The Determination of Toxic Metals in Waters and Wastes by Furnace Atomic Absorption

Similar documents
The Determination of Trace Elements in Stainless Steel by Forked Platform GFAAS

DETERMINATION OF ELEMENTS IN DRINKING WATER AS PER BUREAU OF INDIAN STANDARDS 10500, & USING THE AGILENT 5100 ICP-OES

Trace Metals in Waters by GFAAS, in Accordance with U.S. EPA and Health Canada Requirements

US EPA Method using the Thermo Scientific icap 7600 ICP-OES Duo

Review on Agriculture and Rural Development vol. 2. (1) ISSN

CERTIFICATE OF ANALYSIS REVISED REPORT

The Determination of Trace Elements in Waters using the US EPA Method and the Thermo Scientific icap 6500 ICP-OES (Duo)

CERTIFICATE OF ANALYSIS

High-Purity Acids & Reagents

REPORT NUMBER REPORT DATE SEND TO ISSUE DATE Apr 18, Apr 18, 2017 RECEIVED DATE Apr 05, 2017

US EPA Method using the XSERIES 2 ICP-MS

Water Analysis Using ICP-OES with an Ultrasonic Nebulizer

METHODS OF ANALYSIS BY THE U.S. GEOLOGICAL SURVEY NATIONAL WATER QUALITY LABORATORY-- DETERMINATION OF CHROMIUM IN WATER BY GRAPHITE

Selenium in Natural Waters by Graphite Furnace Atomic Absorption using EPA Method 200.9

LABORATORY TURNAROUND TIMES AND CHARGES

Agilent 200 Series AA Systems AGILENT ATOMIC ABSORPTION SPECTROMETERS

Strong Acid Leachable Metals (SALM) in Soil - Prescriptive

EU Water Analysis Using the Thermo Scientific icap 7400 ICP-OES Duo

Composition Analysis of Animal Feed by HR ICP-OES

Automated Real-Time Determination of Bromate in Drinking Water Using LC-ICP-MS and EPA Method Application

Removal of Arsenic from Water under Static-State Conditions. Jordan Finneseth. Internship: January 2014 July 2014

Agilent Resources for Workflow Compliance MEASURING ELEMENTAL IMPURITIES IN PHARMACEUTICAL MATERIALS

Cadmium Testing and On Site Calibration for Water Testing Detection Range: ppm

California State University, Sacramento: Lead Study

Standard Test Methods for Iron in Water 1

The Agilent Atomic Spectroscopy portfolio for Environmental applications AA, MP-AES, ICP-OES, ICP-MS & ICP-QQQ

Final Report DMR-QA 36. NPDES Permit: IL DMR-QA Study. Open Date: 03/18/16. Close Date: 07/01/16. Report Issued Date: 07/29/16

SUB-Programs - Calibration range Fe Base for "PMI-MASTER Pro" Spark - mode Fe 000

Corrosion Studies on GeoBrom HG520

Coal Ash Material Safety

Application Note. Abstract. Authors. Introduction. Environmental

SILLIKER, Inc. Southern California Laboratory 6360 Gateway Drive, Cypress, CA Tel. 209/ Fax. 714/

Evaluation of Energy-Dispersive X-Ray Fluorescence for Mobile Analysis of Hazardous Metals in Transuranic Waste

DR 1900 PORTABLE SPECTROPHOTOMETER

Newmont Mining Corporation Water Management Standard

ICP Expert software. Technical Overview. Introduction

METHOD 3031 ACID DIGESTION OF OILS FOR METALS ANALYSIS BY ATOMIC ABSORPTION OR ICP SPECTROMETRY

R = kc Eqn 5.1. R = k C Eqn 5.2. R = kc + B Eqn 5.3

Micro Methods Laboratory Quality Assurance Manual

RE: Split Ash Sample Results October 2017 Charah Brickhaven Facility

High Purity Acids Trace Elemental Analysis. Detect as low as 1 to 100 ppt

EPA Method 543: Selected Organic Contaminants by Online SPE LC/MS/MS Using the Agilent Flexible Cube

Analysis of Trace Elements in Seawater Using the Thermo Scientific icap 7000 Series ICP-OES Duo

Drinking Water Analyses

CARBOWAX SENTRY Polyethylene Glycol (PEG) and Methoxypolyethylene Glycol (MPEG) Products

Testing Report. Applicant: XX Overall Rating: XX Data Satisfactory Contact(s): XX Unsatisfactory Others, See Detail Enclosed

Corrosion Studies on GeoBrom HG520

By Sandra R. Jones and Betty J. McLain U.S. GEOLOGICAL SURVEY. Open-File Report

The image part with relationship ID rid4 was not found in the file. Welcome

Corrosion Studies on GeoBrom HG520

SMS 100 MERCURY ANALYZER SMS 100. Mercury analysis directly on solid samples from the undisputed leader in Atomic Spectroscopy

Analysis of Trace Metals in Surface and Bottled Water with the Optima 7300 DV ICP-OES

Environment Laboratory, Central Mine Planning & Design Institute Limited, Regional Institute-IV, Kasturba Nagar, Jaripatka, Nagpur, Maharashtra

Sprint analysis of lubricating oils using the Thermo Scientific icap 7600 ICP-OES

Surface Water Samples: Containers, Preservation and Hold Times Table

CANADA BRITISH COLUMBIA WATER QUALITY MONITORING AGREEMENT

CANADA BRITISH COLUMBIA WATER QUALITY MONITORING AGREEMENT

SCHEDULE I LIST I GENERAL STANDARDS AND CRITERIA FOR THE DISCHARGE OF INDUSTRIAL EFFLUENTS INTO INLAND SURFACE WATERS

FORM 26R CHEMICAL ANALYSIS OF RESIDUAL WASTE ANNUAL REPORT BY THE GENERATOR INSTRUCTIONS GENERAL INFORMATION

Use of Lime Activated Class F Fly Ash for Full Depth Reclamation (FDR) of Asphalt Pavements

Pilot Study Report. Z-92 Uranium Treatment Process

CHAPTER 4 MINIMUM DESIGN CRITERIA

DRPK-005

Simple, fast and reliable analysis of lead in whole blood using the Thermo Scientific icap Q ICP-MS

Selenium Removal. Caroline Dale

EXPORT SOIL-MATERIAL QUALITY REQUIREMENTS

Nitrogen, Simplified TKN (s-tkn ) DOC

Midwest Analytical Services, Inc. "Where industry comes for answers." PRICE LIST 2009

Analysis of Cast Iron Using Shimadzu PDA-7000

Agilent 5100 ICP-OES DUAL VIEW ICP-OES MINUS THE WAIT

Standard Practice D a for Sampling, Preservation and Mitigating Interferences in Water Samples for Analysis of Cyanide

WHERE DO THE PRETREATMENT PERMIT LIMITS COME FROM??? Amy Varinoski City of Mebane

BLACK STURGEON LAKES WATER QUALITY MONITORING

Chapter 1.6. Polished Single-Crystal Silicon, Prime Wafers (all numbers nominal) Wafer Specification Table. Diameter 100 mm 4-inch 150 mm 6-inch

METHOD 0060 DETERMINATION OF METALS IN STACK EMISSIONS

NATIONAL REGULATION AND BORDER CONTROL ACTIVITIES IN THAILAND

Elemental Impurities Regulations View from a CRO

Table 1. VOCs Detections (all data in µg/l)

Report AQUA METROLOGY SYSTEMS. SafeGuard H2O Pilot Report. Online Cr(VI) Remediation System

CANADA BRITISH COLUMBIA WATER QUALITY MONITORING AGREEMENT

Analysis of elemental impurities in drug products using the Thermo Scientific icap 7600 ICP-OES Duo

Dynamic Waste Water Modeling for Coal Burning Power Plants

Using a Metals Translator for Establishing NPDES Limits: A Case Study

The Release of Base Metals During Acidic Leaching of Fly Ash

TOP INORGANIC STANDARDS

Test Report No. F690101/LF-CTSAYHA Issued Date : August 20, 2013 Page 1 of 5

GENERAL PERMIT WMGR098 BENEFICIAL USE OF WASTE FOUNDRY SAND AND SAND SYSTEM DUSTS

Borough of Hopewell, New Jersey

Schedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK

CLEAN WATER Clean Water PT Scheme Schedule Clean Water PT Scheme Schedule. Schedule subject to change see ERA s website at

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005

Method (0.1 to 8.0 mg/l Cu) TNTplus 860

Avio 200 ICP-OES Flat Plate

National Food Safety Standard

Official Journal of the European Union

INSTRUCTIONS AND SAMPLE FORMAT FOR PREPARING A PETITION FOR VARIANCE FOR AN AQUIFER REMEDIATION ZONE OF DISCHARGE

Edward F. Askew PhD Askew Scientific Consulting

Locked Cycle Leaching Test and Yellowcake Precipitation

Water Quality Permitting Program Monitoring Matrix 1,2,3

Transcription:

The Determination of Toxic Metals in Waters and Wastes by Furnace Atomic Absorption Application Note Atomic Absorption Authors Douglas E. Shrader Lucinda M. Voth Lawrence A. Covick Introduction With the increasing awareness of the significance of many trace metals at ultratrace as well as toxicological levels, there is a continuing need to improve existing methods for determining those elements. Analysts must monitor elements such as arsenic, selenium, lead, and cadmium in a rapidly increasing number of samples. The maximum allowable concentration levels of these toxic metals in waters and wastes are regulated by the U.S. EPA and individual states. In drinking water, maximum levels range from 0.002 mg/l for mercury to 1 mg/l for barium [1]. NPDES permit levels for discharging effluents vary, depending upon the metal and industry involved. Guidelines for liquid and solid wastes are set at 100 times the drinking water levels [1]. Because of the extremely low concentrations encountered in many samples, furnace atomic absorption methods are part of EPA approved methodology [2].

The Case for Careful Methods Development Seven metals (As, Ba, Cd, Cr, Pb, Se, and Ag), are regulated in drinking water and may be determined by furnace AA. These same metals are to be determined in liquid and solid waste samples as part of the EP toxicity test used to characterize a hazardous waste [3]. The only approved method for the eighth primary metal in drinking water, mercury, is the cold vapor generation technique. During operation, a temperature profile is displayed on the VDU, overlaid with the analytical signal from the AA, providing a quick and convenient means of determining optimum furnace parameters. An example of this display is shown in Figure 3. Graphite furnace atomic absorption techniques, typically 100 times more sensitive than traditional flame AA methods, require that the analyst take extra care in developing his method. Since µg/l (ppb) levels of metals are normally being determined, extreme caution must be used to prevent contamination of solutions. Required precision and accuracy, sample and standard preparation, and Number AA-31 June 1983 possible interferences and matrix modifications must be considered. This study was undertaken to ascertain optimum parameters, necessary matrix modifiers, and performance characteristics of the seven toxic metals listed above. Instrumentation The instrumentation used for this study was the Agilent AA- 975 Atomic Absorption Spectrophotometer and GTA-95 Graphite Tube Atomizer with programmable sample dispenser. The GTA-95 control unit contains a video display unit. The VDU is used in setting up furnace operating parameters. Temperatures, times, gas flows, and gas type are easily programmed for up to twenty individual steps in the furnace method. The VDU is also used to program sample dispenser parameters. The sample dispenser contains a computer-controlled stepper-motor-driven syringe, which may be programmed to pick up and deliver selected volumes of up to four solutions to the furnace in a single injection. A typical sample dispenser program is shown in Figure 1. Normal calibration was accomplished by injecting variable volumes of a single stock standard. Variable blank volumes were included to keep total injection volume constant. A matrix modifier was automatically added. The sample dispenser may also be programmed to automatically spike samples for analysis using the method of standard additions for calibration (Figure 2). The sample dispenser therefore saves the analyst time by automatically preparing solutions and provides better answers by eliminating potential dilution errors and contamination. Figure 1. Figure 2. Figure 3. Sample parameters, normal calibration. Sample parameters, standard additions. Temperature profile and analytical signal from the AA. 2

Results For each of the seven elements studied, optimum furnace operating parameters were determined. These parameters were used for dilute nitric acid solutions and acetic acid solutions such as those encountered in the test for characteristic of EP toxicity applied to liquid and solid wastes [3]. Recoveries of the seven elements were determined in drinking water and in a complex, high dissolved solids effluent. All results were obtained using the automatic capabilities of the sample dispenser for solution preparation. Injection volumes varied from 10 to 40 µl. Unless otherwise noted, results were derived from peak height measurements of the atomic signal. Arsenic Table 1 illustrates the optimum temperature program for As. Maximum ramp rate to the atomize temperature of 2500 C, along with stop-flow of the normal gas, nitrogen, was used. Temperature programs for other elements were similar with changes made in ash and atomize temperatures as appropriate. Representative peaks and calibration curves for As are shown in Figure 4. Precisions ranged from 2.5 to 4 %RSD. A matrix modifier of 20 mg/l nickel was used. Responses for As were similar in both dilute HNO 3 and acetic solutions. This was true for most elements with the exception of Cr. For utmost accuracy, however, acid type should be the same for both standards and samples. In drinking water, recoveries ranged from 100 to 104%. Low As recoveries were initially obtained from the effluent. However, by increasing the Ni concentration to 1000 mg/l in the effluent and standards, recoveries greater than 90% were achieved. Figure 4. Barium Arsenic calibration. In order to reduce the sensitivity for Ba and cover the concentration range up to 1 mg/l (the maximum drinking water level), 0.5 L/min. of argon (as alternate gas) was used during the atomize step, rather than stop-flow conditions. Furnace operating parameters for Ba are shown in Table 2. The programmability of the GTA-95 permitted less expensive nitrogen to be used during the dry steps, switching to argon prior to the atomize steps for maximum sensitivity and tube life. An ash temperature of 1300 C and atomize temperature of 2500 C were used. Precisions ranged from 0.9 to 4.1 %RSD. Low recoveries for Ba were obtained in both drinking water and effluent. Matrix modification did not alleviate the problem. Accurate results were obtained, however, when the method of standard additions was used for calibration. Table 1. Furnace Operating Parameters for Arsenic Step Temperature Time Gas Gas Read no. ( C) (sec) flow type command 1 75 5.0 3.0 Normal 2 120 50 3.0 Normal 3 1000 1.0 3.0 Normal 4 1000 4.0 3.0 Normal 5 1000 1.0.0 Normal 6 2500.8.0 Normal * 7 2500 2.0.0 Normal * 8 2500 2.0 3.0 Normal 9 40 11 3.0 Normal 10 Table 2. Furnace Operating Parameters for Barium Step Temperature Time Gas Gas Read no. ( C) (sec) flow type command 1 75 5.0 3.0 Normal 2 130 45 3.0 Normal 3 1300 1.0 3.0 Normal 4 1300 5.0 3.0 ALT. 5 1300 1.0.5 ALT. 6 2500.6.5 ALT. * 7 2500 2.0.5 ALT. * 8 2500 1.0 3.0 ALT. 9 40 12 3.0 ALT. 10 3

Cadmium Optimum ash and atomize temperatures for Cd were 300 C and 1800 C, respectively. The use of a slower ramp rate (500 C/sec) to the atomize temperature and peak area measurement of the atomic signal resulted in precisions better than 1 %RSD. Representative peaks and calibration curves for Cd are shown in Figure 5. Ammonium nitrate was found to be the best matrix modifier for Cd, but standard additions was also necessary to achieve 100% recoveries from the effluent. Chromium Precisions obtained for Cr ranged from 1.1 to 1.6% RSD. Ash and atomize temperatures were 900 C and 2500 C respectively. Recoveries for Cr from drinking water were excellent (99 100%). Recoveries of 96 to 100% were obtained from the effluent when ammonium nitrate was used as a matrix modifier. Lead Figure 5. Cadmium calibration. Pb precisions ranged from 0.9 to 2 %RSD. As with Cd, a slow ramp rate of 500 C/sec from the ash temperature (400 C) to the atomize temperature (200 C) produced the best results. With both the drinking water and effluent, standard additions was necessary to obtain recoveries of 96 to 101% for Pb. It is possible that matrix modification with either ammonium nitrate or ammonium oxalate would remove the requirement of standard additions calibration for Pb. Selenium Ash and atomize temperatures for Se were 700 C and 2500 C, respectively. The analysis of four multiple injections of 10 µl produced signals 170% of those obtained with single 40 µl injections. Over the concentration range of interest (0 12 µg/l Se), precisions were better than 5% RSD. Low recoveries were encountered in both the drinking water and effluent when Ni alone was used for matrix modification. Levels to 5000 mg/l Ni were checked in this study. Other modifiers normally used also produced low recoveries. It was found that adding 200 mg/l silver nitrate to the samples, allowing chlorides to precipitate for 30 minutes, followed by analysis using 200 mg/l Ni as modifier, gave usable recoveries (85 112%). 4

Silver Optimum ash and atomize temperatures for Ag were 600 C and 2000 C, respectively. Figure 6 shows representative peaks and calibration curves for Ag. The excellent precisions shown (0.2 to 1.7 %RSD) illustrate the extreme accuracy and repeatability of the sample dispenser as 1, 2, and 4 µl aliquots of a stock standard (automatically diluted to 10 µl with blank solution) were used to build the calibration curves. Improved reproducibility for the drinking water and effluent was obtained when 5% nitric acid was used for matrix modification. Standard additions was necessary to obtain recoveries of 100% from the effluent. Accurate Determinations Table 3 presents results obtained in a recent ASTM graphite furnace round robin for water. One of the samples supplied was a quality control check sample. The concentrations found for the QC check sample correlate extremely well with reported (true) values. As can be seen from this data (and as shown for the seven toxic metals discussed), accurate results can be obtained with furnace atomic absorption, provided the analyst keeps in mind potential interferences, matches acid matrices if necessary, and uses techniques such as matrix modification and standard additions if interferences are present. The Agilent GTA-95 Graphite Tube Atomizer with programmable sample dispenser can save the analyst precious time by automatically preparing solutions and can provide more accurate answers by eliminating potential dilution errors and contamination. Table 3. ASTM Round Robin, Quality Control Check Sample Reported value Concentration Element (µg/l) found Aluminum 61 65 Arsenic 24 24 Beryllium 24 25 Cadmium 6.5 6.6 Chromium 4.4 4.8 Cobalt 30 30 Copper 8.7 8.5 Iron 16 16 Lead 30 30 Manganese 7.9 8.1 Nickel 8.7 8.5 Selenium 8.7 9.0 Vanadium 78 76 Figure 6. Silver calibration. 5

References 1. US EPA, A Handbook of Key Federal Regulations and Criteria for Multimedia Environmental Control, EPA-600/7-79-175(1975). 2. US EPA, Methods for Chemical Analysis of Water and Wastes, EPA-600/4-79-020 (1979). 3. US EPA, Test Methods for Evaluation of Solid Waste, Physical/Chemical Methods, SW-646 (1980). For More Information For more information on our products and services, visit our Web site at www.agilent.com/chem www.agilent.com/chem Agilent shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material. Information, descriptions, and specifications in this publication are subject to change without notice. Agilent Technologies, Inc., 1983 Printed in the USA November 1, 2010 AA031