Analysis of an OTC Multivitamin Tablet Using X-ray Fluorescence Spectroscopy and Atomic Absorption Spectroscopy

Similar documents
S2 RANGER LE: Analysis of Light Elements in Cement, Slags and Feldspar

S8 TIGER Series 2. Lab Report XRF 144. Innovation with Integrity. Accurate Quality Control of Tool Steels XRF

ElvaX Light for Petrochemical analysis

Analysis of Trace Metals in Tobacco and Tobacco Ash using Atomic Emission Spectroscopy

S4 T STAR Biopharmaceutical production: Rapid nutrient and contaminant analysis of cell culture media by TXRF spectroscopy

National Food Safety Standard

Susan Miller Brick Industry Association. John Sanders National Brick Research Center

DETERMINATION OF ELEMENTS IN DRINKING WATER AS PER BUREAU OF INDIAN STANDARDS 10500, & USING THE AGILENT 5100 ICP-OES

Evaluation of Energy-Dispersive X-Ray Fluorescence for Mobile Analysis of Hazardous Metals in Transuranic Waste

ANALYTICAL PERFORMANCE OF A HANDHELD EDXRF SPECTROMETER WITH MINIATURE X-RAY TUBE EXCITATION

Portable Elemental Analysis Food safety and agriculture solutions. Innovation with Integrity. Handheld XRF

X-ray Fluorescence Spectrometry X-ray Fluorescence Spectrometry (XRF) is a non-destructive, quantitative technique for determining chemical

Analysis of Mineral and Heavy Metal Content in Beverages Using the Teledyne Leeman Labs Prodigy Plus ICP-OES

Core Analysis with the Tracer

VHG Labs. Standards and Supplies for Spectrochemical Analysis QUALITY. TRACEABILITY. DEPENDABILITY. WORLDWIDE

EU water analysis using the Thermo Scientific icap 7400 ICP-OES Duo

Review on Agriculture and Rural Development vol. 2. (1) ISSN

SIMPLIFIED MATERIALS ANALYSIS VIA XRF. Ravi Yellepeddi and Didier Bonvin Thermo Fisher Scientific (Ecublens) SARL, Ecublens/Switzerland

Process and quality control in the aluminum industry using ARL 9900 XRF-XRD integrated spectrometer

Advanced Materials Analysis with Micro-XRF for SEM

Analysis of steel and its alloys using the GB/T standard and an Agilent 5100 ICP-OES in dual view mode

ANALYSIS OF TCLP EXTRACTS BY X-RAY FLUORESCENCE

ICP-AES Determination of 15 Kind of impurity elements in the Vanadium-aluminum alloy

REPORT NUMBER REPORT DATE SEND TO ISSUE DATE Apr 18, Apr 18, 2017 RECEIVED DATE Apr 05, 2017

How to Increase Productivity in Mining with Automated Simultaneous XRF Analysis. Arkady Buman and Dr. Kai Behrens Bruker AXS Inc.

EDX 600. Atomic Absorption Spectrometry

EU Water Analysis Using the Thermo Scientific icap 7400 ICP-OES Duo

Increased Sample Throughput for ICP-OES Applied to U.S. EPA Method 200.7

Analysis of Pharmaceuticals by ICP-MS and WDXRF

ElvaX ProSpector in Exploration & Mining

METHOD 3040A DISSOLUTION PROCEDURE FOR OILS, GREASES, OR WAXES

Background The document will be revised for its five year review and updated per the committee agreed upon changes.

Wavelength-dispersive X-ray fluorescence spectrometer

Determination of iron content in iron oxide used for fireworks and firecrackers based on Energy Dispersive X-ray Fluorescence Spectrometry (EDXRF

DRPK-005

Analysis of 14 Trace Elements in Stainless Steel

Reduction of the Sample Size in the Analysis of Rock by EDXRF

S2 PUMA for Analysis of Geological Materials. Innovation with Integrity. Precise and Reliable Grade Control of Geological Materials EDXRF

Analysis of wear metals and contaminants in engine oils using the 4100 MP-AES

METHOD 7550 OSMIUM (ATOMIC ABSORPTION, DIRECT ASPIRATION)

Cement analysis using FP program (Fundamental Parameters)

Multi-Element Analysis of Trace Metals in Animal Feed using ICP-OES

SPECTRO XRF Report. SPECTRO xsortxhh03. Analysis of Solid Metal Samples. Summary

2017 International Conference on Materials Science and Biological Engineering (ICMSBE 2017)

Bottom Ash Data Week 52

Your Technology Resource

On site heavy metal testing and monitoring using monochromatic XRF methods

EPA Methods Update. What s new and what s coming in EPA ICP-MS methods?

Services SAMPLE SUMMARY REPORT DR. PUMP, LLC Isuzu Rodeo. Atlanta, GA 05/24/2007

IUCLID 5 COMPOSITION AND ANALYSIS GUIDANCE DOCUMENT:

Functionality. Normal Operation

RE-REFINED ENGINE OIL BOTTOMS (REOB) & POLYPHOSPHORIC ACID (PPA) IN HOT MIX. 89 th Annual Transportation Short Course

SEMI C SPECIFICATIONS AND GUIDELINES FOR AMMONIUM HYDROXIDE

Background Statement for SEMI Draft Document 4795 (Yellow Ballot) NEW STANDARD: SPECIFICATIONS FOR NITRIC ACID, USED IN PHOTOVOLTAIC APPLICATIONS

Analysis of Stainless Steel by Dual View Inductively Coupled Plasma Spectrometry

Schedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK

Analysis of Stainless Steel by Dual View Inductively Coupled Plasma Spectrometry

Electron Probe Micro-Analysis (EPMA)

EPA Primary. (mg/l as CaCO3) (mg/l as CaCO3)

ISO INTERNATIONAL STANDARD. Iron ores Determination of various elements by X-ray fluorescence spectrometry Part 1: Comprehensive procedure

Bottom Ash Data Week 45

Improved Antimony Recoveries in Soil Matrices by 3050B/6010B. DataChem Laboratories, Inc. Robert P. Di Rienzo Jeffery S. Ward John T.

XRF DRIFT MONITORS DATA CALIBRATION MATERIAL

New Advances using Handheld XRF Technology for the Prevention of Flow-Accelerated Corrosion (FAC)

US EPA Method using the Thermo Scientific icap 7600 ICP-OES Duo

ELEMENTS by ICP 7301 (Aqua Regia Ashing)

Soil Sampling, Preparation and Analysis

Background Statement for SEMI Draft Document 5134 Revision to SEMI C , Specifications for Hydrofluoric Acid

Bottom Ash Data Week 47

New Developments in Measuring Low Silicon in Process Piping Using Handheld X-ray Fluorescence (XRF)

Analysis of inorganic impurities in lubricating oils by ICP-MS

EDXRF Portable Spectrometer

The Determination of Trace Elements in Stainless Steel by Forked Platform GFAAS

This method is applicable to the following parameters: Nitric Hydrochloric acid digestion, Instrumental analysis.

The image part with relationship ID rid4 was not found in the file. Welcome

Analysis of inhomogeneous samples and trace element detection in alloys using QUANTAX Micro-XRF on SEM

CERTIFICATE OF ANALYSIS 1.0 ACCREDITATION / REGISTRATION 2.0 PRODUCT DESCRIPTION

Background Statement for SEMI Draft Document 4701B New Standard: GUIDE FOR HAFNIUM AMIDES

Bridging new frontiers in Lab - to - Field analysis

ENERGY-DISPERSIVE X-RAY FLUORESCENCE ANALYSIS OF MONO- AND POLYCRYSTALS OF SELENIDE SPINELS BY FUNDAMENTAL PARAMETER METHOD

DETERMINATION OF SULPHUR CONTENT OF UN-ALLOYED AND LOW-ALLOYED STEEL BY ICP-AES SPECTROMETRY USING WET CHEMICAL SAMPLE PREPARATION

ONLINE X-RAY ELEMENTAL ANALYSIS OF ORE AND OTHER BULKY MATERIALS

Analysis of Copper and Copper Base Alloys, Using Shimadzu PDA-7000

Easiest analysis of additives in lubricating oils

Trace Metals in Waters by GFAAS, in Accordance with U.S. EPA and Health Canada Requirements

Examination of Analytical Conditions for Trace Elements Based on the Detection Limit of EPMA (WDS)

X-Ray Fluorescence Lab Report

ELEMENTS by ICP 7301 (Aqua Regia Ashing)

Thermo Scientific X-ray product range. For chemical and phase analysis of solids, liquids and powders

Thermo Scientific X-ray product range. For chemical and phase analysis of solids, liquids and powders

Standard Solutions (Traceable to NIST)

SEMI C SPECIFICATIONS AND GUIDELINES FOR HYDROGEN PEROXIDE

Creating Value in Mining: How XRF Delivers Accuracy and Precision in Daily Routine

INTERNATIONAL STANDARD

Thickness and composition analysis of thin film samples using FP method by XRF analysis

Contents. Preface V List of Contributing Authors. Alfred Golloch 1 Introduction 1 References 3

Determination of Elemental Impurities in Graphite-based Anodes using the Agilent 5110 ICP-OES

ISO INTERNATIONAL STANDARD. Petroleum products Determination of sulfur content Wavelength-dispersive X-ray fluorescence spectrometry

Standard Solutions (Traceable to NIST)

Transcription:

Analysis of an OTC Multivitamin Tablet Using X-ray Fluorescence Spectroscopy and Atomic Absorption Spectroscopy Anne Concepcion G. Pozon, Nathalie P. Wong, Dr. Julita C. Robles* and Dr. Wyona C. Patalinghug Chemistry Department De La Salle University Manila *Corresponding Author: julita.robles@dlsu.edu.ph Abstract: The elemental composition of a branded OTC multivitamin tablet was determined using the spectroscopic methods: atomic absorption spectroscopy (AAS), energy dispersive x-ray fluorescence spectroscopy (EDXRFS) and wave dispersive x-ray fluorescence spectroscopy (WDXRFS). Tablets were randomly chosen from the same batch and their elemental composition was expressed in terms of Ca, Cu and Fe. The quantitative analysis was performed using standard addition method for AAS and standard-less method for both XRFS methods. The elemental composition of the tablet obtained from AAS measurements showed that it contains: 257.9mg ± 5.5 for Ca, 1.972 mg ± 0.090 for Cu and 64.46 mg ± 1.73 for Fe. The analysis of the composition of the tablet obtained using EDXRF and WDXRF was reported in terms of % metal oxide. The semi quantitative standard-less EQUA-ALL method was used for the S2 Ranger EDXRFS and results showed that the tablet contains 28.2%, 7.52% and 0.134% of CaO, Fe2O3 and CuO respectively for the loose powder form while 33.7%, 9.27% and 0.179% of CaO, Fe2O3 and CuO respectively for the pressed pellet form. On the other hand, the composition of the tablet obtained with the use of a standard-less method for S8 Tiger WDXRFS gave the following results: 26.46% CaO, 6.236% Fe2O3 and 0.0455% CuO from the analysis of the loose powder form and 27.87%CaO, 6.741% Fe2O3 and 0.0494% CuO for the pressed pellet form. Although the use of XRF is simple and boasts of the use of the standardless method, the reproducibility of the results gathered from XRFS, as represented by the related methods of EDXRF and WDXRF were inferior compared to the results obtained using AAS. Key Words: AAS;, XRF; standardless method; EDXRF; WDXRF; analysis of multivitamin tablet 1. INTRODUCTION The most common analytical technique used for the determination of the elemental composition of various samples is the atomic absorption spectroscopy (AAS). Standard analytical methods and procedures involving the use of AAS for the determination of metal contents has been tested and established. However, this method requires tedious sample preparation and standardization process. The method involved in the sample preparation increases the risk to personal and systematic errors and the use of concentrated acids for sample digestion makes the method hazardous to the environment. X-ray fluorescence spectroscopy (XRFS) is a new analytical technique designed to analyze elements present in a sample using x-ray as the source of radiation. XRFS was found to be a faster, more practical technique used in analyzing elements due to its non-destructive sample preparation process. It can analyze different types of samples in a shorter span of time. Unlike AAS, it does not require the use of separate hollow cathode lamps as source of radiation and thus XRFS can analyze a wider range of elements. FNH-I-009 1

Due to the non-destructive nature of XRFS, several studies have been conducted to evaluate its efficiency on various analyses of different materials. A study by Richard King proposed the use XRFS for mass analysis of various food samples highlighting the ease of sample preparation, faster running time and efficient results with comparison to the results obtained by standard chemical methods. Also, he states that XRFS is more preferred for mass analysis of samples because of its easy sample preparation and faster running time. i Perring et al. recommended that EDXRF can be used as a quick reliable method for the quantification of iron, copper, and zinc in food premixes. They state that good calibrations should be done in order to produce efficient results. ii XRFS has also been used in determining the quality of rice through quantitative determination of elements such as copper, iron, zinc, cadmium, and vitamin A. It was stated that XRFS was better to use since it does not require sample digestion. The results were found to be precise and reproducible. iii In Japan, cadmium pollution in rice has become a widespread problem. A quick way to analyze trace amount of cadmium in brown rice using high energy dispersive XRF with threedimensional polarization optics was developed by Nagayama, H. Their results exhibited a linear correlation of 0.023 to 1.82 ppm with the certified reference materials. They had successfully determined cadmium concentration in rice at 20ppb using Cd K. ALPHA line after optimizing the optics together with the secondary target. iv The most prominent application of XRFS is in heavy metal analysis especially in water samples. Moriyama, T. used XRFS to analyze lead and arsenic in various river water samples. He preferred the use of XRFS since it is simpler, rapid and reproducible in comparison with inductively coupled plasma mass spectroscopy and AAS. v Melquiades, F. and Appolono C. proposed that XRFS can be used as an alternative technique in analyzing contamination in water samples. They compared XRFS with atomic absorption spectroscopy and inductively coupled plasma atomic emission spectrometry (ICP-AES) in terms of rapidness of analysis, precision, reliability, accuracy and low detection limits for trace elements. They used a field portable XRFS since it uses an efficient radioisotope source excitation and has highly sensitive detectors. They claim that this method is more cost effective and time efficient for on-site analysis. vi FNH-I-009 2 There are fourteen (14) essential minerals which are crucial to one s growth and development. These minerals are calcium (Ca), chromium (Cr), copper (Cu), fluorine (F), iodine (I), iron (Fe), magnesium (Mg), manganese (Mn), molybdenum (Mo), phosphorus (P), potassium (K), selenium (Se), sodium (Na) and zinc (Zn). Too little amounts of these essential minerals can cause deficiency diseases while an excess of these minerals can also be toxic. This study compares the quantitative elemental analysis of an over the counter (OTC) multivitamin tablet for pregnant women using XRFS and AAS based on reproducibility of results, ease of sample preparation and level of support and training required in running the equipment. It aims to conduct a quantitative analysis of the mineral content of the OTC multivitamin sample by measuring the amount of calcium, copper and iron present per tablet using XRFS and AAS. The reproducibility of the results and method will be evaluated by comparing the resulting data obtained from both instrumental techniques. Due to the specific mineral requirement of the human body, multivitamins are properly labeled according to its composition. The results obtained from this study will confirm the mineral composition of the multivitamin sample as stated in its label. The study also aims to validate the use of XRFS as an alternative analytical technique for elemental analysis. The study focused on the determination of calcium (Ca), copper (Cu) and iron (Fe) present in a branded OTC multivitamin tablet for pregnant women using Energy Dispersive XRF (EDXRF), Wavelength Dispersive XRF (WDXRF) and Atomic Absorption Spectroscopy. Both loose powder and pressed pellet sampling methods were used for measurements using the EDXRF and WDXRF while the standard addition method was used for AAS measurements. 2. METHODOLOGY The samples used for this study is a branded OTC multivitamin tablet for pregnant women purchased from a local pharmacy in Manila with the same batch number. All chemicals and reagents were analytical reagent grade and did not require any further purification prior to use. Measurements were conducted using the instruments Varian SpectrAA 200 Atomic Absorption Spectrometer, S2 Ranger Energy Dispersive X-ray Fluorescence Spectrometer and S8

Tiger 1K Wavelength-Dispersive X-ray Fluorescence Spectrometer. 2.1 Quantitative Analysis of Ca, Cu and Fe Using Atomic Absorption Spectroscopy (AAS) Each tablet was accurately weighed, pulverized and digested in hot concentrated nitric acid. A 1.000 L dilute sample solution (DSS) was prepared from each homogenized sample. Standard solutions of Cu, Ca and Fe were prepared by diluting the commercially available 1000 ppm standard solutions (Atomic Absorption Standard Grade). The concentration of the dilute solutions was based on the working range for each element. The quantitative analysis of the multivitamin was conducted using the standard addition method. The instrumental parameters of the AAS for each element are listed in Table 1. Table 1. List of Instrument Parameters Used for the AAS Measurements Element: Calcium (Ca) Working Range (μg/ml) 2-800 Wavelength (nm) 422.7 Slit Width (μg/ml) 0.5 sample using an agate mortar and pestle. Measurement as loose powder was done by placing seven (7) grams of the pulverized sample into a liquid cup containing a 4μm Prolene film at the bottom. The pulverized sample was compacted by carefully knocking the cup on top of a table. The loose powder sample was measured in helium atmosphere. For pressed pellet method, 10 grams of the pulverized sample was weighed and pressed into a pellet (diameter = 40mm) by the application of 20 tons of force for 30 seconds. The measurements were conducted under vacuum. 2.2.1 Energy Dispersive X-Ray Fluorescence (EDXRF) Spectroscopic Analysis The samples were analyzed using a semiquantitative software called EQUA-ALL which is a unique universal calibration that automatically corrects line overlaps without the need for any adjustment. The elements were measured in three different conditions for light, medium and heavy weight elements. Table 2 contains a list of the parameters used for the EDXRF Spectroscopic measurements. Table 2. List of the Instrumental Parameters Used for EDXRF Spectroscopic Measurements Element: Copper (Cu) Working Range (μg/ml) 1-280 Wavelength (nm) 327.4 Slit Width (μg/ml) 0.2 Element: Iron (Fe) Working Range (μg/ml) 1-100 Wavelength (nm) 372.0 Slit Width (μg/ml) 0.2 2.2 Quantitative Analysis of Ca, Cu and Fe Using X-Ray Fluorescence (XRF) Spectroscopy The sample preparation and analytical measurements were performed in Bruker Laboratories (Bruker AXS GmbH) Karlsruhe Germany. The measurements were conducted using two sampling methods namely, as loose powder and as pressed pellet. The samples were prepared by pulverizing eight (8) tablets of the multivitamin Light Elements (Voltage = 20 kv) FNH-I-009 3 Filter (μm) none Measurement Time (s) 35 Current (μa) Optimized Maximum count rate (kcps) 100 Medium Elements (Voltage = 40 kv) Filter (μm) Al 500 Measurement Time (s) 35 Current (μa) Optimized Maximum count rate (kcps) 100 Heavy Elements (Voltage = 50 kv) Filter (μm) Cu-250 Measurement Time (s) 100 Current (μa) Optimized Maximum count rate (kcps) 100

2.2.2 Wavelength Dispersive X-Ray Fluorescence (WDXRF) Spectroscopic Analysis The samples were analyzed using a standard-less method and were directly subjected to the machine for analysis. No details on the instrument parameters were provided. 3. RESULTS AND DISCUSSION Representative tablets of the branded OTC multivitamin tablet were randomly chosen from the same batch and were analyzed for its elemental and quantitative composition specifically Ca, Cu and Fe using the following instrumentation techniques: the standard addition method for AAS, the semi quantitative standard-less EQUA-ALL method for S2 Ranger EDXRF and a standard-less method for S8 Tiger WDXRF. The multivitamin label states that each tablet contains 250mg calcium, 2.0mg copper and 60mg iron in the form of calcium carbonate, cupric oxide and ferrous fumarate respectively. These values will be the basis of comparison for the results that will be generated from the analysis. Figure 1. Quantitative Analysis of Cu Using the Standard Addition Method (Tablet No. 2) Table 3. Summary of the Quantitative Analysis of Copper (Cu) Using AAS Measurements Tablet 1 Tablet 2 Tablet 3 Trial 1 2.114 2.010 1.866 Trial 2 2.084 1.943 1.919 Trial 3 1.950 1.891 Average 2.099±0.021 1.968±0.037 1.892±0.027 Total average 1.972 ± 0.090 3.1 Results of the Elemental Analysis using AAS The absorbance of the spiked sample solutions were plotted as a function of the concentrations of the added standard solution of the element being analyzed. The equation of the calibration curve was determined using the calibration model, least squares method. From the x-intercept of the equation, the concentration of the element per sample was determined and the corresponding amount in grams was calculated. Only the plots of the absorbance vs. concentration of the standard solution of the elements Cu, Ca and Fe added into aliquot portions of the dilute sample solution containing the tablet No. 2 are shown in Figures 1, 2 and 3 respectively. The results of the elemental analysis for each element are summarized and presented in Table Nos. 3-5 According to Table 3, the average composition of Cu present in the tablet is 1.972 mg ± 0.090 with only 1.39% difference from the stated value of the multivitamin in the label. Figure 2. Quantitative Analysis of Ca Using the Standard Addition Method (Tablet No. 2) FNH-I-009 4

Table 4. Summary of the Quantitative Analysis of Calcium (Ca) Using AAS Measurements Tablet 1 Tablet 2 Tablet 3 Trial 1 247.2 259.1 260.9 Trial 2 250.0 260.8 259.5 Trial 3 260.6 258.9 264.2 Average 252.6±7.1 259.6±1.0 261.5±2.4 Total average 257.9 ± 5.5 Based on the summary of the results presented in Table 4, with the use of the standard addition method for AAS, it revealed that the average amount of Ca present in the sample is 257.9mg ± 5.5 with a 3.16% difference from the values indicated in the label. In general, a comparison of the values representing the chemical composition of the tablet showed a good agreement between the specifications of the metals on the label and actual values obtained. 3.2 Results of the Elemental Analysis using EDXRF and WDXRF The results for the two XRFS techniques were reported in terms of the percentage of the oxides of Calcium (Ca), Copper (Cu) and Iron (Fe) in the form of Calcium Oxide (CaO), Cupric Oxide (CuO) and Ferric Oxide (Fe2O3) respectively. Table 6. Results of the EDXRF (S2 Ranger) Analysis of the of the OTC Multivitamin Tablet Loose Powder Pressed Pellet [%] 7 g in liquid cup 10 g CaO 28.2 33.7 Fe2O3 7.52 9.27 CuO 0.134 0.179 Table 7 Results of the WDXRF (S8 Tiger) Analysis of the OTC Multivitamin Tablet Figure 3. Quantitative Analysis of Fe Using the Standard A Loose Powder Pressed Pellet [%] 7g liquid cup 10 g CaO 26.46 27.87 Fe2O3 6.236 6.741 CuO 0.0455 0.0494 Table 5 Summary of the Quantitative Analysis of Iron (Fe) Tablet 1 Tablet 2 Tablet 3 Trial 1 65.63 62.52 63.47 Trial 2 64.75 62.27 65.81 Trial 3 67.07 62.85 65.76 Average 65.82±1.17 62.55±0.29 65.01±1.34 Total average 64.46 ± 1.73 The evaluation of the AAS measurements of the iron content revealed that the amount of Fe present in the sample is 64.46mg ± 1.73 with a 7.43% difference from the values indicated in the label. FNH-I-009 5 The WDXRF results for CaO and Fe2O3 show a close comparison with the AAS results and the label stated by the multivitamin brand with percent differences ranging from 2% to 10% only unlike those of the EDXRF which varied from 4% to 34% difference. While for CuO, more than 70% difference was calculated. This was probably due to the effect of a secondary enhancement where the x-ray emitted by a heavier element stimulates the x-ray emissions of a lighter element. 4. CONCLUSION The quantitative analysis of a branded OTC multivitamin tablet designed for pregnant women was performed using AAS, EDXRF and WDXRF. Results obtained from AAS

measurements reflect a high degree of reproducibility and the values obtained are consistent with those reported from the product label. The AAS results showed agreeable values with the label of the multivitamin sample stating the amount of elements present per tablet. The resulting amount per tablet was 257.9mg±5.5 for Ca, 1.972mg±0.090 for Cu and 64.46mg±1.73 for Fe. The XRF allows the analysis of a wide range of elements and does not involve a complicated sample preparation. However, results showed that significant deviation of the reported values was obtained using XRFS. Furthermore, a comparison of the values for WDXRF and EDXRF lack agreement in results. These uncertainties may due to the difference in the sample matrix of the tablet and the reference used in the preparation of the reference calibration curve and due to secondary enhancements brought about by heavier elements present in the sample. The XRFS as analytical tool is better suited for qualitative analysis unless a suitable blank that has identical properties as the sample matrix of the sample is available. This is also the reason why, the use of the standard-less method is not highly recommended for the analysis of samples that have varied composition. A comparison of the EDXRF and WDXRF reveal that WDXRF was more accurate in elemental analysis since it presented closer values between the loose powder and pressed pellet form. Also, it has a higher resolution and higher count rates with the same standard-less and nondestructive approach to samples. Despite the high agreement of the results obtained using AAS, it is noted that sample preparation was a big concern for the analyst. The use of large amounts of acids and standard solutions introduces wastes to the environment and is a cause for concern. 6. REFERENCES i King, R. Determination of calcium, potassium, chlorine, sulfur, and phosphorus in foodstuff by x-ray fluorescence spectrometry. Journal of Science of Food and Agriculture. 2006, 28, 631-634. ii Perring, L. Andrey, D. et al. Rapid quantification of iron, copper and zinc in food premixes using energy dispersive x-ray fluorescence. Journal of Food Composition and Analysis. 2005, 18, 655-663. iii Mizuhara, Manabu. Spectroscopy Focus. Trace element analysis of rice by XRF. October 2007. Japan. iv Hiroyuki, N. Ryoko, O. et al. Determination of Cd at sub-ppm in Brown Rice by X-Ray Fluorescence Analysis Based on the Cd K. ALPHA. Line. Advances in X-ray Chemical Analysis. 2005, 36, 235-247. v Moriyama, T. Yamada, K. Kono, H. The analysis of Trace Hazardous elements in water samples. Advances in X-ray Chemical Analysis. 36, 201-212. vi Melquiades, R. Appoloni C. Application of XRF and field portable XRF for environmental analysis. Journal of Radioanalytical and Nuclear Chemistry. 2004, 262, 533-541. 5. ACKNOWLEDGEMENT The authors would like to acknowledge George Tang (Bruker AXS Pte. Ltd) and Mr. Uwe Gerhard Schneider (Bruker AXS GmnH) for the WDXRF and EDXRF spectroscopic measurements of the samples. FNH-I-009 6