Analysis of Monoclonal Antibody Digests with the Agilent 1290 Infinity 2D-LC Solution

Similar documents
Characterize Fab and Fc Fragments by Cation-Exchange Chromatography

High Sequence Coverage Peptide Mapping of a Monoclonal Antibody

Disulfide Linkage Analysis of IgG1 using an Agilent 1260 Infinity Bio inert LC System with an Agilent ZORBAX RRHD Diphenyl sub 2 µm Column

Cell Clone Selection Using the Agilent Bio-Monolith Protein A Column and LC/MS

Peptide Mapping: A Quality by Design (QbD) Approach

mab Titer Analysis with the Agilent Bio-Monolith Protein A Column

Application Note. Author. Abstract. Biotherapeutics and Biologics. Sonja Schneider Agilent Technologies, Inc. Waldbronn, Germany

Fraction Analysis of Cysteine Linked Antibody-Drug Conjugates Using Hydrophobic Interaction. chromatography. Agilent 1260 Infinity II Bio-Inert System

Application Note. Author. Abstract. Biopharmaceuticals. Verified for Agilent 1260 Infinity II LC Bio-inert System. Sonja Schneider

Technical Overview. Author. Abstract. Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany

Application Note. Author. Abstract. Small Molecule Pharmaceuticals. Sonja Krieger Agilent Technologies, Inc. Waldbronn, Germany

Maximizing Chromatographic Resolution of Peptide Maps using UPLC with Tandem Columns

Quality-by-Design-Based Method Development Using an Agilent 1290 Infinity II LC

Characterization of mab aggregation using a Cary 60 UV-Vis Spectrophotometer and the Agilent 1260 Infinity LC system

Analysis of Intact and C-terminal Digested IgG1 on an Agilent Bio MAb 5 µm Column

Size Exclusion Chromatography of Biosimilar and Innovator Insulin Using the Agilent AdvanceBio SEC column

Developing Robust and Efficient IEX Methods for Charge Variant Analysis of Biotherapeutics Using ACQUITY UPLC H-Class System and Auto Blend Plus

Size Exclusion Chromatography/ Mass Spectrometry Analysis of Antibody Drug Conjugates Using the Agilent 1260 Infinity II Bio-Inert LC

Precise Characterization of Intact Monoclonal Antibodies by the Agilent 6545XT AdvanceBio LC/Q-TOF

mabs and ADCs analysis by RP

PLRP-S Polymeric Reversed-Phase Column for LC/MS Separation of mabs and ADC

Lifetime stability of size exclusion chromatography columns for protein aggregate analysis

Generating Automated and Efficient LC/MS Peptide Mapping Results with the Biopharmaceutical Platform Solution with UNIFI

A Fast and Robust Linear ph Gradient Separation Platform for Monoclonal Antibody (MAb) Charge Variant Analysis

Method Development Considerations for Reversed-Phase Protein Separations

Improved Peptide Maps Using Core-Shell Media: Explaining Better Biopharmaceutical Applications With Kinetex C18 Columns

ADVANCE ACCURACY AND PRODUCTIVITY FOR FASTER ANALYSIS

Method Translation in Liquid Chromatography

High-Throughput Peptide Mapping with the Vanquish UHPLC System and the Q Exactive HF Mass Spectrometer

HIC as a Complementary, Confirmatory Tool to SEC for the Analysis of mab Aggregates

A universal chromatography method for aggregate analysis of monoclonal antibodies

UFFLC of the Proteome Ultra High Pressure and Ultra Fast Flow Liquid Chromatography: A New Approach To Proteomics

Characterization of Glycosylation in the Fc Region of Therapeutic Recombinant Monoclonal Antibody

N-Glycan Profiling Analysis of a Monoclonal Antibody Using UHPLC/FLD/Q-TOF

Qualification of High-Performance Liquid Chromatography Systems

Fast and High-Resolution Reversed-Phase Separation of Synthetic Oligonucleotides

Choosing the Right Calibration for the Agilent Bio SEC-3 Column

A Unique LC-MS Assay for Host Cell Proteins(HCPs) ) in Biologics

Physical Stability of a Silica- Based Size Exclusion Column for Antibody Analysis

Analysis of USP <467> Residual Solvents using the Agilent 7697A Headspace Sampler with the Agilent 7890B Gas Chromatograph

Proteins. Patrick Boyce Biopharmaceutical Marketing Manager Waters Corporation 1

Protein Quantitation using various Modes of High Performance Liquid Chromatography

EPA Method 543: Selected Organic Contaminants by Online SPE LC/MS/MS Using the Agilent Flexible Cube

Peptide Mapping of Glycoprotein Erythropoietin by HILIC LC/MS and RP-LC/MS

Two-Dimensional LC Protein Separation on Monolithic Columns in a Fully Automated Workflow

An Automated System for At-Line Process Analysis of Biopharmaceutical Fermentation Reactions

Application Note. Author. Abstract. Environmental. Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany

Protein-Pak Hi Res HIC Column and HIC Protein Standard

Development of Ultra-fast ph-gradient Ion Exchange Chromatography for the Separation of Monoclonal Antibody Charge Variants

Determination of Iodinated Contrast Media in Aqueous Samples by Direct-Injection LC-MS/MS

Flexibility with. Preparative HPLC

Trastuzumab Glycan Batch-to-Batch Profiling using a UPLC/FLR/Mass Spectrometry Platform

Rapid Peptide Mapping via Automated Integration of On-line Digestion, Separation and Mass Spectrometry for the Analysis of Therapeutic Proteins

Optimizing the Purification of a Standard Chiral Compound Utilizing a Benchtop, Multi-Purpose, Semi-Preparative to Preparative HPLC System

Analysis and Purification of Polypeptides by Reversed-Phase HPLC

Improving Resolution and Column Loading Systematically in Preparative Liquid Chromatography for Isolating a Minor Component from Peppermint Extract

Confident Pesticides Analysis with the Agilent LC/Triple Quadrupole and TOF/QTOF Solutions

Guide to Trap Cartridge Care and Use

Asish Chakraborty, St John Skilton, Weibin Chen and John C. Gebler Waters Corporation, Milford, MA, U.S. INTRODUCTION METHODS. Sample preparation

MAbPac SCX-10 Columns

NPTEL VIDEO COURSE PROTEOMICS PROF. SANJEEVA SRIVASTAVA

AZURA Bio LC. 2 FPLC and Purification Solutions

SMART Digest compared to classic in-solution digestion of rituximab for in-depth peptide mapping characterization

Technical Overview. Authors. Abstract. Verified for Agilent 1260 Infinity II LC Bio-inert System

Detecting low level sequence variants in recombinant monoclonal antibodies

Analysis of biomolecules by SEC and Ion-Exchange UPLC

Agilent 1220 Infinity II LC GET ON THE FAST TRACK TO HIGHEST EFFICIENCY AT A PRICE YOU CAN AFFORD

Concentration of Human Hormones in Drinking Water Using Solid Phase Extraction and Analysis by High Performance Liquid Chromatography

R R Innovation Way P/N SECKIT-7830 Newark, DE 19711, USA Tel: Fax: Website: Published in November 2013

Reversed Phase Solutions for the Analysis of Proteins, Peptides, and Oligonucleotides

Thermo Scientific BioLC Columns. Monoclonal antibody. characterization

Zwitterion Chromatography ZIC

Bio-Monolith Protein G Column - More Options for mab Titer Determination

A Modular Preparative HPLC System for the Isolation of Puerarin from Kudzu Root Extracts

Why do LC MS and LBA

Preparative HPLC is still the

Automated Online SPE for LC/MS/MS Analysis of Trace Organic Contaminants in Water Using the Agilent 1290 Infinity Flexible Cube Module

MassPREP On-Line Desalting Cartridge

Application Note. Authors. Abstract

Agilent BioHPLC Columns PROTEIN IDENTIFICATION AND IMPURITY PROFILING USING REVERSED-PHASE HPLC/UHPLC

Preparative HPLC: Factors and Parameters that Directly Affect Recovery of Collected Fractions

'Tips and Tricks' for Biopharmaceutical Characterization using SEC

SMART Digest Kit User Manual. Version 2

Exploring Extra Sensitivity Using ionkey/ms with the Xevo G2-XS Q-Tof HRMS for Small Molecule Pharmaceutical Analysis in Human Plasma

BÜCHI Labortechnik AG

Chem 321 Lecture 23 - Liquid Chromatography 11/19/13

Thermo Scientific Peptide Mapping Workflows. Upgrade Your Maps. Fast, confident and more reliable peptide mapping.

Agilent 7693A Series Automatic Liquid Sampler. Inject new performance into your gas chromotography

Agilent 7693A Series Automatic Liquid Sampler INJECT NEW PERFORMANCE INTO YOUR GAS CHROMATOGRAPHY

Agilent 7693A Series Automatic Liquid Sampler INJECT NEW PERFORMANCE INTO YOUR GAS CHROMATOGRAPHY

AZURA Bio purification

The Theory of HPLC Gradient HPLC

Proteomics Technology Note

ZORBAX StableBond HPLC Columns

Highly Confident Peptide Mapping of Protein Digests Using Agilent LC/Q TOFs

Thermo Scientific LC Columns for Biomolecules

Genotoxicity is the property of a compound

Bivalirudin Purification:

Thermo Scientific EASY-nLC 1200 System. Leading in simplicity. and performance

Transcription:

Analysis of Monoclonal Antibody Digests with the Agilent 1290 Infinity 2D-LC Solution Application Note Biotherapeutics & Biosimilars Authors Gerd Vanhoenacker, Koen Sandra, Isabel Vandenheede, Frank David, and Pat Sandra Research Institute for Chromatography Kennedypark 26 B-8500 Kortrijk, Belgium Metablys Kennedypark 26 B-8500 Kortrijk, Belgium Abstract Protein biopharmaceuticals such as monoclonal antibodies and recombinant proteins are currently in widespread use for the treatment of various life-threatening diseases including cancer and autoimmune diseases. Protein therapeutics have a complexity far exceeding that of small molecule drugs, hence, unraveling this complexity represents an analytical challenge. This Application Note demonstrates the potential of comprehensive two-dimensional liquid chromatography (LCxLC) using the Agilent 1290 Infi nity 2D-LC Solution with DAD detection for the analysis of monoclonal antibody tryptic digests. Udo Huber, Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany

Introduction Therapeutic macromolecules produced by recombinant DNA technology are becoming increasingly important in the treatment of various diseases. It is estimated that, at present, the global protein therapeutics market (monoclonal antibodies (mab) and other recombinant proteins) represents approximately 20 % of the total pharmaceutical market. Within the current decade, over 50 % of new drug approvals will be biologics 1. During the development and lifetime of these molecules, an in-depth characterization is required. This Application Note describes a peptide mapping method for identity and purity assessment of the monoclonal antibody trastuzumab. Trastuzumab, marketed as Herceptin, is a 150 kda large molecule used in the treatment of HER2 positive metastatic breast cancer. Following trypsinolysis, over 100 peptides with varying physicochemical properties present in a wide dynamic concentration range are expected (Figure 1). This represents a complex analytical challenge, requiring the highest separation power. Comprehensive LCxLC is known to substantially increase the chromatographic resolution as long as the two dimensions are orthogonal and the separation obtained in the first dimension is maintained upon transfer to the second dimension 2,3. This Application Note reports on the combination of strong cation exchange (SCX) and reversed-phase liquid chromatography (RPLC), two combinations shown to provide good orthogonality towards the analysis of peptides 3. Fractions were transferred from the first to the second dimension using a dual loop interface maintaining the (first dimension) resolution. An Agilent 1290 Infinity 2D-LC Solution system was used. Some key results are highlighted and the performance of the system is demonstrated. T1 T2 T3 T4 DIQMTQSPSSLSASVGDR VTITCR ASQDVNTAVAWYQQKPGK APK T5 T6 T7 LLIYSASFLYSGVPSR FSGSR SGTDFTLTISSLQPEDFATYYCQQHYTTPPTFGQGTK T8 VEIK T9 R T10 TVAAPSVFIFPPSDEQLK T14 VDNALQSGNSQESVTEQDSK T18 VYACEVTHQGLSSPVTK T19 SFNR T11 SGTASVVCLLNNFYPR T15 DSTYSSTLTLSK T20 GEC T16 ADYEK T17 HK T12 EAK T13 VQWK Light chain B(1-214) N-ter EVQLVESGGGLVQPGGSLRLSCAASGFN IKDTYIHWVRQAPGKGLEW Lc Hc N-ter T21 EVQLVESGGGLVQPGGSLR T26 IYPTNGYTR T27 YADSVK T28 GR T22 LSCAASGFNIK T29 FTISADTSK T32 WGGDGFYAMDYWGQGTLVTVSSASTK T23 DTYIHWVR T30 NTAYLQMNSLR T33 GPSVFPLAPSSK T24 QAPGK T25 GLEWVAR T31 AEDTAVYYCSR T34 STSGGTAALGCLVK T35 DYFPEPVTVSWNSGALTSGVHTFPAVLQSSGLYSLSSVVTVPSSSLGTQTYICNVNHKPSNTK T36 VDK T55 EEMTK T60 SR T37 K T38 VEPK T46 T47 VVSVLTVLHQDWLNGK EYK T56 NQVSLTCLVK T39 SCDK T42 TPEVTCVVVDVSHEDPEVK T43 FNWYVDGVEVHNAK T48 CK T49 VSNK T57 GFYPSDIAVEWESNGQPENNYK T61 WQQGNVFSCSVMHEALHNHYTQK T40 THTCPPCPAPELLGGPSVFLFPPKPK T50 ALPAPIEK T62 SLSLSPG T44 TKPR T51 TISK T45 EEQYNSTYR T52 AK T41 DTLMISR T53 T54 GQPR EPQVYTLPPSR T58 TTPPVLDSDGSFFLYSK T59 LTVDK Heavy chain A(1-449) C-ter C-ter NYKTTPPVLDSDGSFFLYSKLTVDKSRW QQGNVFSCSVMHEALHNHYTQKSLSLSPG 62 identity peptides Modifications Incomplete and aspecific cleavages... > 100 peptides Figure 1. Structure and amino acid sequence of the protein. Identity peptides are labeled T1-T62. 2

Experimental Solutions and Samples All solvents were HPLC gradient grade from Biosolve B.V. (Valkenswaard, the Netherlands). Phosphoric acid and sodium chloride were from Sigma-Aldrich (Bornem, Belgium). Sample Preparation The following sample preparation procedure was used. Sample Volume corresponding to 100 µg protein Add surfactant 0.1 % Rapigest in 100 mm Tris ph 7.5 Dilution 1 mm CaCl 2 Reduction 5 mm dithiothreitol (DTT) Alkylation 10 mm iodoacetamide (IAA) Digestion Trypsin (1:25 w/w), 37 C for 16 hours Prepare for analysis Lower ph (TFA), wait, centrifuge The sample was also subjected to the following forced degradation conditions. Oxidation stress ph Stress tert-butyl hydroperoxide (TBHP), 30 C for 23 hours Tris ph 9, 37 C for 3 days Instrumentation An Agilent 1290 Infinity 2D-LC Solution was used. The configuration is shown below. G4220A G4220A G4226A G1330A G1316C G4212A G1170A G4236A Agilent 1290 Infinity Binary pump with seal wash option (1st dimension) Agilent 1290 Infinity Binary pump (2nd dimension) Agilent 1290 Infinity Autosampler Agilent 1290 Infinity Autosampler Thermostat Agilent 1290 Infinity Thermostatted Column Compartment Agilent 1290 Infinity Diode Array Detector with standard flow cell Agilent 1290 Infinity Valve Drive 2-position/4-port duo valve for 2D-LC Software Agilent OpenLAB CDS Chemstation revision C.01.04 with 2D-LC add-on software GC Image LCxLC Edition Software for 2D-LC data analysis (GC Image, LLC., Lincoln, NE, USA) 3

Method 1 st Dimension Column Solvent A Solvent B Flow rate Gradient MIC-15-Polysulfoethyl-Asp, 150 1.0 mm, 5 µm (PolyLC Inc.) 5 mm phosphate ph 3 in 5 % acetonitrile 5 mm phosphate ph 3 in 5 % acetonitrile + 400 mm NaCl 60 µl/min 0 to 30 minutes : 3 to 25 % B 30 to 50 minutes : 25 to 45 % B 50 to 55 minutes : 45 to 100 % B 55 to 58 minutes : 100 % B Post-time : 13 minutes at 3 % B Temperature 25 C 2 nd Dimension Column Agilent ZORBAX Eclipse Plus C18, 4.6 50 mm, 3.5 µm (p/n 959943-902) Solvent A 0.1 % phosphoric acid in water Solvent B acetonitrile Flow rate 3.5 ml/min Idle flow rate 0.5 ml/min Initial gradient 0 to 0.43 minutes : 2 to 35 % B 0.43 minutes : 2 % B Gradient modulation 35 % B at 0.43 minutes to 65 % B at 50 minutes to 100 % B at 51 minutes (Figure 2) Temperature 55 C Modulation Modulation on Loops Modulation time Injection 1 to 53 minutes Two 40-µL loops, cocurrent configuration 0.50 minutes Volume 20 µl Temperature 4 C Needle wash Detection Wavelength Data rate 6 seconds flushport (5 mm phosphate ph 3 in 5 % acetonitrile) Signal 214/4 nm, Reference 360/100 nm 80 Hz Figure 2. Screenshot of the 2D-LC method (OpenLAB CDS Chemstation). 4

Results and Discussion Figure 3 shows the contour plot of a digest generated with the described method and the GC Image software. A good coverage of trastuzumabs sequence was obtained. The identification of the spots was based on an adapted method for combination with mass spectrometry (a G6530B QTOF system was used). The spots are randomly scattered across the surface indicating good orthogonality between both dimensions. Figure 3. LCxLC contour plot for the analysis of a tryptic digest of trastuzumab. 5

The analysis was repeated five times to calculate the precision of the method. Table 1 summarizes the results for a selection of four peptides. The volume and retention time precision is very good, indicating that the method is useful for the comparison of different production batches and between originators and biosimilars. Table 1. Precision data calculated on a selection of spots, n = 5. %RSD T3 T11 T43 T50 Volume 2.04 1.94 4.56 2.59 Height 0.85 1.78 2.15 0.54 RT 2D 0.09 0.05 0.25 0.15 The method can also be applied to detect impurities and modifications. To confirm this, a forced degradation study was carried out on a trastuzumab sample. The product was stressed under oxidative conditions and elevated ph conditions resulting in oxidation and deamidation, respectively. Figure 4 shows a detail of an originator sample and an oxidation stressed sample. The stressed sample clearly contains additional spots. One particular spot was picked out to be identified by LC/MS/MS. The spot could be traced back to peptide T41. It is the result of an oxidation of methionine present in this fragment. DAD originator, not stressed DAD originator, oxidation stress Figure 4. LCxLC contour plot for the analysis of a tryptic digest of non-stressed and oxidatively stressed trastuzumab. Note that the oxidation is already present at marginal levels in the non-stressed sample illustrating the strength of the methodology. 6

Figure 5 presents a similar example, where the result for a ph stressed sample is shown. Peptide 46 is known to be sensitive to deamidation because it contains asparagine. During the ph stress, asparagine is converted to aspartate. An additional spot is visible in the contour plot with identical retention time in the second dimension. This indicates that, under the applied conditions, it will be difficult to separate the identity peptide from its deamidated degradation product with reversed phase LC only. The identity of the modification was confirmed by LC/MS/MS using the adapted method. DAD originator, not stressed DAD originator, ph stressed Figure 5. LCxLC contour plot for the analysis of a tryptic digest of nonstressed and ph stressed trastuzumab. Conclusion The Agilent 1290 Infinity 2D-LC Solution is an ideal solution for the analysis of complex samples such as protein biopharmaceuticals. This method demonstrated great potential for purity and identity assessment. The precision of the set-up was excellent allowing the use of the method for quality control and comparative studies between production batches and originators and biosimilars. 7

References 1. K. Sandra, I. Vandenheede, P. Sandra, The power of liquid chromatography-mass spectrometry in the characterization of protein biopharmaceuticals, LCGC Europe Supplement, May 2013. 2. I. Francois, K. Sandra, P. Sandra, Comprehensive liquid chromatography: fundamental aspects and practical considerations a review, Anal. Chim. Acta. 64 14-31 (2009). 3. K. Sandra, M. Moshir, F. D hondt, R. Tuytten, K.n Verleysen, K. Kas, I. Francois, P. Sandra, Highly efficient peptide separations in proteomics. Part 2: bi- and multidimensional liquid-based separation techniques, J. Chromatogr., B 877, 1019-1039 (2009). www.agilent.com/chem This information is subject to change without notice. Agilent Technologies, Inc., 2013 Published in the USA, September 1, 2013 5991-2880EN