Analysis of TCM injections using the Agilent 1290 Infinity LC system

Similar documents
Technical Overview. Author. Abstract. Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany

Disulfide Linkage Analysis of IgG1 using an Agilent 1260 Infinity Bio inert LC System with an Agilent ZORBAX RRHD Diphenyl sub 2 µm Column

Quality-by-Design-Based Method Development Using an Agilent 1290 Infinity II LC

Method Translation in Liquid Chromatography

Application Note. Author. Abstract. Small Molecule Pharmaceuticals. Sonja Krieger Agilent Technologies, Inc. Waldbronn, Germany

Fast and High-Resolution Reversed-Phase Separation of Synthetic Oligonucleotides

Heart-cut 2D-LC/MS approach for pharmaceutical impurity identification using an Agilent 6540 Q-TOF LC/MS System

Application of Agilent AdvanceBio Desalting-RP Cartridges for LC/MS Analysis of mabs A One- and Two-dimensional LC/MS Study

Confident Pesticides Analysis with the Agilent LC/Triple Quadrupole and TOF/QTOF Solutions

Application Note. Author. Abstract. Small Molecule Pharmaceuticals & Generics. A.G.Huesgen Agilent Technologies, Inc. Waldbronn, Germany

Peptide Mapping: A Quality by Design (QbD) Approach

Agilent AdvanceBio SEC Columns for Aggregate Analysis: Instrument Compatibility

Fast and Efficient Peptide Mapping of a Monoclonal Antibody (mab): UHPLC Performance with Superficially Porous Particles

Size Exclusion Chromatography of Biosimilar and Innovator Insulin Using the Agilent AdvanceBio SEC column

PLRP-S Polymeric Reversed-Phase Column for LC/MS Separation of mabs and ADC

Application Note. Author. Abstract. Biotherapeutics and Biologics. Sonja Schneider Agilent Technologies, Inc. Waldbronn, Germany

Application Note. Author. Abstract. Biopharmaceuticals. Verified for Agilent 1260 Infinity II LC Bio-inert System. Sonja Schneider

Pharmaceutical LC/MS Solutions from Agilent Technologies

Characterize Fab and Fc Fragments by Cation-Exchange Chromatography

ADVANCE ACCURACY AND PRODUCTIVITY FOR FASTER ANALYSIS

High-throughput and Sensitive Size Exclusion Chromatography (SEC) of Biologics Using Agilent AdvanceBio SEC Columns

Development of Ultra-fast ph-gradient Ion Exchange Chromatography for the Separation of Monoclonal Antibody Charge Variants

Controlling Agilent 1200 Series Rapid Resolution LC systems through Waters Empower chromatography data software. Technical Overview.

Application Note. Author. Abstract. Environmental. Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany

N-Glycan Profiling Analysis of a Monoclonal Antibody Using UHPLC/FLD/Q-TOF

Agilent 1220 Infinity II LC GET ON THE FAST TRACK TO HIGHEST EFFICIENCY AT A PRICE YOU CAN AFFORD

mab Titer Analysis with the Agilent Bio-Monolith Protein A Column

Fraction Analysis of Cysteine Linked Antibody-Drug Conjugates Using Hydrophobic Interaction. chromatography. Agilent 1260 Infinity II Bio-Inert System

Qualification of High-Performance Liquid Chromatography Systems

Agilent 6430 Triple Quadrupole LC/MS System

Cell Clone Selection Using the Agilent Bio-Monolith Protein A Column and LC/MS

ACQUITY UPLC. I-Class PLUS WHAT SEPARATES YOU FROM EVERYONE ELSE

Characterization of mab aggregation using a Cary 60 UV-Vis Spectrophotometer and the Agilent 1260 Infinity LC system

Application Note. Abstract. Authors. Introduction. Environmental

C196-E067A. Prominence UFLCXR. Shimadzu Ultra Fast Liquid Chromatograph

EPA Method 543: Selected Organic Contaminants by Online SPE LC/MS/MS Using the Agilent Flexible Cube

Improving Resolution and Column Loading Systematically in Preparative Liquid Chromatography for Isolating a Minor Component from Peppermint Extract

Peptide Mapping of Glycoprotein Erythropoietin by HILIC LC/MS and RP-LC/MS

Puzzled About LCMS? Sample. Sensitivity in Mass Spec Analysis. Prep. Adapting LC-UV. Optimizing and Maintaining Your Mass Spec LCMS

easy. HPLC has never been so Agilent 1120 Compact LC Our measure is your success.

APPLICATIONS TN Overview of Kinetex 2.6 µm Core-Shell Technology

Antibody Analysis by ESI-TOF LC/MS

Protein-Pak Hi Res HIC Column and HIC Protein Standard

Precise Characterization of Intact Monoclonal Antibodies by the Agilent 6545XT AdvanceBio LC/Q-TOF

Minimize Method Development Time for Pesticide Screening Using LC/MS Application Kit

Chem 321 Lecture 23 - Liquid Chromatography 11/19/13

Improved SPE for UPLC/MS Determination of Diquat and Paraquat in Environmental

HIC as a Complementary, Confirmatory Tool to SEC for the Analysis of mab Aggregates

Concentration of Human Hormones in Drinking Water Using Solid Phase Extraction and Analysis by High Performance Liquid Chromatography

A Modular Preparative HPLC System for the Isolation of Puerarin from Kudzu Root Extracts

Technical Note. Introduction

Laboratory Reagents. for life science research.

Analysis of Monoclonal Antibody Digests with the Agilent 1290 Infinity 2D-LC Solution Part 2: HILIC RPLC-MS

Analysis of Intact and C-terminal Digested IgG1 on an Agilent Bio MAb 5 µm Column

Overview of preparative HPLC. Analytical Technologies Limited

Fast mass transfer Fast separations High throughput and improved productivity Long column lifetime Outstanding reproducibility Low carryover

Characterization of Glycosylation in the Fc Region of Therapeutic Recombinant Monoclonal Antibody

Optimizing the Purification of a Standard Chiral Compound Utilizing a Benchtop, Multi-Purpose, Semi-Preparative to Preparative HPLC System

Lifetime stability of size exclusion chromatography columns for protein aggregate analysis

An Automated System for At-Line Process Analysis of Biopharmaceutical Fermentation Reactions

INTERNATIONAL PHARMACOPOEIA MONOGRAPH ON ARTEMETHER AND LUMEFANTRINE CAPSULES REVISED DRAFT FOR DISCUSSION

Genotoxicity is the property of a compound

The Theory of HPLC Gradient HPLC

Development of Analysis Methods for Therapeutic Monoclonal Antibodies Using Innovative Superficially Porous Particle Biocolumns

MS Certified Vials. Pre-cleaned and certified vials for mass spectrometry

A universal chromatography method for aggregate analysis of monoclonal antibodies

ZORBAX StableBond HPLC Columns

Agilent 6000 Series LC/MS Solutions CONFIDENCE IN QUANTITATIVE AND QUALITATIVE ANALYSIS

Fast Preparative Column Liquid Chromatography (PCLC)

RP-HPLC METHOD FOR QUANTITATIVE ESTIMATION OF GLATIRAMER ACETATE FOR INJECTION IN PHARMACEUTICAL DOSAGE FORMS

ISSN India; g,secunderabad. Abstractt. a flow rate. of 1ml/min. di hydrogen. which acts. and chronic. including minimize (5) Figure

Exploring Extra Sensitivity Using ionkey/ms with the Xevo G2-XS Q-Tof HRMS for Small Molecule Pharmaceutical Analysis in Human Plasma

The clearly better choice

Available Online through (or) IJPBS Volume 2 Issue 3 JULY-SEPT Research Article Pharmaceutical Sciences

Bivalirudin Purification:

Improved Peptide Maps Using Core-Shell Media: Explaining Better Biopharmaceutical Applications With Kinetex C18 Columns

CQA Assessment by Peptide Mapping Using LC/MS with an AdvanceBio Peptide Plus Column

Developing Robust and Efficient IEX Methods for Charge Variant Analysis of Biotherapeutics Using ACQUITY UPLC H-Class System and Auto Blend Plus

EtG / EtS Analysis UCT INNOVATION THROUGH CHEMISTRY FORENSICS

Maximizing Chromatographic Resolution of Peptide Maps using UPLC with Tandem Columns

Analysis of the bioactive components from different growth stages of Fritillaria taipaiensis P. Y. Li

performance and throughput

Analysis of Monoclonal Antibody Digests with the Agilent 1290 Infinity 2D-LC Solution

Automated Online SPE for LC/MS/MS Analysis of Trace Organic Contaminants in Water Using the Agilent 1290 Infinity Flexible Cube Module

MAbPac SCX-10 Columns

Agilent solutions for contract research and manufacturing organizations

Accucore. Ultimate Core Performance LC Column Technology to Maximize Your Investment. Dave Jarzinski. October 2011

Fundamentals and Techniques of Preparative HPLC. Parto Zist Behboud Tel: 42108

HIGHER IMPACT LC/MS DIAGNOSTICS AGILENT IN VITRO DIAGNOSTIC MEDICAL DEVICES FOR CLINICAL LABORATORIES

Preparative HPLC is still the

Laboratory Water Quality Affects Protein Separation by 2D Gel Electrophoresis

Gas Chromatography Assignment Chem 543/443

High-Throughput, Simplified Protein Precipitation: A Novel 96-Well Format for Precipitation and Filtration EAS 2003

Thermo Scientific Peptide Mapping Workflows. Upgrade Your Maps. Fast, confident and more reliable peptide mapping.

HPLC to UPLC Method Migration: An Overview of Key Considerations and Available Tools

Validated Stability-indicating assay method for determination of Ilaprazole in bulk drug and tablets by high performance liquid chromatography

Liquid Chromatograph Liquid Chromatograph Mass Spectrometer. CL Series. for In Vitro Diagnostic Use

Determination of Adenovirus pviii (31K) Concentration for Estimation of the Empty Capsid Concentration of the Adenovirus Reference Material (ARM)

Analysis of Suspected Flavor and Fragrance Allergens in Cosmetics Using the 7890A GC and Capillary Column Backflush Application

Transcription:

Analysis of TCM injections using the Agilent 1290 Infinity LC system Application Note Traditional Chinese Medicine Authors Zhixiu Xu Agilent Technologies Shanghai Shanghai 200131 China mau 400 350 300 250 200 150 100 50 0 0 2 4 6 8 10 min Abstract An ultra high performance liquid chromatography (UHPLC) method using an Agilent 1290 Infinity LC system with an Agilent 1290 Infinity Diode Array Detector (DAD) was developed for the separation of Shenmai injections. Good separation of complex systems was achieved by connecting two columns: an Agilent ZORBAX Eclipse Plus C18 column (2.1 mm 100 mm, 1.8 µm) and an Agilent ZORBAX SB C18 (2.1mm 50 mm, 1.8 µm) column. These were chosen for their separation efficiency and good chromatogram peak shape. The results demonstrated that this analytical method was simple, sensitive and reliable for rapidly analyzing TCM injections, and is helpful in comprehensively evaluating the quality of the TCM injection.

Introduction Traditional Chinese Medicine (TCM) injections, also known as Chinese herbal injections, have been used for several decades in China. However, evidence of adverse drug reactions (ADR) through the years, have attracted attention to TCM injections. The ADRs have several causes. 1 The consequences of these ADRs have led to discussions regarding the safety of TCM, because many complex components of TCM injections not yet been identified. However, TCM injections have the advantage of higher bioavailability, faster therapeutic effects, and low cost. Eliminating TCM injections is not the best solution for reducing health issues caused by ADR. Therefore, government agencies and manufacturers have held several conferences and summits discussing how to improve safety standards, through better component understanding and stricter quality control. A Shenmai injection composed of Radix Ginseng Rubra (red ginseng) and Radix Ophiopogonis (maidong) is widely used in China to cure the syncope caused by Yin and Qi deficiences and coronary heart disease. Some ADR reports related to Shenmai injection show an urgent need to improve the quality control level of these injections. More research regarding component elucidation is necessary. UHPLC reduces the issues involved with the separation of such complex matrices. Research on Radix Ophiopogonis has shown that the main active components are isoflavoids (for example, Ophiopogonone A-F) 2 and saponins (for example, Ophiopogonin D). The methods discussed in several scientific papers published for the analysis of ginseng 3, ophiopogonis and Shenmai injections require long analysis times 2,4. In this Application Note, a fast analysis method with good separation of the complex components of Shemmai injections was achieved with an Agilent 1290 Infinity UHPLC system. Since research on the therapeutic effects of these injections should be based on good separations, the method published here is applicable. It is a method for improved quality control and for high throughput research. Experimental Equipment: For development of the UHPLC method, an Agilent 1290 Infinity UHPLC system with the following modules was used: Agilent 1290 Infinity Binary Pump with integrated vacuum degasser (p/n G4220A) Agilent 1290 Infinity High Performance Autosampler (p/n G4226A) Agilent 1290 Infinity Thermostatted Column Compartment (p/n G1316C) Agilent 1290 Infinity Diode Array Detector SL with Max-Light flow cell (1.0 µl volume, 10 mm path length) (p/n G4212A) Agilent ChemStation B.04.02 for data acquisition and evaluation Agilent ZORBAX RRHD Eclipse Plus C18 column 2.1 mm 100 mm, 1.8 µm (p/n 857700-902) Agilent ZORBAX RRHD SB C18 column 2.1 mm 50 mm, 1.8 µm (p/n 959758-902) Two columns were connected via a capillary 50 mm 0.12 mm (p/n G1316-87312) Chemicals and materials HPLC-grade acetonitrile, acetic acid and ammoniom acetate were purchased from J&K Chemical LTD. Pure water was obtained from a Millipore pure water system. The reference standards of 20(R)- Propanaxadiol, 20 (R)-Propanaxatriol, Ginsenoside Re, Rg1, Rb1, Rc were purchased from the National Institute for the Control of Pharmaceutical and Biological Products (NICPBP). Ginsenoside Rb2, Rb3, Rg3, Rh2, Rf, F11 were kindly provided by the customer. All the standards were dissolved in methanol to make 1 mg/ml solutions. Shenmai injection and ophiopogonis extraction was provided by customer. Sample preparation A TCM injection sample was taken from the ampoule and filtered through a 0.22 µm membrane filter for injection. 10 ml of n-butyl alcohol saturated with water was added to 2 g of ophiopogonis and stored overnight. Nitrogen was used to remove some of the solvent to obtain a volume of 5 ml. Prior to use, the sample was filtered through a 0.22 µm nylon membrane filter. 2

Chromatographic conditions Mobile phase: A = water, B = acetonitrile Flow rate: 0.6 ml/min Gradient: 20% B; 0 3 min, 20% to 40% B; 3 8 min, 100% B Column temperature: 55 C DAD detector: 254 nm, 202 nm; 360/100 nm Peak width: 0.006 min (40 HZ) Slit: 4 nm Injection: 0.5 µl Autosampler injection with 5 s needle wash with acetonitrile Results and Discussion First an Agilent ZORBAX RRHT SB C18 column (3.0 mm 150 mm) was used for method optimization. Comparing the chromatograms acquired from different wavelengths (202 and 254 nm), it was determined that the 254 nm channel resulted in less peaks. This showed that most of the components have no UV absorbance at 254 nm. Based on previous research 3, saponins only have absorbance at low wavelengths. So 202 nm was selected as the data collecting channel for the remainder of the sample. A mass spectrometer is best as an additional detector for further confirmation if no other detector has been specified for qualitative analysis 5,6. The structure of ginsenoside Re and Rg1 have very similar retention behavior in reverse phase chromatography. It is critical, therefore, that the method is able to separate them. In this study, separation of ginsenoside Re and Rg1 was achieved but with some tailing peaks. When the ZORBAX RRHT Eclipse Plus C18 column (3.0 mm 150 mm) was tested for the separation, good peak shapes were achieved. Good separation was not achieved for some difficult saponin pairs such as ginsenoside Rg1 and Re. In order to achieve good peak shape but also satisfactory separation, two columns were connected forming a hybrid column. The results of the hybrid column compared to the two separate columns are listed in Table 1. From these results, it can be seen that no distinct efficiency was lost when the hybrid column was used. Agilent ZORBAX RRHD Agilent ZORBAX RRHD Eclipse Plus Agilent ZORBAX RRHD SB C18, SB C18, 2.1 mm 50 mm C18, 2.1 mm 100 mm 2.1 mm 50 mm connected with ZORBAX Eclipse RRHD Plus C18, 2.1 mm 100 mm Flow rate Flow rate Flow rate 0.2 ml/min 0.4 ml/min 0.2 ml/min 0.4 ml/min 0.2 ml/min 0.4 ml/min Pressure (bars) 132 262 258 483 374 718 RT Plates RT Plates RT Plates RT Plates RT Plates RT Plates Peak 1 0.890 2799 0.444 2415 1.751 9223 0.890 8057 2.563 15053 1.306 14159 Peak 2 1.276 4291 0.635 3698 2.574 12513 1.306 11776 3.753 19978 1.911 18676 Peak 3 2.725 7313 1.351 7870 5.711 16257 2.877 19624 8.249 23649 4.242 28078 Peak 4 5.841 8400 2.902 9081 12.945 17612 6.540 21667 18.380 24747 9.505 29618 Isocratic sample: p/n 01080-68704 Peak 1: Dimethyl phthalate Peak 2: Diethyl Phthalate Peak 3: Biphenly Peak 4: o-terphenyl Conditions: Isocratic: 65% acetonitrile in water Data rate: 40 HZ, Ambient temperature: (23 C) Detection: 254, 202 nm, 360/100 nm, slit: 4 nm Injection volume: 1 µl Table 1 Column test comparing two separated columns to two connected columns. 3

Figure 1 shows the chromatogram of the Shenmai injection. These results show good separation and fast analysis. Compared to a previously developed method that requires 100 minutes, the method developed in this Application Note is quicker, requiring only 11 minutes. The "difficult" ginsenoside pairs Re and Rg1 also achieved good separation. Ophiopogonis extract was analyzed under the same optimized conditions (Figure 2) 7. This analysis proved that ophiopogonis extracts contain compounds that elute later than the components of red ginseng. However, when ophiopogonis and red ginseng were combined through processing, the components changed and in the corresponding position of the Shenmai injection, some peaks got smaller and some peaks disappeared. mau 400 350 300 250 200 150 100 50 0 Re Rg1 Rf Rb1 0 2 4 6 8 10 Figure 1 Chromatogram of Shenmai injection. mau Rc Rb2 F11 Rg3 min 600 500 400 300 200 100 0 0 2 4 6 8 10 min Figure 2 Chromatogram of extractions from ophiopogonis. 4

Five runs were performed to achieve good retention time and peak area (height) reproducibility (Table 2). The system is ideal for fast analysis of TCM samples with complex metrices. Figure 3 shows an overlaid chromatogram of five runs of Shenmai injection. The ultra high performance of the Agilent 1290 Infinity LC system is proven by the consistency of the results. In these analyses the temperature was evaluated from 55 C to ambient. The results showed that 55 C can reduce the backpressure, produce faster analyses, and provide good peak separation. Flow rate is another parameter evaluated in this study. The flow rate was set from 0.2 ml/min to 0.6 ml/min. The sub-2-µm particle size of the packing material has the advantage of retaining good performance even with higher flow rates. A flow rate of 0.6 ml/min was chosen to leverage fast speed and MS compatibility. RSD (%) Compounds Retention time Peak height Peak area Re 0.042 1.36 5.94* Rg1 0.053 1.49 0.54 Rf 0.038 2.64 0.93 Rb1 0.047 1.29 0.64 Rc 0.045 2.47 0.37 Rb2 0.042 1.59 0.51 F11 0.026 0.96 1.22 Rg3 0.019 0.80 1.40 Table 2 Statistical results of repeatability. Norm. 800 700 600 500 400 300 200 100 0 0 2 4 6 8 10 min Figure 3 Overlaid chromatograms of five TCM injection sample. 5

Conclusion A UHPLC method was established for the separation of Shenmai injection. This method is fast with high efficiency, obtaining both good separation and peak shape. This method can also be used for the analysis of red ginseng and ophiopogonis. This study showed that the connection of two columns did not decrease efficiencies through the capillary connection. The validation indicated that the developed method was sensitive, reliable and rapid. In addition, column temperature and flow rate were also tested to optimize these parameters. The method developed here can be used for high-throughput analysis when therapeutic effect research is in process. Further works will be done, based on the current separation method to obtain identification of additional components. References 1. Kunmei Ji, Jiajie Chen, Meng Li, Zhigang Liu, Lixin Xia, Chunbo Wang, Zhasngke Zhan, Xuli Wu, Comments on serious anaphylaxis caused by nine Chinese herbal injections used to treat common colds and upper respiratory tract infections. Regulatory Toxicology and Pharmacology 55 (2009) 134 138. China 2. Min Ye, Dean Guo, Guan Ye and Chenggang Huang, Analysis of Homoisoflavonoids in Ophiopogon japonicus by HPLC-DAD-ESI-MSn, J Am Soc Mass Spectrom 2005, 16, 234-243 4. Fan Xiaohui, Wang Yi, Cheng Yiyu, LC/MS fingerprinting of Shenmai injection: A novel approach to quality control of herbal medicines, Journal of Pharmaceutical and Biomedical Analysis 40 (2006) 591-597 5. Edgar Naegele, Agilent MassHunter - Fast, computer-aided analysis of LC/ESI-TOF data from complex natural product extracts Part 2: Comparison of Agilent 6210 TOF data from different, biological origin with the Mass Profiler in MassHunter software, Agilent Technologies publication 5989-6076EN 6. Edgar Naegele, Analysis of complex natural product extract from ginseng: Part I: Structure elucidation of ginsenoside by rapid resolution LC-ESI TOF with accurate mass measurement Agilent Technologies publication 5989-4506EN Part II: Structure elucidation of ginsenosides by high resolution ion trap LC/MS, Agilent Technologies publication 5989-4705EN Part III: Species of differentiation of ginseng plants and authentication of ginseng products by LC/MS, Agilent Technologies publication 5989-5493EN 7. Bai Jing, Yu Zhiguo, HPLC fingerprint analysis of Radix ophiopogonis Chinese Traditional Patent Medicine, 29 (1), 2007, 12 13 3. Fuzzati N., "Analysis methods of ginsenosides." J. Chrom. B1-2, 812, 114 133, 2004. 6

7

www.agilent.com/chem/tcm Agilent Technologies, Inc., 2010 February 1, 2010 Publication Number 5990-5254EN