AAPM IGRT Recommendations

Similar documents
Treatment Quality Assurance Cone Beam Image Guided Radiation Therapy. Jean-Pierre Bissonnette, PhD, MCCPM

Quality Assurance for particle beam therapy

E2E SBRT Thorax Phantom

2017 ACR Computed Tomography Quality Control Manual FAQS

03/08/2017. Advances and Challenges in Contour QA for Adaptive RT. Objective. Disclosures

RapidArc: Clinical Implementation

PerFRACTION 3D. 3D Pre-Treatment QA and In-Vivo Monitoring. Your Most Valuable QA and Dosimetry Tools. SunCHECK

Ingenia MR-RT. MR Systems. The comprehensive MR-sim solution to fit your planning

Stereotactic Body Radiation Therapy: Planning and Delivery

7/31/2017. Implementing a Clinical Practice Guideline; Lessons From An Early Adopter of MPPG 5.a. Disclosures

Quality Control Concerns for New Generations of CT, MRI, & PET

GENESIS Edition Transforming CT

CT QA SOLUTIONS. Ensure Accurate Screening, Diagnosis and Monitoring

Going Clinical with the Accuray Radixact System. The Montefiore Experience

GENESIS Edition. Transforming CT

MRI-Guided On-line Adaptive Radiotherapy The UCLA Physics Experience. Disclosures 8/3/2016

N.MAFFEI, G.GUIDI, C.VECCHI, G.BALDAZZI Physics Department, University of Bologna, via Irnerio Bologna, Italy

Strategies for Adaptive RT

CODING GUIDELINES. Radiation Therapy. Effective January 1, 2018

Shielding evaluation for IMRT implementation in an existing accelerator vault

Regulations governing the application of medical accelerators

Guidance for the Physics Aspects of Clinical Trials

Patient dose from kilovoltage cone beam computed tomography imaging in radiation therapy

Learning Objectives. New Technology Reimbursement. Understand the process for CPT applications. Najeeb Mohideen, MD

Benchmark & FAQ for NRG BR001: A Phase 1 Study of SBRT for the Treatment of Multiple Metastases

Fluoroscopy units come in many different physical configurations but all consist of the main components shown in this slide.

Disclaimer and Notices.

Handzettel 1. CARE Right Computed Tomography, committed to the right dose Ivo Driesser Austin, July 22 nd 2014

Radiation Oncology. Open Access. Abstract. BioMed Central

The Roles and Task of Medical Physicists in Clinical Trial (Therapy Trials: RTOG and NRG)

Optimization of Low-Dose CT Protocol in Pediatric Nuclear Medicine Imaging

Proton Treatment Planning Systems Daniel Yeung

MEDICAL PHYSICS (MED PHYS)

SBRT LUNG CANCER CLINICAL PATHWAY

Impact of patient rotational errors on target and critical structure dose in IMRT: A 3D simulation study


Treatment planning for scanned proton beams Francesca Albertini, Center for Proton Therapy, Paul Scherrer Institut Switzerland

Knowledge-based treatment planning: fundamentally different, or more of the same?

Target point correction optimized based on the dose distribution of each fraction in daily IGRT

Assuring Safety and Quality in Image-guided Delivery of Radiation Therapy

8/2/2017. Key attributes of scientific excellence: rigor, innovation, and relevance. Medical Physics 3.0. Key Attributes of Scientific Excellence

Motiva: a TV-based platform for remote patient management

Treatment planning for scanned proton beams and IMPT

Delivery uncertainties and their consequences

GHG meeting ESTRO 2016 Minutes Sunday May 1st 3:00-5:00 ATENE Room - Green Group Meetings

Adaptive Re-planning for Lung RT with Multi-targets

Workflow and Clinical Decision Support for Radiation Oncology

Computed Tomography. Printed in Japan

New PET/CT from Siemens helps more patients benefit from premium technologies

Link to publication in the UWA Research Repository

Validation of a method for in vivo 3D dose reconstruction in SBRT using a new transmission detector

Artis Q and ArtisQ.zen

TENDER # SUPPLY OF MOBILE C-ARM RADIOGRAPHY/FLUOROCOPY SYSTEM FOR WESTERN HEALTH

Clinical implementation and evaluation of the Acuros dose calculation algorithm

Eric W. Abelquist, Ph.D., CHP President, Health Physics Society. August 2017

Physicists Quality Control for MR Equipment

AN EVALUATION OF THE CONSISTENCY OF IMRT PATIENT-SPECIFIC QA TECHNIQUES

Compliance Guidance for FLUOROSCOPIC QUALITY CONTROL 2 nd Edition

Quality Control in Flow. Dr David Westerman Head of Haematopathology Peter MacCallum Cancer Centre

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May Dose Reference Levels

Healthcare. Philips DoseWise Portal. Radiation exposure management software

LEVELS OF EMPLOYEE PROOF OF COMPENTENCY TO EMPLOYEES RAJINDER SINGH, RT(T) QUALITY COORDINATOR

Low Energy Consumption, High-Speed Productivity LCG 3015 AJ. Fiber Laser Cutting System

Function of this control

A comparative study between the imaging system and the optical tracking system in proton therapy at CNAO

Statistical quality control for volumetric modulated arc. therapy (VMAT) delivery using machine log data

Effects of 16 bit CT imaging scanning conditions for metal implants on radiotherapy dose distribution

An innovative new platform built to transform the way the world thinks about treating cancer.

RECENT PROGRESS OF HEAVY-ION CANCER RADIOTHERAPY WITH NIRS-HIMAC

Failure to Detect and Interlock an Out-Of-Tolerance Photon Beam Symmetry by a Linear Accelerator Monitor Chamber

Alternative Career Opportunities for Medical Physicists

Low Dose, Lightning Speed, Latest Applications

Department At-A-Glance

Optimize efficiency with a universal system.

The SAVI TM Applicator: Breast Brachytherapy Training

Characterisation of an Extendable Multi-leaf Collimator for Clinical Electron Beams

AAPM REPORT NO. 56 Medical Accelerator Safety Considerations

MR Safety Evaluation of Orthopedic Devices: Why It's Important and How to Select Test Device Configuration

QUANLYNX APPLICATION MANAGER

Uncertainties and correction methods when modeling passive scattering proton therapy treatment heads with Monte Carlo

GE Healthcare. Introducing Veo * on Discovery * CT750 HD Great care by design

Act anything done, being done, or to be done; the process of doing. Synonymous with procedure and clinical services.

Department of Radio Diagnosis

3/6/2017 TOMOSYNTHESIS GUIDED BREAST BIOPSY LEARNING OBJECTIVES DISCLOSURES

Note RADIATION QUALITY OF A TOMOTHERAPY PHOTON FAN BEAM

VOLUMETRIC MODULATED ARC THERAPY (VMAT): ADVANCED DELIVERY TECHNIQUES FOR STATIC AND MOVING TARGETS. A Dissertation. presented to

ACTIVITY-BASED COSTING: A PRACTICAL MODEL FOR COST CALCULATION IN RADIOTHERAPY

Design Criteria for an IBT facility

The Department of Radiation Oncology. In loving Memory of Penny Slattery ( ) Medical Physics Journal Manager ( )

RADIATION EXPOSURE FROM IMAGING GUIDANCE USING THE EXACTRAC IMAGERY SYSTEM ESTIMATED BY PCXMC SOFTWARE. Kathleen Smith EPFL, M.Sci.

Jonathon A. Nye, Ph.D Emory University School of Medicine Atlanta, GA

Magnetic Resonance Imaging at 7T in Glasgow. A unique opportunity!

Virtual Mold Room. The Future of Custom Radiation Therapy Devices

Transcription:

AAPM IGRT Recommendations Joint ICTP-IAEA International Workshop on the Implementation of Image Guided Radiotherapy (IGRT) 8-12 May 2017 J. Daniel Bourland, PhD, DABR Professor, Departments of Radiation Oncology, Physics and Biomedical Engineering Wake Forest School of Medicine Winston-Salem, North Carolina, USA bourland@wakehealth.edu JD Bourland and Wake Forest University Outline AAPM Report 179: CT-based IGRT Mechanical Quality Assurance Imaging Quality Assurance Dose Quality Assurance AAPM Practice Guideline 2.a Commissioning of x-ray based IGRT QA of x-ray based IGRT ACR-AAPM Tech Standard for IGRT Summary

TG-179: QA for CT-based IGRT Bissonnette JB et al. Quality assurance for image-guided radiation therapy utilizing CT-based technologies: A report of the AAPM TG-179. Medical Physics, Vol. 39, No. 4, April 2012 Bissonnette JB et al. Quality assurance for image-guided radiation therapy utilizing CT-based technologies: A report of the AAPM TG-179. Medical Physics, Vol. 39, No. 4, April 2012

Daily Quality Assurance Frequency Quality Metric Quality Check Tolerance Daily Safety Collision and Interlocks Functional System Operation and Accuracy Warning lights Laser/image/treatment isocenter coincidence OR Phantom localization and repositioning with couch shift Functional +/- 2 mm +/- 2 mm Bissonnette JB et al. Quality assurance for image-guided radiation therapy utilizing CT-based technologies: A report of the AAPM TG-179. Medical Physics, Vol. 39, No. 4, April 2012 Daily Laser/Image/Treatment Isocenter Coincidence Agreement Align phantom with 3 implanted gold seeds to lasers and CBCT Check alignment and acquire 2 orthogonal MV port films Use Setup Intelligence in Mosaiq to check alignment of MV ports with reference images

Images courtesy of Department of Radiation Oncology, Wake Forest Baptist Medical Center, Winston-Salem, NC USA Images courtesy of Department of Radiation Oncology, Wake Forest Baptist Medical Center, Winston-Salem, NC USA

Phantom Localization and Table Shift Test Align Quasar phantom to known offset and CBCT Perform translation and check laser alignment Could be done Daily, typically, is performed Weekly Place small angle under 1 corner to test Hexapod rotational corrections Monthly Quality Assurance Frequency Quality Metric Quality Check Tolerance Monthly Geometric Geometric calibration map OR Save new map Image Quality kv / MV laser alignment Couch shifts, accuracy of motions Scale, distance, and orientation accuracy Uniformity, noise High contrast spatial resolution Low contrast detectability +/- 1 mm +/- 1 mm Baseline Baseline < 2 mm, or 5 > lp/cm Baseline If used for dose calc CT number accuracy and stability Baseline Bissonnette JB et al. Quality assurance for image-guided radiation therapy utilizing CT-based technologies: A report of the AAPM TG-179. Medical Physics, Vol. 39, No. 4, April 2012

MV and kv Isocenter Coincidence Ball bearing test performed Procedure outlined in manufacturer acceptance documents Variation of Winston-Lutz test Recommend that procedure be done with physicist Can do in conjunction with an automated MLC calibration MV EPID Stage Align ball bearing phantom (8mm) with lasers Acquire MV images (gantry at 0, 90, 180, 270) Transfer images to XVI and evaluate Reposition ball bearing relative to jaws Repeat until ball bearing is within 0.25mm of MV isocenter

Renstrom, Master of Science Thesis, 2005 http://lup.lub.lu.se/luur/download?func=downloadfile&recordoi d=2156983&fileoid=2157374 Acquire CBCT scan kv CBCT Stage Determine position of ball bearing in each projection Plot position as a function of gantry angle to determine head sag (flex map) Flex maps must be generated for all acquisition modes (CW and CCW) Acquire CBCT scan Compare to reference set Discrepancy is variation between isocenters

Renstrom, Master of Science Thesis, 2005 http://lup.lub.lu.se/luur/download?func=downloadfile&recordoi d=2156983&fileoid=2157374 MV Isocenter Stability Ball bearing image with gantry, collimator, table rotations Winston-Lutz test Isocenter variation should be within 1 mm Note: go into room and look

Annual Quality Assurance Frequency Quality Metric Quality Check Tolerance Annual Dose Imaging dose Baseline Image System Performance X-ray generator (kv systems) kv, ma, exposure time, linearity Baseline Geometric System Operation Anteroposterior, mediolateral, and craniocaudal orientations are maintained Long and short term resource planning (disk space, effort, etc.) Accurate designations In support of clinical use Bissonnette JB et al. Quality assurance for image-guided radiation therapy utilizing CT-based technologies: A report of the AAPM TG-179. Medical Physics, Vol. 39, No. 4, April 2012 Image Quality CT Imaging Quality Assurance AAPM Report 74 AAPM Report 142 Typical Tests Scale and distance accuracy Low contrast resolution Spatial Resolution Uniformity and noise

Image Quality CatPhan (done annually) Image uniformity Low contrast High resolution Image quality checks for uniformity, heterogeneity, CT number accuracy, resolution, artifacts Now You Are a CT Physicist!

But Wait, Even TG142 Has Imaging QA!

Image Dose Previously discussed assess dosimetric aspects of defined imaging protocols Accuracy of CT Numbers If IGRT technique is used for re-computation of dose, and for general purposes, the CT numbers must be 1) determined and set accordingly, and 2) measured routine to insure stability. Typical approach: electron density phantom, differing imaging parameters according to defined imaging protocols.

Physics Effort for QA Procedures Bissonnette JB et al. Quality assurance for image-guided radiation therapy utilizing CT-based technologies: A report of the AAPM TG-179. Medical Physics, Vol. 39, No. 4, April 2012 End-to-End Test Verify the fidelity and accuracy of the system

AAPM Practice Guideline 2.a Staffing: AAPM Practice Guideline 2.a

Staffing: AAPM Practice Guideline 2.a Staffing: AAPM Practice Guideline 2.a

Commissioning and QA: AAPM PG 2.a Commissioning and QA: AAPM PG 2.a Procedures Described Customer acceptance procedures TPS configuration and connectivity OIS integration and connectivity Routine QA baselines IGRT QA program documentation Safety and interlocks Image contrast Spatial resolution Image scale Image uniformity Imaging-treatment isocenter coincidence Table positioning, repositioning Image dose

ACR-AAPM Technical Standard Medical Physics and IGRT ACR-AAPM Technical Standard Medical Physics and IGRT

ACR-AAPM Technical Standard Medical Physics and IGRT ACR-AAPM Technical Standard Medical Physics and IGRT End-to-end Test

ACR-AAPM Technical Standard Medical Physics and IGRT ACR-AAPM Technical Standard Medical Physics and IGRT

Summary CT-based IGRT devices are imaging devices Quality assurance should proceed based on the specific CT technology Quality assurance comprises: Mechanicals, Interlocks/Safety, Image Quality and Fidelity, Image Dose, Communications/Data Transfer with/to Other Systems Physics effort and instrumentation are required Summary Key AAPM and Other Guidance Documents AAPM Task Group 179. Quality assurance for image-guided radiation therapy utilizing CT-based technologies. 2012 AAPM Task Group 75. Murphy MJ, Balter J, Balter S, et al. The management of imaging dose during image-guided radiotherapy: report of the AAPM Task Group 75. Med Phys. 2007;34(10):4041-4063. AAPM Task Group 104. The role of in-room kv X-Ray imaging for patient setup and target localization. 2009 AAPM Task Group 142. Quality assurance of medical accelerators. Klein et al. Med. Phys. 36(9): 4197-4212. 2009 ACR AAPM Technical Standard for Medical Physics Performance Monitoring of Image-guided radiation therapy (IGRT). Revised 2014 (Resolution 36) ACR-ASTRO Practice Parameter for Image-Guided Radiation Therapy (IGRT). Revised 2014 (CSC/BOC)