Improving Sensitivity in Bioanalysis using Trap-and-Elute MicroLC-MS

Similar documents
Pharmacokinetic Analysis of the mab Adalimumab by ELISA and Hybrid LBA/LC/MS: A Comparison Study

On-line SPE-LC/MS/MS to Detect Organonitrogen and Triazine Pesticides at 10ng/L in Drinking Water

Exploring Extra Sensitivity Using ionkey/ms with the Xevo G2-XS Q-Tof HRMS for Small Molecule Pharmaceutical Analysis in Human Plasma

Rapid Peptide Catabolite ID using the SCIEX Routine Biotransform Solution

Improved Peptide Maps Using Core-Shell Media: Explaining Better Biopharmaceutical Applications With Kinetex C18 Columns

Rapid Soft Spot Analysis using the SCIEX Routine Biotransform Solution

Investigating Biological Variation of Liver Enzymes in Human Hepatocytes

Quantitation of Microcystins and Nodularins in Water Samples Using LC-MS/MS

MultiQuant Software Version 3.0

Peptide Mapping: A Quality by Design (QbD) Approach

A Sub-picogram Quantification Method for Desmopressin in Plasma using the SCIEX Triple Quad 6500 System

A Unique LC-MS Assay for Host Cell Proteins(HCPs) ) in Biologics

Enabling Systems Biology Driven Proteome Wide Quantitation of Mycobacterium Tuberculosis

Disulfide Linkage Analysis of IgG1 using an Agilent 1260 Infinity Bio inert LC System with an Agilent ZORBAX RRHD Diphenyl sub 2 µm Column

mab Titer Analysis with the Agilent Bio-Monolith Protein A Column

IgG Purity/Heterogeneity and SDS-MW Assays with High- Speed Separation Method and High Throughput Tray Setup

Cell Clone Selection Using the Agilent Bio-Monolith Protein A Column and LC/MS

A Complete Workflow Solution for Intact Monoclonal Antibody Characterization Using a New High-Performance Benchtop Quadrupole- Orbitrap LC-MS/MS

EPA Method 543: Selected Organic Contaminants by Online SPE LC/MS/MS Using the Agilent Flexible Cube

A Generic Kit-Based Approach for Quantifying Monoclonal Antibody Drugs Through Direct Digestion of Discovery Study Samples

MassPREP On-Line Desalting Cartridge

Liquid Chromatograph Liquid Chromatograph Mass Spectrometer. CL Series. for In Vitro Diagnostic Use

Tyler Trent, Applications Sales Specialist; Teledyne Tekmar Page 1

Application Note. Abstract. Authors. Introduction. Environmental

MasterView Software. Turn data into answers reliable data processing made easy ONE TOUCH PRODUCTIVITY

The LC-MS/MS Workhorse. SCIEX 4500 Series Mass Spectrometers

Application Note. Author. Abstract. Biotherapeutics and Biologics. Sonja Schneider Agilent Technologies, Inc. Waldbronn, Germany

Characterize Fab and Fc Fragments by Cation-Exchange Chromatography

PLRP-S Polymeric Reversed-Phase Column for LC/MS Separation of mabs and ADC

Maximizing Chromatographic Resolution of Peptide Maps using UPLC with Tandem Columns

Application Note. Author. Abstract. Biopharmaceuticals. Verified for Agilent 1260 Infinity II LC Bio-inert System. Sonja Schneider

Application Note. Author. Abstract. Environmental. Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany

Fraction Analysis of Cysteine Linked Antibody-Drug Conjugates Using Hydrophobic Interaction. chromatography. Agilent 1260 Infinity II Bio-Inert System

On-Line. User s Guide SPE CARTRIDGES. for Rapid Cleanup and Extraction of Analytes

MS Certified Vials. Pre-cleaned and certified vials for mass spectrometry

EASY-Spray Columns. Guidance for column set up and installation Tips to maximize column lifetime

Antibody Analysis by ESI-TOF LC/MS

Size Exclusion Chromatography of Biosimilar and Innovator Insulin Using the Agilent AdvanceBio SEC column

Fast and High-Resolution Reversed-Phase Separation of Synthetic Oligonucleotides

A Fast and Robust Linear ph Gradient Separation Platform for Monoclonal Antibody (MAb) Charge Variant Analysis

Simultaneous in vivo Quantification and Metabolite Identification of Plasma Samples Using High Resolution QTof and Routine MS E Data Analysis

Rapid Peptide Mapping via Automated Integration of On-line Digestion, Separation and Mass Spectrometry for the Analysis of Therapeutic Proteins

Generating Automated and Efficient LC/MS Peptide Mapping Results with the Biopharmaceutical Platform Solution with UNIFI

Precise Characterization of Intact Monoclonal Antibodies by the Agilent 6545XT AdvanceBio LC/Q-TOF

Genotoxicity is the property of a compound

Analysis of Intact and C-terminal Digested IgG1 on an Agilent Bio MAb 5 µm Column

EtG / EtS Analysis UCT INNOVATION THROUGH CHEMISTRY FORENSICS

Improved SPE for UPLC/MS Determination of Diquat and Paraquat in Environmental

Two-Dimensional LC Protein Separation on Monolithic Columns in a Fully Automated Workflow

Protocol for Quantitative Determination of Residual Solvents in Cannabis Concentrates Prepared by: Amanda Rigdon, May 23 rd, 2016

Development of a Fast and Cost-Effective Multi-Residue Method to Determine Pesticides in Tobacco by LC/MS/MS

Proteomics. Proteomics is the study of all proteins within organism. Challenges

Featuring Analyst software under Windows NT for enhanced performance and ease of use. API 150EX. LC/MS System

Method Development Considerations for Reversed-Phase Protein Separations

Application Note. Kwasi Antwi, Amanda Ribar, Urban A. Kiernan, and Eric E. Niederkofler Thermo Fisher Scientific, Tempe, Arizona

Converting a Liquid-Liquid Extraction Method for Vitamin D to a 96-Well Plate Supported Liquid Extraction Format

[ product solution ] Waters Oasis µ Elution PlateS. Patented Innovation. Elution volume as low as 25 μl. No evaporation and reconstitution

Accucore. Ultimate Core Performance LC Column Technology to Maximize Your Investment. Dave Jarzinski. October 2011

Quantitation of cyanotoxins in drinking water according to EPA 544 guidelines

APPLICATIONS TN Overview of Kinetex 2.6 µm Core-Shell Technology

Proudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment QTRAP 5500

Fast mass transfer Fast separations High throughput and improved productivity Long column lifetime Outstanding reproducibility Low carryover

R R Innovation Way P/N SECKIT-7830 Newark, DE 19711, USA Tel: Fax: Website: Published in November 2013

Automated Online SPE for LC/MS/MS Analysis of Trace Organic Contaminants in Water Using the Agilent 1290 Infinity Flexible Cube Module

Confident Pesticides Analysis with the Agilent LC/Triple Quadrupole and TOF/QTOF Solutions

Developing Robust and Efficient IEX Methods for Charge Variant Analysis of Biotherapeutics Using ACQUITY UPLC H-Class System and Auto Blend Plus

mabs and ADCs analysis by RP

Trastuzumab Glycan Batch-to-Batch Profiling using a UPLC/FLR/Mass Spectrometry Platform

Quality-by-Design-Based Method Development Using an Agilent 1290 Infinity II LC

N-Glycan Profiling Analysis of a Monoclonal Antibody Using UHPLC/FLD/Q-TOF

Physical Stability of a Silica- Based Size Exclusion Column for Antibody Analysis

Improving Resolution and Column Loading Systematically in Preparative Liquid Chromatography for Isolating a Minor Component from Peppermint Extract

Agilent 6430 Triple Quadrupole LC/MS System

Concentration of Human Hormones in Drinking Water Using Solid Phase Extraction and Analysis by High Performance Liquid Chromatography

Preparative HPLC is still the

Protein-Pak Hi Res HIC Column and HIC Protein Standard

Chem 321 Lecture 23 - Liquid Chromatography 11/19/13

Utilizing novel technology for the analysis of therapeutic antibodies and host cell protein contamination

ENHANCED MS WITH THE TURN OF A KEY

Analysis of biomolecules by SEC and Ion-Exchange UPLC

High Resolution Accurate Mass Peptide Quantitation on Thermo Scientific Q Exactive Mass Spectrometers. The world leader in serving science

AB SCIEX TripleTOF 5600 SYSTEM. High-resolution quant and qual

Development of Analysis Methods for Therapeutic Monoclonal Antibodies Using Innovative Superficially Porous Particle Biocolumns

C196-E067A. Prominence UFLCXR. Shimadzu Ultra Fast Liquid Chromatograph

Application Note. Authors. Abstract

Bivalirudin Purification:

Bio-Monolith Protein G Column - More Options for mab Titer Determination

CaptiveSpray nanobooster

Analysis of Monoclonal Antibody Digests with the Agilent 1290 Infinity 2D-LC Solution Part 2: HILIC RPLC-MS

Guide to Trap Cartridge Care and Use

Proteins. Patrick Boyce Biopharmaceutical Marketing Manager Waters Corporation 1

Peptide Mapping of Glycoprotein Erythropoietin by HILIC LC/MS and RP-LC/MS

Agilent AdvanceBio SEC Columns for Aggregate Analysis: Instrument Compatibility

Thermo Scientific Peptide Mapping Workflows. Upgrade Your Maps. Fast, confident and more reliable peptide mapping.

Investigation of Solid Phase Microextraction as an Alternative to Dried Blood Spot

AppNote 4/2014. Prep-and-Shoot : The Completely Automated Workfl ow for the Hydrolysis and Analysis of Urine Samples by LC/MS/MS.

LC/MS/MS Solutions for Biomarker Discovery QSTAR. Elite Hybrid LC/MS/MS System. More performance, more reliability, more answers

Optimizing the Purification of a Standard Chiral Compound Utilizing a Benchtop, Multi-Purpose, Semi-Preparative to Preparative HPLC System

A Modular Preparative HPLC System for the Isolation of Puerarin from Kudzu Root Extracts

Transcription:

Improving Sensitivity in Bioanalysis using Trap-and-Elute MicroLC-MS Using the SCIEX M3 MicroLC system for Increased Sensitivity in Antibody Quantitation Remco van Soest and Lei Xiong SCIEX, Redwood City, CA, USA Quantitation of monoclonal antibodies (mabs) in biological fluids is important during all stages of antibody drug development. While traditionally immunoassays are used, more recently LC- MS has been adopted because of its high selectivity, accuracy, and precision. The antibodies can be enriched from the sample using different approaches, e.g. solid phase extraction or immunocapture, and are then digested using trypsin. Unique signature peptides are selected based on criteria such as digestion efficiency, stability after digestion, chromatographic behavior and MS-MS sensitivity, and measured using LC-MS in MRM mode. As the amount of sample that can be drawn from a small animal during DMPK studies is limited, sensitivity of an LC- MS based method becomes very important. MicroLC, using 0.3 mm ID columns at 5-20 µl/min flow rate, can improve sensitivity in LC-MS by a factor of up to 10 compared to using the more commonly used 2.1-3 mm ID columns at flow rates of 0.5-1 ml/min. 1,2 In this technical note we explore the use of MicroLC- MS using 0.3 mm ID columns at 8 µl/min flow rate to achieve lower LLOQ s for the quantitation of infliximab, a mab used for the treatment of Crohn s disease. A Trap-Elute MicroLC method was used to maintain the same throughput as with High Flow LC-MS, while injecting the same sample volume. Sample is loaded at 35 µl/min onto a short trap column, followed by switching the trap column in-line with the separation column for analysis at 8 µl/min. Key Benefits of using the M3 MicroLC system for Antibody Quantitation Quantitate antibodies at levels up to 10 x lower than what can be measured with High Flow LC-MS High throughput by using a Trap-Elute workflow Increased column lifetime and reduced need for cleaning of the MS by protecting the analytical column and MS from salts and other impurities with the Trap-Elute workflow Materials and Methods Sample preparation: was acquired from Myoderm (Norristown, PA, USA). As internal standard SILuMab, a recombinant stable isotope labeled human mab, was used (Sigma-Aldrich, St. Louis, MO, USA). Stock tryptic digests of both mab s were prepared using N-octylglucoside (OGS) as denaturant, TCEP for reduction and MMTS for alkylation, all included in the SCIEX Protein Preparation kit, and TPCK treated trypsin (SCIEX) for digestion. Standards were diluted from stock with 98/2 water/acetonitrile 0.1% formic acid. HPLC conditions High Flow LC: A Shimadzu Prominence HPLC system was used, consisting of two LC-20AD pumps, a SIL-20AC autosampler and CTO-20A column oven. The column used was a 50 x 3 mm Kinetex C18 2.6 µm 100 Å column from Phenomenex (Torrance, CA, USA). Mobile phase A was water with 0.1% formic acid, and mobile phase B was acetonitrile with p 1

0.1% formic acid. Wash solvent for the autosampler was 20/20/60 methanol/acetonitrile/ipa. The gradient method used is listed in Table 1. Flow rate was 0.7 ml/min. Injection volume was 20 µl, and the column was kept at 40 C. Table1. Gradient used with the High Flow LC-MS workflow Time (min) %B 0 5 0.8 10 3.5 25 5 40 5.1 95 5.9 95 6 5 7 5 HPLC conditions MicroLC: A SCIEX M3 MicroLC-TE system, consisting of two MicroLC gradients and an integrated autosampler, was used in combination with a source mounted column oven (SCIEX). As the trap, a 10 x 0.3 mm 5 µm 120 Å ChromXP C18 CL column (SCIEX) was used, and the analytical column was a 50 x 0.3 mm HALO Peptide ES-C18 2.7 µm 160 Å column (SCIEX). Trapping conditions Mobile phase A in the loading gradient was water with 0.1% formic acid, Mobile phase B was acetonitrile with 0.1% formic acid. Sample was loaded from the injection loop onto the trap column using 100% A for one minute at 35 µl/min. The trap was washed with 90% B at 70 µl/min for 2 minutes after every injection. Separation conditions Mobile phase A in the analytical gradient was water with 0.1% formic acid, mobile phase B was acetonitrile with 0.1% formic acid. The gradient used was 3-40% B in 3 min, with a 1.5 min 90% B wash step. Flow rate was 8 µl/min. The column temperature was 40 C. Injection volume was 20 µl, and the autosampler needle and valve wash consisted of two cycles using mobile phase B, followed by one cycle using mobile phase A. Table 2. Signature peptides used for and SILuMab quantitation Antibody Signature Peptide SILuMab DTLMIS[R] SILuMab VVSVLTVLHQDWLNG[K] ASQFVGSSIHWYQQR GLEWVAEIR YASESMSGIPSR p 2

Separation and Analysis time Analysis time and separation resolution are similar between the direct inject High Flow LC-MS method and the trap-elute microlc-ms method, when taking into account the 1 minute trap loading time for the microlc method (see Figure 2). Figure 2A, XIC chromatogram for the High Flow LC-MS method. Figure 1. Valve positions for the trap loading part of the method (top) and the analysis part of the method (bottom). Mass Spectrometry A SCIEX QTRAP 6500 LC-MS/MS system was used. For the MicroLC experiments the standard electrode was replaced with a 25 µm ID electrode (SCIEX). MRM transitions were developed for the peptides listed in table 2, and the source and gas parameters are listed in table 3. MultiQuant 3.0.2 software was used for data analysis. Table3. Source and gas parameters High Flow LC MicroLC Electrode ID 100 µm 25 µm Curtain Gas 30 20 Collision GAS High High IonSpray Voltage 5500 5000 Temperature ( C) 650 300 Ion Source Gas 1 60 40 Ion Source Gas 2 60 10 Figure 2B, XIC chromatogram for the Trap-elute MicroLC-MS method. Sensitivity Improvement In order to determine the improvement in sensitivity that can be achieved using MicroLC Trap-Elute versus High Flow LC, two standard quantitation curves were obtained with each concentration injected five times. The peptide used to quantitate infliximab was YASESMSGIPSR, while DTLMIS[R] was used as an internal standard. For the High Flow experiment 5 5000 p 3

ng/ml infliximab was measured, while 1-1000 ng/ml was measured with MicroLC. The internal standard was maintained at 100 ng/ml. An improvement in S/N of 5x was seen for the infliximab signature peptide with the MicroLC-MS method at the 5 ng/ml level. S/N at the 1 ng/ml level using MicroLC was approximately the same as S/N with the High Flow method at 5 ng/ml (see figure 3). ng/ml to 2 ng/ml, a 5-fold improvement. In order not to exceed 20% response in the first blank after an injection at the ULOQ level, the ULOQ for the MicroLC Trap-Elute method was reduced from 1000 ng/ml to 200 ng/ml. Extra washing steps may be required to reduce carry-over if a larger linear range is required. Table 4. Standard curve date for High Flow LC-MS Actual Concentration (ng/ml) Mean calculated concentration (ng/ml) Accuracy (%) CV (%) 5.0 6.1 121.32 13.49 10.0 9.8 98.43 6.23 20.0 16.3 81.62 11.63 50.0 45.5 90.97 6.72 100.0 94.7 94.73 5.31 500.0 547.3 109.46 3.23 1000.0 1052.2 105.22 4.77 5000.0 4913.1 98.26 9.30 Table 5. Standard curve date for MicroLC-MS Actual Concentration (ng/ml) Mean calculated concentration (ng/ml) Accuracy (%) CV (%) 2.0 1.6 80.25 18.70 3.0 3.1 101.96 9.54 5.0 6.2 123.14 11.07 10.0 11.3 113.14 11.01 20.0 22.9 114.73 5.81 Figure 3. Sensitivity comparison between the High Flow LC-MS and MicroLC trap-elute methods. Accuracy and precision data from both curves are summarized in Tables 4 and 5. Linear regression and a weighting of 1/x was used. R 2 for both the High Flow and MicroLC curves were 0.993. The LLOQ for both the High Flow and MicroLC-MS methods was determined using the criteria that accuracy at the LLOQ should be between 80 and 120 %, and precision should be < 20% CV. Using the MicroLC-MS method the LLOQ was improved from 10 50.0 58.8 117.57 6.92 100.0 110.3 110.34 7.04 500.0 515.3 103.07 7.35 1000.0 961.1 96.11 2.65 p 4

Robustness In order to determine the robustness of the MicroLC trap-elute method, a high concentration of infliximab (10 µg/ml) and SILuMab (1 µg/ml) was digested. A total of 1000 20 µl injections were made over a period of 5 consecutive days. No clogging of tubing, electrode or columns was observed. All 1000 injections were completed using the same trap and analytical column. CV % on peak area for the signature peptide used for quantitation was 4.35%, while the CV % for the SILuMab standard peptide was 6.13%. Intensity 7e7 6e7 5e7 4e7 3e7 2e7 1e7 0e0 Injection # 1000 Injection # 1 1.5 2.0 2.5 3.0 3.5 Time, min Figure 5. Total Ion Chromatograms for injection 1 and 1000 of the robustness test. Separation and peak shapes are identical. Conclusions We have shown that quantitation of infliximab using its signature peptide YASESMSGIPSR can be performed with an approximately 5x higher sensitivity using a trap-elute MicroLC- MS method at 8 µl/min, compared to using a direct inject highflow LC-MS method at 700 µl/min. The trap-elute method used ensures similar throughput while injecting the same 20 µl of sample, and protecting the MicroLC column and MS from contamination. The trap-elute MicroLC-MS workflow proved to be robust over a 1000 injections. Figure 4. Area reproducibility for the peptide YASESMSGIPSR over 1,000 injections was 4.35 %. This workflow offers a solution for applications where Mab s need to be quantitated in small volume samples and/or low concentrations. References 1. Covey,T.R. Thomson B.A., Schneider, B.B. Atmospheric Pressure Ion Sources; Mass Spec. Reviews, 2009, 00, 1-29. 2. Covey T.R., Schneider B.B., Kovarik, Corr, Javahari et al.;the Central Analytical Figures of Merit of ESI, MALDI, and APCI.; Cole RB, ed. Electrospray and MALDI mass spectrometry. 2010, Chapter 13. Hoboken: John Wiley & Sons, Inc. AB Sciex is doing business as SCIEX. 2015 AB Sciex. For Research Use Only. Not for use in diagnostic procedures. The trademarks mentioned herein are the property of AB Sciex Pte. Ltd. or their respective owners. AB SCIEX is being used under license. Document number: RUO-MKT-02-2849-B p 5