VICH Topic GL2 (Validation: Methodology) GUIDELINE ON VALIDATION OF ANALYTICAL PROCEDURES: METHODOLOGY

Similar documents
Verification of Compendial Methods

European Medicines Agency Evaluation of Medicines for Human Use

Draft regional guidelines on stability testing of active substances and pharmaceutical products

VICH Topic GL26 BIOLOGICALS: TESTING OF RESIDUAL MOISTURE

VICH Topic GL32 STUDIES TO EVALUATE THE SAFETY OF RESIDUES OF VETERINARY DRUGS IN HUMAN FOOD: DEVELOPMENTAL TOXICITY TESTING

WHO GUIDELINE Stability testing of active pharmaceutical ingredients and finished pharmaceutical products

Bioanalytical method validation: An updated review

API Stability Protocols and. Chris Byrne Tasmanian Alkaloids

ICH Topic M 4 Q Location issues for Common Technical Document for the Registration of Pharmaceuticals for Human Use Quality Questions and Answers

Reflection paper on the dissolution specification for generic solid oral immediate release products with systemic action

Questions and answers on the 'Guideline on the limits of genotoxic impurities'

Statistical Evaluation Of Stability Data

Application note. Guideline for validation of analytical methods using Cedex Bio, Cedex Bio HT, and Cedex HiRes Analyzers.

IMPURITIES IN NEW DRUG PRODUCTS

COMMITTEE FOR MEDICINAL PRODUCTS FOR HUMAN USE (CHMP)

EVALUATION FOR STABILITY DATA

GUIDELINES ON THE STABILITY DATA REQUIRED FOR REGISTRATION OF STOCK REMEDIES IN SOUTH AFRICA

Examples of regulatory expectations for analytical characterization and testing

Guideline on requirements for the production and control of immunological veterinary medicinal products

Guidance for individual laboratories for transfer of quality control methods validated in collaborative trials with a view to implementing 3Rs

Guideline on the chemistry of active substances for veterinary medicinal products

Reflection paper on the requirements for selection and justification of starting materials for the manufacture of chemical active substances

International Journal of Pharma and Bio Sciences DEVELOPMENT OF ACCELERATED STABILITY PROTOCOL FOR SILDENAFIL TABLETS A EUROPEAN PERSPECTIVE REVIEW

COMMITTEE FOR MEDICINAL PRODUCTS FOR HUMAN USE (CHMP)

Fusion Analytical Method Validation

New Drug Product Impurities

Guideline on the requirements for quality documentation concerning biological investigational medicinal products in clinical trials

GUIDELINE FOR THE STABILITY TESTING

COMMITTEE FOR VETERINARY MEDICINAL PRODUCTS NOTE FOR GUIDANCE:

Adoption by CHMP for release for 3-month public consultation 18 November End of consultation (deadline for comments) 28 February 2011

Size Exclusion Chromatography of Biosimilar and Innovator Insulin Using the Agilent AdvanceBio SEC column

EUROPEAN COMMISSION ENTERPRISE AND INDUSTRY DIRECTORATE-GENERAL. EudraLex The Rules Governing Medicinal Products in the European Union

ICH Guideline S3A: Note for guidance on toxicokinetics: the assessment of systemic exposure in toxicity studies - questions and answers

COMMITTEE FOR MEDICINAL PRODUCTS FOR HUMAN USE (CHMP) GUIDELINE ON THE NON-CLINICAL DEVELOPMENT OF FIXED COMBINATIONS OF MEDICINAL PRODUCTS

BIOPHARMACEUTICAL PROCESS EVALUATED FOR VIRAL CLEARANCE

Validation of a concentration assay using Biacore C

1225 VALIDATION OF lytical performance characteristics that should be considered COMPENDIAL PROCEDURES

Official Letter from the DOH

BRIEFING. . Over time, 466 may be used less frequently and may be withdrawn.

Variation Regulations (EU)

Draft agreed by Scientific Advice Working Party 5 September Adopted by CHMP for release for consultation 19 September

Identification of the need and uses of a reference standard

Understanding and accounting for product

Method Development and Validation for Online UV-Dissolution Methods Using Fiber-Optic Technology

COMMITTEE FOR VETERINARY MEDICINAL PRODUCTS NOTE FOR GUIDANCE 1 : DNA VACCINES NON-AMPLIFIABLE IN EUKARYOTIC CELLS FOR VETERINARY USE

Tony Mire-Sluis Vice President, Corporate, Product and Device Quality Amgen Inc

A2LA. R231 Specific Requirements: Threat Agent Testing Laboratory Accreditation Program. December 6, 2017

Analytical Methods Development and Validation

STABILITY TESTING OF NEW DRUG

ICH guideline Q12 on technical and regulatory considerations for pharmaceutical product lifecycle management

A GUIDELINE ON DOSSIER REQUIREMENTS FOR TYPE I VARIATIONS November 1999

Structure and content of an IMPD. What is required for first into man trial?

QUESTIONS AND ANSWERS TO ICH S3A: NOTE FOR GUIDANCE ON TOXICOKINETICS: THE ASSESSMENT OF SYSTEMIC EXPOSURE IN TOXICITY STUDIES FOCUS ON MICROSAMPLING

Lifecycle Management of Process Analytical Technology Procedures

QUANLYNX APPLICATION MANAGER

Application Note. Authors. Abstract

TECHNICAL AND REGULATORY CONSIDERATIONS FOR PHARMACEUTICAL PRODUCT LIFECYCLE MANAGEMENT Q12

Current Features of USFDA and EMA Process Validation Guidance

Quantitative determination of residual 2-(2-chloroethoxy) ethanol (CEE) in quetiapine fumarate by gas chromatogaraphy

Interested parties (organisations or individuals) that commented on the draft document as released for consultation.

Available Online through (or) IJPBS Volume 2 Issue 3 JULY-SEPT Research Article Pharmaceutical Sciences

Recommendation on elements required to support the medical plausibility and the assumption of significant benefit for an orphan designation

STABILITY TESTING: PHOTOSTABILITY TESTING OF NEW DRUG SUBSTANCES AND PRODUCTS

PHARMACEUTICAL TESTING

ICH guideline Q12 on technical and regulatory considerations for pharmaceutical product lifecycle management - Annexes

Organisation de Coopération et de Développement Économiques Organisation for Economic Co-operation and Development

Research Paper Development of Stability Indicating Reverse Phase HPLC Method for Aripiprazole from Solid Dosage form

Application of PAT for Tablet Analysis. Case examples from Novartis Lorenz Liesum, Lead PAT Hamburg, 19 th of April 2013

TEST PROCEDURES AND ACCEPTANCE CRITERIA FOR NEW VETERINARY DRUG SUBSTANCES AND NEW MEDICINAL PRODUCTS: CHEMICAL SUBSTANCES

Draft agreed by SWP, vswp and GMP/GDP Inspectors WG September Adopted by CVMP for release for consultation 8 November 2016

Composition Analysis of Animal Feed by HR ICP-OES

Developing and Validating Dissolution Procedures for Improved Product Quality

Validation of Thin Layer Chromatographic Procedures

MASTER FILE PROCEDURES

Reflection paper on co-development of pharmacogenomic biomarkers and Assays in the context of drug development

Proposed New USP General Chapter: The Analytical Procedure Lifecycle 1220

Essentials in Stability Analysis and Expiry Determination

USP Chapter 823 USP 32 (old) vs. USP 35 (new)

COMMITTEE FOR THE MEDICINAL PRODUCT FOR HUMAN USE (CHMP)

The rules governing medicinal products in the European Union. Presentation and content of the dossier Edition

Further Stability Considerations

STANDARD OPERATING PROCEDURES. Handling and working with Analytical Standards

Questions and answers on 'Guideline on the environmental risk assessment of medicinal products for human use'

European Medicines Agency practical guidance on the application form for centralised type IA and IB variations

PROPOSAL FOR REVISION OF THE SUPPLEMENTARY GUIDELINES ON GOOD MANUFACTURING PRACTICES: VALIDATION, APPENDIX 7:NON-STERILE PROCESS VALIDATION

Roles and responsibilities of members and alternates, rapporteur and peer reviewers, experts and observers of the Paediatric Committee (PDCO)

Setting Specifications for Biotech Products

Manual 058 Out of Specification Results Investigation

PHARMACEUTICAL REFERENCE STANDARDS

Investigating OOS for Finished Product on the Stability Program. Presented by: Nicole Chang, QA Manager, Apotex Pty Ltd

Pharmacopoeial Reference Standards

Validation of an Endpoint Enzyme Activity Assay to Evaluate Potency for Lot Release and Stability Testing. Loc Vo, PhD

Development of non-substantially manipulated cell-based ATMPs 1 : flexibility introduced via the application of the risk-based approach

IMPURITIES. Antony Fake API Focal Point, PQTm WHO PREQUALIFICATION TEAM MEDICINES

Validated Stability-indicating assay method for determination of Ilaprazole in bulk drug and tablets by high performance liquid chromatography

APPLICATION NOTE. UOP : Analysis of Trace CO and CO 2 in bulk H 2 and Light Gaseous Hydrocarbons by GC

HPLC METHODOLOGY MANUAL

Guidance Document 01 January 2016 CONTENTS. 1. Introduction Background 1.2. Objectives 1.3. Scope and application 1.4 APIMF holder obligations

European Medicines Agency Evaluation of Medicines for Human Use COMMITTEE FOR MEDICINAL PRODUCTS FOR HUMAN USE (CHMP) DRAFT

ANNEX V ASEAN GUIDELINES ON STABILITY STUDY AND SHELF-LIFE OF HEALTH SUPPLEMENTS

Transcription:

The European Agency for the Evaluation of Medicinal Products Veterinary Medicines Evaluation Unit CVMP/VICH/591/98-FINAL London, 10 December 1998 VICH Topic GL2 (Validation: Methodology) Step 7 Consensus Guideline GUIDELINE ON VALIDATION OF ANALYTICAL PROCEDURES: METHODOLOGY TRANSMISSION TO CVMP December 1997 TRANSMISSION TO INTERESTED PARTIES January 1998 COMMENTS REQUESTED BEFORE April 1998 FINAL APPROVAL BY CVMP December 1998 DATE FOR COMING INTO OPERATION BY October 1999 7 Westferry Circus, Canary Wharf, London E14 4HB, UK Switchboard: (+44-171) 418 8400 Fax: (+44-171) 418 8447 E_Mail: mail@emea.eudra.org http://www.eudra.org/emea.html Reproduction and/or distribution of this document is authorised for non-commercial purposes only provided the EMEA is acknowledged

VICH GL2 (VALIDATION METHODOLOGY) October 1998 For implementation Step 7 VALIDATION OF ANALYTICAL PROCEDURES: METHODOLOGY Recommended for Implementation at Step 7 of the VICH Process on 22 October 1998 by the VICH Steering Committee THIS GUIDELINE HAS BEEN DEVELOPED BY THE APPROPRIATE VICH EXPERT WORKING GROUP ON THE BASIS OF THE ICH GUIDELINES ON THE SAME SUBJECT AND HAS BEEN SUBJECT TO CONSULTATION BY THE PARTIES, IN ACCORDANCE WITH THE VICH PROCESS. AT STEP 7 OF THE PROCESS THE FINAL DRAFT IS RECOMMENDED FOR ADOPTION TO THE REGULATORY BODIES OF THE EUROPEAN UNION, JAPAN AND USA. FOR IMPLEMENTATION AT STEP 7-22/10/1998 Page 1 of 10

VALIDATION OF ANALYTICAL PROCEDURES: METHODOLOGY TABLE OF CONTENTS INTRODUCTION...3 1. SPECIFICITY...3 1.1. Identification...3 1.2. Assay and Impurity Test(s)...4 2. Linearity...4 3. Range...5 4. Accuracy...5 4.1. Assay... 6 4.2. Impurities (Quantitation)...6 4.3. Recommended Data...6 5. Precision... 6 5.1. Repeatability...7 5.2. Intermediate Precision...7 5.3. Reproducibility...7 5.4. Recommended Data...7 6. DETECTION LIMIT...7 6.1. Based on Visual Evaluation...7 6.2. Based on Signal-to-Noise...7 6.3 Based on the Standard Deviation of the Response and the Slope...7 6.4 Recommended Data...8 7. QUANTITATION LIMIT...8 7.1. Based on Visual Evaluation...8 7.2. Based on Signal-to-Noise Approach...8 7.3. Based on the Standard Deviation of the Response and the Slope...8 7.4 Recommended Data...9 8. ROBUSTNESS...9 9. SYSTEM SUITABILITY TESTING...9 FOR IMPLEMENTATION AT STEP 7-22/10/1998 Page 2 of 10

VALIDATION OF ANALYTICAL PROCEDURES: METHODOLOGY INTRODUCTION This document is complementary to the parent document which presents a discussion of the characteristics that should be considered during the validation of analytical procedures. Its purpose is to provide some guidance and recommendations on how to consider the various validation characteristics for each analytical procedure. In some cases (for example, demonstration of specificity), the overall capabilities of a number of analytical procedures in combination may be investigated in order to ensure the quality of the drug substance or drug product. In addition, the document provides an indication of the data which should be presented in a registration application. All relevant data collected during validation and formulae used for calculating validation characteristics should be submitted and discussed as appropriate. Approaches other than those set forth in this guideline may be applicable and acceptable. It is the responsibility of the applicant to choose the validation procedure and protocol most suitable for their product. However it is important to remember that the main objective of validation of an analytical procedure is to demonstrate that the procedure is suitable for its intended purpose. Due to their complex nature, analytical procedures for biological and biotechnological products in some cases may be approached differently than in this document. Well-characterized reference materials, with documented purity, should be used throughout the validation study. The degree of purity necessary depends on the intended use. In accordance with the parent document, and for the sake of clarity, this document considers the various validation characteristics in distinct sections. The arrangement of these sections reflects the process by which an analytical procedure may be developed and evaluated. In practice, it is usually possible to design the experimental work such that the appropriate validation characteristics can be considered simultaneously to provide a sound, overall knowledge of the capabilities of the analytical procedure, for instance: specificity, linearity, range, accuracy and precision. 1. SPECIFICITY An investigation of specificity should be conducted during the validation of identification tests, the determination of impurities and the assay. The procedures used to demonstrate specificity will depend on the intended objective of the analytical procedure. It is not always possible to demonstrate that an analytical procedure is specific for a particular analyte (complete discrimination). In this case a combination of two or more analytical procedures is recommended to achieve the necessary level of discrimination. 1.1. Identification Suitable identification tests should be able to discriminate between compounds of closely related structures which are likely to be present. The discrimination of a procedure may be confirmed by obtaining positive results (perhaps by comparison with a known reference material) from samples containing the analyte, coupled with negative results from samples which do not contain the analyte. In addition, the identification test may be applied to materials structurally similar to or closely related to the analyte to confirm that a positive FOR IMPLEMENTATION AT STEP 7-22/10/1998 Page 3 of 10

response is not obtained. The choice of such potentially interfering materials should be based on sound scientific judgement with a consideration of the interferences that could occur. 1.2. Assay and Impurity Test(s) For chromatographic procedures, representative chromatograms should be used to demonstrate specificity and individual components should be appropriately labelled. Similar considerations should be given to other separation techniques. Critical separations in chromatography should be investigated at an appropriate level. For critical separations, specificity can be demonstrated by the resolution of the two components which elute closest to each other. In cases where a non-specific assay is used, other supporting analytical procedures should be used to demonstrate overall specificity. For example, where a titration is adopted to assay the drug substance for release, the combination of the assay and a suitable test for impurities can be used. The approach is similar for both assay and impurity tests: 1.2.1 Impurities are available For the assay, this should involve demonstration of the discrimination of the analyte in the presence of impurities and/or excipients; practically, this can be done by spiking pure substances (drug substance or drug product) with appropriate levels of impurities and/or excipients and demonstrating that the assay result is unaffected by the presence of these materials (by comparison with the assay result obtained on unspiked samples). For the impurity test, the discrimination may be established by spiking drug substance or drug product with appropriate levels of impurities and demonstrating the separation of these impurities individually and/or from other components in the sample matrix. 1.2.2 Impurities are not available If impurity or degradation product standards are unavailable, specificity may be demonstrated by comparing the test results of samples containing impurities or degradation products to a second well-characterized procedure e.g.: pharmacopoeial method or other validated analytical procedure (independent procedure). As appropriate, this should include samples stored under relevant stress conditions: light, heat, humidity, acid/base hydrolysis and oxidation. - for the assay, the two results should be compared. - for the impurity tests, the impurity profiles should be compared. Peak purity tests may be useful to show that the analyte chromatographic peak is not attributable to more than one component (e.g., diode array, mass spectrometry). 2. LINEARITY A linear relationship should be evaluated across the range (see section 3) of the analytical procedure. It may be demonstrated directly on the drug substance (by dilution of a standard stock solution) and/or separate weighings of synthetic mixtures of the drug product components, using the proposed procedure. The latter aspect can be studied during investigation of the range. Linearity should be evaluated by visual inspection of a plot of signals as a function of analyte concentration or content. If there is a linear relationship, test results should be FOR IMPLEMENTATION AT STEP 7-22/10/1998 Page 4 of 10

evaluated by appropriate statistical methods, for example, by calculation of a regression line by the method of least squares. In some cases, to obtain linearity between assays and sample concentrations, the test data may need to be subjected to a mathematical transformation prior to the regression analysis. Data from the regression line itself may be helpful to provide mathematical estimates of the degree of linearity. The correlation coefficient, y-intercept, slope of the regression line and residual sum of squares should be submitted. A plot of the data should be included. In addition, an analysis of the deviation of the actual data points from the regression line may also be helpful for evaluating linearity. Some analytical procedures, such as immunoassays, do not demonstrate linearity after any transformation. In this case, the analytical response should be described by an appropriate function of the concentration (amount) of an analyte in a sample. For the establishment of linearity, a minimum of 5 concentrations is recommended. Other approaches should be justified. 3. RANGE The specified range is normally derived from linearity studies and depends on the intended application of the procedure. It is established by confirming that the analytical procedure provides an acceptable degree of linearity, accuracy and precision when applied to samples containing amounts of analyte within or at the extremes of the specified range of the analytical procedure. The following minimum specified ranges should be considered: - for the assay of a drug substance or a finished (drug) product: normally from 80 to 120 percent of the test concentration; - for content uniformity, covering a minimum of 70 to 130 percent of the test concentration, unless a wider more appropriate range, based on the nature of the dosage form, is justified; - for dissolution testing: +/-20 % over the specified range; e.g., if the specifications for a controlled released product cover a region from 20%, after 1 hour, up to 90%, after 24 hours, the validated range would be 0-110% of the label claim. - for the determination of an impurity: from the reporting level of an impurity 1 to 120% of the specification; for impurities known to be unusually potent or to produce toxic or unexpected pharmacological effects, the detection/quantitation limit should be commensurate with the level at which the impurities must be controlled. Note: for validation of impurity test procedures carried out during development, it may be necessary to consider the range around a suggested (probable) limit; - if assay and purity are performed together as one test and only a 100% standard is used, linearity should cover the range from the reporting level of the impurities 1 to 120% of the assay specification; 4. ACCURACY Accuracy should be established across the specified range of the analytical procedure. 1 see chapters Reporting Impurity Content of Batches of the corresponding VICH- Guidelines: Impurities in New Drug Substances and Impurities in New Drug Products FOR IMPLEMENTATION AT STEP 7-22/10/1998 Page 5 of 10

4.1. Assay 4.1.1 Drug Substance Several methods of determining accuracy are available: a) application of an analytical procedure to an analyte of known purity (e.g. reference material); b) comparison of the results of the proposed analytical procedure with those of a second well-characterized procedure, the accuracy of which is stated and/or defined (independent procedure, see 1.2.); c) accuracy may be inferred once precision, linearity and specificity have been established. 4.1.2 Drug Product Several methods for determining accuracy are available: a) application of the analytical procedure to synthetic mixtures of the drug product components to which known quantities of the drug substance to be analysed have been added; b) in cases where it is impossible to obtain samples of all drug product components, it may be acceptable either to add known quantities of the analyte to the drug product or to compare the results obtained from a second well-characterized procedure, the accuracy of which is stated and/or defined (independent procedure, see 1.2.). c) accuracy may be inferred once precision, linearity and specificity have been established. 4.2. Impurities (Quantitation) Accuracy should be assessed on samples (drug substance/drug product) spiked with known amounts of impurities. In cases where it is impossible to obtain samples of certain impurities and/or degradation products, it is considered acceptable to compare results obtained by an independent procedure (see 1.2.). The response factor of the drug substance can be used. It should be clear how the individual or total impurities are to be determined e.g., weight/weight or area percent, in all cases with respect to the major analyte. 4.3. Recommended Data Accuracy should be assessed using a minimum of 9 determinations over a minimum of 3 concentration levels covering the specified range (e.g. 3 concentrations/3 replicates each of the total analytical procedure). Accuracy should be reported as percent recovery by the assay of known added amount of analyte in the sample or as the difference between the mean and the accepted true value together with the confidence intervals. 5. PRECISION Validation of tests for assay and for quantitative determination of impurities includes an investigation of precision. FOR IMPLEMENTATION AT STEP 7-22/10/1998 Page 6 of 10

5.1. Repeatability Repeatability should be assessed using: a) a minimum of 9 determinations covering the specified range for the procedure (e.g. 3 concentrations/3 replicates each) or b) a minimum of 6 determinations at 100% of the test concentration. 5.2. Intermediate Precision The extent to which intermediate precision should be established depends on the circumstances under which the procedure is intended to be used. The applicant should establish the effects of random events on the precision of the analytical procedure. Typical variations to be studied include days, analysts, equipment, etc. It is not considered necessary to study these effects individually. The use of an experimental design (matrix) is encouraged. 5.3. Reproducibility Reproducibility is assessed by means of an inter-laboratory trial. Reproducibility should be considered in case of the standardization of an analytical procedure, for instance, for inclusion of procedures in pharmacopoeias. These data are not part of the marketing authorization dossier. 5.4. Recommended Data The standard deviation, relative standard deviation (coefficient of variation) and confidence interval should be reported for each type of precision investigated. 6. DETECTION LIMIT Several approaches for determining the detection limit are possible, depending on whether the procedure is a non-instrumental or instrumental. Approaches other than those listed below may be acceptable. 6.1. Based on Visual Evaluation Visual evaluation may be used for non-instrumental methods but may also be used with instrumental methods. The detection limit is determined by the analysis of samples with known concentrations of analyte and by establishing the minimum level at which the analyte can be reliably detected. 6.2. Based on Signal-to-Noise This approach can only be applied to analytical procedures which exhibit baseline noise. Determination of the signal-to-noise ratio is performed by comparing measured signals from samples with known low concentrations of analyte with those of blank samples and establishing the minimum concentration at which the analyte can be reliably detected. A signal-to-noise ratio between 3 or 2:1 is generally considered acceptable for estimating the detection limit. 6.3 Based on the Standard Deviation of the Response and the Slope The detection limit (DL) may be expressed as: DL = 3.3 σ S FOR IMPLEMENTATION AT STEP 7-22/10/1998 Page 7 of 10

where σ = the standard deviation of the response S = the slope of the calibration curve The slope S may be estimated from the calibration curve of the analyte. The estimate of σ may be carried out in a variety of ways, for example: 6.3.1 Based on the Standard Deviation of the Blank Measurement of the magnitude of analytical background response is performed by analyzing an appropriate number of blank samples and calculating the standard deviation of these responses. 6.3.2 Based on the Calibration Curve A specific calibration curve should be studied using samples containing an analyte in the range of DL. The residual standard deviation of a regression line or the standard deviation of y-intercepts of regression lines may be used as the standard deviation. 6.4 Recommended Data The detection limit and the method used for determining the detection limit should be presented. If DL is determined based on visual evaluation or based on signal to noise ratio, the presentation of the relevant chromatograms is considered acceptable for justification. In cases where an estimated value for the detection limit is obtained by calculation or extrapolation, this estimate may subsequently be validated by the independent analysis of a suitable number of samples known to be near or prepared at the detection limit. 7. QUANTITATION LIMIT Several approaches for determining the quantitation limit are possible, depending on whether the procedure is a non-instrumental or instrumental. Approaches other than those listed below may be acceptable. 7.1. Based on Visual Evaluation Visual evaluation may be used for non-instrumental methods but may also be used with instrumental methods. The quantitation limit is generally determined by the analysis of samples with known concentrations of analyte and by establishing the minimum level at which the analyte can be quantified with acceptable accuracy and precision. 7.2. Based on Signal-to-Noise Approach This approach can only be applied to analytical procedures that exhibit baseline noise. Determination of the signal-to-noise ratio is performed by comparing measured signals from samples with known low concentrations of analyte with those of blank samples and by establishing the minimum concentration at which the analyte can be reliably quantified. A typical signal-to-noise ratio is 10:1. 7.3. Based on the Standard Deviation of the Response and the Slope The quantitation limit (QL) may be expressed as: 10 σ QL = S where σ = the standard deviation of the response S = the slope of the calibration curve FOR IMPLEMENTATION AT STEP 7-22/10/1998 Page 8 of 10

The slope S may be estimated from the calibration curve of the analyte. The estimate of σ may be carried out in a variety of ways for example: 7.3.1 Based on Standard Deviation of the Blank Measurement of the magnitude of analytical background response is performed by analyzing an appropriate number of blank samples and calculating the standard deviation of these responses. 7.3.2 Based on the Calibration Curve A specific calibration curve should be studied using samples, containing an analyte in the range of QL. The residual standard deviation of a regression line or the standard deviation of y-intercepts of regression lines may be used as the standard deviation. 7.4 Recommended Data The quantitation limit and the method used for determining the quantitation limit should be presented. The limit should be subsequently validated by the analysis of a suitable number of samples known to be near or prepared at the quantitation limit. 8. ROBUSTNESS The evaluation of robustness should be considered during the development phase and depends on the type of procedure under study. It should show the reliability of an analysis with respect to deliberate variations in method parameters. If measurements are susceptible to variations in analytical conditions, the analytical conditions should be suitably controlled or a precautionary statement should be included in the procedure. One consequence of the evaluation of robustness should be that a series of system suitability parameters (e.g., resolution test) is established to ensure that the validity of the analytical procedure is maintained whenever used. Examples of typical variations are: - stability of analytical solutions, - extraction time In the case of liquid chromatography, examples of typical variations are - influence of variations of ph in a mobile phase, - influence of variations in mobile phase composition, - different columns (different lots and/or suppliers), - temperature, - flow rate. In the case of gas-chromatography, examples of typical variations are - different columns (different lots and/or suppliers), - temperature, - flow rate. 9. SYSTEM SUITABILITY TESTING System suitability testing is an integral part of many analytical procedures. The tests are based on the concept that the equipment, electronics, analytical operations and samples to be analyzed constitute an integral system that can be evaluated as such. System suitability test parameters to be established for a particular procedure depend FOR IMPLEMENTATION AT STEP 7-22/10/1998 Page 9 of 10

on the type of procedure being validated. See Pharmacopoeias for additional information. FOR IMPLEMENTATION AT STEP 7-22/10/1998 Page 10 of 10