ADVANCES IN MICRO GAS CHROMATOGRAPHY

Similar documents
Fast Analysis of Extended Natural Gas using the New Transportable INFICON Micro GC Fusion

SIMPLIFY AND ACCELERATE GAS ANALYSIS. Micro GC Fusion GAS ANALYZER

MICRO GC FUSION ADVANCING GAS ANALYSIS. Abstract # 128 Paper - 10/11/2016-1:10 PM - Room 380 A Micro GC Fusion Advancing Gas Analysis

MicroSAM / Sitrans CV

System Overview. Valve oven. Restrictor. Valve. Sample in and out

SCION GC Series. The Gas Chromatographers Choice for Separations

SCION GC Series. The Gas Chromatographers Choice for Separations

MAX300-RTG PRODUCT NOTE. Real-Time Gas Analysis Process Automation and Control Product Quality Analysis. Industrial Process Mass Spectrometer

Agilent Gas Phase Solutions

Agilent 7820A GC System THE EASY CHOICE FOR ROUTINE ANALYSES

MIDSTREAM GAS QUALITY

APPLICATION NOTE. UOP : Analysis of Trace CO and CO 2 in bulk H 2 and Light Gaseous Hydrocarbons by GC

Heat Value Gas Chromatograph ISO standard version

The Analysis of 4-Vinyl-Cyclo-Hexene (VCH) in Butadiene with the DVLS LGI Injector

Residual Solvents in medical cannabis analysis using the SRI 8610C FID GC June 2014

Waarom moeilijk doen als het makkelijk(er) kan

Improving Natural Gas Liquefaction Plant Performance with Process Analyzers

Final DRAFT API TECHNICAL REPORT. Carbon Content, Sampling, & Calculation

Carbonyl compounds and VOCs analysis from vehicle exhaust: Conventional and alternative methods of detection, analysis and quantification

(L)NG composition measurements

Fast Detailed Hydrocarbon Analysis by Modified ASTM Method D6730

Select Al2O3 MAPD : a new Inert Al2O3 column for the trace analysis of polar reactive hydrocarbons.

Protocol for Quantitative Determination of Residual Solvents in Cannabis Concentrates Prepared by: Amanda Rigdon, May 23 rd, 2016

Gas Chromatography (GC) Operation, Calibration, Troubleshooting and Practice Maintenance

Thomas G. Braga Manager, Research and Development. SulfaTreat, a Business Unit of M I L.L.C. A Smith/Schlumberger Company

Analysis of Suspected Flavor and Fragrance Allergens in Cosmetics Using the 7890A GC and Capillary Column Backflush Application

Gas Chromatography Assignment Chem 543/443

Fuel quality measurements at small-scale LBG refuelling stations Anders Karlsson, SP LNG Workshop , NPL, London, UK

NATURALLY BETTER CHROMATOGRAPHY SOLUTIONS. Clarus Gas Chromatographs Clarus 680/580/480

Capillary GC Troubleshooting: a Practical Approach

Natural Gas Monitoring

Gas vs. Diesel Generator Sets Performance Cost & Application Differences

Qualification of High-Performance Liquid Chromatography Systems

Totalflow analyzer products Precisely what you expect

A Single Analyzer Solution for Refinery Flare Emissions Monitoring (40 CFR 60 Subpart Ja) Using Mass Spectrometry

Separation and Quantitation of a Mixture of Alcohols by Gas Chromatography

MOLECULAR GATE TECHNOLOGY FOR (SMALLER SCALE) LNG PRETREATMENT

Inland Technologies Inc

Safer, More Efficient Combustion Control for Fired Heaters

Automatic Determination of Greenhouse Gases by GC

ICAC MARAMA Portable Emission Analyzers

UNIFI: The user environment Ken Eglinton Nordic User Training, September 2013

Quick Manual to the Waters UPLC System

HYDROGEN AS A CARRIER GAS. Gas Chromatography new frontiers: can Hydrogen substitute Helium?

QUANLYNX APPLICATION MANAGER

GC8000 Process Gas Chromatograph. Bulletin 11B08A01-01E

Maximizing Chromatographic Resolution of Peptide Maps using UPLC with Tandem Columns

Chambers and Instrumentation CHAMBERS. STORA: 22 m 3 chamber

AKAKUS OIL OPERATIONS- LIBYA Gas Utilization& Flare Emission Reduction Project

Inlet-Residue-NGL Solutions for Natural Gas Processing Applications

Technical Overview. Author. Abstract. Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany

Calculation of LNG enthalpies and calorific values at different reference conditions

GAS SUPPLIER INTERCONNECTION PROJECT FACT SHEET

Impact of GC Parameters on The Separation Part 4: Choice of Film Thickness

Analysis of USP <467> Residual Solvents using the Agilent 7697A Headspace Sampler with the Agilent 7890B Gas Chromatograph

Elemental Analysis (EA)

8 stream capability with dual mode (H 2 S & total Sulphur)

Minimize Method Development Time for Pesticide Screening Using LC/MS Application Kit

Complexity. C9+, H2S, methanol, wet gas. Fiscal / Custody transfer. Simpler installation /operation

Agilent Ultimate Plus Fused Silica Tubing

Fusion Analytical Method Validation

An Improved Method of Updating Retention Times for GCMS

SRU-TGTU Analyzer Best Practices. RefComm Galveston 2016 Sulfur Recovery

Sensor Technologies used in Portable Gas Monitors. HA University

PROCESS. Process Analysers For Gas and Liquid Phase

easy. HPLC has never been so Agilent 1120 Compact LC Our measure is your success.

Designed for the Modern Chromatography Lab. Agilent OpenLab CDS

Carbon / Sulfur Analyzers

COMBIMASS Thermal Gas Flow

Chem 321 Lecture 23 - Liquid Chromatography 11/19/13

ABB MEASUREMENT & ANALYTICS ANALYTICAL MEASUREMENT. Ethylene plant production excels with the PGC1000

Agilent 7693A Series Automatic Liquid Sampler. Inject new performance into your gas chromotography

Techniques for Measuring Oil in Water

Purification Workflow Management

STANDARD OPERATING PROCEDURES

Agilent 7693A Series Automatic Liquid Sampler INJECT NEW PERFORMANCE INTO YOUR GAS CHROMATOGRAPHY

Qualitative Phase Behavior and Vapor Liquid Equilibrium Core

Overview Gas & NGL Processing

WEATHER MONITORING. Competitive rates available for long term hire. Hire rate Per week Per month

Teknologi Pemrosesan Gas (TKK 564) Instructor: Dr. Istadi ( )

mab Titer Analysis with the Agilent Bio-Monolith Protein A Column

Excellence in gas analyzers since Over 186 years of Jobin Yvon optical experience. The expertise of HORIBA combustion technology

PROCESSING NATURAL GAS Leontev A.A. Vladimirskiy State University named after the Stoletov brothers Vladimir, Russia

VortexMaster and SwirlMaster Natural gas measurement in power plants, boilers and incinerators

The Alberta Energy Regulator (AER) has approved this directive on February 17, 2016.

Using the Pulsed Flame Photometric Detector (PFPD) for Low-level Analysis of Organophosphorus Pesticides

Electron Capture Detector for GC-2010 Plus

Total Petroleum Hydrocarbons in Vapor: Seeing the Forest for the Trees

Proline Prosonic Flow B 200 For reliable biogas flow measurement without compromise

It works as hard as you do.

PROCESS MOISTURE ANALYZERS Measuring moisture in gas or HC liquids in hazardous areas

Issues around field measurements with electronic noses : low concentration and specificity of odorous compounds Jacques NICOLAS Julien DELVA -

Prima PRO process mass spectrometer

Development of Key- Enabling Technologies for a Variable-blend Natural Gas Vehicle

Greenhouse Gas Measurement Technology

The Apogee Scientific. Technology

HYDROCARBONS, BP C 1500

10/2/2013. Gas CHEMICAL PLANTS AMMONIA METHANOL UTILITIES TOWN GASS SUPPLIES ENERGY INTENSIVE INDUSTRIES. Power Generation

EPA Method 543: Selected Organic Contaminants by Online SPE LC/MS/MS Using the Agilent Flexible Cube

Guidelines on Quality Assurance/Quality Control for The Environmental Measurements of Dioxins

Transcription:

ADVANCES IN MICRO GAS CHROMATOGRAPHY

Fast and Accurate Analysis of Extended Natural Gas Composition and Physical Properties Using a Temperature Programmable Micro GC Fusion Gulf Coast Conference 2015 Debbie Alcorn INFICON, Inc. 2 Technology Place East Syracuse, NY 13057, USA INFICON Advances in Micro Gas Chromatography 2

Outline Natural Gas Analysis Natural Gas Composition Micro GC History Micro GC Fusion Features Micro GC Configuration Diablo EZReporter for Physical Property Calculation Chromatograms Compliance with GPA Standards Conclusion INFICON Advances in Micro Gas Chromatography 3

INFICON Global Presence 900 employees; offices in 17 countries INFICON Advances in Micro Gas Chromatography 11/18/15 4

Natural Gas Analysis Natural gas composition analysis is critical for : Producers and gatherers Gas distribution companies Gas utility companies Gas engine and appliance manufacturers Independent testing laboratories Due to variations in natural gas composition, it is important to measure physical properties, especially for custody transfer situations Software automatically calculates physical properties such as: Heating Value (BTU/cu. ft) The amount of heat released during the combustion of a specified amount of a substance Relative Density/Specific Gravity The ratio of the density of a substance to the density of water INFICON Advances in Micro Gas Chromatography 11/18/15 5

Natural Gas Composition Component Gross Heating Value (BTU/Cu. Ft) Wellhead % Pipeline % Nitrogen 0 <10 < 2 Carbon Dioxide 0 <10 < 1-2 Methane 1010 70-90 93-96 Ethane 1770 <10 < 1 Propane 2516 <10 < 0.5 Isobutane 3252 <2 < 0.5 n-butane 3262 <2 < 0.5 Isopentane 4001 <2 < 0.5 n-pentane 4009 <2 < 0.5 Hexanes Plus 5129 <2 < 0.2 INFICON Advances in Micro Gas Chromatography 11/18/15 6

Micro GC History 3000 Micro GC (µgc) has over ten years of proven excellent analytical performance Micro GC Fusion expands the capability of the 3000 Micro GC INFICON Advances in Micro Gas Chromatography 11/18/15 7

3000 Micro GC Microelectromechanical systems (MEMS) based GC Isothermally heated capillary columns Micro Thermal conductivity detector (µtcd) Analysis can be conducted directly on-site or in a lab 3000 Micro GC in the field 3000 Micro GC in the lab INFICON Advances in Micro Gas Chromatography 11/18/15 8

The Next Generation Micro GC Fusion Micro GC Fusion builds on proven 3000 Micro GC technology Utilizes the same MEMS injectors and TCD Traditional Micro GC technology is limited due to: Isothermal column temperature, leading to potential carryover issues Ambient temperature effects Software and operating system compatibility Micro GC Fusion addresses these issues using new technology INFICON Advances in Micro Gas Chromatography 11/18/15 9

Micro GC Fusion Features Micro GC Fusion new features include: Temperature programmable columns for fast temperature ramping Additional heated zones to minimize ambient temperature effects Web-based interface to eliminate software compatibility issues Front panel display Solid-state hard drive for data storage Optional integrated sample conditioner INFICON Advances in Micro Gas Chromatography 11/18/15 10

Benefits of Temperature Programming Temperature ramping based on resistive column heating allows for: Faster runs Rapid column cleaning Sharper peaks Expanded application range (ex. extended natural gas analysis) Ramping profile is independently optimized for each module Cooldown time is optimized to achieve short cycle times

Isothermal vs. Temperature Programming Isothermal runs result in broad, late eluting peaks Fast temperature ramping improves peak shape, run time, and column cleaning Example Propane Isothermal ~160 seconds Temperature ramping ~50 seconds Increase in peak height Column: RT-Q-Bond Injector: Variable Ramp: 50 C 90 C 240 C (2 C/s, 2.2 C/s)

Integrated Sample Conditioner An optional integrated sample conditioner can be configured The sample conditioner allows for: Reduction of sample pressures of up to 1000 psi down to approximately 5-10 psi Filtering of particulates Heated sample at 100 C A quick connect replaces the standard 1/16 in. inlet

Single Module Natural Gas Analysis Column: RT-Q-Bond Injector: Variable Ramp: 45 C 250 C (2.5 C/s) INFICON Advances in Micro Gas Chromatography 11/18/15 14

Extended Natural Gas Often, it is desirable to analyze C 6 to C 12 components in order to obtain more information about the gas sample A second channel using a Rxi-1ms column provides additional information for extended hydrocarbon analysis Column: Rxi-1ms Ramp: 50 C 130 C 280 C (2 C/s, 5 C/s) INFICON Advances in Micro Gas Chromatography 03/17/14 15

Extended Natural Gas C12 Analysis In order to analyze C12 in the gas phase, the extended natural gas calibration standard was heated For retention time comparison, a heated sample bag containing straight chain hydrocarbons (C6 to C12) is overlaid with the calibration standard Column: Rxi-1ms Ramp: 50 C 130 C 280 C (2 C/s, 5 C/s) INFICON Advances in Micro Gas Chromatography 03/17/14 16

Diablo EZReporter for Physical Property Calculation Allows customized viewing of results Calculates physical properties such as: Component, gross, and net heating values (BTU/cu.ft) Relative density (specific gravity) Wobbe Index And more Creates reports in PDF, CSV, or TXT format INFICON Advances in Micro Gas Chromatography 11/18/15 17

Natural Gas Standards Method Standard GPA 2261 (Industry standard) Portable Micro GC statement TCD Grouped hexanes plus Repeatability Calculation Standards used by EZReporter: ASTM 3588 GPA 2145 GPA 2172 ISO 6976 INFICON Advances in Micro Gas Chromatography 11/18/15 18

Micro GC Fusion Repeatability Percent relative standard deviation (%RSD) is used to measure instrument performance 10 runs were conducted using the same gas standard Component Calibration Gas Concentration Retention Time %RSD Concentration/Area %RSD Nitrogen 1.550% 0.054 0.236 Carbon Dioxide 1.201% 0.061 0.177 Methane 88.654% 0.132 0.163 Ethane 3.001% 0.120 0.252 Propane 1.996% 0.067 0.211 Isobutane 1.003% 0.054 0.187 n-butane 0.998% 0.053 0.178 Isopentane 0.299% 0.05 0.295 Property Gross Heating Value Ideal (BTU/cu. ft) Gross Heating Value Real (BTU/cu. ft) %RSD 0.015 0.015 n-pentane 0.300% 0.053 0.276 C6s 0.598% Group range 0.309 C7s 0.332% Group range 0.391 C8 Plus 0.068% Group range 2.76

Compliance with NGA Standard GPA 2261 calls out repeatability for compound concentrations Component Repeatability Nitrogen 2% Carbon Dioxide 3% Methane 0.2% Ethane 1% Propane 1% Isobutane 2% n-butane 2% Isopentane 3% n-pentane 3% Hexanes Plus 10% INFICON Advances in Micro Gas Chromatography 11/18/15 20

Compliance with NGA Standard Component Repeatability Calibration Gas Concentration Acceptable Range 10 Run Average Normalized Concentration Pass/Fail Nitrogen 2% 1.550% 1.519-1.581 1.548 Pass Carbon Dioxide 3% 1.201% 1.117-1.225 1.200 Pass Methane 0.2% 88.654% 88.477-88.831 88.667 Pass Ethane 1% 3.001% 2.971-3.031 2.996 Pass Propane 1% 1.996% 1.976-2.016 1.998 Pass Isobutane 2% 1.003% 0.983-1.023 1.003 Pass n-butane 2% 0.998% 0.978-1.018 0.998 Pass Isopentane 3% 0.299% 0.290-0.309 0.299 Pass n-pentane 3% 0.300% 0.291-0.309 0.299 Pass Hexanes Plus 10% 0.998% 0.898-1.098 0.990 Pass INFICON Advances in Micro Gas Chromatography 11/18/15 21

Ten Run Overlay Column: RT-Q-Bond Injector: Variable Ramp: 45 C 250 C (2.5 C/s)

Conclusion Micro GC Fusion is an ideal measurement for natural gas analysis across a variety of applications Micro GC Fusion complies with industry standard methods and has excellent retention time, area, and concentration repeatability Combined with Diablo EZReporter, Micro GC Fusion is a powerful tool that can be used in the field or in the lab INFICON Advances in Micro Gas Chromatography 11/18/15 23

For more information, visit: www.inficon.com Or Booth #440 24

Questions?