The Advantages of Automatic Quality Control (AQC) for Blood Gas Testing QC

Similar documents
I-STAT POCT System PRINCIPLE:

Completing the performance picture.

AAG PRINCIPLE SPECIMEN. REF (150 tests) ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE

Global Siemens Headquarters Siemens AG Wittelsbacherplatz Muenchen Germany

AU680 Chemistry Analyzer In-Lab Training Manual

NovoCyte Flow Cytometer

YSI 2900 Series Biochemistry Analyzers BIOPROCESS FOOD & BEVERAGE BIOFUELS MEDICAL RESEARCH

CAMAG. Jack Henion, Emeritus Professor Cornell University and VP of R&D. at Quintiles Bioanalytical and ADME Labs, USA

Q-STRESS CARDIAC STRESS TESTING SYSTEM

Welch Allyn Connex Vital Signs Monitor 6000 Series and Accessory Cable Management Stand

determine optimum instrument settings for their own instruments and establish their own daily values.

Automated Breast Volumes. Simplified.

LifeCycle. manages the entire maternal health program

ACUSON P500 Ultrasound System FROSK edition FROSK Edition

Dry Hematology Analyzer Fast, Simple CBC Analysis Designed for the Point of Care

siemens.com/luminos-select Luminos Select don t compromise, be select.

Interventional CT: More Procedures, Reduced Dose, Improved Speed at Minimal Cost

testo 340 Speed, power & precision in industrial combustion analysis.

Is there a way to unify economical and ecological demands in healthcare?

The Simple Solution for CBC Testing

LabGuide 13 How to Verify Performance Specifications

Explore the future of osmometry. Multi-Sample Micro-Osmometer

OCCUPATIONAL HEALTH MANAGER

Meets Standards AFTER REVIEW ( 2.5 pts) Fail to meet the standard safety precautions, counseled one time

Compounding Solutions. An advanced system for multi-ingredient solution compounding

Strategies for Enhancing the Clinician s Workstation Productivity

Automated Hematology System. Your Complete Choice

Technical Compendium CDI. Blood Parameter Monitoring System 500

CAPABILITIES. Element LIMS. LABORATORY INFORMATION MANAGEMENT for Analytical Testing Laboratories

Expert workflow management cobas p 512 pre-analytical system Professional management of laboratory processes

Stack Gas Analyzer ENDA-7000 ULTIMATE. Ultimate Reliability. Ultimate. Ultimate. Dependability. Accessibility

IQCP for the Akers Bio PIFA Heparin Platelet Factor 4 Rapid Assay

End-to-End Management of Clinical Trials Data

Strengthening Laboratory Management Toward Accreditation. Module 6: Quality Assurance

Outsourcing for IVD Manufacturers

XS-1000i New Sysmex 5-part diff haematology analyser with fluorescence technology

TANGO. ANALYSIS TO GO. Innovation with Integrity. The next generation FT-NIR spectrometer. FT-NIR

Dictation & Transcription Solutions

HIGHER IMPACT LC/MS DIAGNOSTICS AGILENT IN VITRO DIAGNOSTIC MEDICAL DEVICES FOR CLINICAL LABORATORIES

Porcine IgM (Immunoglobulin M) ELISA Kit

CareCenter MD The first PC-based, wireless, diagnostic workstation

t-doc cycle machine data logging digitize your cycle processes

Applied Biosystems 7500 Fast, 7500 and 7300 Real-Time PCR Systems

Choosing a Water Level Logger: 5 Things You Should Know

icap 7000 Plus Series ICP-OES Streamlined performance and ultra-low ICP-OES detection limits for routine laboratories

DICTATION & TRANSCRIPTION

XN-SERIES AUTOMATED HEMATOLOGY SYSTEMS

Overcome inventory management challenges with our comprehensive suite of solutions

SORTEX B range. Optimising conventional sorting.

Simplicity and substrate versatility. Thermal Inkjet. Videojet 8610

Contents OVERVIEW... 3

CARESTREAM DIRECTVIEW Max CR System

// XPLORER-NS. Full range Total Nitrogen, Total Sulfur instrument for the modern Laboratory.

PCR SYSTEMS. a new era in high-productivity qpcr. Applied Biosystems ViiA 7 Real-Time PCR System

Biofluids Analysis: Human Serum with High Throughput NMR

XN-9100 AUTOMATED HEMATOLOGY SYSTEMS

Lytec Features Evolution Matrix

AK 95 S Dialysis machine. Part of Gambro s complete HD system

Indian medicine manufacturer Medreich PLC improves production and quality with coding solutions from Videojet

APFCB. Improving Laboratory Performance Through Quality Control. Designing an Appropriate QC Procedure for Your Laboratory QUALITY CONTROL

New Product Brief PreVAS Single-use Dosing System. Packaging Technology. 3 Disposable Dosing Systems

cobas 8100 automated workflow series 3-D intelligence in lab automation

Sum Parameter Analysis AOX/TOX. AOX Analyzer multi X 2500

QuantStudio Dx Real-Time PCR Instrument

Online data sheet MCS100E HW-C CEMS SOLUTIONS

ATM MODERNIZATION FOUR REASONS TO MODERNIZE YOUR AGING ATM INSTALL BASE. An NCR white paper

Chloride/Sulfate Analyzer

Formulation / Recipe Weighing

HEMOCHRON. Whole Blood Coagulation Systems

Expert workflow management cobas p 612 pre-analytical system. Professional management of laboratory processes

The new AK 96 machine

Anti- Tissue Transglutaminase TTG IgG ELISA Kit

Training. Advancing yourself

This package insert contains information to run the Albumin BCP assay on the ARCHITECT c Systems and the AEROSET System.

Medical Pouch & Bag Sealers

Planning Optimized. Building a Sustainable Competitive Advantage WHITE PAPER

Note: The taller 100 x 13 mm tube is not designed to run on the ACL TOP. Plasma after centrifugation of these tubes must be placed in sample cups.

cobas POC IT solutions Bringing it all together

Free, Total and Double SO2 in Wine

Sahara Pro Advanced colorant performance

Human myelin basic protein(mbp) antibody ELISA Kit

QUALITY CONTROL/QUALITY ASSURANCE IN THE MOLECULAR MICROBIOLOGY LABORATORY

Streamline workflow for higher productivity and patient throughput. DirectView. Classic CR System

COLORIMETRIC SANDWICH ELISA KIT INSTRUCTION MANUAL

Falcon Cell Culture Multi-Flask An easier, more productive way to culture cells

SIWAREX FTC The belt scales in SIMATIC More than just standard. siwarex ftc

For In Vitro Diagnostic Use

Cedex Bio Bioprocess Analyzer. Be Confident. Be Assured.

ICP Expert software. Technical Overview. Introduction

Image guided therapy. Azurion 7. With Azurion, performance and superior care become one

T-DOC EndoSolutions Manage your Endoscope workflow

FER PRINCIPLE SPECIMEN. REF (150 tests) ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE

What it takes to make a new state of the art diagnostics platform for Point of Care use.

Transcription:

Siemens Healthcare Diagnostics, the leading clinical diagnostics company, is committed to providing clinicians with the vital information they need for the accurate diagnosis, treatment and monitoring of patients. Our comprehensive portfolio of performance-driven systems, unmatched menu offering and IT solutions, in conjunction with highly responsive service, is designed to streamline workflow, enhance operational efficiency and support improved patient care. RAPIDComm, RAPIDLab, RAPIDLink, RAPIDQC, RAPIDPoint, and all associated marks are trademarks of Siemens Healthcare Diagnostics Inc. All other trademarks and brands are the property of their respective owners. Product availability may vary from country to country and is subject to varying regulatory requirements. Please contact your local representative for availability. Global Siemens Headquarters Siemens AG Wittelsbacherplatz 2 80333 Muenchen Germany Global Siemens Healthcare Headquarters Siemens AG Healthcare Sector Henkestrasse 127 91052 Erlangen Germany Phone: +49 9131 84-0 www.siemens.com/healthcare Global Division Siemens Healthcare Diagnostics Inc. 511 Benedict Avenue Tarrytown, NY 10591-5005 USA www.siemens.com/diagnostics The Advantages of Automatic Quality Control (AQC) for Blood Gas Testing QC White Paper www.siemens.com/diagnostics Answers for life. Order No. A91DX-110396-GC1-4A00 Printed in USA 02-2011 All rights reserved 2011 Siemens Healthcare Diagnostics Inc.

Introduction For over 40 years, Siemens Healthcare Diagnostics has been developing, manufacturing and marketing critical care products, including blood gas analyzers that now incorporate the innovative Automatic Quality Control (AQC) systems. Our Quality Control (QC) products and services portfolio is designed to help laboratories monitor the reliability and performance of their instrumentation, measure the variability of tests performed, and evaluate the integrity of all results reported. Our commitment to consistency and stringent quality control helps laboratories around the world to excel in the face of ever-increasing regulatory demands and higher performance goals. An overview of Quality Control An analyzers QC system helps to ensure and verify that analytical processes are performing according to expectations, and more importantly, that all patient test results obtained are accurate. Any instrument QC program generally incorporates a variety of processes to help ensure performance quality. These processes include instrument calibration, calibration verification, system functionality checks, preventative maintenance, proficiency testing, and the routine analysis of materials of known concentration, referred to as controls. 1964 First Solid-State ph 1968 16 ph/blood Gas 1971 161 ph/blood Gas 1972 165 ph/blood Gas 1975 175 ph/blood Gas 1979-81 100 Series ph/blood Gas s 1982 2500 CO-oximeter 1987 600 Series Electrolyte s 1988 200 Series Systems 1990 RAPIDLab 238 Blood Gas 1992 270 CO-oximeter 1994 RAPIDLab 800 Systems 1996 RAPIDLab 348 Blood Gas 1997 RAPIDLink Data Management System 2001 RAPIDPoint 400 System 2003 RAPIDPoint 405 System 2005 RAPIDLab 1200 Series Systems 2007 RAPIDComm Data Management System 2010 RAPIDPoint 300 Systems 1988 Ready Sensor Electrodes 1994 RAPIDLab 248 Blood Gas Corning Glassworks Ciba-Corning Chiron Diagnostics Bayer Diagnostics Siemens Figure 1: Siemens Blood Gas Testing Milestones 2 3

Why perform Quality Control? There are three important reasons for test sites to run QC checks: 1. The measured values of control analytes will indicate to the operator if the analyzer is reporting results within acceptable limits. Acceptable QC performance, parameter is ready Failed QC Parameter failed calibration: Parameter should be available after repeat calibration Parameter failed calibration: Parameter is unlikely to become available Required QC analysis was not performed when scheduled 2. To satisfy regulatory requirements that may be unique to certain regions of the world. At the direction of the laboratory director or point-of-care supervisor, and/or according to established local protocols, both RAPIDLab 1200 and RAPIDPoint 400 Series analyzers automatically run the desired level of QC at a frequency that meets or exceeds regulatory requirements. Figure 2. Siemens RAPIDLab 1200 and RAPIDPoint 400 Series analyzers automatically report acceptable and unacceptable QC results for multiple analytes Figure 3. Siemens blood gas analyzers can be configured to meet user-defined QC schedules. Fully customizable QC schedules include selecting the day, time interval and level of QC. Siemens RAPIDLab 1200 and RAPIDPoint 400 Series blood gas analyzers feature fully automated QC routines that measure control samples and evaluate patient test results. If QC results (based on user-defined criteria) are acceptable, then patient values are reported. If a QC test result is out of control, the analyzer will attempt to rectify the situation by automatically repeating the QC procedure. If the analyzer continues to be unsuccessful with the repeated QC procedure, the test channel will be shut down and operator intervention will be required. 3. To evaluate the operational performance of the analyzer over time. Day-to-day and month-on-month performance of Siemens blood gas analyzers can be assessed by laboratory directors and point-of-care supervisors through tabulated reports of QC data, up-to-date QC statistics and softwaregenerated Levey-Jennings graphs. Figure 4: Levey-Jennings graph showing Level 1 AQC results for ph over time on a RAPIDLab 1200 system 4 5

How has blood gas QC evolved over recent years? Blood gas QC has evolved over time as new technologies and materials have become available. Previously, a blood gas control material was made of a buffer solution spiked with varying concentrations of electrolytes, metabolites and hemoglobin (and the hemoglobin fractions). The buffer was also equilibrated with a gas mixture containing oxygen, carbon dioxide and nitrogen. This procedure is known as tonometry. While this QC approach worked well, the procedure was a tedious one, and difficult to maintain at the laboratory level. What is AQC? AQC is an acronym for the Automatic Quality Control system used on Siemens RAPIDLab 1200 and RAPIDPoint 400 Series blood gas analyzers. This cartridge-based system provides QC in a standardized and unique manner that removes the need for operator intervention, and ensures the running of specific levels of QC at specified time intervals, as regulatory requirements dictate or according to individual hospital protocol. Recognizing the need for a commercially available blood gas QC product, manufacturers replaced the manual tonometry process with a blood gas QC product in a sealed glass ampule, impervious to the effects of room air. Although easy to use, it is important that this type of control material be tested as soon as possible after opening, since control values can change upon exposure to room air. RAPIDLab 1200 Blood Gas Series System The possibility of shifting control values due to room air exposure, coupled with the operator interaction required to run QC checks, led manufacturers to search for an alternative solution RAPIDQC Complete product, packaged 30 ampules per level, three unique levels Figure 6: On-board AQC requires no operator intervention during its 28-day cartridge life RAPIDPoint 400/405 Blood Gas Series System The response from Siemens was the Automatic Quality Control (AQC) cartridge. It is important to note that cartridgebased AQC is, just like ampuled QC, an external form of quality control. It is simply packaged differently, and is more convenient. The AQC cartridge is value assigned to a master lot, and is entirely external to the measurement system. The consistency of the manufacturing process allows the laboratory to consider their AQC as one long lot number with no expiration date. The elimination of room air contamination allows the laboratory to consider using the data as one long shift on a day-to-day basis. This technology allows all laboratories to apply the same expected ranges over time for a given type of analyzer. Figure 7: Actual reagent and QC manufacturing line for the cartridge-based products. 6 7

What does the AQC system consist of? How does the AQC system work? The AQC system is made up of the following components: AQC Cartridge a single, self-contained cartridge that attaches to the side of the analyzer and remains viable for up to 28 days. AQC Levels contained in three (Level 1, Level 2, Level 3) separate, sealed foil bags inside the cartridge. A buffer amount of QC is added to each one of the bags to account for repeats. The three levels of control material span the clinically significant ranges. AQC fluidics is designed so that the AQC manifold provides a fluid path for QC materials to flow from the AQC cartridge to the patient sample entry port of the instrument reagent/measurement cartridge. In Figure 8, the red line represents the patient sample flow path and the blue line represents the QC flow path. Both the patient sample and QC materials follow the exact same sample path in the reagent/measurement cartridge (also see Figure 9). Foam wash AQC sample flow Wash 200 cal RCX Zero Data Manager this is built into the blood gas analyzer software, and includes a fully customizable, automatic schedule and decision rules that determines the availability and frequency for testing each parameter. Figure 9: The AQC pathway is the same as the patient sample pathway Light source/ Filter stack Spectrometer Optic head/cell driver mechanism Connector Peri-pump System sensors AQC Cartridge Reagent Sample and QC aspiration Sample collection Waste Calibration/wash fluids Measurement Cartridge Waste Cartridge Figure 8: RAPIDLab 1200 system fluidics both patient sample and AQC pathways RAPIDPoint 400/405 System with AQC cartridge 8 9

QC statistical application Siemens recommends the evaluation method of 3SD (Standard Deviations) Absolute Limits for RAPIDLab 1200 and RAPIDPoint 400 Series analyzers when using on-board AQC. In simple terms, the QC recovery is either in or out of established default range limits. Since each reagent/measurement cartridge contains all new reagents, pump tubes, and fluidic pathways, when a changeover is implemented (every 28 days) it is generally accepted that there will be variability and shifts associated with the change. A similar shift is seen when a sensor or reagent lot is changed on conventional blood gas analyzers (for example, the RAPIDLab 800 Series analyzers) every 6-12 months. However, the amount of change seen with RAPIDLab 1200 or RAPIDPoint 400 Series analyzer cartridges is minimal compared to that seen with more traditional blood gas systems. This is due to two unique system characteristics: Each AQC batch is manufactured to be identical in numerical values to previous batches. This allows operators to compare data over years rather than just months. There is no exposure to room air. Exposure to atmosphere when using ampuled QC tends to generate the most variability. With these two sources of potential error virtually eliminated, the use of traditional 2SD QC limits could cause any small data change to look significant, despite not being clinically significant. The suggested fixed or absolute QC limits are designed to be consistent with the medical needs of clinicians. These limits reduce to an absolute minimum the number of false rejections due to erroneous QC signals, as noted by comparing the small actual standard deviations to the absolute QC limits in Table 1. If operators were to evaluate RAPIDLab 1200 or RAPIDPoint 400 Series analyzer QC results using traditional 2SD statistical limits, they might overreact to small shifts or trends associated with a change of the measurement cartridge, and possibly request unnecessary and potentially costly service interventions. Additionally, traditional QC management only addresses the process capabilities of the instrument based upon recent history. Absolute limits evaluation allows users to apply clinical decision-making to the measurement process over extended periods of time, thereby avoiding an overreaction and expenditure of time and energy resulting from a change that is statistically, but not clinically, significant or relevant. In summary, the fixed QC limits represent the largest possible error that may be present in a test result. These are sometimes referred to as medical need or medical usefulness limits. Because these limits are much larger than the standard deviation of highly precise RAPIDLab 1200 or RAPIDPoint 400 analyzer results, false positives and erroneous out of control signals are essentially eliminated. AQC Level 1 AQC Level 2 AQC Level 3 Target Range (+/-) Target Range (+/-) Target Range (+/-) ph 7.150 0.020 7.350 0.020 7.55 0.026 pco 2 (mmhg) 70.0 6.4 40.0 5.0 22.0 3.0 po 2 (mmhg) 150.0 11.0 100.0 7.8 65.0 9.2 Na + (mm) 115.0 5.0 135.0 5.0 155.0 7.0 K + (mm) 3.00 0.30 5.00 0.30 7.00 0.30 Ca ++ (mm) 1.60 0.12 1.20 0.10 0.80 0.10 Cl - (mm) 80.0 6.0 100.0 6.0 120.0 8.0 Glucose (mg/dl) 200 14 100 10 50 10 Lactate (mm) 12.00 2.00 0.90 0.24 3.00 0.60 thb (g/dl) 18.0 1.6 14.0 1.6 8.0 1.0 % O 2 Hb 78.0 3.0 92.0 3.0 60.0 3.0 % HHb 2.5 4.0 2.5 4.0 16.0 4.0 % COHb 3.5 5.0 3.5 5.0 18.0 5.0 % MetHb 16.0 3.6 2.0 4.6 6.0 3.6 n-bilirubin (mg/dl) 20 4.0 12 2.8 5 2.4 Table 1: AQC target and acceptable range for each reported parameter and each level of QC 10 11

Checks and balances to ensure quality results QC checks, performed either manually or by means of the AQC system, represent an important and independent verification of the stability of both analyzers and overall analyte performance. Nevertheless, calibrations, multiple algorithm checks on patient samples, and multiple quality checks post-patient sample analysis remain the primary guardians of patient result quality. Siemens blood gas systems incorporate a variety of analyzer functions and flagging mechanisms to ensure the quality of the patient results: Calibration routines are set to run after every sample for the first eight hours after installation, and then run at a minimum of once every half hour thereafter. The drift limits for calibrations are set well below clinically significant limits. Any calibration that fails is repeated a maximum of two times, or until the analyte passes. If the analyte does not pass, the parameter is turned off until the next scheduled calibration. The sensors for each analyte have hard-coded limits for slope and offset. Any changes are corrected and adjusted through a calibration cycle. Once any of these reach their limits, the analyte is permanently disabled. In addition to traditional calibration cycles, all sample, calibration, QC and AQC responses are quality-checked with up to eight internal software algorithms. If any of these fail, the sample will be flagged and no result will be reported. A solution is run between each sample to execute multiple sensor quality checks on the analyzers. Additional evaluations of results include checks for bubbles in the sample, and temperature. All function checks are performed automatically. All actions are recorded and documented in the event log. All raw data may be downloaded and forwarded to Siemens for further analysis. The totality of independent quality checks available with Siemens blood gas analyzers, in combination with the Automatic Quality Control system, helps to ensure the integrity of all reported patient test results. 12 13

Peace of mind with AQC Siemens AQC system offers: The flexibility to set a quality control schedule that meets your individual hospital standards and complies with local regulatory guidelines. A unique design that ensures control of the entire sample path the AQC material is introduced at the same point as the patient sample. Three different levels of aqueous material that cover the critical reporting ranges as well as the normal ranges. QC materials that are truly independent from on-board calibrators. The ability to interrupt the QC routine to run a patient sample. Cost-effectiveness and ease of use versus ampuled QC. What should I do now? If your organization is using a Siemens analyzer equipped with AQC, you have to decide your approach to QC limits based on the facts presented above. You should also discuss the information with your medical director. Ultimately the final decision rests with your medical director. Based on the statistical data, the most practical solution would be the use of fixed limits with AQC, since this avoids unnecessary sample rejection while still allowing for error detection, when applicable. This will ensure that you maintain maximum instrument uptime and operational cost-effectiveness. AQC Specifications AQC Cartridge Use-life: Shelf-life: Storage: Humidity QC Method Time to Result User Interface QC Lock-out Automatic Repeat Constant Expected Ranges Customizable Ranges Maintenance Sensitive to Operator Handling Data Management 28 days 4 months Temperature: 2 8 C 0 95%, non-condensing Aqueous QC Material 60 Seconds Color Touch Screen Yes Yes, if enabled Yes Yes None required No Yes, with external RAPIDComm Data Management System 14 15