NEMA Standards Publication NEMA/MITA 1. Good Refurbishment Practices for Medical Imaging Equipment

Similar documents
NEMA Standards Publication NEMA/MITA 1. Good Refurbishment Practices for Medical Imaging Equipment

ANSI C (R2002, R2012) Reaffirmation of ANSI C (R2002) American National Standard. Thermal Demand Meters

ICEA STANDARD FOR DISTRIBUTION FRAME-WIRE TECHNICAL REQUIREMENTS. Published By INSULATED CABLE ENGINEERS ASSOCIATION, INC. Post Office Box 1568

ISO/IEC INTERNATIONAL STANDARD. Systems and software engineering System life cycle processes IEEE

AWS D14.3/D14.3M:2010 An American National Standard. Specification for Welding Earthmoving, Construction, and Agricultural Equipment

PROJECT MANAGEMENT BODY OF KNOWLEDGE

ISO 22000:2005 SYSTEMKARAN ADVISER & INFORMATION CENTER SYSTEM KARAN ADVISER & INFORMATION CENTER FOOD SAFETY MANAGEMENT SYSTEM ISO 22000:2005

IECQ PUBLICATION QC IEC Quality Assessment System for Electronic Components (IECQ System)

Specification for Quality Programs for the Petroleum, Petrochemical and Natural Gas Industry (Draft 10)

CORPORATE QUALITY MANUAL

INTERNATIONAL STANDARD

ISO /TS 29001:2010 SYSTEMKARAN ADVISER & INFORMATION CENTER SYSTEM KARAN ADVISER & INFORMATION CENTER

Association of American Railroads Quality Assurance System Evaluation (QASE) Checklist Rev. 1/12/2017

CONSOLIDATED VERSION IEC Medical device software Software life cycle processes. colour inside. Edition

Specification for Quality Programs for the Petroleum, Petrochemical and Natural Gas Industry

EVANS CAPACITOR COMPANY

CEN Keymark Scheme Rules for Thermostatic Radiator Valves. 2 nd Edition

ISO/TS TECHNICAL SPECIFICATION

Quality Management System MANUAL. SDIX, LLC Headquarters: 111 Pencader Drive Newark, Delaware 19702

FINAL DOCUMENT. International Medical Device Regulators Forum. Medical Device Regulatory Audit Reports

HP Standard Supplier Requirements for Safe and Legal Products

Specification for Quality Programs for the Petroleum, Petrochemical and Natural Gas Industry

AEROSPACE STANDARD. Quality Systems - Aerospace - Model for Quality Assurance in Design, Development, Production, Installation and Servicing

MALAYSIAN STANDARD QUALITY MANAGEMENT SYSTEMS - REQUIREMENTS (FIRST REVISION) (ISO 9001:2008, IDT) (PUBLISHED BY STANDARDS MALAYSIA IN 2009)

Reliance Aerospace Solutions

SHELTON INDUSTRIAL PATTERN INC.

AWS D8.2M:2017 An American National Standard. Specification for Automotive Weld Quality Resistance Spot Welding of Aluminum

ISO/IEC INTERNATIONAL STANDARD. Conformity assessment General requirements for accreditation bodies accrediting conformity assessment bodies

FDA 21 CFR Part 820 vs. ISO 13485:2016 Comparison Table created by greenlight.guru

ISO INTERNATIONAL STANDARD. Clinical investigation of medical devices for human subjects Good clinical practice

QUALITY SYSTEM MANUAL

TECHNICAL DIRECTIVE PART ONE: STANDARDIZATION AND RELATED ACTIVITIES GENERAL VOCABULARY

This is a preview - click here to buy the full publication GUIDE 67. Conformity assessment Fundamentals of product certification. First edition 2004

ASQ Tappan Zee Section Medical Devices, New Regulations and Standards. 26 th September / V. Fischer / Rev. 01

Specification for Quality Programs for the Petroleum, Petrochemical and Natural Gas Industry

QUALITY MANUAL DISTRIBUTION. Your Logo Here. Page 1 1 of 57. Rev.: A

ISO INTERNATIONAL STANDARD

IMDRF. Final Document. Regulatory Authority Assessor Competence and Training Requirements. IMDRF MDSAP Work Group

D8.14M:2017 An American National Standard. Specification for Automotive Weld Quality Arc Welding of Aluminum

General user conditions for supplier s applications and the AUMA supplier portal

QM-1 QUALITY MANAGEMENT SYSTEMS MANUAL. Revision 10

Quality Manual ISO Cedarburg, Wisconsin Revision: L

Medical Device Regulatory Framework 9 SEPTEMBER 2015 FUNDISA CONFERENCE JANE ROGERS

ISO INTERNATIONAL STANDARD

Expert Commentary on BS EN ISO 13485:2016, Medical devices Quality management systems Requirements for regulatory purposes

INTERNATIONAL STANDARD

ISO/IEC INTERNATIONAL STANDARD. Conformity assessment General requirements for third-party marks of conformity

PROForms. Specifications SECTION FIBERGLASS REINFORCED POLYMER (FRP) PULTRUDED PRODUCTS

Medical Device Purchasing Controls Challenges of Compliance in a World Market June, 2013 OMTEC

APS Cleaning Quality Management System Scope of Certification The provision of commercial and industrial cleaning services throughout Queensland.

ISO INTERNATIONAL STANDARD. Quality requirements for fusion welding of metallic materials Part 3: Standard quality requirements

Appendix QI. Intra-Lockheed Martin Work Transfer Agreement (IWTA) Quality Requirements for Deliverable Items

ISO 28002: RESILIENCE IN THE SUPPLY CHAIN: REQUIREMENTS WITH GUIDANCE FOR USE

AAMI Quality Systems White Paper: Comparison of 21 CFR Part 820 to ISO 13485:2016 1

EA-7/04 Legal Compliance as a part of accredited ISO 14001: 2004 certification

TERMS AND DEFINITIONS

Luminus Devices, Inc Quality Management Systems Manual ISO 9001:2008

ISO INTERNATIONAL STANDARD. Cleanrooms and associated controlled environments Part 4: Design, construction and start-up

Document: ISO/TC 176/SC 2/N 1147

0. 0 TABLE OF CONTENTS

4. RELATED DOCUMENTS AND FORMS 4.1. AS ISO Cambridge Valley Machining, Inc. QMS procedures, including, but not limited to:

DRAFT MEDICAL DEVICE GUIDANCE DOCUMENT

Supplier Quality Requirements Q Clauses

Quality Manual. Specification No.: Q Revision 07 Page 1 of 14

ISO 22000:2005 Standard INTERNATIONAL STANDARDS REGISTRATIONS

AWS D9.1/D9.1M:2018 An American National Standard. Sheet Metal Welding Code

OVERVIEW OF THE PREQUALIFICATION OF MALE CIRCUMCISION DEVICES ASSESSMENT PROCESS

HACCP audit checklist

SQF 2000 Guidance. Guidance for Food Sector Category 4 Fresh Produce Pack House Operations. 1st Edition

QUALITY SYSTEM MANUAL

MEDICAL DEVICE CLINICAL INVESTIGATIONS AND ISO 14155

QUALITY MANAGEMENT SYSTEM POLICIES AND PROCEDURES

Supplier Quality Manual (QQS )

Medical Devices. The technical requirements also apply when the submitter is not the legal manufacturer 1 (i.e: a distribution company).

14620 Henry Road Houston, Texas PH: FX: WEB: QUALITY MANUAL

Supplier Quality Manual

ISO 9001 QUALITY MANUAL

ANSI/NEMA WC 71 ICEA S Nonshielded Cables Rated V for Use in the Distribution of Electric Energy

QUALITY MANAGEMENT SYSTEM MANUAL ISO 9001:2008

A REVIEW OF MARKET ENTRY REQUIREMENTS FOR RISK MANAGEMENT, WITH SPECIAL EMPHASIS ON FDA AND ISO COMPLIANCE

Regulations governing the application of medical accelerators

Technical Manufacturing Corporation (TMC)

FDA s Final Rule on Medical Device Data Systems Signals Active Regulation of Data Communication Technologies

Product safety and conformity in the automotive supply chain in the case of product nonconformities 1 st Edition, February 2018 Online-Download-Docume

HACCPEUROPA PUBLICATIONS ISO 22000:2005 FOOD SAFETY QUALITY MANUAL. ISO 22000:2005 Quality Manual

Revision. Quality Manual. Multilayer Prototypes. Compliant to ISO / AS9100 Rev C

ISO/IEC TR TECHNICAL REPORT

ISO/TS TECHNICAL SPECIFICATION

Osprey Technologies, LLC. Quality Manual ISO9001:2008 Rev -

25 D.L. Martin Drive Mercersburg, PA (717)

ISO INTERNATIONAL STANDARD

WHO Prequalification of In Vitro Diagnostics Programme

SKYCOAT, LLC QUALITY CONTROL MANUAL

MDSAP AUDIT PROCESS. A Manufacturer s Perspective. Connie Hoy EVP Regulatory Affairs Cynosure, Inc.

EPICOR, INCORPORATED QUALITY ASSURANCE MANUAL

Space product assurance

GE Healthcare. GoldSeal. Refurbished imaging systems. Reliable quality. Certified confidence.

Document Number: QM001 Page 1 of 19. Rev Date: 10/16/2009 Rev Num: 1. Quality Manual. Quality Manual. Controlled Copy

OPAC EN Operating Procedure for the Attestation of Conformity of Structural Bearings in compliance with Annex ZA of EN 1337/3/4/5/6/7

INTERNATIONAL STANDARD

Transcription:

NEMA Standards Publication NEMA/MITA 1 Good Refurbishment Practices for Medical Imaging Equipment Published by: National Electrical Manufacturers Association 1300 North 17 th Street, Suite 900 Rosslyn, Virginia 22209 www.nema.org 2015 National Electrical Manufacturers Association. All rights, including translation into other languages, reserved under the Universal Copyright Convention, the Berne Convention for the Protection of Literary and Artistic Works, and the International and Pan American copyright conventions.

Page 2 NOTICE AND DISCLAIMER The information in this publication was considered technically sound by the consensus of persons engaged in the development and approval of the document at the time it was developed. Consensus does not necessarily mean that there is unanimous agreement among every person participating in the development of this document. NEMA standards and guideline publications, of which the document contained herein is one, are developed through a voluntary consensus standards development process. This process brings together volunteers and/or seeks out the views of persons who have an interest in the topic covered by this publication. While NEMA administers the process and establishes rules to promote fairness in the development of consensus, it does not write the document and it does not independently test, evaluate, or verify the accuracy or completeness of any information or the soundness of any judgments contained in its standards and guideline publications. NEMA disclaims liability for any personal injury, property, or other damages of any nature whatsoever, whether special, indirect, consequential, or compensatory, directly or indirectly resulting from the publication, use of, application, or reliance on this document. NEMA disclaims and makes no guaranty or warranty, express or implied, as to the accuracy or completeness of any information published herein, and disclaims and makes no warranty that the information in this document will fulfill any of your particular purposes or needs. NEMA does not undertake to guarantee the performance of any individual manufacturer or seller s products or services by virtue of this standard or guide. In publishing and making this document available, NEMA is not undertaking to render professional or other services for or on behalf of any person or entity, nor is NEMA undertaking to perform any duty owed by any person or entity to someone else. Anyone using this document should rely on his or her own independent judgment or, as appropriate, seek the advice of a competent professional in determining the exercise of reasonable care in any given circumstances. Information and other standards on the topic covered by this publication may be available from other sources, which the user may wish to consult for additional views or information not covered by this publication. NEMA has no power, nor does it undertake to police or enforce compliance with the contents of this document. NEMA does not certify, test, or inspect products, designs, or installations for safety or health purposes. Any certification or other statement of compliance with any health- or safety-related information in this document shall not be attributable to NEMA and is solely the responsibility of the certifier or maker of the statement.

Page i Foreword Unless otherwise noted, this publication has been approved as a NEMA standard. The purpose of this standard is to provide requirements that lead to a practice of good refurbishment for medical imaging equipment. This Standards Publication was developed by the Good Refurbishment Practices Working Group of the National Electrical Manufacturers Association and approved by the Ultrasound, Magnetic Resonance Imaging, Molecular Imaging, X-Ray, and Radiation Therapy Sections. Section approval of the standard does not imply that all section members voted for its approval or participated in its development. At the time it was approved, the sections were composed of the following members: Acertara Acoustic Laboratories Advanced Accelerator Applications, USA Agfa HealthCare AllTech Medical Systems America, Inc. Bayer Radiology and Interventional Blue Earth Diagnostics Bracco Diagnostics Inc. Brainlab AG Capintec, Inc. Cardinal Health CIRS Digirad EIZO Corporation Eli Lilly & Company EOS imaging FluoroPharma Medical, Inc. FUJIFILM Medical Systems U.S.A., Inc. FUJIFILM Sonosite, Inc. Gamma Medica, Inc. GE Healthcare Hitachi Medical Systems America, Inc. Hologic Inc. IBA IMRIS Ionetix Corporation Jubilant DraxImage Inc. Konica Minolta Medical Imaging USA, Inc. Landauer, Inc. Lantheus Medical Imaging, Inc. LAP of America Laser Applications LLC MEDIAN Technologies, Inc.

Page ii Medtronic, Inc. Mevion Medical Systems, Inc. Navidea Biopharmaceuticals, Inc. NeuroLogica A Subsidiary of Samsung Electronics Neusoft Medical Systems, USA, Inc. Numa, Inc. Philips Photo Diagnostic Systems Inc. Piramal Imaging Limited Samsung Medison Shimadzu Medical Systems USA Siemens Healthcare Spectrum Dynamics Medical, Inc. Swissray International, Inc. The Phantom Laboratory Toshiba America Medical Systems Toshiba Medical Research Institute USA, Inc. Zevacor Molecular Ziehm Imaging, Inc. In the preparation of this standards publication, input of users and other interested parties has been sought and evaluated. Inquiries, comments, and proposed or recommended revisions should be submitted to the concerned NEMA product subdivision by contacting the following address: Senior Technical Director, Operations National Electrical Manufacturers Association 1300 North 17 th Street, Suite 900 Rosslyn, Virginia 22209

Page iii Foreword CONTENTS... i Introduction... iv Section 1 GENERAL... 1 Section 2 SCOPE... 2 Section 3 NORMATIVE REFERENCES... 3 Section 4 TERMS AND DEFINITIONS... 4 4.1 Expected Service Life... 4 4.2 Intended Use / Intended Purpose... 4 4.3 Manufacturer... 4 4.4 Medical Imaging Equipment... 5 4.5 Normal Use... 5 4.6 Operator... 5 4.7 Refurbisher... 5 4.8 Patient... 5 4.9 Process... 5 4.10 Refurbishment... 6 4.11 Repair... 6 4.12 Rework... 6 4.13 Risk... 6 4.14 Used Medical Imaging Equipment... 6 Section 5 GENERAL REQUIREMENTS FOR REFURBISHMENT OF USED MEDICAL DEVICES... 7 5.1 Quality Management System... 7 5.2 Resource Management... 7 5.3 Corrective and Preventive Action... 7 5.4 Customer Complaints... 7 5.5 Production and Service Provision... 7 5.6 Control of Non-Conforming Product... 8 5.7 Post-Market Surveillance Process... 8 5.8 Document Control... 8 5.9 Purchasing... 8 5.10 Control of Design and Design Changes... 8 5.11 Risk Management Process... 8 Section 6 SPECIFIC REQUIREMENTS FOR GOOD REFURBISHMENT PRACTICE... 9 6.1 General... 9 6.2 Selection of Medical Imaging Equipment for Refurbishment... 9 6.3 Evaluating Market Access Requirements... 9 6.4 Preparation for Refurbishment, Disassembly, Packing, and Shipment... 10 6.5 Planning... 10 6.6 Installation of Safety Updates (Hardware / Software)... 10 6.7 Performance and Safety Test... 10 6.8 Packing, Shipment, and Installation of Refurbished Medical Imaging Equipment... 10 6.9 Record of Refurbishment... 10 6.10 Refurbishment Label... 11

Page iv Introduction Keeping up with the latest innovations in medical technology often involves replacing equipment in medical practice before it reaches the end of its expected service life. This is because innovation cycles for medical technology are much shorter than the functional lifecycle of capital equipment. As a result, a sustainable resource management model is required: early replacement of installed medical imaging equipment by newer technology is more economically feasible if the residual value of the existing medical imaging equipment is utilized. Conserving assets is a fundamental principle of ecological thinking in a recycling economy. Several medical imaging equipment companies have already set up quality management systems to refurbish used medical imaging equipment and have delivered this refurbished equipment across the healthcare sector for many years. Refurbishment addresses the high demand for affordable and reliable products. Customers of this used equipment are not only small hospitals with limited budgets but also leading medical institutes. The EU and the US represent by far the two largest markets for refurbished medical equipment. Refurbishment of used medical imaging equipment is a well-established element of the healthcare economy. If used medical imaging equipment is not accurately maintained according to requirements defined by the manufacturer, it may result in additional risk for patients and operators. Consequently, some countries have imposed bans on the importation of used medical imaging equipment to protect public health. These bans fail to distinguish between quality-assured refurbished medical imaging equipment and second-hand medical imaging equipment of undefined quality, with the effect that patients may be denied access to the safe and economical medical imaging equipment they need. Safety and effectiveness are the most important aspects to consider with medical imaging equipment, including used equipment. To ensure safety and effectiveness, used medical imaging equipment has to be refurbished in a highly specialized and quality-assured way.

Page 1 Section 1 GENERAL The purpose of this standard is to provide requirements that lead to a practice of good refurbishment for medical imaging equipment.

Page 2 Section 2 SCOPE This standard describes and defines the process of refurbishment of used medical imaging equipment and applies to the restoring of used medical imaging equipment to a condition of safety and effectiveness comparable to that of new equipment. This restoration includes actions such as repair, rework, software/hardware updates, and the replacement of worn parts with original parts. This standard enumerates the actions that must be performed and the manner consistent with product specifications and service procedures required to ensure that the refurbishment of medical imaging equipment is done without changing the finished medical imaging equipment s performance, safety specifications, or intended use according to its original or applicable valid registration.

Page 3 Section 3 NORMATIVE REFERENCES The following referenced documents are indispensable for the application of this document. For dated references, only the cited edition applies. For undated references, the latest edition of the referenced document (including any amendments) applies. ISO 13485:2003 Medical devices Quality management systems Requirements for regulatory purposes IEC 60601-1:2005+A1:2012 Medical electrical equipment Part 1: General requirements for basic safety and essential performance ISO 14971:2007 Medical devices Application of risk management to medical devices IEC 62353, ed. 2.0 (2014-09) Recurrent Test and Test After Repair of Medical Electrical Equipment 21CFR820.3(x), FDA QSR - Current Good Manufacturing Practices (CGMP) requirements for medical devices

Page 4 Section 4 TERMS AND DEFINITIONS For the purpose of this standard the following terms and definitions apply. 4.1 EXPECTED SERVICE LIFE Expected service life is the time period specified by the manufacturer during which the medical imaging equipment or system is expected to remain safe for use (i.e., maintain basic safety and essential performance). NOTE Maintenance may be necessary during the expected service life. [IEC 60601-1:2005+A1:2012, 3.28] 4.2 INTENDED USE / INTENDED PURPOSE The intended purpose is the use for which a product, process, or service is intended according to the specifications, instructions and information provided by the manufacturer. NOTE Intended use should not be confused with normal use. While both include the concept of use as intended by the manufacturer, intended use focuses on the medical purpose while normal use incorporates not only the medical purpose but also aspects such as maintenance, service, and transport. [IEC 60601-1:2005+A1:2012, 3.44] 4.3 MANUFACTURER Natural or legal person with responsibility for the design, manufacture, packaging, labelling, assembling, or adapting medical imaging equipment, regardless of whether these operations are performed by that person or on that person's behalf by a third party. NOTE 1 ISO 13485:2003 defines labelling as written, printed, or graphic matter as: a) affixed to a medical device or any of its containers or b) accompanying a medical device, related to identification, technical description, and use of the medical device, but excluding shipping documents. In this standard, that material is described as markings and accompanying document. NOTE 2 Adapting includes making substantial modifications to medical imaging equipment already in use. NOTE 3 In some jurisdictions, the responsible refurbisher can be considered a manufacturer when involved in the activities described. NOTE 4 Adapted from ISO 14971:2007, definition 2.8. [IEC 60601-1:2005+A1:2012, 3.55 modified]

Page 5 4.4 MEDICAL IMAGING EQUIPMENT Medical imaging equipment is medical electric equipment that provides images for clinical applications. NOTE See IEC 60601-1:2005+A1:2012 definition 3.63 for a definition of medical electric equipment. 4.5 NORMAL USE Normal use refers to operation, including routine inspection and adjustments by any operator, and stand-by, according to the instructions for use. NOTE Normal use should not be confused with intended use. While both include the concept of use as intended by the manufacturer, intended use focuses on the medical purpose while normal use incorporates not only the medical purpose, but maintenance, transport, etc. as well. [IEC 60601-1:2005+A1:2012, 3.71] 4.6 OPERATOR An operator is a person handling the medical imaging equipment. [IEC 60601-1:2005+A1:2012, 3.73 modified] 4.7 REFURBISHER A refurbisher is a natural or legal person who conducts refurbishment of medical imaging equipment. 4.8 PATIENT A patient is a living being (person or animal) undergoing a medical, surgical, or dental procedure. NOTE A patient can be an operator. [IEC 60601-1:2005+A1:2012, 3.76] 4.9 PROCESS A process is a set of interrelated or interacting activities that transform inputs into outputs. [IEC 60601-1:2005+A1:2012, 3.89]

Page 6 4.10 REFURBISHMENT Refurbishment is the process or combination of processes applied during the expected service life to restore used medical imaging equipment to a condition of safety and effectiveness comparable to when new. NOTE Refurbishment can include activities such as repair, rework, replacement of worn parts, and update of software/hardware but shall not include activities that result in regulatory submissions. 4.11 REPAIR Repair refers to restoring equipment to safe, functional, normal condition. [IEC 62353, ed. 2.0 (2014-09), 3.39] 4.12 REWORK Rework means action taken on a nonconforming product so that it will fulfill the specified Device Master Record requirements before it is released for distribution. 4.13 RISK [21CFR820.3(x), FDA QSR] Risk is the combination of the probability of the occurrence of harm and the severity of that harm. [IEC 60601-1:2005+A1:2012, 3.102] 4.14 USED MEDICAL IMAGING EQUIPMENT Used medical imaging equipment is medical imaging equipment that has been put into service.

Page 7 Section 5 GENERAL REQUIREMENTS FOR REFURBISHMENT OF USED MEDICAL DEVICES 5.1 QUALITY MANAGEMENT SYSTEM Refurbishment of used medical imaging equipment shall be conducted under a quality management system (QMS) of the refurbisher in compliance with ISO 13485. In addition to ISO 13485, the provisions in paragraphs 5.2 to 5.11 shall be applied. 5.2 RESOURCE MANAGEMENT The refurbisher shall determine, and provide adequate resources, including trained and qualified personal, maintained and calibrated equipment, and instructions, procedures, files, records, or documents to perform the refurbishment, as well as an environment for refurbishment that is in complete compliance with the applicable environmental, occupational health and safety, and pest control requirements. 5.3 CORRECTIVE AND PREVENTIVE ACTION The refurbisher shall implement a comprehensive corrective action and preventive action (CAPA) process, addressing the specific aspects of the refurbishment of used medical imaging equipment. In addition, in the event that the refurbisher identifies, through its CAPA system, safety-related issues that are the responsibility of the original manufacturer and not related to the refurbishment, it shall inform the original manufacturer accordingly. 5.4 CUSTOMER COMPLAINTS The refurbisher shall have in place a system for managing complaints. In addition, the refurbisher shall communicate to the original manufacturer all customer complaints that are not related to the refurbishment of the equipment. 5.5 PRODUCTION AND SERVICE PROVISION The refurbisher shall have documented procedures for refurbishment and service including but not limited to process validation, disinfection processes, identification, traceability and packaging. In addition the organization shall make provisions to have the knowledge and the ability for installing and servicing medical imaging equipment, or to ensure that servicing can be made available in those markets where the refurbisher makes refurbished medical imaging equipment available on the market.

Page 8 5.6 CONTROL OF NON-CONFORMING PRODUCT The refurbisher shall ensure that a product that does not conform to product requirements is identified during refurbishment and controlled to prevent its unintended use or delivery. When a nonconforming product is corrected during refurbishment it shall be subject to re-verification to demonstrate conformity to the requirements of the original manufacturer. 5.7 POST-MARKET SURVEILLANCE PROCESS The refurbisher shall collect feedback from customers and establish documented procedures to notify regulatory authorities of adverse events. The process shall also determine if the adverse event is related to the refurbishment of the used medical imaging equipment or needs to be reported to the original manufacturer. The refurbisher shall also establish his or her own post-market surveillance process to monitor whether the additional risks resulting from refurbishment have been adequately mitigated. The refurbisher shall enable monitoring of its installed base of refurbished medical imaging equipment to allow for update management for safety and effectiveness. 5.8 DOCUMENT CONTROL The refurbisher shall control all work instructions and procedures used to refurbish medical imaging equipment. 5.9 PURCHASING The refurbisher shall establish dedicated supplier management capabilities when components or services are purchased. 5.10 CONTROL OF DESIGN AND DESIGN CHANGES The refurbisher shall review, verify, and validate potential design changes to ensure that the safety and effectiveness requirements of the equipment are not changed from its original or applicable valid registration. All changes, including parts, must be evaluated to determine if the refurbisher needs registration, as he or she may become the legal manufacturer. 5.11 RISK MANAGEMENT PROCESS The refurbisher shall also establish a risk management process that includes any risk introduced by the refurbishment of used medical imaging equipment. This includes changes that would affect parts.

Page 9 6.1 GENERAL Section 6 SPECIFIC REQUIREMENTS FOR GOOD REFURBISHMENT PRACTICE The refurbisher shall establish a specific process for the refurbishment of used medical imaging equipment that, in addition to the general requirements described in Section 5 of this standard includes the following specific requirements. 6.2 SELECTION OF MEDICAL IMAGING EQUIPMENT FOR REFURBISHMENT The refurbisher shall determine the criteria that used medical imaging equipment must meet to qualify for refurbishment, based on an assessment of the risk (in accordance with ISO 14971) associated with refurbishment, for any type of medical imaging equipment it wishes to process. This determination shall consider the following items: a) Intended use and normal use of the equipment b) Expected service life c) Applicable standards d) Service/maintenance history for the equipment e) Existing procedures for the refurbishment of medical imaging equipment, such as service, repair, production, and maintenance Used medical imaging equipment that is at the end of expected service life or that cannot be restored to at least the original safety and performance levels, including all mandatory safety updates, shall not be refurbished. 6.3 EVALUATING MARKET ACCESS REQUIREMENTS To ensure regulatory compliance, the refurbisher shall have a process in place to evaluate market access requirements, such as valid registrations and licenses or restrictions, and to provide instructions for use in the appropriate languages, safety information, warnings, and labels. Refurbished medical imaging equipment for which the registrations or licenses of the original or refurbished medical imaging equipment have been discontinued or where there is no original license may require the refurbisher to obtain a valid registration prior to commercialization.

Page 10 6.4 PREPARATION FOR REFURBISHMENT, DISASSEMBLY, PACKING, AND SHIPMENT The refurbisher shall have procedures in place to ensure that the medical imaging equipment has been suitably cleaned and disinfected to avoid harming any person involved in the disassembly, packing, and shipment. The medical imaging equipment shall be adequately disassembled (if necessary) and packed to prevent damage during shipment. Appropriate procedures shall be in place to avoid violation of privacy rules concerning patient data possibly stored on the relevant equipment. 6.5 PLANNING A refurbishment plan shall be developed and followed to restore the medical imaging equipment to a condition of safety and effectiveness comparable to when new. 6.6 INSTALLATION OF SAFETY UPDATES (HARDWARE / SOFTWARE) The refurbisher shall install all safety updates released by the manufacturer for the relevant medical imaging equipment since it was placed on the market. 6.7 PERFORMANCE AND SAFETY TEST Tests specified for the original medical imaging equipment shall be conducted to verify that original performance and safety specifications are met, including all mandatory safety updates. 6.8 PACKING, SHIPMENT, AND INSTALLATION OF REFURBISHED MEDICAL IMAGING EQUIPMENT Packing and shipment shall be adequate to prevent damage during transit and load/unload operations. Installation, inspection, and any required testing shall be performed according to documented procedures of the manufacturer. 6.9 RECORD OF REFURBISHMENT The record shall reflect for the relevant medical imaging equipment that all operations and processes described in the refurbishment plan have been accomplished. In addition, the record of refurbishment is specifically required to contain, or refer to the location of, the following information: a) date of refurbishment b) any medical imaging equipment identification and control numbers used c) the primary identification label and labelling used for each refurbished unit d) the acceptance records that demonstrate that the equipment has been refurbished in accordance with the refurbishment plan e) list of replaced parts and their identification information The record shall authenticate any refurbished medical imaging equipment through means that allow inspection by authorities and verification by customers as requested.

Page 11 6.10 REFURBISHMENT LABEL The refurbisher shall label all refurbished medical imaging equipment in proximity to the original label identifying that the medical imaging equipment has been refurbished by the refurbisher and the date of refurbishment. Bibliography [2] ISO 14971:2007, Medical devices Application of risk management to medical devices