Table of Contents. 1. Introduction. 2. Authorisation Process. 3. X-Ray Facility Design Considerations and Minimum Requirements. 3.

Similar documents
SECTION RADIATION SHIELDING ENCLOSURE Low Intensity Radiation Protection

Types of Radiation Units

Radiation Shielding Glass RD 30 RD 50

Function of this control

Silent Gliss. Room Divider/Cubicle/ Shower Rail Systems

Premier Inn, Wilson Street, Middlesbrough, TS1 1AE. Site progress report. February 2017

Department of Radio Diagnosis

G. UNDERWRITERS LABORATORY, INC.: 1. UL 325: Electrical Door, Drapery, Gate, Louver, and Window Operators and Systems

Fire Inspection and Code Enforcement, 7 th Edition. Chapter 6 Building Construction: Components Inspector II

CARESTREAM DIRECTVIEW Max CR System

2 / 1. SUMMARY public, should be well documented, and should be reviewed on a regular basis to determine whether it continues to meet the operational

Third WHO Global Forum on Medical Devices Geneva May 2017

Compliance Guidance for FLUOROSCOPIC QUALITY CONTROL 2 nd Edition

Basic Information for Obtaining Permits, Submitting Plans, Scheduling Inspections, etc.

Ticking Form - Lodge - 1 -

MNL, SEL DIMENSIONS, SITE AW PLANNING GUIDE ARTWORK SIZE A SHEET 1 OF 1

Homebuyers Survey. Final Report. Mr Client. A Home, A Street, A Town. 3rd July 2014

Building Construction Occupations

CITY OF RICHMOND PINE CAMP COMMUNITY CENTER ROOF REPLACEMENT PROJECT RCS PROJECT NO

Notification of a Body pursuant to article 18 of the Construction Products Directive 89/106/EEC

MGT undercut THE INVISIBLE GLASS FACADE FIXING SYSTEM

COLUMN SPACING TRUCK ACCESS/DOORS EAVE HEIGHT UTILITIES PARKING & DRIVES

For more information - A guide to the use of lead for radiation shielding

SECTION ALUMINUM WINDOWS

Demolition Plan A1 A3. Scale: 3/32" = 1'-0" Arch. Men N.I.C. Elev. Lobby. Stairs N.I.C. N.I.C. Women N.I.C. Issued for Tender.

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May Dose Reference Levels

HD-Swing Series 4000 Heavy Duty Aluminum Swing Door System with Electric Operator

Radiography Curriculum Analysis

SITE ANALYSIS, DESIGN BRIEF AND CASE STUDY

MRI Safety & Site Planning. Lawrance Yip Senior Radiographer, DR, QMH 16 June 2007

SCREEN PORCHES 101 DEVELOPMENT RESOURCES, INC Permit Requirements

Basement Renovation Guide For Residential Dwellings

Streamline workflow for higher productivity and patient throughput. DirectView. Classic CR System

Movibloc Mobile Shelving / Mobile Lockers

ASTATT1 Prepare to carry out air tightness testing

Reference Value Assessment for DAP Measured with a Transmission Ion Chamber in Gastrointestinal Tract Examinations

Asbestos Management Program April 27, 2017

Asbestos Management Plan

Designed & Engineered by. Protect health and safety on your next project involving ionizing radiation (X-Ray or Gamma)

PORTA-FAB Cleanroom Wall Systems

Safety and Health Office phone , fax GUIDELINES FOR USE OF NUCLEAR MOISTURE/DENSITY GAUGES

HOT SPRINGS FIRE DEPARTMENT EVALUATION OF DOWNTOWN LEGACY STRUCTURES Savoy Hotel Thompson Building Dugan Stewart Building May 16, 2012

SOLID FUEL HEATER APPLICATION FOR PROJECT INFORMATION MEMORANDUM AND /OR BUILDING CONSENT Section 33 or 45, Building Act 2004

The CLAYTEC clayboard is a dry lining board for internal use made of reed-reinforced

WORK PROCESSES SCHEDULE ACOUSTICAL CARPENTER SPECIALIST O*NET/SOC CODE: RAIS CODE: 0861

FEATURED PRODUCT BROCHURE

FLUOROSCOPIC SAFETY AND PATIENT AND OPERATOR RADIATION DOSE REDUCTION: CREATING A FOCUSED FLUOROSCOPIC CURRICULUM FOR RADIOLOGY RESIDENTS

LIMITED POTENTIALLY HAZARDOUS MATERIAL BUILDING INSPECTION REPORT AND ASBESTOS BUILDING INSPECTION REPORT. for

Acoustic Design Criteria

RADIATION SAFETY MANUAL. RADIATION-PRODUCING DEVICES (Academic)

SPACE TYPE: OUTSIDE PARKING (STRUCTURE)

STREET LIGHTING. Springville City Power standard street lights will be installed in accordance with Springville City Power's Line Extension Policy.

Animal Facility Biosafety Level 3 Checklist (date: April 16, 1998)

SPECIFICATION FOR THE DEVELOPMENT OF LPA CLEAN ROOM INFRASTRUCTURE Description of work

1.1 Fruits shall be treated by using one of the following schedules: 2. Approval of facilities for pre-cooling, refrigeration and fumigation:

PATTAYA POSH. A chance to own the best location in North Pattaya. North Pattaya Road, Pattaya, Thailand

Type II Inspection Worksheet

Interior Refurbishment Method Statement

Syllabus. Construction Technology 100 Building Fundamentals

SECTION ALTERNATES

Selecting partitions

Specialists in Acoustic Flooring and House to Flat Conversions

Element Z General Design Requirements Room Standards

Building Systems and Design Elements included in SliDeRulE

Knauf Safeboard. X-ray partitions. Lead-free X-ray shielding plasterboard. 100% lead-free board. Economical radiation protection.

The Ashland Project. Total Area: 1,778 Sq.Ft. 3 Bedroom, 2 Bath, 2 Car Garage

TOWN OF ROTTERDAM RESIDENTIAL BUILDING PERMIT APPLICATION Ext. 395 Needed to Obtain Permit:

DoseWise Portal. Take control of dose management. across your organization

OPERABLE PARTITIONS. Signature

Level 2 NVQ Certificate in Wall and Floor Tiling (Construction) Qualification Specification

Minnesota Healthcare Engineers Association, Fall Conference. Bob Dehler, P.E. Engineering Program Manager September 14, 2017

Take control of dose management

Interim Guidance on the Regulation of In-situ Disposals of Radioactive Waste and Residual Radioactive Contamination on Nuclear Authorised Premises

CASE STUDY OF THE SOUND REDUCTION OF VARIOUS RESIDENTIAL GLAZING TREATMENTS

Regulation respecting X-ray Safety - made under the Occupational Health and Safety Act

Get in front of the Q SBD and Approved Document Q

Stainless steel. 1.6 Stainless steel IP 69 K. Premium Panel, protection category IP 69K

SECTION UNIT PRICES. A. PART A and DIVISION 1 of PART B are hereby made a part of this SECTION.

Qualifying Process for Handyman Online Qualifying Questions. Name: Street Address: City: State: Zip Code. Home Phone: Cell Phone: Address:

NS&I Glasgow [Cross] Re inspection Asbestos Survey Report Copy 1. February Carillion G / B070211LM1

BUILD A SHOWER CUBICLE

Practice Advisory 14. Long span steel roof trusses: welding and section checks needed for some

SECTION LAGUNA SERIES TOP HUNG SLIDING DOOR SYSTEM

Radiation Protection Adviser (RPA) Register

BUILDING PERMIT GUIDELINES - RESIDENTIAL

Land Reclamation Site Siting and Application Guidelines

Property Condition Assessment

GERMICIDAL ULTRAVIOLET ABOUT US. Since 1963, Atlantic Ultraviolet. Ultraviolet is a unique and rapid method. of disinfection. It utilizes germicidal

Asbestos Consulting Radon Measurements Lead Assessments Indoor Air Quality Testing Mold Investigations. Project Number: Page 1 of 22

r e s i d e n t i a l o u t s i d e v i l l a g e c e n t e r

CONSTRUCTION TRADES LABORATORY September 30, 2010

Technical Features Sound Construction Design Aid For Office Acoustics

ADMINISTRATION CHAPTER 1

Tender Document and Specification

ONLY ONLY BUILDING AND PARKING ENVELOPE RAMP STAIRS NEW SURFACE PARKING PROPOSED HOTEL PROPOSED TWO STORY PARKING STRUCTURE NEW SURFACE PARKING

Office Interiors. Office partitions, furniture, shelving, storage, space planning and refurbishment.

WM2012 Conference, February 26 March 1, 2012, Phoenix, Arizona, USA

Optimize efficiency with a universal system.

Transcription:

Table of Contents 1. Introduction 2. Authorisation Process 3. X-Ray Facility Design Considerations and Minimum Requirements 3.1 Location 3.2 Room Size 3.3 Walls 3.4 Viewing Window Requirements 3.5 Change Cubicles 3.6 Access Points 3.7 Public Protection 4. X-Ray Room Accessories Pg 1 1 2 2 2 3 3 4 4 4 5 pg i NRPA_2014

1. Introduction The National Radiation Protection Authority is mandated by the Atomic Energy and Radiation Protection Act, Act 5 0f 2005 to regulate the use of radiation generating equipments. This Act is further supported by the Radiation Protection and Waste Disposal Regulations. The Act requires that a notification be made by any person or institutions that wish to engage in the use of radiation generating equipment. Such notification has to be accompanied by an application and subsequent authorisation of the premises and equipment that will be used for the intended practises, including a diagnostic x-ray facility. Both the Act and the regulations are not specific in respect to the design and requirements of an x-ray diagnostic facility. However the authorisations are issued subject to conditions made by the issuing authority, which includes design and safety requirements that will ensure that protection against ionising radiation is guaranteed. This document sets out minimum requirements for applicants and users in the establishment of a diagnostic x-ray facility and also forms the basis along which an inspector will conduct an inspection in consideration of a premises license. 2. Authorisation Process The authorisation process and requirements apply to any person who uses, modifies or disposes of an electronic product, including x-ray generating equipment or premises on which it is housed. The chronological order for an authorisation should be as follow: a) The person/institution that intends to introduce or modify a practise using ionising radiation must declare his intention on prescribed application form obtainable form the National Radiation Protection Authority (NRPA). b) The application form should include all necessary information such as facility layout and accompanied by a Radiation Management Plan (RMP) as outlined in document NRPA_AG_01. The application will be reviewed by the NRPA and followed by NRPA_2014 pg 1

consultations to reach agreement on details of the application, including premises layout. c) Upon approval granted by the authority for a premises license construction may start. d) An inspection must be conducted mid-way through the construction phase to ascertain that agreed plans are followed and implemented. e) Following the satisfactory safety assessment of the physical facility, the installation of the equipment may be started. f) As soon as the equipment has been installed, the applicant must notify the NRPA to conduct a generic safety assessment. g) It is only after a license has been issued that the electronic product/x-ray equipment may be commissioned and brought into use for the intended practice. 3. X-Ray Facility Design Considerations and Minimum Requirements 3.1 Location a. X-ray rooms should preferably be on the ground floor b. It is however acceptable that the x-ray room be a levelled floor provided that evidence indicates that the structural design can support the weight of the x-ray unit. c. For x-ray rooms on levelled floors, there must be proper shielding for above and below the room. (typically 150 mm concrete floor of 2.3 g/cm 3 and 100mm concrete ceiling, if the space above is occupied ). 3.2 Room Size (including control panel area) The x-ray room layout should at minimum be: a. 20 m 2 for a general radiographic room b. 30m 2 for a fluoroscopy room pg 2 NRPA_2014

c. 35m 2 for CT room d. a parking space of 3m² must be added if patients arrive in bed. e. 8m 2 for mammography 3.3 Walls The walls of the x-ray room should at minimum be constructed with: a. 230 mm brick (2.3 g/cm 3 ) or b. 2 mm leadsheet sandwiched between partitioning alternatively for CT rooms 2.5 mm leadsheet sandwiched between partitioning or c. 115 mm brick (2.3 g/cm 3 ) with 4 mm barium plaster (1 part coarse barium sulphate, 1 part fine barium sulphate, 1 part cement). d. 115 mm brick or 1mm leadsheet wall protection is sufficient for mammography, dental and fixed mobile x-ray units. e. the ceiling height should be at least 2.5m f. Windows are acceptable only if positioned so that nobody outside can be exposed to radiation (or look into the room). A window height of 2m above ground level is recommended. g. Control cubicle wall should be protected with the same wall protection as above and should have a height of atleast 2.5m. 3.4 Viewing window requirements a. The lead glass window should be at least 35 cm x70 cm to give full view of the installed x-ray unit. b. A protective lead screen glass for the control console inside the X-ray room should contain 0.5 2mm of lead (depending on its distance from the patient, the main source of secondary radiation). c. The lead glass and protective material (wall) must overlap each other by at least 25mm d. The x-ray system control panel should be at least 1 m from the NRPA_2014 pg 3

open end nearest to the examination table 3.5 Change Cubicles a. Protection for wall and doors leading into the x-ray room from the change cubicles must be the same as provided for the x-ray room, control cubicle and entrance door above. 3.6 Access points a. Access points must have necessary radiation shielding. Doors should be lined with at leadsheets of at least 2mm. CT room doors must be lined with 2.5mm lead sheets, while 1 mm leadsheet should suffice for dental or fixed mobile x-ray rooms. b. Access doors should preferably be of the sliding type with at least 1.5m clearing space and 100mm overlap on each side. c. Radiation warning signs/notices must be displayed at all entrances into the x-ray room. d. A red light coupled to the x-ray unit exposure switch must be installed above the entrance door. e. The darkroom must ideally have an entrance from the control panel area through a light-tight door or a light trap with double doors. f. Doors leading into the x-ray room must be lockable from the inside to prevent inadvertent entry during examinations. 3.7 Public Protection a. The dose rate in the reception area, adjacent offices, waiting area should be comparable with background levels even when the x-ray tube primary beam is directed to the area of interest. b. The equipment should be positioned so that the primary radiation beam is not directed at the operator s console, windows or doors. pg 4 NRPA_2014

4. X-Ray Room Accessories The following are accessories that must be in an x-ray room as part of the radiation protection requirements a. Gowns, aprons and thyroid protectors made of a material (such as vinyl) which contains lead should be part of protective gear in the x-ray room. b. Aprons should be equivalent to at least 0.25 mm of lead if the X Ray equipment operates up to 100 kv and 0.35 mm of lead if it operates above 100 kv. c. Aprons may be the wrap-around type or the height-relieving type. d. Gauntlets are heavy gloves. They have limited value because they are difficult to use and should therefore only be used where appropriate. e. Additional protective devices should be available in fluoroscopy and interventional radiology rooms which include: Ceiling suspended protective screens. Protective lead curtains mounted on the patient table. Protective lead curtains for the operator if the X- Ray tube is placed in an over couch geometry and if the radiologist must stand near the patient. f. Individual dose monitoring should be undertaken for workers who are normally exposed to radiation in controlled areas. g. Additional accessories such as lead stand, lead strips, exposure charts and casette positioning devices should be available. NRPA_2014 pg 5