Automatic Descending Aorta Segmentation in Whole-Body PET-CT Studies for PERCIST-based Thresholding

Similar documents
Quantification in emission tomography: challenges, solutions, performance and impact

The time has come. Philips GEMINI TF PET/CT with TruFlight technology


Optimisation of Clinical Protocols

Biograph TruePoint PET CT. The World s First HD PET Platform.

Feasibility of state of the art PET/CT systems performance harmonisation

CQIE PET PROCEDURES. American College of Radiology Clinical Research Center. Centers for Quantitative Imaging Excellence LEARNING MODULE

New PET/CT from Siemens helps more patients benefit from premium technologies

AAPM Scientific Meeting Imaging Symposium. State of the Art in Quantitative Imaging CT, PET and MRI. Which Imaging Modality is the Most Quantitative

Biophysical Considerations in the Precision of Quantitative 18 F-FDG PET/CT DISSERTATION

The Unique, New MRI Philips Ingenia 3 Tesla is now in Ayios Therissos! The first-ever digital broadband MR system has been installed in Ayios

GE Healthcare. Introducing Discovery MI DISCOVERY MI

RADIATION ONCOLOGY RESIDENCY PROGRAM Competency Evaluation of Resident

Mirada Case Study. VentureFest, July

GE Healthcare. PET/CT + MR Trimodality Imaging

Propagating Uncertainty in Multi-Stage Bayesian Convolutional Neural Networks with Application to Pulmonary Nodule Detection

RSNA RCA22 Quantitative Imaging: Structured Reporting in 3D Slicer and beyond

Tissue-Based Affine Registration of Brain Images to form a Vascular Density Atlas

PET in clinical trials

Precision in Quantitative Imaging: Trial Development and Quality Assurance

Clinical Applications. ImagingRite. Interventional Radiology

Clarity CT Technology

Multicenter oncology clinical trials are increasingly using

Role of PET/CT Imaging

Disclosures 7/31/2017. Clinical Impact and Applications of 4D Imaging (in RT)

CT procedure with needle guidance for heptatocellular carcinoma

The Efficiency of Respiratory-gated 18 F-FDG PET/CT in Lung Adenocarcinoma: Amplitude-gating Versus Phase-gating Methods

Progress Towards an International Image Quality Monitoring Framework for Quantitative Imaging

SPECIFICATION CT SIMULATOR FOR BLACK LION HOSPITAL, ADDIS ABEBA, ETHIOPIA

The quenching effect in PRESAGE dosimetry of proton beams: Is an empirical correction feasible?

Medical instrumentationi 11/19/2010

Optimizing Adaptive Workflows Using RapidPlan at the Beatson West of Scotland Cancer Centre

BodyFIX Maximize positional accuracy. Support delivery with precise patient positioning and immobilization

R/F. Experiences Using SONIALVISION safire and the Utility of Tomosynthesis. 1. Introduction. 2. Basics of Tomosynthesis.

Connect and Deliver Optima CT580 RT

Target point correction optimized based on the dose distribution of each fraction in daily IGRT

Handzettel 1. CARE Right Computed Tomography, committed to the right dose Ivo Driesser Austin, July 22 nd 2014

Advanced molecular imaging. Proven accuracy inspires confidence. Philips Vereos Digital PET/CT

Positron Emission Tomography Present status and future prospects

Jonathon A. Nye, Ph.D Emory University School of Medicine Atlanta, GA

Jeffrey T. Yap, Ph.D. Dana-Farber Cancer Institute Brigham & Women s Hospital Harvard Medical School

Respiratory-phase phase correlated imaging, treatment planning and delivery Clinical Implementation. Disclosures. Session Objectives.

Abdominal CT with Single-Energy Metal Artifact Reduction (SEMAR): Initial Experiences

Molecular Imaging. Professor Sir Michael Brady FRS FREng Department of Engineering Science Oxford University

Philips PET/CT Gemini GXL. Total performance. Total confidence.

03/08/2017. Advances and Challenges in Contour QA for Adaptive RT. Objective. Disclosures

NIST Medical Imaging Informatics Activities. Medical Imaging

Introducing a new take on efficient workflow: Deeply integrated clinical applications

The EARL FDG-PET/CT Accreditation Programme & Guideline Developments: Results of more than 65 Successfully Accredited Sites and Future Perspectives

One exam. One workflow. Immeasurable versatility.

Quantitative Techniques in PET-CT Imaging

APPROACH METHODS FOR BIOMEDICAL IMAGE PROCESSING USING DOSE BK Y90

An automatic delineation method for bone marrow absorbed dose estimation in

Poster No.: C-1848 Congress: ECR Educational Exhibit. Authors:

Registration of 3D Ultrasound to Computed Tomography Images of the Kidney

Quantitative biomarker for clinical decision making

THE NEW 640 SLICE CT SCANNER

In-vivo Targeting of Liver Lesions with a Navigation System based on Fiducial Needles

Imaging. Steve Rusckowski CEO Philips Healthcare

Emerging Applications and Trends Across Medical Imaging

Computed tomography. High. performance. no trade-offs. Philips Ingenuity CT family

COMMITTEE FOR MEDICINAL PRODUCTS FOR HUMAN USE (CHMP)

Diagnostic Medical Image Processing

CT to Ultrasound Registration: A Porcine Phantom Study. J. Xiang, S. Gill, C. Nguan, P. Abolmaesumi and R. N. Rohling University of British Columbia

Toward Human-oriented Life

80 Years of Innovation

Imaging/Imagine Needs for Proton Therapy: Treatment Planning. Lei Dong, Ph.D. Scripps Proton Therapy Center San Diego, CA

Occams Business Research & Consultancy

At that time, do not reference or utilize the initial plan in any manner or study data may be compromised

Trusted Performance. Smart Investment. 80 detector row Ultra Helical CT

Accuracy Evaluation of a 3D Ultrasound-guided Biopsy System

Features, Functions, and Specifications of MultiImaging System. Page 1

Automated Breast Volumes. Simplified.

CHAPTER-6 HISTOGRAM AND MORPHOLOGY BASED PAP SMEAR IMAGE SEGMENTATION

Recent clinical advances and applications for medical image segmentation

CHAPTER 17 Brain Image Analysis and Atlas Construction

Ingenia MR-RT. MR Systems. The comprehensive MR-sim solution to fit your planning

Computer-aided diagnostic scheme for the detection of lung nodules on chest radiographs: Localized search method based on anatomical classification

LIVER EXTRACTION USING HISTOGRAM AND MORPHOLOGY

Translational & Molecular Imaging Institute

Computed Tomography Based Attenuation Correction in PET/CT: Principles, Instrumentation, Protocols, Artifacts and Future Trends

Image guided Radiation Therapy (IGRT) a perspective

Comparison of quantitative methods for the assessment of interstitial lung disease in MDCT

Fast enough to stop the Capable of delineating Unprecedented imaging power for the. Virtual endoscopy. The gatewa

Imaging Systems. Steve Rusckowski CEO Imaging Systems

arxiv: v1 [cs.cv] 8 Jan 2018

High performance comes easily

Quantitative PET/CT imaging and dosimetry of 89Zr labelled compounds Makris, Nikolaos E.

Determination of Dose to Individual Patients in Radionuclide Imaging Procedures including Planar, SPECT and PET. Estimation of dose and not dosimetry

Section: Magnetic Resonance Imaging

Imaging anytime, anywhere.

A Methodology for Optimizing the Acquisition Time of a Clinical PET Scan Using GATE

Customer Experience Insight

Low Dose, Lightning Speed, Latest Applications

The FUTURE has ARRIVED NeuViz 128

Making the difference with Live Image Guidance

Compartmental model of 18 F-choline

Transcription:

Automatic Descending Aorta Segmentation in Whole-Body PET-CT Studies for PERCIST-based Thresholding Lei Bi, Jinman Kim, Member, IEEE, Lingfeng Wen, Member, IEEE and David Dagan Feng, Fellow, IEEE Abstract Medical imaging is a fundamental component of modern healthcare where majority of medical conditions can benefit from some kinds of imaging. A dual-modal positron emission tomography computed tomography (PET-CT) has increasingly become the preferred imaging method to stage the common cancers and to assess treatment response. For example, PET image can quantitatively assess the treatment before morphological changes that can be detected on anatomical CT image. PET response criterion (PERCIST) is a widely recognised thresholding method for detecting malignant lesions (high metabolic value) in treatment response. It is based on the calculation of standardised uptake value with lean body mass (SUV LBM ), together with a volume of interest (VOI) reference placed on the right lobe of the liver or the descending aorta (when the liver is abnormal e.g., liver cancer). These two structures are considered to have stable metabolism among the PET patient population and therefore can be used to normalise the SUV. The current practice of PERCIST thresholding depends on the manual delineation of the VOI reference which is a time consuming and operator-dependent process. Furthermore, such VOI selection is more difficult on the smaller descending aorta structure when compared to the liver. In this study, we propose a fully automatic approach to the segmentation of the descending aorta for use in the calculation of the PERCIST thresholding. A multi-atlas registration coupled with weighted decision function was used in the whole-body CT for the segmentation of the descending aorta, with the resulting VOI reference mapped to the co-aligned PET counterpart. We evaluated our method with 30 clinical PET-CT studies with preliminary results demonstrating reliability and robustness. Image Segmentation; Multi-atlas Image Registration; Multi-modal PET-CT; Aorta Segmetaton; PERCIST I. INTRODUCTION Medical imaging is becoming ever more important in modern medicine due to its ability to non-invasively visualise Lei Bi, Jinman Kim, Lingfeng Wen, and David Feng are with the School of Information Technologies, University of Sydney, Australia. Lei Bi (e-mail: lei@it.usyd.edu.au); Jinman Kim (phone: +61 2 9036 9708; fax: +61 2 9351 3838; e-mail: jinman.kim@sydney.edu.au). Lingfeng Wen is also with Department of PET and Nuclear Medicine, Royal Prince Alfred Hospital, Australia (e-mail: wenlf@ieee.org). David Dagan Feng is also with CMSP, Department of EIE, Hong Kong Polytechnic University, Hong Kong, and Med-X Research Institute, Shanghai Jiao Tong University, China (e-mail: david.feng@sydney.edu.au). This research was funded in part by ARC grants. internal anatomical structure or functional information for diagnosis, treatment planning and assessment [1]. Computerised imaging techniques have been applied in medical imaging to provide digitalised images for visualisation [2]. Functional imaging modality of Positron Emission Tomography (PET) is capable of visualising early functional change; for example, varied metabolic activity of malignant lesions, prior to significant morphological abnormalities that can be seen by using anatomical imaging such as computed tomography (CT) [3]. Multi-modal PET-CT scanner, which combines PET and CT in one imaging device, provides detailed anatomical information of CT data to supplement functional information of PET data in a co-aligned imaging space. PET image can derive quantitative functional parameters for assessing metabolism of abnormal structures such as tumours. Standardised uptake value (SUV) which is to normalise PET images by injected tracer dose and patient body weight, has been widely accepted in oncology to provide approximate estimate of metabolic rate of glucose consumption in identifying malignant lesions by using the tracer of FDG. The use of lean body mass (LBM) has been proposed to address the inaccurate estimation of body mass with the use of body weight in deriving SUV [4]. PET Response Criterion (PERCIST) has been well accepted in deriving metabolic changes of malignant lesions in assessing treatment response. It suggests that the use of a volume of interest (VOI) reference placed on the right lobe of liver, or the descending aorta in the case of abnormality in the liver. These two structures were considered to have stable SUV among the population of PET studies. The average pixel values of the VOI reference from PET image was then used in thresholding PET data to separate malignant lesions from nonlesions. It is of great benefits to automate the PERCIST-based thresholding for objective evaluation of treatment response and reduce operator-dependent bias. There are extensive literatures which present the automatic segmentation of anatomical structures needed in PERCIST such as the liver [5], and also the aorta [6]. Our recent work [7] has demonstrated a framework of automated liver segmentation specifically for optimised PERCIST with the use of multi-atlas registration of liver. However, the presented framework cannot be applied for the segmentation of the descending aorta when liver is abnormal. Furthermore, the descending aorta is a small structure which introduces greater challenges in the VOI placement when compared to the liver. In this paper, we propose a descending aorta segmentation algorithm for its application to the PERCIST-based thresholding framework. A

fully automatic approach to the segmentation of the descending aorta with the use of multi-atlas registration coupled with weighted decision function was applied. The resultant segmentation was then used as a VOI reference for PERCIST thresholding of PET. II. METHODS A. Overview of the proposed descending aorta segmentation Fig.1 illustrates the overview of our proposed segmentation process of the descending aorta. Initially CT images were pre-processed to remove the bed/linen artifacts automatically by adaptive thresholding and image subtraction from a bed template [8]. Empty spaces were then removed to reduce the computation. After that, the range of descending aorta was estimated by identifying the lung structures in the CT image. Then multi-atlas registration followed by the weighted decision function [6] was applied to the detected range to segment the descending aorta. In addition, the segmented result (label) of the descending aorta was transferred to the co-registered PET images and a VOI reference with the 1cm diameter and 2cm height was placed within the label volume. Finally, by using the VOI reference, a PERCIST-based threshold was calculated and used to separate the structures of high metabolic value in the SUVLBM normalised map of the PET counterpart. Figure1. Overview of our PERCIST-based thresholding of the PET image based on the automated segmentation of the descending aorta from the whole-body PET-CT studies. B. Patient Data Whole body PET-CT studies were acquired from the department of PET and Nuclear Medicine, Royal Prince Alfred hospital with a Siemens Biograph TruePoint PET-CT scanner or a mct PET-CT scanner. All the studies have either 326 slices or 368 slices at a slice thickness of 3 mm to cover the body from the top of the head to the upper thigh of the human body. PET slice had a 200 200 matrix at pixel size of 4.07 mm 2 and CT slice had a 512 512 matrix at the pixel size of 0.98 mm 2. In total, 30 randomly selected patient studies (anonymised) were used in the evaluation, consisting of 5 studies with abnormal activities in the liver and further 25 nonsmall lung cancer studies. No abnormalities were observed in the descending aorta for all the studies. For the construction of the training data, additional 10 patients with no abnormalities in the lung and the aorta regions were used. C. Training Set Construction In our training set, labels of the descending aorta were semi-automatically segmented from 10 CT studies by an experienced operator using geodesic contour modelling in the medical imaging interaction toolkit [9]. Segmentation was done in transaxial slices from the bottom of the aortic arch to the bottom of the lungs. Under segmented descending aorta was preferred when the boundary was unclear from the low-contrast CT images. D. Descending Aorta Segmentation 1) Descending Aorta Localisation Lung structures were segmented from whole-body CT image initially for estimating the spatial range of descending aorta. These lung structures cover the thorax region of the body where the descending aorta structure is localised within the lung pairs. In the segmentation of the lungs, an adaptive thresholding was applied on the CT data to separate air from

other tissue structures. Left and right lungs were then separated by identifying the two largest connected regions [10]. Cropped thorax image covers the range of the descending aorta. 2) Multi-Atlas based Segmentation with Weighted Decision Function The multi-atlas based descending aorta segmentation with decision function was adopted from [6]. In our study, we have N descending aorta atlases A 1, A 2,, A N, and its corresponding manually delineated labels S 1, S 2,, S N. Let cropped thorax image U be the target image and S be the label from the segmentation results. All the atlases were registered to U using affine translation followed by B-Spline registration from the Elastix toolkit [11], resulting in transformation parameters u 1, u 2,, u N. These parameters were then used to map the labels S 1, S 2,, S N, to U. After the transformations, difference D i between U and each of the transformed atlases were calculated and used as weight functions λ for each pixel p according to: 1 i ( p). (1) D ( p) g ( p) i Here, g is a Gaussian filter with σ equal to 4 and ϵ is a constant set to 0.001 in order to avoid division by 0. The probabilistic label S p was generated by computing the average of the weighting functions where: N 1 (2) S p ( p) i( ) i( i( )). N p S u p i 1 ( p) i i 1 The resulting probabilistic label represents the weighted decision function, which was then subject to a Gaussian filter with σ equals to 4. The filtered result was then thresholded at 50% of its pixel values. The largest connected component (pixels) was extracted as the final segmented label. In our experiments N was set to 10. E. VOI Reference Segmentation From the segmented label of the descending aorta in Section D, the VOI reference was placed inside this label by centring the VOI at a slice located at 20% from the top of the label in the transaxial view. For each of the slices in the VOI reference, a circle was placed using the centroid of the pixels in the label, so as to the VOI reference is equally distant from the boundary of the labels. The VOI reference which is defined as a 1cm diameter over 2cm height is only a fraction in volume of the entire label, thus ensuring the complete placement of the VOI reference even with errors in the descending aorta segmentation label. F. PERCIST-based Thresholding The value of LBM, according to [12], was calculated using the body parameters of weight (kg), height (m), age (year old) and gender (1 for male, 0 for female): 2 LBM ( kg) weight { weight [(1.2 weight / height ) (3) 0.23 age 10.8 Gender 5.4] / 100} The derived LBM was then applied to convert PET image to SUVLBM map according to [13] where: CPET ( t) SUVLBM. (4) Injected dose / LBM Here, C PET(t) is the radioactivity concentration at time t, expressed in kbq/ml and injected dose in MBq. The VOI reference from the PET was then used to derive the value of a threshold according to [14]: Th SUV (2 mean) (2 SD) (5) LBM where mean and standard deviation SD were calculated from the pixels within the segmented VOI reference.

Figure 2. Descending aorta and its segmented VOI reference on 4 selected patient studies (columns). The blue boundary outlines the descending aorta segmentation and the red boundary represents the segmented VOI reference. The descending rows are transversal, coronal and saggital views respectively. III. RESULTS AND DISCUSSIONS A. Descending Aorta Segmentation Fig. 2 shows the results of the descending aorta segmentations together with the VOI references in selected challenging cases among the 30 PET-CT studies. In all the cases, descending aortas were correctly identified and segmented according to our visual inspection. However, as expected in a completely automated approach, some cases resulted in slight over-segmentation. Fig 2(b) is a case when a small portion of the segmentation label is outside the descending aorta and occupies part of the mediastinum. Another over-segmentation result is shown in Fig 2(c), where the abnormality seen in the left lung s diaphragm resulted in registration errors which translated to the descending aorta segmentation to include part of the ribs, as depicted in the coronal view. Nevertheless, even with such over segmentations, the placement of the VOI references were always completely within the descending aorta. Further, in all the investigated cases, segmentation errors where always from over-segmentation and not under, which are an important attribute in the VOI reference segmentation. This is attributed to the fact that our weighted decision function was set to 50% which gives preference to over estimation of the atlas registration. Fig. 3 illustrates the importance of the weighted decision function in the final segmentation results of the descending aorta. On the right is the sum of the labels of the registration results prior to the weighting function. Blue line indicates the 50% common region among the 10 atlas labels. The middle result shows the Gaussian smoothed probabilistic label calculated from the weighted decision function where the red line indicates the 50% common region. From the left image, we can clearly see that the weighted decision function corrects the under segmentation results from the multi-atlas result, especially in the top-right and bottomright corners. Such correction ensures that under segmentation is always avoided. As a further test to the accuracy of the VOI reference estimation, we measured the homogeneity of the voxels residing in the VOI reference in PET. It is expected that if the VOI reference is entirely within the descending aorta, it will consists of homogeneous pixel values and thus have low SUV LBM standard deviation from the mean. For each study, we calculated the standard deviation and plotted the results in Fig. 4. Among our 30 studies, the average standard deviations SUV LBM was within 13.3 ± 5.2% to its mean, which was within the expected variation of SUV LBM found in the descending aorta structure from our experiments. Figure 4. Average (blue labels) and standard deviation (line bars) of the pixels (in SUV) measured within the VOI references in our 30 clinical PET studies. B. Thresholding of PET Image The segmented VOI references were used in the PERCISTbased thresholding to calculate the SUVLBM normalised PET. Fig. 5 shows an example of an aligned PET-CT study with mapped descending aorta (blue line) and its corresponding VOI reference (red line). Visually in both the PET and CT, the VOI reference resides completely within the descending aorta. Figure 3. An example of the stages in descending aorta segmentation. (a) shows the CT image of the descending aorta with the segmentation labels from the weighted decision function from (b) and from only the multi-atlas registration (c).

Figure 5. A PET-CT study demonstrating the segmentation of the descending aorta and its use in the placement of the VOI reference for subsequent PERCISTbased thresholding. Here in both PET and CT, we can clearly depict the VOI reference (red outline) from the descending aorta segmentation (blue outline) in orthogonal views of coronal (a), saggital (b) and axial (c), respectively from left to right. Fig. 6 illustrates an example of using the VOI reference in the calculation of SUVLBM thresholding for two PET-CT studies with liver tumour(s). The results of the SUVLBM thresholding were able to depict all the abnormalities for all studies, which were consistent to the findings in the physician s reports. C. Computational Cost The whole-body PET-CT segmentation process took in average of 36±3 minutes running on a consumer PC equipped with Core i7 3.40 GHz CPU and 16.0GB of memory. The biggest time consuming task was the multiatlas registration which took an average of 30minutes (approx. 83% of the total time). We used 10 atlases in our multi-atlas registration thus resulting in this high computation time. IV. CONSLUSION AND FUTURE WORK In this study, we developed a fully automatic multi-atlas segmentation of the descending aorta from low-contrast CT in PET-CT studies. The segment results were then used in the calculation of the PERCIST-based thresholding in the normalization of PET counterpart. In our 30 clinical PET- CT studies, we correctly place the VOI reference within the descending aorta, demonstrating reliability and robustness of our algorithm. We will also further evaluate the performance of our method using greater number of clinical studies, and its integration to PERCIST-based thresholding using the liver VOI reference. Our current solution targeted the robustness and reliability of our segmentation. This resulted in high computation cost which we will attempt to reduce in our future study by code optimisations. ACKNOWLEDGMENT We like to thank our collaborators at the Royal Prince Alfred (RPA) hospital. This research was funded in part by ARC grants. Figure 6. Two random patient studies (rows) with tumour(s) in the liver. The PERCIST-based thresholding based on our descending aorta segmentation reveals the high metabolic regions in various places in the whole-body PET images. Left is the coronal (a) and right is the saggital (b) views. REFERENCES [1] R. Bar-Shalom, et al., "Clinical performance of PET/CT in evaluation of cancer: additional value for diagnostic imaging and patient management," J. Nucl. Med., vol. 44, pp. 1200-09, 2003.

[2] Doi. K, et al., "Computer-aided diagnosis in medical imaging: historical review, current status and future potential," Comput. Med. Imag. Grap., vol. 31, (4-5), pp. 198-211, 2007. [3] N.G. Mikhaeel, et al., "Use of FDG-PET to monitor response to chemotherapy and radiotherapy in patients with lymphomas," Eur. J. Nucl. Med., vol. 33, pp. 22-26, 2006. [4] R.L. Wahl, et al., "From RECIST to PERCIST: evolving considerations for PET response criteria in solid tumors," J. Nucl. Med., vol. 50, pp. 122S-50S, 2009. [5] E. van Rikxoort, et al., "Automatic segmentation of the liver in computed tomography scans with voxel classification and atlas matching," MICCAI segmentation workshop, pp. 101-8, 2007. [6] I. Isgum, et al., "Multi-atlas-based segmentation with local decision fusion Application to cardiac and aortic segmentation in CT scans," IEEE T. Med. Imag., vol. 28(7), pp. 1000-10, 2009. [7] L. Bi, et al., "Automated and Robust PERCIST-based Thresholding framework for whole body PET-CT studies," IEEE Eng Med Bio, (in press), 2012. [8] J. Kim, et al., "A fully automatic bed/linen segmentation for fused PET/CT MIP rendering," J. Nucl. Med., 49(Suppl 1), 387P, 2008. [9] D. Maleike, et al., "Interactive segmentation framework of the medical imaging interaction toolkit," Comput. Meth. Prog. Bio., vol. 96, (1), pp. 72-83, 2009. [10] S. Hu, et al., "Automatic lung segmentation for accurate quantitation of volumetric X-ray CT images," IEEE T. Med. Imag., vol. 20, pp. 490-498, 2001. [11] S. Klein and M. Staring, Elastix. http://www.isi.uu.nl/elastix/. [12] R. Pieterman, et al., "Visualisation and assessment of the protein synthesis rate of lung cancer using carbon-11 tyrosine and positron emission tomography," Eur. J. Nucl. Med., vol. 29, pp. 243-7, 2002. [13] C.K. Kim., et al., "Standardized uptake values of FDG: body surface area correction is preferable to body weight correction, " J. Nucl. Med., vol. 35, (1), pp. 164, 1994. [14] P. Ghosh et al., "Expanding the Power of PET with PERCIST," Siemens Medical Solutions USA, Inc., 2010.