The Effect of MS/MS Fragment Ion Mass Accuracy on Peptide Identification in Shotgun Proteomics

Similar documents
Antibody Analysis by ESI-TOF LC/MS

Highly Confident Peptide Mapping of Protein Digests Using Agilent LC/Q TOFs

Precise Characterization of Intact Monoclonal Antibodies by the Agilent 6545XT AdvanceBio LC/Q-TOF

N-Glycan Profiling Analysis of a Monoclonal Antibody Using UHPLC/FLD/Q-TOF

Application Note. Authors. Abstract

Využití cílené proteomiky pro kontrolu falšování potravin: identifikace peptidových markerů v mase pomocí LC- Q Exactive MS/MS

Investigating Biological Variation of Liver Enzymes in Human Hepatocytes

A Unique LC-MS Assay for Host Cell Proteins(HCPs) ) in Biologics

Rapid Peptide Catabolite ID using the SCIEX Routine Biotransform Solution

Rapid Soft Spot Analysis using the SCIEX Routine Biotransform Solution

PLRP-S Polymeric Reversed-Phase Column for LC/MS Separation of mabs and ADC

Proteomics. Proteomics is the study of all proteins within organism. Challenges

LC/MS/MS Solutions for Biomarker Discovery QSTAR. Elite Hybrid LC/MS/MS System. More performance, more reliability, more answers

A Complete Workflow Solution for Intact Monoclonal Antibody Characterization Using a New High-Performance Benchtop Quadrupole- Orbitrap LC-MS/MS

Peptide Mapping of Glycoprotein Erythropoietin by HILIC LC/MS and RP-LC/MS

Enabling Systems Biology Driven Proteome Wide Quantitation of Mycobacterium Tuberculosis

AB SCIEX TripleTOF 5600 SYSTEM. High-resolution quant and qual

Characterize Fab and Fc Fragments by Cation-Exchange Chromatography

Identification of Microprotein-Protein Interactions via APEX Tagging

Agilent 6430 Triple Quadrupole LC/MS System

Metabolic Changes in Lung Tissue of Tuberculosis-Infected Mice Using GC/Q-TOF with Low Energy EI

High Resolution Accurate Mass Peptide Quantitation on Thermo Scientific Q Exactive Mass Spectrometers. The world leader in serving science

LTQ Orbitrap XL Hybrid FT Mass Spectrometer Unrivaled Performance and Flexibility

Thermo Scientific Peptide Mapping Workflows. Upgrade Your Maps. Fast, confident and more reliable peptide mapping.

Proteomics And Cancer Biomarker Discovery. Dr. Zahid Khan Institute of chemical Sciences (ICS) University of Peshawar. Overview. Cancer.

Increasing MS Sensitivity for Proteomics

Rapid Peptide Mapping via Automated Integration of On-line Digestion, Separation and Mass Spectrometry for the Analysis of Therapeutic Proteins

Detecting Challenging Post Translational Modifications (PTMs) using CESI-MS

Thermo Scientific Solutions for Intact-Protein Analysis. Better, Faster Decisions for. Biotherapeutic Development

The clearly better choice

Advanced QA/QC characterization MS in QC : Multi Attribute Method

Fast and High-Resolution Reversed-Phase Separation of Synthetic Oligonucleotides

CaptiveSpray nanobooster

Advantages of Ion Mobility QTOF for Characterization of BioPharma Molecules

TechnicalNOTE METABOLYNX EXACT MASS DEFECT FILTER: RAPIDLY DISCRIMINATE BETWEEN TRUE METABOLITES AND FALSE POSITIVES INTRODUCTION

Monitoring for Microcystins in Raw Water Supply Reservoirs Using the Agilent 6410 Triple Quadrupole LC/MS

Proteomics: A Challenge for Technology and Information Science. What is proteomics?

rapiflex Innovation with Integrity Designed for Molecules that Matter. MALDI TOF/TOF

Key Words Q Exactive Focus, SIEVE Software, Biomarker, Discovery, Metabolomics

Xevo G2-S QTof and TransOmics: A Multi-Omics System for the Differential LC/MS Analysis of Proteins, Metabolites, and Lipids

New Sample Fractionation Strategies for Proteomic Analyses by LC MS

Confident Pesticides Analysis with the Agilent LC/Triple Quadrupole and TOF/QTOF Solutions

HepG2 Labeling and Cell Growth Protocol for IROA Phenotypic Metabolic Profiling

Utilizing novel technology for the analysis of therapeutic antibodies and host cell protein contamination

Monoclonal Antibody Characterization on Q Exactive and Oribtrap Elite. Yi Zhang, Ph.D Senior Proteomic Marketing Specialist Oct.

Pharmaceutical LC/MS Solutions from Agilent Technologies

Cell Clone Selection Using the Agilent Bio-Monolith Protein A Column and LC/MS

Puzzled About LCMS? Sample. Sensitivity in Mass Spec Analysis. Prep. Adapting LC-UV. Optimizing and Maintaining Your Mass Spec LCMS

Genotoxicity is the property of a compound

EPA Method 543: Selected Organic Contaminants by Online SPE LC/MS/MS Using the Agilent Flexible Cube

Generating Automated and Efficient LC/MS Peptide Mapping Results with the Biopharmaceutical Platform Solution with UNIFI

Improving Sensitivity in Bioanalysis using Trap-and-Elute MicroLC-MS

New Stringent Two-Color Gene Expression Workflow Enables More Accurate and Reproducible Microarray Data

Application Note. Author. Abstract. Biotherapeutics and Biologics. Sonja Schneider Agilent Technologies, Inc. Waldbronn, Germany

Application Note. Abstract. Authors. Introduction. Environmental

Modification Site Localization Scoring Integrated into a Search Engine

About OMICS Group Conferences

Application Note. Kwasi Antwi, Amanda Ribar, Urban A. Kiernan, and Eric E. Niederkofler Thermo Fisher Scientific, Tempe, Arizona

Exploring Extra Sensitivity Using ionkey/ms with the Xevo G2-XS Q-Tof HRMS for Small Molecule Pharmaceutical Analysis in Human Plasma

Featuring Analyst software under Windows NT for enhanced performance and ease of use. API 150EX. LC/MS System

How to view Results with Scaffold. Proteomics Shared Resource

Algorithm for Matching Additional Spectra

Proudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment QTRAP 5500

Heart-cut 2D-LC/MS approach for pharmaceutical impurity identification using an Agilent 6540 Q-TOF LC/MS System

Rapid Extraction of Therapeutic Oligonucleotides from Primary Tissues for LC/ MS Analysis Using Clarity OTX, an Oligonucleotide Extraction Cartridge

Thermo Scientific Mass Spectrometric Immunoassay (MSIA) Pipette Tips. Next generation immunoaffinity. Robust quantitative platform

An Unbiased Approach to Increasing Proteome Coverage Using Ion Mobility with MS E

N- The rank of the specified protein relative to all other proteins in the list of detected proteins.

How to view Results with. Proteomics Shared Resource

Application Note. Authors. Abstract. Clinical Research

Application of Agilent AdvanceBio Desalting-RP Cartridges for LC/MS Analysis of mabs A One- and Two-dimensional LC/MS Study

Center for Mass Spectrometry and Proteomics Phone (612) (612)

Low pg/ml Detection of rhuepos in Horse Plasma Employing High-Resolution MS

Two-Dimensional LC Protein Separation on Monolithic Columns in a Fully Automated Workflow

MasterView Software. Turn data into answers reliable data processing made easy ONE TOUCH PRODUCTIVITY

HDX-MS at Waters Waters Corporation 1

Method Development Considerations for Reversed-Phase Protein Separations

Investigation of a Mammalian Cellular Model for Differential Protein Expression Analysis Using 1D PAGE and Cleavable ICAT Reagents

Method Translation in Liquid Chromatography

Agilent 6500 Series Q-TOF LC/MS System

Proteomics Technology Note

ICP Expert software. Technical Overview. Introduction

MetaboScape 2.0. Innovation with Integrity. Quickly Discover Metabolite Biomarkers and Use Pathway Mapping to Set them in a Biological Context

The Five Key Elements of a Successful Metabolomics Study

Thermo Scientific TSQ Quantum Access MAX

MassPREP On-Line Desalting Cartridge

MBios 478: Mass Spectrometry Applications [Dr. Wyrick] Slide #1. Lecture 25: Mass Spectrometry Applications

Agilent solutions for contract research and manufacturing organizations

Improved Peptide Maps Using Core-Shell Media: Explaining Better Biopharmaceutical Applications With Kinetex C18 Columns

Automated Online SPE for LC/MS/MS Analysis of Trace Organic Contaminants in Water Using the Agilent 1290 Infinity Flexible Cube Module

Characterization of Glycosylation in the Fc Region of Therapeutic Recombinant Monoclonal Antibody

Tony Mire-Sluis Vice President, Corporate, Product and Device Quality Amgen Inc

Agilent 6400 Series. System. Concepts Guide. The Big Picture. Agilent Technologies

Application Note. Author. Abstract. Biopharmaceuticals. Verified for Agilent 1260 Infinity II LC Bio-inert System. Sonja Schneider

PAVING THE WAY FOR ASSESSING IN VIVO DYNAMICS OF MULTIPLE QUALITY ATTRIBUTES FOR PROTEIN THERAPEUTICS

ProSEC 300S. Protein Characterization columns

Simultaneous in vivo Quantification and Metabolite Identification of Plasma Samples Using High Resolution QTof and Routine MS E Data Analysis

UC San Diego UC San Diego Electronic Theses and Dissertations

HIGHER IMPACT LC/MS DIAGNOSTICS AGILENT IN VITRO DIAGNOSTIC MEDICAL DEVICES FOR CLINICAL LABORATORIES

Transcription:

The Effect of MS/MS Fragment Ion Mass Accuracy on Peptide Identification in Shotgun Proteomics Authors David M. Horn Christine A. Miller Bryan D. Miller Agilent Technologies Santa Clara, CA Abstract It is widely accepted that high precursor ion mass measurement accuracy significantly improves confidence in MS/MS database search results through the reduction of false positives. In contrast, the impact of high fragment ion mass accuracy has been largely ignored. The Agilent Accurate Mass Q-TOF, a recent advancement in MS instrumentation, now enables high-throughput accurate mass measurement of both precursor and fragment ion data. This study examines the effects of precursor and fragment ion mass accuracy on protein identification using the Accurate Mass Q-TOF. Results demonstrate that fragment ion mass accuracy is as important (if not more important) than precursor ion mass accuracy in reducing false positive identification rates in MS/MS database searches, and plays a critical role in improving the throughput of proteomics research projects. Introduction Perhaps the most fundamental goal of proteomics research is the identification and characterization of proteins, particularly in the context of cellular expression. The shotgun technology of tandem mass spectrometry paired with advanced liquid chromatography has emerged as the standard technique for high-throughput protein identification (Liu, J. et al., 2007; VerBerkmoes, N.C. et al., 2002). To perform a typical shotgun experiment, a sample is first fractionated and the resulting mixture of proteins is then digested into peptides by an enzyme such as trypsin. The peptide mixture is then separated by high performance liquid chromatography (HPLC) and subsequently analyzed by mass spectrometry to determine the mass/charge (m/z) ratio of each peptide. Peptides of interest are selected for further fragmentation in a collision cell to produce tandem (MS/MS) mass spectra. This resultant data is then searched against a protein database to identify the peptide sequences and further infer the protein content of the sample.

It is well known that high mass accuracy measurement of precursor ions significantly improves confidence in MS/MS protein database search results. However, the impact of high mass accuracy measurement of fragment ions has been largely ignored. Up until recently, MS instruments that possessed sufficient throughput capability to capture both precursor and fragment ion data with high mass accuracy were not readily available. The advent of the Agilent Accurate Mass Q-TOF LC/MS provides the sensitivity, speed, and accurate mass capabilities to enable the assessment of precursor and fragment ion mass tolerance as the maximum mass deviation, or MMD (Zubarev, R. and Mann, M., 2007). The Accurate Mass Q-TOF LC/MS quickly captures information from complex samples with attomole-level sensitivity, a wide in-scan dynamic range that covers 3.5 orders of magnitude, and a fast acquisition rate (20 spectra/sec) at up to 15,000 resolving power. Here, we apply the Q-TOF LC/MS system and demonstrate that fragment ion mass accuracy can be at least as significant as precursor ion mass accuracy in reducing false positive identification rates in MS/MS protein database searches. In addition, we compare false positive rates generated by the mass tolerance parameters particular to other instrument types used for protein identification research. Materials and Methods The experimental workflow for this analysis used a 3100 OFFGEL Fractionator to fractionate the HeLa sample, a 1200 HPLC-Chip/6510 Q-TOF LC/MS to acquire the data, and Spectrum Mill Protein Identification software to perform analysis of the collected data and assess the MMD for HeLa cells (Figure 1). HeLa Cell Culture HeLa S3 cells were grown in Ham s F12 medium (F12K) with 2 mm L-glutamine and 1.5 g/l sodium bicarbonate supplemented with 10% fetal bovine serum until 90% confluency. Cells were washed in PBS, lysed in hypotonic solution containing protease inhibitors cocktail, then homogenized with a Dounce homogenizer. Soluble HeLa lysate was prepared by centrifugation (1 hr at 16,000 rpm at 4 C). An aliquot (~300 µg total protein) of the supernatant was digested with trypsin using a 2,2,2-trifluoroethanol (TFE) based protocol for solubilization and denaturation. The digest was dried to remove the TFE and ammonium bicarbonate. OFFGEL Fractionation The trypsinized lysate was fractionated into 23 fractions in the ph range from 3 to 10. The current was limited to 50 µa and fractionation was stopped after 50 kvh (about 24 hr). LC/MS Analysis Approximately 5% of each fraction (5 µl) was analyzed using a microfluidics based HPLC-Chip connected to an Agilent 6510 Quadrupole Time-of-Flight (Q-TOF) mass spectrometer or an Agilent 6330 Ion Trap mass spectrometer. HPLC-Chip: Protein ID chip with 150 x 0.075 mm analytical column and 160 nl enrichment column. Sample load: 5 µl of fractions obtained from OFFGEL. Flow: 300 nl/min analytical pump, 4 µl/min loading pump. Mobile phases A: 0.1% formic acid (FA), B: 90% acetonitrile, 0.1% FA. Gradients: 3%B to 6%B at 2 min, then 10% B at 10 min, 30%B at 65 min, 40%B at 75 min, 80%B at 80 min until 85 min, then 3%B at 85.1 min. Stoptime: 90 min and post time 10 min. Ion Trap MS Conditions: Drying gas: 4 L/min, 300 C; Skim 1: 30 V; Capillary exit: 75 V; Trap Drive: 85; Averages: 1; ICC: On; Max. Accumulation time: 50 ms; Smart Target: 500,000; MS Scan range: 300-1800; Ultra Scan; MS/MS Scan range: 100-2000; Ultra Scan. Q-TOF MS Conditions: Drying gas: 4 L/min, 300 C; Skimmer: 65 V; Fragmentor: 175 V; Collision energy: slope 3.7 V, offset 2.5 V; MS Scan range and rate: 300-2000 at 3 Hz; MS/MS scan range and rate: 50-3000 at 3Hz. AutoMS/MS: 8 precursors, active exclusion on with 1 repeat and release after 0.17 min. Preferred charge state: 2, 3, >3, unknown. Database Searches Protein database searches were performed with Spectrum Mill Protein Identification software. All searches used the IPI Human Version 3.28 database with trypsin specificity, 2 missed cleavages, 50% minimum scored peak intensity, and dynamic peak thresholding. The forward database search of the ion trap data was performed using 2.5 Da precursor and 0.7 Da fragment ion tolerance while the Q-TOF search used 10 ppm precursor and 40 ppm fragment ion tolerance. Protein identifications were validated manually. A reversed version of 2 This item is intended for Research Use Only. Not for use in diagnostic procedures.

The Effect of MS/MS Fragment Ion Mass Accuracy on Peptide Identification in Shotgun Proteomics the IPI Human database was created using a Perl script. Searches against this reversed database were performed using a variety of different precursor and product ion mass tolerances. The false positive rates for a given Spectrum Mill score and MMD regime were estimated by comparison of the distribution of Spectrum Mill scores for the forward Q-TOF search and reversed database search results. All results were exported to Microsoft Excel and database search scores were divided into 0.5 unit bins and plotted. Results and Discussion Selection of MMD values for Q-TOF data As a first step in this study, it was necessary to determine the optimal mass tolerances of the 6510 Q-TOF to use for the subsequent protein database search. Using various precursor and product ion mass tolerances, a series of forward database searches were performed. The search results that produced the largest number of validated protein identifications used 10 ppm precursor ion and 40 ppm fragment ion tolerance. Figure 1. The experimental workflow consisting of a 3100 OFFGEL Fractionator, a 1200 HPLC-Chip/6510 Q-TOF LC/MS, and Spectrum Mill software for data analysis. As part of this study, the performance of the Q-TOF was compared with the 6330 Ion Trap mass spectrometer. www.proteomics-lab.com 3

Figure 2 shows the distribution of precursor and fragment ion mass errors for the validated peptides. These results indicate that fragment ion mass tolerance displays a wider range than that of the precursor due to significantly reduced signal-tonoise ratios. However, the higher abundance fragments demonstrate similarly high mass accuracy and improvements to the MS/MS search algorithm could take this into account to improve the results. Comparison of ion trap and Q-TOF results Having established the optimal mass tolerance values for the Q-TOF, we next compared the performance of the Q-TOF with the 6330 Ion Trap mass spectrometer, an instrument that is commonly used for protein identification research. Both instruments analyzed aliquots of the same HeLa sample. Table 1 displays the results generated from both instruments. In the ion trap, there were a number of very high scoring random protein hits. Further, the number of spectra with random peptide identifications was more than 10-fold for the ion trap compared to the Q-TOF system. This is presumably partly due to the higher precursor MMD of the ion trap, but also potentially due to the effect of high fragment ion tolerance as well. Also, the total number of MS/MS acquired on the Q-TOF was lower by 90,000 spectra, largely because the high resolution of the Q-TOF better distinguished real peptide signal from background noise. Finally, a greater number of validated peptide identifications were found in the Q-TOF data. Precursor Fragment -20-16 -12-8 -4 0 4 8 12 16 20-50 -40-30 -20-10 0 10 20 30 40 50 ppm Error ppm Error Figure 2. Mass measurement deviation distributions for precursor (3.9 rms ppm) and fragment ions (13.8 rms ppm) using the 6510 Q-TOF. 4 This item is intended for Research Use Only. Not for use in diagnostic procedures.

The Effect of MS/MS Fragment Ion Mass Accuracy on Peptide Identification in Shotgun Proteomics This combination of factors led to a significantly higher identification yield for the Q-TOF compared to the ion trap. Besides the higher mass accuracy, other factors that might have contributed to the high number of valid spectra generated by the Q-TOF include the ability to assign charge state for the fragment ions and the general cleanliness of the Q-TOF MS/MS data, especially when compared to the ion trap data for MS/MS acquired and the number of extracted files. Regardless, the reduction in matches to random proteins exhibited by the data obtained from the Q-TOF enables a vast improvement in confidence related to peptide identification. Precursor and fragment ion mass tolerance Based on the results of these preliminary studies indicating the significance of mass accuracy on the ability to identify peptides, we next proceeded to determine the contributing effects of precursor and fragment ion mass tolerances on confident protein identification. Past research has indicated that high stringency in the precursor ion tolerance can dramatically reduce the rate of false positive identifications for a given search (Liu, T. et al., 2007). To demonstrate this effect, a series of reversed database searches were performed holding the fragment ion mass tolerance constant at 40 ppm (the established threshold for the Q-TOF data obtained in this Total number spectra acquired MS/MS acquired Number of extracted files (.pkl) Number of search result files (.spo) Number of validated spectra Number of validated unique peptides Number of protein IDs Number of unvalidated spectra Ion Trap 355,136 271,507 92,081 72,028 10,618 4151 873 61,410 Q-TOF 363,692 186,049 129,752 20,871 16,072 6253 994 4799 Table 1. Comparison between 6330 Ion Trap and 6510 Q-TOF search results and hits. www.proteomics-lab.com 5

study) while varying the precursor ion tolerance. It is known that reversed database searches are a good method for determining the false positive rate, because searching a reversed database would generate a number of matches that would be known to be false (Elias, J.E. and Gygi, S.P., 2007). As precursor ion tolerance decreased below the fragment ion tolerance of 40 ppm, the false positive rate decreased as expected (Figure 3). Interestingly, as the precursor ion tolerance was increased above the fragment ion tolerance, there was little change from 50 to 200 ppm and only a slight increase with 500 and 1000 ppm. The increase in the number of false positives for 500 and 1000 ppm may be attributed to matches of peptides with 1 Da higher and lower nominal mass. These results suggested that while precursor ion mass tolerance most certainly does affect the false positive rate in database searching, the effect of a low fragment ion mass tolerance is also an important factor in limiting the false positive rate. To determine the significance of low fragment ion mass tolerance on false positive rates, a series of reversed database searches were performed holding the precursor mass 1200 Number of False Positive Matches 1000 800 600 400 200 1000 ppm 500 ppm 20 ppm 10 ppm 5 ppm 200 ppm 100 ppm 50 ppm 0 0 2 4 6 8 10 12 14 Spectrum Mill Reversed DB Search Score Figure 3. Spectrum Mill score distributions of reversed database hits for variable precursor tolerance at constant 40 ppm fragment tolerance using the 6510 Q-TOF. 6 This item is intended for Research Use Only. Not for use in diagnostic procedures.

The Effect of MS/MS Fragment Ion Mass Accuracy on Peptide Identification in Shotgun Proteomics tolerance at 10 ppm and varying the fragment mass tolerance (Figure 4). In contrast to results obtained when the fragment mass tolerance was held constant, the results of this experiment showed that the number of false positives continued to increase as the fragment ion mass tolerance increased. Importantly, the number of false positive matches decreased quite dramatically as the fragment ion mass tolerance was decreased, indicating that high precursor ion mass accuracy is not sufficient to mitigate the effect of large fragment mass measurement errors. We next examined the effects of mass accuracy on database search efficacy more closely by performing reversed database searches with 6510 Q-TOF data using different sets of precursor and fragment mass accuracy settings that represent three different MS instruments an ion trap, a high-resolution 4000 1000 ppm Number of False Positive Matches 3500 3000 2500 2000 1500 1000 500 500 ppm 200 ppm 100 ppm 50 ppm 20 ppm 10 ppm 5 ppm 0 0 2 4 6 8 10 12 14 Spectrum Mill Reversed DB Search Score Figure 4. Spectrum Mill score distributions of reversed database hits for variable fragment tolerance at constant 10 ppm precursor tolerance using the 6510 Q-TOF. www.proteomics-lab.com 7

hybrid linear ion trap, and the Q-TOF itself (Figure 5). As expected, when the precursor ion mass tolerance was improved from 1000 ppm (ion trap) to 5 ppm (high-resolution hybrid linear ion trap), the distribution of false positive matches decreased proportionally as well. However, when the fragment tolerance was lowered from 1000 ppm (ion trap & high-resolution hybrid linear ion trap) to 40 ppm (Q-TOF), this led to a substantial decrease in the score distribution of random matches. These results indicate that the decrease in false positive identification was attributable to not just the precursor ion mass tolerance but also just as strongly to the product ion mass tolerance. In the final part of this study, we examined how both the precursor and fragment ion mass tolerances could ultimately affect confidence in protein identification. Using the Q-TOF data generated at 10 ppm precursor MMD and 40 ppm fragment MMD, the reverse score distributions and the Number of False Positive Matches 8000 7000 6000 5000 4000 3000 2000 1000 0 0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 Spectrum Mill Reversed DB Search Score 10 ppm Precursor, 40 ppm Fragment 5 ppm Precursor, 1000 ppm Fragment 1000 ppm Precursor, 1000 ppm Fragment Figure 5. Spectrum Mill score distributions of reversed database hits using common precursor and fragment mass accuracy parameters on the 6510 Q-TOF. 8 This item is intended for Research Use Only. Not for use in diagnostic procedures.

The Effect of MS/MS Fragment Ion Mass Accuracy on Peptide Identification in Shotgun Proteomics validated protein identifications from the forward database search were compared (Figure 6). A false positive rate was estimated for each scoring bin by comparing the number of random matches in the reversed database search to the number of confidently identified peptide matches. The results indicated that high mass accuracy led to a downward shift in the score distribution of the reversed database matches relative to the forward search. As a result, the Spectrum Mill score required for a particular confidence level was lowered, leading to an increase in the number of confident hits. In fact, the number of confident matches increased up to 6.5-fold with the use of both highly accurate precursor and fragment ion mass measurements on the 6510 Q-TOF (Table 2). 800 50% 75% 95% Number of Matches 700 600 500 400 300 200 100 0 10 ppm Precursor, 40 ppm Fragment reversed database search Q-TOF Forward Database Search (10 ppm Precursor, 40 ppm Fragment) 0 2 4 6 8 10 12 14 16 18 20 22 24 26 28 30 Spectrum Mill DB Search Score Figure 6. Calculation of peptide identification confidence by comparison of forward and reversed search results of 6510 Q-TOF data. The Spectrum Mill score required for a particular confidence level (denoted at the top of the chart) was lowered, leading to an increase in the number of confident hits. www.proteomics-lab.com 9

Table 2 also shows the number of protein matches with 2 or more unique peptide identifications with at least 95%, 75%, or 50% confidence. Using both high-accuracy precursor and fragment ion mass measurements, the 6510 Q-TOF demonstrated roughly a 5-fold improvement in the number of proteins identified with confidence (95% confidence interval). As a result, a much larger proportion of sample identifications can be automatically validated, greatly increasing the automated throughput of proteomics experiments. Conclusions The results of this study indicate that fragment ion mass accuracy is an important component to accurate protein identification. Overall, fragment ion mass accuracy was found to render a significant impact on false positive rates. When the fragment tolerance was lowered from 1000 ppm to 40 ppm, a substantial decrease in the score distribution of random matches in database searches was realized. The positive effect of low fragment mass tolerance has at least two explanations. First, large fragments generated by MS/MS of a peptide will exhibit a similar distribution of masses for a given nominal mass as the precursor. Thus, lowering the fragment tolerance will have a similar effect as lowering the precursor tolerance for these fragments, with the result of limiting the distribution of potential database matches. Second, a wider fragment mass tolerance will greatly increase the chance that random peaks in the MS/MS spectrum will be matched against random entries in the database. Decreasing the fragment tolerance will reduce the chance of such random matches. These data demonstrate that high fragment ion mass accuracy can significantly increase valid protein identification by reducing the number of false positives when searching large databases for proteins, and consequently decreasing the score required for a high-confidence match. Other protein database search engines, such as Mascot, that can accommodate accurate mass data would be expected to yield similar results. In addition, these data demonstrate that the Agilent Accurate Mass Q-TOF LC/MS provides high mass accuracy for both the precursor and product ions, while generating MS/MS spectra at a rate that is well matched to complex sample LC/MS, unlike other high-resolution hybrid linear ion trap instruments. ppm Tolerance Number of Valid Peptide Matches with % Confidence Number of Valid Protein Matches with % Confidence (minimum of 2 or more unique peptides/protein) Precursor Fragment >95% >75% >50% >95% >75% >50% 1000 1000 954 2220 3361 102 213 282 5 1000 4181 6590 7645 338 478 524 10 40 6590 11552 15509 478 728 829 Table 2. The number of confident hits for three different combinations of precursor and fragment mass accuracy settings on the 6510 Q-TOF. 10 This item is intended for Research Use Only. Not for use in diagnostic procedures.

The Effect of MS/MS Fragment Ion Mass Accuracy on Peptide Identification in Shotgun Proteomics References 1. Elias, J.E., S.P. Gygi (2007). Target-decoy search strategy for increased confidence in large-scale protein identifications by mass spectrometry. Nat Methods, 4(3):207-14. 2. Liu, J., A.W. Bell, J.J. Bergeron, C.M. Yanofsky, B. Carrillo, C.E. Beaudrie, R.E. Kearney (2007) Methods for peptide identification by spectral comparison. Proteome Sci, 5:3. 3. Liu, T., M.E. Belov, N. Jaitly, W.J. Qian, R.D. Smith (2007) Accurate Mass Measurements in Proteomics. Chem. Rev., 107:3621-3653. 4. VerBerkmoes, N.C., J.L. Bundy, J.L. Hause, K.G. Asano, J. Razumovskaya, F. Larimer, R.L. Hettich and J.L. Stephenson (2002) Integrating top-down and bottom-up mass spectrometric approaches for proteomic analysis of Shewanella oneidensis. J of Proteom Res, 1(3):239-252. 6. Zubarev, R., and M. Mann (2007) On the proper use of mass accuracy in proteomics. Mol Cell Proteomics, 6(3):377-81. www.proteomics-lab.com 11

About Agilent Technologies Agilent Technologies is a leading supplier of life science research systems that enable scientists to understand complex biological processes, determine disease mechanisms, and speed drug discovery. Engineered for sensitivity, reproducibility, and workflow productivity, Agilent s life science solutions include instrumentation, microfluidics, software, microarrays, consumables, and services for genomics, proteomics, and metabolomics applications. Buy online: www.agilent.com/chem/store Find an Agilent customer center in your country: www.agilent.com/chem/contactus U.S. and Canada 1-800-227-9770 agilent_inquiries@agilent.com Asia Pacific adinquiry_aplsca@agilent.com Europe marcom_center@agilent.com This information contained in this publication is intended for Research Use Only and not to be followed as a diagnostic procedure. Information, descriptions, and specifications in this publication are subject to change without notice. Agilent Technologies shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance or use of this material. Agilent Technologies, Inc. 2008 Printed in the U.S.A. January 10, 2008 5989-7820EN