ANALYSIS OF PESTICIDE RESIDUES IN DRINKING WATER AS PER BUREAU OF INDIAN STANDARDS USING THE AGILENT 7000 GC/MS/MS WITH PESTICIDES ANALYZER

Similar documents
Analysis of 16 Pesticide Residues in Broccoli Using CarbonX dspe QuEChERS AOAC Kits Using GC/MS

Accurately Identify and Quantify One Hundred Pesticides in a Single GC Run

Application Note. Abstract. Authors. Introduction. Environmental

Determination of organochlorine pesticide residues in eggs by gel-permeation chromatography (GPC) cleanup and GC-ECD

Determination of Persistent Organic Pollutants (POPs) in Fish

Application Note: Multi-Residue Pesticides in Fruit and Vegetables (9/16/08)

EPA Method 543: Selected Organic Contaminants by Online SPE LC/MS/MS Using the Agilent Flexible Cube

Optimization of Volatile Petroleum Hydrocarbons Analysis by Automated Headspace Using Method Development Tools

Analysis of Suspected Flavor and Fragrance Allergens in Cosmetics Using the 7890A GC and Capillary Column Backflush Application

Olga Shimelis, Katherine Stenerson, Michael Ye, Carmen T. Santasania, and Ed Mauney. Supelco, Div. of Sigma-Aldrich, Bellefonte, PA USA

Size Exclusion Chromatography of Biosimilar and Innovator Insulin Using the Agilent AdvanceBio SEC column

Application Note. Author. Abstract. Environmental. Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany

On-line SPE-LC/MS/MS to Detect Organonitrogen and Triazine Pesticides at 10ng/L in Drinking Water

Tyler Trent, Applications Sales Specialist; Teledyne Tekmar Page 1

Analysis of USP <467> Residual Solvents using the Agilent 7697A Headspace Sampler with the Agilent 7890B Gas Chromatograph

Quantitative determination of residual 2-(2-chloroethoxy) ethanol (CEE) in quetiapine fumarate by gas chromatogaraphy

DETERMINATION OF ELEMENTS IN DRINKING WATER AS PER BUREAU OF INDIAN STANDARDS 10500, & USING THE AGILENT 5100 ICP-OES

CHROMATOGRAPHY/MASS SPECTROMETRY

Determination of Monocrotophos, Diazinon, Malathion, EPN, and Methamidaphos from Aqueous Samples Using Atlantic HLB SPE Disks

With the globalization of trade, food production

Gas Chromatography Assignment Chem 543/443

Automated Online SPE for LC/MS/MS Analysis of Trace Organic Contaminants in Water Using the Agilent 1290 Infinity Flexible Cube Module

Using the Pulsed Flame Photometric Detector (PFPD) for Low-level Analysis of Organophosphorus Pesticides

Disulfide Linkage Analysis of IgG1 using an Agilent 1260 Infinity Bio inert LC System with an Agilent ZORBAX RRHD Diphenyl sub 2 µm Column

MS Certified Vials. Pre-cleaned and certified vials for mass spectrometry

Capillary GC Troubleshooting: a Practical Approach

Electron Capture Detector for GC-2010 Plus

Improved SPE for UPLC/MS Determination of Diquat and Paraquat in Environmental

Quantitation of cyanotoxins in drinking water according to EPA 544 guidelines

Development of a Fast and Cost-Effective Multi-Residue Method to Determine Pesticides in Tobacco by LC/MS/MS

International Journal of Pharma Research & Review, Feb 2014; 3(2):11-16 ISSN:

A green HPLC technique with a 100% water mobile phase for detecting imidacloprid and its metabolite 6-chloronicotinic acid

Protocol for Quantitative Determination of Residual Solvents in Cannabis Concentrates Prepared by: Amanda Rigdon, May 23 rd, 2016

Validation Report 19

AppNote 5/2006 ABSTRACT

Volatile Organic Compounds in Solids PBM

Separation and Quantitation of a Mixture of Alcohols by Gas Chromatography

Chem 321 Lecture 23 - Liquid Chromatography 11/19/13

Agilent 7820A GC System THE EASY CHOICE FOR ROUTINE ANALYSES

HYDROGEN AS A CARRIER GAS. Gas Chromatography new frontiers: can Hydrogen substitute Helium?

Formulas: See Table 1 MW: Table 1 CAS: Table 1 RTECS: Table 1. METHOD: 2555, Issue 1 EVALUATION: PARTIAL Issue 1: 15 March 2003

APPLICATION NOTE. UOP : Analysis of Trace CO and CO 2 in bulk H 2 and Light Gaseous Hydrocarbons by GC

Low-level Detection of Ethanol, 1,4-Dioxane, and Other Oxygenates Using the Eclipse Purge-and-Trap Sample Concentrator

SCION GC Series. The Gas Chromatographers Choice for Separations

Co-ordination of an organochlorine pesticides in drinking water proficiency testing scheme in South(ern) Africa

The Agilent 5975C Series GC/MSD Performance, productivity and confidence. Our measure is your success.

Implement High Quality, Routine GC-MS/MS Multi-Residue Pesticides Analysis with Ease Paul Silcock & Jason Cole

Quality-by-Design-Based Method Development Using an Agilent 1290 Infinity II LC

The Importance of High Speed Data Acquisition in the Design and Implementation of GC/MS Applications

Application Note. Author. Abstract. Biopharmaceuticals. Verified for Agilent 1260 Infinity II LC Bio-inert System. Sonja Schneider

C146-E096 TD-20. Shimadzu Thermal Desorption System

Extraction and Analysis of Neonicotinoid Pesticides using QuEChERS Approach

MultiQuant Software Version 3.0

Confident Pesticides Analysis with the Agilent LC/Triple Quadrupole and TOF/QTOF Solutions

Extraction of POPs from Environmental Samples using Accelerated Solvent Extraction (ASE )

Determination of Iodinated Contrast Media in Aqueous Samples by Direct-Injection LC-MS/MS

TurboMass GC/MS Software

STANDARD OPERATING PROCEDURES

PAHs in Surface Water by PDA and Fluorescence Detection

Automated Real-Time Determination of Bromate in Drinking Water Using LC-ICP-MS and EPA Method Application

Determination of Oxalate and Other Anions in Bayer Liquor Using Ion Chromatography

Minimize Method Development Time for Pesticide Screening Using LC/MS Application Kit

Validation Report 16

Certified Reference Material - Certificate of Analysis

Composition Analysis of Animal Feed by HR ICP-OES

EU Document on Analytical Quality Control and Method Validation Procedures for Pesticide Residues Analysis in Food and Feed An update

Glyphosate and AMPA in drinking water

Heart-cut 2D-LC/MS approach for pharmaceutical impurity identification using an Agilent 6540 Q-TOF LC/MS System

Agilent Ultimate Plus Fused Silica Tubing

THE SOURCE OF NEW POSSIBILITIES

Preparative HPLC is still the

Determination of Perchlorate in Drinking Water by Ion Chromatography

Agilent 6430 Triple Quadrupole LC/MS System

Fraction Analysis of Cysteine Linked Antibody-Drug Conjugates Using Hydrophobic Interaction. chromatography. Agilent 1260 Infinity II Bio-Inert System

Validated Stability-indicating assay method for determination of Ilaprazole in bulk drug and tablets by high performance liquid chromatography

Research Paper Development of Stability Indicating Reverse Phase HPLC Method for Aripiprazole from Solid Dosage form

A Fast and Effective Approach for Running EPA Method 539: Determination of Hormones in Drinking Water using SPE and LC/MS/MS

Fast and High-Resolution Reversed-Phase Separation of Synthetic Oligonucleotides

1-(2-fluorophenyl)-2-(methylamino)propan-1-one

Exploring Extra Sensitivity Using ionkey/ms with the Xevo G2-XS Q-Tof HRMS for Small Molecule Pharmaceutical Analysis in Human Plasma

Monitoring for Microcystins in Raw Water Supply Reservoirs Using the Agilent 6410 Triple Quadrupole LC/MS

Characterization of mab aggregation using a Cary 60 UV-Vis Spectrophotometer and the Agilent 1260 Infinity LC system

CONTENTS Alternate Retention Time Window Determination

Automated Solid Phase Extraction of PAH Compounds Utilizing the SPE-DEX 5000

KEYWORDS ABSTRACT AppNote 7/2000

Appendix A. Drinking Water Standards and Guidelines for Pesticides Sampled in Oregon Ground and Surface Water

Passive Monitoring: A Guide to Sorbent Tube Sampling for EPA Method 325. Nicola Watson July 2015

Guidelines on Quality Assurance/Quality Control for The Environmental Measurements of Dioxins

Chem-Lab High Purity Solvents

Agilent 7693A Series Automatic Liquid Sampler INJECT NEW PERFORMANCE INTO YOUR GAS CHROMATOGRAPHY

PROFICIENCY TESTING PROGRAM. June 2014

Agilent Sample Preparation Portfolio ACCURATE, REPRODUCIBLE RESULTS RIGHT FROM THE START

Concentration of Human Hormones in Drinking Water Using Solid Phase Extraction and Analysis by High Performance Liquid Chromatography

DETERMINATION OF MYCOTOXINE DEOXYNIVALENOL IN WHEAT FLOUR AND WHEAT BRAN

Available Online through (or) IJPBS Volume 2 Issue 3 JULY-SEPT Research Article Pharmaceutical Sciences

N. Tamilselvi *, Dona Sara Kurian. Department of Pharmaceutical Analysis. KMCH college of pharmacy Coimbatore

RP-HPLC METHOD FOR QUANTITATIVE ESTIMATION OF GLATIRAMER ACETATE FOR INJECTION IN PHARMACEUTICAL DOSAGE FORMS

Agilent 7693A Series Automatic Liquid Sampler. Inject new performance into your gas chromotography

Super Clean Gas Filters Simply the best

Physical Stability of a Silica- Based Size Exclusion Column for Antibody Analysis

Transcription:

ENVIRONMENTAL ANALYSIS ANALYSIS OF PESTICIDE RESIDUES IN DRINKING WATER AS PER BUREAU OF INDIAN STANDARDS USING THE AGILENT 7000 GC/MS/MS WITH PESTICIDES ANALYZER Solutions for Your Analytical Business Markets and Applications Programs Authors Indrajit Sen & Dr. Samir Vyas Agilent Technologies Inc., India ABSTRACT An Agilent 7890B GC with 7000C MSD method for the analysis of residual pesticides in drinking water was developed, validated and successfully implemented for routine analysis as per the Bureau of Indian Standards (BIS). The method used the multi-mode inlet in PTV solvent vent mode, retention time locking and concurrent backflushing for the time effective elimination of less volatile matrix components from the GC column, which reduced cycle time and also prevented contamination of the MS source. This method used a simple sample preparation technique of liquid liquid extraction. Method detection limits of 5 ng/l were achieved for all pesticides, with linearity correlation coefficients > 0.995. INTRODUCTION The presence of organic contaminants in the environment are a result of pollution from various anthropogenic activities [1]. Pesticides, generated by the intensification of agriculture, are regarded as some of the most dangerous contaminants of the environment, despite their numerous merits. Not only are they toxic, they are also mobile and capable of bioaccumulation. In addition, they can take part in various physical, chemical and biological processes. Many of these pesticides are characterized by a strong persistence which explains their wide presence in the environment [2-4]. Due to their physicochemical characteristics and extensive use, many of these pesticides are common in surface water and a growing number of aquifers. Their presence in water is considered a potential risk not only for drinking water quality and human health, but also for ecosystems [3]. India is one of the largest pesticide consumers in the world. Due to the possible risks associated with pesticide use, water needs to be monitored. The Bureau of Indian Standards (BIS) sets regulatory limits for residual pesticides at trace levels (sub ppb) in drinking water (IS 10500) [5], packaged drinking water (IS 14543) [6] and natural mineral water (IS 13428) [7]. This study describes a method to determine pesticide residues at trace levels in water using liquid-liquid extraction and analysis using large volume injection (LVI) on the Agilent 7000C GC/MS/MS. Twenty eight pesticides were analyzed in water samples, including isomer and metabolite compounds. This technique is based on retention time locking and back flushing mode [8-9]. A method detection limit of 5 ng/l and quantitation limit of 10 ng/l were determined for each pesticide, with correlation coefficients > 0.999 (10-100 ng/ml). Parameters for the method validation were established by the European Commission (SANCO and SANTE) [10].

ANALYTICAL TECHNIQUE Reagents and chemicals Ethyl acetate and Dichloromethane (HPLC grade) were purchased from J.T. Baker. Water was purified in-house with a Milli-Q system (Millipore, Tokyo, Japan). Reagent grade anhydrous sodium sulfate, K2HPO4, sodium hydroxide and hydrochloric acid were obtained from Sd Fine Chemicals, Mumbai, India. All pesticides were of greater than 99% purity and were obtained from Sigma-Aldrich (USA). Stock solutions of individual pesticides at 100µg/mL were prepared by dissolving each pesticide standard in ethyl acetate and storing the solution at -20 C. Preparation of Calibration Standards and Extracted Calibration Standards Working standard solutions were prepared by diluting the stock solutions to 10µg/mL in ethyl acetate. Appropriate aliquots were taken and further diluted with ethyl acetate to give a series of calibration standard solutions with concentrations of 10 ng/ml, 20 ng/ml, 50 ng/ml, 75 ng/ml and 100 ng/ml. Extracted calibration standards were prepared by spiking blank water samples (1 L) with the working standard solution, prior to extraction, to give concentrations of 10 ng/l, 20 ng/l, 50 ng/l, 75 ng/l and 100 ng/l. After extraction and reconstitution (1 L to 1 ml) these equate to concentrations of 10 ng/ml, 20 ng/ml, 50 ng/ml, 75 ng/ml and 100 ng/ml. 2

Instrumentation Agilent 7890B Gas Chromatograph Agilent 7000C Mass Selective Triple Quadrupole Detector GC/MS/MS Operating Conditions GC Conditions Agilent 7890B Columns 1. HP-5ms: 15 m x 0.25 mm x 0.25 μm 2. HP-5ms: 15 m x 0.25 mm x 0.25 μm Inlet Agilent Multimode Inlet (MMI) Split Mode and Ratio Solvent vent mode (purge flow to split vent 60 ml/min at 2.62 psi & vent 25mL/min at 14.31 psi for 0.12 min) Inlet Temperature 85 C (0.12 min), 600 C/min to 325 C Oven 60 C (1 min), 40 C/min to 120 C, 5 C/min to 310 C Carrier Gas Run Time 99.9995% Helium constant flow mode at 1.5 ml/min (2 psi during backflush) for column 1 (In front MMI, Out Aux EPC). Column 2 flow 1.7mL/min (In Aux EPC, Out MSD). Aux EPC pressure: 3.0 psi during run, 50 psi during backflush. 40 minutes MS Conditions Quadrupole Temperature 180 C Ion Source Temperature 300 C Transfer Line Temperature 280 C Acquisition Mode EI MRM Agilent 7000C 3

RESULTS AND DISCUSSIONS The selectivity or specificity of the analytical method for the determination of 28 pesticides in water is shown in Figure 1 and Table 1. The calibration curve for each pesticide was plotted using peak area against concentration of pesticide in both the water matrix and aqueous solvent. They were linear with correlation coefficients (r2) > 0.995, as shown in Table 1. Method detection limits of 5 ng/l were determined for all pesticides, by using the signal to noise (S/N) ratio of 3:1 for the strongest mass transition (MRM) with respect to the background noise obtained from the blank sample and method quantitation limits (MQL) of 10 ng/l were determined similarly by considering the signal to noise ratio (S/N) ratio of 10:1. Recovery and precision studies were carried out both inter-day and intra-day at three different spiked concentration levels (10 ng/l, 20 ng/l and 50 ng/l). Recoveries of the 28 pesticide residues were in the range of 88-99%, with RSDs of less than 8%. Results are shown in Table 1. Figure 1. Extracted pesticides standard TIC. 4

Pesticide R.T Min Spiking Level ng/l % Recovery % RSD MRM (Aqueous) (Matrix) Monocrotophos 13.162 10.0 92.18 5.78 20.0 94.24 4.82 50.0 94.71 4.39 127.1>109.0 (10) 127.0>95.0 (15) 0.9994 0.9992 Phorate 13.278 10.0 88.29 6.89 20.0 89.64 7.45 50.0 92.57 5.81 121.0>47.0 (30) 121.0>65.0 (10) 0.9985 0.9956 Alpha-BHC 13.343 10.0 96.84 3.89 20.0 97.69 5.82 50.0 98.58 4.60 216.9>181.0 (5) 218.9>183.0 (5) 0.9994 0.9994 Atrazine 14.433 10.0 95.71 4.91 20.0 96.64 3.86 50.0 96.79 3.73 214.9>58.1 (10) 214.9>200.2 (10) 0.9995 0.9991 Beta-BHC 14.579 10.0 96.60 3.81 20.0 96.91 4.01 50.0 98.12 3.66 216.9>181.1 (5) 181.0 >145.0 (15) 0.9993 0.9991 Gamma-BHC 14.581 10.0 97.42 3.66 20.0 97.29 3.86 50.0 98.94 2.57 216.9 >181.0 (5) 181.0>145.0 (15) 0.9989 0.9990 Delta-BHC 15.517 10.0 95.35 3.53 20.0 97.69 3.39 50.0 97.77 3.11 217.0>181.1 (5) 181.1>145.1(15) 0.9997 0.9994 Paraoxon-methyl 15.548 10.0 90.11 5.82 20.0 90.52 4.60 50.0 93.60 4.91 229.9>106.0 (15) 229.9>200.1 (5) 0.9986 0.9974 Parathion-methyl 17.103 10.0 91.10 4.82 20.0 91.92 4.29 50.0 92.67 4.22 262.9>109.0 (10) 125.0>47.0 (10) 0.9988 0.9987 Malaoxon 17.333 10.0 89.55 5.96 20.0 93.51 5.88 50.0 94.39 5.13 127.0>99.0 (5) 127.0>55.0 (5) 0.9991 0.9992 Alachlor 17.434 10.0 94.76 3.39 20.0 95.01 3.11 50.0 96.53 2.50 181.1>160.2 (10) 160.0>132.1 (10) 0.9996 0.9986 Phorate sulfoxide 18.480 10.0 90.22 4.98 20.0 89.10 5.19 50.0 92.54 4.23 153.0>96.9 (10) 96.9>64.9 (20) 0.9997 0.9994 Aldrin 18.525 10.0 96.28 3.57 20.0 98.12 2.94 50.0 98.58 2.22 262.9>192.9 (35) 254.9>220.0 (20) 0.9992 0.9996 Malathion 18.689 10.0 93.35 3.66 20.0 94.20 3.58 50.0 94.85 2.88 126.9>99.0 (5) 172.9>99.0 (15) 0.9984 0.9971 5

Pesticide R.T Min Phorate sulfone 18.770 Chlorpyrifos 18.998 o,p-dde 21.207 α-endosulfan 21.351 Butachlor 21.709 Dieldrin 22.307 p,p-dde 22.381 o,p-ddd 22.669 β-endosulfan 23.420 p,p-ddd 23.883 o,p-ddt 23.980 Ethion 24.167 Endosulfan sulfate 24.981 p,p-ddt 25.212 Spiking Level ng/l % Recovery % RSD 10.0 88.49 6.43 20.0 91.26 6.19 50.0 91.71 5.80 10.0 96.83 3.73 20.0 97.85 2.50 50.0 98.56 2.44 10.0 97.08 3.58 20.0 97.87 2.99 50.0 98.90 2.52 10.0 93.48 4.75 20.0 95.92 4.39 50.0 96.21 3.99 10.0 94.80 3.86 20.0 96.55 3.22 50.0 97.69 3.09 10.0 95.86 4.33 20.0 97.01 2.89 50.0 97.52 3.24 10.0 96.86 3.79 20.0 97.54 3.11 50.0 98.01 2.80 10.0 97.32 2.91 20.0 97.66 3.07 50.0 98.07 2.16 10.0 95.53 4.91 20.0 96.27 4.66 50.0 96.92 2.49 10.0 96.77 3.27 20.0 97.51 2.33 50.0 98.39 2.19 MRM 153.0>97.0 (10) 124.9>96.9 (5) 196.9>169.0 (15) 198.9>171.0 (15) 246.0>176.2 (30) 248.0>176.2 (30) 194.9>159.0 (5) 194.9>160.0 (5) 176.1>147.1 (10) 188.1>160.2 (10) 262.9>191.0 (35) 277.0>241.0 (5) 246.1>176.2 (30) 315.8>246.0 (15) 235.0>165.2 (20) 237.0>165.2 (20) 206.9>172.0 (15) 194.9>158.9 (10) 234.9>165.1 (20) 236.9>165.2 (20) (Aqueous) (Matrix) 0.9965 0.9969 0.9992 0.9989 0.9992 0.9999 0.9997 0.9997 0.9990 0.9991 0.9997 0.9998 0.9999 0.9995 0.9999 0.9999 0.9989 0.9992 0.9999 0.9997 10.0 97.44 2.98 234.9>165.1 (20) 20.0 98.76 2.55 236.9>165.2 (20) 0.9998 0.9999 50.0 98.99 2.42 10.0 96.85 3.14 20.0 96.79 2.86 50.0 97.93 2.45 10.0 93.08 5.43 20.0 95.21 5.11 50.0 95.34 3.37 10.0 96.88 3.81 20.0 97.83 2.93 50.0 98.71 2.42 230.9>129.0 (20) 230.9>175.0 (10) 271.9>237.0 (15) 273.8 >238.9 (15) 235.0>165.2 (20) 237.0>165.2 (20) 0.9996 0.9994 0.9984 0.9983 0.9998 0.9999 6

CONCLUSIONS The design of the Agilent 7000C Triple Quadrupole GC/MS enables pesticides to be determined at lower detection limits when combined with an inert sample path and GC column backflushing. The high sensitivity EI Extractor Ion Source with improved thermal characteristics delivers confident trace analysis even in complex matrices, and the Triple-Axis HED-EM Detector reduces neutral noise by the doubly off-axis position of the HED-EM. The simple, easy, economical and efficient method was developed for the determination of 28 pesticides in water. The method demonstrates good sensitivity, precision, accuracy and allows for rapid analysis. The results demonstrate that pesticide residues can be detected below the current maximum residue levels (MRL) required by the BIS specifications. Method detection limits of 5 ng/l and quantitation limits of 10 ng/l were achieved for all pesticides. The method is ideally suited for use in a regulatory laboratory for the determination of pesticides in surface, drinking and packaged drinking water. REFERENCES 1. C-K. Meng, Improving productivity and extending column life with backflush, Agilent Technologies Publication 5989-6018EN, 2006. 2. K. Maštovská, K and P. L. Wylie, Evaluation of a new column backflushing set-up in the gas chromatographic-tandem mass spectrometric analysis of pesticide residues in dietary supplements, Journal of Chromatography A, 1265, 2012, 155-164. 3. Pesticide Analysis Reference Guide, GC/MS/MS Pesticide Residue Analysis, Agilent Technologies Publication 5991-2389EN, 2013. 4. European Food Safety Authority, Conclusion on the peer review of the pesticide risk assessment of the active substance etridiazole, EFSA Journal 2010; 8(10):1823. 5. Bureau of Indian Standards, IS 10500 (2012): Drinking Water [FAD 25: Drinking Water], with latest amendments. 6. Bureau of Indian Standards, IS 14543 (2004): Packaged Drinking Water (Other than Packaged Natural Mineral Water) [FAD 14: Drinks and Carbonated Beverages], with latest amendments. 7. Bureau of Indian Standards, IS 13428 (2005): Packaged Natural Mineral Water [FAD 14: Drinks and Carbonated Beverages], with latest amendments. 8. P. L. Wylie and C-K. Meng, A method for the trace analysis of 175 pesticides using the Agilent triple quadrupole GC/MS/MS, Agilent Technologies Publication 5990-3578 EN, 2009. 9. C-K. Meng, The GC/MS/MS Analyzer and the Pesticides and Environmental Pollutants MRM Database, Agilent Technologies Publication 5990-9453EN, 2011. 10. European Commission Health & Consumer Protection Directorate-General, Guidance document on analytical quality control and validation procedures for pesticide residues analysis in food and feed, SANCO/12571/2013, 19 November 2013. Solutions for your analytical business Markets & Applications Programs www.solutions-to-win.com This information is subject to change without notice. Agilent Technologies, Inc. 2016 Printed in EU, 24 May, 2016 5991-6999EN