Heart-cut 2D-LC/MS approach for pharmaceutical impurity identification using an Agilent 6540 Q-TOF LC/MS System

Similar documents
Genotoxicity is the property of a compound

Technical Overview. Author. Abstract. Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany

Characterize Fab and Fc Fragments by Cation-Exchange Chromatography

EPA Method 543: Selected Organic Contaminants by Online SPE LC/MS/MS Using the Agilent Flexible Cube

Quality-by-Design-Based Method Development Using an Agilent 1290 Infinity II LC

Confident Pesticides Analysis with the Agilent LC/Triple Quadrupole and TOF/QTOF Solutions

Fast and High-Resolution Reversed-Phase Separation of Synthetic Oligonucleotides

Application Note. Author. Abstract. Biotherapeutics and Biologics. Sonja Schneider Agilent Technologies, Inc. Waldbronn, Germany

Disulfide Linkage Analysis of IgG1 using an Agilent 1260 Infinity Bio inert LC System with an Agilent ZORBAX RRHD Diphenyl sub 2 µm Column

Size Exclusion Chromatography of Biosimilar and Innovator Insulin Using the Agilent AdvanceBio SEC column

PLRP-S Polymeric Reversed-Phase Column for LC/MS Separation of mabs and ADC

Precise Characterization of Intact Monoclonal Antibodies by the Agilent 6545XT AdvanceBio LC/Q-TOF

Application Note. Author. Abstract. Environmental. Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany

Cell Clone Selection Using the Agilent Bio-Monolith Protein A Column and LC/MS

Characterization of mab aggregation using a Cary 60 UV-Vis Spectrophotometer and the Agilent 1260 Infinity LC system

Agilent 6500 Series Q-TOF LC/MS System

Application Note. Abstract. Authors. Introduction. Environmental

Application Note. Author. Abstract. Biopharmaceuticals. Verified for Agilent 1260 Infinity II LC Bio-inert System. Sonja Schneider

Application Note. Author. Abstract. Small Molecule Pharmaceuticals. Sonja Krieger Agilent Technologies, Inc. Waldbronn, Germany

Peptide Mapping: A Quality by Design (QbD) Approach

Exploring Extra Sensitivity Using ionkey/ms with the Xevo G2-XS Q-Tof HRMS for Small Molecule Pharmaceutical Analysis in Human Plasma

Peptide Mapping of Glycoprotein Erythropoietin by HILIC LC/MS and RP-LC/MS

Rapid Soft Spot Analysis using the SCIEX Routine Biotransform Solution

Minimize Method Development Time for Pesticide Screening Using LC/MS Application Kit

Characterization of Glycosylation in the Fc Region of Therapeutic Recombinant Monoclonal Antibody

Method Translation in Liquid Chromatography

N-Glycan Profiling Analysis of a Monoclonal Antibody Using UHPLC/FLD/Q-TOF

Chem 321 Lecture 23 - Liquid Chromatography 11/19/13

Controlling Agilent 1200 Series Rapid Resolution LC systems through Waters Empower chromatography data software. Technical Overview.

Fraction Analysis of Cysteine Linked Antibody-Drug Conjugates Using Hydrophobic Interaction. chromatography. Agilent 1260 Infinity II Bio-Inert System

Antibody Analysis by ESI-TOF LC/MS

Automated Online SPE for LC/MS/MS Analysis of Trace Organic Contaminants in Water Using the Agilent 1290 Infinity Flexible Cube Module

Qualification of High-Performance Liquid Chromatography Systems

Bivalirudin Purification:

Analysis of Monoclonal Antibody Digests with the Agilent 1290 Infinity 2D-LC Solution Part 2: HILIC RPLC-MS

CAMAG TLC-MS INTERFACE 2

Fundamentals and Techniques of Preparative HPLC. Parto Zist Behboud Tel: 42108

Concentration of Human Hormones in Drinking Water Using Solid Phase Extraction and Analysis by High Performance Liquid Chromatography

Improving Resolution and Column Loading Systematically in Preparative Liquid Chromatography for Isolating a Minor Component from Peppermint Extract

Agilent 6430 Triple Quadrupole LC/MS System

Agilent 1220 Infinity II LC GET ON THE FAST TRACK TO HIGHEST EFFICIENCY AT A PRICE YOU CAN AFFORD

On-line SPE-LC/MS/MS to Detect Organonitrogen and Triazine Pesticides at 10ng/L in Drinking Water

Monitoring for Microcystins in Raw Water Supply Reservoirs Using the Agilent 6410 Triple Quadrupole LC/MS

Determination of Iodinated Contrast Media in Aqueous Samples by Direct-Injection LC-MS/MS

Key Words Q Exactive Focus, SIEVE Software, Biomarker, Discovery, Metabolomics

Agilent AdvanceBio SEC Columns for Aggregate Analysis: Instrument Compatibility

MassPREP On-Line Desalting Cartridge

Optimizing the Purification of a Standard Chiral Compound Utilizing a Benchtop, Multi-Purpose, Semi-Preparative to Preparative HPLC System

Using accurate mass LC/MS/MS in forensic toxicology screening

Enabling Systems Biology Driven Proteome Wide Quantitation of Mycobacterium Tuberculosis

Overview of preparative HPLC. Analytical Technologies Limited

Size Exclusion Chromatography/ Mass Spectrometry Analysis of Antibody Drug Conjugates Using the Agilent 1260 Infinity II Bio-Inert LC

Liquid Chromatograph Liquid Chromatograph Mass Spectrometer. CL Series. for In Vitro Diagnostic Use

Protein-Pak Hi Res HIC Column and HIC Protein Standard

Improved SPE for UPLC/MS Determination of Diquat and Paraquat in Environmental

Available Online through (or) IJPBS Volume 2 Issue 3 JULY-SEPT Research Article Pharmaceutical Sciences

The clearly better choice

Automated Real-Time Determination of Bromate in Drinking Water Using LC-ICP-MS and EPA Method Application

Analysis of Suspected Flavor and Fragrance Allergens in Cosmetics Using the 7890A GC and Capillary Column Backflush Application

Rapid Peptide Catabolite ID using the SCIEX Routine Biotransform Solution

EASY-Spray Columns. Guidance for column set up and installation Tips to maximize column lifetime

Preparative HPLC: Factors and Parameters that Directly Affect Recovery of Collected Fractions

The Importance of High Speed Data Acquisition in the Design and Implementation of GC/MS Applications

MS Certified Vials. Pre-cleaned and certified vials for mass spectrometry

Thermo Scientific TSQ Quantum Access MAX

A Unique LC-MS Assay for Host Cell Proteins(HCPs) ) in Biologics

Accurately Identify and Quantify One Hundred Pesticides in a Single GC Run

High Resolution Accurate Mass Peptide Quantitation on Thermo Scientific Q Exactive Mass Spectrometers. The world leader in serving science

ADVANCE ACCURACY AND PRODUCTIVITY FOR FASTER ANALYSIS

An Automated System for At-Line Process Analysis of Biopharmaceutical Fermentation Reactions

Flexibility with. Preparative HPLC

Certified Reference Material - Certificate of Analysis

AZURA Bio LC. 2 FPLC and Purification Solutions

Using HRAM Survey Analysis Combined with Rapid MS2 Data to Develop a Fragmentation Based Detection Workflow for Structure ID Acquisition

Rapid Peptide Mapping via Automated Integration of On-line Digestion, Separation and Mass Spectrometry for the Analysis of Therapeutic Proteins

QbD approach for analytical method development of anti-pschotic drug

INTERNATIONAL JOURNAL OF PHARMACEUTICAL RESEARCH AND BIO-SCIENCE

Fast Preparative Column Liquid Chromatography (PCLC)

Analysis of USP <467> Residual Solvents using the Agilent 7697A Headspace Sampler with the Agilent 7890B Gas Chromatograph

The Theory of HPLC Gradient HPLC

A Complete Workflow Solution for Intact Monoclonal Antibody Characterization Using a New High-Performance Benchtop Quadrupole- Orbitrap LC-MS/MS

High Resolution Analysis of Heparin and Heparin-like Impurities on Glycomix TM an Anion Exchange Column

HPLC/FLASH Purification ADVANCE THE PACE OF YOUR DISCOVERY TO ACHIEVE THE RESULTS OF TOMORROW

TurboMass GC/MS Software

The Use of LC-MS in Systematic Toxicology Analysis

A Fast and Robust Linear ph Gradient Separation Platform for Monoclonal Antibody (MAb) Charge Variant Analysis

Proteins. Patrick Boyce Biopharmaceutical Marketing Manager Waters Corporation 1

International Journal of Pharma Research & Review, Feb 2014; 3(2):11-16 ISSN:

mabs and ADCs analysis by RP

APPLICATIONS TN Overview of Kinetex 2.6 µm Core-Shell Technology

AZURA Bio purification

Two-Dimensional LC Protein Separation on Monolithic Columns in a Fully Automated Workflow

A Modular Preparative HPLC System for the Isolation of Puerarin from Kudzu Root Extracts

Preparative HPLC System

Analysis of Glyphosate and AMPA in Environmental Water Samples by Ion Chromatography and On-Line Mass Spectrometry with Minimal Sample Preparation

SEC-MS with Volatile Buffers for Characterization of Biopharmaceuticals

Optimization of Volatile Petroleum Hydrocarbons Analysis by Automated Headspace Using Method Development Tools

ICP Expert software. Technical Overview. Introduction

Generating Automated and Efficient LC/MS Peptide Mapping Results with the Biopharmaceutical Platform Solution with UNIFI

Transcription:

Heart-cut 2D-LC/M approach for pharmaceutical impurity identification using an Agilent 640 Q-TF LC/M ystem Application ote Pharmaceuticals Author iji Joseph Agilent Technologies, Inc. Bangalore, India Abstract This Application ote describes a heart-cut two-dimensional LC/M method for the identification of a close eluting impurity observed in Duloxetine drug substance using a non-m compatible reverse phase LC method. The analysis was performed using an Agilent 1260 Infinity LC system coupled with anagilent 640 UHD Q-TF LC/M system. Agilent ZRBAX Eclipse Plus columns were used for the first and second dimensional separation. In the first LC dimension, a non-m compatible liquid chromatography (LC) method was performed. Using a column switching valve, the specific unknown impurity was diverted to the second LC column, and was separated using an M compatible mobile phase from Pump 2. Accurate mass M and M/M data of the impurity were acquired using a 640 UHD Q-TF. Data processing using Agilent MassHunter and Agilent Molecular tructure Correlator (MC) software enabled quick identification of the unknown impurity. Efficient and unambiguous impurity profiling is possible with minimum user intervention when using this method.

Introduction Detection and identification of impurities in the active pharmaceutical ingredients (API s) are critical in the pharmaceutical industry. Increasing stringency in the regulatory environment demands sensitive and convincing analytical approaches for the determination of impurities in API s. ICH guidelines require that impurities at or above 0.1% in the drug substance are identified 1. Better understanding of the impurity gives insights to control the chemical reaction process and achieve higher API purity. At times, the impurities may be structurally similar to the API, therefore, sophisticated and selective analytical methods are critical. Conventional analytical approaches for impurity profiling often involve multiple instrument platforms and can be timeconsuming and laborious. Techniques such as LC/M/M are widely used for impurity identification in pharmaceutical industries due to its high sensitivity, selectivity, and speed of analysis. Coupling high pressure liquid chromatography (HPLC) system with quadrupole time-of-flight (Q-TF) enables accurate mass measurement of both precursor and fragment ions. Data processing using advanced algorithms such as molecular feature extraction (MFE) and molecular formula generation (MFG) of the Agilent Mass Hunter Qualitative Analysis software along with MassHunter Molecular tructural Correlator software (MC) allows impurity identification and structure elucidation. Identification of eight European pharmacopeia specified impurities in atenolol API using this approach is well described in Agilent publication number 991-137E 2. There are limits to LC/M with respect to the use of nonvolatile buffers as mobile phases. If an original LC method is developed with nonvolatile buffers, then the chromatographer has to invest additional time to develop an equivalent LC method with an M compatible mobile phase. Furthermore, it may add additional challenges and uncertainty if an impurity elution order changes with the newly developed M-compatible LC method. This limitation can be eliminated with a heart-cut approach employing two reversed phase LC conditions. The approach described in this Application ote uses a two dimensional LC technique to transfer an impurity of interest from the first column eluted with non-m compatible mobile phase to the second column which is under M compatible LC condition. This approach can be extended to analyses of many pharmaceutical ingredients and related impurities. Experimental Instruments LC/M consisted of an Agilent 640 UHD Q-TF with Jet tream source, and an Agilent 1260 Infinity Binary LC system. The individual modules and columns used are as follows: Agilent 1260 Infinity eries Degasser (G1379B) for the first and second dimensions Two Agilent 1260 Infinity Binary Pumps (G1312B) for the first and second dimension Agilent 1260 Infinity Highperformance Autosampler (G1367D) Agilent 1260 Infinity Thermostatted Column Compartment with 2-Position/6-Port column switching valve (G1316B) Agilent 1290 Infinity Diode Array detector (G4212A) with Max-Light flow cell (4.0 µl volume, 60-mm path length) (G4212 A) Column 1: Agilent ZRBAX Eclipse Plus C-18, 4.6 20 mm,.0 µm (p/n 99990-902) Column 2: Agilent ZRBAX Eclipse Plus C-18, 4.6 7 mm, 3. µm (p/n 99933-902) oftware Agilent Chemtation software B.04.03 was used to acquire LC-UV data and to quantify area percentage of impurities, and MassHunter Workstation (version B.04.00) was used for LC mass spectrometry data acquisition. MassHunter Qualitative Analysis software (B.04.00) was used for data processing. MassHunter MC software (version B.0.00) was used to facilitate structure elucidation of impurities. Reagents and Materials LC/M grade acetonitrile, methanol and formic acid were purchased from Fluka (Germany). Highly purified water from a Milli Q system (Millipore Elix 10 model, UA) was used for mobile phase preparation. Potassium dihydrogen phosphate was purchased from Fluka (Germany). tandards of duloxetine and its impurity were purchased from Varda biotech (India). Workflow Figure 1 shows the instrument block diagram used for the experiment. The impurity analysis of Duloxetine API was performed as per the HPLC method described in Reference 3. The overall chromatographic elution profile using Agilent ZRBAX Eclipse Plus C-18, 4.6 20 mm,.0 µm column was evaluated. 2

The impurity was eluted at 8.7 minutes using the HPLC method 1 (Table 1). Later, in a separate injection, the unknown impurity was transferred to Column 2 by changing the valve position between 8.2 minutes to 9.0 minutes using a 2-Position/6-Port column switching valve. During this heart-cut time, Column 2 is also connected to Pump 2 which was running at 9% aqueous of M compatible mobile phase with a T-connector. The total flow through Column 2 during this time was the sum of the flows from Pumps 1 and 2. The flow from Pump 2 contains high percentage of aqueous component that reduces the overall organic content in the mobile phase, and, thus, impurity can be retained in Column 2. After 9 minutes, the column switch valve was changed back to the original position and eluent from Column 1 using Pump 1 was redirected to DAD and continued the UV analysis. olvent tray 1 Degasser 1 Pump 1 olvent tray 2 Degasser 2 Pump 2 DAD Column thermostat Auto sampler AL Thermostat Agilent 640 Q-TF Figure 1 Instrument block diagram used for the experiment. The column compartment includes two columns and connected with a six port switch over valve. From 9 minutes on, the flow of Column 2 depended entirely on Pump 2. Pump 2 remained at 9% aqueous for 12 minutes. This was found to be sufficient to wash out all buffers from Column 2. Up to this time, Column 2 flow was directed to the waste using Q-TF inlet rheodyne. From 12 minutes on, the gradient was operated using Pump 2 to elute the trapped impurity from Column 2. The eluent was directed to the Q-TF mass spectrometer and data were acquired in an auto M/M mode. 3

The illustrative representation of the valve positions during the heart-cut method is shown in Figure 2 and the entire workflow is summarized in Figure 3. DAD 2 6 port column switch valve 1 6 Pump 2 Q-TF A Column 2 Column 1 3 4 Pump 1 AL tep A (from 0 to 8.2 minutes): Flow path 1: Performing validated reference HPLC UV method using non-m compatible mobile phase. Flow path 2: Column 2 is getting conditioned from Pump 2 using M compatible mobile phase. DAD 1 6 Pump 2 B 2 Q-TF Column 1 Column 2 3 4 Pump 1 AL tep B (from 8.2 to 9 minutes): Flow path 1: Performing Heart-cut for an unknown impurity. During this time, solvents from both the pumps are flowing through Column 2 using a T-connector. C DAD 1 6 Pump 2 2 Q-TF Column 1 Column 2 3 4 Pump 1 AL tep C (from 9 to 12.0 minutes): Flow path 1: HPLC UV analysis continues using Pump 1 with Column 1. Flow path 2: Washing out the buffer from Column 2 using Pump 2. DAD 1 6 Pump 2 D 2 Q-TF Column 1 Column 2 3 4 Pump 1 AL tep D (from 12 minutes to end): Flow path 1: As step C. Flow path 2: Elution of trapped impurity from Column 2 and directed to QTF. Figure 2 Valve positions before, during, and after the heart-cut process. Red color indicates the flow path (1) with non-m compatible mobile phase and green color indicates the flow path (2) with M compatible mobile phase. 4

HPLC eparation (Column1, Pump 1) DAD analysis with non-m compatible LC method using Pump 1 and Column 1 Divert the impurity peak (only) to Column 2 using column switch valve. Use Pump 2 to wash out Column 2. LC/M analysis (Column 2, Pump 2) Analyze the trapped impurity using Agilent 640 Q-TF with a full M scan followed by an auto M/M. Identify impurity by MFE and MFG algorithms based on the accurate mass M and MM data. MC facilitates the structure elucidation of the impurities. Figure 3 Workflow for impurity identification using a heart-cut approach.

Instrument parameters Tables 1 and 2 summarize the instrument parameters. Parameter HPLC:1 (non-m compatible) conditions HPLC: 2 (M compatible) conditions Column Agilent ZRBAX Eclipse Plus C-18, 4.6 20 mm,.0 µm Agilent ZRBAX Eclipse Plus C-18, 4.6 7 mm, 3. µm Flow rate 1.0 ml/min As gradient Mobile phase A Isocratic: 20 mm Phosphate buffer: acetonitrile: methanol; :37:8 0.1% formic acid in water Mobile phase B ot applicable 0.1% formic acid in acetonitrile Detection DAD:229 nm Q-TF Injection vol µl ot applicable eedle wash Activated for 8 seconds using methanol ot applicable Pump mode Pump:1, Isocratic for 30 minutes Pump:2, Gradient Time %B Flow 0 0. 9 0. 10 0.7 17 60 0.7 17.1 0. 2 0. Table 1 Agilent 1200 eries LC instrument parameters. Q-TF M and auto M/M conditions M instrument Agilent 640 Q-TF Ion source AJ EI Acquisition mode 2 GHz, Ext dynamic range Ion polarity Positive mode Drying gas temperature 32 C Drying gas 10 L/min ebulizer 4 psig heath gas temperature 37 C heath gas flow 12 L/min VCap 4,000 V ozzle voltage 00 V Fragmentor 90 V Acquisition M followed by auto M/M Table 2 Agilent 640 Q-TF parameters. 6

Results and Discussion Figure 4 shows the elution profile of duloxetine API with LC-UV analysis. The area percentage of the unknown impurity (retention time: 8.7 minutes) was approximately 0.1%. Figure shows the total ion chromatogram (TIC) obtained from the Q-TF during the heart-cut analysis. mau 20 10 0 DAD1 A, ig=229, Blank DAD1 A, ig=229, Duloxetine API 2.603 4.392 0 2 4 6 8 10.498 7.672 8.783 min Figure 4 Elution profile of duloxetine API using LC-UV analysis. (The heart cut region of the baseline is marked). 10 6 1.6 Total Ion Chromatogram (TIC) Abundance 1.2 0.8 0.4 0 8. 9 9. 10 10. 11 11. 12 12. 13 13. 14 14. 1 1. 16 16. Counts versus acquisition time (min) Figure The total ion chromatogram (TIC) obtained from QTF analysis. 7

Data analysis was performed using the MFE and MFG algorithms of the MassHunter Qualitative Analysis software. Using an MFE algorithm, high resolution accurate mass spectral information of all the sample components was extracted. From the acquired data, the MFE algorithm listed two entities, the m/z values of the entities were 298.127 and 312.1416 (Figure 6). The m/z 298.127 corresponds to the duloxetine API and the second entity corresponded to the unknown impurity. Using the MFG algorithm, a relevant list of candidate molecular formulas for each entity was tabulated and it ranked them according to the relative probabilities (Figure 7). 10 1 1 0.8 Total compound chromatogram (TCC) from MFE +EI TCC can Frag = 90.0 V Duloxetine Counts (%) 0.6 0.4 0.2 0 14.4 14.8 1.2 1.6 16 16.4 16.8 Acquisition time (min) Counts (%) +EI can (16.63 min) Frag = 90.0 V Duloxetine 10 10 2 298.127 6 1.6 m/z of Duloxetine (M+H) + 1.2 4 3 14.0682 0.8 2 0.4 02.3733 1 0 0 100 200 300 400 00 600 700 800 900 1000 Mass-to-charge (m/z) Counts (%) +EI can (16.708 min) Frag = 90.0 V Duloxetine 312.1416 m/z of unknown impurity 168.0838 200 300 400 00 600 700 800 900 Mass-to-charge (m/z) Figure 6 Mass spectra of the API (298.127) and impurity (312.1416). 8

MFE MFG + + + _ how/hide Cpd Label Formula core Mass Avg Mass Mass (MFG) Diff (MFG, ppm) Diff (MFG, mda) Base Peak RT TRUE 1 Cpd 1: 16. 297.1184 297.2948 298.127 16.7 TRUE 2 Cpd 2: 16.701 311.1343 311.372 312.1416 16.701 Mass Diff (MFG, Diff (MFG, how/hide Cpd Label Formula core Mass Avg Mass (MFG) ppm) mda) Base Peak RT TRUE 1 Cpd 1: C18 H19 C18 H19 98.18 297.1184 297.3848 297.1187 1.07 0.32 298.127 16.7 TRUE 2 Cpd 2: C19 H21 C19 H21 98.4 311.1343 311.399 311.1344 0.29 0.09 312.1416 16.701 _ Best Formula core Mass Mass (MFG) core (MFG) Diff (abs. ppm) Diff (mda) ID ource Diff (ppm) RT TRUE C19 H21 98.4 311.1343 311.1344 98.4 0.29 0.09 MFG 0.29 16.701 _ pecies Ion Formula m/z Height core (MFG) core (MFG, M) core (MFG, mass) core (MFG, abund) core (MFG, iso. spacing) (M+H)+ C19 H22 312.1417 62137.9 98.4 98.4 99.94 9.18 99.79 Height (Calc) Height um% (Calc) Height % (Calc) m/z (Calc) Diff (mda) Height Height % Height um % m/z Diff (ppm) 4200.1 76.8 100 312.1417 0.1 62137.9 100 79.2 312.1416 0.29 119921.6 16.9 22 313.1448 0.1 11726.6 20.6 16.3 313.1447 0.3 3807.8.4 7 314.1412 0. 27089.3 4.8 3.8 314.1417 1.62 622. 0.9 1.1 31.1424 0.6 492.6 0.9 0.7 31.1419 1.82 693.8 0.1 0.1 316.1441 0.9 264.4 0 0 316.1432 2.8 Figure 7 MFE/MFG results for Duloxetine and impurity. 9

The accurate mass information of the precursor ion and fragment ions of API and unknown impurity from the MFE/MFG algorithm were uploaded to the MC software and searched against the Chempider database to retrieve all possible structures. The details of the MC software and multiple approaches to identify impurities can be found in Agilent publication 991-137E 2. Multiple candidate structures were retrieved for both entities with their calculated correlation scores. ne of the listed entities was confirmed to be Duloxetine API. Figure 8 shows a screen shot of results from MC software for unknown impurity. The overall MFG score for the selected precursor ion, the rank of the MC proposed structure, and the structure correlation score for impurity are highlighted in red circles. The structure with the highest correlation score (93.26%) corresponds to methyl derivative of Duloxetine API. A C B D E Figure 8 creen shot of MC results for Identification of unknown impurity. A: List of possible molecular formulas for the precursor ion of impurity B: MFG results for the product ions of the impurity C: Candidate structures for the unknown impurity D: Fragment ions for the candidate structure selected in panel C E: ubstructure assignments for a selected fragment ion in panel D 10

Figure 9 summarizes the structure elucidation of the unknown impurity from the MC. This impurity was further confirmed by spiking the impurity standard. Main fragments Mass dm (ppm) Best score 1 8.06 _ 7.1 97.0 2 72.0803 6.0 96.9 3 8.0893 _ 7.9 93.9 4 123.021 9.8 96.7 17.0629 12.0 8.7 6 169.0907 7.8 94.1 Final structure of the impurity m/z: 8.06 m/z: 72.0803 m/z: 8.0893 H 3 C H 3 C H 3 C H 3 C m/z: 169.0907 m/z: 17.0629 m/z: 123.021 Molecular formula C 19 H 21 Calculated mass 311.1344 Measured mass 311.1343 dm (ppm) 0.3 H 3 C H 3 C H 3 C Figure 9 tructure elucidation of duloxetine impurity (m/z: 311.1344) demonstrating wide usability of MC software to assign structures for each fragment ions. Figure 10 shows the increased area percentage from the API sample spiked with the impurity standard. mau 20 1 10 0 Duloxetine API Duloxetine API spiked with impurity 0 2 4 6 8 10 12 14 min Figure 10 Area percentage increment of impurity peak using DAD detection when methyl derivative of Duloxetine was spiked into the API. 11

Conclusion This Application ote demonstrates a workflow for the selective identification of an impurity. This workflow used a heart cut two-dimensional LC/M method to identify an impurity in Duloxetine API. This method eliminated the need to develop an M compatible LC method, for a non-m compatible pharmacopeia method to perform accurate mass analysis of the impurity of the interest. It enabled specific selection of a close eluting, low abundant unknown impurity in the API for M analysis. An Agilent 640 Q-TF along with advanced software tools (MC, MFE, and MFG) software enabled quick and reliable identification of unknown compounds. References 1. International Conference on Harmonization of Technical Requirements for Registration of Pharmaceuticals for Human Use, ICH teering Committee, March 30, 199. 2. Agilent publication 991-137E Pharmaceutical Impurity Identification and Profiling Using Agilent Q-TF LC/M Combined with Advanced MassHunter Data Processing oftware dated ovember 21, 2012. 3. Validated reverse phase method for duloxetine from International Journal of Comprehensive Pharmacy (IJCP), Vol 01, and Issue 03. Viriyala R. K. et al. Pharmacie Globale (IJCP), 3 (03) 2010. www.agilent.com/chem Agilent Technologies, Inc., 2013 Published in the UA, March 1, 2013 991-1873E