Sensitive and flexible bacterial endotoxin testing with the EndoZyme II Recombinant Factor C Assay using the Synergy HTX Microplate Reader

Similar documents
Pierce LAL Chromogenic Endotoxin Quantitation Kit

Limulus Amebocyte Lysate (LAL) QCL-1000

PyroGene Recombinant Factor C Endotoxin Detection System

ab Plasmin Activity Assay Kit (Fluorometric)

ImmuLux Human IL-6 Fluorescent ELISA Kit

ab Factor IXa Activity Assay Kit (Fluorometric)

LIMULUS AMEBOCYTE LYSATE PYROGENT Catalog No. N383, N384 License No. 709

ab Factor Xa Inhibitor Screening Assay Kit (Fluorometric)

ab Factor VIIIa Activity Assay Kit (Fluorometric)

ab Plasmin Inhibitor Screening Assay Kit (Fluorometric)

The New FDA Guidance for Industry Pyrogen and Endotoxins Testing: Questions and Answers. LALUpdate

Product datasheet. Storage recommendations Store the kit at 2-8 C. The kit is stable for a period of up to 3 months from the date of receipt.

Porcine IgM (Immunoglobulin M) ELISA Kit

Application note. Guideline for validation of analytical methods using Cedex Bio, Cedex Bio HT, and Cedex HiRes Analyzers.

Validation of a concentration assay using Biacore C

IDTox Lactate Dehydrogenase (LDH) Enzyme Cytotoxicity Kit For Cell Culture Supernatant Samples

ab Porcine VEGF-A ELISA Kit

Limulus Amebocyte Lysate (LAL) Rapid Endotoxin Detection

ab GFP ELISA Kit Instructions for Use For the quantitative measurement of GFP protein expression

ab99978 BDNF Human ELISA Kit

Data Sheet. PD-1[Biotinylated]:PD-L2 Inhibitor Screening Assay Kit Catalog # Size: 96 reactions

HIT302 LIMULUS AMEBOCYTE LYSATE CHROMOGENIC ENDPOINT ASSAY

Protein Quantitation Kit (Bradford Assay)

High Throughput Quantitation of Cytokine Biomarkers using LANCE Ultra TR-FRET Assays

ab IL-1 alpha (Interleukin-1 alpha) Human ELISA Kit

Human connective tissue growth factor (CTGF) ELISA Kit. MyBioSource.com. This package insert must be read in its entirety before using this product.

ab83355 ATP Assay Kit (Colorimetric/ Fluorometric)

ab TNF alpha Human ELISA Kit

A Comparison of AlphaLISA and TR-FRET Homogeneous Immunoassays in Serum-Containing Samples

RayBio Rat IL-6 ELISA Kit (For Lysates)

Human Serum Albumin Assay

Mouse ALP (Alkaline Phosphatase) ELISA Kit

Human myelin basic protein(mbp) antibody ELISA Kit

CHO Host Cell Proteins 3 rd Generation

MRC5 Host Cell Proteins

Your Analyte ELISA Kit Instruction

Test Your Plate Reader Set-up Before Using LanthaScreen Eu Assays

Human IL-6 EasyTest TM ELISA Kit

Bovine Serum Albumin (BSA) Assay

Global Headquarters 86 Cummings Park Woburn, MA Tel:

Mouse Heparan Sulfate(Heparan Sulfate) ELISA Kit

Limit of Detection. Theoretical Lowest Concentration

Rat α-melanocyte stimulating hormone (α-msh) ELISA Kit

RayBio Human Caspase-3 ELISA Kit

Data Sheet. PD-1:PD-L1[Biotinylated] Inhibitor Screening Assay Kit Catalog # Size: 96 reactions

ab VEGF Receptor 2 Human ELISA Kit

Bovine prolactin/luteotropic hormone (PRL/LTH) ELISA Kit

Human placenta lactogen (HPL) ELISA Kit

Human Leukotriene E4 (LTE4) ELISA Kit

Bovine Prostaglandin E2 (PG-E2) ELISA Kit

Automated Protocol for ANTI-FLAG High Sensitivity, M2 coated 96-well plate Using the Sciclone ALH 3000 Workstation (Caliper Life Sciences)

HTRF (Cisbio) Human Interleukin beta (IL1β) assay Tecan Infinite F500, TR-FRET (HTRF )

Human cross linked N-telopeptide of type I collagen (NTX) ELISA Kit

Mouse Collagen Type III ELISA

Human vascular endothelial cell growth factor A (VEGF-A) ELISA Kit

GSI Equine IL-10 ELISA Kit- Cell Lysate DataSheet

RayBio Custom ELISA Kit

ab Human IGF1 SimpleStep ELISA Kit

Nori TM Canine TGFβ2 ELISA Kit DataSheet

CytoScan WST 1 Cell Cytotoxicity Assay

Mouse Luteinizing Hormone (LH) ELISA

Global Headquarters 86 Cummings Park Woburn, MA Tel:

ab Adenosylhomocysteinase (AHCY) Inhibitor Screening Assay Kit (Fluorometric)

TACS MTT Assays. Cell Proliferation and Viability Assays. Catalog Number: TA tests. Catalog Number: TA tests

Universal Kinase Assay Kit (Fluorometric) For monitoring ADP formation, which is directly proportional to enzyme phosphotransferase activity

Part. no. PV4319.pps Rev. date 29 June 2006

Human immunoglobulin G(IgG) ELISA Kit

Human IL-10 ELISA MAX Set Deluxe

Human Alpha-2-Macroglobulin ELISA

Human IgM Ready-SET-Go!

Synergy NEO HTS Multi-Mode Microplate Reader and Life Technologies Reagents

Canine Creatinine(Cr) ELISA kit. Catalog No. MBS (96T)

LDH-Cytotoxicity Assay Kit II

LDH-Cytox Assay Kit. A Colorimetric Cytotoxicity Measuring Kit. Cat. No LDH-Cytox Assay Kit can be used to measure cytotoxicity in vitro

Enzyme unit definitions and assay design

ab46042 IL-6 (Interleukin-6) High Sensitivity Human ELISA Kit

Mouse Factor XII Total ELISA Kit

NAG-1 (GDF-15, MIC-1) Prostate Cancer ELISA kit

RayBio Human HGF ELISA Kit

QuantiFluor ONE dsdna System

MyBioSource.com. Human VEGF ELISA Kit

alamarblue Technical Datasheet

Erin Patton, MS Senior Product Specialist Charles River Labs, Microbial Solutions

Technical Note. Housekeeping Protein Validation Protocol

Human Amyloid Beta Peptide 1-42 (Aβ1-42) ELISA Kit

PyroSense PAT for WFI By Ken Nichols and Maribeth Donovan Janke

Porcine Transferrin Receptor(TFR) ELISA Kit

ab65354 Superoxide Dismutase Activity Assay kit (Colorimetric)

Glucose 6 Phosphate Dehydrogenase Assay Kit (Colorimetric)

HTRF KinEASE STK discovery kit

ab Interferon-gamma (IFNG) Human SimpleStep ELISA Kit

Mouse TNF alpha ELISA Kit

Human immunoglobulin G(IgG) ELISA Kit

Rat Tumor Necrosis Factor Alpha (TNF-α) ELISA

Converting your ELISA from horseradish peroxidase to alkaline phosphatase using NovaBright chemiluminescence detection reagents

Tnf (Rat) ELISA Kit. Catalog Number KA assays Version: 02. Intended for research use only.

Simple Conversion of ELISA to PerkinElmer s High Sensitivity DELFIA Technology

Transcription:

217 Photo Ruder, Hyglos GmbH - a biomérieux company. Sensitive and flexible bacterial endotoxin testing with the EndoZyme II Recombinant Factor C Assay using the Synergy HTX Microplate Reader Rudolf Wedmann, Thomas Uhlig EndoZyme II is a bacterial endotoxin test for pharmaceuticals and medical devices as well as their in-process intermediates and raw materials. The assay relies on recombinant horseshoe crab Factor C (rfc), an exact synthetic copy of the endotoxin-sensitive enzyme naturally harboured by the blood of horseshoe crabs. Compared with methods sourced from these animals, rfc methods offer better specificity, flexibility, lot-to-lot consistency and, importantly: a sustainable and secure source. This article covers how the EndoZyme II assay is established, optimised and validated using the Synergy TM HTX Multi-Mode Microplate Reader with Gen5 TM software. Introduction Endotoxin consists of lipopolysaccharides (LPS) that are ubiquitous among Gram-negative bacteria. Hence, endotoxin can be found in any place inhabited by these microbes, for instance contaminated water systems. Since endotoxin is a pyrogen, i.e. a fever-inducing agent, it may cause severe immune reactions upon contact with the systemic circulation of a patient. Therefore, injectable pharmaceuticals and implantable medical devices must be tested for endotoxin to guarantee patient safety. To this end, bacterial endotoxin tests (BET), such as the recombinant Factor C (rfc) assay EndoZyme II, are used. rfc is an exact synthetic copy of the endotoxin-detecting enzyme harboured by the blood cells, specifically amebocytes, of horseshoe crabs. In EndoZyme II, rfc is activated by endotoxin and in turn processes a fluorogenic substrate. The more endotoxin rfc detects, the stronger the increase in fluorescence. This signal development is measured by a fluorescence microplate reader such as the Synergy HTX from BioTek Instruments, Inc.

Classical bacterial endotoxin tests (BETs) like the Limulus Amebocyte Lysate (LAL) test still rely on sourcing reagents from the blood of horseshoe crabs that are currently considered as vulnerable and their population is in decline [1]. LAL quality may vary across individual animals and seasons. Furthermore, they show false-positive reactions in the presence of betaglucans. In contrast to the classical BETs, EndoZyme II is entirely sustainable, more consistent across lots (Figure 2 B) and specific to endotoxin, but nevertheless equivalent or better in terms of sensitivity and the determination of endotoxin concentration. Assay handling and data analysis For performing EndoZyme II, initially, all kit components are brought to room temperature. The supplied control standard endotoxin (CSE) is reconstituted in a volume of water as labelled on the respective bottle and the Certificate of Analysis. Serial dilutions of the standard and, if necessary, the samples are prepared. µl of each standard dilution, the negative control and the samples are transferred onto a microplate, at least in duplicate. Defined amounts of the CSE are spiked into aliquots of the samples as positive product controls. The plate is heated to 37 C in the Synergy HTX. Subsequently, µl assay reagent, consisting of previously mixed eight parts Assay Buffer, 217 Photo Ruder, Hyglos GmbH - a biomérieux company. Synergy is a registered trademark of BioTek Instruments, Inc. Furthermore, the blank drfu are subtracted from all other drfu () to adjust the instrument-specific baseline (Figure 1C). For calculating the endotoxin activity of standards, samples and positive controls in endotoxin units per millilitre (EU/mL), a linear equation is fit to the logarithm of each plotted against the logarithm of the EU/mL of each standard point in a standard series (Figure 2A). This is the standard curve which must demonstrate a correlation RFU 47 46 45 44 43 4 A B C 42-4 Replicate 1-2 Replicate 2 4 Replicate 3-3 15 3 45 6 15 3 45 6 drfu 3 2 drfu 4 3 2 - -2-3 Blank.5 EU/mL 15 3 45 6 Figure 1 EndoZyme II reaction kinetics. A The fluorescence intensity of three replicates of.5 EU/mL with EndoZyme II reagents was measured over time and quantified in relative fluorescence units (RFU). B The difference in RFU from time point zero (drfu) was calculated for each of the same replicates of.5 EU/mL and each measurement time point, respectively. C The average drfu of blank (water) and.5 EU/mL was calculated for each time point. one part Substrate and one part rfc-enzyme, are added to the filled wells of the microplate. The Synergy HTX accommodates the microplate, shakes it for 15 seconds, incubates it at 37±.5 C and measures fluorescence intensity with excitation at 38/2 nm and emission at 44/4 nm in relative fluorescence units (RFU) at several time points. For all filled wells, the difference in RFU between the time point zero and 6 minutes is calculated (drfu, Figure 1A and B). coefficient (r) of.98 as defined by the internationally harmonized Pharmacopoeia chapters on the bacterial endotoxin test [2 4]. Per the same regulation, the positive product control (PPC) must show an endotoxin spike recovery of 5 2 % of the expected concentration. Generally, EndoZyme II s linearity is very high up to 5 EU/mL (r.998), in turn guaranteeing excellent accuracy. A B Lot 18238 Lot 18434 Lot 18587,1,1,1 1,1,1,1 1 Standard endotoxin concentration Standard endotoxin concentration Figure 2 EndoZyme II standard curves. A The standard curve of EndoZyme II lot 18434 demonstrates the linear correlation of endotoxin concentration and fluorescence intensity increase in. Error bars denoting the standard deviation of each standard. B Standard curves from different EndoZyme II lots are compared.

Limit of quantitation 1,1,1,1,1 15 3 45 6 75 9 5 12 A Limit of quantitation,1,5 5 6 7 8 9 1 12 Gain B Limit of quantitation,1,5 C 1 5 175 255 Number measurements per data point Limit of quantitation 1,1,1,1,1 44/3 44/4 46/4 485/2 38/2 Emission filter (peak/bandwidth) [nm] 36/4 D 34/3 Excitation filter (peak/bandwidth) [nm] Figure 3 EndoZyme II limit of quantitation (LOQ) depends on different measurement parameters. The LOQ was calculated as the blank s -fold standard deviation added to the average of the blank and converted into EU/ /ml per a standard curve covering the linear range at the respectivee time point. Six blank replicates were used, respectively. Unless indicated otherwise, the fluorescence intensity development in the course of 6 min was recorded, gain 7, 5 measurements per data point as well as filters excitation 38/2 and emission 44/4 were used. A In the course of 12 min assay reaction time, the LOQ was calculated every 5 min. The turquoise line indicates a log-log-linear function fitted to the mean LOQ of the EndoZyme II lots 18238, 18434 and 18587. The turquoise area covers the range between the lowest and highest LOQ determined with these reagent lots. Orange lines represent.5,.5 and.1 EU/mL. B The LOQ was determined depending on the instrument gain. C The LOQ was determined depending on the number of measurements that each data point represents. D The LOQ was calculated for combinations of three different excitation filters and four different emission filters. Instrument and software set-up Setting up the Synergy HTX is simple and fast. It only requires connecting the device to a Windows PC via USB, installing the co-delivered analysis software Gen5 and choosing the instrument from a list in the software. Gen5 Secure is a 21 CFR part 11-compliant version for regulated environments of the pharmaceutical and medical device industry. For qualifying the instrument, we recommend directly running EndoZyme II assays. The readout of fluorescence intensity is relative and depends on the sensitivity of the light detector, the so-called gain. Ideally, the gain is adjusted to yield a drfu of 5 for an endotoxin concentration of.5 EU/mL. A serial dilution of EndoZyme II s control standard endotoxin, applied in triplicate, should show good linearity, i.e.. r of minimum.98 (see above). Meanwhile, a single dilution of the standard, e.g..5 EU/mL, should demonstrate appropriate uniformity across the entire plate, i.e. a drfu coefficient of variation (CV) of maximum 25% for a total 96 replicates. biomérieux provides Gen5 templates, so-called protocols, for the above-mentioned qualification trials as well as for routine endotoxin testing. Users only need to enter trial-specific information such as the reagent lot number, the list of sample identities and their respective dilution factors. Each protocol specifies all measurement parameters for the Synergy HTX, extracts the RFU and processes them. Finally, it returns all relevant endotoxin testing data in customizable reports that can be printed, signed and filed. The Synergy HTX can also contain a module that measures light absorbance at flexible wavelengths, a so-called monochromator. Thus, users who intend to gradually switch to EndoZyme II may also use it for any classical BET or other absorbance-based assays. Assay optimisation For BETs, the quantitation limit is commonly termed sensitivity. In classical BETs, the sensitivity is limited by the background activity of the reagents, i.e. reaction in the absence of endotoxin. EndoZyme II, however, does not display such background activity. Hence, the sensitivity mostly depends on the measurement background of the instrument and the assay reaction time. Indeed, the standard protocol of EndoZyme II exhibits a state-of-the-art sensitivity of.5 EU/mL after 6 minutes reaction time. Non-interferinwater, do not have to be tested at overly sensitive levels. For samples such as instance, water for injection has an endotoxin limit of.25 EU/mL. After a reaction time of only 15 min, EndoZyme II clearly resolves.5 EU/mL (Figure 3A), sufficient for such a sample. On the other hand, after 12 minutes, the assay can reliably resolve as little as.1 EU/mL. Such a high sensitivity is particularly important for medical devices. The higher the sensitivity, the higher the volume of water used for rinsing medical devices before testing the water for endotoxin. Thus, bacterial endotoxin testing can cover larger devices or more

devices within a single water pool. Furthermore, pharmaceuticals exhibiting strong test interference can be tested in higher and in turn less interfering dilutions. In fact, the sensitivity of EndoZyme II does not only depend on the reagents, but also on the instrument, here the Synergy HTX. Increasing the gain (Figure 3B) and the number measurements per data point (Figure 3C) can also enhance the sensitivity. In fact, different sets of light filters can be used for fluorescence excitation and emission reading, but 38/2 and 44/4 nm represent the respective optima (Figure 3D). When EndoZyme II is validated for testing further products, a test for interfering factors with several product dilutions can be used to establish, if necessary, the dilution which allows for 5 2 % spike recovery. Using this dilution or non-interfering undiluted product, accuracy, precision, quantitation limit, linearity and range can be demonstrated by a single method suitability test. biomérieux provides validation protocols to capture user testing results for both initial validation and method suitability, thus reducing the documentation effort for EndoZyme II users. Recovery of reference endotoxin concentration [%] 2 15 5 18 6 In contrast to the LAL test, EndoZyme II s combined assay reagents and the plate with samples can be incubated for 18 min without major influence on the accuracy of the endotoxin concentration determination; it remains well in the range of 5 2 % of the expectation (Figure 4). The same holds true for slight deviations in the timing of the reading, as the endotoxin standards always provide an equally affected internal reference. Given the low time-sensitivity, EndoZyme II plates can be held in a 37 C incubator for the time of the reaction, while another plate is analysed in the Synergy HTX in parallel. Accordingly, EndoZyme II can be performed with high throughput. Method validation For establishing bacterial endotoxin testing using EndoZyme II, a recognized alternative to the LAL test, the assay needs to be validated as a quantitative analytical procedure [5,6,7] following exactly the same methodology as classical BETs as per global Pharmacopoeias [2 4]. This encompasses the demonstration of suitable accuracy, precision, specificity, quantitation limit, linearity, range and robustness. Accuracy, precision, linearity and range can be evaluated by a single set up that is repeated four times. Analysts, days and reagent lots are cross-varied for this purpose. The quantitation limit only needs to be demonstrated once for two reagent lots each. Robustness has sufficiently been tested by biomérieux. Therefore, the user does not need to provide respective data. Specificity has already comprehensively been shown to be equivalent between classical and recombinant BETs such as EndoZyme II by the Japanese Pharmaceuticals and Medical Devices Agency (PMDA). Their multi-centre ring study [8] featured three assays per category and a total 29 endotoxin samples from diverse bacterial strains and of different extent of purification. Thus, the user only needs to show that endotoxin can be quantified in their test samples with acceptable accuracy and precision. Sample hold time in plate [min] 6 18 Assay reagent hold time [min] Figure 4 Influence of different hold times of samples and reagents on EndoZyme II accuracy. EndoZyme II assay reagent was incubated for different hold times and added to.15 EU/mL which was also incubated for different hold times in the assay plate. Each combination of assay reagent and endotoxin sample hold time was tested in triplicate. A standard curve from.5 to 5 EU/mL with min hold time of both assay reagent and endotoxin was used for endotoxin concentration quantitation. Conclusion EndoZyme II is a sensitive, flexible and sustainable bacterial endotoxin test with an unprecedented lot-to-lot consistency that is ideally used with the Synergy HTX. The assay can be optimised for the user s specific needs. Finally, the validation is straightforward and concluded within a few days. RUDOLF WEDMANN is Endotoxin Application Specialist at Hyglos-bioMérieux. He holds a PhD in chemistry from the University of Erlangen, with focus on protein biochemistry. THOMAS UHLIG is Endotoxin Technical Support Manager at Hyglos-bioMérieux and holds a Professional Doctorate of Engineering from Delft University of Technology in bioproduct design. www.biomerieux-industry.com BIOMERIEUX, the blue logo, hyglos and EndoZyme are used and/or pending and/or registered trademarks belonging to biomérieux, or one of its subsidiaries, or one of its companies.

References [1] Smith et al., Limulus polyphemus. The IUCN Red List of Threatened Species 216: e.t11987a815983. 216. Downloaded on 3 November 217. http://dx.doi.org/.235/iucn.uk.216-1.rlts.t11987a815983.en. [2] The United States Pharmacopeial Convention, <85> Bacterial Endotoxins Test, in The United States Pharmacopeia 39 and The National Formulary 34 (USP 4 NF39), Rockville, MD, USA, 216, pp. 161-166. [3] EDQM Council of Europe, 2.6.14. Bacterial endotoxins, in European Pharmacopoeia 8.8, Strasbourg, France, 216, pp. 194-198. [4] The Ministry of Health, Labour and Welfare, Pharmaceuticals and Medical Devices Agency, 4.1 Bacterial Endotoxins Test, in The Japanese Pharmacopoeia, Seventeenth Edition (JP17), Tokyo, Japan, 216, pp. 1-114. [5] International Council for Harmonisation of Technical Requirements for Pharmaceuticals for Human Use, ICH Harmonised Tripartite Guideline - Validation of Analytical Procedures: Text and Methodology Q2(R1), 25. [6] The United States Pharmacopeial Convention, <1225> Validation of Compendial Procedures, in The United States Pharmacopeia 39 and The National Formulary 34 (USP 4 NF39), Rockville, MD, USA, 216, pp. 161-166. [7] Bolden et al., Application of recombinant Factor C reagent for the detection of bacterial endotoxins in pharmaceutical products, PDA Journal of Pharmaceutical Science and Technology, 217 Jul 2. [8] Kikuchi et al., Collaborative Study on the Bacterial Endotoxins Test Using Recombinant Factor C-based Procedure for Detection of Lipopolysaccharides, Pharmaceutical and Medical Device Regulatory Science, 48(4), 217, pp- 252-26.