Preparative Purification of Corticosteroids by HPLC; Scalability and Loadability Using Agilent Prep C18 HPLC Columns Application

Similar documents
Fast Analysis of Environmental Phenols with Agilent Poroshell 120 EC-C18 Columns

Reversed-phase Separation of Intact Monoclonal Antibodies Using Agilent ZORBAX Rapid Resolution High Definition 300SB-C8 1.

Scale-up with the Agilent SD-1 Purification System analytical and preparative runs on a single system

Application Note. Authors. Abstract. Introduction. Pharmaceutical

Simple Techniques for Improving the Isolation of Synthetic Peptides Jo-Ann Jablonski Principal Scientist Waters Corporation

Method Translation in Liquid Chromatography

Faster Separations Using Agilent Weak Cation Exchange Columns

Agilent 218 Purification System. Purify your way

Fast and High-Resolution Reversed-Phase Separation of Synthetic Oligonucleotides

Practical Aspects of Preparative HPLC in Pharmaceutical Development and Production

Rapid UHPLC Analysis of Reduced Monoclonal Antibodies using an Agilent ZORBAX Rapid Resolution High Definition (RRHD) 300SB-C8 Column

A method to pack Agilent Prep C18 Media on Load & Lock Column

Agilent SD-1 Purification System. Purify your way SD-1

Separation of Recombinant Human Erythropoietin (repo) Using Agilent Bio SEC-3

Performance Characteristics of the Agilent 1220 Infinity Gradient LC system

Encompassing Method Development, Small Scale Purification, and Post-Purification Analysis with the Investigator SFC System

10. Validated Normal Phase HPLC Method for the Determination. Fulvestrant is primarily used in the treatment of hormone receptor

Overview of preparative HPLC. Analytical Technologies Limited

Bivalirudin Purification:

Application Note. Biopharma. Authors. Abstract. James Martosella, Phu Duong Agilent Technologies, Inc Centreville Rd Wilmington, DE 19808

High-Throughput LC/MS Purification of Pharmaceutical Impurities

What differentiates Dynamic Extractions?

Emulation of the Agilent 1100 Series LC Through Waters Empower Software Analysis of an Analgesic Mixture

Performance characteristics of the 1260 Infinity Quaternary LC system

Fast Preparative Column Liquid Chromatography (PCLC)

Agilent Prep LC Columns for Small Molecules and Biomolecules MAINTAIN RAPID, RELIABLE SEPARATIONS AS YOU SCALE-UP

Agilent AdvanceBio SEC Columns for Aggregate Analysis: Instrument Compatibility

Automated fraction re-analysis does it really make sense? Application. Udo Huber. Abstract

Benefits of 2D-LC/MS/MS in Pharmaceutical Bioanalytics

Chem 321 Lecture 23 - Liquid Chromatography 11/19/13

Effective Scaling of Preparative Separations Using Axial Compression Columns

Fast, Low Pressure Analysis of Food and Beverage Additives Using a Superficially Porous Agilent Poroshell 120 EC-C18 Column

Prefractionator for Reliable Analysis of the Light Ends of Crude Oil and other Petroleum Fractions

Techniques for Improving the Efficiency of Large Volume Loading in Preparative Liquid Chromatography

Method Development Considerations for Reversed-Phase Protein Separations

Quality-by-Design-Based Method Development Using an Agilent 1290 Infinity II LC

Optimizing Protein Separations with Agilent Weak Cation-Exchange Columns

Fraction Analysis of Cysteine Linked Antibody-Drug Conjugates Using Hydrophobic Interaction. chromatography. Agilent 1260 Infinity II Bio-Inert System

Developing Quantitative UPLC Assays with UV

Improved Gas Chromatograph Method for the Analysis of Trace Hydrocarbon Impurities in 1, 3-Butadiene

EPA 8330A Analysis of Explosives Using Agilent Poroshell 120 EC-CN and EC-C18 Columns

Biotherapeutic Method Development Guide

Separate and Quantify Rituximab Aggregates and Fragments with High-Resolution SEC

Improving Resolution and Column Loading Systematically in Preparative Liquid Chromatography for Isolating a Minor Component from Peppermint Extract

A Modular Preparative HPLC System for the Isolation of Puerarin from Kudzu Root Extracts

Fundamentals and Techniques of Preparative HPLC. Parto Zist Behboud Tel: 42108

Technical Overview. Author. Abstract. Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany

Zwitterion Chromatography ZIC

Biotherapeutic Method Development Guide

Sean M. McCarthy and Martin Gilar Waters Corporation, Milford, MA, U.S. INTRODUCTION EXPERIMENTAL RESULTS AND DISCUSSION

A Simple Rapid and Sensitive Method Development for Quantification of Quetiapine Fumarate in Bulk and Dosage Forms Using RP-HPLC

Characterize Fab and Fc Fragments by Cation-Exchange Chromatography

EVOLUTE ABN SPE Columns for Solid Phase Extraction of Environmental Samples

and for the Development of Isolation Strategies

Biomolecule Purification. Purification columns and media for peptides, oligonucleotides, and proteins

Analysis of Atrazine in Drinking Water at the ppb Level Using New Agilent Reversed Phase LC Columns. Application. Author. Abstract.

Application Note. Author. Abstract. Pharmaceuticals. Detlef Wilhelm ANATOX GmbH & Co. KG. Fuerstenwalde, Germany mau

Developing Purification Strategies for the Agilent 1260 Infinity II Preparative LC/MSD System

Improving Retention Time Precision and Chromatography of Early Eluting Peptides with Acetonitrile/Water Blends as Solvent B

A Primer. Principles in preparative HPLC

Technical Overview. Author. Abstract. A.G.Huesgen Agilent Technologies, Inc. Waldbronn, Germany

Preparative HPLC: Factors and Parameters that Directly Affect Recovery of Collected Fractions

A Rapid Method for Trace Analysis of Organophosphorus Pesticides in Drinking Water

The Use Of Automated Backflush on the 7890A/5975A GC-MS System. Application. Setup Example Using Biomarkers. Authors. Abstract.

ITSP Extraction of Cortisol from Plasma

Peptide Mapping: A Quality by Design (QbD) Approach

Optimizing the Purification of a Standard Chiral Compound Utilizing a Benchtop, Multi-Purpose, Semi-Preparative to Preparative HPLC System

Introduction Ron Majors is a Senior Scientist at Agilent. Bill Champion is a chemist in Agilent s HPLC Columns tech support group.

Fast Method Development Using the Agilent 1290 Infinity Quaternary LC System with Column Selection Valve

Application Note. Author. Abstract. Pharmaceuticals. A.G.Huesgen Agilent Technologies, Inc. Waldbronn, Germany

High-resolution Analysis of Charge Heterogeneity in Monoclonal Antibodies Using ph-gradient Cation Exchange Chromatography

Analysis of amoxicillin and five impurities on the Agilent 1220 Infinity LC System

Maximizing Sample Recovery on the ACCQPrep HP125

Flexibility with. Preparative HPLC

Monitoring for Microcystins in Raw Water Supply Reservoirs Using the Agilent 6410 Triple Quadrupole LC/MS

Peptide Mapping of Glycoprotein Erythropoietin by HILIC LC/MS and RP-LC/MS

A Comprehensive Workflow to Optimize and Execute Protein Aggregate Studies

Monitoring for Microcystins in Raw Water Supply Reservoirs Using the Agilent 6410 Triple Quadrupole LC/MS

Agilent MetaCarb 87H Organic Acids Column H + Form User Manual Part No. A5214

Agilent 1290 Infinity Quaternary LC Support of Columns with 2.1 to 4.6 mm ID to 1200 bar

Determination of Pesticides in Water by SPE and LC/MS/MS in Both Positive and Negative Ion Modes. Application Note. Author. Abstract.

DIDANOSINE ORAL POWDER Final text for addition to The International Pharmacopoeia

Fast and High Resolution Analysis of Intact and Reduced Therapeutic Monoclonal Antibodies (mabs)

A Practical User Guide for the Determination of Optimal Purification Gradients for the Gilson PLC 2020

ENVIRONMENTAL ANALYSIS. Solutions for Your Analytical Business Markets and Applications Programs. Solution Note

Determination of Asarinin in Xixin (Asari Radix Et Rhizoma)

Clone Selection Using the Agilent 1290 Infinity Online 2D-LC/MS Solution

Optimizing Purified Sample Recovery on Reverse-Phase HPLC Systems

for Acclaim Carbamate Column

mab Titer Analysis with the Agilent Bio-Monolith Protein A Column

Mobile Phase Optimization in SEC Method Development

Analysis and Purification of Polypeptides by Reversed-Phase HPLC

ADVANCE ACCURACY AND PRODUCTIVITY FOR FASTER ANALYSIS

Automated QbD-Based Method Development and Validation of Oxidative Degraded Atorvastatin

AdvancedTools in HPLC methoddevelopment

High-throughput and Sensitive Size Exclusion Chromatography (SEC) of Biologics Using Agilent AdvanceBio SEC Columns

Genotoxicity is the property of a compound

Agilent 1260 Infinity Purification Systems. Infinitely better purity and recovery

Analysis of Intact and C-terminal Digested IgG1 on an Agilent Bio MAb 5 µm Column

Solubility Studies Utilizing a Phase Monitor

Transcription:

Preparative Purification of Corticosteroids by PLC; Scalability and Loadability Using Agilent Prep C18 PLC Columns Application Pharmaceuticals Authors Cliff Woodward and Ronald Majors Agilent Technologies, Inc. 2850 Centerville Road Wilmington, DE 19808-1610 USA Abstract Agilent Prep reversed-phase columns with different diameters displayed the nearly linear scaleup for the gradient separation of corticosteroids. The loadability of the columns was quite good with as much as 8 mg of total corticosteroid injected onto a 4.6-mm column and 320 mg onto a 30-mm column. Introduction Corticosteroids are a family of drugs that include cortisol (hydrocortisone) and corticosterone, adrenal hormones found naturally in the body, as well as synthetic drugs such as triamcinolone and prednisone. Though natural and synthetic corticosteroids are both potent anti-inflammatory compounds, the synthetics exert a stronger effect. The oral forms of corticosteroids are used to treat numerous autoimmune and inflammatory conditions including bursitis, tendonitis, and Crohn s disease. Corticosteroids are available for inhalation by mouth to treat asthma conditions of restricted breathing as well as symptoms of nasal allergies. Topical forms are available to treat skin conditions such as eczema, psoriasis, and insect bites. Since corticosteroids are used in so many ways, their analysis in a variety of matrices is important. PLC represents a convenient, rapid and sensitive technique for the analysis and quantitation of these important drugs. When investigating new synthetic or natural corticosteroids or attempting to identify trace impurities by large injections, preparative quantities are needed. So scaling up to larger columns is required. The purpose of this application note is to demonstrate the ability to perform a direct and linear scaleup from analytical to larger scale preparative PLC using a matched set of columns. Corticosteroids are large bulky organic molecules with very similar structures (Figure 1). Their solubilities in water vary from insoluble to very sparingly soluble. Their solubility in organic solvents such as methyl alcohol and acetonitrile is much better. Thus, these compounds are good candidates for and should be easily separated by reversedphase liquid chromatography (LC). Even though their structures are similar, as depicted in Figure 1, chromatographic retention amongst the corticosteroids studied is quite different and a gradient is required for their separation in a reasonable time. Gradient elution is often a preferred approach to speed up analytical and preparative separations and at the same time to provide an increase in sample capacity. Since our goal was to eventually isolate tens of milligram amounts of a single corticosteroid, we initially explored analytical PLC conditions where a minimum of solvent and sample would be required during method development.

C 3 C 3 C 3 C 3 ydrocortisone Corticosterone C 3 C 3 C 3 C 3 F Triamcinolone Prednisone Figure 1. Structures of corticosteroids. Chromatographic Separation of Corticosteroids on an Analytical Scale Goals in preparative chromatography include maximizing throughput, purity and yield. The scaleup process involves injecting increasingly larger amounts of material on column until the resolution becomes so poor that purity is compromised. There are two approaches to the scaleup process. The first is to scaleup on an analytical column by injecting progressively larger sample masses until resolution is barely acceptable (arbitrarily set at R s <1.2); then, based on the preparative mass requirement, move to a suitably-sized prep column; and adjust the chromatographic conditions dictated by the analytical separation but matched to the preparative column. The second approach is to choose the preparative column needed, then perform method development directly on this column until resolution and the resultant analyte purity is achieved. The first approach is generally preferred since the analytical method development requires less solvent, sample, and time. If the preparative column has the same chemistry as the analytical column, then the scaleup should be a linear process. Usually the same particle size is used in both the analytical separation and the preparative separation steps. In preparative chromatography, one may work under slightly overloaded conditions where resolution may be sacrificed for yield and the efficiency advantage of small particle sizes ( ~ 5 µm) may not be required. In addition, larger particle sizes ( ~ 10 µm) will provide sufficient resolution but generate less backpressure allowing flow rates to be increased to improve separation speed. n the other hand, sometimes chromatographic resolution for critical pairs is insufficient with a 10-µm particle and having the ability to use 5-µm prep columns is advantageous. In such cases, a compromise among resolution, required purity, and speed will be required for optimum preparative performance. Figure 2 shows the initial binary mobile phase gradient separation of the four corticosteroids on a column with reversed phase C18 packing. The column used was an Agilent Scalar C18, 4.6 mm 150 mm, a typical size for analytical work. During method development, initial experiments involving various water (solvent A ) and neat water-miscible organic solvent [that is, acetonitrile (ACN), 2

methanol (Me), or isopropanol (IPA)] (solvent B ) combinations were performed. owever, when using pure organic solvents in a binary pumping system, selectivity was not adequate so mixtures of organic solvents were investigated. The mobile phase found to yield the best overall resolution of the corticosteroids consisted of water as the weak component (solvent A ) and 50% ACN:17% Me:33% IPA as the strong component (solvent B ). The chromatographic conditions used in all the experiments described in this application note are shown in Table 1. Since the corticosteroids are neutral compounds, no buffer was required in the mobile phase system. In preparative work, since the goal is to collect pure fractions, it is preferable to avoid the use of buffers, especially nonvolatile ones, so that further work of buffer removal from the collected fraction is not required. If the LC conditions do require a buffer to provide better chromatography, then the use of a volatile buffer, such as ammonium acetate or formic acid, is recommended. Volatile buffers may be removed by rotary evaporation or by nitrogen purging. The solvent chosen for injection was dimethylsulfoxide (DMS), particularly useful for preparative chromatography [1]. The DMS is an excellent solvent for dissolution of many organic compounds, especially at higher sample concentrations often needed for preparative work. In addition, DMS elutes rapidly from a typical reversed-phase column (see Figure 2) and does not overlap or interfere with the collected fractions. Near the column void volume, a small DMS peak may appear depending on the wavelength used for effluent monitoring and on the flow cell path length. mau 225 200 Corticosterone 175 150 125 100 Prednisone ydrocortisone 75 Triamcinolone 50 25 DMS 0 0 2.5 5 7.5 10 12.5 15 17.5 20 22.5 Retention time, min Figure 2. Analytical separation of corticosteroids on Agilent Scalar-C18 Column. 3

Table 1. Chromatographic Conditions for Scalability and Loadability Experiments For 4.6-mm id Mobile phase A: Water Mobile phase B: 50% Acetonitrile: 17% Methanol: 33% Isopropanol (v/v) Gradient: Time 0, 25% B; 7 min, 25% B; 15 min, 34% B;15.1 min, 45% B; 20 min, 45% B; 20.1 min, 25% B; Stop Time: 24 min Flow rate: 1 ml/min, 4.6-mm id Injection solvent: DMS Injection volume: 4.6-mm id: 80 µl Injection loop: No external loop DAD settings: Detector wavelength: 244, 8; Reference 360, 100 Flowcell: 0.06-mm pathlength For 21.2-mm id Mobile phase A: Water Mobile phase B: 50% Acetonitrile: 17% Methanol:33% Isopropanol (v/v) Gradient: Time 0, 25% B; 7 min, 25% B; 15 min, 34% B;15.1 min, 45% B; 20 min, 45% B; 20.1 min, 25% B; Stop Time: 24 min Flow rate: 21.2 ml/min, 21.2-mm id Injection solvent: DMS Injection volume: 21.2-mm id, 1600 µl (700-µL loop and 900-µL needle fill Injection loop: 1500 µl DAD settings: Detector wavelength: 244, 8; Reference 360, 100 Flowcell: 0.06-mm pathlength For 30.0-mm id Mobile phase A: Water Mobile phase B: 50% Acetonitrile: 17% Methanol:33% Isopropanol (v/v) Gradient: Time 0, 25% B; 7 min, 25% B; 15 min, 34% B;15.1 min, 45% B; 20 min, 45% B; 20.1 min, 25% B; Stop Time: 24 min Flow rate: 42.4 ml/min, 30.0-mm id Injection solvent: DMS Injection volume: 30.0-mm id, 3200 µl Injection loop: 5000 µl DAD settings: Detector wavelength: 244, 8; Reference 360, 100 Flowcell: 0.06-mm pathlength Loadability of Corticosteroids on Agilent Preparative Reversed-Phase Columns Loadability for a preparative separation is defined as the maximum amount of analyte that can be injected onto a column that no longer permits the isolation of product at the desired level of purity or recovery levels; in other words, it is the maximum sample capacity of a particular column for the analyte(s) of interest. In the reversed-phase mode for a column of fixed dimensions, loadability is affected mostly by bonded phase coverage, chromatographic resolution, and chromatographic conditions such as mobile phase, injection solvent, temperature, etc. In preparative chromatography, the size of the column is dictated by the amount of purified material needed for further investigation. Generally, the larger the column, the greater 4

amount of sample can be injected onto the column. aving a wide range of column dimensions with the same chemistry allows the chromatographer greater flexibility in obtaining the desired amount of purified material. For a fixed chromatographic method with the same stationary phase and column length, the loadability (sample capacity) is directly related to the mass of packing in a column as depicted in Equation 1. 2 x 1 r 2 C L x 2 = (1) 2 r 1 where: x2 is the maximum amount (mass) of sample injected on column 2 (the larger id column) x1 is the maximum amount (mass) of sample injected onto column 1 (the smaller id column) r 1 and r 2 are the radii of columns 1 and 2, respectively C L is the ratio of column lengths of column 2: column 1 (in this case both lengths are the same) It can easily be deduced that the sample mass that can be injected is directly proportional to the radius ratios squared. For a linear scaleup, the chromatographic flow conditions should be adjusted to reflect this ratio and maintain an equivalent linear velocity. For example, for the scaleup of a separation done in a 4.6-mm id analytical column with a flow rate of 1.0 ml/min to an equivalent preparative column with a 21.2-mm id, the flow rate should be adjusted to (21.2/4.6) 2 1.0 ml/min or to 21.2 ml/min in order to have the same separation time. To investigate the loadability of Agilent Prep C18 columns of various dimensions for these corticosteroids, we obtained three 150-mm long columns of 4.6, 21.2, and 30-mm internal diameters (ids) (Agilent Technologies, Inc., Wilmington, DE, USA) and injected increasing volumes of a stock solution of the four test corticosteroids (50-mg/mL concentration of each). For one of the dimensions, the 30-mm id column, Figure 3 shows a series of total corticosteroid injected sample masses from 2.5 mg up to 320 mg. At 320 mg of total injected corticosteroid (80 mg each), column overloading began to appear and peaks became distorted, typical of overload conditions. ere overloading is defined, for a given column dimension, as the mass of sample injected where the resolution of critical compound pairs appears to deteriorate thus precluding collection of pure fractions. mau 2000 1500 1000 500 0 320 mg 160mg 80 mg 40 mg 20 mg 10 mg 5 mg 2.5 mg 0 5 10 15 20 25 30 35 40 min Figure 3. Corticosteroids on Agilent Prep-C18, 30.0 150 mm Loadability: 2.5 320 mg total. 5

To further illustrate this behavior, Figure 4 shows the resolution for peaks 2 and 3 (critical pair) as a function of loading on the three different sized columns. For each column, the injected mass was equivalent to the radius ratios squared so that an apples to apples comparison could be made. As can be seen in Figure 4, as the sample size increases, at low loading, resolution remains relatively constant. owever, as the injection mass increases, resolution decreases. Eventually around R s = 1.2, peaks become severely overlapped so that individual peak purity is compromised. Although there are some variations in the absolute resolution due to slight selectivity differences, it is apparent that the columns behave quite similarly and that column id is really a determinant of how much can be injected onto a preparative column prior to overload. When resolution becomes less than 1.2, overlap between the adjacent peaks occurs, precluding collection of a pure fraction. bviously, even with poor resolution, use of heartcutting techniques enables pure analyte to be obtained, but at the cost of yield. Scalability for a Matched Set of Columns The scalability of a chromatographic process is the reproducibility of results on columns of different ids when using the same particle size and bonded phase. In order to assess scalability for the Agilent Prep columns, we determined the maximum loadability of total analyte for each column dimension; and then compared all three column dimensions to see if the performance was similar. The effect of overloading each size column can clearly be seen in Figure 5 that shows the maximum amount of corticosteroids 2 and 3 that can be injected under the chromatographic conditions employed without significant peak overlap. At the point of overload, peak shapes also begin to deteriorate with visible tailing occurring. Nevertheless, at least 80 mg of each steroid (320 mg total) could be successfully injected onto the largest (30 mm) id prep column. This scalability experiment demonstrated the excellent sample capacity of the Agilent Prep phase as well as reproducibility for the corticosteroids at all column dimensions. 3 2.5 2 Resolution 1.5 1 4.6-mm 21.2-mm 0.5 30.0-mm 0 0 2 4 6 8 10 Relative Mass Injected, mg (4.6-mm data shown) Figure 4. Resolution vs. loading for corticosteroids on different id columns. 6

Triamcinolone Prednisone ydrocortisone Corticosterone 30.0-mm id - 320 mg on column 21.2-mm id - 160 mg on column 4.6-mm id - 8 mg on column Figure 5. Scalability of Agilent Prep-C18: overlay of chromatographic results from 4.6 mm id to 30.0 mm id. Conclusions The use of Agilent Prep reversed-phase columns with different diameters displayed the nearly linear scaleup for the gradient separation of corticosteroids. The loadability of the columns was quite good with as much as 8 mg of total corticosteroid injected onto a 4.6-mm column and 320 mg onto a 30-mm column. Reference 1. Angelika Kaufmann, Ulrike Jegle, Alexander Dreyer, and Ronald E. Majors, (2005) Using DMS as an Injection Solvent to Increase Sample Load in Preparative LC Applications, Agilent Technologies, publication 5989-2485EN www.agilent.com/chem For More Information For more information on our products and services, visit our Web site at www.agilent.com/chem. 7

www.agilent.com/chem Agilent shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material. Information, descriptions, and specifications in this publication are subject to change without notice. Agilent Technologies, Inc. 2005 Printed in the USA March 21, 2005 5989-2484EN