A protocol for the determination of total cell concentration of natural microbial communities in drinking water with FCM

Similar documents
A flow cytometric method for AOC determination

Application Note. Yeast Analysis with a life - dead staining kit (Yeast control - viability)

Identification of red and white blood cells from whole blood samples using the Agilent 2100 bioanalyzer. Application Note

Phagocytosis Assay Kit (IgG FITC)

PERFECT-COUNT MICROSPHERES

An evaluation of volatile suspended solids as a true measure of. metabolic activity in activated sludge

Titration of Fluorochrome-Conjugated Antibodies for Labeling Cell Surface Markers on Live Cells

Application Information Bulletin: Set-Up of the CytoFLEX Set-Up of the CytoFLEX* for Extracellular Vesicle Measurement

High Purity Simultaneous Cell Sorting of Nano and Large Phytoplankton Using the MoFlo Astrios EQ

Flow Cytometry - The Essentials

READ ME FIRST. In order to assure high quality results, pay particular attention to:

SYTOX Green Dead Cell Stain

ab Hypoxic Response Human Flow Cytometry Kit

determine optimum instrument settings for their own instruments and establish their own daily values.

Flow Cytometry Support Reagents

Lab 2-Microbial Enumeration

INVESTIGATING BIOCHEMICAL INDICATORS OF TREATMENT EFFICIENCY IN WASTE STABILISATION PONDS THROUGH THE USE OF FLOW CYTOMETRY

Nuclear Condensation Assay Kit Green Fluorescence

This Document Contains:

ab Apoptosis/Necrosis Detection Kit (blue, red, green)

CyFlow Space Your flexible flow cytometer

Electronic Supplementary Material (ESI) for Journal of Materials Chemistry This journal is The Royal Society of Chemistry 2011.

CyFlow Space Your flexible flow cytometer

Does the High Nucleic Acid Content of Individual Bacterial Cells Allow Us To Discriminate between Active Cells and Inactive Cells in Aquatic Systems?

CyFlow Cube series Appealing from every angle

ab Cellular Reactive Detection Assay Kit (Deep Red Fluorescence) Version 3 Last updated 22 December 2016

PROCEDURE FOR USE NICKEL NTA Magnetic Agarose Beads (5%)

Microbiological tests not just for safe drinking water. PRACTICE REPORT MICROBIOLOGY Author: Dr. Elmar Grabert, HACH LANGE GmbH, Germany

FACS Blue LacZ beta Galactosidase detection kit

IDEXX Summary. IDEXX Laboratories. Date: April Report Highlights:

!! PLEASE READ BEFORE USE!!

Franzens-Universitaet Graz, Humboldtstrasse 50, 8010 Graz. Phone: ++43 (0) Fax: ++43 (0)

TITLE: LIVE/DEAD VIABILITY FOR CLINICAL SAMPLES

Department of Microbiology, Lab 016 instructions

QS S Assist KINASE_MSA Kit

Highly Accurate, Reliable and Real-time Enumeration of Individual Bacteria for the Determination of Raw milk Hygienic Quality

No-wash, no-lyse detection of phagocytic cells via a phrodo BioParticles functional assay in human whole blood on the

Flow Cytometry for Ploidy and Genome Size analysis - 1. Functional principles

Lab Exercise: Examining Water Quality: Most Probable Number & Colilert Test Kit Lab

OxiSelect ROS Assay Kit

Neutrophil/Monocyte Respiratory Burst Assay Kit

CRITICAL ASPECTS OF STAINING FOR FLOW CYTOMETRY

Different Potential of Extracellular Vesicles to Support Thrombin Generation: Contributions of Phosphatidylserine, Tissue Factor, and Cellular Origin

Immunofluorescence Staining Protocol for 3 Well Chamber, removable

Before You Begin. Calibration Protocols

ISO INTERNATIONAL STANDARD

SANTA CRUZ BIOTECHNOLOGY, INC.

1. Materials Provided Additional Materials Required General Advice About Treatment and Sampling Times... 2

TACS MTT Assays. Cell Proliferation and Viability Assays. Catalog Number: TA tests. Catalog Number: TA tests

Introduction. Technical Note

qpcr Kit, DNA-free Product components 100 rxn 250 rxn Product description

NovoCyte Flow Cytometer

Outline. Sampling purpose D-2 standard Ballast water biology Indicative vs detailed analysis Organisms detection methods Viability Minimum dimension

Application Note. Assay Portability on the BD FACSVerse System. Summary. Maria Jaimes, Yibing Wang, Catherine McIntyre, and Dev Mittar

Laboratoire Arago Observatoire Océanologique Banyuls-sur-mer

QuantiFluor ONE dsdna System

ExoGlow -NTA Fluorescent Labeling Kit

EdU Flow Cytometry Kit. User Manual

Summary of Mutagenic Toxicity Test Results for EvaGreen

Whole Mount IHC Protocol

Methodology. Protocol for plant genome size estimation by flow cytometry (2) GSAD: a genome size in the Asteraceae database

Invertase Assay Kit. Catalog Number KA assays Version: 03. Intended for research use only.

PowerSoil DNA Isolation Kit

FLOW CYTOMETRY. CyAn ADP. Analyzer

Why use Individual Bacteria Count instead of Colony Forming Units?

CyFlow Space. Your Flexible Flow Cytometer.

Flow Cytometry SOP: 14 color flow for immune activation, senescence, and exhaustion

Modified mtec Agar, Colilert, and M-FC Agar - Field Trial Comparison of Bacteria Enumeration Methods in Surface Waters of Eastern Wyoming

Optimization of a LanthaScreen Kinase assay for BRAF V599E

Electronic Supplementary Information

Large Volume Serial Dilutions:

Tali Viability Kit Dead Cell Green

EZ-Vision DNA Dye as Loading Buffer

The preparation of native chromatin from cultured human cells.

Flow Cytometry. Cell Sorting 34 Instruments 34 Consumables 38

Qdot nanocrystal. wide range of biological investigations, Qdot nanocrystals are powerful complements

QIAfilter Plasmid Midi Kit (Cat #: 12243)

LDH-Cytox Assay Kit. A Colorimetric Cytotoxicity Measuring Kit. Cat. No LDH-Cytox Assay Kit can be used to measure cytotoxicity in vitro

Protocol for FACS analysis of HeLa cell transfectants

Cytotoxicity LDH Assay Kit-WST

Analytical Quality Control:

ab Fatty Acid Oxidation Human Flow Cytometry Kit

NOTE ACRYLAMIDE IS NEUROTOXIN YOU MUST WEAR GLOVES.

Bacterial Counts - Quantitative Analysis of Microbes

2. SUMMARY AND EXPLANATION

Environmental. National Legionella Conference Sample Taking Workshop. Right Solutions Right Partner

NADP+/NADPH Assay Kit

PrepSEQ Residual DNA Sample Preparation Kit

Biological Instability in a Chlorinated Drinking Water Distribution Network

Microbiological Quality: Understanding Drinking Water Quality and Management

Technical Note Detection of post-immunoprecipitation proteins by Western blot using the Quick Western Kit IRDye 680RD

QS S Assist STK_FP Kit

Improved Monitoring of P. aeruginosa on Agar Plates

In-Cell Western Kits I and II

XS-1000i New Sysmex 5-part diff haematology analyser with fluorescence technology

sirna Transfection Reagent

Lab Exercise 13: Growth Curve

A Comparison of the Bacteriological Quality of Drinking Water for the City of Johannesburg in the Wet and Dry Season

Technology for Real-Time Detection of Microbes in Water

PARP-1 (cleaved) Human In-Cell ELISA Kit (IR)

Transcription:

Techneau June 2007 A protocol for the determination of total cell concentration of natural microbial communities in drinking water with FCM Deliverable 3.3.7

Techneau June 2007 A protocol for the determination of total cell concentration of natural microbial communities in drinking water with FCM Deliverable 3.3.7 2006 TECHNEAU TECHNEAU is an Integrated Project Funded by the European Commission under the Sixth Framework Programme, Sustainable Development, Global Change and Ecosystems Thematic Priority Area (contractnumber 018320). All rights reserved. No part of this book may be reproduced, stored in a database or retrieval system, or published, in any form or in any way, electronically, mechanically, by print, photoprint, microfilm or any other means without prior written permission from the publisher

Colofon Title A protocol for the determination of total cell concentration of natural microbial communities in drinking water with FCM. Author(s) Frederik Hammes Michael Berney Marius Vital Thomas Egli. Quality Assurance By J. van den Broeke (s::can) Deliverable number Deliverable 3.3.7 This report is: PU = Public

Summary The conventional approach to characterize the general microbial quality of drinking water is the use of heterotrophic plate counts (HPC), which enumerate the bacterial cells which can grow up to form visible colonies on specific semi-solid media at selected incubation temperatures within a selected time period. On average, about 1 % of the total bacteria in drinking water are culturable with standard HPC methods. A method for the rapid determination of total bacterial cell concentrations in drinking water, using fluorescent staining and flow cytometry, has been established and standardised. The method uses SYBR Green I which binds to microbial nucleic acids, is excited by a laser at 488 nm, and emits fluorescence at 520 nm, which can then be detected with a flow cytometer. The FCM method is fast (10 minutes required for staining and 3 minutes for the analysis). The calibrated flow cytometer has an overall error of less than 5%, inclusive of instrument and operator error. It is possible to detect cell concentrations as low as 200 cells/ml with this method, but all evidence up to now suggests that this is well below the typical range of drinking water (c.a. 1 x 10 5 cells/ml). The value of accurate total cell counting is evident, e.g. when treatment plant operators aim to monitor the affectivity of specific processes such as membrane filtration or ozonation, or when regrowth during treatment (e.g. GAC filtration) is studied. TECHNEAU - 1 - June, 6 2007

Contents Contents 2 1 Introduction 3 1.1 Background 3 1.2 Goals 4 1.3 Acknowledgement 4 2 Total cell concentration 5 2.1 General overview 5 2.2 Apparatus 5 2.3 Reagents 5 2.4 Procedure 6 2.5 Calculations 6 3 Discussion 7 4 Conclusion 10 5 References 11 TECHNEAU - 2 - June, 6 2007

1 Introduction 1.1 Background High concentrations of planktonic bacterial cells (> 1 x 10 4 cells ml -1 ) of broadly diverse populations are commonly found in both bottled mineral water and tap water upon consumption (Figure 1) (Leclerc and Moreau, 2002; Hoefel et al., 2003; Berry et al., 2006; Hammes et al., 2007). The primary objectives of drinking water treatment (from a microbiological perspective) are to ensure the absence of any pathogenic bacteria in the finished product, and to limit any uncontrolled regrowth during distribution of the water. Irrespective of the different disinfection processes, it is common for bacteria to regrow during treatment and distribution. ground water bottled mineral water tap water river/lake water 1.E+00 1.E+01 1.E+02 1.E+03 1.E+04 1.E+05 1.E+06 1.E+07 total cell concentration (cells/ml) Figure 1. Typical total cell concentrations for groundwater, bottled mineral water, non-chlorinated tap water and lake water. All data typical for Swiss water sources. The cultivation-dependant method of heterotrophic plate counts (HPC) has been in use for more than 100 years (reviewed in e.g., Bartram et al. (2003) and Sartory (2004)). HPC still remain the primary parameter for assessment of the general microbiological quality of drinking water, even though it has become dogged by questions about what the method actually measures, how the method should be performed, and how the results should be interpreted (Allen et al., 2004; Sartory, 2004). Most critically, in what became commonly known as the great plate count anomaly, HPC detect only a small fraction of the total bacterial cells in aquatic environments (Staley and Konopka, 1985; Allen et al., 2004). Typical HPC guidelines for drinking water is in the range TECHNEAU - 3 - June, 6 2007

of 100 500 cfu/ml, while typical total cell concentrations in non-chlorinated drinking water is in the range of 100,000 cells/ml (Figure 1.1). Even though the immense discrepancy between total bacterial concentrations and cultivable bacterial cell concentrations (HPC) in aquatic samples has been shown numerous times (Staley and Konopka, 1985), the former is still not included in drinking water legislation, guidelines or everyday operational decision-making. The value of accurate total cell counting is evident, e.g. when treatment plant operators aim to monitor specific processes such as membrane filtration or ozonation. Flow cytometry (FCM) has tremendous potential as a tool for the analysis of bacteria in drinking water. This method can be used for direct enumeration of the total cell concentrations in water (Hammes and Egli, 2005; Lebaron et al., 1998), staining and detection of specific cellular features such as viability (Hoefel et al., 2003; Phe et al., 2005; Berney et al., 2006), or specific detection of targeted cells with antibodies (Vital et al., 2007). 1.2 Goals The aim of this work was to establish a standard flow cytometry protocol for the determination of total cell concentrations of natural microbial populations in drinking water. The specific target was to establish a method which is rapid (< 30 minutes per analysis), accurate and user friendly. 1.3 Acknowledgement This report contains the protocol for total cell counting with flow cytometry. An extended version of this work with details of the methodology has been accepted for publication in Water Research (Hammes et al., in press). TECHNEAU - 4 - June, 6 2007

2 Total cell concentration 2.1 General overview The flow cytometric method produces direct total cell counts with high speed (10 15 minutes), high accuracy (less than 5 % instrument error), and high sensitivity (lower detection limit of about 200 cells/ml). It does not permit differentiation of bacterial cells on the basis of taxonomy, metabolic activity, culturability or viability, and estimations of biomass or biovolume from light scattering data are complex and subjective and difficult to do. The method requires a flow cytometer operator that can distinguish stained microbial cells from background noise signals on the basis of relevant control samples and experience. The method assumes proper maintenance, standardization and calibration of the flow cytometer used. The method consists of staining an aqueous sample of suspended bacteria with a DNA-targeting fluorochrome and enumeration of the stained cells by counting with a flow cytometer. Fixatives for storage of samples are optional (see Standard Methods 9216 B, Greenberg et al. (1993)) but long-term storage before analysis is not recommended. 2.2 Apparatus a. Flow cytometer, equipped with a 488 nm blue laser with a minimum power of 15 mw 1. b. Filters, a 520 nm bandpass filter for green fluorescence detection, a 630 nm bandpass or longpass filter for red fluorescence detection. c. Counting modus, volumetric counting hardware are inclusive to some instruments 2 and preferable, though this can be compensated for by the use of commercially available bead count standards 3. d. Flow cytometry measuring tubes, comes standard for different brands of instruments. 2.3 Reagents a. Fluorochrome, SYBR Green I 4, diluted 1:100 from original stock solution in 0.2 µm filtered DMSO, and stored at -20 C until use. b. Dilution liquid, preferably from the same solution in which the bacteria are suspended, filtered with a 0.1 µm sterile syringe filter prior to use. c. EDTA stock solution, 500 mm EDTA suspended in nanopure water and sterile filtered (0.2 µm) prior to use. d. Lysis buffer stock solution, comprising 10 % v/v Triton X-100, 5 % v/v Tween 20, 10 mm Tris-HCl, and 1 mm EDTA. 1 Partec GmbH or equivalent 2 Partec GmbH or equivalent 3 CountBright (Invitrogen) or equivalent 4 Or equivalent DNA binding stains e.g. SYBR Green II or SYTO9 TECHNEAU - 5 - June, 6 2007

2.4 Procedure Collect water samples as directed in Section 9060 of Standard Methods (Greenberg et al. 1993). Fixation can be performed as described in Section 9216 B of Standard Methods (Greenberg et al. 1993), but direct processing of samples is highly recommended. Take 1 ml of the sample in a sterile Eppendorf tube and add 10 µl of the fluorochrome and 10 µl of EDTA stock solution, mix thoroughly by vortexing and incubate in the dark at room temperature for at least 10 minutes. If the bacteria concentrations are expected (or found) to exceed 1 x 10 7 cells/ml, it is advised to increase the concentration of fluorochrome accordingly (10 µl of respectively fluorochrome and EDTA for every 1 x 10 7 cells per ml). If the flow cytometer is not equipped with volumetric counting hardware, add just before measurement a commercially available bead standard of known concentration as prescribed by the manufacturer 5. Run the sample through a flow cytometer using the instrumental settings appropriate for the filter set-up and trigger on the green fluorescence channel. If the bacterial concentration in the sample exceeds 2 x 10 5 cells/ml, dilute the sample to a concentration between 0.1 2 x 10 5 cells/ml after staining and just before measurement with the dilution liquid. If the bacteria are present in clusters (can be controlled with fluorescence microscopy or evidenced through FCM light scatter data), dispersion of clustered cells can be achieved with the addition of 1 % v/v of lysis buffer prior to staining, coupled with vortexing or gentle sonication. 2.5 Calculations Optimal separation of stained bacteria from background noise is achieved on a two dimensional dotplot of green fluorescence (520 nm) and red fluorescence (630 nm). Background noise and stained bacteria are separated from each other by means of electronic gating (see section 3 of this report), using the flow cytometry software. Depending on the flow cytometer used, the software usually performs the counting and calculations automatically. Any sample dilution should be accounted for by the operator. 5 CountBright (Invitrogen) or equivalent TECHNEAU - 6 - June, 6 2007

3 Discussion Flow cytometry is able to clearly discern stained bacterial cells from background. Figure 3.1a shows a typical FCM dot-plot of green fluorescence (FL1) and red fluorescence (FL3) in which each individual dot represents the corresponding green and red fluorescence signals for each particle that was detected. The bacterial populations are separated from the instrument noise and background particles (e.g. inorganic crystals) by means of electronic gates which are constructed manually using the supplied software of the flow cytometer. The distinction between background noise signals and bacteria is based on experience with the instruments and relevant controls (e.g. a water sample before and after filtration with 0.1 µm). We have found that the best separation is achieved using the combination of green and red fluorescence (Figure 3.1a), although some authors use green fluorescence alone, or green fluorescence in combination with scattered light (particle size; Figure 3.1b) (Hammes and Egli, 2005; Hammes et al., 2007). The scatter signals (SSC) shown in Figure 3.1b provides additional information of the relative size of the particles (reference), and in the example displayed below, one can clearly recognise the tow distinct populations, different in both fluorescent intensity and size. These represent the so-called high nucleic acid (HNA) and low nucleic aid (LNA) bacteria (Lebaron et al., 2001). electronic gate bacteria instrument noise Fig. 3.1a. Flow cytometry dot-plot of bacteria from the effluent of a granular active carbon reactor. Each dot represents corresponding green fluorescence (FL1) and red fluorescence (FL3) of each particle which was analysed. Fig. 3.1b. Flow cytometry dot-plot of bacteria from the effluent of a granular active carbon reactor. Each dot represents corresponding green fluorescence (FL1) and scattered light signals (SSC) of each particle which was analysed. Note that the red fluorescence (FL3) in Figure 3.1a also originates from the SYBR Green I fluorochrome which is used for staining (Figure 3.2). Even though the emission of SYBR Green I is 10-fold higher at 520 nm (FL1) than at 615 nm (FL3). This difference, however, is compensated for by using higher voltage settings for the FL3 detectors (section 2, this report). TECHNEAU - 7 - June, 6 2007

FL1 detector range FL3 detector range Figure 3.2. The excitation and emission spectrum of SYBR Green I used for the staining of bacteria. The optimum emission is at 520 nm, similar to the filters used in the flow cytometer for detection. A tail of the stain is also visible above 600 nm (range of the red FCM filters). The method is rapid it requires 10 minutes for staining and 3 minutes per analysis. The result is that a large volume of samples can easily be processed and the quick results allow treatment plant operators to take immediate action in case the results suggest a problem in the system. The method is also very accurate. The standard deviation on triplicate FCM measurement is in the range of 5 %, of which half can be ascribed to user error and half to instrument error. The method is sensitive (Figure 3.3). It is possible to measure accurately in a concentration range of 200 cells/ml, although the optimal range of measurement lies between 10 000 100 000 cells/ml. 10 Total cell concentration (cells ml -1 x 10 4 ) 8 6 4 2 6000 5000 4000 3000 2000 1000 0 0 1 2 3 4 5 0 20 40 60 80 100 120 % of bottled mineral water Figure 3.3. Sensitivity and reproducibility of FCM measurements of bacteria from bottled mineral water diluted with cell-free water (n = 9). All data points are average values of triplicate samples (data from Hammes et al. (2007)). TECHNEAU - 8 - June, 6 2007

The method is statistically sound. In the analysis of one sample, 200 ul of solution is actually measured. This means that in a sample with a cell concentration of 100 000 cells/ml, and effective amount of 20 000 events would be recorded (Hammes and Egli, 2005). TECHNEAU - 9 - June, 6 2007

4 Conclusion The total cell count method with flow cytometry has been standardised and tested to a level where it is deemed usable for the analysis of total bacteria cell concentrations in drinking water. The method can be used directly for samples with bacterial concentrations in the range of 200 200,000 cells/ml, which is also the typical range for drinking water bacteria. In case of higher cell concentrations, I dilution step would be required. The FCM method represents a fast and accurate way to measure total microbial cell concentrations in drinking water. The values obtained this way will be particularly useful for the accurate description of the microbiology in drinking water treatment trains (Hammes et al., 2007), and for the correct assessment of the microbial stability of drinking water. Given enough experience and data gathered with this manner, it is foreseen that it might bring about a paradigm shift in the legislative guidelines for assessing the general microbial quality of drinking water. The method will be tested on pilot scale and full scale treatment and distribution systems for final validation and interpretation of the results. TECHNEAU - 10 - June, 6 2007

5 References Allen, M.J., Edberg, S.C., and Reasoner, D.J. (2004) Heterotrophic plate count bacteria what is their significance in drinking water? International Journal of Food Microbiology, 92, 265 274. Bartram, J., Cotruvo, J., Exner, M., Fricker, C. and Glasmacher, A. (2003) Heterotrophic plate counts and drinking-water safety. London, UK: IWA Publishing on behalf of the World Health Organization. Berney, M., H. U. Weilenmann, and T. Egli. 2006. Flow-cytometric study of vital cellular functions in Escherichia coli during solar disinfection (SODIS). Microbiology 152:1719-29. Berry, D., Xi, C. and Raskin, L. (2006) Microbial ecology of drinking water distribution systems. Current Opinion in Biotechnology 17, 297 302. Hammes F.A. and Egli, T. (2005) New method for assimilable organic carbon determination using flow-cytometric enumeration and a natural microbial consortium as inoculum. Environmental Science and Technology, 39, 3289 3294. Hammes, F., M. Berney, W. Yingying, M. Vital, O. Köster, and T. Egli. (in press). Total bacterial cell counts as a descriptive microbiological parameter for drinking water treatment process. Accepted for publication in Water Research. Hoefel, D., W. L. Grooby, P. T. Monis, S. Andrews, and C. P. Saint. 2003. Enumeration of water-borne bacteria using viability assays and flow cytometry: a comparison to culture-based techniques. Journal of Microbiological Methods 55, 585-97. Lebaron, P., Parthuisot, N. and Catala, P. (1998). Comparison of blue nucleic acid dyes for flow cytometric enumeration of bacteria in aquatic systems. Applied and Environmental Microbiology 64, 1725-1730. Lebaron, P., P. Servais, H. Agogue, C. Courties, and F. Joux. 2001. Does the high nucleic acid content of individual bacterial cells allow us to discriminate between active cells and inactive cells in aquatic systems? Appl. Environ. Microbiol. 67:1775-82. Leclerc, H. and Moreau, M. (2002) Microbiological safety of natural mineral water. FEMS Microbiology Reviews 26(2), 207 222. Reasoner, D.J. and Geldreich, E.E. (1985) A new medium for the enumeration and subculture of bacteria from potable water. Applied and Environmental Microbiology 49, 1 7. TECHNEAU - 11 - June, 6 2007

Sartory, D.P. (2004). Heterotrophic plate count monitoring of treated drinking water in the UK: a useful operational tool. International Journal of Food Microbiology 92, 297 306. Staley, J. T. & Konopka, A. (1985). Measurement of in situ activities of nonphotosynthetic microorganisms in aquatic and terrestrial habitats. Annual Reviews in Microbiology 39, 321-346. Vital M., Füchslin H., Hammes F. A. and Egli T. (2007) Growth of Vibrio cholerae O1 Ogawa Eltor in freshwater. Microbiology SGM, 153, 1993 2001. TECHNEAU - 12 - June, 6 2007