Molecular Characterization of Biotherapeutics The Agilent 1260 Infi nity Multi-Detector Bio-SEC Solution with Advanced Light Scattering Detection

Benefits of 2D-LC/MS/MS in Pharmaceutical Bioanalytics

A Comprehensive Workflow to Optimize and Execute Protein Aggregate Studies

Charge Heterogeneity Analysis of Rituximab Innovator and Biosimilar mabs

Emulation of the Agilent 1100 Series LC Through Waters Empower Software Analysis of an Analgesic Mixture

Identifying Monoclonal Antibody Mutation Sites Using the Agilent 1290 Infinity II 2D-LC Solution with Q-TOF LC/MS

High-resolution Analysis of Charge Heterogeneity in Monoclonal Antibodies Using ph-gradient Cation Exchange Chromatography

N-Glycan Profiling Analysis of a Monoclonal Antibody Using UHPLC/FLD/Q-TOF

Separate and Quantify Rituximab Aggregates and Fragments with High-Resolution SEC

Method Transfer from an Agilent 1200 Series LC to an Agilent 1260 Infinity II LC

Application Note. Authors. Abstract. Biopharmaceuticals

Protein Separation with ph Gradients Using Composite Buffer Systems Calculated by the Agilent Buffer Advisor Software

AdvanceBio HIC: a Hydrophobic HPLC Column for Monoclonal Antibody (mab) Variant Analysis

Fast and Efficient Peptide Mapping of a Monoclonal Antibody (mab): UHPLC Performance with Superficially Porous Particles

Reproducible LC/MS Peptide Separation Using Agilent AdvanceBio Peptide Plus Columns

A streamlined drug-to-antibody ratio determination workflow for intact and deglycosylated antibody-drug conjugates

A Comprehensive Approach for Monoclonal Antibody N-linked Glycan Analysis from Sample Preparation to Data Analysis

Fast and High-Resolution Reversed-Phase Separation of Synthetic Oligonucleotides

Advanced Characterization of Antibody Drug Conjugates (ADCs) by Liquid Chromatography and Mass Spectrometry (LC/MS) John Gebler, Ph.D.

Application Note. Author. Abstract. Biopharmaceuticals. Verified for Agilent 1260 Infinity II LC Bio-inert System. Sonja Schneider

Technical Overview. Author. Abstract. Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany

Fast and High Resolution Analysis of Intact and Reduced Therapeutic Monoclonal Antibodies (mabs)

Automated Scouting of Stationary and Mobile Phases Using the Agilent 1290 Infinity II Method Development Solution

Application Note. Author. Abstract. Small Molecule Pharmaceuticals. Sonja Krieger Agilent Technologies, Inc. Waldbronn, Germany

Method Transfer from an Agilent 1260 Infinity LC to an Agilent 1260 Infinity II LC

Application Note. Authors. Abstract. Ravindra Gudihal Agilent Technologies India Pvt. Ltd. Bangalore, India

Put the PRO in Protein Characterization

High-throughput and Sensitive Size Exclusion Chromatography (SEC) of Biologics Using Agilent AdvanceBio SEC Columns

Characterization of Glycosylation in the Fc Region of Therapeutic Recombinant Monoclonal Antibody

Application Note. Abstract. Author. Biopharmaceutical

Antibody Analysis by ESI-TOF LC/MS

Minimizing Sample Carryover Using the Multiwash Function of the Agilent 1290 Infinity II Multisampler

mab Titer Analysis with the Agilent Bio-Monolith Protein A Column

Quantification of Host Cell Protein Impurities Using the Agilent 1290 Infinity II LC Coupled with the 6495B Triple Quadrupole LC/MS System

Analysis of Monoclonal Antibodies and Their Fragments by Size Exclusion Chromatography Coupled with an Orbitrap Mass Spectrometer

Characterizing Antibody-Drug Conjugates Using Micro Pillar Array Columns Combined with Mass Spectrometry (µpac -MS)

Robustness of the Agilent Ultivo Triple Quadrupole LC/MS for Routine Analysis in Food Safety

Monoclonal Antibody Charge Variants Identification by Fully Automated Capillary Isoelectric Focusing Mass Spectrometry

An Integrated Workflow for Intact and Subunits of Monoclonal Antibody Accurate Mass Measurements

LC/MS Based Quantitation of Intact Proteins for Bioanalytical Applications

High-Throughput LC/MS Purification of Pharmaceutical Impurities

Characterization of mab aggregation using a Cary 60 UV-Vis Spectrophotometer and the Agilent 1260 Infinity LC system

Ensure your Success with Agilent s Biopharma Workflows

Chromatography column for therapeutic protein analysis

Rapidly Characterize Antibody- Drug Conjugates and Derive Drug-to-Antibody Ratios Using LC/MS

Thermo Scientific MAbPac HIC Columns. Novel Hydrophobic Interaction HPLC Columns. Designed for Monoclonal Antibody Analysis

Advantages of Ion Mobility QTOF for Characterization of BioPharma Molecules

Agilent 1290 Infinity II 2D-LC Solution Biopharmaceutical Polymer Analysis. WCBP Jan 2017 Washington, DC

Developing Purification Strategies for the Agilent 1260 Infinity II Preparative LC/MSD System

Peptide Mapping of Glycoprotein Erythropoietin by HILIC LC/MS and RP-LC/MS

Application Note. Authors. Abstract. Ravindra Gudihal Agilent Technologies India Pvt. Ltd. Bangalore India

Rapid UHPLC Analysis of Reduced Monoclonal Antibodies using an Agilent ZORBAX Rapid Resolution High Definition (RRHD) 300SB-C8 Column

MAbPac RP Column. High-performance reverse phase chromatography column for monoclonal antibody analysis

Towards an in vivo Stability Assay for ADCs and Their Metabolites in Serum by Affinity Capture LC-MS

Technical Overview. Authors. Abstract. Verified for Agilent 1260 Infinity II LC Bio-inert System

Intact & Subunit Purity

Analysis of Intact and C-terminal Digested IgG1 on an Agilent Bio MAb 5 µm Column

Intact and reduced mab/adc fragment separation on reversed phase chromatography with mass spec detection

Size Exclusion Chromatography of Biosimilar and Innovator Insulin Using the Agilent AdvanceBio SEC column

Analysis of Nucleosides Using an Agilent Infinity II High Speed UHPLC with the 6130 Single Quadrupole Mass Selective Detector

Agilent AdvanceBio SEC Columns for Aggregate Analysis: Instrument Compatibility

Antibody-drug Conjugates: Characterization and Control Strategies of Lysine-linked Products

Peptide Mapping: A Quality by Design (QbD) Approach

Complete characterization of a lysine-linked antibody drug conjugate by native LC/MS intact mass analysis and peptide mapping

Heart-cut 2D-LC/MS approach for pharmaceutical impurity identification using an Agilent 6540 Q-TOF LC/MS System

Analysis of Monoclonal Antibody N-glycans by Fluorescence Detection and Robust Mass Selective Detection Using the Agilent LC/MSD XT

Application Note. Biopharma. Authors. Abstract. James Martosella, Phu Duong Agilent Technologies, Inc Centreville Rd Wilmington, DE 19808

Application Note LCMS-87 Automated Acquisition and Analysis of Data for Monitoring Protein Conjugation by LC-MS Using BioPharma Compass

Mobile Phase Optimization in SEC Method Development

Separation of Recombinant Human Erythropoietin (repo) Using Agilent Bio SEC-3

Enabling routine characterization of proteins. Agilent MassHunter BioConfirm software

Analysis of Monoclonal Antibodies and Their Fragments by Size-Exclusion Chromatography Coupled with an Orbitrap Mass Spectrometer

Highly Confident Peptide Mapping of Protein Digests Using Agilent LC/Q TOFs

Improved Sensitivity with the Agilent 1290 Infinity Evaporative Light Scattering Detector (ELSD)

Columns for Biomolecules BioLC Column Lines

Monitoring of Mammalian Cell Culture Media with HILIC LC/MS

NISTmAb characterization with a high-performance RP chromatography column

Reversed-phase Separation of Intact Monoclonal Antibodies Using Agilent ZORBAX Rapid Resolution High Definition 300SB-C8 1.

Chromatographic Workflows for Biopharmaceutical Characterization

Quantification of genotoxic "Impurity D" in Atenolol by LC/ESI/MS/MS with Agilent 1200 Series RRLC and 6410B Triple Quadrupole LC/MS

2012 Waters Corporation 1

Transcription:

Size Exclusion Chromatography/ Mass Spectrometry Analysis of Antibody Drug Conjugates Using the Agilent 126 Infinity II Bio-Inert LC Application Note Biologics & Biosimilars Author Sonja Schneider Agilent Technologies, Inc. Waldbronn, Germany Abstract Recombinant monoclonal antibodies (mabs) have emerged as important therapeutic agents for the treatment of different diseases 1. These biotherapeutics have been extended by the development of antibody drug conjugates (ADCs). ADCs are mabs to which cytotoxic payloads are covalently attached, enabling delivery of potent drugs to a specific target. This Application Note describes the analysis by size exclusion chromatography with online mass spectrometry (SEC/MS) of an mab (Trastuzumab) and an ADC (Trastuzumab emtansine) using the Agilent 126 Infinity II Bio-inert LC with quadrupole time-of-flight (Q-TOF) mass spectrometry (MS) detection. SEC is a straightforward technique of introducing mabs and ADCs into the LC/MS system, requiring little method optimization. From the MS data collected for the ADC, it was possible to calculate a drug-to-antibody ratio (DAR) of 3.1 using the Agilent DAR calculation software. We demonstrate that the 126 Infinity II Bio-inert LC is an optimal front end for LC/MS analysis of biopharmaceuticals. 1 4 DAR 2 147,87 DAR 3 148,45 1.1 1..9.8.7 DAR 1 146,13 DAR 4 149,2.6.5.4 DAR 5 149,962.3.2.1 DAR 145,172 DAR 6 15,917 DAR 7 151,873 DAR 8 152,844 141, 143, 145, 147, 149, 151, 153, 155, Deconvoluted mass (Da)

Introduction Antibody drug conjugates (ADCs) are recombinant monoclonal antibodies (mabs) conjugated to cytotoxic drug molecules using various available linker chemistries 2. The combination of highly potent drugs with antibodies targeted to specific antigens is expected to enhance the therapeutic action by the sensitive discrimination between healthy and diseased tissue 3. The advantage of ADCs as therapeutic agents is the increased control of the pharmacokinetics of the drug 4. The delivery to the target diseased tissue is improved, while the exposure of healthy tissues to toxic drugs is minimized, leading to selective elimination of tumor cells expressing the target antigen 2. Figure 1 shows the structure of the ADC Trastuzumab emtansine (T DM1, Kadcyla). It consists of three components: mab (Trastuzumab) A bifunctional linker ([N-maleimidomethyl]cyclohexane- 1-carboxylate, MCC) Small molecule drug (DM1, derivative of maytansine) T-DM1 is typically administered for the treatment of human epidermal growth factor receptor 2 (HER2)-positive breast cancer 5. The mab part of T-DM1 binds selectively to HER2-positive cells, leading to internalization of the ADC antigen complex into the cell. Inside the cell, the complex is degraded, releasing the cytotoxic drug, which destroys the cancer cell 6. Emtansine is the combination of the cytotoxic maytansinoid DM1 (which inhibits tubulin polymerization to induce cell-cycle arrest and cell death) and a bifunctional MCC linker. MCC is chemically linked to the antibody through lysine residues on the one side and bound to DM1 by nonreducible thioether linkages on the other side 7. The resulting ADC is typically highly heterogeneous regarding both the distribution and the loading of the cytotoxic drug species on DM1 Figure 1 - Structure of Trastuzumab emtansine. the antibody. Therefore, physicochemical characterization of ADCs is more complex, compared to the analysis of the corresponding mab 8. Two types of ADC are currently available on the market; lysine-linked ADCs (such as T-DM1) and cysteine-linked ADCs 9. Some cysteine conjugates can be analyzed using hydrophobic interaction chromatography (HIC) to determine the drug-to-antibody ratio (DAR). The analysis of lysine conjugates, however, is not feasible using HIC. Due to the higher heterogeneity, UV-Vis spectroscopy and MS are generally the methods of choice to determine the DAR. In contrast to cysteine conjugates, lysine-conjugated ADCs do not dissociate under acidic or high organic conditions (such as those commonly used in MS) due to their intact interchain disulfides. The analysis of ADCs using reversed-phase (RP) liquid chromatography (LC) coupled to MS detection can be problematic due to irreversible binding of the proteins to the stationary phase 8,1. Elevated column temperatures (6 to 8 C) and propanol containing buffers can reduce these effects. However, these conditions might lead to problems regarding temperature-sensitive proteins. SEC is usually carried out at room temperature or slightly above without the need for high temperatures. Thioether linker Light chain This Application Note describes the SEC/MS analysis of T-DM1 as well as Trastuzumab using an Agilent 126 Infinity II Bio-inert LC coupled to quadrupole time-of-flight (Q-TOF) mass spectrometry (MS). Experimental Instrumentation The SEC-MS experiments were carried out on a 126 Infinity II Bio-inert LC with online Q-TOF detection comprising the following modules: Agilent 126 Infinity II Bio-inert LC Agilent 126 Infinity II Bio-inert Pump (G5654A) Agilent 126 Infinity II Bio-inert Multisampler (G5668A) Agilent 126 Infinity II Multi Column Thermostat (G7116A) with bio-inert heat exchanger (Option #19) Agilent 126 Infinity II Diode Array Detector WR (G7115A) with bio inert flow cell (Option #28) MS System CL NH Heavy chain CH 1 Asn-297-- CH 2 CH 3 Trastuzumab Agilent 655 ifunnel quadrupole time of flight LC/MS 2

Column Agilent Bio SEC-3, 3 Å, 7.8 3 mm, 3 µm (p/n 519-2511) Software Agilent MassHunter Workstation software with Bioconfirm software, version B.8.. The DAR was calculated using Agilent DAR Calculator software, version B.1., build 1.15.2.. Samples T-DM1 (Kadcyla) and Trastuzumab (Herceptin) Chemicals All solvents used were LC/MS grade. LC/MS grade acetonitrile and formic acid (FA) were purchased from Sigma-Aldrich, St. Louis, Missouri, US. N-glycanase and MS-friendly deglycosylation (Tris) Buffer was purchased from Prozyme, Hayward, CA, USA). Fresh ultrapure water was obtained from a Milli-Q Integral system equipped with LC-Pak Polisher and a.22 µm membrane point-of-use cartridge (Millipak). Sample preparation Both samples (mab and ADC) were deglycosylated by adding 25 mu of N-glycanase to 1 µl of 1 mg/ml protein solution (in Tris), and incubated overnight at 37 C. Table 1. Chromatographic and MS conditions. Parameter Mobile phase Flow rate Stop time Needle wash mode Injection volume 5 µl Condition Column temperature 3 C Diode array detection MS detection Acquisition mode MS1 Minimum range Maximum range Ion polarity Source parameters 49 % Water, 5 % ACN, 1 % FA.5 ml/min 25 minutes Standard wash 28 nm/ 4 nm, Ref.: OFF >.25 minutes (.5 seconds response time) (1 Hz) 1 m/z 7, m/z Positive Gas temperature 29 C Gas flow Nebulizer 14 L/min 2 psig Sheath gas temperature 4 C Sheath gas flow Scan source parameters Vcap Nozzle voltage Fragmentor Skimmer 1 12 L/min 5, V 2, V 5 V Octopole RF peak 75 Reference masses 922.98, 1,821.9523 Column preparation Agilent Bio SEC-3 columns are shipped in.1 M sodium phosphate buffer, ph 7.. Sodium phosphate buffer is not compatible with MS analyses as it is a nonvolatile salt. The column was extensively flushed with mobile phase for approximately 48 hours before LC/MS analysis. 3

Results and Discussion Typically, SEC of mabs is carried out in a physiological buffer such as phosphate buffered saline (PBS) with a ph of approximately 7.4. The buffer maintains the native structure of the protein and reduces undesired binding of the proteins to the column. However, PBS is incompatible with MS analysis. Instead, for the SEC/MS analysis, a 126 Infinity II Bio inert LC was used together with a solvent mixture of 5 % ACN, 49 % water, and 1 % FA. Although a denaturing solvent was used for the analysis, a highly reproducible analysis was possible, resulting in minimal relative standard deviations of.12 % for the retention time and.486 % for area precision (Figure 2). The SEC/MS analyses of intact and deglycosylated Trastuzumab are displayed as deconvoluted spectra in Figure 3. The insets show the Gaussian distribution of the charge state envelope for the mabs. The spectra were deconvoluted using the maximum entropy deconvolution algorithm in Agilent MassHunter BioConfirm software. The deconvoluted spectrum of the intact Trastuzumab (Figure 3A) shows multiple mass peaks corresponding to different glycoforms of the mab. As a comparison, Figure 3B shows the deconvoluted spectrum of the deglycosylated Trastuzumab, observed as a single main peak with a mass of 145,173 Da. Using SEC before MS analysis, adduct peaks (which would arise from buffer salts when performing static infusion experiments) are minimized 7. Mass spectra of ADCs are even more complex due to the combination of both glycosylation and drug loading heterogeneities, which pose a challenge for interpretation (Figure 4). Therefore, the ADC was deglycosylated prior to further SEC/MS analysis to reduce the spectrum complexity. Response units 1 1 4. 3.8 3.6 3.4 3.2 3. 2.8 2.6 2.4 2.2 2. 1.8 1..8.6.4.2 Peak RSD RT (%) RSD area (%) Main.12.486 6.4 6.8 7.2 7.6 8. 8.4 8.8 9.2 9.6 1. 1.8 1 1 12. 12.4 12.8 Acquisition time (min) Figure 2. Overlay of seven consecutive SEC/UV runs of T-DM1 at 28 nm. 4

Figure 3. A) Deconvoluted spectrum of intact Trastuzumab. B) Deconvoluted spectrum of deglycosylated Trastuzumab. The insets show the raw mass spectra of the analyte. 1 3 2.7 1 3 1.8 151,97.18 15,139.84 1..8 152,55.41.6.4.2 1 4 6. A 148,224.51 5.5 5. 4.5 4. 3.5 3. 2.5 2. 1.5 1. 147,916..5 148,78.69 146,777.85 147,772.36 148,836.82 149,134.81 146,8 147, 147,2 147,4 147,6 147,8 148, 148,2 148,4 148,6 148,8 149, 149,2 149,4 149,6 149,8 15, 1 5 Deconvoluted mass (amu) 1.9 1 4 1.8 B 3.2 2,963.7774 145,172.7 2.8 1.7 2.4 2. 1.5 1.3 3,377.186.8 1,821.9511 3,457.5172.4 1.1 1. 1,6 1,9 2,2 2,5 2,8 3,1 3,4 3,7 4, 4,3 4,6.9.8.7.6.5.4.3 145,334.58.2.1 145,492.77 142,5 143, 143,5 144, 144,5 145, 145,5 146, 146,5 147, 147,5 148, 148,5 Deconvoluted mass (amu) 2.6 2.5 2.4 2.3 2.2 2.1 2. 1.9 1.8 1.7 1.5 1.3 1.1 1..9.8.7.6.5.4.3.2.1 148,62.76 149,181 148,385.66 148,547.25 Figure 4. Deconvoluted spectrum of T-DM1. The inset shows the raw mass spectra of the analyte. 1 4 3,26.245 1..8.6 1,821.9517.4.2 1,6 1,9 2,2 2,5 2,8 3,1 3,4 3,7 4, 4,3 4,6 153,16.39 3,23.835 3,148.9445 2,5 2,25 2,45 2,65 2,85 3,5 3,25 3,45 3,65 3,85 153,971.78 148,63.47 154,93.32 155,855.87 143, 144, 145, 146, 147, 148, 149, 15, 151, 152, 153, 154, 155, 156, 157, 158, Deconvoluted mass (amu) 158,134.25 5

The SEC/MS analysis of the deglycosylated T-DM1 is shown in Figure 5A (raw mass spectrum) and Figure 5B (deconvoluted spectrum). The spectrum was also deconvoluted using the maximum entropy algorithm in MassHunter BioConfirm software. The spectrum of the deglycosylated ADC is much easier to interpret, and the resulting deconvolution (Figure 5B) is considerably simpler. The ADC spectrum is composed of nine equally spaced groups of peaks, 1 3 1.8 where the masses of the most intense peaks are separated by 958 Da, on average. This mass corresponds to the mass of one DM1 drug (738.3 Da) and one MCC linker (219.7 Da), representing the drug load species11. The first peak in the deconvoluted ADC spectrum matches the mass of the deglycosylated Trastuzamab in Figure 3B (145173 Da). Therefore, D (Trastuzumab with zero drug load) can be assigned to the first peak of the spectrum. A 2,885.619 The following peaks represent species with DARs ranging from to 8. A second minor series of peaks, marked with an asterisk, is separated from the main peaks with a mass approximately 221.7 Da higher. These peaks can be assigned to species with an attached linker but without conjugated DM1, as shown by Marcoux et al.9. 3,22.2872 3,85.3888 1..8.6.4.2 2,5 2,15 2,25 2,35 2,45 2,55 2,65 2,75 2,85 2,95 3,5 3,15 3,25 3,35 3,45 3,55 3,65 3,75 3,85 3,95 1 3 7 B D3 D2 6 D4 5 D1 4 D5 3 2 1 D6 D D7 D8 141,5 142,5 143,5 144,5 145,5 146,5 147,5 148,5 149,5 15,5 151,5 152,5 153,5 Figure 5. SEC-MS analysis of the deglycosylated T-DM1 A) raw mass spectrum and B) deconvoluted spectrum representing DAR species D to D8 plus satellite peaks (marked with an asterisk) representing Trastuzumab + linker signals. 6

The deconvoluted spectra were then exported as a.csv file, and imported into the Agilent DAR calculator. After inputting/selecting D mass and drug/linker mass, the DAR calculator automatically selects, annotates, and integrates mass peak groups of ADCs with various drug loadings. It then calculates the average DAR, and generates a peak list table 11. The average DAR was calculated as approximately 3.1, matching previously determined values for this molecule 12. Figure 6 shows a representative graph of the DAR calculation of the deglycosylated ADC. Figure 7 displays the DAR peak list ranging from DAR to DAR 8 together with the masses, area, and area % from which the average DAR of 3.1 was calculated. Conclusions SEC/MS using MS-compatible buffer (solvent mix of 5 % ACN, 49 % water, and 1 % FA) was performed for the analysis of Trastuzumab and T-DM1 on the Agilent 126 Infinity II Bio-inert LC with Q-TOF detection. SEC prior to MS was useful because mabs and ADCs can easily be desalted before introduction into the MS ion source. The spectrum of Trastuzumab revealed multiple mass peaks corresponding to different glycoforms of the mab. The spectrum of T-DM1 revealed a highly complex peak pattern due to both glycosylation and drug loading heterogeneities. Deglycosylation of Trastuzumab and T-DM1 permitted a more facile interpretation of the data. The deconvolution of the deglycosylated T-DM1 revealed nine equally spaced peaks that were assigned to nine versions of Trastuzumab with zero to eight molecules of DM1 attached. The average DAR of 3.1 was calculated using the Agilent DAR calculator. 1 4 1.1 1..9.8.7.6.5.4.3.2.1 DAR 145,172 DAR 1 146,13 DAR 2 DAR 3 147,87 148,45 DAR 4 149,2 DAR 5 149,962 DAR 6 15,917 DAR 8 DAR 7 152,844 151,873 141, 143, 145, 147, 149, 151, 153, 155, Deconvoluted mass (Da) Figure 6. DAR calculation of the deglycosylated T-DM1 with the Agilent DAR calculator, resulting in an average DAR of 3.1. Figure 7. DAR Peak List, screenshot from the Agilent DAR calculator. 7

References 1. Sandra, K.; Vandenheede, I.; Sandra, P. Modern chromatographic and mass spectrometric techniques for protein biopharmaceutical characterization. J. Chromatogr. A 214, 1335, 81 13. 2. McCombs, J. R.; Owen, S. C. Antibody drug conjugates: design and selection of linker, payload and conjugation chemistry. Journal of American Association of Pharmaceutical Scientists 215, 17(2), 339 351. 3. Firer, A.; Gellerman, G. Targeted drug delivery for cancer therapy: the other side of antibodies. Journal of Hematology and Oncology 212, 5:7. 4. Rohrer, T. Antibody drug conjugates: Potent weapons for the oncology arsenal. Chemistry Today 29, 27(5), 56 6. 5. Koen, S.; et al. Multiple heart-cutting and comprehensive two-dimensional liquid chromatography hyphenated to mass spectrometry for the characterization of the antibody-drug conjugate ado-trastuzumab emtansine. J. Chromatogr. B 216. 6. Panowski, S. Site-specific antibody drug conjugates for cancer therapy. mabs 214, 6:1, 34 45. 7. Krop, I.; et al. Phase I Study of Trastuzumab-DM1, an HER2 Antibody Drug Conjugate, Given Every 3 Weeks to Patients With HER2 Positive Metastatic Breast Cancer. American Society of Clinical Oncology 21, Volume 28, Number 16, 2689 274. 8. Wakankar, A.; et al. Analytical methods for physiochemical characterization of antibody drug conjugates. Landes BioScience, mabs 211, 3:2; 161 172. 9. Marcoux, J. Native mass spectrometry and ion mobility characterization of Trastuzumab emtansine, a lysine linked antibody drug conjugate. Protein Science 215, Volume 24, 121 1223. 1. Lazar, A. C.; et al. Analysis of the composition of immunoconjugates using size-exclusion chromatography coupled to mass spectrometry, Rapid Commun. Mass Spectrom. 25, 19, 186 1814. 11. Chen, J.; Murphy, S. Drug-to Antibody Ratio (DAR) Calculation of Antibody-Drug Conjugates (ADCs), Agilent Technologies Application Note, publication number 5991-6263EN, 215. 12. Kim, M. T.; et al. Statistical Modeling of the Drug Load Distribution on Trastuzumab Emtansine (Kadcyla), a Lysine-Linked Antibody Drug Conjugate. Bioconjugate chemistry 214, 25, 1223 1232. www.agilent.com/chem For Research Use Only. Not for use in diagnostic procedures. This information is subject to change without notice. Agilent Technologies, Inc., 216 Published in the USA, November 1, 216 5991-7553EN