On-line SPE-LC/MS/MS to Detect Organonitrogen and Triazine Pesticides at 10ng/L in Drinking Water

Similar documents
Application Note. Author. Abstract. Environmental. Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany

EPA Method 543: Selected Organic Contaminants by Online SPE LC/MS/MS Using the Agilent Flexible Cube

Featuring Analyst software under Windows NT for enhanced performance and ease of use. API 150EX. LC/MS System

Changing the Game in LC-MS/MS. New Tools for Unprecedented Performance in Residue Analysis. Dr. Thomas Glauner

Development of a Fast and Cost-Effective Multi-Residue Method to Determine Pesticides in Tobacco by LC/MS/MS

Improved SPE for UPLC/MS Determination of Diquat and Paraquat in Environmental

Automated Online SPE for LC/MS/MS Analysis of Trace Organic Contaminants in Water Using the Agilent 1290 Infinity Flexible Cube Module

Exploring Extra Sensitivity Using ionkey/ms with the Xevo G2-XS Q-Tof HRMS for Small Molecule Pharmaceutical Analysis in Human Plasma

Tyler Trent, Applications Sales Specialist; Teledyne Tekmar Page 1

LC/MS/MS Solutions for Biomarker Discovery QSTAR. Elite Hybrid LC/MS/MS System. More performance, more reliability, more answers

Quantitation of Microcystins and Nodularins in Water Samples Using LC-MS/MS

A Sub-picogram Quantification Method for Desmopressin in Plasma using the SCIEX Triple Quad 6500 System

Confident Pesticides Analysis with the Agilent LC/Triple Quadrupole and TOF/QTOF Solutions

Chem 321 Lecture 23 - Liquid Chromatography 11/19/13

Thermo Scientific TSQ Quantum Access MAX

Rapid Soft Spot Analysis using the SCIEX Routine Biotransform Solution

Concentration of Human Hormones in Drinking Water Using Solid Phase Extraction and Analysis by High Performance Liquid Chromatography

Application Note. Author. Abstract. Biotherapeutics and Biologics. Sonja Schneider Agilent Technologies, Inc. Waldbronn, Germany

Liquid Chromatograph Liquid Chromatograph Mass Spectrometer. CL Series. for In Vitro Diagnostic Use

Accurately Identify and Quantify One Hundred Pesticides in a Single GC Run

Extraction and Analysis of Neonicotinoid Pesticides using QuEChERS Approach

Size Exclusion Chromatography of Biosimilar and Innovator Insulin Using the Agilent AdvanceBio SEC column

Automated Real-Time Determination of Bromate in Drinking Water Using LC-ICP-MS and EPA Method Application

Proudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment QTRAP 5500

Converting a Liquid-Liquid Extraction Method for Vitamin D to a 96-Well Plate Supported Liquid Extraction Format

Practical guidance for switching between anion and cation analysis systems

The LC-MS/MS Workhorse. SCIEX 4500 Series Mass Spectrometers

Quality-by-Design-Based Method Development Using an Agilent 1290 Infinity II LC

Investigation of Solid Phase Microextraction as an Alternative to Dried Blood Spot

Disulfide Linkage Analysis of IgG1 using an Agilent 1260 Infinity Bio inert LC System with an Agilent ZORBAX RRHD Diphenyl sub 2 µm Column

Technical Overview. Author. Abstract. Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany

High Temperature Liquid Chromatography

AB SCIEX TripleTOF 5600 SYSTEM. High-resolution quant and qual

Validation Report 19

MS Certified Vials. Pre-cleaned and certified vials for mass spectrometry

MassPREP On-Line Desalting Cartridge

State of the art in chromatographic analysis (LC MS/MS) of steroidal estrogens in surface water Status and Outlook

Optimizing the Purification of a Standard Chiral Compound Utilizing a Benchtop, Multi-Purpose, Semi-Preparative to Preparative HPLC System

EtG / EtS Analysis UCT INNOVATION THROUGH CHEMISTRY FORENSICS

User Bulletin. Applied Biosystems 3400 DNA Synthesizer. New Features in 3400 DNA Synthesizer Firmware v In This User Bulletin

Genotoxicity is the property of a compound

Development of a Clinical Research Method for the Measurement of Testosterone. Dihydrotestosterone

PLRP-S Polymeric Reversed-Phase Column for LC/MS Separation of mabs and ADC

EASY-Spray Columns. Guidance for column set up and installation Tips to maximize column lifetime

CHROMATOGRAPHY/MASS SPECTROMETRY

APPLICATION NOTE. UOP : Analysis of Trace CO and CO 2 in bulk H 2 and Light Gaseous Hydrocarbons by GC

Determination of Iodinated Contrast Media in Aqueous Samples by Direct-Injection LC-MS/MS

Validation Report 16

Application Note. Authors. Abstract

Method Translation in Liquid Chromatography

This User Bulletin does not apply to PROCISE clc instruments.

AppNote 4/2014. Prep-and-Shoot : The Completely Automated Workfl ow for the Hydrolysis and Analysis of Urine Samples by LC/MS/MS.

Analysis of 16 Pesticide Residues in Broccoli Using CarbonX dspe QuEChERS AOAC Kits Using GC/MS

Determination of Oxalate and Other Anions in Bayer Liquor Using Ion Chromatography

Fast and High-Resolution Reversed-Phase Separation of Synthetic Oligonucleotides

Rapid Peptide Catabolite ID using the SCIEX Routine Biotransform Solution

TechnicalNOTE METABOLYNX EXACT MASS DEFECT FILTER: RAPIDLY DISCRIMINATE BETWEEN TRUE METABOLITES AND FALSE POSITIVES INTRODUCTION

Agilent 1220 Infinity II LC GET ON THE FAST TRACK TO HIGHEST EFFICIENCY AT A PRICE YOU CAN AFFORD

Fusion Analytical Method Validation

QUANLYNX APPLICATION MANAGER

Peptide Mapping: A Quality by Design (QbD) Approach

Maximizing Chromatographic Resolution of Peptide Maps using UPLC with Tandem Columns

mab Titer Analysis with the Agilent Bio-Monolith Protein A Column

Preparative HPLC is still the

Analysis of Glyphosate and AMPA in Environmental Water Samples by Ion Chromatography and On-Line Mass Spectrometry with Minimal Sample Preparation

A Modular Preparative HPLC System for the Isolation of Puerarin from Kudzu Root Extracts

TurboMass GC/MS Software

Pharmacokinetic Analysis of the mab Adalimumab by ELISA and Hybrid LBA/LC/MS: A Comparison Study

The Importance of High Speed Data Acquisition in the Design and Implementation of GC/MS Applications

The Theory of HPLC Gradient HPLC

Fraction Analysis of Cysteine Linked Antibody-Drug Conjugates Using Hydrophobic Interaction. chromatography. Agilent 1260 Infinity II Bio-Inert System

ENHANCED MS WITH THE TURN OF A KEY

A Unique LC-MS Assay for Host Cell Proteins(HCPs) ) in Biologics

MasterView Software. Turn data into answers reliable data processing made easy ONE TOUCH PRODUCTIVITY

Comparison of Extraction Methods to Monitor Pesticide Residues in Surface Water

Agilent 7820A GC System THE EASY CHOICE FOR ROUTINE ANALYSES

Aflatoxin & Ochratoxin Analysis. A Completely Automated Solution with the Freestyle, ThermELUTE, and SMART Immunoaffinity Cleanup Columns

Rapid Peptide Mapping via Automated Integration of On-line Digestion, Separation and Mass Spectrometry for the Analysis of Therapeutic Proteins

SUBJECT: System Software Patch Provides Support for Fast SYBR Green Master Mix and RNA-to-C T

Peptide Mapping of Glycoprotein Erythropoietin by HILIC LC/MS and RP-LC/MS

Automated Solid Phase Extraction of PAH Compounds Utilizing the SPE-DEX 5000

Key Words Q Exactive Focus, SIEVE Software, Biomarker, Discovery, Metabolomics

Proteomics Technology Note

BÜCHI Labortechnik AG

Thermo Scientific Mass Spectrometric Immunoassay (MSIA) Pipette Tips. Next generation immunoaffinity. Robust quantitative platform

UFFLC of the Proteome Ultra High Pressure and Ultra Fast Flow Liquid Chromatography: A New Approach To Proteomics

Determination of Monocrotophos, Diazinon, Malathion, EPN, and Methamidaphos from Aqueous Samples Using Atlantic HLB SPE Disks

N-Glycan Profiling Analysis of a Monoclonal Antibody Using UHPLC/FLD/Q-TOF

Improving Resolution and Column Loading Systematically in Preparative Liquid Chromatography for Isolating a Minor Component from Peppermint Extract

A green HPLC technique with a 100% water mobile phase for detecting imidacloprid and its metabolite 6-chloronicotinic acid

Thermo Scientific Peptide Mapping Workflows. Upgrade Your Maps. Fast, confident and more reliable peptide mapping.

Fast Preparative Column Liquid Chromatography (PCLC)

Physical Stability of a Silica- Based Size Exclusion Column for Antibody Analysis

HPLC/FLASH Purification ADVANCE THE PACE OF YOUR DISCOVERY TO ACHIEVE THE RESULTS OF TOMORROW

Accucore. Ultimate Core Performance LC Column Technology to Maximize Your Investment. Dave Jarzinski. October 2011

A Fast and Robust Linear ph Gradient Separation Platform for Monoclonal Antibody (MAb) Charge Variant Analysis

Quantitative determination of residual 2-(2-chloroethoxy) ethanol (CEE) in quetiapine fumarate by gas chromatogaraphy

Certified Reference Material - Certificate of Analysis

Characterization of mab aggregation using a Cary 60 UV-Vis Spectrophotometer and the Agilent 1260 Infinity LC system

Fluensulfone on Potatoes and Sugar Beets

Transcription:

application note On-line SPE-LC/MS/MS of Pesticides On-line SPE-LC/MS/MS to Detect Organonitrogen and Triazine Pesticides at 10ng/L in Drinking Water API 3200 LC/MS/MS System Overview The feasibility of on-line Solid Phase Extraction (SPE) coupled to Liquid Chromatography with tandem Mass Spectrometry (LC/MS/MS) to detect organonitrogen and triazine pesticides at concentrations of 10ng/L in drinking water was studied. Sensitivity, reproducibility, linearity and recovery in tap water samples and water samples spiked with humic acid were investigated to evaluate sample preparation efficiency and possible matrix effects. Introduction Current legislation for drinking waters in the UK requires a lower Limit of Quantitation (LOQ) for pesticides of 100ng/L (ppt) with a Coefficient of Variation (CV) below 12.5% and a Limit of Detection (LOD) of 20ng/L. 1 In order to achieve these criteria, traditional methods usually use different sample pre-treatments for each pesticide class, such as different procedures, e.g. SPE, to concentrate the analytes and remove matrix components. After extraction and specific chemical derivatization processes several different analytical techniques are often needed to perform the actual detection and determination of pesticides at these low levels. If the analytes are thermally stable pesticides, then they are analyzed by Gas Chromatographic (GC) techniques whilst thermally labile compounds are usually determined by Liquid Chromatography (LC). This paper describes an automated SPE sample clean-up and pre-concentration step utilizing large volume injection (5mL of untreated water samples) prior to LC/MS/MS analysis. With this approach, all investigated 26 pesticides are first trapped on an SPE cartridge and then back flushed onto a reverse phase LC column. The pesticides are then separated on the analytical column using a binary LC gradient and are determined on the API 3200 LC/MS/MS System in Multiple Reaction Monitoring (MRM).

Trapping the analytes on an on-line SPE cartridge has three major advantages over methods employing direct on column injection: 1. Analytes can be rapidly enriched on the SPE column, thus enabling detection at lower LOD. 2. Matrix components, such as humic acid or inorganic salts are washed away prior to the analytical column. The longevity of the column life is therefore extended whilst maximizing assay robustness. Table 1. LC gradient used for separation of organonitrogen and triazine pesticides Step Total Time (min) A (%) B (%) 0 0.1 90 10 1 4.0 90 10 2 4.5 60 40 3 5.5 60 40 4 12.0 0 100 5 14.0 0 100 6 14.5 90 10 7 16.0 90 10 3. The reproducibility of the analysis is improved by automation and thus minimizing human errors during sample preparation. Table 2. MRM transitions with compound dependant MS/MS parameters for each detected pesticide Compound Q1 Q3 DP in V CE in V Dwell Time (ms) Ametryn 228.2 96.1 46 33 30 Atraton 212.1 100.1 41 35 30 Atrazine 216 104.1 46 39 30 Desmetryn 214.1 82.2 45 43 30 Epoxiconazole 330 121 36 27 30 EPTC 190.1 86.1 26 17 100 Fenpropidin 274 147 51 37 30 Fenpropimorph 304 117 61 71 30 Flusilazole 316 247 36 25 30 Flutriafol 302 123 41 39 30 Metazachlor 278.2 210.2 31 17 30 Pendimethalin 282 212 6 15 30 Prometon 226.2 184.2 51 23 30 Prometryn 242.2 200.2 41 23 30 Propachlor 212 94 36 37 30 Propazine 230.2 188.3 51 23 30 Propiconazole 342 159 46 37 30 Propyzamide 256 190 41 19 30 Simazine 202.1 132.1 46 25 30 Simetryn 214 124 31 27 30 Tebuconazole 308 70 21 39 30 Terbuthylazine 230.1 174.1 46 23 30 Terbutryn 242.2 186 41 23 30 Triadimefon 294 197 36 21 30 Tri-allate 304 143 35 34 100 Trietazine 230 99 26 33 30

Figure 1. MRM transitions illustrating differences in the relative sensitivities of 10ng/L Terbutryn (S/N = 498), Atrazine (S/N =65) and for EPTC (S/N =9) calculated with peak-to-peak algorithm Experimental HPLC 5mL of tap water were loaded onto a Phenomenex Strata C18 on-line SPE cartridge using a CTC HTC PAL autosampler (fitted with a 5mL syringe, a 5mL loop and a 32 x 10mL sample rack). The trap cartridge was fitted into load position A of a high pressure, Valco 10 port switching valve. An Agilent 1200 isocratic pump was used to load the sample at 3mL/min onto the trapping cartridge. After 2.5 minutes the valve was switched to position B where an Agilent 1200 Binary pump was used to provide the gradient elution of pesticides. For the separation of pesticides an Agilent Zorbax 1.8µm 4.6x50mm with a flow rate of 600µL/min at ambient temperature was used. The gradient profile is shown in Table 1 where mobile phase A was 2mM ammonium formate with 0.1% formic acid in water and mobile phase B was 2mM ammonium formate with 0.1% formic acid in methanol. After 14.5 minutes the switching valve was switched back to the loading position to analyze the next sample. Local tap water without any pretreatment was spiked with pesticides and injected directly onto the SPE-LC/ MS/MS system. To study matrix effects commercially available humic acid (Sigma-Aldrich) was dissolved in 0.2% formic acid at a level of 10 parts per million (ppm) and pesticides were spiked into this matrix. Mass Spectrometry An API 3200 LC/MS/MS System with a Turbo V Source and Electrospray Ionization probe was used for all experiments. Curtain gas 25psi Gas1 70psi Gas2 30psi CAD gas 4 Temperature 700 C IonSpray voltage 5500V

Table 3. Retention times, Signal-to-Noise, reproducibility and linearity data for each detected pesticide in tap water without the use of internal standards Compound Retention time S/N of 10ng/L in Tap Water Standard Deviation of 10ng/L (n=6) %CV of 100ng/L (n=6) r Value for Calibration Line (10-500ng/L) Ametryn 10.1 51 1.01 2.53 0.9997 Atraton 8.6 46 0.52 2.93 0.9998 Atrazine 10.3 65 0.95 1.76 0.9988 Desmetryn 9.3 47 0.65 2.59 0.9992 Epoxiconazole 11.6 88 0.51 1.05 0.9987 EPTC 11.9 9 1.99 4.47 0.9993 Fenpropidin 10.4 178 0.58 1.7 0.9996 Fenpropimorph 12.9 56 0.66 6.72 0.9979 Flusilazole 11.6 169 0.44 1.69 0.9987 Flutriafol 10.2 12 1.13 1.92 0.9994 Metazachlor 10.3 117 0.63 1.82 0.9997 Pendimethalin 11.1 64 0.93 3.7 0.9946 Prometon 9.6 55 0.75 7.54 0.9998 Prometryn 10.9 128 0.43 1.75 0.9991 Propachlor 10.2 28 0.65 1.95 0.9994 Propazine 11.3 75 1.39 2.82 0.9994 Propiconazole 11.9 55 0.51 2.37 0.9998 Propyzamide 11.3 160 0.46 3 0.9994 Simazine 9.5 39 0.98 3.05 0.9991 Simetryn 9.3 43 0.51 3 0.9994 Tebuconazole 11.9 116 0.64 2.18 0.9996 Terbuthylazine 10.5 171 0.28 2.12 0.9995 Terbutryn 10.9 498 0.39 3.13 0.9991 Triadimefon 11.2 42 0.62 2.02 0.9998 Tri-allate 12.9 22 1.21 4.09 0.9985 Trietazine 11.7 13 1.76 2.02 0.9993 Detected MRM transitions with compound dependant parameters which were optimized using Analyst Software Quantitative Optimization for each pesticide are displayed in Table 2 [Declustering Potential (DP) and Collision Energy (CE)]. During any chromatographic measurement small drifts in retention time will inevitably be caused by gradual column degradation or small changes in temperature and mobile phase composition in between running samples. When incorporating time windows (periods or segments) into the chromatogram a small change in retention may result in pesticides falling outside of the window and therefore going undetected during the MS/MS acquisition. This phenomena becomes more profound as the number of compounds being monitored and/or windows within a given experiment increases. Therefore, the ion path of the API 3200 System using a Linear Accelerator (LINAC ) Collision Cell was set to constantly monitor all pesticides throughout the entire chromatogram. In addition with this approach, it is possible to subsequently add more pesticides to the suite of compounds screened for without the prior need to investigate the elution profile of the added analytes.

Figure 2. Calibration lines of the weakly ionizing organonitrogen pesticide EPTC (top) with r=0.9993, and the strongly ionizing triazine pesticide Terbuthylazine (bottom) with r=0.9995 over a linear range of 10ng/L to 500ng/L Results and Discussion The analyte sensitivity varied markedly from pesticide to pesticide and the corresponding S/N (peak-to-peak) ratio for the pesticides varied from 498:1 (Terbutryn) down to 9:1 (EPTC), see Figure 1 and Table 3. Therefore, this qualifies the need to use 5mL of sample for SPE so as to include weakly ionizing compounds, such as EPTC, Flutriafol and Trietazine within the same sample injection. Reproducibility was assessed by the analysis of tap water samples which had been spiked at two levels 10ng/L and 100ng/L (the LOQ needed for drinking water monitoring). Table 3 shows excellent standard deviations and coefficients of variation (%CV) for all pesticides. Linearity was studied over a concentration range form 10 to 500ng/L. Correlation coefficients were found to be greater than 0.994 for all calibration lines when a linear fit with 1/x weighting was applied as presented in Figure 2 and Table 3. No internal standards were used when reproducibility and linearity were investigated.

Figure 3. Carry over assessment using an injection of a 500ng/L standard of Fenpropidin and EPTC followed by an injection of blank water directly afterwards (corresponding MRM transitions for the same compounds are shown for the two injections) To further test the robustness of the method 70 injections were made of the same 100ng/L standard of Terbutryn over a 20 hour period which gave a CV of 6.6%. A 500ng/L standard was injected followed by a blank water sample to assess possible carry over using the developed method. Figure 3 shows a typical result for 2 pesticides and indicates that carry over is less than 0.2%. Figures 4 shows examples of Total Ion Chromatograms (TIC) obtained from the direct injection of different types of spiked water. In these examples it can be seen that there is a very low effect on the intensity of the response observed when tap water and water containing a high level of humic acid is compared to distilled water. The recoveries of all pesticides in tap water ranged from 77 to 115% (Table 4). Also when pesticides were spiked into water containing 10ppm of humic acid there was a decrease in sensitivity but recoveries compared to distilled water were still over 67%. The effect of humic acid was to decrease recovery in comparison to tap water samples by typically less than 15% (the only exception being Tri-allate with 20%). Conclusion The method described provides a simple, robust and cost effective approach for the routine analysis of organonitrogen and triazine pesticides in potable waters by on-line SPE-LC/MS/ MS. The use of on-line SPE eliminates the additional cost and time required for traditional SPE methods and offers improved robustness and reproducibility. Limits of Detection (LOD) have been calculated using both peak-to-peak measurements and also by calculating the LOD as 4.65 times the standard deviation of a 10ng/L spike. Both calculations show that all the pesticides

Figure 4. Total Ion Chromatograms of a 100ng/mL spike of organonitrogen and triazine pesticides in distilled water (top), tap water (middle) and 10ppm humic acid (bottom) to assess ion suppression effects tested can be detected below 10ng/L with good CV below 8% at 100ng/L and excellent linearity from 10 to 500ng/L. The sensitivity of the method was only slightly effected by the type of water sample analyzed and has been shown to be sensitive enough for all the pesticides in potable waters. In water samples with a high total organic carbon, such as humic acid at 10 parts per million, some ion some suppression results but it was less than 20%. The suitability of the application of the API 3200 LC/MS/MS System for the routine analysis of these pesticides has therefore been demonstrated at the required levels and precisions. Currently, the application of this technique to the determination of other pesticides such as acidic herbicides and organophosphorus pesticides is being investigated. Author Stephen J. Lock (Applied Biosystems, Warrington, Cheshire, UK) Reference 1 A manual of Analytical Quality control for the water industry (NS30), Drinking Water Inspectorate, UK

Table 4. Recovery data at 100ng/L for each detected pesticide in tap water and water spiked with 10ppm humic acid Compound Peak Area of 100ng/L in Distilled Water Peak Area of 100ng/L in Tap Water Peak Area of 100ng/L in Water with 10ppm Humic Acid Recovery (%) from Tap Water Recovery (%) from Water with 10ppm Humic Acid Effect of Humic Acid on Recovery (%) Ametryn 2.68E+05 2.08E+05 1.94E+05 77.8 72.3-5 Atraton 2.68E+05 2.12E+05 2.04E+05 78.9 75.8-3 Atrazine 1.53E+05 1.22E+05 1.02E+05 79.9 66.5-13 Desmetryn 4.32E+05 3.37E+05 2.95E+05 77.9 68.4-10 Epoxiconazole 3.87E+05 4.14E+05 3.58E+05 107.2 92.5-15 EPTC 1.57E+04 1.65E+04 1.56E+04 105.2 99.1-6 Fenpropidin 1.03E+06 8.51E+05 8.15E+05 82.8 79.2-4 Fenpropimorph 4.59E+05 3.89E+05 3.80E+05 84.7 82.6-2 Flusilazole 2.59E+05 2.88E+05 2.56E+05 111.2 98.7-12 Flutriafol 6.93E+04 7.47E+04 6.88E+04 107.9 99.4-8 Metazachlor 5.19E+05 4.41E+05 4.04E+05 84.9 77.8-7 Pendimethalin 1.24E+05 1.41E+05 1.27E+05 113.7 102.6-11 Prometon 6.82E+05 5.39E+05 5.18E+05 79.1 76-3 Prometryn 8.61E+05 7.74E+05 7.00E+05 89.8 81.2-9 Propachlor 2.59E+05 2.27E+05 2.05E+05 87.9 79.3-9 Propazine 3.97E+05 3.51E+05 2.95E+05 88.3 74.4-14 Propiconazole 2.18E+05 2.40E+05 2.17E+05 110.2 99.6-11 Propyzamide 1.31E+05 1.22E+05 1.06E+05 93.1 81.4-12 Simazine 1.52E+05 1.20E+05 1.16E+05 79.1 76.3-3 Simetryn 4.87E+05 3.73E+05 3.50E+05 76.6 71.9-5 Tebuconazole 2.09E+05 2.24E+05 2.15E+05 107.2 102.8-4 Terbuthylazine 9.68E+05 8.86E+05 7.43E+05 91.5 76.7-15 Terbutryn 2.45E+06 2.13E+06 1.93E+06 86.7 78.6-8 Triadimefon 1.50E+05 1.71E+05 1.60E+05 114.5 107-7 Tri-allate 4.13E+04 4.59E+04 3.76E+04 111.1 91.1-20 Trietazine 2.74E+05 2.39E+05 2.08E+05 87.3 76-11 For Research Use Only. Not for use in diagnostic procedures. Applera Corporation is committed to providing the world s leading technology and information for life scientists. Applera Corporation consists of the Applied Biosystems and Celera Genomics businesses. Copyright 2007, Applera Corporation and MDS Inc. All rights reserved. Information subject to change without notice. Applera, Applied Biosystems and AB (Design) are registered trademarks of Applera Corporation or its subsidiaries in the US and/or certain other countries. API 3200 and Turbo V are trademarks and Analyst and LINAC are registered trademarks of Applied Biosystems/MDS SCIEX, a joint venture between Applera Corporation and MDS Inc. All other trademarks are the sole property of their respective owners. Printed in the USA. 07/2007 Publication 114AP63-01 Headquarters 850 Lincoln Centre Drive Foster City, CA 94404 USA Phone 650.638.5800 Toll Free 800.345.5224 www.appliedbiosystems.com International Sales For our office locations please call the division headquarters or refer to our Web site at www.appliedbiosystems.com/about/offices.cfm