Laser measurement of KCl vapor in 4 MW CFB boiler during straw combustion and ferric sulfate injection

Similar documents
FIRING ESTONIAN OIL SHALE OF HIGHER QUALITY IN CFB BOILERS ENVIRONMENTAL AND ECONOMIC IMPACT

Modeling the use of sulfate additives for potassium chloride destruction in biomass combustion

Fpollutant-free zero air for analytical

Retrofit of Rodenhuize 4 power station: The Max Green and Cold Back-up-projects

Emissions from wood-fuelled equipment. Senior research scientist Heikki Oravainen Technical Research Centre of Finland

CEMS for Measurement of Ammonia, SO 3, and Low Level NOx

Carbon / Sulfur Analyzers

EXPERIMENT 5. Molecular Absorption Spectroscopy: Determination of Iron with 1,10-Phenanthroline

High efficient multi-fuel CYMIC concept for biomass, rejects and coal for Hamburger Hungaria Katriina Jalkanen Valmet Technologies Oy

Application of CFB (Circulating Fluidized Bed) to Sewage Sludge Incinerator

Advanced Processes Analysis and Control Methods for CFB Power Plants Project Overview

Discipline Chemical Testing Issue Date Certificate Number T-3553 Valid Until Last Amended on - Page 1 of 9

SPECIATION OF ALKALI METALS IN BIOMASS COMBUSTION AND GASIFICATION PAVANKUMAR BAJRANG SONWANE HENG BAN, COMMITTEE CHAIR THOMAS K. GALE PETER M.

Combustion Master CS Carbon and Sulphur Analyser

Which of these is the formula for disulfur heptoxide? A. S 2 O 7 B. S 7 O 2 C. SO 2 D. N 2 O

Model for the Forecasting of the Fusion Temperature of Biomass Ashes

Mikko Hupa Åbo Akademi Turku, Finland

Trace Moisture Contamination in Ultra-High Purity Phosphine: Techniques for Measurement and Control

Foster Wheeler Advanced Bio CFB Technology for Large Scale Biomass & Peat Firing Power Plants

REPORT NUMBER REPORT DATE SEND TO ISSUE DATE Apr 18, Apr 18, 2017 RECEIVED DATE Apr 05, 2017

*20GSD5201* Double Award Science: Chemistry. Unit C2 Higher Tier TUESDAY 9 JUNE 2015, AFTERNOON [GSD52] *GSD52* *G5802* TIME 1 hour 15 minutes.

Improving Steel Reheat Furnace Performance with the ZoloSCAN-RHT Combustion Monitoring System

8 th SGC International Seminar on Gasificaton October 15 th -16 th, 2014 Malmö, Sweden

IMPACT OF OPERATING CONDITIONS ON SO 2 CAPTURE IN A SUPERCRITICAL CFB BOILER IN POLAND

Measuring trace impurities in HyCO, regeneration hydrogen and inert gases Los Gatos Research (LGR)

Diode laser analyser LS4000 Ammonia slip measurement for DeNOx process

Samcheok Green Power 4 x 550 MW e Supercritical Circulating Fluidized-Bed Steam Generators in South Korea

Clean Energy. from. Renewables Biomass & Waste

KKB Micro Testing Labs Pvt. Ltd., , 2 nd Floor, Tarun Plaza, NFC Main Road, Krishna Nagar Colony, Moula Ali, Hyderabad, Telangana

Kazushige KUROSAWA*, Zhibao ZHANG**, and Zhengbing WANG** [Delivered Products & Systems] 1. Introduction. 2. Overview of Nanjing

Green Fuel Nordic The Smart Way. Utilising RTP TM technology to produce sustainable 2 nd generation bio-oil from local feedstocks

Co-combustion in fluidised bed influence of K, Cl and S

Excerpt of Thermal Power Guidelines for New Plants

Smart CHP from Biomass and Waste

Hydrogen Sulphide and Mercaptan Removal

Welcome to. Kendal Power Station

Biomass Power Plants

The world s largest biofuel CHP plant Alholmens Kraft, Pietarsaari

(a) To find out which is the more reactive metal, zinc or tin, the following experiment could be carried out. piece of zinc shiny surface

// XPLORER-NS. Full range Total Nitrogen, Total Sulfur instrument for the modern Laboratory.

Greenhouse Gas Measurement Technology

Fouling of the Cooling Surfaces in Biofuel-Fired Fluidized Bed Boilers

AUTOMATED WATER ANALYSIS

Hard water. Hard and Soft Water. Hard water. Hard water 4/2/2012

Fact Sheet. Feb 2011

SMS 100 MERCURY ANALYZER SMS 100. Mercury analysis directly on solid samples from the undisputed leader in Atomic Spectroscopy

Combustion Control and Safety A comparison between Zirconium Oxide/Catalytic and Tunable Diode (TDL) Technologies Dr Stephen Firth / Rhys Jenkins

WHITE PLUME REMOVAL USING FUEL ADDITIVES IN THERMAL POWER PLANTS

FlueGas_071029_GB. Flue Gas Cleaning


Center for Advanced Sustainable Iron & Steel Making

S.E. (Chemical) (First Semester) EXAMINATION, 2012 PROCESS CALCULATIONS (2008 PATTERN) Time : Three Hours Maximum Marks : 100

Lectio Praecursoria. Lappeenranta University of Technology. From the SelectedWorks of Kari Myöhänen

Can your unit pass a Particulate Emission Compliance Test?

The firing and co-firing of difficult biomass fuels W R Livingston Doosan Babcock R&D

Name: Class: Date: 6. Most air pollution is produced by a. thermal inversions. c. ozone layer depletion. b. fuel burning. d. volcanic eruptions.

Gasification of Municipal Solid Waste

THE CHALMERS GASIFIER

Fluidized Bed Combustion of Biomass and Waste-derived Fuels Current Status and Challenges

Investigators: R. E. Mitchell, Associate Professor, Mechanical Engineering Department; P. A. Campbell and L. Ma, Graduate Researchers

S2 RANGER LE: Analysis of Light Elements in Cement, Slags and Feldspar

Pyrolysis of cotton stalks and utilization of pyrolysis char for sustainable soil enhancement and carbon storage

TruePeak TDLS200. NH 3 Slip Measurement. <Document Number> Copyright Yokogawa Electric Corporation <date/time>

BITUMINOUS COAL ASH SULFUR RETENTION IN CIR5^ DEPARTMENT OF REGISTRATION AND EDUCATION ILLINOIS STATE GEOLOGICAL SURVEY CIRCULAR

Circulating Fluidized Bed Technology for Large Scale Power Generation Using Coal and Petroleum Coke

Biomass and Biofuels

The ANDRITZ GROUP Overview

INTEGRATED HEAT, ELECTRICITY AND BIO-OIL PRODUCTION. IEA Biomass Task 34 Meeting in Chicago Jani Lehto, Metso Pekka Jokela, UPM

REACHED DEVELOPMENTS OF BIOMASS COMBUSTION TECHNOLOGIES AND FUTURE OUTLOOK

Quartz Glass for Ultra High Pressure and High Intensity Discharge Lamps. Heraeus Quarzglas

Technical Note TN-114 rev 16 wy/wh.11-04

Clean energy, natural solutions. Understanding rice husk as a biomass fuel

Methods for determination of mercury in LP gas

Towards New Milestones In CFB Boiler Technology CFB 800MWe

Behaviour of gaseous chlorine and alkali metals during biomass thermal utilisation

DETERMINATION OF ELEMENTS IN DRINKING WATER AS PER BUREAU OF INDIAN STANDARDS 10500, & USING THE AGILENT 5100 ICP-OES

Poultry Litter as a Renewable Resource. Fibrominn Biomass Power Plant

Environment Protection Engineering NEGATIVE EFFECTS OF BIOMASS COMBUSTION AND CO-COMBUSTION IN BOILERS

Halogens, dioxins/furans

GCE Environmental Technology. Energy from Biomass. For first teaching from September 2013 For first award in Summer 2014

Technical Notes for Biomass and Air Quality Guidance

Two Great Technologies Together At Last. The 700 Series NanoTrace Moisture And Oxygen Analyzers

Surface Flammability, Smoke and Toxic Gas Generation of "3M Polyurethane Adhesive Sealant 560"

Hamm MW Pyrolysis Plant. Integrated Pyrolysis into Power Plant Plant capacity 100,000 t/a Pre-processed Waste Materials

New Developments in Measuring Low Silicon in Process Piping Using Handheld X-ray Fluorescence (XRF)

Standard Optics Information

Traditional and Advanced Air Quality Control Solutions

The Importance of Water Absorbance/Transmittance on the Efficiency of Ultraviolet Disinfection Reactors

Circulating Fluidized Bed Technology Towards 800 MWe Scale Lagisza 460 MWe Supercritical CFB Operation Experience

Effect of Flue Gas Recirculation on the Formation of Fine Particulate Matter in a Domestic Pellet-Fired Boiler

Catalytic gasification of biomass for hydrogen production with in-situ CO 2 absorption using novel bi-functional Ni-Mg-Al-CaO catalyst

2015 O LEVEL CHEMISTRY 5073/02

FRANCE S BIGGEST BIOMASS CHP PLANT

TWEED RIVER HIGH SCHOOL 2006 PRELIMINARY CHEMISTRY. Unit 2 Metals

Experimental Study on Combustion of Biomass in a Boiler with Gasification

Using the Pulsed Flame Photometric Detector (PFPD) for Low-level Analysis of Organophosphorus Pesticides

Gasification of Biomass and SulfurContaining Carbonaceous Fuels. Majid Charmchi

Continous Emission Monitoring: Methods, Regulations and QA/QC. Anders Arvidsson, Opsis AB, Sweden

Surface Flammability, Smoke and Toxic Gas Generation of "3M Polyurethane Adhesive Sealant 550 Fast Cure"

Surface Flammability, Smoke and Toxic Gas Generation of "3M Scotch-Weld Acrylic Adhesive DP8410NS Green"

Transcription:

Laser measurement of KCl vapor in 4 MW CFB boiler during straw combustion and ferric sulfate injection Juha Toivonen, Tapio Sorvajärvi Department of Physics, Tampere University of Technology, P.O. Box 692, FI-33101 Tampere, Finland Tel: +358 40 8490 490, E-mail: juha.toivonen@tut.fi, http://www.tut.fi/optics/ Joni Maunula, Jaani Silvennoinen Valmet Technologies, P.O. Box 109, FI-33101 Tampere, Finland http://www.valmet.com/ KEY WORDS potassium chloride, laser measurement, straw combustion, ferric sulfate injection ABSTRACT Combustion of the renewable solid fuels and the mixed fuels causes serious corrosion and slagging problems in combustion boilers. One cause of the problems is alkali chloride vapors, e.g. KCl, that is formed during combustion due to the high content of alkali and chlorine in the biomass. In order to prevent these problems with proper fuel content and/or additive injection, a sensitive and reliable monitoring technique for the alkali chlorides is required. A laser measurement technique was applied to monitor the KCl concentration through the flue gas channel of a 4 MW circulating fluidized bed boiler during combustion of straw and wood. The straw composition in the fuel was changed from 0% to 100 %, and ferric sulfate injection was used to control the KCl concentration in the furnace. Chlorine is known to occur mainly in forms of KCl and HCl in the flue gas due to the high content of potassium in straw. The vaporous KCl can be converted to less harmful potassium sulfate and HCl with ferric sulfate injection. Single measurement curves are discussed and the operation of the laser measurement in two longer-term experiments is demonstrated. The laser measurement resulted stable concentration values in steady combustion conditions and followed transients due to the furnace operational changes with time resolution less than a minute. The online laser measurement enables a stable and controlled KCl concentration due to actively adjusted ferric sulfate injection rate. 1 INTRODUCTION Environmental aspects and increasing price of the fossil fuels have increased the interest of the renewable fuels in power production. One good alternative as a largely CO 2 neutral fuel, is biomass that can be harvested, for example, as agricultural residues. However, the combustion of the renewable solid fuels and the mixed fuels causes serious corrosion and slagging problems in combustion boilers /1/. One cause of the problems is the alkali chloride vapors that are formed due to the high content of alkali and chlorine in the biomass. In order to prevent these problems with proper fuel content or a control substance a sensitive and reliable monitoring technique for the alkali chlorides is required. Recently, an optical method called collinear photofragmentation and atomic absorption spectroscopy (CPFAAS) for detection of KCl vapor at parts per billion (ppb) level was introduced /2/. CPFAAS is a derivative of Beer- Lambert law and its calibration in gas sensing is directly based on the physical constants /3/. The method combines the benefits of the previously presented KCl detection techniques, excimer laser induced fragmentation fluorescence (ELIF) /4/ and differential optical absorption spectroscopy (DOAS) /5/, in sensitivity and in diagnostics of large sample volumes. In this study, a KCl detector based on CPFAAS was integrated to the flue gas channel of 4 MW th circulating fluidized bed boiler. The performance of the KCl detector was studied during the start-up of the boiler and during KCl mitigating ferric sulfate additive injections. The measured KCl values were compared to HCl readings from FTIR in order to verify their reliability.

2 EXPERIMENTAL METHODS 2.1 Measurement environment The 4 MW th combustion test facility used for the tests has a height of 13 m and a furnace cross-section area of 1.21 m 2. The boiler is water cooled, and the output heat is used in the local district-heating network. The CFB furnace is provided with optional cooling surfaces to facilitate the heat transfer for various test configurations. The furnace was combusted with straw during the ferric sulfate injection experiments. The composition of the used straw is presented in Table 1. Due to the high content of the potassium in the fuel, the chlorine occurs mainly in forms of KCl and HCl in the flue gases. Table 1. Composition and heating value of straw used as fuel. Total moisture 9.7 % Gross calorific value, dry basis 18 MJ/kg Bulk density 570 kg/m 3 Net calorific value, dry basis 17 MJ/kg Analysis moisture 5.9 % Net calorific value, as received 15 MJ/kg Ash 550 C, dry basis 8.1 % Ash 815 C, dry basis 7.5 % Aluminium (Al) 280 mg/kg Volatile matter, dry 74 % Phosporous (P) 300 mg/kg Potassium (K) 7600 mg/kg Carbon, C dry 46 % Calcium (Ca) 3700 mg/kg Hydrogen, H dry 5.9 % Magnesium (Mg) 590 mg/kg Nitrogen, N dry 0.61 % Manganese (Mn) 21 mg/kg Oxygen, O dry, calculated 40 % Silicon (Si) 23000 mg/kg Fluoride, F dry 0.01 % Chloride, Cl dry 0.29 % Sulfur, S dry 0.13 % Iron (Fe) 160 mg/kg 2.2 Ferric sulfate reaction with KCl The superheater corrosion risk of high- potassium and high- chlorine fuels is reduced through changing of the flue gas atmosphere with fuel fractions containing sulfur or additives containing elemental sulfur or sulfate /6,7/. In this study the ferric sulfate was used to eliminate alkali chlorides in the flue gases. Ferric sulphate is thermally dissociated in high temperature followed by the formation of SO 3. Sulfur trioxide reacts with KCl through reaction 2 KCl + SO 3 + H 2 O K 2 SO 4 + 2 HCl (1) to less corrosive potassium sulfate and hydrogen chloride. Equation 1 indicates that the absolute changes in HCl concentration can be used to estimate the absolute changes in KCl concentration. 2.3 Online laser measurement The physical steps in collinear photofragmentation and atomic absorption spectroscopy (CPFAAS) are presented in Fig. 1. The method is based on the continuous monitoring of the potassium atoms in the sample with an infrared laser having wavelength of 766.5 nm that corresponds to the absorption line of the potassium atoms. Fig. 1 a) shows how he IR laser beam is aligned to go through the sample containing KCl molecules. IR transmission may fluctuate significantly, when the sample itself is unstable. The actual measurement event starts when a UV pulse is triggered to pass the sample along the same optical path with the IR laser beam (Fig 1. b)). The wavelength of the UV pulse is selected to correspond to the dissociative UV absorption band of KCl molecules and, therefore, the amount of K atoms is temporarily increased. The increased amount of K atoms decreases the transmission of the IR laser due to the atomic absorption. Fig 1. c) shows how the K atoms recombine with Cl atoms or react with ambient gas compounds, when the IR transmission recovers to same level where it was just before the triggering of the UV pulse. A single measurement event lasts typically less than 5 µs. Transmission fluctuations, due to e.g. fly ash movement on measurement path, are negligible within short measurement time window and can be ignored.

Figure 1. The principles of KCl detection using CPFAAS. a) IR laser measures the temporal transmission through the sample. b) UV pulse fragments KCl molecules inducing a fast decrease in IR transmission. c) The released K atoms react with ambient gases and IR transmission recovers its original level. The recovery is fast compared with sample fluctuations. CPFAAS is shown to be capable of measuring accurate KCl concentrations from pure samples in laboratory conditions /3/. The power plant environment sets new challenges to CPFAAS technique due to other gas species absorbing the same 266 nm UV wavelength than KCl. The absorption by the other gases decreases the intensity of fragmenting UV pulse faster than it would be decreased in pure KCl sample and, therefore, a smaller amount of KCl molecules is fragmented. In this study the distortion due to the other gases was minimized by measuring the UV intensities before and after the sample. The UV attenuation was approximated to be even through the sample, when a profile for the UV intensity through the furnace could be solved. By adding the approximated UV intensity profile to the previously introduced equation /3/ the relation between the maximum absorbance αl max, achieved right after the UV pulse has passed the sample, and a true KCl concentration x KCl could be solved. The highest measurement repetition rate in CPFAAS technique is limited by the recovery time of the K atoms and the maximum pulse repetition rate of the UV laser. In this study KCl detector was set to collect data with 7 HZ repetition rate and in 200 signal sequences corresponding to ~30 s measurement windows. 3 RESULTS AND DISCUSSION Three single transmission curves resulting KCl concentrations of 5 ppm, 50 ppm and 230 ppm are presented in Fig. 2. The fragmentation of KCl molecules was induced with average UV input energy of 23 µj and UV beam cross sectional area of 3.6 cm 2. The presented single transmission curves demonstrate the potential of the technique. In all example signals signal to noise ratio was found to be 25 or better. The fast drop at t=0 becomes deeper as sample concentration is increased. The recovery times were found to vary, which indicates the fluctuations in gas atmosphere of the flue gas. The fluctuation in the base levels of the signals at t<0 indicates the changing visibility through the sample. The evolution of the KCl concentrations during the start-up of the furnace is presented in Fig. 3 a). The fuel was composed of 15 % of straw and 85 % of wood. The KCl readings are presented as distributions of KCl concentrations measured within 30 measurement time windows. The distributions are normalized with their maxima in order to enhance the contrast of the figure. The startup is done in three steps, which are seen as three

Figure 2. Three single CPFAAS transmission curves recorded in the vicinity of the UV fragmentation. The concentrations are calculated from the depths of dips at t=0. different KCl levels. Within stable combustion conditions KCl detector resulted readings with small ~5 % variation. The lowest reliably detected KCl concentration was 0.5 ppm. Fig. 3 b) presents an experiment, in which the load and the fuel of the furnace remained constant, but the flue gas atmosphere was modified with ferric sulfate injections. Ferric sulfate was injected with four different injection rates. The injection times are highlighted in Fig. 3 b) with double arrays and numbers 1, 2, 3 and 4. At the beginning of the measurement the furnace was fired with 100 % of straw. The KCl and HCl concentrations were 235 ppm and 71 ppm, respectively. When the injection was turned on KCl concentration decreased within one minute to 85 ppm and simultaneously HCl increased up to 179 ppm. Similar KCl decrease and HCl increase behavior was observed when injection rate was increased in phases 2, 3 and 4. At the end of the experiment the injection was turned off, when KCl increased to 249 ppm and HCl decreased to 72 ppm. The temperature of the flue gas decreased 20 C when injection was started and increased 20 C when injection was turned off. The difference between the flue gas temperature at start point and the end point was ~10 C. a) b) Figure 3. a) Evaluation of KCl concentration during start-up of furnace, and b) averige KCl (black) and HCl (gray) concentrations during ferric sulfate injections.

Table 2. Measured gas concentratrions at different injection phases. Phase HCl [ppm] KCl [ppm] beginning 71 235 1 179 85 2 199 69 3 249 39 4 268 25 end 71 249 The measured gas concentrations at the different ferric sulfate injection phases are listed in Table 2. The most dramatic changes are seen when injection is started and ended, but these reference points are invalid for comparison of ΔKCl and ΔHCl due to changed temperature of flue gas. Otherwise, the temperatures remained nearly constant between the additive injection rates changes and comparisons between ΔKCl and ΔHCl was reasonable. The comparison of the absolute changes showed approximately relation of ΔKCl 0.75*ΔHCl. The observed relation justifies the order of the measured KCl values. The 25 % difference between ΔKCl and ΔHCl is reasonable when the gas sampling is taken into account. FTIR collected gas sample from the single point in the middle of the flue gas channel while the KCl detector measured the average gas concentration through the whole channel. Therefore, already small differences in spatial spreading of the injected additive caused different kind of measurement conditions for the two detectors. The 25 % difference can also be partly explained with estimated 1 m sample length. The accuracy of the sample length was 10 % that scales directly to the calculated KCl concentrations. 4 CONCLUSIONS CPFAAS method was applied to monitor KCl vapor concentration in the flue gas channel of 4 MW th circulating fluidized bed boiler. Single measurement curves were discussed and the operation of the detector in two longer term experiments was demonstrated. The detector measured stable concentration values in steady combustion conditions and followed the changing concentrations due to the furnace operational changes with time resolution less than a minute. The measured KCl concentration changes during the additive injections were also compared to HCl concentrations measured with FTIR detector. The relation between the detected KCl and HCl concentration changes was found to be ΔKCl 0.75*ΔHCl, which verifies the correct order of magnitude for the measured KCl readings. The presented online laser measurement of KCl is suitable for continuous online monitoring as the calibration of the device can be maintained by following the UV laser pulse energies at the laser output and after the transmission through the process volume. Thus, the technique enables a stable and controlled KCl concentration in biomass combustion due to actively adjusted ferric sulfate injection rate. ACKNOWLEDGMENTS The financial support of the graduate school of Tampere University of Technology for TS is acknowledged. REFERENCES 1. Nielsen H., Frandsen F., Dam-Johansen K., Baxter L.: The implications of chlorine-associated corrosion on the operation of biomass-fired boilers. Progress in Energy and Combustion Science 26 (2000) 283 298. 2. Sorvajärvi T., Saarela J., Toivonen J.: Optical detection of potassium chloride vapor using collinear photofragmentation and atomic absorption spectroscopy. Optics Letters 37 (2012) 4011 4013. 3. Sorvajärvi T., Toivonen J.: Principles and calibration of collinear photofragmentation and atomic absorption spectroscopy. Applied Physics B 115 (2014) 533-539. 4. Gottwald U., Monkhouse P.: Single-port optical access for spectroscopic measurements in industrial flue gas ducts. Applied Physics B 69 (1999) 151-154.

5. Forsberg C., Bröström M., Backman R., Edvardsson E., Badiei S., Berg M., Kassman H.: Principle, calibration, and application of the in situ alkali chloride monitor. Review of Scientific Instruments 80 (2009) 023104. 6. Silvennoinen J., Hedman M.: Co-firing of agricultural fuels in a full-scale fluidized bed boiler. Fuel Processing Technology 105 (2013) 11-19. 7. Aho M., Vainikka P., Taipale R., Yrjas P.: Effective new chemicals to prevent corrosion due to chlorine in power plant superheaters. Fuel 87 (2008) 647-654.