HIC as a Complementary, Confirmatory Tool to SEC for the Analysis of mab Aggregates

Similar documents
Analysis of Monoclonal Antibody and Related Substances using a New Hydrophobic Interaction Chromatography (HIC) Column

Thermo Scientific MAbPac HIC Columns. Novel Hydrophobic Interaction HPLC Columns. Designed for Monoclonal Antibody Analysis

High-Resolution Charge Variant Analysis for Top-Selling Monoclonal Antibody Therapeutics Using a Linear ph Gradient Separation Platform

High-Resolution Separation of Intact Monoclonal Antibody Isoforms

Lifetime stability of size exclusion chromatography columns for protein aggregate analysis

A Fast and Robust Linear ph Gradient Separation Platform for Monoclonal Antibody (MAb) Charge Variant Analysis

ph gradient analysis of IgG1 therapeutic monoclonal antibodies using a 5 µm WCX column

A Novel ph Gradient Separation Platform for Monoclonal Antibody (MAb) Charge-Variant Analysis

Thermo Scientific. MAbPac HIC-10. Product Manual. P/N: June Part of Thermo Fisher Scientific

Monoclonal Antibody Analysis on a Reversed-Phase C4 Polymer Monolith Column

A universal chromatography method for aggregate analysis of monoclonal antibodies

Analysis of Monoclonal Antibodies and Their Fragments by Size Exclusion Chromatography Coupled with an Orbitrap Mass Spectrometer

mab and ADC Analysis Shanhua Lin, Ph.D. The world leader in serving science

Application Note. Authors. Abstract. Biopharmaceuticals

Separate and Quantify Rituximab Aggregates and Fragments with High-Resolution SEC

MAbPac RP Column. High-performance reverse phase chromatography column for monoclonal antibody analysis

mabs and ADCs analysis by RP

NISTmAb characterization with a high-performance RP chromatography column

High-speed, High-resolution Oligonucleotide Separations Using Small Particle Anion-Exchangers

Analysis of Monoclonal Antibodies and Their Fragments by Size-Exclusion Chromatography Coupled with an Orbitrap Mass Spectrometer

Development of Ultra-fast ph-gradient Ion Exchange Chromatography for the Separation of Monoclonal Antibody Charge Variants

Separation of Intact Monoclonal Antibody Sialylation Isoforms by ph Gradient Ion-Exchange Chromatography

Characterization of mab aggregation using a Cary 60 UV-Vis Spectrophotometer and the Agilent 1260 Infinity LC system

High-resolution Analysis of Charge Heterogeneity in Monoclonal Antibodies Using ph-gradient Cation Exchange Chromatography

A Comprehensive Workflow to Optimize and Execute Protein Aggregate Studies

Analysis of Intact and C-terminal Digested IgG1 on an Agilent Bio MAb 5 µm Column

ProPac Elite WCX, 5 μm Particle, for Fast, High Resolution Protein and mab Analysis. July 2018

High-throughput and Sensitive Size Exclusion Chromatography (SEC) of Biologics Using Agilent AdvanceBio SEC Columns

Simple charge variant profile comparison of an innovator monoclonal antibody and a biosimilar candidate

Chromatography column for therapeutic protein analysis

How to Create Your Own Affinity Titer Column. Kelly J. Flook*, Yury Agroskin and Chris Pohl Thermo Fisher Scientific, Sunnyvale, CA, USA

AdvanceBio HIC: a Hydrophobic HPLC Column for Monoclonal Antibody (mab) Variant Analysis

Columns for Biomolecules BioLC Column Lines

Concentration of Human Hormones in Drinking Water Using Solid Phase Extraction and Analysis by High Performance Liquid Chromatography

Mobile Phase Optimization in SEC Method Development

Improving Retention Time Precision and Chromatography of Early Eluting Peptides with Acetonitrile/Water Blends as Solvent B

Size Exclusion BioHPLC Columns

Agilent AdvanceBio SEC Columns for Aggregate Analysis: Instrument Compatibility

CX-1 ph Gradient Buffer

Molecular Characterization of Biotherapeutics The Agilent 1260 Infi nity Multi-Detector Bio-SEC Solution with Advanced Light Scattering Detection

Phenyl Membrane Adsorber for Bioprocessing

Charge Heterogeneity Analysis of Rituximab Innovator and Biosimilar mabs

Chromatographic Workflows for Biopharmaceutical Characterization

High Throughput Sub-4 Minute Separation of Antibodies using Size Exclusion Chromatography

columns P r o P a c H I C C o l u m n S o l u t i o n s f o r P r o t e i n A n a l y s i s

R R Innovation Way P/N SECKIT-7830 Newark, DE 19711, USA Tel: Fax: Website: Published in November 2013

Determination of Inorganic Anions in Acid Rain Using a Dedicated High-Pressure Capillary Ion Chromatography System

Determination of Manganese in Brine ( 30% NaCl)

Analysis of Fracking Flowback Water From the Marcellus Shale Using In-line Conductivity, Automated Dilution, and Ion Chromatography

Separation of Monoclonal Antibodies Using TSKgel HPLC Columns

Implementation of a Walk-Up High- Pressure Capillary Ion Chromatograph for the Fast Separation of Pharmaceutical Relevant Inorganic Anions and Cations

Application Note. Author. Abstract. Biopharmaceuticals. Verified for Agilent 1260 Infinity II LC Bio-inert System. Sonja Schneider

High-Throughput, High-Resolution Monoclonal Antibody Analysis with Small Particle Size HPLC Columns

Developing Robust and Efficient IEX Methods for Charge Variant Analysis of Biotherapeutics Using ACQUITY UPLC H-Class System and Auto Blend Plus

User Guide. SMART Digest Kit

Reducing Cycle Time for Charge Variant Analysis of Monoclonal Antibodies

Separation of Recombinant Human Erythropoietin (repo) Using Agilent Bio SEC-3

Multiple Detector Approaches to Protein Aggregation by SEC

Gentle Bioanalysis of Proteins

Phenyl Membrane Adsorber for Bioprocessing

TSK-GEL BioAssist Series Ion Exchange Columns

Protein-Pak Hi Res HIC Column and HIC Protein Standard

An Automated System for At-Line Process Analysis of Biopharmaceutical Fermentation Reactions

Determination of Trace Concentrations of Oxyhalides and Bromide in Municipal and Bottled Waters Using a Compact Ion Chromatography System

Fast mass transfer Fast separations High throughput and improved productivity Long column lifetime Outstanding reproducibility Low carryover

ION EXCHANGE KIT FOR MAB SEPARATIONS

MAbPac SCX-10 Columns

Clone Selection Using the Agilent 1290 Infinity Online 2D-LC/MS Solution

Higher Order mab Aggregate Analysis using New Innovative SEC Technology

UHPLC for Large Bio-Therapeutics

Parallel LC with Capillary PS-DVB Monolithic Columns for High-Throughput Proteomics

Using the NISTmAb reference standard to demonstrate a simple approach to charge variant analysis

Fast protein separations with ion exchange columns

Application of Agilent AdvanceBio Desalting-RP Cartridges for LC/MS Analysis of mabs A One- and Two-dimensional LC/MS Study

Determination of Oxalate and Other Anions in Bayer Liquor Using Ion Chromatography

High-Resolution Analysis and Purification of Oligonucleotides with the DNAPac PA-100 Column

Seamless Method Transfer from an Agilent 1260 Infinity Bio-inert LC to an Agilent 1260 Infinity II Bio-inert LC

Thermo Scientific Solutions for Intact-Protein Analysis. Better, Faster Decisions for. Biotherapeutic Development

Thermo Scientific BioLC Columns. Monoclonal antibody. characterization

Intact and reduced mab/adc fragment separation on reversed phase chromatography with mass spec detection

Analysis of biomolecules by SEC and Ion-Exchange UPLC

Fast and High Resolution Analysis of Intact and Reduced Therapeutic Monoclonal Antibodies (mabs)

columnsi o n P a c C S 1 6 C a t i o n - E x c h a n g e C o l u m n

Introduction. Yan Liu, Victor Barreto, Rosanne Slingsby, Chris Pohl Thermo Fisher Scientific, Sunnyvale, CA, USA

for Acclaim Carbamate Column

Column for High Performance,High-Binding Capacity Ion Exchange Chromatography:TSKgel SuperQ-5PW and Its Applications

Thermo Scientific Dionex ICS-900 Ion Chromatography System

mab Titer Analysis with the Agilent Bio-Monolith Protein A Column

BioHPLC columns. Tim Rice Biocolumn Technical Specialist

In-Cell and External Hydrolysis of Biomass for Carbohydrate Determinations Using Automated Solvent Extraction

Size Exclusion BioHPLC columns Ion Exchange BioHPLC columns

ProSEC 300S. Protein Characterization columns

Monoclonal Antibody Characterization Achieving Higher Throughput and Productivity

Effect of Temperature on IC Signal Noise

Separation Science Built for Biopharma

Developing Quantitative UPLC Assays with UV

Determination of additives and byproducts in an acid copper plating bath by liquid chromatography

An effective platform for purification of IgM monoclonal antibodies using Hydroxyapatite

Transcription:

HIC as a Complementary, Confirmatory Tool to SEC for the Analysis of mab Aggregates Julia Baek, Shanhua Lin, Xiaodong Liu, Thermo Fisher Scientific, Sunnyvale, CA Application Note 2126 Key Words Size exclusion chromatography, SEC, hydrophobic interaction chromatography, HIC, mab aggregate, MAbPac SEC-1, MAbPac HIC-1 Goal To demonstrate the analysis of monoclonal antibody (mab) aggregates using the Thermo Scientific MAbPac SEC-1 and MAbPac HIC-1 columns. The MAbPac SEC-1 column provides fast and high-resolution separation of mab aggregates from the monomer. The MAbPac HIC-1 column provides excellent separation of mab aggregates from the monomer and the ability to resolve mab variants. Introduction Monoclonal antibodies and related products are widely used as therapeutic agents to treat various cancers and autoimmune diseases. The production of mab therapeutics can be highly challenging due to various chemical degradation processes and aggregation. MAb aggregates may form during protein expression, downstream processing, and storage, and may be caused by high concentration, elevated temperature, abrupt change in ph, shear strain, freeze-thaw cycle, or surface adsorption. Aggregates in mab therapeutics can result in an undesired immune response that affects the safety and the potency of the drug due to incorrect dosage. 1 Therefore, monitoring mab aggregation is important for their production and quality assurance. Size-exclusion liquid chromatography (SEC) has been the method of choice for the detection and quantification of mab aggregates. However, the use of both standard hydrophobic interaction chromatography (HIC) and mixed-mode HIC for separation of protein aggregates has been reported. 2-4 HIC separates proteins based on the hydrophobicity in the native state and can often detect changes in protein structure as well as aggregates. 5 The MAbPac SEC-1 column is based on high-purity, spherical, porous (3 Å), 5 µm silica particles covalently modified with a proprietary diol hydrophilic layer, designed for mab aggregate analysis. The MAbPac HIC-1 column is a wide-pore (1 Å), 5 µm silicabased HIC column with polyamide functionality, providing desired selectivity for intact mab and mab variants. Both columns can be used under non-denatured conditions to preserve the bioactivity of mabs/proteins. HIC can be used as a secondary assay to confirm aggregation or for removal of aggregates during purification. Here, we demonstrate the use of MAbPac SEC-1 and MAbPac HIC-1 columns for high-resolution mab aggregate analysis.

2 Experimental Chemicals and Reagents Deionized (DI) water, 18.2 MΩ-cm resistivity Fisher Scientific Isopropanol (P/N A461-4) Sodium phosphate monobasic monohydrate (NaH 2 O, 98.%) Ammonium sulfate [(NH 4 ) 2 SO 4, 99.%] Sample Handling Equipment Thermo Scientific vial and closure kit (P/N 55428) Sample Preparation The mab sample was provided by a customer. HIC: The 8 mg/ml sample was diluted two-fold with mobile phase A. SEC: The 8 mg/ml sample was directly injected onto the column. Separation Conditions Instrumentation Thermo Scientific Dionex UltiMate 3 BioRSLC system equipped with: SR-3 Solvent Rack (without degasser) (P/N 535.92) LPG-34RS Biocompatible Quaternary Rapid Separation Pump (P/N 54.36) WPS-3TBRS Biocompatible Rapid Separation Thermostatted Autosampler (P/N 5841.2) TCC-3RS Rapid Separation Thermostatted Column Compartment (P/N 573.) VWD-34RS Rapid Separation Variable Wavelength Detector (P/N 574.1) equipped with a micro flow cell Columns MAbPac HIC-1, 4.6 1 mm (P/N 8848) MAbPac SEC-1, 4 15 mm (P/N 75592) SEC Mobile phase 5 mm sodium phosphate, 3 mm NaCl, ph 6.8 Preparation of mobile Dissolve 6.9 g of sodium phosphate monobasic, phase monohydrate (NaH 2 O) and 17.5 g of sodium chloride in 9 ml of DI water, adjust the ph to 6.8 with 5% NaOH solution and bring the volume to 1 ml with DI water. Filter the mobile phase through a.22 µm filter. Flow rate.2 ml/min Column temperature 3 C UV detector 28 nm wavelength HIC Mobile phase A 2 M ammonium sulfate, 1 mm sodium phosphate, ph 7. Mobile phase B 1 mm sodium phosphate, ph 7. Preparation of Dissolve 13.8 g of sodium phosphate mobile phase A monobasic, monohydrate (NaH 2 O) and 264.2 g of ammonium sulfate in 8 ml of DI water, adjust the ph to 7. with 5% NaOH solution and bring the volume to 1 ml with DI water. Filter the mobile phase through a.22 µm filter. Preparation of Dissolve 13.8 g of sodium phosphate mobile phase B monobasic, monohydrate (NaH 2 O) in 9 ml of DI water, adjust the ph to 7. with 5% NaOH solution and bring the volume to 1 ml with DI water. Filter the mobile phase through a.22 µm filter. Gradient Time (min) %A %B -5. 6 4. 6 4 1. 6 4 29. 1 34. Flow rate.5 or 1. ml/min Column temperature 3 C UV detector 28 nm wavelength Software Thermo Scientific Dionex Chromeleon 6.8 Chromatography Data System

Results and Discussion The mab aggregates were separated within 8 minutes on a MAbPac SEC column (Figure 1); two aggregate peaks and a monomer peak were observed. The first aggregate peak was confirmed as a dimer using the Thermo Scientific Exactive Plus EMR mass spectrometer. 6 7 6 Monomer 22 2 15 1 mab variants Monomer 3 5 4 3 2 1 Aggregate 1 2 2 4 6 8 1 12 Column: MAbPac SEC-1, 5 µm Format: 4. 15 mm Mobile Phase A: 5 mm sodium phosphate,3 mm NaCl, ph 6.8 Flow Rate:.2 ml/min Inj. Volume: 1 µl (8 mg/ml) Temp.: 3 ºC Detection: UV (28 nm) Figure 1. Separation of mab aggregates using MAbPac SEC-1 column. Figure 2 shows the separation of two mab aggregate peaks from the monomer form using the MAbPac HIC column. In HIC, aggregates typically elute later than the monomer due to the exposed hydrophobic groups in the denatured chains of aggregate. 7 The intensity ratios of the aggregate peaks to the monomer peak were consistent with the MAbPac SEC-1 result. The first aggregate peak was broad with obvious additional peaks, most likely due to different conformations of the mab aggregate. 5 7 Aggregate 1 2 1 15 2 25 Column: MAbPac HIC-1, 5 µm Format: 4.6 1 mm Mobile Phase A: 2 M ammonium sulfate, 1 mm sodium phosphate, ph 7. Mobile Phase B: 1 mm sodium phosphate, ph 7. Gradient: Time (min) %A %B -5. 6 4. 6 4 1. 6 4 29. 1 34. 1 Flow Rate:.5 ml/min Inj. Volume: 1 µl (4 mg/ml) Temp.: 3 ºC Detection: UV (28 nm) Figure 2. Separation of mab aggregates and hydrophilic variants using MAbPac HIC-1 column. The SEC column separates by size and gives one peak for a mab dimer, whereas the HIC column separates by hydrophobicity and resolves the different isomeric forms of the dimer. This gives additional information of the aggregate forms present which you cannot get from SEC. Aggregates observed on the MAbPac HIC column were collected and re-injected onto the MAbPac SEC-1 column (Figure 3). The retention time of the collected fraction was compared to the retention time of the mab aggregate peak. The retention times of the monomer and the first aggregate peak fractions matched the retention time of the chromatogram obtained by directly injecting the mab sample onto the MAbPac SEC-1 column.

4 7 a 22 a 14 14 b c 1 22 1 b 2 4 6 8 1 12 1 15 2 25 Column: MAbPac SEC-1, 5 µm Format: 4. 15 mm Mobile Phase A: 5 mm sodium phosphate,3 mm NaCl, ph 6.8 Flow Rate:.2 ml/min Inj. Volume: (a) 1 µl (8 mg/ml) (b) 2 µl (c) 2 µl Temp.: 3 ºC Detection: UV (28 nm) Sample: (a) mab (b) Monomer fraction from MAbPac HIC-1 run (c) Aggregate fraction from MAbPac HIC-1 run Figure 3. Separation of mab fractions obtained from MAbPac HIC- 1 column on the MAbPac SEC-1 column. In addition to the aggregate peaks, the MAbPac HIC-1 column was able to separate two hydrophilic mab variant peaks. Several chromatographic conditions were investigated including mobile phase, temperature, and flow rate. Separation of both mab aggregate and hydrophilic variants was best achieved using a 1.2 M to M ammonium sulfate gradient in 28 minutes at.5 ml/min flow rate. Addition of organic solvent to the mobile phases caused the hydrophilic variant peaks to disappear while the aggregate peaks were still separated. Lower flow rates (.5 ml/min) with longer gradient time gave better resolution of the aggregates and hydrophilic variants compared to 1 ml/min with shorter gradient time (Figure 4). Column: MAbPac HIC-1, 5 µm Format: 4.6 1 mm Mobile Phase A: 2 M ammonium sulfate, 1 mm sodium phosphate, ph 7. Mobile Phase B: 1 mm sodium phosphate, ph 7. Gradient (a) Time (min) %A %B -5. 6 4. 6 4 1. 6 4 15. 1 2. 1 Gradient (b) Time (min) %A %B -5. 6 4. 6 4 1. 6 4 29. 1 34. 1 Flow Rate: (a) 1. ml/min (b).5 ml/min Inj. Volume: 1 µl (4 mg/ml) Temp.: 3 ºC Detection: UV (28 nm) Figure 4. Separation of mab aggregates using MAbPac HIC-1 column with different flow rates.

Conclusion MAbPac SEC-1 column provides fast and highresolution separation of mab and its aggregates. MAbPac HIC-1 column provides excellent separation of mab and its aggregates and can be used as a secondary method for mab aggregate analysis. MAbPac HIC-1 column can resolve mab variants and mab aggregates within the same analysis. References 1. Vázquez-Rey, M.; Lang, D.A.: Aggregates in monoclonal antibody manufacturing processes, Biotechnol Bioeng, 211, 18, 1494 158. 2. Chen, J.; Tetrault, J.; Ley, A. Comparison of standard and new generation hydrophobic interaction chromatography resins in the monoclonal antibody purification process, J. Chromatogr. A., 28, 1177, 272-281. 3. Lu Y.; Williamson, B.; Gillespie, R. Recent advancement in application of hydrophobic interaction chromatography for aggregate removal in industrial purification process, Curr. Pharm. Biotechnol., 29, 1, 427-433. 4. Gagnon, P. IgG aggregate removal by chargedhydrophobic mixed mode chromatography, Curr. Pharm. Biotechnol., 29, 1, 434-439. 5. Mahn, A. Hydrophobic interaction chromatography: fundamentals and applications in biomedical engineering, Biomedical Science, Engineering and Technology, Prof. Dhanjoo N. Ghista Ed. (212) 63-623. 6. Lin, S.; Wang, H.; Hao, Z.; Horn, D.; Tracy, M.; Liu, X. Analysis of Monoclonal Antibodies, Aggregates, and Their Fragments by Size Exclusion Chromatography Coupled with an Orbitrap Mass Spectrometer, ASMS 214. 7. Rathore, A.S.; Joshi, V.; Yadav, N. Aggregation of monoclonal antibody products: formation and removal, BioPharm International, 213, 26, 4-45. Application Note 2126 For Research Use Only thermoscientific.com/biolc 216 Thermo Fisher Scientific Inc. All rights reserved. All other trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is presented as an example of the capabilities of Thermo Fisher Scientific products. It is not intended to encourage use of these products in any manners that might infringe the intellectual property rights of others. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details. USA and Canada +1 8 332 3331 Australia 13 735 292 (free call domestic) China 8 81 5118 (free call domestic) 4 65 5118 France +33 ()1 6 92 48 34 Germany +49 () 2423 9431 2 or 21 India +91 22 6742 9494 +91 27 1766 2352 AN2126-EN 616S Japan 12 753 67 (free call domestic) 12 753 671 (fax) Korea +82 2 342 86 United Kingdom +44 () 1928 534 11 New Zealand 8 933 966 (free call domestic) Singapore +65 6289 119 All Other Enquiries +44 () 1928 534 5 Technical Support For advice and support, please visit our website: www.thermoscientific.com/chromexpert