Organohalide Pesticides in Water by GC/µECD with Agilent J&W DB-CLP1 and DB-CLP2

Similar documents
Contract Laboratory Program (CLP) Pesticide Analysis with 0.18 mm ID High-Efficiency GC Columns Utilizing Helium Carrier Gas. Application.

A Direct Column-Performance Comparison for Rapid Contract Laboratory Program (CLP) Pesticide Analysis. Application. Authors. Abstract.

Analysis of 16 Pesticide Residues in Broccoli Using CarbonX dspe QuEChERS AOAC Kits Using GC/MS

Analysis of USP <467> Residual Solvents using the Agilent 7697A Headspace Sampler with the Agilent 7890B Gas Chromatograph

ANALYSIS OF PESTICIDE RESIDUES IN DRINKING WATER AS PER BUREAU OF INDIAN STANDARDS USING THE AGILENT 7000 GC/MS/MS WITH PESTICIDES ANALYZER

Agilent Ultimate Plus Fused Silica Tubing

Application Note. Abstract. Authors. Introduction. Environmental

FULLY AUTOMATED SPE-DEX Horizon Technology. understands the. challenges today s. laboratories face. The automated. extraction systems

SPE-DEX 4790 Automated Extraction System. A Fully Automated Sample Preparation System. votre distributeur

CHROMATOGRAPHY/MASS SPECTROMETRY

Application Note: Multi-Residue Pesticides in Fruit and Vegetables (9/16/08)

Analysis of Intact and C-terminal Digested IgG1 on an Agilent Bio MAb 5 µm Column

Accurately Identify and Quantify One Hundred Pesticides in a Single GC Run

Capillary GC Troubleshooting: a Practical Approach

Extraction of POPs from Environmental Samples using Accelerated Solvent Extraction (ASE )

Gas Chromatography Assignment Chem 543/443

Analysis of Suspected Flavor and Fragrance Allergens in Cosmetics Using the 7890A GC and Capillary Column Backflush Application

EPA Method 543: Selected Organic Contaminants by Online SPE LC/MS/MS Using the Agilent Flexible Cube

Determination of Monocrotophos, Diazinon, Malathion, EPN, and Methamidaphos from Aqueous Samples Using Atlantic HLB SPE Disks

Application Note. Author. Abstract. Biopharmaceuticals. Verified for Agilent 1260 Infinity II LC Bio-inert System. Sonja Schneider

High-throughput and Sensitive Size Exclusion Chromatography (SEC) of Biologics Using Agilent AdvanceBio SEC Columns

EXPORT SOIL-MATERIAL QUALITY REQUIREMENTS

MS Certified Vials. Pre-cleaned and certified vials for mass spectrometry

Application Note. Author. Abstract. Environmental. Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany

SURPLUS SOIL-MATERIAL REUSE REQUIREMENTS

Separation and Quantitation of a Mixture of Alcohols by Gas Chromatography

UHPLC Separation of Triazine Herbicides at Elevated Temperature

PerkinElmer Supplies/Parts. GC Accessories. Autosampler Syringes

SCION GC Series. The Gas Chromatographers Choice for Separations

Optimization of Volatile Petroleum Hydrocarbons Analysis by Automated Headspace Using Method Development Tools

Fast and High-Resolution Reversed-Phase Separation of Synthetic Oligonucleotides

Protocol for Quantitative Determination of Residual Solvents in Cannabis Concentrates Prepared by: Amanda Rigdon, May 23 rd, 2016

SCION GC Series. The Gas Chromatographers Choice for Separations

Choosing the Right Calibration for the Agilent Bio SEC-3 Column

Monitoring for Microcystins in Raw Water Supply Reservoirs Using the Agilent 6410 Triple Quadrupole LC/MS

mab Titer Analysis with the Agilent Bio-Monolith Protein A Column

PROFICIENCY TESTING PROGRAM. June 2014

Total Petroleum Hydrocarbons By Gas Chromatography in Colombian Waters and Soils

Optimizing the Purification of a Standard Chiral Compound Utilizing a Benchtop, Multi-Purpose, Semi-Preparative to Preparative HPLC System

Determination of Persistent Organic Pollutants (POPs) in Fish

Quantitative determination of residual 2-(2-chloroethoxy) ethanol (CEE) in quetiapine fumarate by gas chromatogaraphy

Method Translation in Liquid Chromatography

Determination of organochlorine pesticide residues in eggs by gel-permeation chromatography (GPC) cleanup and GC-ECD

Volatile Organic Compounds in Solids PBM

PAL SPME Fibers Optimized for Automation

Improved SPE for UPLC/MS Determination of Diquat and Paraquat in Environmental

CORESTA RECOMMENDED METHOD N 61

Converting a Liquid-Liquid Extraction Method for Vitamin D to a 96-Well Plate Supported Liquid Extraction Format

PROFICIENCY TESTING PROGRAM. June 2013

Technical Overview. Author. Abstract. Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany

Chem 321 Lecture 23 - Liquid Chromatography 11/19/13

Formulas: See Table 1 MW: Table 1 CAS: Table 1 RTECS: Table 1. METHOD: 2555, Issue 1 EVALUATION: PARTIAL Issue 1: 15 March 2003

Co-ordination of an organochlorine pesticides in drinking water proficiency testing scheme in South(ern) Africa

Certified Reference Material - Certificate of Analysis

Peptide Mapping of Glycoprotein Erythropoietin by HILIC LC/MS and RP-LC/MS

Environmental Management Chapter ALABAMA DEPARTMENT OF ENVIRONMENTAL MANAGEMENT WATER DIVISION - WATER SUPPLY PROGRAM ADMINISTRATIVE CODE

GERBER PURE Water. Primary Inorganics Antimony

Automated Online SPE for LC/MS/MS Analysis of Trace Organic Contaminants in Water Using the Agilent 1290 Infinity Flexible Cube Module

The Determination of Toxic Metals in Waters and Wastes by Furnace Atomic Absorption

Alex J. Zhang, Pranathi Perati, and Monika Verma; Thermo Fisher Scientific, Sunnyvale, CA, USA

The Importance of High Speed Data Acquisition in the Design and Implementation of GC/MS Applications

2013 Water Analysis Report

Quantitation of cyanotoxins in drinking water according to EPA 544 guidelines

Agilent 7820A GC System THE EASY CHOICE FOR ROUTINE ANALYSES

HYDROGEN AS A CARRIER GAS. Gas Chromatography new frontiers: can Hydrogen substitute Helium?

Size Exclusion Chromatography of Biosimilar and Innovator Insulin Using the Agilent AdvanceBio SEC column

Agilent AdvanceBio SEC Columns for Aggregate Analysis: Instrument Compatibility

Automated Solid Phase Extraction of PAH Compounds Utilizing the SPE-DEX 5000

Validation Report 19

Preparation. Sample. Extraction Glassware Liquid/Liquid Extraction SUPELCO

Olga Shimelis, Katherine Stenerson, Michael Ye, Carmen T. Santasania, and Ed Mauney. Supelco, Div. of Sigma-Aldrich, Bellefonte, PA USA

Disulfide Linkage Analysis of IgG1 using an Agilent 1260 Infinity Bio inert LC System with an Agilent ZORBAX RRHD Diphenyl sub 2 µm Column

Automated Real-Time Determination of Bromate in Drinking Water Using LC-ICP-MS and EPA Method Application

Inorganic Chemicals (IOCs)

Agilent 7693A Series Automatic Liquid Sampler INJECT NEW PERFORMANCE INTO YOUR GAS CHROMATOGRAPHY

CLEAN WATER Clean Water PT Scheme Schedule Clean Water PT Scheme Schedule. Schedule subject to change see ERA s website at

Physical Stability of a Silica- Based Size Exclusion Column for Antibody Analysis

Low-level Detection of Ethanol, 1,4-Dioxane, and Other Oxygenates Using the Eclipse Purge-and-Trap Sample Concentrator

CONTENTS Alternate Retention Time Window Determination

Agilent 7693A Series Automatic Liquid Sampler INJECT NEW PERFORMANCE INTO YOUR GAS CHROMATOGRAPHY

Cell Clone Selection Using the Agilent Bio-Monolith Protein A Column and LC/MS

On-line SPE-LC/MS/MS to Detect Organonitrogen and Triazine Pesticides at 10ng/L in Drinking Water

Electron Capture Detector for GC-2010 Plus

Improving Sensitivity in Bioanalysis using Trap-and-Elute MicroLC-MS

Protein-Pak Hi Res HIC Column and HIC Protein Standard

Using the Pulsed Flame Photometric Detector (PFPD) for Low-level Analysis of Organophosphorus Pesticides

Determination of Perchlorate in Drinking Water by Ion Chromatography

Bio-Monolith Protein G Column - More Options for mab Titer Determination

Agilent 7693A Series Automatic Liquid Sampler. Inject new performance into your gas chromotography

Lifetime stability of size exclusion chromatography columns for protein aggregate analysis

It works as hard as you do.

Developing Robust and Efficient IEX Methods for Charge Variant Analysis of Biotherapeutics Using ACQUITY UPLC H-Class System and Auto Blend Plus

Investigation of Solid Phase Microextraction as an Alternative to Dried Blood Spot

HYDROCARBONS, BP C 1500

Concentration of Human Hormones in Drinking Water Using Solid Phase Extraction and Analysis by High Performance Liquid Chromatography

BÜCHI Labortechnik AG

APPLICATION NOTE. UOP : Analysis of Trace CO and CO 2 in bulk H 2 and Light Gaseous Hydrocarbons by GC

International Atomic Energy Agency. Department of Nuclear Sciences and Applications. IAEA Environment Laboratories

BOTTLED WATER QUALITY REPORT. FIJI WATER COMPANY W. Olympic Boulevard Los Angeles, CA INTRODUCTION OUR SOURCE FOR OUR WATER

Transcription:

Organohalide Pesticides in Water by GC/µECD with Agilent J&W DB-CLP and DB-CLP Application Note Environmental Authors Doris Smith and Ken Lynam Agilent Technologies, Inc. Abstract Organohalide pesticides were successfully extracted from water samples using liquid-liquid extraction and analyzed by a dual column GC/µECD method. An Agilent J&W DB-CLP and DB-CLP column set provided suitable separation of the targeted analytes, while enabling reliable detection at and below established maximum contaminant level concentrations. The method was calibrated over a range of 0.0-.0 µg/l, demonstrating the effectiveness of this application. Introduction The widespread agricultural application of organohalide pesticides and herbicides has led to their detection in many environmental ground and surface waters. Pesticide residues can enter water supplies through runoff from direct applications and by leaching through the soil into groundwater. Because of their toxicity and potential health risks, monitoring pesticides in water for human consumption is of particular concern. The European Union (EU) and United States Environmental Protection Agency (EPA) have established regulations for maximum pesticide levels in drinking waters [,,3]. One of the most common analytical techniques for the determination of pesticides in water is gas chromatography. Many of the EPA methods established for the analysis of pesticides in water are determined by gas chromatography with electron capture detection (ECD), which provides excellent sensitivity and selectivity for trace organohalide pesticides. These methods also recommend the use of GC columns; a primary analytical column along with a second column having a different selectivity for confirmatory analysis.

This application note uses an Agilent pesticide column set for pesticide analysis using EPA Method 505 [4]. The Agilent J&W DB-CLP column is employed for primary analysis, while an Agilent J&W DB-CLP column provides confirmatory analysis with a less polar stationary phase and different selectivity to help verify the analyte s identity. Several of the targeted analytes of EPA Method 505 present separation problems on some manufacturers pesticide column sets. Difficulties in separation have been noted based on published competitor chromatograms between transnonachlor/a-chlordane, simazine/atrazine/g-bhc, and alachlor/aldrin. The Agilent J&W DB-CLP column offers resolution of all pesticides, allowing for more accurate determination. The procedural standard calibration technique is employed by EPA Method 505 to establish the calibration curves. This technique requires that the calibration standards be treated as samples and processed in the same manner as a sample prior to analysis. This method is used to compensate for inefficiencies in the sample preparation procedure. Experimental An Agilent 90A GC equipped with dual µecd detection and an Agilent 3B Automatic Liquid Sampler was used for this study. An inert tee was used to split the effluent : to the primary and confirmation columns. Table lists the chromatographic conditions used for these analyses. Table lists flow path consumable supplies. Table. Chromatographic conditions Column : Agilent DB-CLP 30 m x 0.3 mm, 0.5 µm (p/n 3-3) Column : Agilent DB-CLP 30 m x 0.3 mm, 0.5 µm (p/n 3-33) Carrier: Helium, constant flow.5 ml/min Oven: 90 C (0.5 min), 35 C/min to 5 C, C/min to 300 C (.5 min) CFT device: Inert tee (p/n G34-005), split ratio : Retention gap: 5 m x 0.3 mm id deactivated fused silica tubing Inlet: µl splitless 50 C, purge flow 0 ml/min at 0.5 min GC/µECD: Agilent 90A GC Sampler: Agilent 3B Automatic Liquid Sampler, 5.0 µl tapered syringe (p/n 5-3) µecd: 35 C, constant column + makeup (N ) = 30 ml/min Table. Flow path supplies Vials and caps: MS certified amber crimp top glass vials and caps kit (p/n 590-3) Vial inserts: 50 µl glass/polymer feet (p/n 5-) Syringe: 5 µl tapered (p/n 5-3) Septum: Advanced Green (p/n 53-459) Inlet liner: Ultra inert double tapered liner (p/n 590-393) Ferrules: 0.5 mm id short; 5/5 Vespel/graphite (p/n 50-354) CFT fittings: Internal nut (p/n G55-0530) CFT ferrules: SilTite ferrules, 0.3 mm id (p/n 5-53) 0x magnifier: 0x magnifier loop (p/n 430-00) Reagents and chemicals All reagents and solvents were ACS or Ultra Resi grade. Methanol (MeOH) and hexane from JT Baker were purchased through VWR International (West Chester, PA). Sodium sulfite (Na SO 3 ) and sodium chloride (NaCl) were purchased from Sigma-Aldrich (St. Louis, MO). The EPA 505 analyte standard was purchased from Ultra Scientific (North Kingston, RI, USA). Solutions and standards An aqueous sodium sulfite solution was prepared at 5 mg/ml. This solution was added to the sample during collection to reduce any residual chlorine. The analyte primary dilution standard (PDS) was prepared by diluting the commercially prepared pesticide stock solution with methanol to yield the analytes at a varied concentration of 0.05-.5 µg/ml. Individual analyte concentrations in the PDS can be found in Table 3. This solution was used to fortify reagent water for method calibration. Table 3. Individual analyte concentrations for the primary dilution standard based on the commercially prepared neat standard concentrations. Compound Commercial standard µg/ml Alachlor 0.00 0.50 Aldrin.00 0.05 Atrazine 50.00.50 g-bhc.00 0.05 a-chlordane.00 0.05 g-chlordane.00 0.05 Dieldrin.00 0.05 Endrin.00 0.05 Heptachlor.00 0.05 Heptachlor epoxide.00 0.05 Hexachllorocyclopentadiene.00 0.05 Methoxychlor 5.0 0.5 cis-nonachlor.00 0.05 trans-nonachlor.00 0.05 Simazine 5.00.55 Primary dilution std (PDS) µg/ml

Sample preparation A 35 ml water sample was extracted using liquid-liquid extraction with hexane, and the top hexane layer used for GC analysis. Figure illustrates the sample extraction procedure graphically. A 35 ml aliquot of water was collected in a 40 ml vial and 0.35 ml of 5 mg/ml aqueous Na SO 3 was added to convert any residual free chlorine. The calibration standards were prepared by adding the appropriate amount of primary dilution standard to yield the desired analyte concentrations. Six grams of NaCl were added to the sample, followed by ml hexane. The sample was sealed and shaken vigorously by hand for minute. The organic and aqueous layers were allowed to separate, and the hexane layer transferred to GC vials for analysis. Results and Discussion The targeted organohalide pesticides were resolved on the DB-CLP primary analysis column and DB-CLP confirmation column in less than 4 minutes. Figure depicts the dual column GC/µECD chromatograms of a.0 µg/l extracted standard. Atrazine and simazine co-elute on the DB-CLP, but are resolved on the primary analysis column. The DB-CLP was able to sufficiently separate trans-nonachlor from the chlordane isomers, and atrazine from simazine. The resolution, as calculated by the half width method, for trans-nonachlor relative to gamma-chlordane, was.03 and.9 relative to alpha-chlordane. The resolution between atrazine and simazine was determined to be.. Sample extraction procedure Collect 35 ml water in 40 ml vial Calibration and recovery samples Add 0.35 ml of 5 mg/ml aq. sodium sulfite solution, cap and shake well Fortify with appropriate amount of Primary Dilution Standard to yield desired sample concentration, cap, and shake well Add g NaCl Add ml Hexane, cap, and shake vigorously for minute Transfer Hexane layer to GC vial for analysis Transfer to GC vial and analyze Figure. Flow chart for the extraction of organohalide pesticides in water. 3

Separation of EPA 505 organohalide pesticides 000 5000 4000 3000 000 000 0 Agilent DB-CLP 3 4 5 9 0. Hexachlorocyclopentadiene. Hexachlorobenzene 3. Atrazine (50 µg/l) 4. Simazine (50 µg/l) 5. g-bhc. Heptachlor. Alachlor (0 µg/l). Aldrin 4 0 min 5 4 9. Heptachlor epoxide 0. g-chlordane. trans-nonachlor. a-chlordane. Dieldrin 4. Endrin 5. cis-nonachlor. Methoxychlor (5 µg/l) 000 5000 4000 3000 000 Agilent DB-CLP 3,4 5 9 0 5 4 Analysis time 4 min 000 0 4 0 min Figure. GC/µECD chromatogram of a µg/l fortified standard analyzed on an Agilent J&W DB-CLP 30 m 0.3 mm, 0.5 µm column (part number 3-3) and DB-CLP 30 m 0.3 mm, 0.5 µm column (part number 3-33). This sample was prepared and extracted according to the sample preparation procedure detailed in Figure. Chromatographic conditions are listed in Table. 4

A point calibration curve was generated to test the linearity of the method. Linearity as defined by the correlation coefficient (r ) of the calibration curve can be used to evaluate the performance of a gas chromatographic column. The calibration standards were fortified at 0.0, 0.05, 0., 0., 0.5, and µg/l. These standards were processed and extracted in the same manner as samples to account for any deficiencies in the sample preparation procedure. A nonlinear response can be indicative of breakdown or adsorption of the compound in the inlet or column. The performance of the DB-CLP and DB-CLP columns yielded correlation coefficient (r ) values 0.99 over the calibration range of this study, with the exception of simazine which was slightly lower. The individual pesticide analyte values are shown in Table 4. Figure 3 depicts the GC/µECD chromatogram for a 0. µg/l fortified sample. This level is at the maximum contaminant level (MCL) set by current EU regulations, and below MCLs established by the EPA. The chromatogram shows excellent response and separation at this level which enables the accurate quantification needed for reliable pesticide monitoring. Table 4. Correlation coefficients (r ) for EPA 505 organohalide pesticides calibration standards determined by GC/µECD. r values Analyte DB-CLP DB-ClP Hexachlorocyclopentadiene.0000 0.9999 Hexachlorobenzene.0000 0.993 Atrazine 0.999 0.999 Simazine 0.99 * g-bhc 0.999 0.999 Heptachlor 0.999 0.9999 Alachlor 0.999 0.9993 Aldrin 0.9995 0.999 Heptachlor epoxide 0.9999.0000 g-chlordane 0.999.0000 trans-nonachlor 0.9999 0.9999 a-chlordane 0.9999.0000 Dieldrin 0.999.0000 Endrin 0.999 0.999 cis-nonachlor 0.9999 0.9999 Methoxychlor 0.9999 0.999 *Co-elution Linearity results 0. µg/l Extracted reagent water sample 900 00 00 00 500 400 300 00 00 900 00 00 00 500 400 300 00 00 Agilent DB-CLP 4 0 min Agilent DB-CLP 3 5 4 3,4 5 9 4 0 min 0 9 5 4 0 4 5. Hexachlorocyclopentadiene. Hexachlorobenzene 3. Atrazine (5 µg/l) 4. Simazine (5 µg/l) 5. g-bhc. Heptachlor. Alachlor ( µg/l). Aldrin 9. Heptachlor epoxide 0. g-chlordane. trans-nonachlor. a-chlordane. Dieldrin 4. Endrin 5. cis-nonachlor. Methoxychlor (0.5 µg/l) Figure 3. GC/µECD chromatogram for a water sample fortified at the maximum contaminant level of 0. µg/l set by current EU and EPA regulations This sample was prepared and extracted according to the sample preparation procedure detailed in Figure, and analyzed using an Agilent J&W DB-CLP 30 m 0.3 mm, 0.5 µm column (part number 3-3) and DB-CLP 30 m 0.3 mm, 0.5 µm column (part number 3-33). Chromatographic conditions are listed in Table. 5

The method was able to detect organohalide pesticides with a high level of sensitivity at trace levels. A tap water sample was analyzed for the targeted pesticides to demonstrate the effectiveness of the method. The water sample was collected and prepared according to the sample preparation steps shown in Figure and evaluated under the chromatographic conditions listed in Table. Figure 4 shows the tap water sample relative to a reagent blank and an extracted 0.0 µg/l fortified reagent water sample on the DB-CLP and DB-CLP columns. The 0.0 µg/l sample was fortified at a level well below the established EU and EPA threshold for pesticides in drinking water. The targeted compounds were not detected in the tap water sample at the calibrated range of this study. Extracted tap water sample Relative to reagent blank and low level standard 3. Hexachlorocyclopentadiene. Hexachlorobenzene 3. Atrazine (5 µg/l) 4. Simazine (5 µg/l) 5. g-bhc. Heptachlor. Alachlor (0. µg/l). Aldrin 9. Heptachlor epoxide 0. g-chlordane. trans-nonachlor. a-chlordane. Dieldrin 4. Endrin 5. cis-nonachlor. Methoxychlor (0. µg/l) 350 300 50 00 5 4 9 0 5 4 0.0 µg/l standard Tap water Reagent water 50 Agilent DB-CLP 4 0 min 350 300 3,4 5 9 0 5 4 0.0 µg/l standard 50 00 Tap water Reagent water 50 Agilent DB-CLP 4 0 min Figure 4. GC/µECD chromatograms for a tap water sample, reagent water blank, and a standard fortified at a level of 0.0 µg/l. The chromatographic traces are same scale, but offset for illustration purposes. These samples were prepared and extracted according to the sample preparation procedure detailed in Figure, and analyzed using an Agilent J&W DB-CLP 30 m 0.3 mm, 0.5 µm column (part number 3-3) and Agilent J&W DB-CLP 30 m 0.3 mm, 0.5 µm column (part number 3-33). Chromatographic conditions are listed in Table.

Conclusions This application demonstrates an effective analytical method to extract and detect sub-µg/l level organohalide pesticides in water samples. The Agilent J&W DB-CLP and DB-CLP column set resolved the targeted analytes, while providing excellent sensitivity and reliable quantitation at sub-µg/l levels. An analysis time just under 4 minutes allows for a higher sample throughput and increased laboratory productivity. Calibration standards showed good linearity on both columns, yielding regression coefficients r 0.99 with the exception of simazine, which was slightly lower at 0.99. This method was able to detect pesticides at an order of magnitude below the EU and EPA maximum contaminant levels for pesticides in water. A water sample fortified at 0.0 µg/l was prepared and successfully analyzed by this application exhibiting the capability of DB-CLP and DB-CLP columns for low level pesticide determination. Analysis of a tap water sample did not detect any pesticides at the calibrated levels of this method. References. ECC Council Directive 0//ECC, Off. J. European Communities, No. L 9, August 30 90, p.. EEC, Drinking Waters Directive, Official Journal N9/, Directive 0//EEC (9). 3. EPA, List of Drinking Water Contaminants & MCLs, U.S. Environmental Protection Agency, Washington, DC, Retrieved October 4, 003 from http://www.epa.gov/safewater/mcl.html#mcls. 4. US EPA Method 505. Analysis of Organohalide Pesticides and Commercial Polychlorinated Biphenyl (PCB) Products in Water by Microextraction and Gas Chromatography. Revision.. (995). For More Information These data represent typical results. For more information on our products and services, visit our Web site at www.agilent.com/chem.

www.agilent.com/chem Agilent shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material. Information, descriptions, and specifications in this publication are subject to change without notice. Agilent Technologies, Inc., 0 Printed in the USA March 9, 0 599-00EN