Application Note. Genomics. Abstract. Authors. Kirill Gromadski Ruediger Salowsky Susanne Glueck Agilent Technologies Waldbronn, Germany

Similar documents
Performance characteristics of the High Sensitivity DNA kit for the Agilent 2100 Bioanalyzer

Evaluating the Agilent 4200 TapeStation System for High Throughput Sequencing Quality Control

Implementation of Automated Sample Quality Control in Whole Exome Sequencing

Evaluation with RNA ScreenTape Assays

Comparing the Agilent 2100 Bioanalyzer Performance to Traditional DNA Analysis Techniques

TruSeq ChIP Sample Preparation

Quantitative analysis of PCR fragments with the Agilent 2100 Bioanalyzer. Application Note. Odilo Mueller. Abstract

User Manual. NGS Library qpcr Quantification Kit (Illumina compatible)

Agilent Bioanalyzer and TapeStation Systems. systems, consumables and supplies

KAPA Library Preparation Kit Ion Torrent Platforms

Technical Review. Real time PCR

Optimizing FFPE DNA preparation for use in SureSelect XT2

RIPTIDE HIGH THROUGHPUT RAPID LIBRARY PREP (HT-RLP)

Electrophoresis and the Agilent Bioanalyzer. Advanced Biotechnology Lab I Florida Atlantic University January 23, 2008

Lab methods: Exome / Genome. Ewart de Bruijn

Next Generation Sequencing. Target Enrichment

Automated size selection of NEBNext Small RNA libraries with the Sage Pippin Prep

Non-Organic-Based Isolation of Mammalian microrna using Norgen s microrna Purification Kit

NEBNext. for Ion Torrent LIBRARY PREPARATION KITS

Single Cell Genomics

LabChip GXII: Antibody Analysis

User Bulletin. Veriti 96-Well Thermal Cycler AmpFlSTR Kit Validation. Overview

Whole genome Bisulfite Sequencing for Methylation Analysis Preparing Samples for the Illumina Sequencing Platform

Hybridization capture of DNA libraries using xgen Lockdown Probes and Reagents

E.Z.N.A. Size Select-IT Kit. D preps D preps D preps

Fast and reproducible amplicon library preparation for HLA typing with the 454 sequencing system

XactEdit Cas9 Nuclease with NLS User Manual

Cancer Genetics Solutions

Ion AmpliSeq Exome RDY Library Preparation

KAPA Library Amplification Kit Illumina Platforms

Gene Regulation Solutions. Microarrays and Next-Generation Sequencing

Amplicon Sequencing Template Preparation

SureSelect Target Enrichment System for Illumina Paired-End Sequencing Library

QuantiFluor ONE dsdna System

Amplicon Library Preparation Method Manual. GS FLX Titanium Series October 2009

Absolute Human Telomere Length Quantification qpcr Assay Kit (AHTLQ) Catalog # reactions

Certificate of Analysis of the Holotype HLA 24/7 Configuration A1 & CE

NEBNext Magnesium RNA Fragmentation Module

Thermo Scientific GeneJET NGS Cleanup Kit

One platform, endless possibilities

BenchSmart 96. Semi-automated Pipetting Higher Accuracy, Greater Flexibility

3C bioscience. ProbeSure Genotyping Master Mix User Guide. P a g e 1 ProbeSure User Guide v1.0

Improved Chemistry for NGS Library Cleanup and Size Selection Speakers: Charles Cowles, PhD & Curtis Knox

Quantifiler Human DNA Quantification Kit Quantifiler Y Human Male DNA Quantification Kit

TECHNICAL BULLETIN. SeqPlex DNA Amplification Kit for use with high throughput sequencing technologies. Catalog Number SEQX Storage Temperature 20 C

Welcome to the NGS webinar series

QIAGEN Whole Genome Amplification REPLI-g Eliminating Sample Limitations, Potential Use for Reference Material

Application Note Detecting low copy numbers. Introduction. Methods A08-005B

Applications for DNA, RNA, and Protein Analysis

Gene Expression Technology

SOLiD Total RNA-Seq Kit SOLiD RNA Barcoding Kit

KAPA HiFi Real-Time PCR Library Amplification Kit

Introduction. Technical Note

SureSelect Target Enrichment System for Roche 454 GS FLX and GS Junior Sequencing Platforms

CRISPR/Cas9 Gene Editing

Single Cell 3 Reagent Kits v2 Quick Reference Cards

qpcr Quantitative PCR or Real-time PCR Gives a measurement of PCR product at end of each cycle real time

Laboratory #7 PCR PCR

T7-Based RNA Amplification Protocol (in progress)

KAPA Human Genomic DNA Quantification and QC Kit

SPRIworks Systems. Push button. Walk away. Fully automated library construction systems with built-in size selection and cleanup BR-15981B

TaqPath ProAmp Master Mixes

DNA SMART ChIP- Seq Kit User Manual

Accel-NGS 2S DNA Library Kit for Illumina Platforms

Generating Exome Enriched Sequencing Libraries from Formalin-Fixed, Paraffin-Embedded Tissue DNA for Next-Generation Sequencing

DNA and RNA Clean-up Guide

Stratagene QPCR Human Reference Total RNA

TruSeq Sample Preparation Best Practices and Troubleshooting Guide

Bacterial 16S rrna V1-3 Amplicon Sequencing Standard Protocol

Echo-Enhanced SMART-Seq v4 for RNA Sequencing

Supplemental Materials. DNA preparation. Dehalogenimonas lykanthroporepellens strain BL-DC-9 T (=ATCC

Principals of Real-Time PCR. Amira A. T. AL-Hosary Lecturer of Infectious Diseases, Faculty of Veterinary Medicine, Assiut University, Egypt

P HENIX. PHENIX PCR Enzyme Guide Tools For Life Science Discovery RESEARCH PRODUCTS

High Cross-Platform Genotyping Concordance of Axiom High-Density Microarrays and Eureka Low-Density Targeted NGS Assays

KAPA Library Preparation Kit with Real-time PCR Library Amplification for Illumina Platforms

Targeted Sequencing Using Droplet-Based Microfluidics. Keith Brown Director, Sales

Next Gen Sequencing. Expansion of sequencing technology. Contents

Accel-NGS 2S Plus DNA Library Kit NGS Prep with Single Indexing and Molecular Identifiers (MIDs) Cat. No /21096

Myers Lab ChIP-seq Protocol v Modified January 10, 2014

MHC Region. MHC expression: Class I: All nucleated cells and platelets Class II: Antigen presenting cells

Brilliant II SYBR Green QPCR Master Mix

Targeted Sequencing in the NBS Laboratory

Lecture Four. Molecular Approaches I: Nucleic Acids

FFPE DNA Extraction kit

NEBNext Multiplex Small RNA Library Prep Set for Illumina (Set 1)

Investigator Quantiplex HYres Kit

QuantStudio 3D Digital PCR System

Lecture 8: Sequencing and SNP. Sept 15, 2006

Methods of Biomaterials Testing Lesson 3-5. Biochemical Methods - Molecular Biology -

Uncertainties and certainties in GMO analytics using qpcr

Using Low Input of Poly (A) + RNA and Total RNA for Oligonucleotide Microarrays Application

TruePrime Single Cell WGA Kit

1. COMPONENTS. PyroStart Fast PCR Master Mix (2X) (#K0211 for 250 reactions of 20µl) 2. STORAGE 3. DESCRIPTION

Introductory Next Gen Workshop

SureSelect Target Enrichment System

Getting high-quality cytogenetic data is a SNP.

KAPA Adapter Kits Ion Torrent Platforms

HiPer Real-Time PCR Teaching Kit

High-Resolution Oligonucleotide- Based acgh Analysis of Single Cells in Under 24 Hours

Transcription:

Improving sample quality for target enrichment and next-gen sequencing with the Agilent High Sensitivity DNA Kit and the Agilent SureSelect Target Enrichment Platform Application Note Genomics Authors Kirill Gromadski Ruediger Salowsky Susanne Glueck Agilent Technologies Waldbronn, Germany Abstract Next-generation sequencing (NGS) has revolutionized the genetic landscape. It is a lengthy, labor-intensive process that yields results never before achieved. As a result, it is imperative that the quality of the DNA sample be evaluated from the start, as most NGS sample preparation protocols require PCR amplification to generate DNA libraries prior to sequencing. The likelihood of artifact generation could contribute to bias, affecting the potential results. The High Sensitivity DNA Kit used with the Agilent 2100 Bioanalyzer has been optimized with improved levels of detection. The improved sensitivity allows the numbers of library PCR cycles to be reduced, removing amplification bias and significantly improving the quality of NGS data with increased accuracy. This Application Note describes how the Agilent 2100 Bioanalyzer High Sensitivity DNA Kit can be used to provide quantitative and qualitative information about the DNA samples used in the Agilent SureSelect Target Enrichment System.

Introduction The Agilent 2100 Bioanalyzer, an automated on-chip electrophoresis system, has already proven to be a valuable tool for automated sizing and quantification of various doublestranded DNA sample types relevant for the next-generation sequencing (NGS) sample preparation workflow. 1 The Agilent 2100 Bioanalyzer with the DNA 1000 kit is recommended by NGS platform providers for measuring DNA sample quality prior to sequencing runs. These quality checks reduce time and resources wasted by lowquality samples. Recently, a High Sensitivity DNA kit was developed which offers improved sensitivity for checking the size and quantity of precious low concentrated DNA starting material or DNA libraries down to a concentration of 100 pg/µl. Next-generation sequencing technology has brought high throughput to genome sequencing, but the new processes lack the ability to target specific areas of a genome. The SureSelect Target Enrichment System, enables genomic areas of interest to be sequenced exclusively. This creates process efficiencies that reduce costs and allow more samples to be analyzed per study. 2 The Agilent High Sensitivity DNA Kit and the Agilent 2100 Bioanalyzer can be used for quality control at several steps during the SureSelect Target Enrichment workflow. During the sample preparation, the Agilent 2100 Bioanalyzer is used for quality control and sizing selection of the sheared genomic DNA, and to assess the quality and size distribution of the PCR amplified sequencing library DNA. After post-hybridization amplification, the Agilent 2100 Bioanalyzer can be used to determine the quality and the concentration of the PCR-amplified capture DNA before sequencing. This Application Note describes how the High Sensitivity DNA kit and the Agilent 2100 Bioanalyzer can be used before sequencing to reduce the number of required PCR cycles. This reduces amplification bias, thus improving the quality of DNA libraries created during the SureSelect Target Enrichment workflow. Experimental DNA library preparation The DNA library was prepared for Illumina s Genome Analyzer II sequencers according to manufacturer s instructions. SureSelect Target Enrichment The SureSelect Target Enrichment for the Illumina single-end sequencing platform, consisting of three main steps; sample preparation, hybridization and post hybridization amplification, was carried out as described in the manual. 3 16 DNA samples obtained after the post-hybridization amplification with different numbers of PCR cycles (4-18) were used for DNA analysis with the Agilent 2100 Bioanalyzer. High Sensitivity DNA analysis with the Agilent 2100 Bioanalyzer The on-chip DNA electrophoresis was performed on the Agilent 2100 Bioanalyzer in combination with the Agilent High Sensitivity DNA kit, according to the High Sensitivity DNA kit guide. 4 A dedicated High Sensitivity DNA assay is available with the Agilent 2100 Expert software (revision B.02.07 or higher). An integration region from 100 to 2000 bp was used for all samples for smear quantification. DNA quantification In addition to the fluorescence-based DNA quantification on the Agilent 2100 Bioanalyzer, the Qubit fluorometer and the Qubit Quant-iT dsdna BR Assay kit were used for DNA quantification according to the manufacturer s instructions. 2

Results and discussion This Application Note describes how the Agilent High Sensitivity DNA kit and the Agilent 2100 Bioanalyzer can be used to further improve the quality of DNA sequencing libraries enriched by the SureSelect kit. For this purpose, 16 amplified and purified DNA samples from the post-hybridization PCR amplification step were analyzed with the High Sensitivity DNA kit and the 2100 Bioanalyzer prior to sequencing on the Illumina platform. Figure 1 shows electropherograms of typical PCR amplified DNA libraries. The electropherograms show a typical smear from 150 to 350 nucleotides. The primers/primer-dimers migrated very close to the lower marker, but did not affect the analysis. The excellent sensitivity of the High Sensitivity DNA kit allowed the amplified DNA to be detected and reliably quantified, even after only four PCR cycles (figure 2). As expected, DNA concentration increased with the number of PCR cycles. Above 14 PCR cycles, the increase in DNA concentration was no longer linear and becomes saturated (figure 1B). When using 10 or more cycles, the DNA concentration was outside the quantitative range of the High Sensitivity assay. These samples were diluted with TE buffer in the indicated dilution ratios (figure 1B) prior to the analysis on the Agilent 2100 Bioanalyzer. [FU] 350 300 250 200 150 100 50 0 [FU] 450 400 350 300 250 200 150 100 50 0 B A 35 10 cycles 8 cycles 6 cycles 4 cycles 100 200 300 400 600 1000 2000 10380 [bp] 16 cycles (1:10 dil.) 14 cycles (1:10 dil.) 12 cycles (1:10 dil.) 35 100 200 300 400 500 600 1000 2000 10380 [bp] Figure 1 PCR-amplified DNA library derived from the SureSelect Target Enrichment workflow, analyzed with the High Sensitivity DNA kit. (A) Overlay of DNA electropherograms obtained after 4 to 10 PCR cycles as well as TE buffer blank (black). The number of PCR cycles is indicated in the electropherogram overlay. (B) Overlay of DNA electropherograms obtained after 12 to 18 PCR cycles. The number of PCR cycles and the dilution ratios are indicated in the electropherogram overlay. 3

The key observation clearly shown in figure 1B, is that the quality of the PCR product depended on the number of PCR cycles performed. After 14 PCR cycles, an additional DNA smear at approximately 500 bp was detected in the electropherogram. This PCR artifact could potentially affect the efficiency of an NGS experiment. When running amplifications with PCR cycles below 14 this PCR artifact was not observed. DNA library analysis with the High Sensitivity DNA kit allows the number of required PCR cycles to be reduced. This results in fewer amplification-related artifacts, significantly improving the DNA sample quality for downstream sequencing. It should be expected that the improved DNA sample quality due to the reduced number of PCR cycles will positively affect the outcome of the downstream sequencing by reducing allelic bias, single-stranded DNA generation, and duplicate sequences. 5 The DNA concentrations determined with the High Sensitivity DNA kit were also compared with the DNA concentrations measured with a fluorometer (table 1). The DNA samples obtained after 4 to 10 PCR cycles were measured directly without dilution; all other samples were diluted as indicated in figure 1. The selected assay on the fluorometer permitted DNA quantification only after 10 or more PCR cycles. The High Sensitivity DNA kit provided a reproducible DNA quantification after only four PCR cycles. For added confidence in our results, each DNA sample was measured four times on two different High Sensitivity DNA chips with the Agilent 2100 Bioanalyzer. [FU] 150 100 50 0 35 Region 1 Region 1 4 cycles From To Corrected % of Average Size distribution Concentration Molarity [bp] [bp] area total size [bp] in CV [%] [pg/µl] [pmol/l] 100 2,000 395.8 51 254 12.4 283.33 1,713.8 PCR cycles Fluorometer 2100 Bioanalyzer Deviation from DNA conc. [ng/µl] DNA conc. [ng/µl] fluorometer [%] 4-0.090 ± 0.015 - - 0.083 ± 0.007-6 - 0.304 ± 0.017 - - 0.301 ± 0.010-8 - 1.13 ± 0.07 - - 1.34 ± 0.07-10 3.4 4.49 ± 0.16 32.1 3.4 4.54 ± 0.17 33.8 3.3 3.68 ± 0.22 10.3 4.1 4.03 ± 0.23-1.0 12 13.4 14.5 ± 0.5 8.0 12.5 13.9 ± 0.6 11.8 14 36.4 39.8 ± 3.5 9.4 43.6 51.6 ± 3.3 18.3 18 51.0 63.7 ± 3.4 24.9 61.4 70.0 ± 5.4 14.0 Table 1 Comparison of DNA quantification with a fluorometer and the Agilent 2100 Bioanalyzer. 16 different DNA samples obtained from the post-hybridization amplification step of the SureSelect Target Enrichment workflow were quantified with the fluorometer and the Agilent 2100 Bioanalyzer. The DNA samples were measured directly (4 to 10 PCR cycles) or after dilution (12 to 18 cycles) as indicated in figure 1. The standard deviation for the total DNA concentration measured with the Agilent 2100 Bioanalyzer was calculated from four data points measured on two different DNA chips. [ p] 100 150 200 300 400 500 600 10002000 10380 [bp] Figure 2 Electropherogram obtained after 4 PCR cycles. The integration region from 100 to 2000 bp was used for smear quantification. 4

Above 10 PCR cycles, overall DNA concentrations are generally similar for both methods. Variances in the DNA concentration determined with the fluorometer and the on-chip electrophoresis could be due to the differences in technologies. Different fluorescent dyes are used, and the quantification by the Agilent 2100 Bioanalyzer is preceded by an electrophoretic separation of the sample. Additional variances were introduced by diluting the samples. Figure 3 graphically summarizes the data from table 1. The DNA concentrations obtained through both DNA quantification methods were plotted against the number of PCR cycles. Both methods, the fluorometer and the on-chip electrophoresis, clearly show the expected sigmoid amplification rate. The results obtained with both methods are in good agreement with each other. The Agilent 2100 Bioanalyzer provides DNA quantification as well as additional valuable information on the quality of the enriched DNA library. The insert in figure 3 shows the same data in double logarithmic scale to demonstrate the linearity of both methods. The linear dynamic range for the SureSelect DNA samples analyzed with the High Sensitivity DNA assay and the Agilent 2100 Bioanalyzer was determined to be 80 to 5000 pg/µl with r 2 = 0.9996. The last data point, after 18 PCR cycles, was not taken into account for this linearity analysis, as the PCR begins to saturate after 14 cycles. Concentration [pg/µl] 60 40 20 0 Log of concentration [pg/µl] 100 10 1 0.1 0 5 10 15 20 Log of PCR cycles number Agilent 2100 Bioanalyzer Fluorometer 0 5 10 15 20 Number of PCR cycles Figure 3 Comparison of DNA quantification with a fluorometer and the Agilent 2100 Bioanalyzer. The obtained DNA concentrations (table 1) were plotted against the number of PCR cycles, Agilent 2100 Bioanalyzer ( black), fluorometer ( blue). The insert shows the same data in double logarithmic scale to demonstrate the linearity of both methods. 5

Conclusion Quality control of DNA samples after library generation derived from the SureSelect Target Enrichment workflow can easily be performed with the High Sensitivity DNA kit and the Agilent 2100 Bioanalyzer. The High Sensitivity DNA kit offers increased sensitivity for DNA analysis down to pg/µl concentrations across a broad linear dynamic range. This enhanced performance allows the number of required PCR cycles to be significantly decreased, eliminating PCR artifacts, while still reliably quantifying the sample. The improved DNA quality will improve the outcome of downstream sequencing analysis, maximizing throughput efficiency while minimizing cost per sample Therefore, the number of required PCR cycles can be significantly decreased, eliminating PCR artifacts, while still reliably quantifying the sample. References 1. Performance characteristics of the High Sensitivity DNA assay for the Agilent 2100 Bioanalyzer, Agilent Technologies Technical Note, publication number 5990-4417EN, 2009. 2. Gnirke, A., Melnikov, A., Maguire, J., Rogov, P., LeProust, E.M., Brokman, W., Fenell, T., Giannoukos, G., Fisher, S., Russ, C., Gabriel, S., Jaffe, D.B., Lander, E.S. and Nusbaum, Solution hybrid selection with ultra-long oligonucleotides for massively parallel targeted sequencing, C., Nature Biotechnology, 27 (2), 182-189, 2009. 3. SureSelect Target Enrichment System, Illumina Single-End Sequencing Platform Library Prep, Agilent Technologies Manual, reference number G3360-90010, 2009. 4. Agilent High Sensitivity DNA Kit Guide, Agilent Technologies Manual, reference number G2938-90321, 2009. 5. Quail, M.A., Kozarewa, I., Smith, F., Scally, A., Stephens, P.J., Durbin, R., Swerdlow, H., and Turner, D.J.; A large genome center s improvements to the Illumina sequencing system ; Nature Methods, 5 (12) 1005-10, 2008. 6

7

www.agilent.com/chem/dna Agilent Technologies, Inc., 2010, 2016 Published March 1, 2016 Publication Number 5990-5008EN