DETERMINATION OF HYDROCARBON DEW POINT IN NATURAL GAS. Andy Benton. Michell Instruments, Inc 11 Old Sugar Hollow Road Danbury, CT 06810

Similar documents
PROCESS MOISTURE ANALYZERS Measuring moisture in gas or HC liquids in hazardous areas

PROCESS. Process Analysers For Gas and Liquid Phase

Eric Kelner and Darin L. George. Southwest Research Institute 6220 Culebra Road San Antonio, Texas, USA

Eric Kelner and Darin L. George. Southwest Research Institute 6220 Culebra Road San Antonio, Texas, USA

OptiPEAK TDL600 Moisture in Natural Gas Analyzer

Qualitative Phase Behavior and Vapor Liquid Equilibrium Core

Condumax II Hydrocarbon Dew-Point Analyzer

Effective & Sustainable SVE tool to extract PHC vapours. Bogdan Pawlak P.Geol & Ole Mrklas, PhD

MIDSTREAM GAS QUALITY

MOSITURE MEASUREMENT IN NATURAL GAS

Waarom moeilijk doen als het makkelijk(er) kan

Final DRAFT API TECHNICAL REPORT. Carbon Content, Sampling, & Calculation

Fast Detailed Hydrocarbon Analysis by Modified ASTM Method D6730

HydrocarbonEngineering. May2009. Reprinted from.

Analysis of Permanent Gases and Extended Hydrocarbons in Natural Gas and Natural Gas Liquids by Capillary Gas Chromatography

Global Calibration Gases LLC

(L)NG composition measurements

Precision Humidity Solutions

Residual Solvents in medical cannabis analysis using the SRI 8610C FID GC June 2014

Natural Gas Monitoring

Name: Organic chemistry. Class: Foundation revision questions. Date: 67 minutes. Time: 67 marks. Marks: Comments: Page 1 of 27

System Overview. Valve oven. Restrictor. Valve. Sample in and out

Hydrocarbon Processing Industry Gas Chromatographs. Standard GC Analyzer Solutions

Heat Value Gas Chromatograph ISO standard version

ETHYLENE PLANT DRY CRACKED GAS Product Code 54801

MicroSAM / Sitrans CV

Production Allocation Deployment, from Concept to Operation

Thomas G. Braga Manager, Research and Development. SulfaTreat, a Business Unit of M I L.L.C. A Smith/Schlumberger Company

Development of cryogenic LNG reference liquids Dr Paul Holland EffecTech Group

Simple Dew Point Control HYSYS v8.6

Technical Guidance Note (Monitoring) Medium Combustion Plant Directive and Generator Controls: monitoring point source emissions

ARE PRODUCED WATER EMISSION FACTORS ACCURATE?

TDLAS. Application Note. Tunable Diode Laser Absorption Spectroscopy (TDLAS) for Carbon Dioxide Measurements In Natural Gas Processing

Improving Natural Gas Liquefaction Plant Performance with Process Analyzers

ADVANCES IN MICRO GAS CHROMATOGRAPHY

Natural Gas Processing Unit Modules Definitions

PROCESSING NATURAL GAS Leontev A.A. Vladimirskiy State University named after the Stoletov brothers Vladimir, Russia

New advances in mine site gas analysis using gas chromatographs

Fast Analysis of Extended Natural Gas using the New Transportable INFICON Micro GC Fusion

GAS CONDITIONING FOR GAS STORAGE INSTALLATIONS

MEASURING HYDROCARBON DEW POINT, THE ECONOMIC IMPACT

Schedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK

Simple Dew Point Control HYSYS v10. When the simulation is set up the overall PFD should look like the following figure.

VE Technology Sampling System for the (Manual) Measurement of Trace levels of Mercury in Natural Gas. The Background

Heat Effects in Gas Systems

APPLICATION NOTE. Real- time inline process monitoring. Refinery, fuel gas V

Real-Time Measurement for Unparalleled Safety and Product Quality Control

THE AMERICAN SOCIETY OF MECHANICAL ENGINEERS 345 E. 47th St., New York, N.Y GT-227

Stack Sampling. requires airflow through the sampling probe to be at the same rate as that flowing in the waste gas stream isokinetic

The Proper Mechanism for Field Validating Moisture Analyzers. GE Measurement & Sensing GE Oil & Gas

Large Valve Oven Analyzers for Refinery Gas Analysis. ASTS 2015 Houston Shannon Coleman Application Specialist Agilent Technologies

TRACE ANALYTICAL. ta7000 Series UHP Gas Purity Monitors At the critical link in your manufacturing process. Features /Benefits

SVENSK STANDARD SS-ISO 18453:2004. Naturgas Korrelation mellan vattenhalt och daggpunkt (ISO 18453:2004, IDT)

Bulk Calculations - Gas BCG 3.0. Supported Standards Manual. Lists the standards supported by BCG 3.0 at the time of document publication

FLARE GAS MEASUREMENT AND RECOVERY OF FUEL FEED GAS WITH RESIDUAL OXYGEN CALORIMETRY

Guidance Document for the Introduction of Landfill- Derived Renewable Gas into Natural Gas Pipelines

Title slide. LNG Technology. Compiled by PD.Supriyadi

Reprinted from HydrocarbonEngineering December

EDEXCEL NATIONAL CERTIFICATE/DIPLOMA. PRINCIPLES AND APPLICATIONS of THERMODYNAMICS NQF LEVEL 3 OUTCOME 3 -COMBUSTION

Monitoring for Toxic VOCs in Oil Industry Applications

2.1 This procedure provides steps and conditions for the determination of Vapor Space Organic Compounds in ground waters, and industrial waste waters.

Compact liquefied gas expander technological advances

2011 Laurance Reid Gas Conditioning Conference

Portable Instruments. Fast, accurate and stable measurements in the most demanding industrial environments.

EFFECT OF MIXTURE CONSTITUENTS ON THE LAMINAR BURNING VELOCITY OF LPG-CO2-AIR MIXTURES

Chemistry of Petrochemical Processes

Improved Gas Chromatograph Method for the Analysis of Trace Hydrocarbon Impurities in 1, 3-Butadiene

Aurora Moisture Analyzer New Light for Moisture Measurement in Carbon Dioxide

STANDARD OPERATING PROCEDURES

AGC NovaPRO Process Gas Chromatograph Refinery Applications

APPLICATION NOTE. UOP : Analysis of Trace CO and CO 2 in bulk H 2 and Light Gaseous Hydrocarbons by GC

Integrated Sampling Solutions

Totalflow analyzer products Precisely what you expect

REPORT. Hobré HIGAS, a Portable Natural Gas Quality Meter. Wobbe Index Analyzer. Project report September 2015

ADVANCES IN MICRO GAS CHROMATOGRAPHY

10/2/2013. Gas CHEMICAL PLANTS AMMONIA METHANOL UTILITIES TOWN GASS SUPPLIES ENERGY INTENSIVE INDUSTRIES. Power Generation

HILASE - QUALITY CONTROL INSTRUMENT

Customer-Owned Distributed Methane Interconnection Issues

Meet the airgc. Ambient air carrier. Top performance. catalytics accelerated

Impact of Elemental Sulfur in Natural Gas

MOLECULAR GATE TECHNOLOGY FOR (SMALLER SCALE) LNG PRETREATMENT

Dave Abbott Independent GT Combustion Specialist

MAX300-RTG PRODUCT NOTE. Real-Time Gas Analysis Process Automation and Control Product Quality Analysis. Industrial Process Mass Spectrometer

Development of an LNG sampling and composition measurement reference standard. Jianrong Li LNG II Workshop 22 August 2017, VSL, Delft, NL

Novel Method for Gas Separation By: Chris Wilson and Dr. Miguel Bagajewicz

Online dissolved gas analysis sensor

Contents Principles of Operation General Specifications... 4 Outline Drawing and construction.. 5 Performance Measurements. 6 Gas Sensitivity/Li

Optimising. the LNG process. The rapidly expanding global LNG industry continues. Projects

University of Zagreb, Faculty of Mining, Geology & Petroleum Engineering

Modular Oil & Gas Equipment Onshore & Offshore

Measuring moisture in ultra-high-purity gases

Calibration gas cylinders Calibration gas cylinders for the calibration or the periodical checkings of the gas detectors

Fpollutant-free zero air for analytical

Shur-Shot X-Proof Hydrogen Fluoride Alarm Operations Manual

CHAPTER 3 BENEFITS OF BETTER BURNING

Available online at Energy Procedia 100 (2009) (2008) GHGT-9. Allan Hart and Nimalan Gnanendran*

TECHNIQUES OF COMPOSITE SAMPLING. Matthew S. Parrott K2controls, Inc FM 1960 Rd West Houston, TX

Promet EExd Process Moisture Analyzer

WHAT IS LIQUEFIED NATURAL GAS?

Field Operations & Inlet Receiving. Chapter 8

Transcription:

DETERMINATION OF HYDROCARBON DEW POINT IN NATURAL GAS Andy Benton Michell Instruments, Inc 11 Old Sugar Hollow Road Danbury, CT 06810 ABSTRACT This paper considers the requirements for control of hydrocarbon dew point in natural gas and how measurement of this important gas quality parameter can be achieved. A summary of the commercially available on-line instrumentation is provided covering: Manual, visual technique with chilled mirror dewpointmeter Equation of state calculation from extended composition analysis by gas chromatograph Automatic, optical condensation dewpointmeter The role of each measurement technology is described and assessed in terms of the effectiveness of the measurement method utilised together with other technical considerations as well as initial and operating cost implications. Full consideration is given to the specific difficulties to be confronted resulting from the complex nature of the parameter concerned. Such peculiarities include the effects of pressure, fractional condensation, the minute proportion of heaviest molecular weight components within the gas composition that contributes to the formation of condensate at the hydrocarbon dew point, and the overall subjectivity of the measurement itself where no absolute reference or definition is possible. A case is presented for the use of advanced optical techniques in an adaptation of the fundamental chilled mirror principle to provide automatic on-line measurement with a degree of objectivity and repeatability unobtainable with other measurement techniques. INTRODUCTION Hydrocarbon dew point in natural gas is an important quality parameter, stipulated in contractual specifications and enforced throughout the supply chain, from producers through transmission and distribution companies to final end users. Accurately monitoring hydrocarbon dew point temperature in natural gas is therefore vital if the integrity and quality of the gas are to be maintained, and contractual agreements complied with. Hydrocarbon dew point is the temperature at which hydrocarbon condensates first begin to form when natural gas is cooled, at constant pressure, and thus ceases to be wholly gaseous. In many cases, however, avoiding the formation of such liquids in natural gas is of critical importance. For example, in gas pipeline transmission the presence of liquid hydrocarbons in combination with traces of moisture leads to the formation of hydrates as virtual solid masses. Under pipeline operating conditions of high flow and pressure, these hydrates can cause impact damage to compressors and restrict or even block pipelines. 1200 1050 900 Pressure 750 PSIG 600 Water and HC liquids Wholly gaseous 450 HC liquids present in gas present in 300 gas HC Dewpoint 150 H 2 O Dewpoint 0-40 -35-30 -20-10 0 Temperature, Temperature, F Pressure Figure 1. Typical retrograde phase envelope for natural gas In addition, certain end users at electricity power stations have critical needs with respect to hydrocarbon dew point. The latest generation of natural gas fuelled turbines designed for low NOX emissions require at least 50 F of 'superheat' - the differential between dew point temperature and the delivery temperature of the gas on entry to the fuel gas/air pre-mixing system. If this is not achieved then the cooling effect of the expansion of large flows of fuel gas can cause liquid to be formed, resulting in possible flashback of the flame from the ignition nozzle into the mixing chamber. The cost repercussions of such a problem can be serious given the resultant sudden interruption to electricity generation, potential damage to the mixing system and increased maintenance cost/downtime to the power station. For the natural gas producer, the control of hydrocarbon dew point can be a costly aspect of natural gas processing. The chiller plants, which are commonly used for low temperature separation of natural gas liquids, operate by expansion of the gas flow to develop a Joules Thomson effect. The lower the gas must be cooled, the lower the

pressure by which the process gas has to fall, typically by 400 to 600 psi. This has a direct impact on production costs in terms of running the turbine compressors to recompress the gas to transmission pressures, typically 750 to 1500 psig. A further direct link to the economics of natural gas production is that the greater the proportion of natural gas liquids (condensates) separated out, the lower the heating value for the gas and thus the metered value of the gas sold. Well Head 10-20MPa Glycol Contactor Water dewpoint reduction Separator HC liquids and water Hydrocarbon Dewpoint reduction P = 3 to 5 MPa J.T. Valve Low Temperature Seperator Condensate Figure 2: Natural gas processing method Figure 3: Hydrocarbon dew point reduction plant Transmission Pipeline 7 to 15 MPa Compressor It is therefore in the interest of producers to monitor and control hydrocarbon dew point reduction plants within tight limits, to minimise processing costs while maximising the earnings potential for the gas produced. Such commercial pressures, however, often mean that plants run as close to the contractual limit for hydrocarbon dew point as possible. This in turn raises the possibility of straying outside contractual specifications, leading to the requirement for continuous monitoring as part of tariff metering by operators downstream. This may include measurements being made by the pipeline transmission company initially receiving the gas and at all custody transfer points thereafter, through to final retail distribution or large industrial end user. With increased global competition among natural gas producers, particularly as world markets deregulate and the network of interconnecting pipelines between countries continues to expand, the need to improve production efficiency whilst administering vigilant quality controls has never been greater. As such, the methods of analysis for hydrocarbon dew point are a matter currently receiving the attention of companies operating throughout the natural gas industry. THE METHODS FOR HYDROCARBON DEW POINT MEASUREMENT Manual, visual technique with chilled mirror dewpointmeter The use of manually operated chilled mirror dewpointmeters is the simplest and most widely applied method of hydrocarbon dew point measurement. These instruments have a metallic mirror surface mounted within a pressurised sample chamber, with a glass viewing port through which an eyepiece allows the operator to observe the mirror surface. The instruments are used for periodic spot check measurements. This type of instrument utilise expanded carbon dioxide gas, supplied from a cylinder under operator control, to cool the underside of the mirror when performing each measurement. A flow valve manually adjusts the rate and degree of cooling, while the operator observes the mirror surface in order to detect the point at which condensate appears. The aim is to detect the very first visible signs of condensate. This can, however, be extremely problematic given the virtually invisible film-like properties of natural gas liquids. At the same time, the decreasing temperature measurement reading, projected into the eyepiece, must be observed as the mirror temperature is cooled. The rate of mirror cooling is of critical importance and should be as slow as possible through the region in which the dew point is likely to be found, in order to achieve the best sensitivity and repeatability of measurement. The subjective nature of such a measurement technique means that anything less than the greatest of care and an experienced eye can result in great variability in results. Even given these reservations, which question the objectivity of this measurement technique, this method is established as the de facto standard for this parameter by virtue of fundamental measurement principle and is in almost universal usage across the global gas industry. So, in the vast majority of cases, it is against this manual visual method that all forms of on-line automatic hydrocarbon dew point measurement are periodically compared and expected to agree irrespective of whether such automated techniques may have the capability to achieve a greater level of detection sensitivity and repeatability.

Strengths: The traditional and most widely used measurement technique The de facto standard for hydrocarbon dew point measurement Low capital investment - Instrument costs US$5,000 to US$10,000. Weaknesses: Periodic spot checking only 'Subjective', operator dependant measurement of variable sensitive and repeatability Labour intensive, therefore high running costs Equation of state calculation from extended GC analysis Process gas chromatographs are widely used in the natural gas industry at custody transfer points for the calculation of metering parameters such as heating value and Wobbe index. Such process GC s commonly provide a composition analysis to C6+ hydrocarbons. The use of an equation of state calculation to derive the hydrocarbon dew point of natural gas requires an extended analysis quantifying the trace quantities of individual heavy end components present. It is widely recognised that the variation in C10 and higher components, typically in the single figure ppm(mol) region or lower concentrations, has a significant effect in terms of hydrocarbon dew point. To obtain the extended analysis required a more specialist form of GC might be used, with twin columns/ovens to enable separation and thus detection of the heavy end components. In such a GC, a higher temperature oven can be used to separate the heavier molecular weight components to help minimise the additional time taken for such an extended analysis. Even so, the extension of analysis typically reduces the frequency of measurement to between five to six times per hour, compared with up to twenty times per hour for a conventional Process GC performing a normal C1 to C6+ analysis. As chromatography is an inferential form of measurement, a reference gas of certified composition is required for regular calibration. As such, to achieve accuracy in an extended analysis the reference gas used should be certified for the trace heavy end components, up to the highest molecular weight hydrocarbon of the extended analysis to be performed. Analysis of heavier components is critical, with C10 and higher components being of particular concern as they have such a significant influence on the hydrocarbon dew point. Quite simply, an analysis reporting single figure ppm(mol) of these components will have a calculated dew point 10 F or more higher than that obtained from an analysis in which no detectable traces are found. It is suggested that a limit Mol% C 1 Methane CH 4 88 C 2 Ethane C 2 H 6 5 C 3 Propane C 3 H 8 2 C 4 Butane C 4 H 10 1 C 5 Pentane C 5 H 12 0.3 C 6 Hexane C 6 H 14 0.1 C 7 Heptane C 7 H 16 0.03 C 8 Octane C 8 H 18 0.01 C 9 Nonane C 9 H 20 0.001 C 10 Decane C 10 H 22 0.0002 C 11 Undecane C 11 H 24 0.0000 C 12 Dodecane C 12 H 26 0.0000 CO 2 1.5 H 2 O 0.005 (50ppm) H 2 S 0.004 N 2 2.0 O 2 0.05 Table 1. Typical natural gas composition of detection of 0.1 ppm(mol) is recommended for such an analysis to be suitable for the determination of hydrocarbon dew point 1. The skills required to maintain performance of this method of hydrocarbon dew point determination are therefore more suited to an analytical chemist s laboratory than to a natural gas process plant. The significance of the extended composition analysis when calculating hydrocarbon is illustrated, albeit in an extreme case, by the example shown in table 2 of a laboratory analysis for actual pipeline natural gas taken in Canada. In this case, the calculated value for hydrocarbon dew point is underestimated by around 46 o F based on C9+ compared to the C14 analysis, and more than 100 o F if the analysis is shortened to C6+. Strengths: Potential to combine a number of gas quality/fiscal metering parameters into one analyzer The components contributing to a high dew point level may be identified and so help to determine the reason/source Possibility to provide a theoretical phase envelope curve Weaknesses: Accuracy of analysis, and thus hydrocarbon dew point calculation, is dependant on correct and regular use of specialist reference gases An indirect method of determining hydrocarbon dew point relying on the correct application and suitability of the equation of state being used Susceptible to measurement errors due to limit of analysis sensitivity and composition changes (presence of aromatics: benzene & toluene)

Specialist staff required to operate/maintain performance High initial outlay (US$150,000 plus for GC with C12 capability), installation costs (analyzer house) and running costs (personnel and reference gases) Automatic, optical condensation dewpointmeter A number of automatic, optical condensation dewpointmeters have been designed for on-line hydrocarbon dew point measurement over the past thirty years. All of these instruments function on the fundamental principle of direct measurement of the temperature at which hydrocarbon liquids start to form on a chilled surface exposed to the gas sample to be measured by definition the hydrocarbon dew point. The main difficulties in making such a direct fundamental measurement, however, are associated with two characteristics of hydrocarbon liquid formation in natural gas. Firstly, the colourless appearance and low surface tension of natural gas condensates means that the liquid film, which forms as the sample is cooled through the dew point temperature, is almost invisible to the human eye. Likewise, it is difficult to detect using the opto-electronic systems required by an automatic measurement system. Secondly, natural gas has a spectrum of potentially condensable components that contribute to the liquid formed when the gas is cooled through the hydrocarbon dew point. This effect is complex in that some of the hydrocarbon compounds exhibit an affinity for one another while others repel, but effectively the heaviest end components condense first followed by the lighter compounds in descending order of molecular weight. The difficulty here is that, unlike the measurement of water dew point where the objective is to determine the temperature at which a single condensable component (water vapour) becomes saturated, hydrocarbon dew point is, by its very nature, indistinct, as condensation occurs gradually across a range of temperature. LGR mg/nm 3 160 140 120 mg/m 3 100 80 60 40 20 0-30 -20 A -10 0 +10 +20 Temperature, Temperature. C Figure 4. Typical liquid to gas ratio curve (at constant pressure, 400 PSIG) The liquid to gas ratio (LGR) curve (Figure 4), which shows the region referred to commonly as the hydrocarbon tail, best illustrates this characteristic (points B to C). This shows that the theoretical (thermodynamic) dew point, if defined as the temperature at which the first of the heaviest end components bond together from gaseous to liquid phase (point C). However, this would be some 36 o F(20 C) or more higher than the temperature at which the first visible quantity of condensate could be detected using the manual cooled mirror technique and the temperature at which detectable formations of condensates may occur of significance to producers, transporters and end users of the gas. It is therefore accepted practice to take a regression of the linear portion of the LGR curve, to eliminate the otherwise exaggerated effects of the hydrocarbon tail, in order to determine an operational (applied) hydrocarbon dew point of significance in both process and gas quality terms 2 (point A). Any form of automatic hydrocarbon dew-point analyzer must therefore allow the user to adjust the sensitivity of measurement in an informed manner. Ideally, this will be achieved by periodically providing a 'picture' for the condensate formation characteristic of the gas, from the first minute liquid formations into the linear region of the characteristic as shown on the LGR plot. In such a way, the true sensitivity of measurement can be periodically judged and verified by the user and, if desired by the user, could be set to achieve agreement with any analytical method stipulated in contractual agreement with their suppliers, fore example, a carefully performed manual visual chilled mirror measurement by an experienced operator. Due to this fractional condensation effect, the repeatability and sensitivity of measurement will be adversely affected by the inter-relationship of sample flow rate and cooling rate during measurement cooling cycles. The amount of condensate formed on a cooled surface during a measurement cycle is not only dependant on the composition of the gas, but also the supply of components being condensed at that temperature, which in turn is influenced by both the cooling rate and sample flow rate. As such, to achieve good measurement performance, any such automatic technique should operate with a fixed sample volume trapped within the analysis cell prior to the start of each measurement cycle. This method decouples the sample flow-rate/cooling rate interdependency and maintains the integrity of the sample throughout the measurement cycle. B C The successful application of on-line, automatic measurement has been achieved by developing specialist measurement technologies that confront these issues. One of the most extensively applied methods over the last two decades is the Dark Spot optical technique utilised by Michell Instruments CONDUMAX and latest CONDUMAX II Hydrocarbon Dew-Point Analyzers.

The Dark Spot automatic measurement technique Operating with sensitivity of the order of 1mg/m³ of condensed liquid from gas, the Dark Spot optical principle is radically different to that of any other chilled mirror instrument. The CONDUMAX II Hydrocarbon Dew-Point Analyzer comprises an EExd explosion proof certified Main Unit housing the measurement Sensor Cell, and all associated electronics. This Main Unit is located within the hazardous zone with a suitable sampling system either outdoors close to the pipeline or, if preferred, within an existing analyzer house. The selfcontained Main Unit provides all automatic measurement functions, with touch screen interface and alphanumeric display, and providing analogue outputs and MODBUS communications for integration to SCADA/DCS. Figure 6. CONDUMAX II Main Unit: Internal view The Sensor Cell contains an optical surface, mounted on a Peltier heatpump with an embedded miniature precision thermocouple and optical detection components. Figure 5. CONDUMAX II Main Unit and Condumax II within Sampling System Figure 6. CONDUMAX Sensor Cell Assembly and exploded view

Figure 7. Dark Spot principle - Illustration The optical surface is acid etched and semi-matt with a central conical shaped depression. As illustrated in figures 7 and 8, a well-collimated beam of visible spectrum red light is focused onto this central region, which, in dry conditions, reflects most of the beam from the surface to form an annulus ring of light. Optical detection is focused on light dispersed into the centre of the annulus ring - the Dark Spot. During a cooling cycle, as hydrocarbon condensates form on the abraded surface, its optical properties are modified; the reflected light intensity of the annulus ring increases marginally whilst, more significantly, there is a dramatic reduction in the scattered light intensity within the Dark Spot region. Monitored by a photo-detector, the exact signal change depends on the amount of condensate formed on the surface, enabling the measurement sensitivity of CONDUMAX II to be set up to the user s specific requirements. Figure 8. Dark Spot principle Schematics Dry and wetted optical surface To enable the user to apply such judgement, the Main Unit software provides the facility to periodically run a Sensitivity Calibration (figure 9) whereby a graph akin to the LRG plot is produced for the actual natural gas being analysed. In this case, the curve relates signal change in millivolts from 0 to 1500mV, which increases dependant on the amount of condensate formed and is thus directly proportional to LGR for temperatures through the whole cooling range. As with the LGR plot, a regression of the linear portion of the Sensitivity Calibration curve can be taken to determine the dew point of the gas and thus verify the level of sensitivity set for the CONDUMAX II Analyzer, in terms of a trip value for subsequent automatic measurement cycles. Such measurement cycles are highly repeatable for the set signal change measured by the optical system. A factory default trip value of 275mV has been proven to provide comparative measurements to those obtained by an experienced operator applying the manual, visual chilled mirror technique. Figure 9. Typical Sensitivity Calibration Signal Change, mv 1500 1400 1300 1200 1100 1000 900 800 700 600 500 400 300 200 100 0 Trip point 275mV (Factory default) 10 15 20 25 30 35 40 45 50 Optical Surface Temperature, deg. F Furthermore, the high detection sensitivity of the Dark Spot technique counters problems associated with executing continuous, on-line measurements. It enables

CONDUMAX to trap a fixed volume sample of gas for each analysis cycle, de-coupling flow rate and mirror cooling rate effects, and giving a liquid phase composition that is totally representative of the gas being measured. This flexibility allows CONDUMAX II to be used for both condensate quantity and dew point temperature analysis. In the former, the optical surface is cooled to a fixed temperature pre-set by the user and the change in optical intensity is monitored. This data is interpreted as a liquid to gas ratio (mg/m 3 or lb/mmscf) by periodic cross calibration against a recognised condensate analysis technique. In both operating modes the cycle time is normally 10 minutes, so 6 measurements are made each hour. Pressure, bara 60 50 40 30 20 10 0-16 -15-14 -13-12 -11-10 Figure 10. Example of phase envelope produced by CONDUMAX In addition to such on-line monitoring, the CONDUMAX II can also but used to produce a phase envelope plot through a series of measurements performed at a number of pressure levels over the range of interest, which can be completed in one to two hours. An example for a typical European transmission pipeline gas is shown in figure 10. Strengths: Direct fundamental 'objective' measurement of high sensitivity and repeatability Stand alone operation - In-built verification routines Sensitivity may be harmonized with measurement techniques/practices agreed between gas supplier and receiver at custody transfer No specialist operation/maintenance staff requirements Ability to produce phase envelopes through direct measurement rather than theoretical estimation Weaknesses: Temperature, de g. C Dewline Significant initial investment (US$35,000 to 45,000) though low installation and running costs CONCLUSION As a process of global de-regulation of the natural gas industry continues, resulting in an increase in the commercial complexity of the gas supply chain, often involving transfer of custody between a number of independent companies and crossing national or even continental boundaries, gas quality specifications are being enforced more rigorously as each company must ensure that the gas they receive will meet the requirements of their customer onward through the supply chain to the end user. With the resultant rise in the number of parties needing to address gas quality issues, the effective, accurate measurement of hydrocarbon dew point grows in importance. As such, there is marked increase in the use of automatic, on-line instruments such as Michell Instruments CONDUMAX II. The combination of a fundamental measurement principle with the high sensitivity and objectivity afforded by the Dark Spot technique have led to CONDUMAX II being applied extensively throughout the whole supply chain. These applications range from performance monitoring and operational control of hydrocarbon dew point reduction plant, where producers look to achieve improved processing efficiency in order to remain competitive, through gas quality measurements at fiscal monitoring points, to reservoir engineering research work to improve extraction techniques. Furthermore, in the absence of any recognised international standard or even scientific definition for this parameter, the ability of the CONDUMAX II to inform the user of the condensate formation characteristic of the gas being measured, enables the actual sensitivity of measurement to be judged and, importantly, harmonised with measurement techniques/practices agreed between suppliers and receivers at custody transfer points. REFERENCES 1. Cowper, C. J., Natural gas hydrocarbon dew point; comparison of measurement and calculation nd methods, 2 Gas Analysis Symposium & Exhibition 2002, Maastricht, Belgium, January 2002. 2. Bannell, J. L. K., Dixon A. G. and Davies T. P., The Monitoring of Hydrocarbon Dewpoint, paper presented at the International Congress of Gas Quality, Groningen, The Netherlands, 1986.

Table 2. Example of equation of state (Peng Robinson) calculation of HC dew point value using extended C14, intermediate C9+ and standard C6+ analysis of the same natural gas composition Extended to C14 Intermediate to C9+ Standard C6+ helium mol He 4.9807 4.9807 4.9807 hydrogen H2 0.9961 0.9961 0.9961 nitrogen N2 183.2910 183.2910 183.2910 carbon dioxide CO2 67.7380 67.7380 67.7380 Methane C1 9298.0325 9298.0325 9298.0325 Ethane C2 283.9018 283.9018 283.9018 Propane C3 80.6879 80.6879 80.6879 I-butane ic4 12.9499 12.9499 12.9499 n-butane nc4 15.9383 15.9383 15.9383 I-pentane ic5 3.9846 3.9846 3.9846 n-pentane nc5 2.9884 2.9884 2.9884 cyclopentane C5H10 0.2391 0.2391 0.2391 n-hexane nc6 2.7892 2.7892 5.6879 cyclohexane C6H12 0.1992 0.1992 benzene C6H6 0.1692 0.1692 n-heptane nc7 0.8965 0.8965 toluene C7H8 0.0598 0.0598 n-octane nc8 0.7471 0.7471 p-xylene C8H10 0.0100 0.0100 n-nonane nc9 0.3287 0.8169 n-decane nc10 0.2889 n-undecane nc11 0.1096 n-dodecane cc12 0.0598 n-tridecane nc13 0.0199 n-tetradecane nc14 0.0100 9961.4162 9961.4162 9961.4162 HC dew point, o F at 400 psig 79 33-23