HemoCheck TM Sample Policy. SUBJECT: Detection of blood residue on various surfaces

Similar documents
One such problem is blood residual inside the channels of a robotic arm.

Guide for Cleaning, Sterilization and Storage of Introducers for Caldera Medical Implants

Best Practices for Reprocessing Endoscopes

Competency Guide: Care and Handling of Rigid Endoscopes

Water for Instrument Processing

Standards of Infection Prevention in Reprocessing Flexible Gastrointestinal Endoscopes Webinar FAQ

Best Practices for Reprocessing Robotic Instruments

QUALITY ASSURANCE IN AN MDRD

Flexible Endoscopes: The A,B,C s of Monitoring Manual Cleaning Efficacy!

Aesculap SterilContainer System Instructions for Use Steam, Ethylene Oxide and Vapor Hydrogen Peroxide Sterilization Modalities

CAUTION: FEDERAL LAW (USA) RESTRICTS THESE DEVICES TO SALE BY OR ON THE ORDER OF A PHYSICIAN.

Endoscope Washer Disinfector Cleaning Process Challenge Devices Research & Validation

Sterilization and Disinfection: Getting It Right. Instructions

Guidelines for Selection and Use of Disinfectants

INSTRUMENT PROCESSING FROM PREP/PACKAGING TO STORAGE. January 2016

Instructions for use. Reusable surgical instruments. Hersteller:

Sterilization Service Revolution Service Enhancement in HA Hospitals

Responses to Questions for SYSTEM 1E Process Monitoring and Validation Webinar

Getinge da Vinci Solution

St Boniface Research Centre Winnipeg, Manitoba Canada

Safe to Handle? Comparing Manually and Machine-Washed Medical Devices

Is contamination of bronchoscopes really a problem?

01.48 Receiving and Processing Vendor Owned Surgical Instrumentation in the Sterile Processing Department

In use test of a Washer-disinfector reprocessing of flexible endoscopes

What Dentistry Can Learn from AAMI ST79. Best Practices for device and instrument reprocessing

8/27/2015. Learning Objectives. Purpose of Sterilization. To destroy all living pathogenic microorganisms

UNIVERSITY OF TOLEDO HEALTH SCIENCE CAMPUS

Best Practices for High Level Disinfection Part I

Contaminated or clean? Your eye can t tell. The 3M Clean-Trace ATP Monitoring System can.

Sterilization of Schweickhardt Instruments

CSA Z314 Series Standards Understanding the Standards

New Requirements of ISO 17664/AAMI ST 81

Comparison of ANSI/AAMI ST-91 Standards, SGNA Standards and AORN Guidelines on Processing Endoscopes March 9, 2017

Symphion 6.3 Endoscope

In the Know about.appropriate Use of 3M Attest Rapid Readout Biological Indicators

Porcine IgM (Immunoglobulin M) ELISA Kit

AS/NZS 4187 RELEVANT STANDARDS. Robyn Williams NUM CSSD LBHHS & Standards Australia Chair HE-023

Volume 13, Issue 5/ Medtech

Designing an effective cleaning procedure for medical devices through laboratory studies

Validation of cleaning and sterilization of reusable medical devices

CAN STERILE REALLY BE STERILE?: EFFECTIVE AUTOCLAVING METHODS

GENERAL SURGERY DEVICES

Products made in Switzerland

Note: for laboratory research use only. RNA High-purity Total RNA Rapid Extraction Kit (Spin-column) Signalway Biotechnology

Packaging Systems for Central Service Operations

Testing of routinely sterilized products should

The Role of Cleaning and Sterilization in Infection Control:

35.0 Sterile Supply Unit (SSU)

Clostridium difficile Update Dr. Michelle Alfa, Winnipeg Sponsored by ARJO

Instruction Guide Re-Usable Surgical Instruments. Instruction Guide Re-usable surgical instruments (GA 30.1)

STERILIZATION VALIDATION IN THIS SECTION REPROCESSING VALIDATIONS FOR REUSABLE MEDICAL DEVICES. Radiation Sterilization Validation

Human Anti-Hepatitis A Virus IgG (HAV-IgG) ELISA kit

Standard Operating Procedure for performing a muscle biopsy

Probe Manual EL Rev D. Handling, cleaning and sterilization of the Medistim EL Imaging Probe

SAMPLE LITERATURE Please refer to included weblink for correct version.

Experience of soil tests used in two French hospitals : detergent validation

INFECTION PREVENTION PROCESSES 1

DIAGNOSTIC AUTOMATION, INC.

Cleaning Procedures. These guidelines are intended for Bulk Pharmaceutical Chemicals but almost every word applies to Drug Products

Human IL-6 ELISA Set

Sterilization & Disinfection

Mouse ALP (Alkaline Phosphatase) ELISA Kit

8/13/2013. Keys to Facilities Planning. 3M Sterile U Network 3M Sterile U Web Meeting August 15th, 2013

HBeAg and HBeAg Ab ELISA Kit

Updates to AAMI s ST79 Steam Sterilization Standard Tuesday, January 15, :00 noon 1:30 pm Eastern

Applying Cleaning Validation Principles in the Manufacture of Medical Devices and Diagnostics August 20, 2014 IVT San Diego

PREVENTING WOUND INFECTION BEGINS BEFORE THE PATIENT

Biomedical Waste Handling and Disposal

Christina Bradley Endoscope Decontamination A Webber Training Teleclass. Christina Bradley

Recommended Procedure for Sampling Gastrointestinal Flexible Endoscopes

Unloading, Storage, Distribution, Transportation and Aseptic Presentation of Sterile Items

2. Describe correct actions to take to address sterilization process failures. 3. List possible causes of sterilization process failures.

MyBioSource.com. Human VEGF ELISA Kit

MyBioSource.com. Human Total Bilirubin (TBB) Elisa kit. (Competitive ELISA)

Bovine IgG ELISA Kit

Human anti-hepatitis A virus (HAV) antibody (IgM) ELISA Kit

Microbiological Cleaning Method Validation

Draft Guidance for Industry and FDA Staff Processing/Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling

Guidelines for the use of disinfectants in hospital settings. Eleni Tomprou Infection Control Nurse, MSc General Hospital of Athens Polykliniki

BSL3 and BSL4 Autoclaves. Laboratory Line

Instructions for use

Are you on track for success?

RIGID ENDOSCOPE OPERATING INSTRUCTIONS

ATTENTION: READ THESE INSTRUCTIONS VERY CAREFULLY BEFORE YOU PREPARE AND USE THE PRODUCT FOR THE FIRST TIME.

Biosafety Protocol. Biodesign Swette Center for Environmental Biotechnology Standard Operating Procedures

PRODUCT CATALOGUE. for gke Steri-Record and gke Clean-Record Chemical indicators, PCDs and accessories for cleaning and sterilization

Infection control doctor. Mrs Muneera Al. Training supervisor. Abdulsalam. Dr Naser Yahya. Dr Haifa Al Mousa

Guidelines for the use of blood warming devices(aabb) Reference: 2002 AABB

EH&S. Sheet. Fact. Safe and Effective Use of Autoclaves. What are autoclaves? Factors for effective sterilization. Dry heat cycle - when to use

What does ECRI consider the #1 safety issue in Healthcare for 2010? Increasing Scrutiny on Endoscopes

Human BDNF ELISA. For the precise measurement of BDNF in human serum, plasma, body fluids, tissue homogenate or cell culture supernates.

Instructions for Use Reprocessed Balloon Inflation Devices

2016. All rights reserved.

PeliClass human IgG subclass ELISA kit Enzyme-linked immunosorbent assay

37 th Annual Food Safety & Sanitation Workshop Don Jones. Basic Cleaning. Basic Cleaning

I-STAT POCT System PRINCIPLE:

Transcription:

HemoCheck TM Sample Policy SUBJECT: Detection of blood residue on various surfaces DEPARTMENT: Central Service APPROVED BY: EFFECTIVE: REVISED: 07/2016 PURPOSE: To test for detection of blood residue on surfaces and to help ensure proper cleaning and reduce risk to personnel or patients 1 POLICY: The HemoCheck TM test detects blood residues on various surfaces. Random testing of various surfaces is to be done according to the manufacturer's guidelines to ensure that the cleaning process is being done properly. 2 Test for residues of blood-based test soils on: Chamber walls of automatic washers, Ultrasonic cleaner, Operating room tables, work bench, surfaces of surgical instruments (both surface and lumen area) are all areas that can be tested for blood residues. RATIONALE: "A problem analysis should be completed for any problem with any aspect of decontamination that can pose a risk to personnel or patients. The problem analysis should define and resolve the problem and the system should be monitored to ensure that the problem has been corrected.. One such problem is blood residue on any surface in the Central Service and Operating room settings. It is often difficult to distinguish between blood and other types of stains. A stain left on a surface that has been cleaned is never good, but if that stain is blood the implications can be very serious. There has been a growing concern about the effectiveness of decontamination technique for reusable medical instrumentation in healthcare facilities. Studies have shown the ability of sterilization technologies, which under normal conditions achieve acceptable sterility assurance levels, to be greatly impaired by the presence of residual soil containing serum and salt 3. Residual organic debris on processed surgical instruments is a concern and visual inspection is not a 100% accurate. 4 It is important to test any surface that is suspect of blood residue. The danger of unclean surfaces in a hospital or of handling instruments contaminated with blood is obvious in this age of hepatitis, CJD, and HIV. The procedures for 1

sterilizing instruments are based on years of scientific testing of clean instruments. If surgical instruments are not clean, the procedures are ineffective. Dried blood on instruments is hazardous to the employees of the hospital and to the next surgical patient upon which the instruments are used. Many types of soil could be present on reusable medical devices, but dried blood is especially difficult to clean. As a liquid, blood tends to flow over and into joints, hinges, grooves, and other difficult-to-clean locations. It then coagulates and dries to create a significant challenge to cleaning. 5 Blood is composed primarily of water-soluble proteins. If blood remains on an instrument after the wash phase of a mechanical cleaning equipment cycle, the water-soluble proteins will be denatured in any subsequent heated process, such as the hot disinfection phase of a washer-disinfector or washer-sterilizer cycle or a steam sterilization process. Denaturing proteins creates a soil that is much harder to clean the next time through the washing process. It is imperative that all traces of blood, body fluids, and debris be removed during the wash phase of a mechanical cleaning equipment cycle. Failure to do so could result in undetected bioburden that could pose a risk to employee health or result in a patient infection.. 6 Visual detection alone does not allow one to detect residual bioburden that may remain on cleaned devices 7 16% of the loaner instruments tested positive for blood. 8 Stains come in all sizes, shapes and intensity. Sometimes a stain can appear quite benign, or not be easily detected by the human eye. Finding a stain on an instrument which has gone through reprocessing is never good; one needs to understand what that stain is to improve their process. Technicians who are processing medical devices are looking for a fast effective test they can perform in their work space that will provide them feedback and give them assurance they are doing their reprocessing of the medical devices properly. Since the current standard is visually clean, it is assumed that you can see dirt or the stain ( bioburden) on a medical device and then re-clean it. The real goal is to detect a level low enough so one can be certain that the item tested is clean in case nothing is visually detected. A stain the size of only 1μg is visible to the naked eye on an instrument and no one would even consider using an instrument on the next patient with a stain that size. As a result the detection limit to be achieved needs to be at least around this visibility limit of 1μg. 9 2

The use of a surface detection tests like the HemoCheck is supported in the AORN Recommended Practices and Guidelines as well as AAMI ST 79. The HemoCheck is described in Annex D of ST 79 as one such test. The Hemocheck is a product that helps you verify that your process for cleaning is working properly. Some of the areas that can be tested on a medical device are: the flat surface of any medical devices, box lock (hinge) area, inside a lumen or cannulated area of a device, the distal tip area of a robotic arm, the ECRP elevator wire on the distal tip, these are just some of the areas you can test with the HemoCheck. Anywhere you suspect that a medical device might not be clean you should swab and test to ensure your process is doing what it should getting medical devices clean. Special swabs (various sizes and lengths) for these areas are available to test for residual bioburden (stains). A quality improvement system that monitors stains (residual bioburden) suspected to be blood is an important function of any Infection Control program. Testing stains with the HemoCheck TM and recording results in a log is one such system. Examples of area to test on medical devices Distal tip area of robotic arms Other Areas of a medical device 3

PROCEDURE FOR USING THE HEMOCHECK : 1. Gloves must be worn throughout the test procedure to avoid contamination of the test. Open the protective pouch of the HemoCheck test kit. Included are: A: Indicator-vial (transparent cap), B: activator-vial (green cap) and C: cotton swab. 2. Wet surfaces are swabbed with the dry cotton swab. Dry surfaces are swabbed by moistening the swab with a drop of clean water (Do not use chlorinated water!) Swab the sample area thoroughly and vigorously (on a hard surface like an instrument). Concentrate swabbing action on difficult to clean areas like joints, crevices, (signs of visible spots or discolorations). Also the ECRP elevator wire on the distal tip, any distal tip of a scope such as the air water channel (jet) or the distal tip of a robotic arm. Any areas that you think are hard to clean and should be tested for cleanliness. If a lumen is going to be checked; you swab the inner channel vigorously back and forth. Note: Swabs come in various sizes and lengths. This allows you to test varies surface and the inner lumens of many devise like a flexible scope, suction and almost any lumen medical device. 3. Open the indicator vial (A transparent cap) and transfer the liquid into the activator vial (B green cap). 4. Place the sample swab (C) into the vial (head down into the liquid) and shake at least 5 times. If using a longer type swab was used for testing lumens, cut the tip off into the vial with a clean scissor. 5. Monitor the swab for a period of 30 seconds for a color change to blue-green, which will indicate blood residues on the tested surface. In the presence of large amount of blood the whole indicator solution will change to dark blue. Record your result since a small amount of blood residue will not form a stable color. The used test liquid may change to a light blue-green after several hours, which does not indicate residues. 6. Record results of all tests on Quality form and report results to the appropriate person. 4

Note: In case of jointed surgical instruments blood residues are most common inside joints, which can be sampled with a swab. Longer narrower swabs can be used for checking inside cannulated instruments. Swab A B C A: Indicator-vial (transparent cap), B: activator-vial (green cap) and C: cotton swab. Please note the following: PRINCIPLE Due to the high content of Peroxidases in blood, an enzymatic reaction is used for detection of blood residues. MEASURING RANGE The test kit can detect 0,1 µg of blood by showing a slight blue-green spot. 1 µg of blood in the test will already give a dark blue color. INTERFERENCES Oxidising agents like chlorine or hypochlorite (present in some disinfecting agents and detergents) will also produce a color change. In this case the test cannot be used to detect blood residues. STORAGE Store HemoCheck TM in closed pouches in a cool place 2 C- 25 C. Keep away from light and heat. RESPONSIBILITY: The Central Service manager is responsible for training and for assuring initiation, completion, and analysis of the monitoring assessment activity for testing of blood residuals on various surfaces. 5

Result of the HemoCheck test, amount of blood shown from left to right: 100µg, 10µg, 1µg, 0.1µg, Blind control HemoCheck TM Test Log Record the Color Change for HemoCheck TM Result A color change to blue-green, which will indicate blood residues on the tested surface from a HemoCheck test. Test Date Tester Initials Item Tested HemoCheck TM Result Action Taken Comments 6

1 ANSI/AAMI ST 35 2003: 9.1-General Rationale;9.2 Quality Process; 9.2e Documentation of Decontamination processing parameters. 2 http://www.healthmark.info/cleaningverification/hemocheck/study_on_hemocheck_and_endocheck.pdf 3 Alfa,M.,et al, Comparison of Ion Plasma, Vaporized Hydrogen Peroxide, and 100% Ethylene oxide Sterilization to the 12/88 Ethylene oxide gas Sterilizer, Infection Control and Hospital epidemiology, 1996; 17:92-100 4 AORR Journal; July 1995,Vol62, NO1;DesCoteaux, Poulin, Julien, Guidoin 5 Page 55 ; ANSI/AAMI ST79:2010 & A1 & A2 & A3 & A4 6 Mechanical cleaning ANSI/AAMI ST79:2010/A3:2012-7.5.3.3;page 58 7 http://www.accessdata.fda.gov/scienceforums/forum06/b-48.htm 8 AORN Journal; March 2007,Volume 85,#3;page 566 9 TOSI man s Best, No. 9, Vol. 1/February 2007( www.pereg.de) 7